Electromagnetic Induction Imaging: Signal Detection Based on Tuned-Dressed Optical Magnetometry

Abstract

A recently introduced tuning-dressed scheme makes a Bell and Bloom magnetometer suited to detect weak variations of a radio-frequency (RF) magnetic field. We envisage the application of such innovative detection scheme as an alternative (or rather as a complement) to RF atomic magnetometers in electromagnetic-induction-imaging apparatuses.

Introduction

In 2001, Griffiths [1] proposed an imaging technique based on inferring one or more of the three passive electromagnetic properties (conductivity σ\sigma, permittivity ε\varepsilon and permeability μ\mu) to produce images on the basis of the response to a position dependent oscillating magnetic field. Several denominations are used to identify this kind of methodologies, among which electromagnetic induction imaging (EII) [2], electro-magnetic tomography (EMT) [3], magnetic induction tomography (MIT) [4, 5], and also mutual inductance tomography (same acronym) [6], the latter three stressing the potential of the technique to provide 3D mapping. A review on the subject was recently authored by Ma and Soleimani [7].

As a general feature, the technique uses an AC magnetic field to excite eddy currents in the specimen. The secondary magnetic field generated by those currents is then detected and analyzed.

Magnetometers can detect and record in-phase (ε\varepsilon dependent) and out-of-phase (σ\sigma dependent) response of specimens subjected to an excitation radio-frequency field BrfB_{\rm rf}. The response obtained while scanning the position of a sample with respect to the field generator and detector enables the registration of 2D maps. The possibility of varying the BrfB_{\rm rf} frequency (and correspondingly the skin depth in the specimen) enables 3D (tomographic) capabilities.

Optical magnetometers, in the so-called radio-frequency (RF) implementation [8, 9] are excellent detectors of weak (variation of) magnetic fields oscillating at a resonant frequency and hence constitute favourite detectors for EII, particularly in the case of low-conductivity and/or small size specimens.

This led recently to important steps toward high resolution EII of weakly conductive materials. Appositely developed RF magnetometers with opportune field-specimen arrangement, [10, 11], demonstrated a sensitivity sufficient to detect and characterize sub S m−1^{-1} conductivity material in small size (5 ml) specimens, despite the operation in unshielded environment) [2].

Such specification level makes the technique suited to develop tools for medical diagnostics [12], where the negligible invasivity of EII represents a valuable attractive, but the low-conductivity and the need of high-spatial resolution constitute a severe requirement.

This work considers the potential of a peculiar configuration of a Bell&Bloom (BB) magnetometer as a detector of eddy currents induced in small and/or weakly conductive specimens, which could be eventually used to produce maps in 2D or 3D scans.

While the basic principle of operation (detection of a secondary magnetic field) is shared with the above mentioned research, the proposed sensor works with a particular arrangement that does not require resonant conditions for the time-dependent excitation field. The proposed arrangement can constitute an alternative as well as a complement to the commonly used RF magnetometers.

The standard BB implementation of atomic magnetometers is based on optically pumping atoms into a given Zeeman sublevel (typically with a maximum magnetic number along the quantization axis defined by the pump light wavevector) that is not stationary due to the presence of a transverse magnetic field.

In the usual picture of the BB operation, the macroscopic magnetization precesses around the magnetic field direction and it is periodically reinforced when its orientation is along the optical axis of the pump beam. The periodic reinforcement is obtained by modulating the pump radiation –its intensity, polarization or wavelength– synchronously with the precession. A weak and unmodulated probe radiation interrogates the evolution of the atomic state and produces an output signal whose dynamics is driven by the field under measurement.

The presence of time dependent field can modify the spin dynamics also in non-resonant conditions. As an example, in conventional BB magnetometry, slow-varying magnetic field variations, oriented along the bias field direction, are detected with a high sensitivity. Field variations along perpendicular directions produce some signal, as well, but with a second-order response [13].

Interesting dynamic responses of precessing spins are observed in a different regime, namely in the case of an intense fast-varying magnetic field. In particular, a strong magnetic field oscillating along the probe-beam axis (i.e. transversely to the static one) at a frequency much above the resonance may effectively freeze the atomic precession, according with a phenomenon originally studied in the late Sixties [14] and generally known as magnetic dressing.

In a recent work [15], we have demonstrated that the dressing phenomenon can be deeply tuned by the application of a secondary (and much weaker) field that oscillates at the same frequency –or at a low-order harmonic– of the dressing one, perpendicularly to the latter. Among the peculiarities of such tuned-dressed configuration, we pointed out that the system shows a remarkable response to small variations of the amplitude and/or of the phase of the tuning field. This work investigates with a proof-of-principle experiment the potential of such a tuning-dressing apparatus to detect tiny field variations due to eddy currents induced by the tuning field.

Experimental setup

The experimental setup is built around a BB magnetometer making use of cesium thermal vapor in a centimetric, gas buffered cell. The atomic vapour is synchronously pumped by D1D_1 radiation (at mW/cm2^2 irradiance) and polarimetrically probed by a weak (at µW/cm2^2 irradiance) D2D_2 radiation, co-propagating with the pump one along the xx direction, further details can be found in Ref. [16].

The atomic sample is merged in a constant field at µT level (let its direction be zz) oriented perpendicularly to the laser beams. This static field B0B_0 is obtained by partially compensating the environmental magnetic field. The task is accomplished by means of three large size (180 cm) mutually orthogonal Helmholtz pairs. Additional adjustable quadrupoles let improve the field homogeneity.

The tuning-dressing effect [15] occurs when two phase-related oscillating fields (BDB_D and BtB_t) are applied along xx to dress the atoms, and along yy to tune the dressing effect, respectively. A solenoidal coil (11 cm in length and 4 cm in diameter) surrounding the cell generates BDB_D and a small (millimetric) solenoid wound on a ferrite nucleus generates BtB_t. The interaction geometry is sketched in Figure 1.

Figure 1. Magnetic field and laser beams geometry. Circularly polarized D1D_1 pump radiation and linearly polarized D2D_2 probe radiation propagate along xx, which is the same direction of a strong oscillating (dressing) field BDB_D. A weaker (tuning) field BtB_t oscillates along yy synchronously with BDB_D; a weak static field B0B_0 is oriented along zz.

The BDB_D and BtB_t coils are supplied by two waveform generators (Agilent 33250A) phase-locked to each other. Series capacitors help to adapt the impedances, and a linear amplifier can be used to enhance the BtB_t amplitude. Each of these coils has 10 Ohm series resistor to precisely monitor the current phases and amplitudes via a 16 bit, 500 kS/s DAQ card (NI 6346).

The tuning-dressed phenomenon occurs and is well modeled and characterized for a tuning-field frequency ωt\omega_t that is an integer multiple of the dressing frequency ωD\omega_D (ωt=pωD\omega_t = p\omega_D). It is worth stressing that, in the considered application, switching among different pp values may enable fast and relevant variations of the skin depth, which in EII applications may constitute an interesting feature.

The coil-specimen-sensor geometry is sketched in Figure 2. The used specimens are one or more Aluminum disks of assigned thickness dd, diverse radii RR, centered on the yy axis, and lying on a xzxz plane between the BtB_t generator and the atomic sensor. The experimental results reported in this work are obtained with aa =40 mm, bb =130 mm, dd =30 µm, and RR ranging from 24 mm to 350 mm. In this proof-of-concept experiment, these values are chosen on the basis of fortuitous constraints of the available setup. In real imaging applications, all of them could be selected in view of optimizing the detection performance, in terms of sensitivity and/or spatial resolution.

Figure 2. Relative positions of specimen, sensor, and tuning/dressing field sources. The specimen is a circular disk of radius RR (thick grey line), which is parallel to the xzxz plane and is centered on the yy axis, at a quote y=by=b over the sensor SS. A solenoidal coil CDC_D surrounds SS and produces a homogeneous field BDB_D along the xx axis. The latter coincides with the optical axis of the pump and probe lasers D1D_1 and D2D_2. A small dipolar source CtC_t located on the yy axis at distance aa from the specimen (and a+ba+b from the sensor) produces a tuning field BtB_t which is oriented along yy on the sensor, and excites tangential eddy currents in the disk. Those currents modify both the amplitude and the phase of the tuning field on SS.

The eddy currents induced in the conductive disks modify the amplitude and the phase of BtB_t in the sensor location. As summarized in Sec.III, both these parameters play a role in shifting the effective Larmor frequency, and this is the key feature at the basis of the proposed detection technique.

Effective Larmor Frequency

As shown and discussed in the Ref.[15], a magnetic field that oscillates in the direction perpendicular to both the transverse static field B0B_0 and the longitudinal dressing field BDB_D at a frequency pp times larger than that of the dressing field, modifies the effective Larmor frequency according to

ΩL=γ[B0J0(ξ)+BtJp(ξ)sin⁡(ϕ)],(1)\Omega_L=\gamma\left[B_0J_0(\xi)+B_tJ_p(\xi)\sin(\phi)\right], \tag*{(1)}

if pp is an odd integer, and to

ΩL=γ[B0J0(ξ)]2+[BtJp(ξ)cos⁡(ϕ)]2,(2)\Omega_L=\gamma\sqrt{\left[B_0J_0(\xi)\right]^2+\left[B_tJ_p(\xi)\cos(\phi)\right]^2}, \tag*{(2)}

if it is an even one. Here, γ\gamma is the atomic gyromagnetic factor (γ≈3.5\gamma\approx3.5 Hz/nT for Cesium), and the argument ξ\xi is set by the amplitude of the dressing field BDB_D and by its frequency ωD\omega_D according to

ξ=γBDωD.(3)\xi=\frac{\gamma B_D}{\omega_D}. \tag*{(3)}

The validity of the eqs.(1) and (2) requires that

ξ≫γBtωD,ω0ωD\xi\gg\frac{\gamma B_t}{\omega_D},\frac{\omega_0}{\omega_D}

In the measurements shown in this paper, the tuning field oscillates at the same frequency of BDB_D (i.e. p=1p=1), while both its amplitude and phase vary either due to different waveform-generator settings or to the presence of conductive samples placed in the proximity of the field source and of the sensor. In particular assuming that ϕ=φ−θ\phi=\varphi-\theta, where φ\varphi is the relative phase of the two RF generators, and θ\theta is the dephasing caused by the eddy currents in the specimen, the effective Larmor frequency can be expressed as

f=γ2π[J0(ξ)B0+J1(ξ)Btsin⁡ϕ]=f0+Asin⁡(φ−θ),(4)f=\frac{\gamma}{2\pi}\left[J_0(\xi)B_0+J_1(\xi)B_t\sin\phi\right]=f_0+A\sin(\varphi-\theta), \tag*{(4)}

where f0f_0 is the dressed frequency in absence of the tuning field. Beside causing the dephasing θ\theta, the eddy currents attenuate BtB_t, and this shielding effect determines a reduction of the parameter AA.

The eq.(4) lets define the conditions in which the system response to either the attenuation or the dephasing is maximized:

∣∂f∂A∣=∣sin⁡(φ−θ)∣(5)\left|\frac{\partial f}{\partial A}\right|=\left|\sin(\varphi-\theta)\right| \tag*{(5)}

is maximal for ϕ=φ−θ=π/2\phi=\varphi-\theta=\pi/2, while

∣∂f∂θ∣=∣Acos⁡(φ−θ)∣(6)\left|\frac{\partial f}{\partial\theta}\right|=\left|A\cos(\varphi-\theta)\right| \tag*{(6)}

is maximal when J1J_1 is maximum, that is at ξ≈1.84\xi\approx1.84 and ϕ=0,π\phi=0,\pi. These relations provide an indication of good experimental working conditions.

Results

The BtB_t variation due to the eddy currents induced in the specimen can be evaluated analytically for large (infinite radius) disks [17], or numerically for finite radius ones. The variation is monotonically dependent on the disk radius, and concerns both the phase and the amplitude of BtB_t. This is verified experimentally as shown in Figure 3

Figure 3. The resonance frequency as a function of the different relative phase φ\varphi, for various disk diameters whose size in mm is reported in the legend. The vertical error bar (≈3\approx3 Hz) is smaller than the symbols used in the graph. The solid lines represent the best fit profile obtained with the target function eq(4), with AA and θ\theta as fitting parameters.

The atomic resonance is experimentally analyzed by scanning the frequency of the pump-laser modulation around the effective Larmor frequency expressed by the eq.(4). The resonance line-width is about 25 Hz, however the high S/N ratio let a best fit targeted to a Lorentzian profile determine the peak frequency with a sub-Hz accuracy. On the other hand, the ambient field fluctuations (the BB magnetometer is operated in an unshielded environment, with deactivated field-stabilization system [18]) are at nT level, and this leads to a few Hz uncertainty in the resonant frequency estimation.

The measurements shown here are performed at B0=1.7B_0 =1.7 µT corresponding to an undressed Larmor frequency of 6 kHz. The dressing field frequency is set at 15 kHz or 30 kHz, and its amplitude BDB_D is set to get a dressing parameter ξ=1.84\xi= 1.84, according to the indication provided by the eq.(6). In these conditions, the dressed frequency with no tuning field, is f0=J0(1.84)γB0/2π≈1900f_0 = J_0(1.84)\gamma B_0/2\pi\approx1900 Hz.

When the tuning field is activated, the measured peak frequency has a remarkable dependence on both the amplitude and the phase of BtB_t at the sensor location, according to eq.(4). Both AA and θ\theta can be determined by scanning the relative phase of the RF generators in the [0,2π][0, 2\pi] interval.

The Figure 3 shows clearly that an increase in disk radius causes both a reduction of the amplitude AA and a phase shift of the sinusoidal profiles. It is also clear that larger variations of the effective Larmor frequency occur when ϕ=±π/2\phi= \pm\pi/2, while operating near the zero-crossings (ϕ=0,π\phi= 0,\pi) improves the sensitivity to ϕ\phi (i.e to θ\theta) variations, consistently with the eqs.(5), (6).

A clearer visualization is facilitated by a best fit procedure targeted to the function expressed in eq.4. Plots of the best-fit parameter AA versus the disk radius are shown with dots in Figure 4. The solid lines in that figure are results from a finite-elements numerical simulation. The latter is based on the geometry of Figure 2, without the CDC_D solenoid and considering BtB_t at the center of the sensor as produced by a point dipolar source and attenuated by on axis Al disks.

Figure 4. Dependence of the BtB_t amplitude on the Al disk radius inferred from experimental measurements (dots) and evaluated from a simplified numerical simulation (line, see text for more details). Both the cases of ωD=2π⋅15\omega_D = 2\pi\cdot15 kHz and 2π⋅302\pi\cdot30 kHz are considered. The thinner skin depth at 30 kHz produces a faster decay to a lower level, when the disk radius increases, as expected.

This analysis has been performed with the set of Al disks described in Sec.II and we report results obtained

at both the tested tuning-dressing frequencies. Selecting different values of the oscillating fields changes the skin depth and hence modifies the shielding properties of the Al samples. As known, the skin depth depends on the absolute permeability μ\mu of the medium and on its resistivity ρ\rho, according to:

δ=2ρωμ(7)\delta= \sqrt{\frac{2\rho}{\omega\mu}} \tag*{(7)}

In the case of Al, this formula gives δ=460\delta= 460 µm and 650 µm, for the two considered frequencies, respectively. Both values are much larger than the thickness of the foil from which the Aluminum disks (30 µm) are cropped.

The general trend of the experimental results reported in Figure 4 is qualitatively reproduced by numerical simulation. In particular both the experiment and the simulation show that the curve width and the asymptotic values decrease at higher ωD\omega_D (smaller δ\delta). However, some quantitative discrepancies emerge: double thickness Al disks are considered in the simulation to produce the well matching profiles reported in Figure 4. Indeed, several details are neglected in the simulation, such as the presence of the solenoid CDC_D and of an electric heater surrounding the Cs cell, and the finite size of the BtB_t source and of the sensor volume. Very likely one of more of these factors are responsible for the mentioned discrepancies. Similar issues are encountered in the phase estimation θ(r)\theta(r).

V Discussion

The proposed tuning-dressing technique, extends to all-optical atomic magnetometric sensors applicability as detectors in EII setups. Interestingly, the proposed arrangement could be easily implemented in pre-existing EII apparatuses, and dual modality (RF and tuning-dressing) setups could be built, with incremental complexity of the available instrumentation.

In arrangements where the pump and probe beams are not parallel, a dual-mode operation would make possible to apply the RF field along different orientation, in such a way to induce variously distributed eddy currents. In fact, while in RF magnetometry the RF field is applied perpendicularly to the pump beam, in the tuning-dressing BB arrangement BtB_t is perpendicular to the probe beam.

Compared to conventional RF apparatuses, the tuning-dressing implementation besides not requiring that the detection field oscillates at a resonant frequency. Moreover, fast selection of different BtB_t frequencies is feasible, via an appropriate selection of the harmonic parameter pp. The possibility of selecting arbitrary values for both ω\omega and pp is relevant for the skin depth dependence on ρω\rho\omega (eq. 7), which is of interest to improve the tomographic (three-dimensional) potential of the EII apparatuses.

Concerning the detection efficiency, the proposed methodology takes advantage from the fact that in the tuning-dressed configuration the weak tuning field borrows strength from the primary dressing field. The linear dependence on BtB_t occurring for odd pp (which can be maximized with appropriate choices of ξ\xi and selecting ϕ=±π/2\phi= \pm\pi/2) makes possible and efficient the use of approaches as that described in ref.[2]. Specifically it would be possible to apply a vanishing BtB_t obtained as a superposition of two opposite contributions, in such way that the specimen causes a BtB_t variation via an unbalance of the two applied terms.

Improved detection techniques could be developed on the basis of the harmonic dependence on ϕ\phi. As an example, if BtB_t oscillates at frequency ωt\omega_t slightly different from pωp\omega, the signal will appear as a low-frequency beating term oscillating at ωt−pω\omega_t-p\omega: a feature that would enable the application of phase-sensitive detection techniques.

Conclusion

We have presented a proof of principle experiment proposing a new kind of magnetometric detector of small or weakly conductive specimens. Specifically, we have used a recently studied dressing-field configuration to make an all-optical Bell-and-Bloom magnetometer suited to detect faint oscillating fields as occurring in magnetic induction tomography setups. The experiment demonstrated that the variation of a non-resonantly oscillating probe field can be effectively measured to detect the presence of conductive objects. The described approach can be regarded as non-resonant alternative to the well known methodologies based on RF magnetometers. It constitutes indeed a complementary approach to the problem and dual-mode (RF and tuning-dressing) apparatuses can be envisaged, with interesting novel features and enhanced flexibility.

References

  1. [1]H. Griffiths, “Magnetic induction tomography,” Measurement Science and Technology, vol. 12, pp. 1126–1131, jul 2001.
  2. [2]C. Deans, L. Marmugi, and F. Renzoni, “Sub-Sm⁻¹ electromagnetic induction imaging with an unshielded atomic magnetometer,” Applied Physics Letters, vol. 116, no. 13, p. 133501, 2020.
  3. [3]Z. Z. Yu, A. T. Peyton, M. S. Beck, W. F. Conway, and L. A. Xu, “Imaging system based on electromagnetic tomography (EMT),” Electronics Letters, vol. 29, no. 7, pp. 625–626, 1993.
  4. [4]A. Korjenevsky, V. Cherepenin, and S. Sapetsky, “Magnetic induction tomography: experimental realization,” Physiological Measurement, vol. 21, pp. 89–94, feb 2000.pubmed.ncbi.nlm.nih.gov/10720003
  5. [5]P. Bevington, R. Gartman, and W. Chalupczak, “Inductive imaging of the concealed defects with radio-frequency atomic magnetometers,” Applied Sciences, vol. 10, no. 19, 2020.
  6. [6]M. Soleimani, W. R. B. Lionheart, and A. J. Peyton, “Image reconstruction for high-contrast conductivity imaging in mutual induction tomography for industrial applications,” IEEE Transactions on Instrumentation and Measurement, vol. 56, no. 5, pp. 2024–2032, 2007.
  7. [7]L. Ma and M. Soleimani, “Magnetic induction tomography methods and applications: a review,” Measurement Science and Technology, vol. 28, p. 072001, jun 2017.DOI
  8. [8]C. Deans, L. Marmugi, S. Hussain, and F. Renzoni, “Electromagnetic induction imaging with a radio-frequency atomic magnetometer,” Applied Physics Letters, vol. 108, no. 10, p. 103503, 2016.DOI
  9. [9]A. Wickenbrock, N. Leefer, J. W. Blanchard, and D. Budker, “Eddy current imaging with an atomic radio-frequency magnetometer,” Applied Physics Letters, vol. 108, no. 18, p. 183507, 2016.DOI
  10. [10]K. Jensen, M. Zugenmaier, J. Arnbak, H. Stærkind, M. V. Balabas, and E. S. Polzik, “Detection of low-conductivity objects using eddy current measurements with an optical magnetometer,” Phys. Rev. Research, vol. 1, p. 033087, Nov 2019.
  11. [11]L. Marmugi, C. Deans, and F. Renzoni, “Electromagnetic induction imaging with atomic magnetometers: Unlocking the low-conductivity regime,” Applied Physics Letters, vol. 115, no. 8, p. 083503, 2019.
  12. [12]H. Scharfetter, R. Casanas, and J. Rosell, “Biological tissue characterization by magnetic induction spectroscopy (mis): requirements and limitations,” IEEE Transactions on Biomedical Engineering, vol. 50, no. 7, pp. 870–880, 2003.pubmed.ncbi.nlm.nih.gov/12848355
  13. [13]G. Bevilacqua, V. Biancalana, Y. Dancheva, A. Fregosi, and A. Vigilante, “Spin dynamic response to a time dependent field,” 2021.DOI
  14. [14]S. Haroche, C. Cohen-Tannoudji, C. Audoin, and J. P. Schermann, “Modified Zeeman hyperfine spectra observed in ¹H and ⁸⁷Rb ground states interacting with a nonresonant RF field,” Phys. Rev. Lett., vol. 24, pp. 861–864, 1970.
  15. [15]G. Bevilacqua, V. Biancalana, A. Vigilante, T. Zanon-Willette, and E. Arimondo, “Harmonic fine tuning and triaxial spatial anisotropy of dressed atomic spins,” Phys. Rev. Lett., vol. 125, p. 093203, Aug 2020.
  16. [16]G. Bevilacqua, V. Biancalana, P. Chessa, and Y. Dancheva, “Multichannel optical atomic magnetometer operating in unshielded environment,” Applied Physics B, vol. 122, no. 4, p. 103, 2016.
  17. [17]C. V. Dodd and W. E. Deeds, “Analytical solutions to eddy-current probe-coil problems,” Journal of Applied Physics, vol. 39, no. 6, pp. 2829–2838, 1968.
  18. [18]G. Bevilacqua, V. Biancalana, Y. Dancheva, and A. Vigilante, “Self-adaptive loop for external-disturbance reduction in a differential measurement setup,” Phys. Rev. Applied, vol. 11, p. 014029, Jan 2019.

Paper details

vlnc:W_53umn32b
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 0

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…
Contents