Enhanced Sensitivity of THz NbN Hot Electron Bolometer Mixers
Abstract
We studied the effect of the NbN/Au contact on the sensitivities of a NbN hot electron bolometer (HEB) mixer by measuring the double sideband (DSB) receiver noise temperature (T_rec_DSB) at three local oscillator frequencies of 1.6, 2.5 and 5.3 THz. The HEB has cleaned contact structures with a thick Au layer. We demonstrated low mixer noise temperatures (T_mixer_DSB) of 240 K and 290 K at 1.6 and 2.5 THz, respectively. The latter reach roughly 3 times the quantum noise at their frequencies. The mixer is developed for the proposed OASIS and SALTUS (concept) missions. The enhanced T_mixer_DSB are more than 30 % better in comparison with published NbN HEB mixers. The improvement can reduce the integration time of a heterodyne instrument roughly by a factor of 2. The T_mixer^DSB of the same HEB has shown limited improvement at 5.3 THz, which is partly due to non-optimized antenna geometry. Besides, the results also help to understand device physics of a wide HEB (4 um) at high frequencies.
Introduction
Hot electron bolometer (HEB) mixers based on a thin NbN superconducting bridge are so far the most sensitive heterodyne detectors for high spectral resolution () spectroscopy in astronomic observations in the frequency range between 1 and 6 THz. Due to water vapor absorption in the atmosphere, such observations can only be realized through air-, balloon-, and space-borne observatories. NbN HEBs have been flown, for example, on HIFI-Herschel [1], SOFIA [2], and STO2 [3], and will be flown on GUSTO [4] and ASTHROS [5]. They are also the choice for FIRSS (or LETO) [6], OASIS [7], and SALTUS [8] space missions, proposed (or to be proposed) to ESA or NASA. In particular, due to limited lifetime of a space mission, a low mixer noise temperature () is highly demanded in order to make optimal use of the observation time since the integration time of the heterodyne instruments is proportional to the square of the noise temperature [9].
Many years of research and development at different research groups in the world have been devoted to realizing, understanding, and improving low noise NbN HEBs [10, 11, 12]. Here we only quote the sensitivities for one particular frequency, namely receiver noise temperatures (, the sensitivity of a mixer including the contribution from all the optical components) measured at 1.6 THz or extrapolated to this frequency, which are, for example, 750 K from HIFI [1], 690 K from STO2 [13], and 760 K from GUSTO [14]. The have not been reduced in the past decade. The HEBs for STO2 [13] and GUSTO [14] were fabricated at TU Delft/SRON using contacts that have a superconducting interlayer of either NbTiN [15] or Nb [16] between thin Au and ultra-thin NbN layers. This interlayer helps to overcome the proximity effect to keep the NbN underneath the Au being superconducting. Thus, it also prevents the reduction of the critical temperature () of the NbN bridge. However, the interlayer may introduce side effects, e.g., on RF loss, which can affect . Here we present the results of a new NbN HEB mixer using contacts that were cleaned and have a direct interface between thick Au and NbN. We show that the DSB mixer noise temperature improves by roughly 30 % at 1.6 and 2.5 THz compared to the results that we have previously reported or that are in the literature.
The aim of this study is also to demonstrate sensitive HEB mixers for the SALTUS mission [8], which requires three arrays of HEBs, operating respectively in the 1-2 THz frequency band for numerous molecules, such as HO lines, at a single frequency of 2.67 THz for HD (1-0) line, and at a single frequency of 5.33 THz for HD (2-1) line.
HEB Geometry and Fabrication
The HEB under study (labeled as “OASIS BM2 7B”) has a 400 nm long and 4 m wide NbN bridge, with a normal state resistance () of 79 and a of 8.8 K, embedded in the center of a spiral antenna. We chose the large size instead of a typical size of 200 nm 2 m NbN bridge as used in [17, 18] because of, as to be discussed, the strong lateral proximity effect due to the contacts. Our fabrication method could bring the to a lower value that can affect the LO power and the operating temperature. We aim for an LO power of around 200 nW, which is desirable in view of the receiver stability and availability of the power from existing LO sources. In addition, we expect wider HEBs to show a lower contact resistance if there is, which could lead to an RF loss.
The antenna used in [17] and in GUSTO [18] was modified to accommodate the two-times wider bridge at its center. The log spiral (defined by ), where is the distance
This work was supported partly by NWO (project no. 614.061.609) and partly by TU Delft Space Institute (Corresponding author: Behnam Mirzaei; Jian Rong Gao).
B. Mirzaei, D. Ren and J. R. Gao are with the Department of Imaging Physics at Delft University of Technology, Delft, The Netherlands (emails: [email protected], [email protected], [email protected]).
J. R. G. Silva, W. J. Vreeling, W. Laauwen, and J. R. Gao are with SRON Netherlands Institute for Space Research, Leiden/Groningen, The Netherlands (emails: [email protected] & [email protected] & [email protected]). from the center, is the inner radius, and is the azimuthal angle) was adapted and optimized in COMSOL with the following parameters for best coupling to the lower frequencies: µm and . The device is fabricated on a 380 µm thick, highly resistive Silicon wafer covered by a 5 nm sputtered NbN film [19] with a of 9.9 K.
The spiral antenna including contact structures was patterned by evaporation of a thick, 200 nm Au on the NbN film and a lift-off. The crucial step of properly in-situ Ar milling (cleaning) of the film prior to Au evaporation was performed to realize the Au/NbN interface as transparent as possible. The optimal milling parameters were determined by extensive pretests on a similar NbN film. In detail, the milling beam voltage is 350 V. The beam current is 38 mA. The acceleration voltage is 600 V. The discharge voltage is 30 V. The sample was rotated by 10 rpm. The etching duration is 50 sec. It etches the native oxide (assumed) and also about 8% of the NbN film, corresponding to 0.4 nm, which is monitored by the change of the sheet resistance of the film.
In contrast to the method of adding a superconducting interlayer (e.g., Nb or NbTiN) between the Au contacts and NbN [15, 16] e.g., for GUSTO’s detectors, we here directly deposit the thick Au (200 nm) on the cleaned NbN, which is expected to provide better contacts and thus a lower loss of the RF current from the antenna to the NbN bridge. Besides, this process also reduces the number of fabrication steps, making the fabrication more reliable and reproducible. A similar approach for the contact structures was reported earlier in [20]. An earlier work of HEBs, whose Au/NbN contacts are deposited in-situ, has shown an extremely low of 600 K at 2.5 THz, suggesting also the importance of the interface [21].
In the next step a sub-µm NbN bridge area is masked by a negative E-beam resist followed by the reactive ion etching of the rest of the surface till the remaining bare NbN is fully removed. An optical micrography of such a detector is shown in the inset of Figure 1.

Figure 1. Resistance versus temperature around the superconducting transition of a 400 nm (length) × 4 µm (width) NbN HEB with cleaned Au/NbN contacts. Inset: optical micrograph of the device showing the spiral arms around the NbN bridge.
We measured the resistance vs temperature (RT) of several HEBs with a length of either 200 or 400 nm. As shown in Figure 1, a of 8.8 K for the 400 nm long HEB was typically found, which is about 1 K lower than the of the original film. Furthermore, from the single transition feature in the RT and non-zero resistance at 4 K, the contact regions of Au/NbN become fully normal, confirmed through a dedicated test structure. We also found a as low as 8.2 K obtained from an HEB with a 200 nm long bridge. This confirms a (lateral) proximity effect, inducing from the thick Au contacts to the superconducting NbN bridge, and weakening its superconductivity. This suggests that we have a transmissive (or clean) interface between Au and NbN in the contacts [12].
Experimental Setup
We applied a heterodyne measurement setup in air to measure receiver noise temperature, which is very similar to what reported in [14, 22]. In detail, a far-infrared (FIR) gas laser generates the LO signal at three frequencies of 1.63, 2.52, and 5.25 THz, which are quoted as 1.6, 2.5, and 5.3 THz respectively in this paper. These frequencies are very close to what are required for SALTUS’s mixers. The amount of the LO power to the HEB is controlled by a swing-arm voice coil actuator, where the swing-arm blocks a part of the LO radiation, and thus acts as an effective variable optical attenuator [23], located in front of the FIR laser. The HEB mixer is mounted at the back of a 10 mm diameter elliptical silicon lens, with an antireflection coating optimized for 1.6 THz [24], which couples the radiation to the antenna. The mixer block is placed in a liquid helium cryostat. The THz signal and LO pass through a UHMW-PE window at 300 K and a heat filter (QMC) at 4 K, with a cut-off frequency of 5.8 THz. A 3 µm thick Mylar beam splitter is used to combine the radiation of a hot or cold (295/77 K) load with that of the LO. The hot and cold loads are needed for the Y-factor measurements to determine . The mixer is biased and read out using a bias-T followed by a cryogenic low noise SiGe amplifier (LNA) inside the cryostat with a noise temperature () of 6.5 K. Between the bias-T and LNA there is a circulator to prevent standing waves between the LNA and the HEB. The IF signal is filtered at 1.7 GHz within a band of 80 MHz and is further amplified by a room temperature LNA before it is measured by a power meter. The mixer block is at a temperature of 4.6 K.
The optical losses (L) of the components in the optical path from the hot/cold load to the HEB antenna are summarized in Table I. The air path is ∼ 30 cm long with a relative humidity of ∼ 40 %. It is worthy to mention that the coupling loss between the antenna and the HEB [17] is neither listed in this table nor included in the calculation of and the mixer conversion loss from the measured and receiver conversion loss . The reason for this will be discussed in section IV.
Table I. TABLE I. OPTICAL LOSSES CREATED BY AIR, 3 µm BEAM SPLITTER,
UHMW-PE WINDOW, QMC HEAT FILTER, AND SILICON LENS WITH ANTIREFLECTION COATING AT 1.6 THz. THE LOSS OF THE LENS AT EACH FREQUENCY WAS SIMULATED USING COMSOL
| LO Frequency | (dB) | (dB) | (dB) | (dB) | (dB) |
| 1.63 THz | 0.64 | 0.09 | 0.38 | 1.14 | 0.35 |
| 2.52 THz | 0.38 | 0.19 | 0.61 | 0.34 | 1.11 |
| 5.25 THz | 0.90 | 0.63 | 1.47 | 0.56 | 1.14 |
Table 1.
Results and Discussion
We focus only on the NbN HEB mixer labeled as “OASIS BM2 7B” although very similar performance was found for two other HEBs from the same batch (wafer) and another. A set of measured current-voltage (IV) curves without and with different LO pumping levels at 1.6 THz are shown in Figure 2, where the optimum operating point resulting in the best is indicated.

Figure 2. Unpumped and pumped IV curves of the NbN HEB at 1.6 THz. The thick, red curve is the optimally pumped IV. The optimum operating region within roughly 5 % of the lowest is indicated with a circle, which covers two bias points of 0.6 and 0.8 mV. The dashed load line is used to estimate LO power at HEB itself.
The of the mixer is obtained using the Y-factor that is the ratio between the two measured receiver output powers responding to the hot () and the cold loads (). By combining the Y-factor and the Callen-Welton temperatures of the blackbody loads, the can be obtained [25]. The and are recorded as a function of HEB current at an optimal DC bias voltage (0.6 mV), where the current variation reflects the scan of LO power by the voice coil attenuator. In this way, the measurement of a is not influenced by the fluctuations and drift of the FIR laser power [25]. The latter are known for FIR gas lasers. Furthermore, this technique using the same-current to determine the Y-factor can mitigate the direct detection effect [26, 27, 28, 29] on it and thus on . In essence, the same current is a result of slight change of LO power () to compensate the difference between the broad band hot and cold load power. As a result, the bias points in two cases are adjusted to be the same, but is no longer the same. So, conceptually this is not fully correct because of a difference in the mixer gain [30, 31]. However, in the case of practical NbN HEBs in this work, where the change of is small and is only 2-4 % of the required , this method can mitigate the direct detection effect effectively without overestimating . This conclusion was based on our earlier experiments in [29, 25], where we made a comparison between the same-current method and the conventional fixed-LO-power-method, where a cold, narrow bandpass filter (CNBPF) centered at the LO frequency was introduced. Although the same current-method was used in [26, 28, 32] and in many of our own publications, ideally, we should apply a CNBPF [25, 27, 28] to mitigate the direct detection effect. Unfortunately, we do not have such CNBPFs for three different frequencies available. It is interesting to note that double-current method measures a slightly lower (~10 %) than the conventional fixed-LO-power-method even for a waveguide HEB (without an CNBPF) because the latter still suffers from the direct detection effect [32]. The receiver output power (IF power) measured in response to the hot/cold loads together with the calculated are shown for all three frequencies in Figure 3. The minimum , taken at a current of around 45 μA, is 530, 640, and 2180 K at 1.6, 2.5, and 5.3 THz, respectively. We have repeated measurements by varying the voltage until 1.2 mV in a step of 0.2 mV. We obtain very similar minimum values at 0.8 mV, but with a slightly higher current.

Figure 3. Receiver output power (IF power) in response to hot/cold loads and the fitted curves (left axis), and receiver noise temperature (right axis) versus current at three LO frequencies 1.6, 2.5, and 5.3 THz and bias voltage of 0.6 mV.
We determine the receiver conversion loss () or total loss of the mixer at similar operating points using a standard U-factor technique [33]. The is 6.6, 7.7, and 11.7 dB at 1.6, 2.5, and 5.3 THz, respectively.
We also estimate the absorbed LO power at HEB from the optimally pumped IV curve using the isothermal technique [34]. The is around 240 nW for all three frequencies, which seems to be a good compromise between availability of the LO sources and stability of the receiver, where a too high requirement of LO can challenge the LO source, while a too low requirement of LO can cause the HEB to be sensitive to any fluctuation of THz power [23]. It is also useful to stress that we will operate the HEB at a current of μA for the given voltage (e.g., 0.6 mV). The reason for this is to be discussed. Table II summarizes all the measured receiver performance parameters of , , and LO power at three frequencies. To be able to fairly judge the mixer performance independent of the measurement setup, we extract and mixer conversion loss from the measured receiver parameters in Table II by taking advantage of Table I in the same way as [35]. In detail, for 1.6 THz we subtract the first four optical loss terms listed in Table I, but not the loss term of the coated Si lens, leading to a . Then we apply Eq.1 in [35] using the noise due to the optics () being 101 K, and K to derive the . It is worthy to stress that our includes both the optical loss of the antireflection coated Si lens (0.35 dB) and the antenna-HEB coupling loss (not known).
| LO frequency | LO power | ||
| 1.63 THz | 530 K | 6.6 dB | 245 nW |
| 2.52 THz | 640 K | 7.7 dB | 235 nW |
| 5.25 THz | 2180 K | 11.7 dB | 233 nW |
Table 2. Measured double sideband (DSB) receiver noise temperatures (), DSB receiver conversion losses (), and absorbed LO powers.
| LO frequency | (Quantum noise unit) | ||
| 1.63 THz | 240 K | 3.1 × hν/k_B | 4.4 dB |
| 2.52 THz | 290 K | 2.4 × hν/k_B | 5.5 dB |
| 5.25 THz | 620 K | 2.5 × hν/k_B | 7.4 dB |
Table 3. Derived double sideband (DSB) mixer noise temperature (including the contribution of the Si lens ideally antireflection coated for each specific frequency and antenna-HEB coupling loss), in unit of hν/k_B (QN), and DSB mixer conversion loss (). Note that all are obtained at an IF of 1.7 GHz and an operating temperature of 4.6 K.
To drive and at 2.5 and 5.3 THz, we do slightly differently using the measured data in Table II and the optical loss in Table I, but assuming that the lens in both cases is optimally coated for each frequency so that the loss is about 0.35 dB, similar to 1.6 THz. So, by using of 116 K and 401 K for 2.5 and 5.3 THz, respectively, we derive both and . Like in the case of 1.6 THz, here the include the contribution of the antenna-HEB coupling loss. Table III summarizes the mixer performance parameters i.e., and , which are likely the lowest values ever reported for HEBs. To get a feeling how close the is to the fundamental noise limit, we also quote it in units of the quantum noise (QN, hν/k_B) [9] in this table, where is the Planck constant, frequency, and the Boltzmann constant. We consider Table III to be the key outcome of this paper.
To better sense the improvement we have made in THz HEBs, we will compare with and reported by others or by our team if they are available, otherwise we will compare only. We now compare and with what we obtained from the HEBs used for GUSTO arrays [14], where they are spiral antenna coupled NbN HEBs of 0.2 μm × 2 μm in size with the contact structures of Nb/Au bilayer. More than 16 HEBs were tested at 1.6 THz and similar numbers were tested at 5.3 THz (from which we extrapolated the 4.7 THz data for GUSTO).
It is useful to note that at 1.6 THz the GUSTO test setup was the same as what was used in this work i.e., all optical losses are similar. However, all the data were measured at an IF of 2 GHz, so we extrapolate the data to 1.7 GHz for a comparison. For GUSTO HEBs we found the average and of 390 K and 6.7 dB, respectively. Comparing with the in table III (240 K), our new result has an improvement of 39 %.
This can reduce the integration time of a heterodyne observation instrument by a factor of 2 if we conservatively take only about 30 % since the integration time is proportional to the square of the noise temperature [9], which is remarkable. In other words, the observation of a telescope can be nearly 2 times faster, which significantly impacts the determination of the lifetime of a space mission.
Not only the is improved, but also the in Table III (1.6 THz) is roughly 2 dB lower than what found for GUSTO HEBs. The latter can reduce the noise contribution of an LNA (referred to Eq.1 in [35]) and thus a receiver (or system) noise temperature of a practical instrument.
No data at 2.5 THz from the HEBs for GUSTO were available to compare. But there are numerous data at this frequency published in the literature. The of 600 to 900 K, measured within 1-2 GHz at IF, was reported in the HEB with in-situ made Au/NbN contacts in [21], where the lowest value is comparable to our 2.5 THz data. Other examples include a of 780 K reported in [36], which used also Ar cleaned Au/NbN contacts, 900 K in [37], 1400 K in [38], and 630 K in [23], all obtained using the same setup but in vacuum [25] at SRON, that means less optical loss than our air measurement setup discussed here e.g., no loss of air and the window. Furthermore, 1100 K using an air setup was reported in [39], 950 K in [40], and 1800 K in [41]. The latter was measured earlier in a Nb diffusion-cooled HEB. So, we can generally conclude that our new at 2.5 THz is at least 30 % lower than what reported so far in the literature [36, 37, 38, 39, 40, 41]. Here we compare instead of because the latter is often not available in the literature or is unable to be derived due to some missing information.
Less improvement of the at 5.3 THz may be partly due to the QN, which plays a more vital role when the frequency increases [42].
At 5.3 THz, the HEBs for GUSTO have shown the average and of 760 K and 9.3 dB at an IF of 1.7 GHz [14], respectively. The new (620 K) shows 18 % improvement, which is marginal. We also compare our new result of (2180 K) at 5.3 THz in the literature, where we find it comparable with what reported in [17] and [43] using the vacuum setup at SRON, by correcting for the difference in the optical loss. In these cases, there is almost no improvement.
There are more measured performance data of HEBs in the literature. However, due to the difference in the LO frequency, we do not make a detailed comparison here, but quoting some recent results of measured : 700 K at 1.9 THz (at 1.7 GHz IF) [36] and 950 K at 4.75 THz (at IF between 1.25-1.75 GHz) [32].
Here we try to address possible questions that might arise from the measured data. First, why the in Table III increases by 3 dB from 1.6 THz to 5.3 THz? We believe this is due to an increase in coupling loss between the antenna and the HEB. This is caused by the antenna modification to accommodate a 4 m wide NbN bridge, leading to a less tight winding than the antenna used in [17, 18] with an inner diameter of 6.6 m, which well performs at higher frequencies. In contrast, the antenna here has an inner diameter of 12 m, being less favorable for 5.3 THz [45]. To discuss this quantitatively, we simulate the coupling loss between the antenna and HEB for our case in the same way as in [17], but using COMSOL. We find the coupling loss of 0.13 dB (the efficiency of 97 %), 0.36 dB (92%), and 1.6 dB (69%) for 1.6, 2.5, and 5.3 THz, respectively. Clearly the coupling loss at 5.3 THz is more than that (0.63 dB) for the antenna used in [17, 18]. This can explain a large part of the increase in in Table III. To further quantify the discussion, we calculate the intrinsic DSB mixer conversion loss using the data of from table III by subtracting the antenna-HEB coupling losses and also the loss of 0.35 dB for coated Si lenses at three frequencies. We find the is 3.9, 4.8, and 5.4 dB for 1.6, 2.5, and 5.3 THz, respectively. These values are considerably lower than what was reported in [42] and lower than others reported in THz HEBs as far as we know. However, the remaining increase in the , observed also in [42], is not expected.
We expect the to be frequency independent in the frequency range of our experiment, where THz photon energies are considerably higher than the gap frequency of the NbN bridge. This is supported by the close-to-optimally pumped IV curves at three frequencies being extremely well overlapped, shown in Figure 4. This result suggests that the effect of the heating of THz current in the bridge is the same at the three frequencies. We can further speculate that despite of the wide NbN HEB (4 m), the THz current flow along the bridge is likely uniform, as found for a narrower NbN HEB (2 m) [39]. This is in particular interesting in view of the fact that the skin depth for the NbN at 5.3 THz is 0.43 m, which is only ~ 10 % of the bridge width. As a byproduct of this discussion, we can state that it is still safe to apply the normal state resistance as the RF impedance of the 4 m HEB at a high frequency of 5.3 THz.
Figure 4. Close-to-optimally pumped IV of the NbN HEB at three frequencies of 1.6, 2.5, and 5.3 THz. They seem to be fully overlapped.
The strong lateral proximity effect observed on the RT curve suggests the transparent or clean contacts in our HEBs. Such contacts can reduce the RF loss, as we discussed in the introduction. Besides, they can also affect the temperature profile of the hot electrons across the bridge as well as in the contact regions. The measured low mixer conversion losses at 1.6 and 2.5 THz in table III seem to support these hypotheses. Following a hot spot model of mixing for an HEB [31, 47], the contact properties are expected to affect the conversion loss and noise qualitatively. Our improved confirms that the contacts play a crucial role experimentally, but not yet theoretically. The reported simulations, such as in [31, 48] based on the hot-spot model are unable to address this effect since most of them do not include the role of the contacts.
Our contacts behave differently from the earlier Au/NbN contacts [15, 40], which do not change the of the contact region and Nb bridge, suggesting poor interfaces as a result of no Ar cleaning.
The enhanced sensitivity of the NbN HEB mixer by engineering the contacts reported here can also stimulate the development of novel MgB HEB mixers [35, 43, 44], which can offer a high operating temperature of K and a large IF bandwidth of GHz. However, little has been explored on the contacts [21].
The optimal operating region of the HEB is indicated in Figure 2. More specifically, at 0.6 mV we will choose the HEB to be operated at the current of 45 A or below as suggested by Figure 3. One should avoid working at a current above 45 A because it moves away from the resistive state, where a HEB is known to be unstable [51] and where relaxation oscillations switching between the superconducting state and resistive state are present [52]. Such oscillations were also recently reported in [53]. The same work also found weaker, intrinsic oscillations when a NbN HEB is in the resistive state without LO power and is operated near its . Although it is unclear whether the intrinsic oscillations can affect our performance of the HEB, it might be worthy to perform a dedicated measurement to localize the region of low , but being free of any unwanted oscillations.
Conclusion
By introducing cleaned thick Au contacts to a NbN HEB mixer we have demonstrated extremely low mixer noise temperatures () of 240 and 290 K at 1.6 and 2.5 THz, respectively, which are about 3 times the quantum noise (h/kB). The mixer is developed for future FIR space observatories, in particular, for SALTUS mission concept, the proposal of which will be submitted to NASA in response to the Probe mission call. The improvement of is more than 30 % in comparison with reported NbN HEB mixers [14]. This can reduce the integration time of a heterodyne instrument roughly by a factor of 2. In other words, such an improvement can make the observation twice as fast, that is highly demanded for cryogenic space missions in view of limited resources. The of the same HEB at 5.3 THz has shown 620 K and thus a limited improvement, which is partly due to the non-optimized antenna geometry, supported by the high antenna- HEB coupling loss, and is likely also due to the QN at this high frequency. Interestingly, despite of a 4 m wide NbN bridge, the pumped IV curves are identical at three different frequencies, suggesting that the RF current flows uniformly along the bridge.
Acknowledgment
The authors thank Yuner Gan for modeling the loss of the coated Si lens. We acknowledge Marc Zuiddam and the staff in Kavli Nanolab Delft for their technical support. We are also grateful for the huge support from Paul Urbach at TU Delft, and Pieter Dieleman, Jan-Willem den Herder, and Jan Geralt bij de Vaate at SRON.
References
- [1]S. Cherednichenko, V. Drakinskiy, T. Berg, P. Khosropanah and E. Kollberg, “Hot-electron bolometer terahertz mixers for the Herschel Space Observatory,” Review of Sci. Instruments, vol. 79, pp. 034501, Mar. 2008.DOI
- [2]C. Risacher, R. Güsten, J. Stutzki, H.-W. Hübers, R. Aladro, A. Bell, C. Buchbender, D. Büchel, T. Csengeri, C. Duran, U. U. Graf, R. D. Higgins, C. E. Honingh, K. Jacobs, M. Justen, B. Klein, M. Mertens, Y. Okada, A. Parikka, P. Pütz, N. Reyes, H. Richter, O. Ricken, D. Riquelme, N. Rothbart, N. Schneider, R. Simon, M. W. Wiedner, H. Wiesemeyer, M. Ziebart, P. Fusco, F. Rosner and B. Wohler, “The upGREAT Dual Frequency Heterodyne Arrays for SOFIA,” J. Astron. Instrum., vol. 7, no. 4, pp. 1840014, Dec. 2018.DOI
- [3]Y. M. Seo, P. F. Goldsmith, C. Walker, D. J. Hollenbach, M. G. Wolfire, C. Kulesa, V. Tolls, P. N. Bernasconi, Ű. Kavak, F. F. S. van der Tak, R. Shipman, J. R. Gao, A. Tielens, M.G. Burton, H. York, E. Young, W. L. Peters, A. Young, C. Groppi, K. Davis, J. L. Pineda, W. D. Langer, J. H. Kawamura, A. Stark, G. Melnick, D. Rebolledo, G. F. Wong, S. Horiuchi and T. B. Kuiper, “Probing ISM Structure in Trumpler 14 and Carina I Using the Stratospheric Terahertz Observatory 2,” ApJ, vol. 878, no. 2, June 2019.DOI
- [4]C. Walker, C. Kulesa, A. Young, W. Verts, J. R. Gao, Q. Hu, J. Silva, B. Mirzaei, W. Laauwen, J. Hesler, C. Groppi, and A. Emrich, “Gal/Xgal U/LDB Spectroscopic/Stratospheric THz Observatory: GUSTO,” in Proc. SPIE 12190, Millimeter, Submillimeter, and Far-Infrared Detectors and Instrumentation for Astronomy X, Montréal, Québec, Canada, 2022, article no. 121900E.DOI
- [5]Jose V. Siles, Jorge Pineda, Jonathan H. Kawamura, Paul F. Goldsmith, et al., “ASTHROS: The astrophysics stratospheric telescope for high spectral resolution observations at submillimeter-wavelengths,” in Proc. SPIE 11445, Ground-based and Airborne Telescopes VIII, Online, California, USA, 2020, article no. 114453G.DOI
- [6]D. Rigopoulou, C. Pearson, B. Ellison, M. Wiedner, V. Ossenkopf Okada, et al., “The far-infrared spectroscopic surveyor (FIRSS). Experimental Astronomy,” Exp Astron., Vol.51, no. 3, pp. 699-728, May 2021.DOI
- [7]C. K. Walker, G. Chin, S. Aalto, C. M. Anderson, J. W. Arenberg, C. Battersby, et al., “Orbiting Astronomical Satellite for Investigating Stellar Systems (OASIS): following the water trail from the interstellar medium to oceans,” in Proc. SPIE 11820, Astronomical Optics: Design, Manufacture, and Test of Space and Ground Systems III, San Diego, California, USA, 2021, article no. 1182000.
- [8]https://live-az-saltus.paonsite.io/sites/default/files/2023-03/SALTUS_Intro_5March23.pdf
- [9]C. K. Walker, Terahertz Astronomy, New York, NY, USA: Taylor & Francis, 2016, pp. 159–227.
- [10]H.-W. Hubers, “Terahertz Heterodyne Receivers,” IEEE J. Sel. Top. Quantum Electron., vol. 14, no. 2, Apr. 2008.DOI
- [11]A Shurakov, Y Lobanov and G Goltsman, “Superconducting hot-electron bolometer: from the discovery of hot-electron phenomena to practical applications,” Supercond. Sci. Technol., vol. 29, no. 2, pp. 023001, Dec. 2015.DOI
- [12]T. M. Klapwijk and A. V. Semenov, “Engineering physics of superconducting hot-electron bolometer mixers,” IEEE Trans. Terahertz Sci. Technol., vol. 7, no. 6, pp. 627–648, Oct. 2017.DOI
- [13]Xuxiang Liu, “Demonstration of 2×2 heterodyne receiver array at 1.4THz using HEB mixers and Fourier phase grating LO,” M.S. thesis, Delft University of Technology, Delft, The Netherlands, 2015
- [14]José R. G. Silva, Wouter M. Laauwen, Behnam Mirzaei, Nathan Vercruyssen, Matvey Finkel, Menno Westerveld, Nikhil More, Vitor Silva, Abram Young, Craig Kulesa, Christopher Walker, Floris van der Tak and Jian Rong Gao, “4×2 Hot electron bolometer mixer arrays for detection at 1.46, 1.9 and 4.7 THz for a balloon borne terahertz observatory,” to be submitted to IEEE Trans. Terahertz Sci. Technol. (2023 special issue).arxiv.org/abs/2311.05755
- [15]M. Hajenius, J. J. A. Baselmans, J. R. Gao, T.M. Klapwijk, P. A. J. de Korte, B. Voronov and G. Gol’tsman, “Low noise NbN superconducting hot electron bolometer mixers at 1.9 and 2.5 THz,” Supercond. Sci. Technol., vol. 17, pp. S224–S228, Mar. 2004.DOI
- [16]T. Aggarwal, P. Khosropanah, W. Zhang, F. D. Tichelaar, J.R. Gao, T. M. Klapwijk, “Superconducting contacts and NbN HEB mixer performance” in Proceedings of the 19th International Symposium on Space Terahertz Technology, Groningen, The Netherlands, 2008, pp. 398-402.
- [17]W. Zhang, P. Khosropanah, J. R. Gao, T. Bansal, T. M. Klapwijk, W. Miao, and S. C. Shi, "Noise temperature and beam pattern of a NbN hot electron bolometer mixer at 5.25 THz," J. Appl. Phys., vol. 108, no. 9, pp. 093102, Nov. 2010.DOI
- [18]J. R. G. Silva, M. Finkel, W. M. Laauwen, S. J. C. Yates, B. Mirzaei, N. Vercruyssen, A. Young, C. Kulesa, C. Walcker, F. van der Tak and J. R. Gao, “Beam waist properties of spiral antenna coupled HEB mixers at Supra-THz frequencies,” IEEE Trans. Terahertz Sci. Technol., vol. 13, no. 2, pp. 167-177, Mar. 2023.DOI
- [19]The NbN film on Si was provided by TERAHERTZ AND INFRARED PHOTONICS, Moscow, Russia. We measured the thickness to be 5 nm using High-resolution transmission electron microscopy (HRTEM) although the nominal thickness was 3.5 nm.
- [20]R. Lefèvre, Y. Jin, A. Féret, T. Vacelet, W. Miao, L. Pelay, F. Dauplay, M. Ba-Trung, G. Gay, J.-C. Villégier, J. Spatazza, J.-M. Krieg, and Y. Delorme, “Terahertz NbN hot electron bolometer fabrication process with a reduced number of steps”, in Proceedings of the 23rd International Symposium on Space Terahertz Technology, Tokyo, April 2012, pp. 122-125.
- [21]I. Tretyakov, S. Ryabchun, M. Finkel, A. Maslennikova, N. Kaurova, A. Lobastova, B. Voronov, and G. Gol’tsman, “Low noise and wide bandwidth of NbN hot-electron bolometer mixers,” Appl. Phys. Lett., vol. 98, pp. 033507, 2011.DOI
- [22]Y. Gan, B. Mirzaei, J. R. G. D. Silva, S. Cherednichenko, F. van der Tak, and J. R. Gao, "Heterodyne performance and characteristics of terahertz MgB2 hot electron bolometers," J. Appl. Phys., vol. 133, no. 7, pp. 074503, Feb. 2023.DOI
- [23]D. J. Hayton, J.R. Gao, J.W. Kooi, Y. Ren, W. Zhang, and G. de Lange, “Stabilized hot electron bolometer heterodyne receiver at 2.5 THz,” Appl. Phys. Lett., vol. 100, no. 8, pp. 081102, Feb 2012. See the footnote 19.DOI
- [24]We use an elliptical Si lens with a semi-minor axis of a 5 mm, the semi-major axis b of 5.235 mm, and the extension c from geometric center of the lens of 1.2 mm. The lens has antireflection coating optimized for 1.6 THz. The lens is the same as the one used in [14, 22].
- [25]P. Khosropanah, J. R. Gao, W. M. Laauwen, M. Hajenius, “Low noise NbN hot electron bolometer mixer at 4.3 THz,” Appl. Phys. Lett., vol. 91, no. 22, pp. 221111, Nov. 2007.DOI
- [26]S. Scherednichenko, A. Vereykin, B. Voronov, E. Menshikov, E. Gershenzon, and G. Gol’tsman, “Quasioptical phonon-cooled NbN hot-electron bolometer mixer at 0.5–1.1 THz,” in Proceedings of the 9th International Symposium on Space Terahertz Technology, Pasadena, CA, USA, 1998, pp. 45–53.
- [27]J. J. A. Baselmans, A. Baryshev, S. F. Reekerm, M. Hajenius, J. R. Gao, T. M. Klapwijk, Y. Vachtomin, S. Maslennikov, S. Antipov, B. Voronov, and G. Gol’tsman, “Direct detection effect in small volume hot electron bolometer mixers,” Appl. Phys. Lett., vol. 86, no. 16, pp. 163503-1–163503-3, 2005.DOI
- [28]S. Cherednichenko, V.Drakinsky, T.Berg, E. L.Kollberg, and I.Angekov, “The direct detection effect in the hot-electron bolometer mixer sensitivity calibration,” IEEE Trans Microw. Theory Tech., vol. 55, no. 3, pp. 504-510, Mar. 2007.
- [29]M. Hajenius, J. J. A. Baselmans, A. Baryshev, J. R. Gao, T. M. Klapwijk, J. W. Kooi, W. Jellema, and Z. Q. Yang, “Full characterization and analysis of a terahertz heterodyne receiver based on a NbN hot electron bolometer,” J. Appl. Phys., vol. 100, pp. 074507-1–074507-10, Oct. 2006.DOI
- [30]B. S. Karasik and A. I. Elant ev, “Noise temperature limit of a superconducting hot-electron bolometer mixer,” Appl. Phys. Lett. vol. 68, no. 6, pp. 853-855, 1996.DOI
- [31]H. F. Merkel, P. Khosropanah, D. Wilms Floet, P. A. Yagoubov, and E. L. Kollberg, “Conversion gain and fluctuation noise of phonon-cooled hot-electron bolometers in hot-spot regime,” IEEE Trans. Microwave Theory Tech., vol. 48, no. 4, pp. 690-699, Apr. 2000.
- [32]D. Büchel et al., “4.7-THz Superconducting Hot Electron Bolometer Waveguide Mixer,” IEEE Trans. Terahertz Sci. Technol., vol. 5, no. 2, pp. 207-214, March 2015.
- [33]S. Cherednichenko, M. Kroug, H. Merkel, P. Khosropanah, A. Adam, E. Kollberg, D. Loudkov, G. Gol’tsman, B. Voronov, H. Richter, and H. Huebers, “1.6 THz heterodyne receiver for the far infrared space telescope,” Phys. C: Supercond. and Its Applicat., vol. 427, pp. 427-431, Mar. 2002.
- [34]H. Ekström, B. S. Karasik, E. Kollberg and K. S. Yngvesson, “Conversion gain and noise of niobium superconducting hot-electron-mixers,” IEEE Trans Microw. Theory Tech., vol. 43, no. 4, pp. 938-947, Apr.1995.
- [35]Y. Gan, B. Mirzaei, J. R. G. D. Silva, J. Chang, S. Cherednichenko, F. van der Tak, and J. R. Gao, "Low noise MgB2 hot electron bolometer mixer operated at 5.3 THz and at 20 K," Appl. Phys. Lett., vol. 119, no. 20, pp. 202601, Nov. 2021.DOI
- [36]Y. Delorme, R. Lefèvre, W. Miao, A. Féret, W. Zhang, T. Vacelet,F. Dauplay, L. Pelay, J. Spatazza, M. Ba Trung, J-M. Krieg, Y.Jin, P. Khosropanah, J. R. Gao and S. C. Shi, “A quasi-optical NbN HEB mixer with 800 K DSB noise temperature at 2.5 THz,” in Proceedings of the 22nd International Symposium on Space Terahertz Technology, Tucson, AZ, USA, 2011, pp. 123-126.
- [37]W. Zhang, J. R. Gao, M. Hajenius, W. Miao, P. Khosropanah, T. M. Klapwijk, and S. C. Shi, “Twin-slot antenna coupled NbN hot electron bolometer mixer at 2.5 THz,” IEEE Trans. Terahertz Sci. Technol., vol. 1, no. 2, pp. 378 -382, Nov. 2011.DOI
- [38]P. Pütz, C. E. Honingh, K. Jacobs, M. Justen, M. Schultz, and J. Stutzki, “Terahertz hot electron bolometer waveguide mixers for GREAT,” A&A., vol. 542, L2, June 2012.
- [39]M. Kroug et al., “NbN hot electron bolometric mixers for terahertz receivers,” IEEE Trans. Appl. Supercond., vol. 11, no. 1, pp. 962-965, March 2001.DOI
- [40]J. J. A. Baselmans, M. Hajenius, J. R. Gao, T. M. Klapwijk, P. A. J. de Korte, B. Voronov and G. Gol’tsman, “Doubling of sensitivity and bandwidth in phonon cooled hot electron bolometer mixers,” Appl. Phys. Lett., vol. 84, no. 11, pp. 1958-1961, Mar. 2004.
- [41]R.A. Wyss, B.S. Karasik, W.R. McGrath, B. Bumble, and H. LeDuc, ”Noise and Bandwidth Measurements of Diffusion-Cooled Nb Hot-Electron Bolometer Mixers at Frequencies Above the Superconductive Energy Gap,” in Proceedings of the 10th International Symposium on Space Terahertz Technology, Charlottesville, Virginia, USA, 1999, pp. 215-228.
- [42]W. Zhang, P. Khosropanah, J. R. Gao, E. L. Kollberg, K. S. Yngvesson, T. Bansal, R. Barends, and T. M. Klapwijk, “Quantum noise in a terahertz hot electron bolometer mixer,” Appl. Phys. Lett., vol. 96, no. 11, pp. 111113, Mar. 2010.DOI
- [43]J. L. Kloosterman, D. J. Hayton, Y. Ren, T. Y. Kao, J. N. Hovenier, J. R. Gao, T. M. Klapwijk, Q. Hu, C. K. Walker, and J. L. Reno, “Hot electron bolometer heterodyne receiver with a 4.7-THz quantum cascade laser as a local oscillator,” Appl. Phys. Lett., Vol. 102, no. 1, pp. 011123, Jan. 2013.DOI
- [44]C. Risacher et al., "First Supra-THz Heterodyne Array Receivers for Astronomy with the SOFIA Observatory," IEEE Trans. Terahertz Sci. Technol., vol. 6, no. 2, pp. 199-211, March 2016.DOI
- [45]A. D. Semenov, H. Richter, H. Hubers, B. Gunther, A. Smirnov, K. S. Il’in, M. Siegel and J. P. Karamarkovic, “Terahertz Performance of Integrated Lens Antennas with a Hot-Electron Bolometer,” IEEE Trans. Microw. Theory Tech., vol. 55, no. 2, pp. 239-247, Feb. 2007.DOI
- [47]D. Wilms Floet, E. Miedema, T. M. Klapwijk, and J. R. Gao, “Hotspot mixing: A framework for heterodyne mixing in superconducting hot-electron bolometers,” Appl. Phys. Lett. vol. 74, pp. 433-435, Jan. 1999.DOI
- [48]W. Miao, W. Zhang, K. M. Zhou, H. Gao, K. Zhang, W. Y. Duan, Q. J. Yao, S. C. Shi, Y. Delorme, and R. Lefevre, “Investigation of the Performance of NbN Superconducting HEB Mixers of Different Critical Temperatures,” IEEE Trans. Appl. Supercond., vol. 27, no. 4, June 2017.DOI
- [51]A. D. Semenov, H.-W. Hubers, J. Schubert, G. N. Gol’tsman, A. I. Elantiev, B. M. Voronov, and E. M. Gershenzon, “Design and performance of the lattice-cooled hot-electron terahertz mixer,” J. Appl. Phys., vol. 88, no. 11, pp. 6758, 2000.DOI
- [52]Y. Zhuang and K. S. Yngvesson, “Detection and Interpretation of Bistability Effects in NbN HEB Devices,” in Proceedings of the 13th International Symposium on Space Terahertz Technology, Cambridge, USA, 2002, pp. 463-472.
- [53]R F Su, Y D Zhang, X C Tu, X Q Jia, C H Zhang, L Kang, B B Jin, W W Xu, H B Wang, J Chen, and P H Wu, “Microwave probing of relaxation oscillations related to terahertz power detection in superconducting hot electron bolometers,” Supercond. Sci. Technol. vol. 32, pp. 105002, 2019.DOI