Josephson Parametric Amplifier based Quantum Noise Limited Amplifier Development for Axion Search Experiments in CAPP

Abstract

This paper provides a comprehensive overview of the development of flux-driven Josephson Parametric Amplifiers (JPAs) as Quantum Noise Limited Amplifier for axion search experiments conducted at the Center for Axion and Precision Physics Research (CAPP) of the Institute for Basic Science. It focuses on the characterization, and optimization of JPAs, which are crucial for achieving the highest sensitivity in axion particle detection. We discuss various characterization techniques, methods for improving bandwidth, and the attainment of ultra-low noise temperatures. JPAs have emerged as indispensable tools in CAPPs axion search endeavors, playing a significant role in advancing our understanding of fundamental physics and unraveling the mysteries of the universe.

Introduction

CAPP was established in October 2013, entering the realm of axion physics following the notable contributions of experiments like the Axion Dark Matter Experiment (ADMX), Haloscope At Yale Sensitive To Axion Cold dark matter (HAYSTAC), etc. [1, 2, 3, 4] These experiments paved the way for axion research, and CAPP aimed to build on these achievements. It focused on establishing itself as a significant player in the field by pushing the boundaries in all relevant parameters and expanding the detection sensitivity.

It is extremely difficult to observe axions because of their weak interactions with Standard Model (SM) fields. In the presence of a magnetic field, an axion with a mass of mam_a and momentum pap_a generates an electric field described by the Lagrangian,

Laγ=−gaγγαDMBext⋅E(1)\mathcal{L}_{a\gamma} = -g_{a\gamma\gamma}\alpha_{\mathrm{DM}}\mathbf{B}_{\mathrm{ext}}\cdot\mathbf{E} \tag*{(1)}

where gaγγg_{a\gamma\gamma} is axion to two-photon coupling constant [5]; aDMa_{\mathrm{DM}} is the axion dark matter field [6, 7]BextB_{\mathrm{ext}} and E\mathbf{E} are external magnetic and generated electrical field, respectively. Alternatively, it can be said that in the presence of a magnetic field field, the axion field mixes with the electromagnetic field of same frequency. The resulting field has frequency faxf_{ax} defined by the simple equation:

fax=(mac2)2+(pac)2h≈mac2h.(2)f_{ax} = \frac{\sqrt{(m_a c^2)^2 + (p_a c)^2}}{h} \approx\frac{m_a c^2}{h}. \tag*{(2)}

Here, hh and cc represent the Planck constant and the speed of light in vacuum, respectively. The kinetic energy (pac)2(p_a c)^2 of virialized axions is small, however, this kinetic energy broadens the line following a Maxwellian distribution with an equivalent quality factor of Qa≈106Q_a \approx10^6 [8]. To enhance the signal, we can adopt the haloscope concept introduced by Sikivie in 1983 [9] and first applied in 1987 by the authors of [10]. Sikivie’s proposed experimental setup involves a detector consisting of a microwave cavity resonator immersed in a strong magnetic field and cooled to cryogenic temperatures. The cavity plays a crucial role in resonantly accumulating the produced photons from the converted axion dark-matter field. When the frequency of the axion signal aligns with the resonance of the cavity, it may be observed using specialized low-noise readout equipment.

Haloscope

The axion haloscope employs a cavity with a high-quality factor, positioned within a strong magnetic field for the detection of signals arising when the axion frequency aligns with the cavity’s resonance. For the output signal power in haloscope experiments a larger magnetic field and cavity volume result in a greater conversion of axions into microwave photons (Eq. (3)) [11]:

P(a)=[(gγαfaπ)2ℏ3c3ρama2][β1+βωc1μ0B02VC010QL11+(2(fc−fax)Δfc)2].(3)P^{(a)} = \left[\left(\frac{g_{\gamma}\alpha f_a}{\pi}\right)^2 \frac{\hbar^3 c^3 \rho_a}{m_a^2}\right]\left[\frac{\beta}{1+\beta}\omega_c\frac{1}{\mu_0}B_0^2VC_{010}Q_L\frac{1}{1+\left(\frac{2(f_c-f_{ax})}{\Delta f_c}\right)^2}\right]. \tag*{(3)}

where gγ=(πfagaγγ)/αg_{\gamma} = (\pi f_a g_{a\gamma\gamma})/\alpha is the coupling constant of the axion-photon interaction with values of −0.97-0.97 and 0.360.36 for the Kim-Shifman-Vainshtein-Zakharov (KSVZ) [12, 13] and Dine-Fischler-Srednicki-Zhitnitskii (DFSZ) [14, 15] models, respectively; α\alpha is the fine structure constant; faf_a, ρa\rho_a and mam_a are the decay constant, local halo density and mass of the axion dark matter, respectively. β\beta is the coupling constant between the cavity and the readout system, with the term β/(1+β)\beta/(1+\beta) defining the fraction of the output power. fc=ωc/2πf_c = \omega_c/2\pi is the cavity resonance frequency; μ0\mu_0 is the magnetic permeability of free space; B0B_0 symbolizes the external magnetic field; VV, C010C_{010} and QLQ_L are the cavity volume, form factor for a particular mode, in our case TM010\mathrm{TM}_{010}, and the cavity loaded quality factor, Δfc=fc/QL\Delta f_c = f_c/Q_L. The form factor characterizes the degree of overlap between the cavity TM010\mathrm{TM}_{010} mode and the external magnetic field [16]:

C010=∣∫E⋅Bext dV∣2∫ϵ′∣E∣2 dV×⟨Bext2⟩V.(4)C_{010} = \frac{\left|\int\mathbf{E}\cdot\mathbf{B}_{\mathrm{ext}}\,dV\right|^2}{\int\epsilon'|\mathbf{E}|^2\,dV \times\langle\mathbf{B}_{\mathrm{ext}}^2\rangle V}. \tag*{(4)}

where ϵ=ϵ′+iϵ′′\epsilon= \epsilon' + i\epsilon'' is the relative permissivity within the cavity. A higher quality factor ensures that converted microwave photons persist within the cavity for extended durations, thereby amplifying the signal. Reducing the physical temperature of the cavity and minimizing noise from the receiver chain, further enhance the efficiency of axion detection.

In haloscope experiments, the typical cavity shape is cylindrical, fitting efficiently into the design of a regular solenoid superconducting magnet, making optimal use of the magnet’s bore volume. The cavity commonly utilizes the TM010\mathrm{TM}_{010} mode, where the electric field aligns with the cylinder’s axis and results in the maximum form factor. The central frequency of the cavity is primarily determined by its inner diameter. To precisely set the resonance frequency, one or more tuning rods, movable inside the cavity, are employed. Dielectric materials or highly conductive metals are chosen for the tuning rod based on the desired frequency.

To access the cavity signal, a movable antenna is employed to optimize the coupling β\beta. As indicated in Eq. (3), the output power strongly depends on cavity parameters C010C_{010}, QLQ_L, and β\beta. This necessitates optimization of the cavity design, considering the impact of tuning rods on C010C_{010} and aiming to avoid mode cross-coupling. The axion conversion power in the experiments performed at CAPP typically lays in a range between 10−2310^{-23} to 10−2210^{-22} W. Such minuscule power levels impose stringent demands on the initial stage of our readout system.

The signal frequency is linked to the axion’s mass mac2≈hfaxm_ac^2 \approx hf_{\mathrm{ax}}, an attribute that is currently unknown. Therefore, it is necessary to scan across an extensive frequency spectrum to maximize the probability of capturing this signal. The speed at which this frequency scan can be executed [8]:

S∝gγ4(SNR)2η1Tsys2B04V2C0102QL2.(5)S \propto\frac{g_{\gamma}^{4}}{(SNR)^{2}\eta}\frac{1}{T_{\mathrm{sys}}^{2}}B_{0}^{4}V^{2}C_{010}^{2}Q_{L}^{2}. \tag*{(5)}

SNRSNR denotes the chosen signal-to-noise ratio (for the average of 5 and the threshold of 3.7 for a confidence level of 90%); η\eta represents the data acquisition (DAQ) efficiency; TsysT_{\mathrm{sys}} is the system noise temperature.

Consequently, the scanning frequency rate is inversely proportional to the square of the total system noise. Hence, one of our primary goals is to minimize it, thereby augmenting of our scanning speed. System noise emanates from multiple sources within our setup, encompassing noise generated by the cavity, other passive components and the amplifiers. To achieve the lowest noise levels, it is imperative to cool down all readout components to temperatures lower than, T=(hfmin⁡)/kBT = (hf_{\min})/k_B, where kBk_B are the Boltzmann constants, respectively, and fmin⁡f_{\min} denotes the minimum scanning frequency of the cavity.

The most advanced amplifiers are capable of approaching noise levels close to the quantum limit. This article is dedicated to the adaptation and application of such amplifiers. In the following sections, we will elaborate on our measurement techniques for the JPAs, present their key parameters, and showcase both single JPA and multiple JPA readouts, including their respective noise characteristics.

Flux-driven JPA

In the earlier stages of CAPP axion search experiments, such as CAPP-9 T and CAPP-PACE [17, 18], we utilized state-of-the-art semiconductor amplifiers. The InP high-electron-mobility transistor (HEMT) is renowned for its ability to provide the lowest noise temperature in cryogenic low noise amplifier (LNA) applications. We employed those manufactured by Low Noise Factory, specifically designed for cryogenic use [19]. These devices exhibit exceptionally low noise temperatures, reaching about 1.2 K when operated at temperatures below 10 K for frequencies around 1 GHz [20]. These experiments demonstrated CAPP’s capability to create a high-quality radio-frequency (RF) tunable cavity, maintain low temperatures, and operate a strong superconducting magnet in persistent mode [17, 18]. It also provided invaluable experience in data collection and processing. However, readouts chained based on cryogenic HEMT amplifiers have noise levels several tens of times higher than the quantum noise limit. For this case it would take several decades to scan the range between 1 to 2 GHz with DFSZ sensitivity.

JPA design

To achieve near-quantum-noise-limited amplification, we employ a flux-driven JPA designed and fabricated for these experiments in RIKEN and the University of Tokyo [21]. This JPA design consists of a λ/4\lambda/4 resonator terminated by a superconducting quantum interference device (SQUID), as depicted in Figure 1. The main features of the devices are defined by using photolithography techniques on Nb film deposited on a 0.3-mm-thick silicon substrate. The SQUID, using Dolan-bridge type junctions [22], is fabricated using electron-beam lithography followed by shadow evaporation of aluminium. Its inner loop has a diameter of roughly 16 μ\mum, and the junctions are designed to have identical critical currents of approximately 2 μ\muA. The SQUID connects the λ/4\lambda/4 resonator to ground. A pump line couples inductively to the SQUID, delivering a modulated flux at twice the resonance frequency. This pump signal is rejected by the λ/4\lambda/4 resonator as it only permits odd harmonics, and therefore does not couple into the signal input-output line. The pump line is shunted to ground close to the SQUID location, improving the inductive coupling. A static magnetic field generated by a superconducting coil is used to tune the frequency of the JPA.

Flux-driven JPA: equivalent circuit diagram, JPA layout, and SQUID image

Figure 1. Flux-driven JPA. (a) Equivalent circuit diagram of the JPA. A flux bias defines the average inductance of the SQUID loop (the Josephson junctions are denoted as ⊠\boxtimes in the loop), while the pump is modulating this flux at twice the frequency of the signal to be amplified. Since the signal’s input and output utilize the same port, a cryogenic circulator is required to operate the amplifier. (b) JPA layout, where CcC_c is the interdigitated capacitor. The λ/4\lambda/4 resonator is implemented as a meander line, whose length determining the maximum JPA frequency. (c) Image of the SQUID (light structure in the center). The pump line is inductively coupled to the SQUID.

Different JPAs were designed for amplifying different signal frequencies by changing the length of the λ/4\lambda/4 line. The instantaneous bandwidth of the device is designed by choosing the coupling capacitor size. A larger instantaneous bandwidth, when keeping the design otherwise constant, necessities in an increased pump strength for a certain gain and will have a reduced saturation power. The tunable bandwidth of these devices was limited by the small ratio of the SQUID inductance to the geometric inductance of the λ/4\lambda/4 coplanar waveguide (CPW), which has been addressed in more recent JPA designs to be used in future experiments. The devices were designed using COMSOL simulations to obtain the capacitance value of the interdigitated capacitor and the resonance frequency of the λ/4\lambda/4 CPW resonator. Numerical simulations using the negative resistance model [23] were employed to predict the needed pump power, achievable gain, tunable range and instantaneous bandwidth.

The JPA is operated in the three-wave mixing mode [24], where the frequencies of the pump fpf_p, signal fsf_s, and idler fif_i are related by:

fp=fs+fi.(6)f_p = f_s + f_i. \tag*{(6)}

The JPA chips have dimensions of 2.5×5×0.3 mm32.5 \times5 \times0.3\ \mathrm{mm}^{3}, see Figure 1(b), and are fixed on top of a copper block. A printed circuit board (PCB) to connect microwave lines to the input and pump lines is fixed on the same copper block and wirebonded to the chip. This copper block is encased within a copper base for the DC bias coil and thermally bonded to it. When not in use, this assembly is stored in an antistatic plastic box at room temperature within an evacuated desiccator. The detailed sample holder design is described elsewhere [25].

JPA measurement setup

RF setup

In Figure 2, one can observe the basic setup for testing and for the axion search experiment, featuring various temperature stages, including 300 K, 4 K, still, cold plate (CP), and mixing chamber plate (MXC) of the dilution refrigerator (DR). This design is adaptable to different types of DRs and the nature of the experiment. It offers inputs for measuring various parameters such as tuning frequency range, gain, instantaneous bandwidth, and noise temperature of the JPAs. All RF connections from the 300 K flange to the 4K plate are made with CuNi coaxial 0.047-inch-diameter cables, provided either by BlueFors (BF) [26] or made in-house. Below 4K, some of the RF lines are done with the same cable, while the signal lines are constructed using superconducting NbTi 0.047-inch-diameter cables to minimize heat load and reduce losses. Each input line is equipped with a series of attenuators, thermally anchored to each temperature stage of the fridge, to reduce Johnson noise originating from higher temperature stages. The attenuator values are chosen by considering the equation below:

JPA readout diagram with either a cavity or a noise source at the input

Figure 2. JPA readout diagram with either a cavity or a noise source at the input. Components arranged from left to right: noise source for tests or cavity for experiment run, SW1: RF switch to toggle between JPA-test and axion-experiment modes; the Coupler: “cold” directional coupler for injecting test signals into the JPA; components C and IS1 are a cryogenic RF-circulator and an RF isolator; AMP1 and AMP2: cryogenic HEMT amplifiers; IS2: RF isolator between HEMTs; AMP3: room temperature amplifier; SW2: RF switch between vector network analyzer (VNA) and spectrum analyzer (SA); VNA, SA, and RF generator.

Tattout=Tattn−1A+Tattn(1−A),(7)T_{\mathrm{att}}^{\mathrm{out}} = T_{\mathrm{att}}^{n-1} A + T_{\mathrm{att}}^n (1-A), \tag*{(7)}

where TattoutT_{\mathrm{att}}^{\mathrm{out}} is the output noise temperature of the attenuator, Tattn−1T_{\mathrm{att}}^{n-1} is the output noise temperature of the attenuator on the previous temperature stage, TattnT_{\mathrm{att}}^n is the physical temperature of the attenuator, and AA is the attenuation coefficient. Here, we assumed that the physical temperature of the attenuator equals the temperature of the corresponding stage of the fridge. This assumption is appropriate since the dissipation and self-heating are negligible when small power signals are involved. The RF chain was installed on the MXC plate and consists of a range of cryogenic microwave components, including couplers, circulators, and isolators. Prior to their use at mK temperatures, each component was preliminary characterized on a cold testing bench. To control the bias current ibi_b we use a Yokogawa GS200 current source [27].

DC wiring, cold DC filters and switch boxes

To test our JPAs and other RF and DC components, we set up a dedicated test system using a BF LD400 dry fridge [28]. To enhance our testing capabilities, we implemented a set of 32 differential lines, each individually shielded, into the fridge. These lines are constructed using twisted pairs, each placed within a Teflon matrix and equipped with its own CuNi braided shield. For pairs connecting the room temperature (RT) flange to the 4 K stage, we utilized cables with 0.1 mm diameter brass wires. For connections between the 4 K stage and the MXC, twisted pairs were made from copper nickel coated 0.1 mm NbTi wires. The cables were thermally connected to each temperature stage through their CuNi shields. On the 4 K stage, they were thermally connected via a specialized interconnection box, providing a transition between brass and superconducting wires. At a base temperature of approximately 11 mK on the MXC, the estimated maximum heat load from wires connected to the upper stage does not exceed 0.8 μW\mu\mathrm{W}.

In the fridge DC lines, we incorporated DC filter assemblies at the 4 K stage. Each of our filter assemblies contained 12 differential T-type RC filters. In the lines used to provide power for the HEMT amplifiers, the RC filters are replaced with LC filters in order to reduce resistive losses. These filters are constructed on a two-layer PCB, utilizing both sides as ground planes to minimize interference and enhance thermalization, see Figure 3(a).

DC and RF wiring

Figure 3. DC and RF wiring. (a) Top and (b) bottom views of the cryogenic filter block for the BF fridge. The block contains 12 differential T-type RC filters designed for standard BF DC lines and is mounted on the 4 K stage of the fridge. (c) DC and (d) RF boxes of the test fridge BF4 located at room temperature.

To safeguard the connectors on the top of the BF fridge from potential damage due to frequent connection and disconnection during testing, we have designed two specialized boxes for DC and RF connections Figure 3(c) and (Figure 3d). These boxes are permanently connected to the fridge, and all connections for setup, testing, and experiments are made within these boxes. The RF box also houses an RF amplifier, while the DC box features switches controlled by a 40-411-001 PXI 64 Channel Relay Driver Module [29] based on a PXES-2590 chassis [30], enabling flexible configurations for testing purposes.

Noise source

Our setup incorporates a noise source (NS), facilitating the measurement of noise parameters for the JPAs and cryogenic HEMT amplifiers. The NS is a separate assembly comprising a wideband matched 50 Ω\Omega terminator, fixed on a bulkhead, labelled as “2” in Figure 4(a), a heater “3”, and a thermometer “4”. These components have been integrated onto a copper fixture (1). To enhance thermal connections, the enclosure was plated with gold of 6 μm\mu\mathrm{m} thick. The NS equivalent circuit is shown in Figure 4(b).

Cryogenic NS components, equivalent circuits, and cooling curves

Figure 4. Cryogenic NS. (a) One-channel NS components on the MXC assembly: 1 - enclosure, 2 - 50 Ω\Omega terminator, 3 - heater, 4 - thermometer, 5 - connector, 6 - thermalization wire, 7 - isolating plastic stand. (b) Equivalent circuit of the one channel NS. (c) Four-channel NS and (d) its equivalent circuit [31]. (e) Cooling of a single-channel NS heated up to 100, 200, 300 and 400 mK. The NS is thermally connected to the MXC with the temperature stabilized at 50 mK.

This NS has been designed for placement on the MXC plate of dry BF or wet Leiden dilution refrigerators. Our design prioritizes compact dimensions, minimizing the footprint on the MXC plate while ensuring adequate heat capacity for thermal stabilization. Bulkhead SMA adapters were employed to establish connections between the 50 Ω\Omega cryogenic wideband matched loads (2) and the measurement chain. These adapters, in conjunction with the wideband XMA cryogenic 50 Ω\Omega terminations boasting a maximum voltage standing wave ratio (VSWR) of 1.15, guarantee 50 Ω\Omega matched connections across the frequency range up to 18 GHz. At temperature TT, the source generates Johnson-Nyquist noise with the noise temperature TnT_n in accordance with the Nyquist theorem:

Tn=hfkB(12+1ehf/kBT−1).(8)T_n = \frac{h f}{k_B}\left(\frac{1}{2}+\frac{1}{e^{h f/k_B T}-1}\right). \tag*{(8)}

For the interconnections between the NS and the rest of the measurement chain, we utilized 0.047-inch-diameter NbTi microwave coaxial cables. Three superconducting NbTi twisted-pair cables within braided CuNi shields were employed to connect the temperature sensor (4) and heater (3) to the micro-Sub-D-9 connector (5) of the NS. The four-point temperature measurements were conducted using a calibrated ruthenium-oxide resistance sensor. The heater was crafted using a 100 Ω\Omega precision resistor, with its housing carefully removed. It was then securely attached to the NS assembly using epoxy glue. The Lake Shore Cryotronics 372 AC resistance bridge and temperature controller enabled precise temperature measurements and control. NbTi superconducting twisted pairs and RF cables were utilized to minimize losses and thermal conductance between the MXC and the NS connectors. Additionally, the NS assembly was securely attached to the MXC plate using two 20-mm-long plastic spacers (7). Since both DC and RF connections to the NS were made using superconducting cables, thermal anchoring to the MXC plate primarily relied on a weak link created with copper wire (6).

Through a proportional-integral-derivative (PID) controller, the NS temperature could be adjusted from 50 mK to 1 K without affecting the MCX plate temperature. Initially, we employed a single-channel NS in our setup. However, as our testing requirements evolved and the demand for increased testing speed became apparent, we transitioned to a four-channel NS configuration [31], all housed within a single enclosure as shown in Figure 4(c). The four channel NS equivalent circuit is shown in Figure 4(d). This allowed us to significantly accelerate our testing processes.

The time it takes for the temperature of the NS to stabilize depends on the heat capacitance of the NS itself, along with all the components mounted on it, and its thermal coupling to the MXC. This thermal conductance of the coupling is determined by the length and diameter of the copper magnet wire, the annealing of the wire, and the quality of the contacts between the wire, the NS enclosure on one side, and the MXC plate on the other. To ensure efficient thermal transfer, we took careful steps to establish good contacts. We removed the insulation clad from the wire and pressed the wire against the enclosure and plate using brass screws. We attempted to enhance contact reliability by introducing clamp contacts on both sides of the wires, but this approach led to a significant reduction in thermal conductivity, preventing us from lowering the temperature of our NS below about 180 mK. We attribute this to the introduction of substantial mechanical defects in the crimped connections, resulting in a significant increase in thermal resistance. A similar effect has been observed in other experiments [32]. After returning to our initial connection method, the thermal connection was restored.

In Figure 4(e), the relaxation of the NS temperature after it was set to target temperatures is shown. The relaxation curve deviates from an exponential shape due to the strong non-linearity of thermal conductance and heat capacities at millikelvin temperatures.

Shielding of JPAs and RF components from magnetic fields

According to Eq. (5), the scanning speed of axion haloscope searches is strongly determined by the magnet’s strength and size, making it one of the pivotal factors in these experiments. The presence of such a magnet also generates a large magnetic field in its vicinity. Contemporary magnets incorporate compensation coils to establish a reduced magnetic field zone near the magnet. Nevertheless, the available space for this purpose is constrained by the size of the fridge. Consequently, it may not be sufficient to accommodate all sensitive components. Our superconducting magnets, which generate a magnetic field ranging from 8 to 18 T, with an aperture up to 320 mm, produces in a magnetic field strength of below 20 mT around the RF-readout components. As such, magnetic shielding is necessary.

RF switch shielding

Within our experimental setup, we face the challenge of potential magnetic field interference affecting a range of critical components, including circulators, isolators, HEMT amplifiers, and notably, the JPA. To surmount this challenge, we have used shielded components, like circulators and isolators. For the unshielded components that can only function in low magnetic field environments, like RF switches Figure 5(a) and (Figure 5b), special shields were implemented.

Two types of RF switches and magnetic-field simulations

Figure 5. Two types of RF switches used at CAPP with and without magnetic shields. (a) Three-position RF switch. (b) Five-position RF switch. (c) and (d) ANSYS [33] simulations of magnetic field strength under an external vertical magnetic field of 100 mT for the three-position and five-position RF switches, respectively.

Onion shield

JPAs contain the most sensitive component to magnetic fields, i.e., the SQUID. In order to safeguard them against the influence of magnetic fields, we have developed a three-layer shielding system [34], referred to as the Onion Shield (OS) due to its nested design, see Figure 6(a). The performance of the OS results from its design and the properties of the materials used.

Onion shield showing the three-layer design, magnetic-field simulation, and JPA resonance frequency

Figure 6. Onion shield. (a) Three-layer design of the shield. (i) Niobium or NbTi superconducting shield, (ii) Cryophy shield, (iii) inner superconducting Al shield. The JPA locations are indicated by the arrows. (b) ANSYS [33] simulation of the magnetic field including only the two outer layers of the OS under an external field of 50 mT and temperature of 100 mK. (c) Resonance frequency frf_{\mathrm{r}} of the JPA in the passive mode depending on the applied DC flux bias, both in the presence and absence of an applied vertical residual magnetic field of the superconducting magnet. The field strength at the components location is expected to be lower that 50 mT. The measured shielding factor corresponding to these conditions ranges between 10510^{5} and 10610^{6}.

The shield has three layers, each with its own function to block and control magnetic fields. The outermost layer [(i) in Figure 6(a)], constructed with Nb or Nb alloy superconducting materials, has a high critical magnetic field value, and the external magnetic field is shielded by it only at cryogenic temperatures. The second layer (ii), consisting of a ferromagnetic material with high magnetic permeability [35, 36], shields magnetic fields penetrating the outer shield through the structural openings. Finally, the innermost layer (3) is also made of superconducting aluminum with a critical temperature of about 1.2 K.

LayerOuterMiddleInner
MaterialNb or NbTiCRYOPHYAl
Superconducting transition, K9.2-11N/A1.2
Maximum magnetic field, T>0.1≈10−4\approx10^{-4} (saturation)<0.01

Table 1. Onion shield layer properties.

At non-cryogenic temperatures, the OS acts like a regular magnetic shield due to its second layer made of ferromagnetic material. In fields with a magnitude close to the Earth’s field, the calculated field inside the OS is a few tens of nT. During the initial cooling, the outermost layer undergoes a superconducting transition at the critical temperature (for our material choice it is between 9.2 and 11 K). During the transition from the normal to the superconducting state, the outer layer of the OS “freezes” the Earth’s magnetic field inside the shield. Throughout this process, the magnetic field inside the second shield, composed of a ferromagnetic material, remains predominantly unaffected.

As the cooling process continues and the temperature approaches the critical point for the innermost layer (iii) of the shield, its phase transition occurs. This innermost layer of the shield, made of superconducting aluminum, freezes the residual magnetic field that penetrated through the previous layers, i.e. a few tens of nT.

Since the first layer of the OS is in a superconducting state, while the field strength of the magnet increases, a persistent current appears on the surface of the first layer of the OS due to Meissner effect. This current freezes the magnetic field within this layer. However, some magnetic field penetration is inevitable through the openings required for cable routing and thermalization.

The second shielding layer plays a pivotal role in substantially reducing the penetration of the magnetic field into the first layer of the OS. However, again due to cable holes, there is a small penetration through it. This change in the residual field of the ferromagnetic shield is further shielded by the aluminum layer. Persistent currents induced on the surface of the third shield counteract minor field fluctuations within the second layer, ensuring that these fluctuations do not impact the magnetic field within the OS internal volume. Consequently, this configuration effectively shields the internal volume from the influence of external magnet field variations. The magnetic field within the third layer of the shield undergoes only marginal changes, even when the magnet is on, remaining within a range of 100 nT. This is confirmed by both simulation and experiments. Simulations in Figure 6(b) and measurement, in Figure 6(c), showed consistent results. The measurements were done with a 5.8 GHz JPA.

JPA characterization

For all of our devices, we conduct thorough testing procedures. Since the JPA functions as a resonant amplifier, our objectives are to determine the tunable range, explore the relationships between gain, instantaneous bandwidth, and noise as a function of pump power and detuning frequency, and establish the input signal range within which the amplifier does not saturate. The key steps in our characterization process include:

  1. Passive resonance identification by measuring the resonance frequency as a function of the DC flux.

  2. Gain parametric map (paramap) construction (gain GG vs. pump power PP and detuning δ=fp/2−fr\delta= f_p/2 - f_r per bias current ibi_b), illustrating the gain concerning both pump power and detuning frequency.

  3. The 1 dB gain compression point (P1dBP_{\mathrm{1dB}}) measurement in order to determine the input saturation power of the amplifier.

  4. Noise paramap generation by measuring the noise temperature TnT_n vs. pump power and detuning.

Resonance frequency vs DC flux bias current

In the passive regime, where no pump signal is applied, the JPA operates as a weakly coupled quarter-wave resonator, shorted to ground through a non-hysteretic SQUID, and coupled to the outside circuit via a capacitor CcC_c [see Fig. 1(a)]. When subjected to a small input signal, the SQUID remains in a superconducting state, effectively acting as a flux-dependent inductor. Consequently, we can treat the JPA as a quarter-wavelength resonator, with its effective inductance tunable by adjusting the flux Φext\Phi_{\mathrm{ext}} going through the SQUID loop,

Leff=Φ02πIc∣cos⁡(πΦext/Φ0)∣(9)L_{\mathrm{eff}} = \frac{\Phi_0}{2\pi I_c \left|\cos\left(\pi\Phi_{\mathrm{ext}}/\Phi_0\right)\right|} \tag*{(9)}

where Φ0=2.067⋅10−15\Phi_0 = 2.067 \cdot10^{-15} Wb represents the flux quantum, and Ic≈Ic1≈Ic2I_c \approx I_{c1} \approx I_{c2} signifies the critical currents of junction 1 and junction 2, respectively.

The operating frequency of the JPA is determined by both the physical dimensions of the transmission line and the inductance of the SQUID. To adjust this frequency within a specific range, we utilize an inductively coupled DC flux bias coil. This coil is constructed as a single-layer coil wound with 0.05 mm copper-coated NbTi wire on a copper core, featuring an inner diameter of 32.5 mm and a length of 11.5 mm. It comprises 200 windings and has an approximate resistance of 200 Ω\Omegas at room temperature.

The impedance of a quarter-wavelength resonator becomes finite and real at resonance, ideally zero. As a result, the reflection coefficient exhibits a dip at resonance. However, in the low loaded quality factor regime, it results in a small and broad dip spread over a wide frequency range (approximately 100–200 MHz). When measuring the JPA through various components, each with its own characteristics, distinguishing this dip becomes challenging. On the other hand, phase measurement is comparatively easier because the resonance induces a 360-degree phase shift, whereas the phase shifts of other components do not exhibit such a strong frequency dependence. For this reason, we utilize phase-frequency measurements. We examine the JPA by sending a tiny signal towards it and recording the reflected signal. To separate the incoming and outgoing waves, a circulator is employed. The signal from the JPA is then amplified using two low-noise cryogenic HEMT amplifiers, AMP1 and AMP2, and RT amplifier AMP3, see Figure 2. The output signal recorded at the VNA input is influenced by several frequency-dependent components within the chain, resulting in the observed signal profile depicted in Figure 7(a). The reflection signal of the JPA is obtained through S21S_{21} measurements using a VNA in the configuration shown in Figure 2. Subsequently, the phase information of this spectrum is employed to extract the resonance frequency, utilizing a parameter fit based on Ref. [37]. The resonance frequency is measured as a function of the coil current, as shown in Figure 7(b). The measurement reveals that the minimum observable resonance frequency was below 1.25 GHz, while the maximum reached 1.33 GHz. However, at lower frequencies, the JPA exhibits large sensitivity to flux noise, attributed to the higher ∂fr∂ib\frac{\partial f_r}{\partial i_b}, and therefore this device was only used above 1.26 GHz.

Passive resonance of the JPA: phase versus frequency and resonance frequency versus coil current

Figure 7. The passive resonance of the JPA. (a) Dependence of S21S_{21} phase vs. frequency for a JPA measurement in the circuit in Figure 2 for a fixed current when the pump signal is switched off. (b) Relationship between the resonance frequency and the DC flux bias for a JPA.

The phase of reflected signal is given by:

arg⁡(S11)=arctan⁡(2Γ0δfδf2+[ΓR2−Γ02])(10)\arg(S_{11}) = \arctan\left(\frac{2\Gamma_{0}\delta f}{\delta f^{2} + [\Gamma_{R}^{2} - \Gamma_{0}^{2}]}\right) \tag*{(10)}

Here, δf=f−fr\delta f = f - f_{r} indicates the frequency deviation from the resonance frequency frf_{r}. ΓR=fr/(2Qi)\Gamma_{R} = f_{r}/(2Q_{i}) signifies the half-width of the resonance curve, determined by the internal quality factor QiQ_{i}. It signifies the bandwidth of the resonator attributed to its internal losses. Similarly, Γ0=fr/(2Qe)\Gamma_{0} = f_{r}/(2Q_{e}) denotes the half-width of the resonance curve, determined by the external quality factor QeQ_{e}. It denotes the bandwidth of the resonator, accounting for both intrinsic and coupling losses.

JPA gain measurement

Upon applying a pump signal with frequency fp≈2fsf_{p} \approx2f_{s}, the three-wave mixing mode obeys Eq. (6). Using the JPA in three-wave mixing mode causes an amplified signal and an idler mode to appear symmetrically around fp/2f_{p}/2. In the circuit shown in Figure 2, we estimate the JPA gain through the change in amplification in the output by comparing transmission measurements with the JPA in both on and off states. This establishes the baseline measurement, denoted as SoffS_{\mathrm{off}}. A separate measurement is performed with a microwave short circuit connector replacing the JPA to assess the accuracy of this baseline. The comparison indicates that the baseline obtained from this off-resonance measurement is at most 0.2 dB less efficient than if an ideal reflection was achieved.

Next, the JPA is tuned to the desired working point, and a subsequent transmission measurement SonS_{\mathrm{on}} is taken. From these two measurements, the power gain is determined via equation:

GJ=10log⁡10∣SonSoff∣2.(11)G_{J} = 10\log_{10}\left|\frac{S_{\mathrm{on}}}{S_{\mathrm{off}}}\right|^{2}. \tag*{(11)}

To find the optimal working point, we measure a “paramap”, which represents the dependence of GJG_{J} measured at fs=fp/2+1f_{s} = f_{p}/2 + 1 kHz on the pump frequency fpf_{p} and pump power PpP_{p}. The 1 kHz offset is enough to avoid the degenerate mode of the JPA. This measurement is repeated for all tuning frequencies of the JPA by varying ibi_{b}.

To determine the operational points across the entire frequency range, we devised a two-step procedure. Optimization included finding values of fpf_{p} and PpP_{p} for minimum noise and acceptable gain across the entire operational frequency range controlled by the ibi_{b} value. Our desired gain falls within the range of 15 to 20 dB, which minimizes the noise contribution of the subsequent HEMT amplifiers.

During paramap measurements, after each adjustment of ibi_{b}, the resonance frequency frf_{r} is established through phase measurements and subsequent parameter fitting. With the detuning defined as δ=fp/2−fr\delta= f_{p}/2 - f_{r}, iso-gain contours exhibit a minimum in required pump power near δ=0\delta= 0, as depicted in Figure 8(a). Initially, this minimum is observed at specific frequencies; however, it gradually shifts towards lower detunings, a phenomenon attributed to pump-induced shifts in the resonance frequency [37]. As illustrated in Figure 8(a), amplification can reach up to 40 dB.

Characterization of the JPAs

Figure 8. Characterization of the JPAs. (a) and (b) Gain and noise parameter maps of the 6 GHz JPA for fr=5.921f_{\mathrm{r}} = 5.921 GHz. The area of optimal working points for fr=5.921f_{\mathrm{r}} = 5.921 GHz is shown with pink circles. δn=1/BW(fp/2−fr)\delta_{\mathrm{n}} = 1/\mathrm{BW}(f_{\mathrm{p}}/2 - f_{\mathrm{r}}) is the normalized detunning, BW refers to the JPA bandwidth. (c) Gain dependence of the 2.3 GHz JPA on input power for one resonance frequency. (d) P1dBP_{\mathrm{1dB}} compression point for three resonance frequencies of the 2.3 GHz JPA.

Based on the data acquired from the gain paramap, we outlined an area of interest and measured a noise paramap [Fig. 8(b)] within that region. The displayed gain and noise paramap are for a 6 GHz JPA. As evident, we recorded an added noise level of around 150 mK for this amplifier, slightly exceeding the quantum limit for a phase insensitive amplifier at this frequency (ℏω)/(2kB)≈141(\hbar\omega)/(2k_{B}) \approx141 mK [38].

For each JPA, we conducted numerous measurements at different frequencies of interest to pinpoint the optimal configuration. The results presented in Figure 8(c) and (Figure 8d) showcase the outcomes of noise measurements for the 6 GHz JPA using different settings at two distinct frequencies. The red vertical lines correspond to the quantum limits for each of these frequencies. As illustrated, for each frequency, we achieved parameters that align with very low noise levels, approaching the quantum limit.

P1dB measurement

Due to the nonlinear nature of the JPA, its gain will saturate beyond certain signal and pump amplitudes. A common method for assessing amplifier saturation involves determining the output power at which the gain decreases by 1 dB. This process involves driving the amplifier with a continuous tone at the desired signal frequency and progressively raising the input power level while tracking the output power. P1dBP_{\mathrm{1dB}} is an indicator that the system deviates from the desired small-signal behavior.

Our measurement of P1dBP_{\mathrm{1dB}} was conducted at δf=1\delta f = 1 kHz across the entire JPA tunable range and various pump powers, each corresponding to different gains. The results are illustrated in Figure 8(e) and (Figure 8f). The power due to the axion signal, which is expected to be around −180-180 dBm, is significantly below the saturation power of the JPA. However, for a few JPAs it was observed that thermal noise at the input can also induce saturation, altering the device behavior. As an example, when we considered a device with operating frequencies around 2.3 GHz at gains exceeding 23 dB, we observed saturation due to thermal noise when the NS temperature exceeded 120 mK. While this may not render the device unusable at these frequencies, we need to to careful during direct noise temperature measurements using a NS in these frequency and gain ranges [39].

Idler mode and added noise

In the three-wave mixing non-degenerative mode, an idler tone appears with frequency fi=fp−fsf_i = f_p - f_s (see Figure 9). The amplification of the signal at the idler frequency is equal to

JPA output power spectrum showing the amplified signal and idler tones

Figure 9. JPA output power measured by the spectrum analyzer including the amplified signal and its idler. A single signal tone is applied to the input of the JPA.

GI=GJ−1.(12)G_I = G_J - 1. \tag*{(12)}

When the input signal PINP_{\mathrm{IN}} of the fsf_s frequency is applied to the JPA, the output of the amplifier will have two power terms, corresponding to fsf_s and fif_i frequencies:

POUT=GJ(PIN+hfs2)+GI(PIN+hfi2).(13)P_{\mathrm{OUT}} = G_J \left(P_{\mathrm{IN}} + \frac{h f_s}{2}\right) + G_I \left(P_{\mathrm{IN}} + \frac{h f_i}{2}\right). \tag*{(13)}

At GJ≫1G_J \gg1, the added noise equals half a photon [40, 41].

Noise measurement

To ascertain the noise temperature of the JPA, we utilize two steps. The first is based on the Y-factor method [42], while the second involves measuring the noise spectrum both with the JPA on and off, a technique known as the spectrum comparison method [43].

Modified Y-factor

To measure the noise temperature, we utilized a methodology akin to the well-established Y-factor method [42], employing the NS discussed earlier in § 4.3. The Y-factor technique is commonly applied in the field of RF and microwave measurements, particularly for evaluating the noise figure of amplifiers or receivers. It involves calculating the ratio of output noise power observed when a system is connected to a calibrated hot load (high temperature) PHOTP^{\mathrm{HOT}} to that observed when connected to a calibrated cold load (low temperature) PCOLDP^{\mathrm{COLD}}. Mathematically, it is expressed as:

Y=PHOTPCOLD=TsysHOTTsysCOLD(14)Y = \frac{P^{\mathrm{HOT}}}{P^{\mathrm{COLD}}} = \frac{T_{\mathrm{sys}}^{\mathrm{HOT}}}{T_{\mathrm{sys}}^{\mathrm{COLD}}} \tag*{(14)}

The right part of Eq. (14) reflects the fact that the noise temperatures TsysHOTT_{\mathrm{sys}}^{\mathrm{HOT}} and TsysCOLDT_{\mathrm{sys}}^{\mathrm{COLD}} are proportional to the corresponding noise powers. The noise temperature of a system in both cases contains the amplifier-added noise TaddT_{\mathrm{add}}:

TsysHOT=TnHOT+Tadd,TsysCOLD=TnCOLD+Tadd(15)T_{\mathrm{sys}}^{\mathrm{HOT}} = T_n^{\mathrm{HOT}} + T_{\mathrm{add}}, \qquad T_{\mathrm{sys}}^{\mathrm{COLD}} = T_n^{\mathrm{COLD}} + T_{\mathrm{add}} \tag*{(15)}

From Eqs. 14 and 15, knowing YY, we obtain:

Tadd=TnHOT−TnCOLDYY−1,Gsys=PHOT−PCOLDkBBW(TnHOT−TnCOLD)(16)T_{\mathrm{add}} = \frac{T_n^{\mathrm{HOT}} - T_n^{\mathrm{COLD}}Y}{Y-1}, \qquad G_{\mathrm{sys}} = \frac{P^{\mathrm{HOT}} - P^{\mathrm{COLD}}}{k_B BW\left(T_n^{\mathrm{HOT}} - T_n^{\mathrm{COLD}}\right)} \tag*{(16)}

where BWBW is the bandwidth. The advantage of this method is that it does not affect the precision of gain measurement GsysG_{\mathrm{sys}}.

So, the original Y-factor method measures the noise temperature TsysT_{\mathrm{sys}} at two points. To mitigate the influence of quantum effects on Johnson noise given by Eq. (8) and ensure control linearity, which can be compromised by JPA saturation, we have modified the standard method by measuring TsysT_{\mathrm{sys}} at multiple temperature points.

For illustration, we considered the measurement of noise with a 2.3 GHz JPA. The noise power emerging from the system was measured with the SA with a 1 kHz resolution bandwidth. The power spectra were recorded at NS temperatures TsT_s of 60, 120, and 180 mK. The power values were converted into power spectral densities (PSD) by dividing them by the resolution bandwidth of the SA. The pump power and resonance frequency were adjusted to achieve a JPA gain of approximately 20 dB.

The lower limit on noise temperature for linear, phase-insensitive amplifiers is described by [44]:

Taddmin⁡=lim⁡S0→0Sn(f,T)kB=hf2kB(17)T_{\mathrm{add}}^{\min} = \lim_{S_0 \to0}\frac{S_n(f,T)}{k_B} = \frac{hf}{2k_B} \tag*{(17)}

which results in approximately 55.2 mK at 2.3 GHz. From the measurement in Figure 10a we obtained a total added noise temperature of Tadd≈120T_{\mathrm{add}} \approx120 mK, at the cavity physical temperature of 50 mK. This cavity temperature at 2.3 GHz frequency, according to Eq. (8), corresponds to a noise temperature of ≈70\approx70 mK. Therefore, the total system noise temperature is estimated to be Tsys≈190T_{\mathrm{sys}} \approx190 mK. The lower bound for TsysT_{\mathrm{sys}} is determined by the standart quantum limit (SQL) [45], which is approximately 110 mK at 2.3 GHz. Therefore, for the 2.3 GHz haloscope experiment, we obtain a total system noise of 1.7 SQL.

Spectrum comparison

The second method we employ in the noise measurement is the spectrum comparison method. An advantage of this method is that it does not require the continuous usage of a NS, resulting in faster measurements as there is no need for changing the temperature. This technique is applicable because the JPA can be switched on and off, allowing noise measurements in both states. The off state of the JPA corresponds to a condition where GJoff≈1G_J^{\mathrm{off}} \approx1 and Toff≈0T^{\mathrm{off}} \approx0 (absence of the pumping signal of the JPA and changing the bias current to move the JPA passive resonance away from the measured frequency range). The noise calculation consists of comparing the power spectra of the JPA in the “on” and “off” states and deducing GsysG_{\mathrm{sys}} and TaddT_{\mathrm{add}} from the results obtained with the JPA in the off state.

Initially, we characterized the HEMT noise by performing the Y-factor method with the JPA turned off. For our experiments around 1 GHz the best noise we have measured with the RF-chain including HEMTs corresponds to 1 K, which is approximately 40 times the half-quanta limit. GsysoffG_{\rm sys}^{\rm off} and TsysoffT_{\rm sys}^{\rm off} are measured using a VNA. After that the JPA is turned on, and GsysonG_{\rm sys}^{\rm on} is measured. The JPA gain GJG_J is calculated using Eq. (11). The VNA emits a signal of ample strength, rendering the noise negligible. By sweeping the frequency of the narrow-band signal, we can obtain the dependence of GJ(fs)G_J(f_s).

2.3 GHz JPA noise measurements

(a)

1.58 GHz JPA noise measurements

(b)

Figure 10. Noise measurements from two different JPAs: (a) 2.3 GHz JPA noise measured using the Y-factor method. The upper plot shows the power spectral densities measured for fr=2.305f_r = 2.305 GHz for three noise source temperatures at the optimal working point [39]. The inset displays the PSD of noise for three frequencies used to calculate the noise temperature. The frequency dependence of the calculated noise temperature is shown in the bottom plot. δf\delta f is a deviation from frf_r, δf=f−fr\delta f = f - f_r. (b) 1.58 GHz JPA noise measured using the spectrum comparison method. The upper plot shows the output power spectra for the JPA ON and JPA OFF. The inset displays the JPA gain curve. The noise temperature depending on the offset frequency is shown in the bottom plot.

The noise power spectrum on the output of AMP3 (see Figure 2), PsysonP_{\rm sys}^{\rm on}, is recorded by a SA when the JPA is on. Upon determining GsysonG_{\rm sys}^{\rm on}, TJT_J can be estimated. Consequently, the total added noise of the JPA measured with the spectrum comparison method, TaddSCT_{\rm add}^{\rm SC}, is computed as:

TaddSC≈PsysonPsysoffTsysoffGJ−(Ts+TaddoffGJ),(18)T_{\rm add}^{\rm SC} \approx\frac{P_{\rm sys}^{\rm on}}{P_{\rm sys}^{\rm off}} \frac{T_{\rm sys}^{\rm off}}{G_J} - \left(T_s + \frac{T_{\rm add}^{\rm off}}{G_J}\right), \tag*{(18)}

where TsT_s is the stable physical temperature of the noise source while conducting the spectrum comparison method.

As confirmed by our measurements, both methods yield results that are consistent within the measurement error. However, the spectrum comparison method allows achieving the same precision much faster. In recent experiments, we have only utilized the modified Y-factor method to measure the frequency dependence of the noise temperature of the HEMT amplifier chain. We have employed spectrum comparison methods to measure the noise of the JPA.

Optimization of working point

Paramap Optimization

Based on the JPA tests described in this section, we obtain an optimal working point (OWP) for each signal frequency fsf_s and JPA gain of choice GJG_J (usually close to 20 dB by three parameters: ibi_b, fpf_p, and PpP_p. To determine the OWP, we used mainly two approaches. Historically the first method is based on the observation that the minimum noise at otherwise equal conditions coincides with the lowest PpP_p.

The JPA gain is a function of frequency, typically with a peak occurring at fs=fp2f_s = \frac{f_p}{2}. Using GJG_J to denote the peak gain value, the tuning is done by following the steps below:

  1. Tune ibi_b such that frf_r is equal to the desired fsf_s.

  2. Set fpf_p to 2fs2f_s.

  3. Increase PpP_p until the desired gain is obtained.

  4. Repeat steps 2 and 3 for small deviations from frf_r by changing ibi_b.

  5. Pick a set of ibi_b, fpf_p, PpP_p that has the lowest PpP_p for the desired gain at fsf_s.

Before running the experiment, we conduct a series of preliminary measurements to determine the optimal values of the JPA parameters (fsf_s, GJG_J), which include ibi_b, fpf_p, and PpP_p. Further details are described in [46].

We have also devised a method to heuristically find an optimized working point using the Nelder-Mead algorithm [47]. This approach allowed us to identify a working point with a low noise temperature near a given initial seed [48].

CAPP JPA collection

Since 2019, we have designed and tested over 40 JPAs for CAPP’s axion search experiments. Parameters for some of these JPAs are presented in Table 2.

JPA#1#2#3#4#5#6#7#8#9
Min Frequency, GHz1.0061.1181.1411.21.321.351.41.481.56
Max Frequency fmax⁡f_{\max}, GHz1.0711.1851.1711.2511.4011.401.451.551.64
Tunable range, MHz656730518150507080
TaddT_{\mathrm{add}}, mK<90<90<60<95<80<80<85<75<85
Tn/(hfmax⁡/(2kB))T_n/(hf_{\max}/(2k_B))3.53.22.133.22.382.382.442.012.15
Bandwidth at 20dB gain (kHz)>150>150>100>200>150>150>150>200>200
JPA#10#11#12#13#14#15#16#17#18
Min Frequency, GHz1.631.642.272.302.352.353.75.15.5
Max Frequency fmax⁡f_{\max}, GHz1.731.742.312.452.52.53.855.46.0
Tunable range, MHz1001004015150150120300500
TaddT_{\mathrm{add}}, mK<80<80<120<130<100<110<120<145<150
Tn/(hfmax⁡/(2kB))T_n/(hf_{\max}/(2k_B))1.931.922.162.321.671.831.31.121.04
Bandwidth at 20dB gain (kHz)>200>150>100>200>200>100>300>1000>1000

Table 2. Characteristics of some of the JPAs measured at physical temperature of 30 mK.

Extension of amplification bandwidth

Our primary challenge stemmed from the constrained tunable range imposed by the necessity to cool down our substantial 1.2-ton magnet using liquid helium. Consequently, we opted for a wet dilution refrigerator as a key component in our experimental setup to ensure safe superconducting magnet operations. Each cooldown of the dilution refrigerator consumes several hundred liters of liquid helium, and replacing the JPA requires several weeks of effort. Therefore, it is practical to match the amplifier’s bandwidth with the cavity’s, which is approximately 300 MHz. This can be achieved by employing a split-band technique, wherein each amplifier operates within its designated frequency band.

Cold multiplexer

Initially we considered a measurement setup with a microwave RF switch as the simplest means for extending the scanning frequency range of the RF chain. The multiplexing design of the multiple JPA readout scheme was based on the five-position RF switch, shown as SW2 in Figure 5a. This way, we designed the setup, such that each channel of the RF switch is connected to the one of five JPAs, utilizing a common output, see Figure 11a. The axion signal detection chain starts from the cavity, connecting to the RF switch SW1, which is used to switch between the cavity (scanning for axions) and the noise source (noise measurements). The first directional coupler DC1 connects to the BYPASS line, which is used for the characterizing of the passive resonances of each JPA. Similar to the circuit diagram shown in Figure 2 the cryogenic circulator C1 is connected to the JPA. The five-position RF switch SW2 switches between JPAs according to the operating frequency range in axion search experiment. The pump line for each JPA utilizes a power divider PD, and the pump frequency and power are set according to the active JPA. The output of multiplexed JPA circuit is connected to the microwave isolator I and an additional directional coupler DC2. DC2 is needed to perform additional measurements of the HEMT chain, see HEMT bypass line in Figure 11a, to measure the losses in the output chain of the JPA, and to precisely characterise the noise temperature. In order to characterize the differences in transmission for each channel of the RF switch, we performed additional measurements with the circuit diagram shown in Figure 11b. We used a four channel noise source [31], shown as 4 CH NS, with 4 identical cryogenic wideband terminators connected to the five-position RF switch. The common port of the RF switch is connected to the cryogenic circulator and to the JPA. We performed measurements of the JPA noise temperature for each of the channels and did not observe any differences.

JPA multiplexing setup at the MXC

Figure 11. JPA multiplexing setup at the MXC: (a) The diagram of the detection chain in the axion search experiment. The setup consists of the cavity readout line and five JPAs measuring line. The switching is provided by the RF switch SW1. Compared with the JPA readout 2, a 5-position RF switch SW2 is added, which connects five JPAs through superconducting cables and protective bias-tees. The setup utilizes one channel noise source 1 CH NS. The pump signal is split using the power divider PD. The output is followed by an additional bias-tee, cryogenic microwave isolator I1, directional coupler DC2, and cryogenic isolator I2. (b) Setup for characterizing the transmission of each channel of the 5-position RF switch.

In our early tests, we encountered an issue where a JPA was damaged. Further investigation revealed the potential for charge accumulation on the switch electrodes during mechanical movement. This accumulated charge could discharge at the moment of switching, generating current and voltage pulses that posed a threat to the amplifiers. At mK temperatures, leakage current can be exceptionally low due to charge carriers freezing in insulating materials, resulting in prolonged charge retention on the electrodes. Once we recognized this issue, we implemented bias tees, shown as Bias-T in Figure 11a, to safely discharge the switch, enabling us to use this method for switching between several JPAs without causing damage. Another challenge associated with this method is the necessity to supply a substantial current (≈250\approx250 mA) inside the fridge to facilitate channel switching and reset the RF switch. To achieve this, we would need to modify the fridge by adding more low-resistance cables. As of now, this modification has been deferred. After addressing these issues, the proposed circuit scheme, which utilizes a 5-position multiplexer combined with the parallel-series connection design described in the next chapter, will allow for the expansion of the operating frequency range beyond 1 GHz.

Series connection of JPAs

In order to increase the bandwidth of our readout, we introduced a series connection involving an additional JPA alongside the first JPA [JPA1 and JPA2, see Figure 12(b)]. During measurements, only one JPA operates at a given time, while the other JPA remains detuned from its operational frequency using a DC flux bias. This configuration seamlessly integrates into the readout chain via the second circulator, resulting in additional losses of approximately 0.4 dB at frequencies around 1 GHz. These losses, corresponding to less than 10% of the signal, lead to a minor acceptable increase in noise temperature.

Simplified amplifier schematics

Figure 12. Simplified amplifier schematics. (a) single JPA connection, (b) serial connection, (c) parallel and (d) parallel-serial connections.

The series connection has enabled us to simultaneously use two JPAs within the same cooldown cycle, a setup we have employed in our test fridges since 2021.

Parallel connection of JPAs

The second idea for the bandwidth extension is to connect multiple JPAs with different frequencies on a single PCB, which appeared feasible given that the wavelength of the signal at 1–2 GHz significantly exceeds the length of the PCB traces of a few mm. The dimensions of the PCB we employ comfortably accommodate three chips, as depicted in Figure 13(a).

Parallel connection of three JPAs

Figure 13. Parallel connection of three JPAs [49]. (a) PCB for the parallel connection of 3 JPAs. The DC flux bias is common for all of the JPAs and is created by a 200-turn coil wound with superconducting wire around the PCB holder (not shown). (b) Dependence of the resonance frequency on bias current for three JPAs connected in parallel, with the JPA pump switched off.

The initial measurements using this parallel configuration demonstrated that the JPAs can operate independently, exhibiting noise and gain characteristics similar to those in a single-JPA operation configuration [49]. However, certain challenges may arise because the two JPAs on opposite sides of the PCB [see Fig. 13(a)] share the same sensitivity to DC bias, causing their flux-to-frequency characteristics to change similarly. In Figure 13(a), we can see three JPAs mounted on a PCB in order of their highest passive resonance frequency (HPRF): 1.28, 1.295, and 1.36 GHz. In Figure 13(b), one can see the dependence of the resonance frequency on bias current for the three JPAs connected in parallel, with the JPA pump power switched off. As we can observe, there is similarity in the DC flux bias sensitivity of the JPAs with maximum resonance frequencies of 1.28 and 1.36 GHz. This similarity can potentially lead to interference in the most sensitive low-frequency range, a matter which is discussed further in § 6.5.

Parallel-serial connection

Combining the two ideas and developing a parallel-serial JPA readout, where two parallel configurations are linked in series was the next logical step [50]. This configuration was initially evaluated in an RF assembly named “Leo-I”, which contained all RF components intended to be placed at MXC. The goal was to test the entire assembly in a dry fridge first, before moving it to the wet CAPP-MAX fridge without any modifications.

Before the dry fridge test, we arranged three JPAs in parallel with HPRFs of 1.07, 1.12, and 1.13 GHz in the first holder based on the Onion shield, while the second holder accommodated a single JPA of 1.185 GHz. Unfortunately, the middle JPA of 1.12 GHz, which was initially mounted and characterized on a single-JPA PCB version, was damaged during the transfer to the three-JPA PCB setup. Subsequent tests on the remaining three JPAs confirmed that all of them exhibited low noise levels while maintaining a required gain of 20 dB.

The complexity introduced into the readout chain led to some uneven frequency response and variations in HEMT noise temperature measured with the JPA off, ranging from 1.09 to 1.7 K [Figure 14(a) and (b)]. This cannot be entirely eliminated due to inherent phase delays in cables and the theoretical impossibility of achieving a perfect match for three-port non-reciprocal components within the chain. However, we accepted this limitation because at a JPA gain of 20 dB, the effect of the next HEMT amplification stage on noise is reduced by a factor of 100, resulting in only a small impact on the system noise temperature.

Characteristics of multi-JPA amplifier based on Leo I assembly

Figure 14. Characteristics of multi-JPA amplifier based on Leo I assembly. (a) HEMT amplifier gain, (b) HEMT amplifier noise. (a) and (b) were measured with all JPAs OFF. (c)–(f) Noise temperatures of JPAs 1–4, respectively. The red dashed line corresponds to the quantum limit for the amplifier added noise.

In Run 6 of the CAPP-MAX experiment, as mentioned earlier, three JPAs were in operation. Two of them were situated within one holder, while the third was mounted in a separate holder. The DC flux bias was shared by connecting the DC flux coils of both holders in series. Further details of the measurements conducted during Run 6 of the CAPP-MAX experiment utilizing this amplifier can be found in [25, 51].

JPA interference in split band amplifier

For Run 7 of the CAPP-MAX experiment, we prepared a new ”Leo II” assembly, comprising a full set of 6 JPAs with HPRFs of 1.25, 1.30, 1.36, 1.40, 1.45, and 1.50 GHz in a parallel-serial connection as shown in Figure 12(d). To minimize possible interference, we ordered the JPAs to avoid close operating frequencies among neighbors. In the first holder, we positioned JPAs starting with the lowest HPRF (1.25 GHz), followed by 1.36 GHz, and finally 1.45 GHz. The second holder followed a sequencing of 1.30 GHz, 1.40 GHz, and 1.50 GHz.

Tests of this assembly (Figure 15) in the dry fridge revealed characteristics similar to those of Leo I. These JPAs allowed us to cover a frequency range of about 300 MHz, spanning from 1.2 to 1.5 GHz. However, interference between some of the JPAs at their lowest operating frequencies emerged during testing.

Six-JPA assembly (Leo II), with (a) assembly installed at the mixing chamber plate of the dry dilution refrigerator, (b) flux-sweep curves of the six JPAs, and (c) added noise temperature depending on the frequency for the six JPAs

Figure 15. Six-JPA assembly (Leo II). (a) Assembly installed at the mixing chamber plate of the dry dilution refrigerator. (b) Flux-sweep curves of the six JPAs. (c) Added noise temperature depending on the frequency for the six JPAs. The red dashed line corresponds to the quantum limit for the amplifier added noise.

This setup was tested with a series connection of DC bias lines for the two JPA holders. This configuration posed challenges, particularly in achieving detuning of the JPA resonance frequencies from each other, especially within the narrow range of the lowest operating frequencies where the sensitivity of JPAs to interference is most pronounced. Due to the use of dc bias coils with the same design and almost identical JPA flux sensitivities, leveraging the Φ0\Phi_{0} periodicity of SQUID characteristics was not feasible. Utilizing another branch of the flux-to-frequency characteristic would have resulted in the same interference problem. We decided not to modify the design due to the risk of damage.

While it would have been feasible to utilize different twisted pairs to supply DC flux bias current for the second holder, we deliberated several ideas to tackle this issue, with the objective of minimizing the interval between runs without requiring time-consuming fridge modifications.

Another approach involved connecting the DC-flux coils in parallel, with small resistors of different values in series with the coils in each branch. In this configuration, the coils will have different sensitivity to input current, but there is still a probability of interference for JPAs in different holders due to overlap at lower frequencies.

Furthermore, we developed and implemented a new circuit design using Schottky diodes. This circuit allowed us to independently apply two DC flux biases to two JPA holders using the same twisted pair of wires. It achieved this by separating the flux biases using currents of different directions to bias one or the other JPA holder. The circuit featured a diode rectifier operating at the 4 K stage of the fridge to address heat dissipation in the diodes [52].

A potential issue arose during the operation of GaAs diodes in proximity to a strong magnetic field due to the field’s effect on the diode’s performance. This effect could lead to changes in properties depending on the orientation of the diode in the magnetic field [53]. To mitigate this concern, we replaced each of the two diodes with three diodes, each oriented along different orthogonal Cartesian axes [52].

After Run 6 of our axion experiment, we re-accessed the fridge design. We decided to upgrade our DC lines and implement a twisted pair to control the second JPA holder independently. As a result, for Run 7, we were able to forego the use of this diode circuit, postponing its installation.

Before Run 7, we conducted a 6-JPA assembly (Leo-II) measurements and confirmed their noise characteristics [see Figure 15(c)] in the experiment fridge.

Conclusions

In this paper, we have presented an overview of the characterization and implementation of JPAs at CAPP, detailing the processes involved in achieving high sensitivity in axion detection. Drawing from the established flux-driven JPA technology developed by RIKEN and the University of Tokyo, we successfully adapted these amplifiers for the CAPP haloscope search experiments through several key developments:

  1. We investigated and devised methods to operate flux-driven JPAs across a continuous frequency range. This involved introducing the paramap method to determine the operating points aimed at achieving specific gains and minimal noise. We established a test setup based on a dilution fridge, complemented by a computer-controlled system for JPA testing and experiment execution.

  1. We introduced two methods for extending the effective amplifier frequency range, utilizing series and parallel connections. Our experiments with multiple JPA configurations, including the 6-JPA assembly, have demonstrated the efficacy of our approaches in achieving a broad frequency range. By selecting JPA sequences and employing different techniques, we successfully minimized interference and maximized sensitivity without compromising noise characteristics. By connecting 6 JPAs, each with an operating frequency band of 48–52 MHz and different central frequencies, we expanded our coverage to approximately 300 MHz, which is almost six times the range of a single JPA. It

enhances our scanning capabilities by minimizing downtime and helium loss during cooldowns for the CAPP-MAX system [25].

  1. We introduced a nested shield based on superconductors and Cryophy, enabling operation in high magnetic fields up to 0.1 T without saturating the Cryophy and achieving a shielding factor of 10510^{5}–10610^{6} with residual fields below 100 nT for stable JPA operation. A patent application for this innovation is currently underway.

  1. Our JPA development has facilitated the implementation of JPA-based amplifiers in various CAPP experiments, as described in [54, 55, 56, 57, 58, 25], and has allowed the CAPP to contribute to the global axion search effort.

These developments pave the way for further breakthroughs in future CAPP experiments.

List of Acronyms

ADMXAxion Dark Matter Experiment
BFBlueFors
CAPPCenter for Axion and Precision Physics Research
CPcold plate
CPWcoplanar waveguide
DAQdata acquisition
DRdilution refrigerator
DFSZDine-Fischler-Srednicki-Zhitnitskii
HAYSTACHaloscope At Yale Sensitive To Axion Cold dark matter
HEMThigh-electron-mobility transistor
HPRFhighest passive resonance frequency
IBSInstitute for Basic Science
JPAsJosephson Parametric Amplifiers
KAISTKorea Advanced Institute of Science and Technology
KSVZKim-Shifman-Vainshtein-Zakharov
LNAlow noise amplifier
MXCmixing chamber plate
NSnoise source
OSOnion Shield
OWPoptimal working point
PCBprinted pircuit board
PIDproportional-integral-derivative
PSDpower spectral densities
RFradio-frequency
RQCRIKEN Center for Quantum Computing
RTroom temperature
SAspectrum analyzer
SMStandard Model
SQLstandart quantum limit
SQUIDsuperconducting quantum interference device
VNAvector network analyzer
VSWRvoltage standing wave ratio

Table 3.

Conflict of Interest Statement

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Author Contributions

SVU led the development of the amplifier and proposed the idea of parallel connection of the JPAs. JMK proposed the idea of series connection of the JPAs. AFL and YN developed and produced the JPAs. SVU, CK, JMK, and BII developed the test setup. SO and SVU contributed to sample preparation. CK and JMK prepared the software control. CK, JMK, and BII performed the tests in the test fridge. JMK, BII, and SA tested the setup in the axion experiment fridge. BII, MK, and SVU contributed to component tests. BII, MK, SVU, JMK, and SA prepared the figures. SVU, CK, JSK, JMK, BII, AFL, YN, SA, and YKS contributed to the interpretation of the results. YKS led the axion search in CAPP. SVU wrote the manuscript, and CK, JSK, JMK, BII, AFL, YN, MSK, SA, and YKS contributed to text refinement. All authors agree to be accountable for the content of the work.

Acknowledgments

This work was supported by the Institute for Basic Science (IBS-R017-D1) and JSPS KAKENHI (Grant No. JP22H04937). Arjan F. van Loo was supported by a JSPS postdoctoral fellowship. The authors thank Dr. Seongtae Park for his assistance with the PCB development.

References

  1. [1]N. Du, N. Force, R. Khatiwada, E. Lentz, R. Ottens, L. J. Rosenberg, G. Rybka, G. Carosi, N. Woollett, D. Bowring, A. S. Chou, A. Sonnenschein, W. Wester, C. Boutan, N. S. Oblath, R. Bradley, E. J. Daw, A. V. Dixit, J. Clarke, S. R. O’Kelley, N. Crisosto, J. R. Gleason, S. Jois, P. Sikivie, I. Stern, N. S. Sullivan, D. B Tanner, and G. C. Hilton. Search for invisible axion dark matter with the axion dark matter experiment. Phys. Rev. Lett., 120:151301, 2018.DOI
  2. [2]T. Braine, R. Cervantes, N. Crisosto, N. Du, S. Kimes, L. J. Rosenberg, G. Rybka, J. Yang, D. Bowring, A. S. Chou, R. Khatiwada, A. Sonnenschein, W. Wester, G. Carosi, N. Woollett, L. D. Duffy, R. Bradley, C. Boutan, M. Jones, B. H. LaRocque, N. S. Oblath, M. S. Taubman, J. Clarke, A. Dove, A. Eddins, S. R. O’Kelley, S. Nawaz, I. Siddiqi, N. Stevenson, A. Agrawal, A. V. Dixit, J. R. Gleason, S. Jois, P. Sikivie, J. A. Solomon, N. S. Sullivan, D. B. Tanner, E. Lentz, E. J. Daw, J. H. Buckley, P. M. Harrington, E. A. Henriksen, and K. W. Murch. Extended search for the invisible axion with the axion dark matter experiment. Phys. Rev. Lett., 124:101303, 2020.DOI
  3. [3]B. M. Brubaker, L. Zhong, Y. V. Gurevich, S. B. Cahn, S. K. Lamoreaux, M. Simanovskaia, J. R. Root, S. M. Lewis, S. Al Kenany, K. M. Backes, I. Urdinaran, N. M. Rapidis, T. M. Shokair, K. A. van Bibber, D. A. Palken, M. Malnou, W. F. Kindel, M. A. Anil, K. W. Lehnert, and G. Carosi. First results from a microwave cavity axion search at 24 μeV. Phys. Rev. Lett., 118:061302, 2017.DOI
  4. [4]K.M. Backes, D. A. Palken, S. Al Kenany, B. M. Brubaker, S. B. Cahn, A. Droster, Gene C. Hilton, Sumita Ghosh, H. Jackson, S. K. Lamoreaux, A. F. Leder, K. W. Lehnert, S. M. Lewis, M. Malnou, R. H. Maruyama, N. M. Rapidis, M. Simanovskaia, Sukhman Singh, D. H. Speller, I. Urdinaran, Leila R. Vale, E. C. van Assendelft, K. van Bibber, and H. Wang. A quantum enhanced search for dark matter axions. Nature, 590:238, 2021.
  5. [5]Giovanni Grilli di Cortona, Edward Hardy, Javier Pardo Vega, and Giovanni Villadoro. The QCD axion, precisely. Journal of High Energy Physics, 2016(34):1–36, 01 2016.DOI
  6. [6]F. Wilczek. Problem of strong p and t invariance in the presence of instantons. Phys. Rev. Lett., 40:229, 1978.DOI
  7. [7]S. Weinberg. A new light boson? Phys. Rev. Lett., 40:223–226, 1978.DOI
  8. [8]Dongok Kim, Junu Jeong, SungWoo Youn, Younggeun Kim, and Yannis K. Semertzidis. Revisiting the detection rate for axion haloscopes. Journal of Cosmology and Astroparticle Physics, 2020(03):066, mar 2020.DOI
  9. [9]P. Sikivie. Experimental tests of the “invisible” axions. Phys. Rev. Lett., 51:1415, 1983.
  10. [10]S. DePanfilis, A. C. Melissinos, B. E. Moskowitz, J. T. Rogers, Y. K. Semertzidis, W. U. Wuensch, H. J. Halama, A. G. Prodell, W. B. Fowler, and F. A. Nezrick. Limits on the abundance and coupling of cosmic axions at 4.5 < m_a < 5.0 μeV. Phys. Rev. Lett., 59:839–842, Aug 1987.pubmed.ncbi.nlm.nih.gov/10035884
  11. [11]Benjamin M. Brubaker. First results from the HAYSTAC axion search. PhD thesis, 2018.arxiv.org/abs/1801.00835
  12. [12]J. E. Kim. Weak-interaction singlet and strong CP invariance. Phys. Rev. Lett., 43:103, 1979.DOI
  13. [13]M. A. Shifman, A. I. Vainshtein, and V. I. Zakharov. Can confinement ensure natural CP invariance of strong interactions? Nucl. Phys. B, 166:4933, 1980.DOI
  14. [14]A. P. Zhitnitsky. Possible suppression of axion–hadron interactions. Sov. J. Nucl. Phys. (Engl. Transl.); (United States), 31:2:260, 1980.
  15. [15]Michael Dine, Willy Fischler, and Mark Srednicki. A Simple Solution to the Strong CP Problem with a Harmless Axion. Phys. Lett. B, 104:199–202, 1981.
  16. [16]Junu Jeong, SungWoo Youn, Sungjae Bae, Dongok Kim, Younggeun Kim, and Yannis K Semertzidis. Analytical considerations for optimal axion haloscope design. Journal of Physics G: Nuclear and Particle Physics, 49(5):055201, Mar 2022.DOI
  17. [17]Junu Jeong, Sung Woo Youn, Sungjae Bae, Jihgeun Kim, Taehyeon Seong, Jihn E Kim, and Yannis K Semertzidis. Search for invisible axion dark matter with a multiple-cell haloscope. Physical Review Letters, 125(22):221302, 2020.DOI
  18. [18]Ohjoon Kwon, Doyu Lee, Woohyun Chung, Danho Ahn, HeeSu Byun, Fritz Caspers, Hyoungsoon Choi, Jihoon Choi, Yonuk Chong, Hoyong Jeong, Junu Jeong, Jihn E. Kim, Jinsu Kim, Çağlar Kutlu, Jihnhwan Lee, MyeongJae Lee, Soohyung Lee, Andrei Matlashov, Seonjeong Oh, Seongtae Park, Sergey Uchaikin, SungWoo Youn, and Yannis K. Semertzidis. First results from an axion haloscope at capp around 10.7 μeV. Phys. Rev. Lett., 126:191802, May 2021.
  19. [19]https://lownoisefactory.com/product/lnf-lnc4_8f/.
  20. [20]J Schleeh, G Alestig, J Halonen, A Malmros, B Nilsson, P ÅNilsson, J P Starski, N Wadefalk, H Zirath, and J Grahn. Ultralow-power cryogenic InP HEMT with minimum noise temperature of 1 K at 6 GHz. IEEE Electron Device Letters, 33(5):664, 2012.DOI
  21. [21]T. Yamamoto, K. Inomata, M. Watanabe, et al. Flux-driven Josephson parametric amplifier. Applied Physics Letters, 93(4):042510, 2008.
  22. [22]G. J. Dolan. Offset masks for lift-off photoprocessing. Applied Physics Letters, 31(5):337–339, 09 1977.DOI
  23. [23]Kyle M. Sundqvist and Per Delsing. Negative-resistance models for parametrically flux-pumped superconducting quantum interference devices. EPJ Quantum Technology, 1(1):1–21, 03 2014.DOI
  24. [24]A. Roy and M. Devoret. Introduction to parametric amplification of quantum signals with Josephson circuits. Comptes Rendus Physique, 17(7):740–755, 2016.DOI
  25. [25]Saebyeok Ahn, JinMyeong Kim, Boris I. Ivanov, Ohjoon Kwon, HeeSu Byun, Arjan F. van Loo, SeongTae Park, Junu Jeong, Soohyung Lee, Jinsu Kim, Çağlar Kutlu, Andrew K. Yi, Yasunobu Nakamura, Seonjeong Oh, Danho Ahn, SungJae Bae, Hyoungsoon Choi, Jihoon Choi, Yonuk Chong, Woohyun Chung, Violeta Gkika, Jihn E. Kim, Younggeun Kim, Byeong Rok Ko, Lino Miceli, Doyu Lee, Jiwon Lee, Ki Woong Lee, MyeongJae Lee, Andrei Matlashov, Pallavi Parashar, Taehyeon Seong, Yun Chang Shin, Sergey V. Uchaikin, SungWoo Youn, and Yannis K. Semertzidis. Extensive search for axion dark matter over 1 GHz with CAPP’s Main Axion eXperiment. arXiv:2402.12892, submitted in PRX, 2024.arxiv.org/abs/2402.12892
  26. [26]https://bluefors.com.
  27. [27]https://tmi.yokogawa.com/kr/solutions/products/generators-sources/source-measure-units/gs200/.
  28. [28]https://bluefors.com/products/dilution-refrigerator-measurement-systems/.
  29. [29]https://www.pickeringtest.com/en-us/product/64-channel-relay-driver-module.
  30. [30]https://www.adlinktech.com/Products/PXI_PXIe_platform/PXIChassis/PXES-2590.
  31. [31]Boris I. Ivanov, Jinmyeong Kim, Çağlar Kutlu, Arjan F. van Loo, Yasubonu Nakamura, Sergey V. Uchaikin, Seonjeong Oh, Violeta Gkika, Andrei Matlashov, Woohyun Chung, and Yannis K. Semertzidis. Four-channel system for characterization of Josephson parametric amplifiers. In Proceedings of the 29th International Conference on Low Temperature Physics (LT29), pages 0112001–0112006. JPS Con. Proc., 2023.DOI
  32. [32]Gray Rybka. Private Communication, May 2023.
  33. [33]Ansys simulation software, https://www.ansys.com/.
  34. [34]Sergey V. Uchaikin, Boris I. Ivanov, Jinmyeong Kim, Çağlar Kutlu, Arjan F. van Loo, Yasubonu Nakamura, Seonjeong Oh, Violeta Gkika, Andrei Matlashov, Woohyun Chung, and Yannis K. Semertzidis. CAPP axion search experiments with quantum noise limited amplifiers. In Proceedings of the 29th International Conference on Low Temperature Physics (LT29), pages 0112011–0112016. JPS Con. Proc., 2023.DOI
  35. [35]https://www.aperam.com/product/cryophy/.
  36. [36]P. Arpaia, M. Buzio, O. Capatina, K. Eiler, S.A.E. Langleslag, A. Parrella, and N.J. Templeton. Effects of temperature and mechanical strain on ni-fe alloy cryophy for magnetic shields. Journal of Magnetism and Magnetic Materials, 475:514–523, 2019.DOI
  37. [37]Philip Krantz, Yarema Reshitnyk, Waltraut Wustmann, Jonas Bylander, Simon Gustavsson, William D Oliver, Timothy Duty, Vitaly Shumeiko, and Per Delsing. Investigation of nonlinear effects in Josephson parametric oscillators used in circuit quantum electrodynamics. New Journal of Physics, 15(10):105002, oct 2013.DOI
  38. [38]H. A. Haus and J. A. Mullen. Quantum noise in linear amplifiers. Phys. Rev., 128:2412, 1962.DOI
  39. [39]Kutlu Ç., van Loo A.F., Uchaikin S.V., Matlashov A. Lee D., Oh S., Kim J.S, Chung W., Nakamura Y., and Semertzidis Y.K. Characterization of a flux-driven Josephson parametric amplifier with near quantum-limited added noise for axion search experiments. Supercond. Sci. Technol., 34:085013, 2021.DOI
  40. [40]L. Kuzmin, K. Likharev, V. Migulin, and A. Zorin. Quantum noise in josephson-junction parametric amplifiers. IEEE Transactions on Magnetics, 19(3):618–621, 1983.DOI
  41. [41]E Babourina-Brooks, A Doherty, and G J Milburn. Quantum noise in a nanomechanical duffing resonator. New Journal of Physics, 10(10):105020, oct 2008.
  42. [42]Glenn F. Engen. A new method of characterizing amplifier noise performance. IEEE Transactions on Instrumentation and Measurement, 19(4):344–349, 1970.DOI
  43. [43]H.T. Friis. Noise figures of radio receivers. Proceedings of the IRE, 32(7):419–422, 1944.DOI
  44. [44]A. A. Clerk, M. H. Devoret, S. M. Girvin, Florian Marquardt, and R. J. Schoelkopf. Introduction to quantum noise, measurement, and amplification. Rev. Mod. Phys., 81:1208, Apr 2010.DOI
  45. [45]Carlton M. Caves. Quantum limits on noise in linear amplifiers. Phys. Rev. D, 26:1817–1839, Oct 1982.DOI
  46. [46]Çağlar Kutlu, Saebyeok Ahn, Sergey V. Uchaikin, Soohyung Lee, Arjan F. van Loo, Yasunobu Nakamura, Seonjeong Oh, and Yannis K. Semertzidis. Systematic approach for tuning flux-driven Josephson parametric amplifiers for stochastic small signals. In Proceedings of the 29th International Conference on Low Temperature Physics (LT29), pages 0112021–0112027. JPS Con. Proc., 2023.DOI
  47. [47]J. A. Nelder and R. Mead. A Simplex Method for Function Minimization. Comput. J., 7(4):308–313, 01 1965.DOI
  48. [48]Younggeun Kim, Junu Jeong, SungWoo Youn, Sungjae Bae, Arjan F. van Loo, Yasunobu Nakamura, Sergey Uchaikin, and Yannis K. Semertzidis. Parameter optimization of Josephson parametric amplifiers using a heuristic search algorithm for axion haloscope search. arXiv:2404.18345, submitted in Electronics, 2024.DOI
  49. [49]Sergey V. Uchaikin, Jinmyeong Kim, Boris I. Ivanov, Arjan F. van Loo, Yasunobu Nakamura, Saebyeok Ahn, Seonjeong Oh, Seongtae Park, Andrei Matlashov, Woohyun Chung, and Yannis K. Semertzidis. Improving amplification bandwidth by combining Josephson parametric amplifiers for active axion search experiments at IBS/CAPP. Acceptet in Journal of Low Temperature Phys, 2023.DOI
  50. [50]Jimmyeong Kim, Minsu Ko, Sergey V. Uchaikin, Boris I. Ivanov, Arjan F. van Loo, Yasunobu Nakamura, Seonjeong Oh, V. Gkika, Yannis K. Semertzidis, and Saebyeok Ahn. Wideband amplification for ibs-capp’s axion search experiments. https://www.researchsquare.com/article/rs-3550142/latest, submitted in Journal of Low Temperature Physics, 2023.DOI
  51. [51]Jimmyeong Kim, Boris I. Ivanov, Çağlar Kutlu, Seongtae Park, Arjan F. van Loo, Yasunobu Nakamura, Sergey V. Uchaikin, Seonjeong Oh, Violeta Gkika, Andrei Matlashov, Woohyun Chung, and Yannis K. Semertzidis. Josephson parametric amplifier in axion experiments. In Proceedings of the 29th International Conference on Low Temperature Physics (LT29), pages 0111991–0111996. JPS Con. Proc., 2023.DOI
  52. [52]MinSu Ko, Sergey V. Uchaikin, Boris I. Ivanov, JinMyeong Kim, Seonjeong Oh, and Violetta Gkika. Expanding scanning frequency range of Josephson parametric amplifier axion haloscope readout with Schottky diode bias circuit. In 2023 IEEE 24th International Conference of Young Professionals in Electron Devices and Materials (EDM), pages 810–813, 2023.DOI
  53. [53]Z. G. Sun, M. Mizuguchi, T. Manago, and H. Akinaga. Magnetic-field-controllable avalanche breakdown and giant magnetoresistive effects in Goldsemi-insulating-GaAs Schottky diode. Applied Physics Letters, 85(23):5643–5645, 2004.DOI
  54. [54]Jinsu Kim, Ohjoon Kwon, Çağlar Kutlu, Woohyun Chung, Andrei Matlashov, Sergey Uchaikin, Arjan Ferdinand van Loo, Yasunobu Nakamura, Seonjeong Oh, HeeSu Byun, Danho Ahn, and Yannis K. Semertzidis. Near-quantum-noise axion dark matter search at CAPP around 9.5 μeV. Phys. Rev. Lett., 130:091602, 2023.DOI
  55. [55]Andrew K. i, Saebyeok Ahn, Çağlar Kutlu, JinMyeong Kim, Byeong Rok Ko, Boris I. Ivanov, HeeSu Byun, Arjan F. van Loo, SeongTae Park, Junu Jeong, Ohjoon Kwon, Yasunobu Nakamura, Sergey V. Uchaikin, Jihoon Choi, Soohyun Lee, MyeongJae Lee, Yun Chang Shin, Jinsu Kim, Doyu Lee, Danho Ahn, SungJae Bae, Jiwon Lee, Younggeun Kim, Violeta Gkika, Ki Woong Lee, Seonjeong Oh, Taehyeon Seong, DongMin Kim, Woohyun Chung, Andrei Matlashov, SungWoo Youn, and Yannis K. Semertzidis. Axion dark matter search around 4.55 μeV with Dine-Fischler-Srednicki-Zhitnitskii sensitivity. Phys. Rev. Lett., 130:070102, Feb 2023.DOI
  56. [56]Andrew K. Yi, Saebyeok Ahn, Çağlar Kutlu, JinMyeong Kim, Byung Rok Ko, Boris I. Ivanov, HeeSu Byun, Arjan F. van Loo, SeongTae Park, Junu Jeong, Ohjoon Kwon, Yasunobu Nakamura, Sergey V. Uchaikin, Jihoon Choi, Soohyun Lee, MyeongJae Lee, Yun Chang Shin, Jinsu Kim, Doyu Lee, Danho Ahn, SungJae Bae, Jiwon Lee, Younggeun Kim, Violeta Gkika, Ki Woong Lee, Seonjeong Oh, Taehyeon Seong, DongMin Kim, Woohyun Chung, Andrei Matlashov, SungWoo Youn, and Yannis K. Semertzidis. Search for the Sagittarius tidal stream of axion dark matter around 4.55 μeV. Phys. Rev. D, 108:L021304, Jul 2023.DOI
  57. [57]Çağlar Kutlu, Soohyun Lee, Sergey V. Uchaikin, Saebyeok Ahn, Sungjae Bae, Junu Jeong, SungWoo Young, Arjan F. van Loo, Yasunobu Nakamura, Seonjeong Oh, and Yannis K. Semertzidis. Search for qcd axion dark matter around 24.5 μeV using an 8-cell microwave resonant cavity haloscope and a flux-driven Josephson parametric amplifier. In Proceedings of Science, - 41st International Conference on High Energy physics (ICHEP2022) - Astroparticle Physics and Cosmology, volume 414, 2023.DOI
  58. [58]Younggeun Kim, Junu Jeong, SungWoo Youn, SungJae Bae, Kiwoong Lee, Arjan F. van Loo, Yasunobu Nakamura, Seonjeong Oh, Taehyeon Seong, Sergey Uchaikin, Jihn E. Kim, and Yannis K. Semertzidis. Experimental search for invisible axions as a test of axion cosmology around 22 μeV. arXiv:2312.11003 [hep-ex], 2023.

Paper details

vlnc:W_7a5nsp39
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 0

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…
Contents