Robust inverse dielectric design of skin phantoms for implantable wireless link evaluation

Abstract

Skin phantoms support reproducible evaluation of implantable wireless links, yet published complex permittivities of skin predict different propagation losses. We formulate a robust inverse design to determine a homogeneous phantom permittivity that minimizes the largest channel-loss difference across selected skin reference models, incidence angles and both transverse-electric and transverse-magnetic polarizations. Computed candidates and analytical lower bounds bracket the optimal mismatch and constrain the critical depth. Across seven frequencies from 5 to 500 GHz, the minimax mismatch at 1.3 mm ranges from 1.13 to 13.72 dB. For a 3 dB tolerance, suitable permittivities exist at every investigated depth at 5 GHz, while critical depths at the other frequencies range from 0.23 to 1.01 mm. Only one of seven material cases from the literature meets a 6 dB tolerance at 1.3 mm. These results translate channel-loss tolerances into quantitative design criteria for skin phantoms used in implantable wireless link evaluation.