Monte Carlo-based secondary dose check platform for the world's first ultra-compact upright proton therapy system
Abstract
Objective. This work aimed to develop and validate an independent Monte Carlo (MC) secondary dose check of the RayStation treatment planning system (TPS) for the Mevion S250-FIT, the world's first ultra-compact upright proton therapy system. Approach. The treatment nozzle, including the energy modulation system of boron carbide and polycarbonate plates and the adaptive aperture, was modeled in the TOPAS (TOol for PArticle Simulation) MC code, and the beam model was commissioned against integrated depth dose curves, in-air spot profiles, and dose per monitor unit measured at 12 energy settings. Based on this model, an in-house converter, RayStation2TOPAS, recalculates the TPS plans by reconstructing each pencil beam spot, the aperture for each energy layer, and the seated patient's rotation. In total, 15 water phantom plans and 25 clinical plans were recalculated and compared with the TPS by the point dose, with a tolerance of 5%, and by three-dimensional gamma analysis at 3%/2mm, with a pass rate of at least 90%. Main results. The simulated proton range agreed with the measurement within 0.25mm, the spot full width at half maximum within 6.8%, and the dose per monitor unit within 1.3%. The point dose differed from the TPS by -0.76% to +1.98% in the water phantom plans and by -3.03% to +3.99% in the clinical plans, with gamma pass rates of 96.39% to 99.93% and 90.08% to 99.98%, so every plan passed both tests. Significance. To our knowledge, this is the first independent MC secondary dose check for the S250-FIT, and it complements measurement-based patient-specific quality assurance of its plans.
Introduction
The availability of proton therapy is limited by the size and cost of its facilities, much of which comes from the rotating gantry that delivers the beam to a supine patient from different angles ([26, 32]). Upright proton therapy replaces this rotating gantry with a fixed horizontal beamline and obtains the treatment angles by rotating the seated patient on a robotic positioner ([17] (Mackie et al 2021)). This design reduces both the footprint and the cost of the facility ([31]), and upright treatments on fixed beamlines are already in clinical use ([15, 16, 18, 13, 6, 7]). The Mevion S250-FIT applies this approach in an ultra-compact system that combines a superconducting synchrocyclotron with an upright patient positioner and an upright CT scanner. The first unit was installed in a converted linear accelerator vault at Stanford and began treating patients in June 2026 ([9]).
As in any proton therapy system, before treatment, the dose calculated by the treatment planning system (TPS) for each plan is verified as part of patient-specific quality assurance, traditionally by measurements in a water phantom, which take beam time and test the calculated dose only in water ([22]). An independent secondary dose check, which recalculates the plan in the patient geometry with a dose engine separate from that of the TPS, can replace these measurements ([23]). Monte Carlo (MC) simulation is suited to this check because it is more accurate than analytical pencil beam algorithms in heterogeneous tissue (Taylor et al 2017). For example, at the Maastro Proton Therapy Center, MC recalculation of more than 6000 plans over five years detected two planning errors that the measurements would have missed (Rinaldi et al [28]).
However, an MC secondary dose check depends on the delivery system, because its beam model is derived from the commissioning measurements of that system. Such checks have therefore been built for individual systems, including the Varian ProBeam (Aitkenhead et al 2020), the IBA Proteus ONE (Liu et al [19]), and the Mevion S250i, for which MC models have been developed with a GPU-based (Gajewski et al [14]) and a CPU-based code (Chiang et al 2021). The S250i is the gantry-mounted predecessor of the S250-FIT, but neither of its models describes the S250-FIT, whose energy modulation system replaces the 18 Lexan plates of the S250i with 10 boron carbide and 2 Lexan plates, which scatter the beam differently and give spots up to 14 % smaller (Charyyev et al [9]). At present, plans for the S250-FIT are calculated with the MC dose engine of RayStation, and to our knowledge, no independent MC dose check has been reported for this system.
To address this gap, this work presents an independent MC secondary dose check for the S250-FIT. First, the treatment nozzle was modeled in TOPAS, a CPU-based MC code (Perl et al [27], Faddegon et al [12]), and the beam model was commissioned against the integrated depth dose curves, in-air spot profiles, and absolute dose per monitor unit measured at 12 energy settings during the clinical commissioning of the system (Charyyev et al [9]). An in-house converter, RayStation2TOPAS, was then built on this commissioned model to recalculate RayStation plans. The converter reconstructs the delivery of each plan spot by spot, including the adaptive aperture, which can change shape between energy layers, and reproduces the upright treatment geometry, in which the beam is fixed and the seated patient is rotated and tilted. For each plan, the recalculated dose is compared with the RayStation dose at the point where RayStation reports the plan dose and by a three-dimensional gamma analysis. Finally, the check was validated on 15 water phantom plans and 25 clinical plans.
Materials and methods
Mevion S250-FIT Upright Proton Therapy System
The system investigated in this work is a Mevion S250-FIT Upright Proton Therapy System (Mevion Medical Systems, Littleton, MA, USA), installed at the Stanford Cancer Center (Stanford FIT Proton Therapy). It is the world’s first ultra-compact upright platform that delivers pencil beam scanning with a fixed horizontal beamline. Treatment angles are obtained by rotating the patient about the vertical axis on a six-degree-of-freedom upright patient positioner rather than rotating a massive gantry, and images are acquired by a diagnostic-quality upright fan-beam CT (Leo Cancer Care Inc., Middleton, WI, USA).
A compact superconducting synchrocyclotron, similar to that of its gantry-mounted predecessor, the S250i platform, accelerates protons to a fixed maximum energy of approximately 227 MeV. Nested – scanning coils steer the beam to each spot position at up to 10 mm/ms before it enters the treatment nozzle. The treatment nozzle contains transmission ionization chambers that monitor the beam, the energy modulation system (EMS), and the dynamic adaptive aperture (AA). The EMS has 12 plates, 2 polycarbonate (Lexan) and 10 boron carbide (BC), unlike the S250i, which uses 18 Lexan plates. The plates vary the beam energy over the range from about 227 MeV to 9 MeV with a specific plate configuration for each energy layer, and the switching time between energy layers is 50 ms to 500 ms. The AA sits downstream of the EMS and trims the beam laterally. It is a nickel mini-multileaf collimator with seven leaf pairs, each 10 cm thick along the beam, carried on a translating stage that can position it at any point in the field. Its five central pairs are 0.5 cm wide, and the two outer pairs, which act as jaws, are 2 cm wide. The AA of the S250-FIT follows the same design as its predecessor. After the treatment nozzle, the beam passes through the exit window, which is a 1.59 mm Lexan plate, before being delivered to a patient.
Monte Carlo Model of the Treatment Nozzle
To develop a secondary dose check platform based on MC simulation, the treatment nozzle was modeled within an MC environment (Figure 1). All MC simulations were performed with TOPAS version 3.9 built on Geant4 10.07.p03 (Agostinelli et al [1], Allison et al [3, 4]). Electromagnetic interactions, including ionization energy loss and multiple Coulomb scattering, were modeled with the g4em-standard_opt4 constructor. Inelastic nuclear interactions were modeled with g4h-phy_QGSP_BIC_HP, and nuclear elastic scattering with g4h-elastic_HP. Production thresholds were set to 0.05 mm for protons and 0.5 mm for electrons, positrons, and photons.

Figure 1. The TOPAS model of the treatment nozzle, drawn by TOPAS from the input the converter wrote for one energy layer of the static plan of the 1000 cm target at 10 cm depth (§ 2.5), with the snout at 200 mm. The arrow shows the direction of the beam, and the lines point from the labels to the components. The EMS plates (red) in this layer are in the beam, while the other plates are retracted, all inside the EMS air box. The AA (yellow) is a nickel cylinder with the opening of the plan (gray), followed by the exit window (green) and the water phantom (blue), whose surface is 50 mm upstream of the isocenter (cross). The virtual source, 1870 mm upstream of the isocenter, is outside the view.
The proton beam was modeled as a TOPAS emittance source with a bivariate Gaussian phase-space distribution on a plane 1870 mm upstream of the isocenter. This distance is the virtual source-to-axis distance (SAD) of the scanning system, taken from the focal length of 187 cm measured during the clinical commissioning of the system [9]. The phase space can be specified with five parameters: the mean energy, relative energy spread, the spatial beam spread distribution (beam spot size), , the angular spread distribution (beam divergence), , and the coefficient of correlation, [8, 10]. The same values were used for both transverse axes, consistent with the rotational symmetry of the measured in-air spots. Spot positioning was modeled as a rotation about the virtual source point as
where is the planned spot position at the isocenter plane and mm. The EMS was modeled as 12 rectangular plates in air. Plate materials were assigned as the Geant4 reference materials for polycarbonate and boron carbide. The thickness of each boron carbide plate was set from its water-equivalent thickness in the system’s range-shifter table. Each plate was placed either along the beam axis or laterally displaced from the field, depending on its state within the energy configuration.
For plans with open collimation, the AA leaves were retracted outside the scanned field. For static and dynamic collimation, the aperture was modeled as a nickel cylinder with a radius of 150 mm (TsAperture) containing a straight-walled polygonal opening taken from the treatment plan. The representation for the collimated plans follows from what the treatment plan contains. The plan does not store the positions of the individual leaves; it stores the aperture opening as a polygon, a single one for static collimation and one per energy layer for dynamic collimation. Modeling the opening directly from this polygon reproduces the aperture as planned, without assuming how the machine converts the outline into leaf positions. The outer shape of the collimator does not affect the dose. Since 100 mm of nickel stops protons of every deliverable energy, only the opening shapes the field, and the 150 mm radius covers the entire scanned field at the aperture plane. Downstream of the AA, the exit window was modeled as a 1.59 mm polycarbonate slab.
Beam Model Commissioning and Benchmarking
The TOPAS model was commissioned and tested against the clinical commissioning measurements of the S250-FIT. The same measurements were used to commission the beam model of the TPS, RayStation v2024A (RaySearch Laboratories, Stockholm, Sweden), and were exported from it. The exported file contains integrated depth dose (IDD) curves for pristine Bragg peaks, in-air spot profiles, and absolute dose per monitor unit (MU), together with the same quantities computed by the RayStation Monte Carlo (RS MC) dose engine. All three quantities were compared at the 12 energy settings of the clinical commissioning, from 49.32 MeV to 227.09 MeV. The setting at 227.09 MeV is the maximum energy, delivered directly from the synchrocyclotron with no plate in the beam, and the other 11 settings were chosen so that each of the 12 EMS plates is in the beam in at least one of them. Each measurement was reproduced in TOPAS using the nozzle model from § 2.2. The mean energy and the energy spread were tuned against the IDD curves. The spot size, divergence, and correlation were obtained by a Bayesian fit to the RS MC in-air spot profiles at 227.09 MeV.
Integrated Depth Dose Curves
A single spot was transported through the nozzle along the central axis, with the snout at 200 mm, into a 600 mm water cube, whose surface was at the isocenter. The DoseToWater quantity was scored in a water-cylinder detector coaxial with the beam, 600 mm long, divided into 1000 depth bins of 0.6 mm each. Its diameter, 81.6 mm, was set equal to that of the detector used in the measurement. The simulated IDDs were compared with the measured and RS MC curves after normalization to their maxima. For the comparison, the range , the depth distal to the peak at which the dose falls to 80 % of the maximum, the peak width , the distance between the proximal and distal 80 % points, and the distal fall-off , the distance between the distal 80 % and 20 % points, were calculated.
In-Air Spot Profiles
A single spot was transported through the nozzle along the central axis, with the snout at 250 mm, into air. The Fluence quantity, restricted to protons, was scored in an air box centered on the isocenter, 400 mm long and divided into 1 mm bins along the beam and 0.5 mm bins laterally, the same spacing as the measured profiles. Its lateral half-width, between 25 and 85 mm, was set for each energy setting to ensure the box contained the entire spot. Profiles were taken in the five measurement planes, located at the isocenter and 100 mm and 200 mm upstream and downstream of it. In each plane, the profile along was the mean of the two 0.5 mm rows of the 1 mm slab on either side of the spot center, which form a line 1 mm wide through the center, and the profile along was taken likewise. The full width at half maximum (FWHM) of each simulated profile was calculated from a two-Gaussian fit to the profile, and that of each measured and RS MC profile by linear interpolation at half of its maximum.
Absolute Dose per MU
The dose per MU was simulated under the conditions of the measurement. A uniform field of 100 × 100 mm, made of 41 × 41 spots of equal weight with a spacing of 2.5 mm, was delivered with the snout at 200 mm into a water phantom of 201 × 201 × 400 mm, divided into 3 mm bins laterally and 1 mm bins in depth. The water surface was placed upstream of the isocenter by the reference depth, 10 mm for energies up to 114.52 MeV and 50 mm for higher energies, so that the reference point was at the isocenter. The dose at the reference point was taken as the mean over the voxels within 10 mm of the central axis, matching the 20 mm diameter detector given in the commissioning data, and was interpolated linearly between the two depth bins around the reference depth.
TOPAS gives the dose per simulated proton, , whereas the measurement gives the dose per MU, . The two are related by the number of protons per MU, :
The monitor chambers of the S250-FIT are upstream of the EMS ([9]), so they measure the protons before the plates slow them down, and is the same at every energy setting. A single value was therefore determined from the measurements, such that the simulated equals the measured on average over the 12 energy settings. With this value, the simulated at each energy setting was compared with the measured and RS MC values. The same converts the MU of the plans into protons in equation (4).
RayStation2TOPAS
RayStation2TOPAS is an in-house converter that recalculates RayStation plans for the S250-FIT in TOPAS as an independent MC secondary dose check. The converter first parses the Digital Imaging and Communications in Medicine (DICOM) export of a plan into a list of spots, and then reconstructs the delivery and its absolute dose from that list in the nozzle model of § 2.2.
Plan Parsing and Delivery Reconstruction
In the first stage, the converter parses the RT Ion Plan. The plan describes the delivery of each beam as a sequence of control points. The control points mark the progress of the delivery by the MU delivered so far, and each one gives the delivery parameters from that point on. Each energy layer is described by two control points, one at its start and one at its end. The start control point of each layer holds the EMS setting of the layer and the position and MU of every spot (“meterset weight” in the plan). The first control point of a beam additionally holds the parameters of the whole beam: the isocenter, the snout position, and the rotation of the upright patient positioner (“patient support angle” in the plan), together with the backrest tilt (“table-top pitch angle” in the plan).
In the second stage, the converter turns the spot list into TOPAS runs. The converter groups the spots of each beam by EMS setting and simulates each group as a separate run. TOPAS steps through the spots in order, and at each step it aims the source at the spot and simulates a number of histories proportional to the spot’s MU. The plan’s total budget determines the number of histories per spot. The converter divides this budget among the runs in proportion to the MU of each run, the sum of the MU of its spots, so that run receives histories, at least one, and then divides among the spots of the run:
where is the MU of spot and the sum runs over the spots of run . Rounding gives whole numbers of histories, and the lower limit of one history, for each run and for each spot, keeps every energy layer and every spot in the simulation.
Each run yields the dose for each simulated history. The converter obtains the absolute dose of the plan by scaling this dose to the number of protons the run represents and summing over the runs:
where is the dose of run and the number of histories simulated in it, so that is the dose per simulated proton. is the MU of run in one fraction, the sum of the MU of its spots, and is the number of protons it delivers, with the number of protons per MU (§ 2.3.3). The sum over the runs therefore gives the dose of one fraction. It is multiplied by the number of planned fractions, , because the plan gives the MU of a single fraction, whereas the exported TPS dose is for the entire treatment.
Aperture Reconstruction
How the aperture is modeled depends on the beam’s collimation mode, which the converter determines from the plan. Open plans carry no aperture, and the AA is modeled with all leaves retracted (§ 2.2). Static plans carry a single opening in the standard aperture block of the beam (IonBlockSequence), stored as a polygon, a list of vertices that the DICOM standard defines as projected onto the isocenter plane ([25]), and the same opening is used for every energy layer. Dynamic plans reshape the AA between layers. RayStation stores the opening of each layer as a polygon of the same form, but in private tags of the layer’s start control point rather than in standard DICOM attributes. The converter reads these vertices and, for both static and dynamic plans, takes the material and thickness of the aperture from the aperture block.
The plan defines each opening at the isocenter plane, whereas the aperture sits upstream of it, so each polygon is scaled back along the diverging beam. Because the aperture walls are straight and the beam diverges, the downstream face of the aperture limits the field, and the polygon is scaled to that face by
where is the distance of the downstream face from the isocenter ( for a snout at 200 mm).
Validation Plans
The converter was tested on 15 water phantom plans and 25 clinical plans. In the water phantom plans, five targets, a target at depths of 5, 10, and 20 cm and a target at depths of 10 and 20 cm, were each planned with open, static, and dynamic collimation. The plans used between 10 and 18 energy layers. Hence, the target depths and sizes tested the EMS model over different sets of plate configurations, and the three collimation modes tested the aperture reconstruction (§ 2.4.2). Each plan consisted of a single field with a patient support angle of , the snout at 200 mm, the isocenter at 50 mm depth, and 30 fractions.
The clinical plans were the 25 plans of the first cohort of patients treated on the S250-FIT at Stanford. The plans covered 10 brain, 5 head and neck, 6 thoracic, 2 spine, and 2 bone treatment sites. Each patient was scanned seated in the treatment position on the upright fan-beam CT, with a slice thickness of 2 mm, and was planned in RayStation v2024A with the MC proton dose engine, a statistical uncertainty of 0.5 %, and a dose grid of 2 mm. The plans had between 1 and 6 fields and backrest tilts between 0 and . Static apertures were used in 14 plans and dynamic apertures in 11 plans, and the plans were delivered in 1 to 30 fractions. Unlike the water phantom plans, the clinical plans contained heterogeneous tissue, several fields at different patient support angles, and immobilization devices, so they also tested the CT conversion and the patient geometry (§ 2.6).
Dose Calculation
Water phantom plans were simulated in a water box of mm, centered on the beam axis and divided into voxels of mm, with the 1 mm side along the beam, which is the voxel size of the RayStation dose grid. The box served as both the phantom and the scoring volume, so its front face was placed at the phantom surface given by the plan. The DoseToWater quantity was scored in each voxel.
Clinical plans were simulated in the planning CT, loaded in TOPAS as a DICOM patient (TsDicomPatient) and positioned so that the plan isocenter coincided with the isocenter of the nozzle model. On the S250-FIT, the beam is fixed, and the patient rotates, so in the simulation the patient was kept fixed and the whole nozzle, including the source, was rotated by the patient support angle of the beam about the vertical axis through the isocenter. When the plan included a backrest tilt, the patient was tilted by that angle about the left–right axis through the isocenter.
In RayStation, the patient geometry used for the dose calculation consists of the volume inside the External region of interest (ROI) and the ROIs of type Support and Fixation, and the rest of the volume is treated as vacuum. The converter applied the same rule, and voxels outside these structures were set to air ( HU). Where the plan overrides the material of a structure, TOPAS was given the RayStation material itself, with its elemental composition, mass density, and mean excitation energy. Where the plan gives only a density, the structure was given the CT number that yields that density in TOPAS. As in RayStation, the override of a Support or Fixation ROI applied to the whole ROI, whereas the override of any other ROI applied only to its part inside the External ROI. The CT was then cropped in-plane to a box around the patient and the support and fixation structures. Figure 2(a) and (Figure 2b) show these steps on the CT of the IROC Houston anthropomorphic head-and-neck phantom.
Figure 2. The patient pipeline, shown on the CT of the IROC Houston anthropomorphic head-and-neck phantom, for one beam of its plan, on the axial slice through the isocenter. (a) Planning CT with the External outline and the box to which the CT was cropped. (b) The CT as TOPAS reads it, with the voxels outside the External set to air, the direction of the beam, and the snout face of the nozzle. (c) RayStation and (d) TOPAS dose of the beam, as physical dose, with TOPAS resampled onto the RayStation grid and set to zero outside the External. Dose below 10 % of the maximum is not shown. (e) Gamma index at 3 %/2 mm. Because a single beam is shown, the dose-difference criterion and the low-dose threshold were percentages of the maximum of the RayStation dose of the beam, not of the dose at the isocenter as for a plan. The cross marks the isocenter.
Each beam was simulated on its own copy of the CT. When the CT extended into the space of the nozzle, rows and columns containing only air were removed from the upstream side of the CT for that beam, so that no voxel values changed. The TOPAS overlap check was enabled for every simulation, and a simulation stopped if the patient still overlapped the nozzle.
CT numbers were converted to materials and mass densities with the stratified scheme of Schneider et al. [29], as implemented in TOPAS, with density correction factors derived from the clinical calibration curve of CT number to stopping-power ratio. The dose was scored in the voxels of the cropped CT itself, so the simulated dose had the same voxels as the CT. In each voxel, TOPAS converted the deposited energy to dose to water with the ratio of the stopping power of water to that of the voxel material (the DoseToWater quantity), and the result was saved as a DICOM RT Dose file, the same format as the RayStation dose. The dose of each run was converted to absolute dose with equation (4), and the doses were summed on a common grid, since the upstream trim could give each beam a different grid.
Dose Comparison
The simulated dose was compared with the RayStation dose on the RayStation dose grid. It was transferred to that grid by volume-weighted resampling, in which each RayStation voxel received the mean of the simulated dose over its volume, weighted by its overlap with each simulation voxel.
RayStation voxels covered by the simulation to less than 99 % of their volume were excluded. For the clinical plans, the simulated dose was set to zero outside the volume in which RayStation computes dose, the External ROI together with the ROIs of type Support and Fixation. A voxel was kept when any part of it was within this volume.
Two tests were applied, following the structure of a calculation-based secondary dose check (Zhu et al 2021), and both used the absolute dose. The first compared the dose at the point where RayStation reports the plan dose. This is the dose reference point of the plan, if it has one; otherwise, the beam dose specification point shared by its beams. In the water phantom plans, this point lies on the beam axis at the center of the spread-out Bragg peak (SOBP), and in the clinical plans at the isocenter. The mean over the voxels centered on it was compared, and a plan passed when the simulated dose agreed with the RayStation dose within 5 %.
The second test compared the dose distributions with the three-dimensional gamma index (Low et al 1998), computed with PyMedPhys [5] at 3 %/2 mm, with a low-dose threshold of 10 %, an interpolation fraction of 10, and a maximum gamma of 2. Neither distribution was rescaled, so the gamma test also detected an error in the overall dose level. Global normalization was used [24], so the dose-difference criterion and the low-dose threshold were expressed as percentages of a single reference dose across the whole plan. The reference dose was the RayStation dose, averaged over voxels, at the same point. Only a plan with neither point would use the maximum of the RayStation dose. A plan passed the gamma test when at least 90 % of the evaluated voxels had a gamma index of 1 or less. Figure 2(c)–(Figure 2e) shows the two doses and the gamma index for one beam of the head-and-neck phantom plan.
Results
Beam Model
Table 1 lists the source parameters obtained by tuning the model against the IDDs and the in-air spot profiles (§ 2.3). Table 2 and Figure 3 compare the RS MC, simulated, and measured IDDs at 12 energy settings, from 49.32 MeV to 227.09 MeV. The simulated differed from the measurement by at most 0.25 mm, at 154.40 MeV, with a mean absolute difference of 0.11 mm, while the RS MC differed by at most 0.15 mm, at 59.11 MeV, with a mean absolute difference of 0.07 mm. The differences in the peak width ranged from −0.19 mm to +0.27 mm for TOPAS and from −0.31 mm to −0.10 mm for RS MC, and those in the distal fall-off from +0.21 mm to +0.40 mm and from +0.03 mm to +0.20 mm.
| Parameter | Symbol | Value |
| Mean energy | 228.12 MeV | |
| Relative energy spread | 0.275 % | |
| Spot size | 4.29 mm | |
| Divergence | 2.65 mrad | |
| Correlation | −0.816 |
Table 1. Source parameters of the beam model, identical for both transverse axes, defined at the virtual source plane 1870 mm upstream of the isocenter.
| Energy (MeV) | Measured (mm) | (mm) RS MC | (mm) TOPAS | (mm) RS MC | (mm) TOPAS | (mm) RS MC | (mm) TOPAS |
| 49.32 | 22.88 | +0.02 | +0.05 | −0.23 | −0.19 | +0.15 | +0.29 |
| 59.11 | 31.04 | −0.15 | +0.09 | −0.30 | −0.17 | +0.18 | +0.33 |
| 84.89 | 57.87 | +0.03 | +0.15 | −0.26 | −0.10 | +0.14 | +0.30 |
| 90.03 | 64.08 | −0.11 | +0.14 | −0.17 | +0.02 | +0.17 | +0.32 |
| 114.52 | 97.39 | −0.06 | +0.16 | −0.19 | −0.05 | +0.16 | +0.32 |
| 135.57 | 130.72 | +0.08 | +0.24 | −0.10 | +0.03 | +0.17 | +0.35 |
| 154.40 | 164.10 | +0.05 | +0.25 | −0.22 | +0.06 | +0.03 | +0.21 |
| 168.51 | 191.34 | +0.09 | +0.08 | −0.16 | +0.06 | +0.15 | +0.36 |
| 171.62 | 197.56 | −0.06 | +0.06 | −0.31 | −0.06 | +0.07 | +0.26 |
| 199.01 | 255.99 | −0.01 | 0.00 | −0.15 | +0.17 | +0.06 | +0.23 |
| 224.58 | 316.45 | +0.03 | −0.04 | −0.16 | +0.27 | +0.11 | +0.28 |
| 227.09 | 322.71 | −0.14 | −0.11 | −0.14 | +0.26 | +0.20 | +0.40 |
Table 2. Measured range of the IDDs at the 12 energy settings, and the differences of RS MC and TOPAS from the measurement in the range , the peak width , and the distal fall-off .

Figure 3. Integrated depth dose curves at the 12 energy settings. (a) Simulated, measured, and RS MC curves, each normalized to its maximum. (b) to (d) Differences from the measurement in , , and for RS MC and TOPAS.
Table 3 and Figure 4 compare the RS MC, simulated, and measured spot FWHM at the same 12 settings. Over all five planes and both axes, 120 comparisons in total, the TOPAS FWHM differed from the measured FWHM by at most 2.49 mm, at 59.11 MeV, with a mean absolute difference of 1.05 mm, while the RS MC FWHM differed by at most 1.75 mm, at 171.62 MeV, with a mean absolute difference of 0.56 mm. The largest relative differences were 6.8 % for TOPAS and 6.2% for RS MC.
| Energy (MeV) | FWHM (mm) Measured | FWHM (mm) RS MC | FWHM (mm) TOPAS | FWHM (mm) Measured | FWHM (mm) RS MC | FWHM (mm) TOPAS |
| 49.32 | 45.75 | −0.57 | −1.15 | 45.73 | −0.52 | −1.89 |
| 59.11 | 42.35 | −0.57 | −1.70 | 42.46 | −0.64 | −1.32 |
| 84.89 | 36.16 | −0.48 | −1.26 | 36.56 | −0.85 | −1.45 |
| 90.03 | 34.86 | −0.39 | −1.26 | 35.10 | −0.58 | −1.14 |
| 114.52 | 29.99 | −0.56 | −1.02 | 30.21 | −0.76 | −1.25 |
| 135.57 | 26.10 | −0.69 | −1.16 | 26.25 | −0.86 | −1.24 |
| 154.40 | 22.78 | −0.94 | −1.16 | 22.85 | −1.03 | −1.37 |
| 168.51 | 20.25 | −0.82 | −0.88 | 20.47 | −1.03 | −1.15 |
| 171.62 | 19.61 | −0.95 | −1.05 | 19.69 | −1.04 | −1.16 |
| 199.01 | 14.51 | −0.45 | −0.83 | 14.53 | −0.48 | −0.79 |
| 224.58 | 10.00 | +0.02 | −0.40 | 10.11 | −0.10 | −0.50 |
| 227.09 | 8.99 | +0.12 | −0.47 | 8.93 | +0.16 | −0.43 |
Table 3. Measured FWHM of the in-air spots at the isocenter plane on both axes at the 12 energy settings, and the differences of RS MC and TOPAS from the measurement.

Figure 4. In-air spot profiles and widths. (a) to (c) Simulated, measured, and RS MC profiles along at the isocenter plane at 49.32 MeV, 135.57 MeV, and 227.09 MeV. The simulated profile is the two-Gaussian fit to the proton fluence along a line through the spot center, normalized to its maximum; the measured and RS MC profiles are normalized to their maxima. Each profile is centered on the midpoint of its FWHM. (d, e) FWHM at the isocenter plane along and at the 12 energy settings. (f) FWHM difference from the measurement, RS MC or TOPAS minus measured, over all five planes and both axes, 120 comparisons each.
Table 4 and Figure 5 compare the RS MC, simulated, and measured dose per MU at the same 12 settings. The number of protons per MU was , and the numbers that reproduced the measurement at each energy setting ranged from to . The simulated dose per MU differed from the measurement by −1.3% to +1.1%, with a mean absolute difference of 0.7%, while the RS MC dose per MU differed by −0.4% to +0.9%, with a mean absolute difference of 0.5%.
| Energy (MeV) | Depth (mm) | Measured (cGy/MU) | RS MC (cGy/MU) | TOPAS (cGy/MU) | Difference from measured (%): RS MC | Difference from measured (%): TOPAS |
| 49.32 | 10 | 0.1090 | 0.1086 | 0.10862 | −0.4 | −0.3 |
| 59.11 | 10 | 0.0902 | 0.0910 | 0.09059 | +0.9 | +0.4 |
| 84.89 | 10 | 0.0697 | 0.0699 | 0.07034 | +0.3 | +0.9 |
| 90.03 | 10 | 0.0672 | 0.0677 | 0.06791 | +0.7 | +1.1 |
| 114.52 | 10 | 0.0601 | 0.0605 | 0.06048 | +0.7 | +0.6 |
| 135.57 | 50 | 0.0659 | 0.0665 | 0.06649 | +0.9 | +0.9 |
| 154.40 | 50 | 0.0621 | 0.0624 | 0.06192 | +0.5 | −0.3 |
| 168.51 | 50 | 0.0603 | 0.0606 | 0.05989 | +0.5 | −0.7 |
| 171.62 | 50 | 0.0603 | 0.0604 | 0.06010 | +0.2 | −0.3 |
| 199.01 | 50 | 0.0589 | 0.0593 | 0.05891 | +0.7 | 0.0 |
| 224.58 | 50 | 0.0593 | 0.0594 | 0.05851 | +0.2 | −1.3 |
| 227.09 | 50 | 0.0595 | 0.0595 | 0.05888 | 0.0 | −1.0 |
Table 4. Dose per MU at the calibration condition: 100 × 100 mm field, 2.5 mm spot spacing, mean over a 20 mm diameter region at the reference depth, with 7.896 × 10^7 protons per MU. Measured and RS MC values are from the clinical commissioning data.

Figure 5. Dose per MU at the calibration condition. (a) Simulated, measured, and RS MC dose per MU at the 12 energy settings. The dotted line separates the settings with a reference depth of 10 mm from those with 50 mm. (b) Difference from the measurement for RS MC and TOPAS. The mean TOPAS difference is zero by construction, because the number of protons per MU was set from these 12 settings (§ 2.3.3). Error bars are the statistical uncertainty of the simulated dose.
Water Phantom Plans
Table Table 5 lists the point dose and gamma results of the 15 water phantom plans, and figure Figure 6 shows their depth doses and lateral profiles. The simulated point dose differed from the RayStation dose by −0.76 to +1.98%, with a mean of +0.64%, and all 15 plans passed the point dose test. The largest difference, +1.98%, was found for the static plan of the 1000 cm target at 10 cm. The gamma pass rate ranged from 96.39 to 99.93%, with a mean of 98.94%, and all 15 plans passed the gamma test.
| Target (cm) | Depth (cm) | Collimation | Layers | Spots | RayStation (cGy) | TOPAS (cGy) | Difference (%) | Gamma (%) |
| 125 | 5 | Open | 10 | 900 | 2775.84 | 2805.05 | +1.05 | 99.47 |
| 125 | 5 | Static | 10 | 686 | 2786.93 | 2809.88 | +0.82 | 99.40 |
| 125 | 5 | Dynamic | 14 | 503 | 2553.02 | 2533.69 | −0.76 | 99.93 |
| 125 | 10 | Open | 10 | 937 | 2774.11 | 2777.39 | +0.12 | 99.51 |
| 125 | 10 | Static | 10 | 789 | 2802.71 | 2817.93 | +0.54 | 99.39 |
| 125 | 10 | Dynamic | 10 | 304 | 2825.09 | 2834.39 | +0.33 | 99.71 |
| 125 | 20 | Open | 10 | 951 | 2763.53 | 2773.12 | +0.35 | 99.91 |
| 125 | 20 | Static | 10 | 935 | 2772.48 | 2768.98 | −0.13 | 99.89 |
| 125 | 20 | Dynamic | 10 | 454 | 2844.37 | 2847.93 | +0.12 | 99.82 |
| 1000 | 10 | Open | 18 | 3729 | 2773.86 | 2809.89 | +1.30 | 97.23 |
| 1000 | 10 | Static | 18 | 3475 | 2759.83 | 2814.53 | +1.98 | 96.39 |
| 1000 | 10 | Dynamic | 18 | 1974 | 2840.60 | 2876.37 | +1.26 | 96.79 |
| 1000 | 20 | Open | 18 | 3658 | 2745.78 | 2774.40 | +1.04 | 99.04 |
| 1000 | 20 | Static | 18 | 3643 | 2761.01 | 2779.33 | +0.66 | 98.98 |
| 1000 | 20 | Dynamic | 18 | 2951 | 2755.56 | 2779.28 | +0.86 | 98.68 |
Table 5. Point dose and gamma results of the water phantom plans. Doses are physical doses summed over the 30 fractions, at the point where RayStation reports the plan dose. The gamma pass rate is for 3 %/2 mm on absolute dose with global normalization.

Figure 6. RayStation and TOPAS dose of the 15 water phantom plans, as physical dose summed over the 30 fractions on the RayStation dose grid, one panel per plan. Columns are the targets and rows the collimation modes. (a) Depth dose along the beam axis. (b) Lateral profile along at the depth of the point where RayStation reports the plan dose, through the voxel nearest to that point. Solid lines are TOPAS and dashed lines RayStation.
For the 125 cm target, the point dose difference ranged from to % (mean %) and the gamma pass rate from 99.39 to 99.93 % (mean 99.67 %). For the 1000 cm target, the point dose difference ranged from to % (mean %) and the gamma pass rate from 96.39 to 99.04 % (mean 97.85 %). The mean point dose difference was % for the open plans, % for the static plans, and % for the dynamic plans, and the mean gamma pass rate was 99.03, 98.81, and 98.99 %, respectively.
Clinical Plans
Table 6 compares the simulated and RayStation doses of the 25 clinical plans. The point dose differed from the RayStation dose by % to %, with a mean of %, and all 25 plans passed the point dose test. The largest difference, %, was for plan 10. The gamma pass rate ranged from 90.08 % to 99.98 %, and all 25 plans passed the gamma test.
| Plan | Site | Aperture | Fields | Fractions | RayStation | TOPAS | Difference (%) | Gamma (%) | Time (min) |
| 1 | Neck | Static | 2 | 15 | 2079.9 | 2067.5 | −0.60 | 99.20 | 115 |
| 2 | Mandible | Static | 4 | 30 | 5555.3 | 5535.0 | −0.37 | 99.64 | 118 |
| 3 | Thorax | Static | 5 | 9 | 2604.2 | 2648.2 | +1.69 | 99.30 | 169 |
| 4 | Skull base | Static | 4 | 5 | 1906.4 | 1890.8 | −0.82 | 99.50 | 202 |
| 5 | Brain | Static | 4 | 30 | 6136.7 | 6208.2 | +1.17 | 99.63 | 166 |
| 6 | Brain | Static | 3 | 6 | 1015.0 | 993.7 | −2.09 | 98.70 | 150 |
| 7 | Bone | Static | 2 | 1 | 744.6 | 756.6 | +1.60 | 97.81 | 101 |
| 8 | Brain | Static | 4 | 30 | 4977.8 | 5030.5 | +1.06 | 99.98 | 249 |
| 9 | Brain | Dynamic | 3 | 20 | 3345.6 | 3414.2 | +2.05 | 99.97 | 123 |
| 10 | Thorax | Dynamic | 3 | 15 | 5617.4 | 5841.3 | +3.99 | 94.08 | 120 |
| 11 | Chest wall | Dynamic | 3 | 5 | 2843.8 | 2866.5 | +0.80 | 97.41 | 136 |
| 12 | Brain | Dynamic | 4 | 17 | 3558.8 | 3666.6 | +3.03 | 90.99 | 173 |
| 13 | Spine | Static | 2 | 27 | 4502.5 | 4516.0 | +0.30 | 95.75 | 87 |
| 14 | Brain | Dynamic | 4 | 17 | 2909.1 | 2863.1 | −1.58 | 99.75 | 157 |
| 15 | Thorax | Static | 3 | 8 | 1584.4 | 1602.5 | +1.14 | 99.60 | 111 |
| 16 | Head and neck | Dynamic | 4 | 25 | 4680.0 | 4789.5 | +2.34 | 92.08 | 179 |
| 17 | Head and neck | Dynamic | 4 | 19 | 3551.3 | 3443.7 | −3.03 | 93.01 | 200 |
| 18 | Brain | Dynamic | 4 | 30 | 6443.0 | 6349.9 | −1.45 | 99.53 | 129 |
| 19 | Lung | Static | 1 | 5 | 1718.4 | 1754.8 | +2.12 | 99.36 | 55 |
| 20 | Lung | Dynamic | 3 | 4 | 4420.7 | 4321.5 | −2.24 | 99.42 | 87 |
| 21 | Rib | Static | 2 | 5 | 3374.2 | 3428.8 | +1.62 | 99.90 | 68 |
| 22 | Brain | Dynamic | 6 | 30 | 6187.6 | 6383.3 | +3.16 | 97.98 | 187 |
| 23 | Spine | Static | 2 | 1 | 178.0 | 180.6 | +1.46 | 90.08 | 92 |
| 24 | Brain | Dynamic | 5 | 10 | 1662.5 | 1661.9 | −0.03 | 99.71 | 190 |
| 25 | Head and neck | Static | 3 | 20 | 3809.8 | 3828.8 | +0.50 | 99.63 | 116 |
Table 6. Point dose and gamma results of the clinical plans. Doses are physical doses for the whole plan. Differences are simulation minus RayStation, relative to RayStation, computed before rounding. Time is the time of the whole check of the plan, including the preparation of the simulation, on a workstation with two Intel Xeon Gold 6248R processors (48 cores, 3.0 GHz) and 376 GB of memory.
For the 14 plans with static apertures, the point dose difference ranged from % to % and the gamma pass rate from 90.08 % to 99.98 %. For the 11 plans with dynamic apertures, the point dose difference ranged from % to % and the gamma pass rate from 90.99 % to 99.97 %.
Discussion
This work developed an independent MC secondary dose check for the Mevion S250-FIT. The TOPAS model of the nozzle agreed with the IDDs, in-air spot profiles, and dose per MU measured at all 12 energy settings during the clinical commissioning. Built on this model, RayStation2TOPAS recalculated 15 water phantom plans and 25 clinical plans. The water phantom plans covered open, static, and dynamic collimation, and the clinical plans added heterogeneous patient anatomy and the upright treatment geometry. Every plan passed both the point dose and the gamma test, so the check can be used for patient-specific quality assurance of the S250-FIT, as an independent MC check has been used for five years for the S250i at the Maastro Proton Therapy Center [28]. The remaining differences between TOPAS, the measurements, and RayStation, and the limitations of the check, are discussed below.
The range of the simulated IDDs agreed with the measurement within 0.25 mm at all 12 energy settings. The spot FWHM agreed with the measurement within 6.8 % in all 120 comparisons, and that of the RS MC model of the clinical TPS within 6.2 %. This agreement is similar to that reported for the TOPAS model of the S250i, whose IDDs agreed with the measurements within 1 mm and whose spot sizes agreed within 10 % (Chiang et al 2021). The simulated dose per MU agreed with the measurement within 1.3 % at every energy setting, and that of RS MC within 0.9 %.
The water phantom plans tested the beam model and the reconstruction of the delivery, including the adaptive aperture, in a homogeneous medium. In these plans, open, static, and dynamic collimation agreed with RayStation with nearly the same mean gamma pass rate, 99.03 %, 98.81 %, and 98.99 %, respectively. The clinical plans added the patient anatomy and the upright geometry to the check. The planning CT, with the material overrides of RayStation, was converted to materials and mass densities, and each field was simulated with the rotation of the seated patient. With these additional steps, the point dose differences spread more widely than in the water phantom plans, from −3.03 to +3.99 %. The spread may also come from the reporting point. In the water phantom plans, this point was the center of the SOBP, where the dose is uniform, whereas in the clinical plans it was the isocenter, where the dose need not be uniform. Even the largest difference was within the 5 % tolerance, and every plan passed the gamma test.
The check has a few limitations. First, it verifies the dose calculation of RayStation, but not the transfer of the plan to the treatment machine or its delivery. Reconstructing the delivered dose from the machine log files, as done at other centers [22, 28], would extend the check to these steps. Second, the shape of the adaptive aperture is taken from the outline planned in RayStation, without modeling how the machine converts this outline into leaf positions (section 2.4.2). All 25 clinical plans passed the check with this approach, but its long-term effect is not yet known and should be monitored as more plans are checked. Finally, checking a clinical plan took 55 to 249 min on a CPU workstation. This time would be long for a center that starts several new patients per day, but one or two new patients per week start treatment on the S250-FIT at Stanford, so each plan can be checked before its first treatment fraction. Most of this time was spent preparing the plan for the simulation rather than in TOPAS itself, so optimizing this preparation would shorten the check in future applications.
Conclusion
An independent MC secondary dose check based on TOPAS was developed and validated for the Mevion S250-FIT. Its model of the nozzle agreed with the IDDs, in-air spot profiles, and dose per MU measured during the clinical commissioning at all 12 energy settings. Built on this model, RayStation2TOPAS recalculates RayStation plans, including the adaptive aperture of each energy layer and the upright treatment geometry. In the 15 water phantom plans and 25 clinical plans, the recalculated point dose differed from the RayStation dose by at most 2.0 % and 4.0 %, the gamma pass rate at 3 %/2 mm was at least 96.39 % and 90.08 %, and every plan passed both tests. With 55 to 249 min per clinical plan on a CPU workstation, the check can be performed for each patient before the first treatment fraction.
References
- [1]Agostinelli S, Allison J, Amako K, Apostolakis J, Araujo H, Arce P, Asai M, Axen D, Banerjee S, Barrand G et al 2003 Geant4—a simulation toolkit Nuclear Instruments and Methods in Physics Research Section A 506 250–303.DOI
- [3]Allison J, Amako K, Apostolakis J, Araujo H, Arce Dubois P, Asai M, Barrand G, Capra R, Chauvie S, Chytracek R et al 2006 Geant4 developments and applications IEEE Transactions on Nuclear Science 53 270–278.
- [4]Allison J, Amako K, Apostolakis J, Arce P, Asai M, Aso T, Bagli E, Bagulya A, Banerjee S, Barrand G et al 2016 Recent developments in Geant4 Nuclear Instruments and Methods in Physics Research Section A 835 186–225.
- [5]Biggs S, Jennings M, Swerdloff S, Chlap P, Lane D, Rembish J, McAloney J, King P, Ayala R, Guan F et al 2022 PyMedPhys: a community effort to develop an open, Python-based standard library for medical physics applications Journal of Open Source Software 7 4555.DOI
- [6]Blumenfeld P, Pryanichnikov A, Hillman Y, Wajnryt E, Berger A, Winograd S, Wygoda M, Salhab A, Fang M, Feldman J and Popovtzer A 2025 A prospective clinical trial of upright image-guided proton therapy for locally recurrent head and neck and brain malignancies Radiotherapy and Oncology 211 111097.DOI
- [7]Blumenfeld P, Pryanichnikov A, Tochner Z, Allen A M, Fried I, Gozal D, Marzeeq S, Ledot S, Winograd S, Salhab A, Wygoda M, Hillman Y, Feldman J and Popovtzer A 2026 Pediatric CNS relapse of neuroblastoma treated with upright proton craniospinal irradiation Clinical and Translational Radiation Oncology 56 101080.DOI
- [8]Charyyev S, Artz M, Szalkowski G, Chang C W, Stanforth A, Lin L, Zhang R and Wang C K C 2020 Optimization of hexagonal-pattern minibeams for spatially fractionated radiotherapy using proton beam scanning Medical Physics 47 3485–3495.DOI
- [9]Charyyev S, Johansson A, Jiang Z, Heng V J, Chuang C F Y, Piantino M R, Graham S, Skinner L B, Kovachuk N, Maradia V, Dai X, Gu X, Han B, Ghomi L E K, Hiniker S M, Rao Y J, Loo B W Jr, Xing L, Surucu M and Yang Y 2026 Clinical commissioning of the first ultra-compact gantry-less upright proton therapy system arXiv:2609.25141.arxiv.org/abs/2609.25141
- [10]Charyyev S, Maradia V, Chang C W, Lin L, Dong L, Li H, Schuemann J, Shen J, Kang M, Taylor P et al 2025 Multi-institutional survey of source modeling parameters for proton pencil beam scanning treatment planning system commissioning Medical Physics 52 e70199.DOI
- [12]Faddegon B, Ramos-Méndez J, Schuemann J, McNamara A, Shin J, Perl J and Paganetti H 2020 The TOPAS tool for particle simulation, a Monte Carlo simulation tool for physics, biology and clinical research Physica Medica 72 114–121.DOI
- [13]Feldman J, Pryanichnikov A, Shwartz D, Hillman Y, Wygoda M, Blumenfeld P, Marash M and Popovtzer A 2024 Study of upright patient positioning reproducibility in image-guided proton therapy for head and neck cancers Radiotherapy and Oncology 201 110572.DOI
- [14]Gajewski J, Schiavi A, Krah N, Vilches-Freixas G, Rucinski A, Patera V and Rinaldi I 2020 Implementation of a compact spot-scanning proton therapy system in a GPU Monte Carlo code to support clinical routine Frontiers in Physics 8 578605.DOI
- [15]Gordon K, Gulidov I, Koryakin S, Smyk D, Makeenkova T, Gogolin D, Lepilina O, Golovanova O, Semenov A, Dujenko S, Medvedeva K and Mardynsky Y 2021 Proton therapy with a fixed beamline for skull-base chordomas and chondrosarcomas: outcomes and toxicity Radiation Oncology 16 238.DOI
- [16]Gordon K, Gulidov I, Smyk D, Semenov A, Golubev K, Lemaeva A, Koryakin S, Jumaniyazova E, Vishnyakova P, Eremina I, Fatkhudinov T and Kaprin A 2024 Upright proton therapy for esthesioneuroblastoma: a single-institution experience Frontiers in Oncology 14 1348291.DOI
- [17]Hegarty S, Hardcastle N, Korte J, Kron T, Everitt S, Rahim S, Hegi-Johnson F and Franich R 2022 Please place your seat in the full upright position: a technical framework for landing upright radiation therapy in the 21st century Frontiers in Oncology 12 821887.DOI
- [18]Lemaeva A, Gulidov I, Smyk D, Agapova Y, Koryakin S, Eremina I, Gantsova E, Fatkhudinov T, Kaprin A and Gordon K 2024 A single-center experience of the upright proton therapy for skull-base chordomas and chondrosarcomas: updated results Clinical and Translational Radiation Oncology 48 100814.DOI
- [19]Liu C, Ho M W, Park J, Hsi W C, Liang X, Li Z, Song Y, Feng H and Zhang Y 2021 Fast MCsquare-based independent dose verification platform for pencil beam scanning proton therapy Technology in Cancer Research and Treatment 20 1530338211033076.DOI
- [22]Matter M, Nenoff L, Meier G, Weber D C, Lomax A J and Albertini F 2018 Alternatives to patient specific verification measurements in proton therapy: a comparative experimental study with intentional errors Physics in Medicine and Biology 63 205014.DOI
- [23]Meijers A, Guterres Marmitt G, Ng Wei Siang K, van der Schaaf A, Knopf A C, Langendijk J A and Both S 2020 Feasibility of patient specific quality assurance for proton therapy based on independent dose calculation and predicted outcomes Radiotherapy and Oncology 150 136–141.DOI
- [24]Miften M, Olch A, Mihailidis D, Moran J et al 2018 Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group no. 218 Medical Physics 45 e53–e83.DOI
- [25]NEMA 2026 Digital imaging and communications in medicine (DICOM), part 3: Information object definitions https://dicom.nema.org/medical/dicom/current/output/chtml/part03/sect_C.8.8.25.html. PS3.3, Sections C.8.8.25 and C.8.8.25.4.
- [26]Peeters A, Grutters J P C, Pijls-Johannesma M, Reimoser S, De Ruysscher D, Severens J L, Joore M A and Lambin P 2010 How costly is particle therapy? Cost analysis of external beam radiotherapy with carbon-ions, protons and photons Radiotherapy and Oncology 95 45–53.
- [27]Perl J, Shin J, Schümann J, Faddegon B and Paganetti H 2012 TOPAS: an innovative proton Monte Carlo platform for research and clinical applications Medical Physics 39 6818–6837.DOI
- [28]Rinaldi I, Cartechini G, Schiavi A, Gajewski J, Krah N, Rucinski A, Vilches-Freixas G, Patera V and Nijsten S 2026 Five years of clinical application of independent Monte Carlo-based patient-specific quality assurance at Maastro Proton Therapy Center Physica Medica 150 107195.DOI
- [29]Schneider W, Bortfeld T and Schlegel W 2000 Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions Physics in Medicine and Biology 45 459–478.pubmed.ncbi.nlm.nih.gov/10701515
- [31]Volz L, Korte J, Martire M C, Zhang Y, Hardcastle N, Durante M, Kron T and Graeff C 2024 Opportunities and challenges of upright patient positioning in radiotherapy Physics in Medicine and Biology 69 18TR02.DOI
- [32]Yan S, Ngoma T A, Ngwa W and Bortfeld T R 2023 Global democratisation of proton radiotherapy The Lancet Oncology 24 e245–e254.DOI