ACUS: An Angular Calibrated Uncertainty Score for Incomplete-View CT

Abstract

Sparse-view and limited-arc computed tomography (CT) are commonly treated as projection completion: a network fills an unmeasured view, and the result is ranked by similarity to one ground-truth image. That score can be high even when the query angle is far from every acquired measurement, because a learned prior may invent anatomy the scan never constrained. Existing protocols therefore leave a basic question unanswered: how does predictive reliability change with distance from the acquired views? We propose the Angular Calibrated Uncertainty Score (ACUS) to answer that question. ACUS is the Winkler interval score of a nominal 90% predictive interval on held-out projections, binned by distance to the nearest acquired view. It charges interval width and adds a large miss penalty when a held-out pixel falls outside the interval, so a narrow band helps only when coverage holds. We use collapse on acquired angles as a necessary consistency check on the predictor. On chest and dental cone-beam orbits, interval width grows with the angular gap. Empirical coverage ranges from 77% to 92%, and the peak signal-to-noise ratio (PSNR) of the point estimate falls as the width rises. Nearest-neighbor, interpolation, and point completion, each given one residual scale, have zero reported far coverage and incur outside-distance penalties. ACUS exposes gap-dependent overconfidence that a point-estimate PSNR alone does not measure.

Introduction

Incomplete-view computed tomography (CT) includes sparse-view and limited-arc acquisition. Figure 1(a) shows the incomplete sinogram: angles with no rays can still be filled from an anatomical prior [13, 24]. The result can look anatomical, and the question is whether it is trustworthy where the scan never measured.

ACUS overview: estimation reconstruction of incomplete CT scan, existing metrics, and angular calibrated uncertainty score

Figure 1. Incomplete-view CT as an interval problem. (a) Sparse-view and limited-arc scans leave angles with no ray coverage. A prior can complete them, and the question is whether the image is trustworthy. (b) PSNR, SSIM, and acquired-view geometry give no error bound on those angles. (c) ACUS reports width and coverage by the gap to the nearest acquired view: narrow nearby, wide far away. A static scale fails at large gaps.

Figure 1(b) shows why the peak signal-to-noise ratio (PSNR) or structural similarity (SSIM) does not answer that

question. Either score compares a prediction with one reference and can stay high far from every acquired view, because the prior can introduce structure the measurements do not determine. A geometry summary of the acquired trajectory has the same blind spot: it describes the rays that were taken and gives no error bound on the held-out projection [14, 16]. Many volumes match the acquired integrals and disagree elsewhere. This set is the measurement fiber.

Figure 1(c) scores that held-out projection. ACUS is the Winkler interval score of a nominal 90% interval, indexed by the gap δ\delta to the nearest acquired view [11, 22]. The interval is narrow near an acquired view and wide far away, and coverage is reported with the width. A static scale, one residual magnitude at every angle, fails at large gaps. Collapse on acquired angles checks that compatible volumes reproduce the measurements that were taken. The score charges width and penalizes pixels outside the interval, so a narrow interval helps only when coverage holds. § 3.1 defines the score, and § 4 records a refinement that can raise volume similarity while agreeing less with the acquired projections.

In this paper, we propose ACUS, a gap-indexed interval score for incomplete-view CT, together with a residual-scale estimator and a held-out evaluation on chest and dental cone-beam orbits. Our contributions can be summarized as:

  • Angular interval score. ACUS is the nominal 90% Winkler score on held-out projections, binned by the angular gap δ\delta. The score charges interval width and a miss penalty, so a narrow interval is rewarded only when coverage holds. Acquired-view collapse is the consistency check.

  • Residual-scale estimator. We train a feed-forward mean and pixelwise scale with held-out likelihood and an acquired-view collapse gate. The published interval opens with the angular gap, instead of reusing the static scale in Figure 1(c).

  • Cone-beam evaluation. We evaluate the score and the estimator on chest and dental cone-beam orbits under sparse and multi-arc K=8K=8 sampling, with 40 test cases per domain. Interval width grows with the gap, empirical coverage ranges from 77% to 92%, and constant-scale point predictors fail the far bins by missing held-out observations.

Related work

Reconstruction and projection prediction. Incomplete-view tomography admits multiple volumes consistent with the acquired line integrals [14, 16]. Learned reconstruction can select a useful estimate through an anatomical prior. Convolutional post-processing [13] and dual-domain metal-artifact reduction [15] illustrate how image and projection information can improve point estimates, although their acquisition and corruption models differ from ours. Implicit attenuation fields [7, 25] and radiative Gaussian representations [6, 26] instead fit a spatial representation through rendered projections. These advances motivate a complementary evaluation question: how reliable is a predicted line integral at an angle absent from the scan? A strong reconstruction score does not by itself answer that question.

Predictive representations. Masked image modeling [12] and joint-embedding prediction [4] learn transferable features rather than calibrated detector intensities. X-WIN applies viewpoint-conditioned latent prediction to chest radiographs and demonstrates projection rendering [24]. This makes predictive representations relevant backbones, but the latent objective does not specify a nominal interval or a penalty for missing a held-out measurement. Our point-completion baseline isolates a point-estimate objective using the same backbone and an ℓ1\ell_1 projection loss. It is not a reproduction of every cited representation learner.

Uncertainty and its evaluation target. Geometric incompleteness quantifies the information supplied by a source trajectory at a voxel [9]. Uncertainty-aware null-space networks combine data-consistent reconstruction with an image-domain residual scale [3]. Recent analysis of radiative Gaussian uncertainty distinguishes posterior spread from localization of realized reconstruction error [27]. The target matters in each case: geometry, voxel error, and a held-out projection are different quantities. ACUS evaluates the last, retaining both interval length and outside distance through the classical interval score [11, 22]. Distribution-free image regression uses a separate calibration procedure to obtain coverage guarantees under its assumptions [2]. Our untempered Gaussian intervals receive an empirical assessment, not such a guarantee. A scoring rule determines what predictions are rewarded, while calibration determines how interval frequencies behave on a specified population. Keeping these roles separate allows the same interval criterion to evaluate different predictors without attributing statistical guarantees to their architecture.

Sequential acquisition. Learned sampling policies [19], sequential experimental design [20], and stopping policies [21] optimize the acquisition sequence or its termination. Perturbed Gaussian Ensemble uses structural variation across density perturbations to select a new view [23]. Those methods evaluate the utility of acquiring a measurement. ACUS instead scores the predictive interval before that measurement is revealed. The distinction permits retrospective comparison without assuming that an uncertainty ranking is already a validated acquisition policy. Testing whether the score improves a closed-loop policy remains a separate experiment.

Method

ACUS scores the nominal 90% interval of a predictive mean and a pixelwise scale. On a covered pixel the score equals the interval width. On a missed pixel it adds twenty times the outside distance, so a narrow interval helps only when coverage holds (Equation (7)). Held-out likelihood trains the scale to open with the angular gap. A collapse loss trains it to close on acquired angles. Figure 2(a) is the predictor, (b) its training, and (c) the score. The gap and the score are defined below. The network reads only the two nearest views. δ\delta and the measurement fiber still depend on the full acquired set.

Interval estimator used to compute ACUS

Figure 2. Interval estimator used to compute ACUS. (a) A frozen MAE encodes acquired views. Distance-weighted softmax fuses the two nearest into a mean μ\mu and a scale σ\sigma. Point completion is the mean-only baseline. δ\delta marks acquired, near, and far queries. (b) Chest and dental specialists train on sparse and multi-arc orbits with dropout and held-out queries. Gaussian likelihood has no variance target. Collapse drives σ\sigma toward zero and μ\mu toward the acquired view. (c) The nominal 90% interval is μ±1.645σ\mu\pm1.645\sigma (width 3.29σ3.29\sigma). Inside, ACUS equals the width. Outside, it adds twenty times the distance to the bound.

Angular interval scoring

A circular cone-beam orbit provides projections ya=Pa(V)y_a = P_a(V) of an unknown volume VV. One scan reveals a set A\mathcal{A}. In the main protocol A\mathcal{A} is either a sparse subset of K=8K=8 views on a 50-view, 180∘180^\circ orbit, or two to three short arcs of 25∘25^\circ to 45∘45^\circ whose total view count is still K=8K=8. A single contiguous arc of 50∘50^\circ to 120∘120^\circ is reported only as a stress test. The measurement fiber is

F(A,yA)={V:Pa(V)=ya, a∈A}.(1)\mathcal{F}(\mathcal{A}, y_{\mathcal{A}}) = \{V : P_a(V) = y_a,\ a \in\mathcal{A}\}. \tag*{(1)}

At a query a′∉Aa' \notin\mathcal{A} the projections still compatible with the acquired integrals are the push-forward {Pa′(V):V∈F}\{P_{a'}(V) : V \in\mathcal{F}\}. The fiber is a set of volumes. It carries no probability by itself, and the network never enumerates it. On an acquired angle the noise-free push-forward is the singleton {ya}\{y_a\}.

What the angular gap controls. For one fixed volume the cone-beam transform is continuous in the source angle, so the true projection returns to the acquired measurement as the query approaches it. That fact does not bound the rest of F\mathcal{F}. Finitely many views leave a null space, and without a regularity class two consistent volumes can separate after an arbitrarily small angle step. Inside a class B\mathcal{B} on which the transform is uniformly continuous in angle, every legal projection at a′a' lies near the nearest acquired measurement, and the diameter of this restricted push-forward tends to zero with the gap δ(a′)\delta(a'). More precisely, if ∥Pb(V)−Pa(V)∥≤ωB(∣b−a∣)\lVert P_b(V)-P_a(V)\rVert\leq\omega_{\mathcal{B}}(|b-a|) uniformly over V∈BV \in\mathcal{B}, the triangle inequality gives diameter at most 2ωB(δ)2\omega_{\mathcal{B}}(\delta) for the restricted push-forward. This conditional observation motivates the boundary check, not a numerical bound on the network’s scale. It does not set how fast the spread opens at a large gap. That rate depends on B\mathcal{B}. In practice the specialist learns an anatomical prior from training data. We do not identify its support with a known regularity class or estimate a uniform modulus from finite tests. We use δ(a′)=min⁡a∈A∣a′−a∣\delta(a') = \min_{a \in\mathcal{A}} |a' - a| on the sampled 180∘180^\circ trajectory, without identifying its endpoints. Figure 2(a) marks three regimes of this gap: acquired (δ=0\delta=0), near (small δ\delta), and far (large δ\delta). The numerical cuts, under 20∘20^\circ and at least 40∘40^\circ, are fixed in § 4. The index is computed from the mask, so it is independent of the predictor’s errors. Pooling solely by view count would mix interpolation near measured support with extrapolation across large gaps. It is not a voxel-incompleteness score and not a condition number of the whole missing wedge, so sparse and multi-arc orbits are reported separately at the same δ\delta.

The published interval. Figure 2(c) draws the interval that ACUS scores. We publish a diagonal Gaussian on detector pixels, as a predictive stand-in for the restricted push-forward,

p(x∣a′,A,yA)=N(μ(a′),diag⁡σ(a′)2),(2)p(x \mid a', \mathcal{A}, y_{\mathcal{A}}) = \mathcal{N}\bigl(\mu(a'), \operatorname{diag} \sigma(a')^2\bigr), \tag*{(2)}

where xx is the predicted detector image and Ω\Omega is the set of detector pixels. We summarize the distribution by the mean scale

u(a′)=1∣Ω∣∑i∈Ωσi(a′).(3)u(a') = \frac{1}{|\Omega|}\sum_{i\in\Omega}\sigma_i(a'). \tag*{(3)}

The conditioning in (2) is written on all of A\mathcal{A}. § 3.2 uses only the two nearest views. The nominal 90% interval at detector pixel ii is [Li,Ui]=μi±1.645σi[L_i,U_i]=\mu_i\pm1.645\sigma_i. ACUS is the mean Winkler score of that interval on the held-out intensity yiy_i [11, 22],

Si={Ui−Li,Li≤yi≤Ui,Ui−Li+2α(Li−yi),yi<Li,Ui−Li+2α(yi−Ui),yi>Ui.(4)S_i = \begin{cases} U_i-L_i, & L_i\le y_i\le U_i,\\ U_i-L_i+\frac{2}{\alpha}(L_i-y_i), & y_i<L_i,\\ U_i-L_i+\frac{2}{\alpha}(y_i-U_i), & y_i>U_i. \end{cases} \tag*{(4)}
ACUS⁡(a′)=1∣Ω∣∑i∈ΩSi,α=0.1.(5)\operatorname{ACUS}(a')=\frac{1}{|\Omega|}\sum_{i\in\Omega}S_i,\qquad\alpha=0.1. \tag*{(5)}

Lower is better. With α=0.1\alpha=0.1 the central interval is μ±1.645 σ\mu\pm1.645\,\sigma, so the width Ui−Li=3.29 σiU_i-L_i=3.29\,\sigma_i is the sharpness penalty in Figure 2(c). A miss adds 20 times its distance outside that bound. For a continuous target cumulative distribution FF, differentiating the expected score with respect to its two endpoints gives

∂LES⁡=−1+20F(L),∂UES⁡=1−20[1−F(U)].(6)\partial_L\operatorname{ES}=-1+20F(L),\qquad\partial_U\operatorname{ES}=1-20[1-F(U)]. \tag*{(6)}

The optima are the 0.05 and 0.95 quantiles. They are unique when these quantiles are unique [11, 22]. Thus the score elicits the central interval, rather than the full predictive density. The Gaussian restriction can prevent μ±1.645σ\mu\pm1.645\sigma from attaining those quantiles for an asymmetric target. Properness is an expected-score statement under the target distribution, not a claim that a fitted interval is calibrated. It does not identify F\mathcal{F}. Let di=(Li−yi)++(yi−Ui)+d_i=(L_i-y_i)_+ +(y_i-U_i)_+ and let dˉ\bar d be its mean. The score then splits as

ACUS⁡(a′)=3.29 u(a′)+20 dˉ(a′).(7)\operatorname{ACUS}(a')=3.29\,u(a')+20\,\bar d(a'). \tag*{(7)}

On a covered pixel the score equals the interval length 3.29 σi3.29\,\sigma_i, not the scale σi\sigma_i. If every pixel is covered, ACUS⁡=3.29 u\operatorname{ACUS}=3.29\,u. Tables list the mean scale uˉ(δ)\bar u(\delta) and the coverage cˉ(δ)\bar c(\delta) so the two terms stay visible. Coverage and mean scale must be read together. Coverage near 90% alone does not bound the miss term, because misses can have different magnitudes. PSNR of μ\mu is written beside them and is not the acceptance test.

A held-out projection is one legal member of the push-forward, the projection of the scanned volume. Coverage is the rate at which that member falls inside the interval. A rate near 90% is evidence of marginal empirical coverage on the evaluated pixels. It does not establish conditional coverage for every case, angle, or anatomical structure. It does not measure the diameter of {Pa′(V):V∈F}\{P_{a'}(V):V\in\mathcal{F}\}. The acquired-angle gate checks the boundary where the restricted push-forward is a point. Passing this check does not certify that an interval contains every compatible projection. An undersized scale incurs the miss term whenever a held-out intensity falls outside the interval, regardless of its point-estimate PSNR. Small uu with coverage holding means μ\mu is a usable point readout. Large uu indicates predictive spread

within the learned anatomy class, which a mean image alone omits. Small uu with failed coverage indicates an overconfident predictor, as in the dental-to-chest case. Pixelwise independence is a modeling choice, and a draw from (2) is not a projection of a volume in F\mathcal{F}. A volume readout V⋆V^\star would also have to match yay_a on A\mathcal{A} and land in a high-probability region of pp at the query. § 4 records a volume readout that leaves the fiber.

Residual-scale learning and collapse

Backbone. Figure 2(a) is this inference step. Each acquired projection is a 224 × 224 cone-beam image. A frozen MAE Vision Transformer (ViT-B/16) encoder [12] tokenizes it. An action-conditioned transformer fuses the two nearest acquired views by distance-weighted softmax at temperature 0.10, conditioned on the signed angle from the query. One head reads the fused tokens as the predictive mean μ(a′)\mu(a'). Point completion is that mean path trained with an ℓ1\ell_1 projection loss and no scale head. A second head reads the same tokens as a per-pixel log-scale, initialized near log 0.08 and clamped so that σ∈[e−6,0.35]\sigma\in[e^{-6},0.35]. This scale head is the only module added to the point-completion backbone. It is what allows the published width to vary with δ\delta. The clamp stops the scale from saturating, and the log term in the likelihood penalizes inflation. At test time KK is fixed at 8 and every unobserved angle on the 50-view orbit is queried. δ\delta is computed from every view in A\mathcal{A}. The other views constrain F\mathcal{F} and do not enter the features. Near-angle mixing is off, so a small gap does not replace μ\mu by a copy of the nearest view.

Held-out likelihood. Figure 2(b) draws KK from 6 to 10, takes a sparse subset with probability 0.55 and a multi-arc subset otherwise, holds out four query views, and drops 15% of the acquired views. Chest and dental are separate specialists. The held-out target is scored with the Gaussian negative log-likelihood (NLL)

LNLL=12E[log⁡σ2+(y−μ)2/σ2].(8)\mathcal{L}_{\mathrm{NLL}} = \frac{1}{2}\mathbb{E}\left[\log\sigma^2 + (y-\mu)^2/\sigma^2\right]. \tag*{(8)}

A light ℓ1\ell_1 and gradient penalty on μ\mu (λ=0.15\lambda=0.15 and 0.05) keeps the mean from drifting while the scale is learning. As marked in Figure 2(b), the scale has no explicit variance target. For a fixed conditioning input and mean, minimizing the expected NLL over an unconstrained scale yields σi2=E[(Yi−μi)2]\sigma_i^2=\mathbb{E}[(Y_i-\mu_i)^2]. Optimizing the mean as well recovers the conditional mean and variance when finite and representable. The scale therefore responds to both residual variability and mean bias. It is not a separately identified epistemic variance. NLL trains the density parameters, whereas ACUS evaluates its interval endpoints. The auxiliary penalties and scale clamp further separate the fitted solution from this population optimum. The squared standardized residual favors larger scales for difficult predictions, while the logarithmic term charges indiscriminate inflation.

Collapse on acquired views. On an acquired angle the push-forward must equal the measurement, even when many volumes remain in F\mathcal{F}. Figure 2(b) enforces that boundary by presenting an acquired view as a query:

Lcol=∥μ(a~)−ya~∥1+E[σ(a~)],a~∈A.(9)\mathcal{L}_{\mathrm{col}} = \lVert\mu(\tilde{a})-y_{\tilde{a}}\rVert_1 + \mathbb{E}[\sigma(\tilde{a})], \qquad\tilde{a}\in\mathcal{A}. \tag*{(9)}

These two terms are the arrows in Figure 2(b): μ\mu is pulled toward the acquired projection and σ\sigma toward zero. The loss is soft, so a finite error and the positive scale floor remain. We check the resulting scale and mean error numerically rather than asserting exact collapse. With measurement noise, collapse would describe consistency with the observed sample, not uncertainty about its latent noise-free value. The present gate is interpreted within the simulated measurement protocol. Acquired-view mean error and scale are checked separately from the unmeasured-view results. The total loss weights LNLL\mathcal{L}_{\mathrm{NLL}} and Lcol\mathcal{L}_{\mathrm{col}} equally. Each specialist runs for 8,000 steps at batch size 1, from a point-completion initialization, with encoder learning rate zero.

What a calibrated interval means. Coverage is the fraction of detector pixels lying in μ±1.645σ\mu\pm1.645\sigma. Residuals are not Gaussian, so the σ\sigma that minimizes (8) need not hit the nominal 90%. We report untempered scales. Temperature scaling would need separate calibration data and a joint

Input: acquired set A\mathcal{A} with views yAy_{\mathcal{A}}, query angle a′a' with held-out projection yy, predictor (μ,σ)(\mu,\sigma), level $\alpha=0.1. |
Output: ACUS(a′)(a') and mean scale u(a′)u(a').
1: δ(a′)←min⁡a∈A∣a′−a∣\delta(a')\leftarrow\min_{a\in\mathcal{A}}|a'-a|
2: (μ,σ)←Predict⁡(a′,A)(\mu,\sigma)\leftarrow\operatorname{Predict}(a',\mathcal{A}) using the two nearest views
3: L←μ−1.645σ,U←μ+1.645σL\leftarrow\mu-1.645\sigma,\quad U\leftarrow\mu+1.645\sigma
4: for each detector pixel ii do
5: Si←Ui−LiS_i\leftarrow U_i-L_i. If yi∉[Li,Ui]y_i\notin[L_i,U_i] then Si←Si+2αdiS_i\leftarrow S_i+\frac{2}{\alpha}d_i
6: end for
7: u(a′)←mean⁡(σ)u(a')\leftarrow\operatorname{mean}(\sigma). Then ACUS(a′)←mean⁡(S)(a')\leftarrow\operatorname{mean}(S)
8: gate: if a′∈Aa'\in\mathcal{A} then require u(a′)≈0u(a')\approx0 and μ≈ya′\mu\approx y_{a'}

Table 3. Computing ACUS for one query angle.

check against the width charge in (7). The static scale in Figure 1(c) is this failure: a scale fit on near residuals pays the miss term on far pixels.

Experiments

Data and protocol. Chest slices come from LUNA16 [17] and dental volumes from ToothFairy [5, 8]. Both are rendered onto the same kind of circular cone-beam orbit: 50 views over 180°. We train one specialist per anatomy. Eight cases per domain set the constant residual scale of the copy baselines. The next 40 cases are the test set. Every unobserved orbit angle is a query. Near queries sit less than 20° from the closest acquired view. Far queries sit at least 40° away. The main tables use K=8K=8. Metrics are mean ±\pm standard deviation over cases. Figure 5 shows the sparse protocol on both anatomies, with unobserved views colored by ACUS.

Baselines. Nearest-neighbor and linear angular interpolation copy local measurements. Point completion is the same backbone trained to match the held-out projection in ℓ1\ell_1. None of the three predicts a scale. We give each a single scale equal to its residual on the eight calibration cases, and we use that constant at every angle. Their acquired-view error is zero by construction, because a query that hits an acquired angle returns the measurement. The interval model must predict that view. We therefore report acquired-view width and PSNR only for ACUS, and we do not treat an exact copy as a competing interval.

Main result. On both anatomies and both orbit families, the interval is narrow near an acquired view and several times wider far from one (Tabs. 1, 2 and 4). On chest sparse orbits the mean scale grows from 0.018 to 0.088, and PSNR of the predictive mean falls from 32.48 dB to 18.64 dB. Coverage is 0.81/0.87 (chest sparse), 0.77/0.80 (chest multi-arc), 0.92/0.86 (dental sparse), and 0.89/0.91 (dental multi-arc). Constant-scale baselines have zero reported far coverage, and acquired-view width is 0.002 in every main cell. Dental multi-arc point completion is slightly ahead on far PSNR (22.47 versus 22.37 dB) and still has zero reported far coverage. Figure 3 shows those far queries: constant-scale baselines leave most pixels outside the interval, while ACUS keeps a structured scan and a sparse

miss mask. Near an acquired view the methods look similar, and the break appears in the far columns. Figure 4 repeats the far query for FDK [10] and SART [1], reprojected from the observed orbit, and for SAX-NeRF, R2^2-Gaussian, NAF, and X-Gaussian [6, 7, 25, 26]. Analytical reprojections lose structure. The learned images stay sharp, and one calibration scale makes their miss masks follow the copy baselines. Supplement, Sec. D gives the near/far grids, and Secs. E to H collect further boards. On chest sparse orbits the copy baselines use constant scales 0.014, 0.012, and 0.013, with near coverage 0.61, 0.70, and 0.72. ACUS uses 0.018 near an acquired view and 0.088 on far queries, and its far PSNR is 18.64 dB against 16.40, 16.34, and 14.14 dB. Dental sparse orbits repeat that split at a smaller far scale, from 0.024 to 0.061, with far PSNR 21.12 dB against 18.51, 18.54, and 20.95 dB. Under multi-arc packing the chest near scales already agree (0.022, 0.022, and 0.026 against 0.023), while the far scale opens to 0.095 and far PSNR is 17.48 dB against 15.51, 15.28, and 13.59 dB. Dental multi-arc scales are 0.027/0.063 at coverage 0.89/0.91. Case-level scale deviations on these ACUS rows are at most 0.003, except 0.008 on chest multi-arc far queries.

Figure 4. Same far-query protocol with classical and learned reconstructors. (a) Chest. (b) Dental. Each method shows prediction, color residual, and miss mask with a shared colorbar. Feldkamp-Davis-Kress (FDK) [10] and Simultaneous Algebraic Reconstruction Technique (SART) [1] reconstruct the observed orbit and are reprojected at the query angle. Structure-Aware X-ray Neural Radiodensity Fields (SAX-NeRF), R2^{2}-Gaussian, Neural Attenuation Fields (NAF), and X-Gaussian use the same orbit geometry [6, 7, 25, 26].

Figure 8 and Table 4 split the official sparse set into angular bins. On these sparse bins, coverage stays between 0.79 and 0.87 on chest and between 0.86 and 0.93 on dental. On chest sparse orbits the mean scale steps through 0.014, 0.027, 0.055, 0.079, and 0.088, and PSNR falls from 33.88 dB to 18.64 dB, with the largest step from 29.49 dB to 23.11 dB between the 10∘10^\circ to 20∘20^\circ and 20∘20^\circ to 30∘30^\circ bins. Dental sparse scales rise more slowly, through 0.021, 0.030, 0.047, 0.057, and 0.061. The multi-arc bins are separate: chest coverage there is 0.750 in the 0∘0^\circ to 10∘10^\circ row, and the 60∘60^\circ to 180∘180^\circ row has limited case support. Chest multi-arc coverage is 0.800 in the 40∘40^\circ to 60∘60^\circ bin, below 0.859 and 0.868 in the two middle bins. That tail row is u=0.111u=0.111, coverage 0.804, and 16.68 dB on chest, and u=0.068u=0.068, coverage 0.914, and 21.59 dB on dental. The pooled multi-arc far cells sit next to the 40∘40^\circ to 60∘60^\circ rows rather than on that thin tail (Supplement, Sec. C). Supplement, Sec. C also shows a 32.4° case from the 30∘30^\circ to 40∘40^\circ bin, which is not a far query. Figure 6 draws the same score on the acquisition ring. The central volume is a spatial anchor.

Gap (°)Chest, sparse uuChest, sparse CoverChest, sparse PSNRDental, sparse uuDental, sparse CoverDental, sparse PSNR
0 to 100.014±0.0010.791±0.06933.88±1.560.021±0.0010.931±0.02433.17±1.04
10 to 200.027±0.0050.848±0.06329.49±2.470.030±0.0020.911±0.04329.15±1.83
20 to 300.055±0.0090.851±0.06623.11±1.710.047±0.0040.900±0.04724.71±1.98
30 to 400.079±0.0040.833±0.08719.18±1.500.057±0.0010.860±0.05122.00±0.57
40 to 600.088±0.0030.869±0.05218.64±0.990.061±0.0020.858±0.08321.12±2.04
Gap (°)Chest, multi-arc uuChest, multi-arc CoverChest, multi-arc PSNRDental, multi-arc uuDental, multi-arc CoverDental, multi-arc PSNR
0 to 100.013±0.0020.750±0.06232.88±1.300.021±0.0010.914±0.02932.70±1.27
10 to 200.034±0.0050.785±0.06625.75±1.650.034±0.0020.853±0.05526.38±1.40
20 to 300.057±0.0080.859±0.05323.15±1.550.047±0.0030.862±0.06424.16±1.51
30 to 400.075±0.0120.868±0.06221.10±2.310.055±0.0030.870±0.06722.85±1.68
40 to 600.094±0.0080.800±0.05417.54±0.530.062±0.0020.907±0.02522.41±0.81
60 to 1800.111±0.0000.804±0.00016.68±0.000.068±0.0000.914±0.00021.59±0.00

Table 4. Official 40-case K=8K=8 angular bins for ACUS. uu is mean scale, not interval length. Cover targets 0.90. PSNR is of the predictive mean, in dB. Sparse protocols have no 60° to 180° bin. Multi-arc far columns in Table 2 pool the 40° to 60° and 60° to 180° bins. The last row has limited case support.

Figure 6. Orbit rings with ACUS-colored borders and a central volume rendering. (a) Chest. (b) Dental. Cold borders sit near acquired views. Warm borders sit on far queries. The central volume is a spatial anchor, not a reconstruction in the measurement fiber.

Figure 8. Sparse K=8K=8 gap bins on the official 40-case split. (a) Chest. (b) Dental. Curves show mean scale, empirical coverage, and PSNR of the predictive mean. The 40° to 60° bin is the far column of Table 1. Error bars are case-level standard deviations, and the dashed line is the nominal 90% rate. Mean ACUS is omitted because the outside-distance term is unreported.

Collapse ablation. Removing LcolL_{\mathrm{col}} on a 20-case multi-arc split, separate from the 40-case tables, leaves far width unchanged at 0.095 (chest) and 0.071 (dental) and raises acquired-view width from 0.002 to 0.018 (chest) and 0.015 (dental) (Figure 7(a)). The term is tied to acquired views. The dental far width on this split is not the main-table value 0.063. Far coverage on the same split moves from 0.80 to 0.79 on chest and from 0.90 to 0.89 on dental. On the official cells, predictive PSNR at acquired angles is 46.4 and 52.6 dB on chest and 49.5 and 54.9 dB on dental, for sparse and multi-arc orbits, at the same acquired width 0.002. Supplement, Sec. I gives the ablation rows and the acquired-view PSNR on all four official cells.

Figure 7. Acquired-view collapse and cross-anatomy far coverage on multi-arc queries. (a) On the 20-case multi-arc ablation split, distinct from the 40-case main tables, removing Lcol\mathcal{L}_{\mathrm{col}} raises the acquired scale from 0.002 to 0.018 (chest) and 0.015 (dental). Far-view scales remain 0.095 (chest) and 0.071 (dental). (b) Far coverage for official 40-case same-domain cells and the official transfer evaluation. The dental-to-chest collapse is a negative control, not a transfer success.

Where the width is anatomy, not geometry. A chest model on dental multi-arc scans keeps a wide far interval (0.093±\pm0.013), coverage 0.84, and far PSNR 19.59±\pm2.08. A dental model on chest scans keeps far width 0.061±\pm0.001, coverage 0.59, and far PSNR 16.58±\pm1.12, against about 0.80 to 0.87 in-domain (Figure 7(b)). Dental-to-chest is the negative control: the narrow interval is an anatomical prior. Supplement, Sec. G shows the corresponding transfer panels.

Long-arc stress test. On 16 cases with one arc of 50∘50^\circ to 120∘120^\circ, far width stays large (0.090 chest, 0.064 dental; near/far PSNR 32.45/18.68 and 31.25/21.28; far coverage 0.82 and 0.88). Near width is 0.027 on both anatomies, against far widths 0.090 and 0.064. The check uses another acquisition family and is not a matched 40-case estimate (Supplement, Sec. J).

Volume readout, not a result. On 12 sparse cases at 64364^3, Figure 9 compares a free residual refinement with sparse FDK and the ground-truth volume. Volume PSNR rises (chest 14.04±\pm1.11 vs. 6.82±\pm0.36, dental 18.26±\pm0.82 vs. 9.41±\pm0.56) and agreement with the acquired projections falls (chest 9.79±\pm1.10 vs. 13.32±\pm1.41, dental 9.13±\pm0.87 vs. 10.41±\pm0.99). The refinement was not trained with query-angle likelihood, and the FDK intensities are not on the measurement scale. Only 0.07 to 0.15 of the query-angle forward projections fall inside the predicted intervals, so the diagnostic does not yield a volume in FF. Supplement, Sec. H pairs these slices with the reprojection curves.

Figure 9. Official 64364^3 slices side by side in one column, chest on the left and dental on the right. (a) Chest. (b) Dental. Columns are axial, coronal, and sagittal. Rows are the ground-truth volume, sparse FDK, and a free residual refinement that can look closer to the anatomy while agreeing less with the acquired projections, and does not yield a volume in F\mathcal{F} (§ 4).

Discussion

The tables separate point-estimate accuracy from interval reliability. Near acquired support, accurate local baselines coexist with a small learned scale, while far away the constant scales have zero reported coverage and the learned scales open. Dental multi-arc point completion leads slightly on far PSNR and still misses the far interval. One residual scale leaves angle-conditioned calibration, a trained ensemble, and a conformal interval on the same outputs untested, and acquired-view collapse leaves far coverage unchecked.

Chest coverage stays below the nominal rate, so further calibration is required before the 90% level is an operating guarantee. Correlated misses can sit on a small structure and barely move the detector average, so pixelwise coverage and the diagonal Gaussian fall short of a simultaneous confidence region for a projection or a volume.

On the far-query boards, FDK and SART lose structure, while SAX-NeRF, R2R^2-Gaussian, NAF, and X-Gaussian stay sharp, and one calibration scale makes their miss masks follow the copy baselines (Figure 4). Under (7), a covered query scores 3.29 uu and every miss adds twenty times its outside distance. Learned rows omit that distance, so zero far coverage activates the penalty without an ACUS rank. A scalar is printed only for the analytical repro­jectors (Supplement, Sec. J); Sec. E splits it into width and miss.

On 12 sparse cases, only 0.07 to 0.15 of the query-angle forward projections fall inside the predicted intervals (Figure 9). Volume similarity can rise while the estimate leaves F\mathcal{F}.

On the shared 50-view, 180° trajectory, a chest model on dental scans keeps coverage 0.84 with a wide far interval, while a dental model on chest scans keeps a far width near 0.061 and coverage 0.59. The opened scale follows the training anatomy. Pooled chest coverage of 0.77 to 0.80 on multi-arc queries is the in-domain counterpart, while the 0° to 10° bin in Table 4 is 0.750.

The present evidence supports a practical audit of gap-conditioned scale and coverage, the acquired boundary, and the miss magnitude when an aggregate interval score is published.

Conclusion

We present ACUS, an angular calibrated uncertainty score for incomplete-view CT. ACUS assigns a nominal predictive interval on held-out projections by distance to the nearest acquired view, with acquired-view collapse as a necessary consistency check. A feed-forward mean and scale trained with held-out likelihood gate wider intervals at larger gaps. ACUS uses a diagonal Gaussian on detector pixels and is evaluated on simulated orbits. Cross-anatomy transfer is a negative control, not a domain-generalization claim. Constant-scale point predictors fail far bins despite strong near-angle PSNR.

LLM Usage Statement

ChatGPT assisted with manuscript editing, LaTeX formatting, figure typography, and checks of references. The authors take responsibility for the content and results of this work.

Broader impact and asset notes

ACUS is an interval-evaluation tool for incomplete-view computed tomography (CT) protocols. It scores a finished acquisition. It is not an acquisition policy, a stopping rule, or a certificate that the mean image is the only volume consistent with the measurements. Section H is the concrete form of that limit: volume PSNR can rise while agreement with the acquired projections falls, and only 0.07 to 0.15 of the query-angle forward projections fall inside the predicted intervals.

Assets. Chest slices come from LUNA16 [17] and dental volumes from ToothFairy [5, 8] under their published licenses. All orbits are simulated cone-beam renderings on one 50-view, 180° geometry, so a transfer experiment changes the object and the specialist, not the angular grid. No patient-identifying metadata appears in the figures. A single displayed case is evidence for the 40-case, 20-case, 16-case, or 12-case table named in its caption.

Protocol details beyond the main text

Shared geometry. Both domains are rendered onto a circular cone-beam orbit of 50 views spanning 180°. Every query is one of those rendered views that the mask left out, so the evaluation never leaves the simulated grid. The gap δ(a′)\delta(a') is the minimum absolute distance on this sampled trajectory from the query to a view that the mask kept. Endpoints of the 180° arc are not glued together. A query near one end is far from a view near the other end even though both sit on the same semicircle. This matches the definition in the main text and is why the multi-arc tables contain a 60° to 180° bin that the sparse tables do not. Sparse sampling spreads eight views around the semicircle, so the largest hole stays inside 40° to 60°. Packing the same eight views into two or three short arcs opens at least one much larger hole, and that hole is the last row of Table 6 and Table 8.

Chest multi-arc bins (40 cases), from the same official split as the main multi-arc table. uu is mean scale, not interval length. Far queries aggregate the 40∘40^\circ to 60∘60^\circ and 60∘60^\circ to 180∘180^\circ bins.

GapuuCoverPSNR
0∘0^\circ to 10∘10^\circ0.013±0.0020.013\pm0.0020.750±0.0620.750\pm0.06232.88±1.3032.88\pm1.30
10∘10^\circ to 20∘20^\circ0.034±0.0050.034\pm0.0050.785±0.0660.785\pm0.06625.75±1.6525.75\pm1.65
20∘20^\circ to 30∘30^\circ0.057±0.0080.057\pm0.0080.859±0.0530.859\pm0.05323.15±1.5523.15\pm1.55
30∘30^\circ to 40∘40^\circ0.075±0.0120.075\pm0.0120.868±0.0620.868\pm0.06221.10±2.3121.10\pm2.31
40∘40^\circ to 60∘60^\circ0.094±0.0080.094\pm0.0080.800±0.0540.800\pm0.05417.54±0.5317.54\pm0.53
60∘60^\circ to 180∘180^\circ0.111±0.0000.111\pm0.0000.804±0.0000.804\pm0.00016.68±0.0016.68\pm0.00

Table 6.

Dental multi-arc bins (40 cases), from the same official split as the main multi-arc table. uu is mean scale, not interval length. Far queries aggregate the 40∘40^\circ to 60∘60^\circ and 60∘60^\circ to 180∘180^\circ bins.

GapuuCoverPSNR
0∘0^\circ to 10∘10^\circ0.021±0.0010.021\pm0.0010.914±0.0290.914\pm0.02932.70±1.2732.70\pm1.27
10∘10^\circ to 20∘20^\circ0.034±0.0020.034\pm0.0020.853±0.0550.853\pm0.05526.38±1.4026.38\pm1.40
20∘20^\circ to 30∘30^\circ0.047±0.0030.047\pm0.0030.862±0.0640.862\pm0.06424.16±1.5124.16\pm1.51
30∘30^\circ to 40∘40^\circ0.055±0.0030.055\pm0.0030.870±0.0670.870\pm0.06722.85±1.6822.85\pm1.68
40∘40^\circ to 60∘60^\circ0.062±0.0020.062\pm0.0020.907±0.0250.907\pm0.02522.41±0.8122.41\pm0.81
60∘60^\circ to 180∘180^\circ0.068±0.0000.068\pm0.0000.914±0.0000.914\pm0.00021.59±0.0021.59\pm0.00

Table 8.

Why the two headline families share a view budget. Sparse K=8K=8 picks eight angles on the 50-view circle. Multi-arc K=8K=8 packs the same budget into two or three short arcs of 25° to 45°. Holding KK fixed keeps angular arrangement from being confounded with the number of measurements. The two families are therefore reported separately at the same δ\delta, including in the supplement tables.

Near, far, and the bins underneath them. Near queries sit under 20° from the closest acquired view. Far queries sit at least 40° away. Angles from 20° to 40° are scored and are excluded from both headline columns. Figure 11 is one such case, at 32.4°. Reported means average per-case query means, and the standard deviations are case-level. Neighboring detector pixels share anatomy, so these deviations are not pixelwise standard errors.

Figure 11. Held-out 32.4° mid-range intervals in the 30° to 40° bin, not a far query of at least 40°. Panels show ground truth (GT), linear interpolation, the ACUS mean, a scale map, the 90% bounds, the residual, and outside-interval pixels.

Calibration of copy baselines. Nearest-neighbor, linear interpolation, and point completion receive one residual scale fit on eight calibration cases per domain. Those eight cases are disjoint from the 40-case test split. The fitted scale is one positive number per method and domain, held fixed at every query angle. This is the constant-scale failure the miss penalty charges. It is not an angle-conditioned calibrator, an ensemble, or a conformal interval on the same outputs. Acquired-view error of a copy baseline is zero by construction, because a query that lands on an acquired angle returns the measurement. The interval model has to predict that view from its neighbors. We therefore report acquired-view width and predicted PSNR only for ACUS. An exact copy is not entered as a competing interval in Table 14.

DomainOrbitAcq. widthPred. PSNR on acquired
ChestSparse0.00246.4
ChestMulti-arc0.00252.6
DentalSparse0.00249.5
DentalMulti-arc0.00254.9

Table 14. Acquired-view collapse check for ACUS (main K=8K=8 cells).

External reconstructors. Structure-Aware X-ray Neural Radiodensity Fields (SAX-NeRF) [7], R2R^2-Gaussian [26], Neural Attenuation Fields (NAF) [25], and X-Gaussian [6] are evaluated on the same far-query geometry used for the ACUS grids. Feldkamp-Davis-Kress (FDK) [10], the Simultaneous Algebraic Reconstruction Technique (SART) [1], and adaptive-steepest-descent projection onto convex sets (ASD-POCS) [18] reconstruct the observed orbit and are reprojected at the query. Each method returns a point image. We attach the same residual visualization used for the copy baselines so that miss masks are comparable across columns. These methods were trained or derived as point reconstructors, not as interval estimators. The scalar ACUS printed in Table 9 is computed only for the three analytical reprojections, which is the statement in the main discussion. Learned reconstructors appear as images and miss masks. They are not given a second, specially tuned scale in this supplement.

Same-geometry analytical reconstructors on chest: Feldkamp-Davis-Kress (FDK) [10], Simultaneous Algebraic Reconstruction Technique (SART) [1], and adaptive-steepest-descent projection onto convex sets (ASD-POCS) [18].

MethodScaleCover near/farPSNR near/farACUS
FDK0.1440.57/0.5011.13/11.600.739
SART0.0860.60/0.5015.58/14.510.509
ASD-POCS0.0870.60/0.5015.57/14.510.509

Table 9.

Gap-conditioned evaluation

The official 40-case bins also appear in the main-text experiments. The same values are repeated here in Table 5 to Table 8. Figure 10 plots those four tables together. The mechanism the main paper claims is visible without pooling. Mean scale rises monotonically with the gap on every official split: chest sparse from 0.014 to 0.088, chest multi-arc from 0.013 to 0.111, dental sparse from 0.021 to 0.061, and dental multi-arc from 0.021 to 0.068. Coverage does not follow that rise. It stays inside 0.75 to 0.93, below the nominal 90% rate on chest and closer to it on dental. The point-estimate PSNR falls over the same bins, which is the separation ACUS is built to report: a wider interval can remain approximately calibrated while the mean image becomes a poorer point estimate. A static scale cannot trace this curve, because it has one magnitude at every gap.

Figure 10. Official K=8K=8 bins, 40 cases per domain. Bands are case-level standard deviations. The dashed line is the nominal 90% rate. The 60° to 180° chest and dental multi-arc points have no case-level deviation because that bin contains one query pattern in the official split. These curves use the printed supplement tables, not a later rerun.

GapuuCoverPSNR
0° to 10°0.014±0.0010.014 \pm0.0010.791±0.0690.791 \pm0.06933.88±1.5633.88 \pm1.56
10° to 20°0.027±0.0050.027 \pm0.0050.848±0.0630.848 \pm0.06329.49±2.4729.49 \pm2.47
20° to 30°0.055±0.0090.055 \pm0.0090.851±0.0660.851 \pm0.06623.11±1.7123.11 \pm1.71
30° to 40°0.079±0.0040.079 \pm0.0040.833±0.0870.833 \pm0.08719.18±1.5019.18 \pm1.50
40° to 60°0.088±0.0030.088 \pm0.0030.869±0.0520.869 \pm0.05218.64±0.9918.64 \pm0.99

Table 5. Chest sparse bins (40 cases), from the same official K=8K=8 split as the main sparse table. uu is mean scale, not interval length. The 40° to 60° row is the far column of that table.

Figure 11 shows one chest case and one dental case at a mid-range gap of 32.4°, which lies in the 30° to 40° bin and is not a far query of at least 40°. The columns follow the same convention as the main far-query boards: ground truth, linear interpolation, the ACUS mean, the scale map, the published 90% bounds, the residual, and the miss mask. The scale maps are already structured at this mid-range gap, and the miss masks are not uniform across the detector.

How the sparse bins sit under the headline cells. On chest sparse orbits, scale rises from 0.014 in the 0°–10° bin to 0.088 in the 40°–60° bin (Table 5). The rise is monotone across the five bins: 0.014, 0.027, 0.055, 0.079, 0.088. The largest single step is from the 10°–20° bin to the 20°–30° bin, where the mean scale roughly doubles and PSNR of the predictive mean falls from 29.49 dB to 23.11 dB. That is the same transition the protocol uses to leave the near group. The pooled near cell in main Table 1 is u=0.018u=0.018, coverage 0.81, PSNR 32.48 dB. It mixes the 0°–10° bin (u=0.014u=0.014, coverage 0.791, PSNR 33.88 dB) with the 10°–20° bin (u=0.027u=0.027, coverage 0.848, PSNR 29.49 dB). The pooled cell and the first bin are different summaries. The far cell u=0.088u=0.088, coverage 0.87, PSNR 18.64 dB matches the 40°–60° row exactly, because sparse orbits have no wider bin. Dental sparse scale rises more slowly and saturates: 0.021, 0.030, 0.047, 0.057, 0.061 (Table 7). The far cell 0.061 is again the last bin. The pooled dental near cell 0.024 likewise sits between the first bin 0.021 and the second bin 0.030. Far PSNR remains higher than on chest (21.12 dB against 18.64 dB) at a narrower scale (0.061 against 0.088). The main conclusion that the opened scale follows the training anatomy is already visible inside these two in-domain curves, before the transfer table is consulted.

Methoduu near/farCover near/farPSNR near/far
Chest
Nearest0.014/0.0140.61/0.0030.27/16.40
Linear0.012/0.0120.70/0.0032.34/16.34
Point (ℓ1\ell_1)0.013/0.0130.72/0.0032.17/14.14
ACUS0.018/0.0880.81/0.8732.48/18.64
Dental
Nearest0.014/0.0140.67/0.0030.24/18.51
Linear0.012/0.0120.76/0.0031.72/18.54
Point (ℓ1\ell_1)0.011/0.0110.73/0.0031.78/20.95
ACUS0.024/0.0610.92/0.8631.90/21.12

Table 1. Sparse K=8K=8, 40 test cases per domain. Entries are near/far. uu is mean scale, not interval length. Cover targets 0.90. Point denotes point completion. ACUS scale standard deviations (near/far): Chest 0.003/0.003, Dental 0.002/0.002.

Dental sparse bins (40 cases), from the same official split as the main sparse table. uu is mean scale, not interval length. The 40∘40^\circ to 60∘60^\circ row is the far column of that table.

GapuuCoverPSNR
0∘0^\circ to 10∘10^\circ0.021±0.0010.021\pm0.0010.931±0.0240.931\pm0.02433.17±1.0433.17\pm1.04
10∘10^\circ to 20∘20^\circ0.030±0.0020.030\pm0.0020.911±0.0430.911\pm0.04329.15±1.8329.15\pm1.83
20∘20^\circ to 30∘30^\circ0.047±0.0040.047\pm0.0040.900±0.0470.900\pm0.04724.71±1.9824.71\pm1.98
30∘30^\circ to 40∘40^\circ0.057±0.0010.057\pm0.0010.860±0.0510.860\pm0.05122.00±0.5722.00\pm0.57
40∘40^\circ to 60∘60^\circ0.061±0.0020.061\pm0.0020.858±0.0830.858\pm0.08321.12±2.0421.12\pm2.04

Table 7.

Coverage does not track scale. Chest sparse coverage across the five bins is 0.791, 0.848, 0.851, 0.833, 0.869. It stays inside the band the main text quotes for the sparse chest curve, between 0.79 and 0.87, and it is not monotone in the gap. The 30°–40° bin is the local dip (0.833±0.087), and it is also the bin with the widest case-level spread in that table. Dental sparse coverage is 0.931, 0.911, 0.900, 0.860, 0.858. It starts near the nominal rate and drifts down to the mid-0.80s, which is why the headline dental sparse pair can be 0.92 near and 0.86 far without either number being a summary of every bin. Under main (7) a fully covered detector scores 3.29 uu, and each missed pixel adds twenty times its outside distance. The mean outside distance is unreported, so these rows cannot be ranked by a scalar ACUS.

Multi-arc bins and the thin last row. Chest multi-arc scale continues the same monotone opening, and it does not stop at 60°: 0.013, 0.034, 0.057, 0.075, 0.094, 0.111 (Table 6). The headline far cell 0.095 pools the 40°–60° row (0.094, coverage 0.800, PSNR 17.54 dB) with the 60°–180° row (0.111, coverage 0.804, PSNR 16.68 dB). An equal mixture of those two rows would sit near 0.102. The published pool sits at 0.095, next to the 40°–60° row, so the thin last bin does not dominate the headline far cell even though its own scale is the widest in the paper. Dental multi-arc scale is 0.021, 0.034, 0.047, 0.055, 0.062, 0.068, and the headline far cell 0.063 is the corresponding pool of 0.062 and 0.068 (Table 8). Dental coverage on those two far bins is 0.907 and 0.914, which is why the pooled far coverage in main Table 2 can sit at 0.91 even though several nearer dental bins are in the mid-0.80s. The last row prints a case-level standard deviation of 0.000 because support is too thin to show scatter. That zero is not a population variance, and 16.68 dB or 21.59 dB is not a high-precision estimate. The chest multi-arc 0°–10° bin is the other pooled quantity that needs the rule underneath it: coverage 0.750±0.062 at scale 0.013, against a headline near cell of coverage 0.77 at scale 0.023. The far cell is 0.80. Undercoverage relative to 0.90 is present on both sides of the gap, and the lowest rate is the nearest bin, where PSNR is highest (32.88 dB).

Methoduu near/farCover near/farPSNR near/far
Chest
Nearest0.022/0.0220.74/0.0029.05/15.51
Linear0.022/0.0220.71/0.0029.18/15.28
Point (ℓ1\ell_1)0.026/0.0260.77/0.0028.79/13.59
ACUS0.023/0.0950.77/0.8029.59/17.48
Dental
Nearest0.020/0.0200.79/0.0029.20/19.83
Linear0.021/0.0210.77/0.0029.13/19.80
Point (ℓ1\ell_1)0.018/0.0180.70/0.0029.55/22.47
ACUS0.027/0.0630.89/0.9129.78/22.37

Table 2. Multi-arc K=8K=8, 40 test cases per domain. Entries are near/far. uu is mean scale, not interval length. Cover targets 0.90. Point denotes point completion. ACUS scale standard deviations (near/far): Chest 0.003/0.008, Dental 0.001/0.002. Acquired scale is 0.002 in both domains (PSNR 52.6/54.9 dB).

The 32.4° case belongs to the 30°–40° row: chest sparse scale 0.079, coverage 0.833, PSNR 19.18 dB; dental sparse scale 0.057, coverage 0.860, PSNR 22.00 dB. Those means sit between the near scales near 0.02 and the chest far scale 0.088. The panel should be cited with that row, not with the far column of main Table 1.

Figure 12 scores acquired angles on the 20-case multi-arc ablation, not the unobserved angles in the bins above. With Lcol\mathcal{L}_{\mathrm{col}}, acquired width is 0.002. Without it, acquired width rises to 0.018 (chest) and 0.015 (dental), while far width stays at 0.095 and 0.071 (Table 10). Chest far width 0.095 matches the 40-case multi-arc cell. Dental far width 0.071 does not match the 40-case cell 0.063. The two splits stay separate.

Figure 12. Acquired-view collapse on the 20-case multi-arc ablation split. The panel is the same acquired-view collapse diagnostic as main Figure 7(a). The values agree with Table 10.

Collapse ablation on a 20-case multi-arc split.

DomainSettingFar widthAcq. widthFar cover
Chestwith Lcol\mathcal{L}_{\mathrm{col}}0.0950.0020.80
Chestwithout Lcol\mathcal{L}_{\mathrm{col}}0.0950.0180.79
Dentalwith Lcol\mathcal{L}_{\mathrm{col}}0.0710.0020.90
Dentalwithout Lcol\mathcal{L}_{\mathrm{col}}0.0710.0150.89

Table 10.

Near versus far medical grids

On a chest sparse near query the four means span 2.2 dB: 30.27, 32.34, 32.17, and 32.48 for nearest, linear, point completion, and ACUS. The far block is where they separate. The chest sparse constants are 0.014, 0.012, and 0.013, the size of the 0°–10° ACUS bin, against an ACUS far scale of 0.088. Main Table 1 records zero far coverage for every constant-scale row. In Figure 13 the far miss masks of those rows fill the detector, and the ACUS miss mask stays sparse. A full mask means the factor-of-20 term in main (7) is active on nearly every pixel. The figure does not supply the outside distance, so it supports the coverage column and not a scalar ranking.

Figure 13. Near versus far readout on one chest sparse case. Left block: near query. Right block: far query. Color panels show scale and absolute residual with a shared colorbar. Near the acquired view every method looks similar. Far away only ACUS reports a non-constant scale and keeps most pixels inside the interval.

Dental keeps a narrower absolute scale. Figure 14 uses the same layout. Dental sparse far scale is 0.061 with coverage 0.86; chest sparse far scale is 0.088 with coverage 0.87. The cooler dental map is a smaller absolute scale at similar far coverage, not a better-calibrated interval. Dental near coverage is 0.92 and chest near coverage is 0.81, so the two maps also have to be read with their own near rates. The same dental scale, evaluated on chest in Table 11, drops far coverage to 0.59.

Figure 14. Dental near versus far with color scale and residual maps and a shared colorbar. The dental specialist keeps a narrower absolute scale than the chest specialist, while far coverage remains near the nominal rate on its own domain.

Cross-anatomy transfer on multi-arc orbits.

Train → TestFar widthFar coverFar PSNR
Chest → Dental0.093±0.0130.093\pm0.0130.8419.59±2.0819.59\pm2.08
Dental → Chest0.061±0.0010.061\pm0.0010.5916.58±1.1216.58\pm1.12

Table 11.

Multi-arc packing moves which angles are local. Figure 15 is the chest multi-arc version of the same near-versus-far readout. Packing eight views into two or three short arcs changes the set of queries that fall under 20°. Inside an arc, neighboring acquired views make a query local even though most of the semicircle is empty. Outside the arcs, a query can sit 60° or more from every measurement, which sparse K=8K=8 does not do. The panel still shows the near-narrow and far-wide pattern, at chest multi-arc scale 0.023/0.095, coverage 0.77/0.80, and PSNR 29.59/17.48. Near PSNR is 2.9 dB below the sparse near value 32.48 dB: a multi-arc near query is local only to a short arc. The near scale stays small (0.023, next to the sparse near scale 0.018) while the far scale opens to 0.095. Sparse and multi-arc rows are not averaged, because their near images come from different neighborhoods.

Figure 15. Chest multi-arc near versus far with color residuals. Multi-arc packing changes which angles are local, but the same near-narrow and far-wide pattern appears.

External methods and residual mosaics

Figure 16 separates prediction, residual, and miss for the learned reconstructors on both anatomies. The analytical rows are a same-geometry point baseline, not a matched rerun of the official ACUS cells. On Table 9, FDK reaches 11.13/11.60 dB with coverage 0.57/0.50 and ACUS 0.739. SART reaches 15.58/14.51 dB with coverage 0.60/0.50 and ACUS 0.509. ASD-POCS matches SART at the displayed precision (15.57/14.51, coverage 0.60/0.50, ACUS 0.509), so the convex-set step does not separate from SART here. Far PSNR of 11.6 to 14.5 dB and far coverage 0.50 both sit below the official chest ACUS cells (17.48 to 18.64 dB, far coverage 0.80 to 0.87).

Reading the printed scalar. These three rows are the only place a scalar ACUS is attached to a point reconstructor. Main (7) splits that scalar into width and outside distance, ACUS=3.29 u+20 dˉ\mathrm{ACUS} = 3.29\,u + 20\,\bar{d}. For FDK, 3.29×0.144≈0.473.29 \times0.144 \approx0.47, so a score of 0.739 leaves about 0.26 to the miss term. For SART, 3.29×0.086≈0.283.29 \times0.086 \approx0.28, and a score of 0.509 leaves about 0.23 to the miss term. The miss terms are similar, about 0.26 and 0.23, so FDK’s larger width raises the scalar while far coverage stays at 0.50. The learned rows do not print a scalar, so the same split cannot be run on SAX-NeRF or on point completion. Zero far coverage says the miss term is active and does not say how large the outside distances are.

SAX-NeRF, R²-Gaussian, NAF, and X-Gaussian stay sharp, and one calibration scale still puts most far pixels outside the interval. Figure 17 localizes that residual on chest edges and soft-tissue boundaries. A constant scale cannot widen only at those boundaries. The ACUS scale is spatially varying, and chest far coverage stays at 0.80 to 0.87. These crops are not a per-method leaderboard and do not assign a scalar ACUS to the learned reconstructors.

Rows of Figure 18 are nearest neighbor, point completion, and ACUS; columns are the four anatomy-by-packing protocols. On dental multi-arc far queries, point completion reaches 22.47 dB and ACUS 22.37 dB, at scales 0.018 and 0.063. The tenth of a decibel favors the constant-scale mean. The miss mask does not: far coverage is reported as zero for point completion and 0.91 for ACUS. The same split appears in all four columns.

Figure 18. Residual and miss mosaic across nearest neighbor (NN), point completion, and ACUS under four protocols. Warm residual mass on far queries is the qualitative signature of an under-wide interval.

Zoom-inset medical panels

The crops keep the marked regions and the display scale of the full panels. They do not change a detector-wide rate. Chest sparse far coverage is 0.87 at PSNR 18.64 dB, 13.8 dB below the near PSNR 32.48 dB, which is why a far residual crop still shows anatomy-shaped error. Chest multi-arc far coverage is 0.80 at scale 0.095. Dental multi-arc far coverage is 0.91 at scale 0.063 and PSNR 22.37 dB. The smaller dental scale and the higher far PSNR are the same dental operating point as Table 8. The boxes were chosen to expose residual texture, not to sample the detector. A clean crop can sit inside the covered majority, and a bright crop can be one of the misses whose outside distance, multiplied by 20, would dominate the score. Either crop is compatible with the main observation that correlated misses can concentrate on a small structure.

Protocol walls and qualitative boards

Figure 22 lines up ground truth, μ\mu, scale, and absolute residual for sparse, multi-arc, and near settings on both anatomies. Near columns match near PSNR of 29.6 to 32.5 dB. Far columns match far PSNR of 17.5 to 22.4 dB. The scale row stays dark at the headline near scales 0.018 to 0.027 and opens at the headline far scales 0.061 to 0.095, with the thin chest tail at 0.111. In domain, scale and residual open together. A dark scale beside a bright residual is the constant-scale and dental-to-chest pattern, and it is absent from this wall.

Figure 23 removes the mean and the residual. Dental stays cooler than chest, 0.061 against 0.088 sparse and 0.063 against 0.095 multi-arc. Chest is still rising from the multi-arc 40°–60° bin (0.094) to the thin tail (0.111). Dental has flattened, 0.062 then 0.068. The widest mean, 0.111, is about one third of the per-pixel ceiling 0.35, so the opening is not an artifact of the clamp.

Panel (a) of Figure 24 is Table 11. Chest to dental keeps far width 0.093±0.013, coverage 0.84, and PSNR 19.59±2.08. That width sits with the chest far scales 0.088 (sparse), 0.095 (multi-arc), and 0.090 (long arc), not with the dental far scale 0.063. Dental to chest keeps far width 0.061±0.001, coverage 0.59, and PSNR 16.58±1.12. That width matches the dental sparse far scale 0.061. Coverage falls to 0.59, from in-domain chest far coverage of 0.80 on the matched multi-arc protocol and 0.87 on sparse orbits. This is the overconfident cell in the main text: small scale, failed coverage. The chest-to-dental width deviation 0.013 is larger than the in-domain chest far deviations in the main captions (0.003 sparse, 0.008 multi-arc). The dental-to-chest deviation 0.001 stays tight. Panels (b) to (d) are matched-specialist boards and are not transfer results.

Figure 24. Cross-anatomy and detailed qualitative panels. (a) Cross-anatomy transfer. (b) to (d) Domain-specific detail boards.

Figure 25 shows five further chest orbits. The predicted projections keep the large structures and lose detail, consistent with far PSNR near 18 dB rather than near PSNR near 32 dB. Five cases do not change the 40-case rates in Table 5 and Table 6.

Figure 25. Additional chest qualitative panels from held-out orbits. (a) to (e) are distinct cases, each showing a query strip, the held-out ground-truth projection, and the predicted projection. These panels are included to show breadth across cases, not to claim a new metric.

Orbit, volume, and collapse diagnostics

The orbit ring in main Figure 6 is the geometric reading of ACUS. Each surrounding projection is bordered by its score. Cold borders are near queries, at official near scales 0.018 to 0.027 and near PSNR 29.6 to 32.5 dB. Warm borders are far queries, at far scales 0.061 to 0.095 and far PSNR 17.5 to 22.4 dB. Acquired views themselves are narrower still, at scale 0.002 (Table 14). The ring is Figure 8 drawn on the gantry. The center is a spatial anchor, not a volume in the measurement fiber. Main Figure 5 marks which angles were kept.

Figure 5. Acquisition orbit schematics for chest and dental sparse protocols. Colored segments mark unobserved views on the 50-view, 180∘180^\circ orbit, using the ACUS color scale.

Figure 16. External learned methods with color residuals on chest and dental far queries. Each row is one reconstructor. Columns alternate prediction, residual, and miss for chest and dental.

On the 12 sparse cases behind main Figure 9 and Table 13, chest volume PSNR rises from 6.82±0.366.82\pm0.36 to 14.04±1.1114.04\pm1.11 and reprojection PSNR falls from 13.32±1.4113.32\pm1.41 to 9.79±1.109.79\pm1.10. Dental volume PSNR rises from 9.41±0.569.41\pm0.56 to 18.26±0.8218.26\pm0.82

DomainMethodVol. PSNRReproj. PSNRQuery interval inclusion
Chestsparse FDK6.82±0.366.82 \pm0.3613.32±1.4113.32 \pm1.41n/a
Chestfree residual refinement14.04±1.1114.04 \pm1.119.79±1.109.79 \pm1.100.07 to 0.15
Dentalsparse FDK9.41±0.569.41 \pm0.5610.41±0.9910.41 \pm0.99n/a
Dentalfree residual refinement18.26±0.8218.26 \pm0.829.13±0.879.13 \pm0.870.07 to 0.15

Table 13. Volume readout diagnostic on 12 sparse cases at 64364^3.

and reprojection PSNR falls from 10.41±0.9910.41\pm0.99 to 9.13±0.879.13\pm0.87. The gains are 7.2 dB and 8.9 dB. The losses are 3.5 dB and 1.3 dB. The refinement was not trained with query-angle likelihood, and the FDK intensities are not on the interval’s measurement scale, so the FDK inclusion cell is marked n/a. Only 0.07 to 0.15 of the refined query-angle forward projections fall inside the predicted intervals, against far coverage of 0.80 to 0.91 when the same intervals are scored on held-out ground truth. Figure 27 separates those slices from acquired-view scale maps, which belong with the acquired PSNR of 46.4 to 54.9 dB in Table 14. Figure 28 plots width against angle beside the reprojection curve that falls while volume PSNR rises. The 60°–180° row combines the largest chest scale, 0.111, with a displayed deviation of zero because the row is thin.

Figure 27. Volume-slice diagnostic and acquired-view collapse maps. (a) and (b) Axial and related slices from the 64364^{3} free residual refinement. (c) and (d) Acquired-view scale maps that should collapse near zero.

Figure 28. Width-versus-angle curve and reprojection diagnostic for the volume readout experiment.

Figure 17. External methods with residual zoom insets on chest far queries. Colored boxes mark local regions. The bottom-right zooms enlarge residual structure that a full-frame thumbnail hides.

Coverage and acquired-view diagnostics

The 77% to 92% range in the main conclusion is the eight headline cells: chest sparse 0.81/0.87, chest multi-arc 0.77/0.80, dental sparse 0.92/0.86, dental multi-arc 0.89/0.91. Bins extend past it, from chest multi-arc 0°–10° at 0.750±0.062 to dental sparse 0°–10° at 0.931±0.024. These are untempered Gaussian frequencies, not a distribution-free guarantee. The widest sparse-chest coverage spread is ±0.087 on the 30°–40° bin.

Figure 29 is the acquired-view gate in Table 14. Predictive PSNR is 46.4, 52.6, 49.5, and 54.9 dB on chest sparse, chest multi-arc, dental sparse, and dental multi-arc, which is 27.8 to 35.1 dB above the far PSNR of the same cell. Removing Lcol\mathcal{L}_{\mathrm{col}} on the 20-case split multiplies acquired width by nine on chest (0.002 to 0.018) and by 7.5 on dental (0.002 to 0.015). Far width is unchanged, and far coverage moves only from 0.80 to 0.79 and from 0.90 to 0.89. Figure 30 and Figure 31 add far-query boards under both packings. They use the same miss convention as main Figure 3 and do not change the 40-case estimates.

Figure 3. Far-angle held-out projections on chest and dental cone-beam orbits. Columns: ground truth (GT), nearest neighbor (NN), point completion, ACUS mean, ACUS scale, color residual maps for NN, point completion, and ACUS, and miss masks, with a shared residual colorbar. Constant-scale baselines miss most far pixels. ACUS keeps a structured scale and a sparse miss mask.

Figure 29. Collapse gate on acquired views for chest and dental specialists.

Figure 30. Sparse far-query detail boards for chest and dental.

Figure 31. Multi-arc far-query detail boards for chest and dental.

Figure 20. Dental multi-arc overview with local zoom insets. The dental scale is smaller in absolute value, and the residual zooms remain structured and locally organized.

Extended tables

The four bin tables reprint main Table 4 in a one-domain-per-table layout so that a curve, a mid-gap figure, and a headline cell can be checked without reconstructing the pool by hand. The remaining tables are the numerical form of the side experiments: analytical reprojection, collapse ablation, cross-anatomy transfer, the long-arc stress test, the volume readout, and the acquired-view gate. This section states how each table lines up with a sentence in the main results or the main discussion. The prose in Secs. C, H and I already walks the bin, volume, and acquired-view numbers. The paragraphs after the tables take the side experiments one at a time.

Figure 21. Chest multi-arc overview with local zoom insets. Comparing Figure 19 to Figure 21 shows that the zoom pattern tracks the gap, not the packing of the acquired arcs.

Figure 19. Chest sparse overview with local zoom insets. Top row: full panels with colored region-of-interest (ROI) boxes. Bottom row: magnified crops of ACUS mean, scale, and residual.

Figure 22. Protocol wall across eight settings. Rows: ground truth, ACUS mean, ACUS scale, absolute residual. Columns: sparse, multi-arc, near, and mixed near protocols on chest and dental.

Table 9 is read in Sec. E. Table 10 and Table 11 are read with Figure 12 and Figure 24. The long-arc table is the remaining comparison. On 16 cases and one arc of 50° to 120°, chest near/far width is 0.027/0.090, near/far PSNR is 32.45/18.68, and far coverage is 0.82. Dental near/far width is 0.027/0.064, near/far PSNR is 31.25/21.28, and far coverage is 0.88. The

Figure 23. ACUS scale heatmap wall across six protocols. The scale opens on far settings and stays dark near acquired views.

Long-arc stress test (16 cases, one arc of 50∘50^\circ to 120∘120^\circ).

DomainNear widthFar widthPSNR near/farFar cover
Chest0.0270.09032.45/18.680.82
Dental0.0270.06431.25/21.280.88

Table 12.

far scales sit with the official far scales of the same anatomy (0.088 and 0.095 on chest, 0.061 and 0.063 on dental). Sixteen cases are not pooled into main Tabs. 1 and 2. Copy baselines are absent from Table 14 because an acquired-angle query returns the measurement exactly.

Figure 26. Combined 3D orbit, volume-slice, collapse, and cross-anatomy board. The volume slice row is a negative diagnostic from the main text. Higher volume PSNR can coexist with worse agreement on acquired projections.

Gap-binned constant scales

The main tables assign nearest neighbor, linear interpolation, and point completion one residual scale, fit on the eight calibration cases and then frozen at every query. That design matches the failure drawn in main Figure 1(c): a static scale can cover near angles and miss far ones. A natural objection is that the zero far coverage is an artifact of refusing the baseline a coarser calibration. If each method may use a different positive constant in each gap bin, the miss penalty might disappear while the point predictor is left unchanged.

Figure 32 tests that objection on a later chest multi-arc rerun with the same view budget, N=40N=40 and K=8K=8. One scale per bin raises nearest-neighbor and linear far coverage from 0 to 0.381. Point completion, whose far PSNR is the highest of the three, reaches only 0.036. The learned scale on that same rerun keeps far coverage 0.829. Bin-wise constants therefore recover part of the coverage that a single constant loses, and they do not reproduce the learned coverage. The mechanism is the same one in the main score. A constant, even a different constant in each bin, has no pixelwise structure, so the pixels that sit far outside the band still dominate the miss term. The learned scale opens with the gap and varies across the detector, which is what keeps coverage near the nominal rate while the point-estimate PSNR continues to fall.

Figure 32. Chest multi-arc far coverage on the later rerun, N=40N=40, K=8K=8. Gray bars use one residual scale for every angle. Blue bars use one residual scale per gap bin. The line is the learned scale from that same rerun, not the official multi-arc cell.

On the sparse split of the same rerun the objection fails more sharply. All three binned constants still have far coverage 0. The calibration residuals in those far bins do not open the scale above the near-angle magnitude, so a gap-indexed constant collapses back to the static scale the main tables already reject.

This rerun is not the official 40-case cell and is not pooled with Table 5 and Table 6. Its learned chest sparse far scale is 0.066 with coverage 0.831, against the official 0.088 and 0.869. Its learned multi-arc far scale is 0.096 with coverage 0.829, against the official 0.095 and 0.80. The comparison is internal to the rerun. It supports the main claim that far coverage is a property of a gap-dependent scale, not a property that any positive constant can be tuned to match, and it does not replace the official numbers.

Implementation and reproducibility notes

Architecture. The encoder is a frozen MAE ViT-B/16 [12] on 224×224224 \times224 projections, with learning rate zero. Fusion uses the two nearest acquired views at softmax temperature 0.10. δ(α′)\delta(\alpha') still uses the full mask. Matched bins of sparse and multi-arc orbits nevertheless have similar scales, which is the empirical content of that split: the unused views change F\mathcal{F} and do not enter the features. The scale head, the only module added over point completion, is initialized near log 0.08, between the official near scales near 0.02 and the official far scales 0.06 to 0.10. It is clamped to σ∈[e−6,0.35]\sigma\in[e^{-6}, 0.35]. The widest reported mean scale is 0.111, about one third of the ceiling. Near-angle mixing is off, so a narrow near interval comes from the scale head and Lcol\mathcal{L}_{\mathrm{col}}, not from copying a neighbor into μ\mu.

Optimization and splits. Each specialist trains for 8,000 steps at batch size 1 from a point-completion initialization. LNLL\mathcal{L}_{\mathrm{NLL}} and Lcol\mathcal{L}_{\mathrm{col}} have equal weights. Table 10 is the check on that choice: removing Lcol\mathcal{L}_{\mathrm{col}} widens the acquired scale and leaves the far scale in place. The ℓ1\ell_{1} and gradient penalties on μ\mu use weights 0.15 and 0.05. The scale has no ground-truth variance target. Training draws KK from 6 to 10, so the test budget K=8K=8 sits inside the training range, uses a sparse subset with probability 0.55, holds out four queries, and drops 15% of acquired views. Chest and dental specialists are separate. There is no shared head that would be forced to average a far scale near 0.06 with one near 0.09. The eight calibration cases are disjoint from the 40 test cases. The 20-case ablation, the 16-case long arc, and the 12-case volume readout are separate from that test set in role, including where a chest far number coincides. A draw from the diagonal Gaussian need not be a projection of a volume in FF. Section H is the volume that fails both acquired-view agreement and query-angle inclusion.

Reading guide for the figure set

Section C reads the pooled cells against the bins, including the 32.4° case and the 20-case collapse panel. Sections D and F are the near/far and inset boards. Section E is the reconstructor comparison and the only width/miss split of a printed scalar. Section G is the protocol wall, the scale heatmaps, and the transfer panels. Sections H and I are the volume diagnostic, the acquired-view PSNR, and the extra far boards. Section J holds the numeric expansions, including the 16-case long arc. Section K is a later chest rerun and is not pooled with those tables. Section N is the scoring argument behind those tables.

Interval scoring and consistency

Why the interval score targets two quantiles. For a scalar target YY with a continuous cumulative distribution FF and finite first moment, write

Sα(L,U;Y)=(U−L)+2α[(L−Y)++(Y−U)+].S_{\alpha}(L,U;Y) = (U-L) + \frac{2}{\alpha}\left[(L-Y)_{+} + (Y-U)_{+}\right].

The derivatives of the expected positive parts are F(L)F(L) and −[1−F(U)]-[1-F(U)]. Consequently, stationarity gives F(L)=α/2F(L)=\alpha/2 and F(U)=1−α/2F(U)=1-\alpha/2. The expected score is convex in each endpoint, so these quantiles minimize it. Unique quantiles give a unique optimum [11, 22]. With atoms, the corresponding subgradient conditions permit non-unique quantiles. Properness concerns the target quantiles. It is not strict propriety for the entire density, because distinct distributions can share both endpoints. The symmetric Gaussian family restricts which pairs can be represented. Properness of the score is therefore not evidence that a fitted Gaussian interval is calibrated. The coverage columns are that evidence: 0.75 to 0.93 across bins, and 0.77 to 0.92 on the eight headline cells. Chest remains below the nominal rate, which is why the main discussion asks for further calibration before 90% is treated as an operating guarantee.

Sharpness and missed observations. At α=0.1\alpha= 0.1, the standard-normal quantile is z0.95≈1.64485z_{0.95} \approx1.64485. The implemented rounded value 1.645 gives interval length 3.29σ3.29\sigma and

an outside-distance multiplier of 20. Averaging is linear, so the detector-average score is exactly the average interval length plus 20 times the average outside distance. Coverage records only whether the outside distance is zero. It cannot identify the score without the magnitudes of the positive distances. This is why the main tables are diagnostics of interval behavior rather than complete numerical ACUS rankings. The factor of 20 makes the missing magnitude consequential. On the chest sparse far cell, u=0.088u=0.088 and the width term is 3.29×0.088≈0.293.29 \times0.088 \approx0.29. An average outside distance of about 0.015, in the same intensity units as σ\sigma, produces a miss term of 20×0.015=0.3020 \times0.015 = 0.30 and already matches that entire width term. On the dental sparse far cell, u=0.061u=0.061 and the width term is about 0.20, matched by an average outside distance of about 0.010. A constant-scale method with zero reported far coverage could therefore sit near the learned ACUS or several times above it, depending on how far outside the narrow band the pixels lie. The miss masks show that those pixels exist and do not show the distance. For the analytical rows the scalar is printed, and Sec. E separates their width and miss terms.

A restricted continuity argument. Suppose a nonempty class B∩F\mathcal{B} \cap\mathcal{F} satisfies a uniform angular modulus ωB\omega_{\mathcal{B}} in a fixed detector norm. For two compatible volumes V1V_1, V2V_2 and a closest acquired angle aa, both project to yay_a at aa. The triangle inequality gives

∥Pa′(V1)−Pa′(V2)∥≤2ωB(∣a′−a∣).\left\lVert P_{a'}(V_1) - P_{a'}(V_2) \right\rVert\le2\omega_{\mathcal{B}}(\lvert a' - a\rvert).

Thus the restricted projection set contracts as the gap vanishes if the assumed modulus vanishes. No modulus ωB\omega_{\mathcal{B}} is estimated. The argument supplies the boundary condition tested by Table 14, acquired scale 0.002 and acquired PSNR 46.4 to 54.9 dB, and Table 10 shows that Lcol\mathcal{L}_{\mathrm{col}} enforces it. The far curves are not that modulus. Chest sparse opens from 0.014 to 0.088 and dental sparse from 0.021 to 0.061 on the same geometry, and Table 11 keeps the geometry fixed while swapping the specialist.

What the training scale estimates. For a fixed conditioning input and mean, the expected Gaussian NLL has scale derivative 1/σ−E[(Y−μ)2]/σ31/\sigma- \mathbb{E}[(Y-\mu)^2]/\sigma^3. Its unconstrained stationary scale is the conditional residual root-mean-square. This includes error in the center and does not separate aleatoric uncertainty, epistemic uncertainty, and approximation bias. The fitted network further imposes a scale floor and ceiling, parameter sharing, finite training, and auxiliary penalties. Empirical coverage must therefore be evaluated rather than inferred from the likelihood objective. The collapse loss tests a different property at acquired angles and cannot calibrate missing angles by itself. The headline cells show what a scale that does track residual size looks like, and where it does not. PSNR is a statement about root-mean-square error of μ\mu. For a fixed peak, a drop of Δ\Delta decibels multiplies that error by 10Δ/2010^{\Delta/20}. Chest sparse falls from 32.48 dB to 18.64 dB, a factor of about 4.9 in amplitude, and the mean scale rises from 0.018 to 0.088, also a factor of about 4.9. Chest multi-arc falls from 29.59 dB to 17.48 dB, a factor of about 4.0, and the scale rises from 0.023 to 0.095, a factor of about 4.1. Dental multi-arc falls from 29.78 dB to 22.37 dB, a factor of about 2.3, and the scale rises from 0.027 to 0.063, a factor of about 2.3. Dental sparse is the exception: PSNR falls from 31.90 dB to 21.12 dB, a factor of about 3.5, while the scale rises from 0.024 to 0.061, a factor of about 2.5. Three of the four headline cells are what the NLL stationary condition leads one to expect when the pixelwise scale follows residual magnitude. The fourth is the gap between that stationary condition and a fitted, clamped, penalized mean of pixelwise scales. None of the four is an identity, and none of them recovers coverage from PSNR. A constant scale fitted where residuals are small, which is what the eight calibration cases supply to the copy baselines, is the same stationary condition applied to the wrong gap. On chest sparse the constant sits near 0.012 to 0.014, the size of the nearest bin, and the far residual is several times larger. Zero reported far coverage is the result the miss penalty was introduced to charge.

References

  1. [1]A. H. Andersen and A. C. Kak. Simultaneous algebraic reconstruction technique (SART): A superior implementation of the ART algorithm. Ultrasonic Imaging, 6(1):81–94, 1984. 6, 7, 11, 21pubmed.ncbi.nlm.nih.gov/6548059
  2. [2]Anastasios N. Angelopoulos, Amit Pal Kohli, Stephen Bates, Michael Jordan, Jitendra Malik, Thayer Alshaabi, Srigokul Upadhyayula, and Yaniv Romano. Image-to-image regression with distribution-free uncertainty quantification and applications in imaging. In International Conference on Machine Learning, pages 717–730, 2022. 2arxiv.org/abs/2202.05265
  3. [3]Christoph Angermann, Simon Göppel, and Markus Haltmeier. Uncertainty-aware null space networks for data-consistent image reconstruction. arXiv preprint arXiv:2304.06955, 2023. 2arxiv.org/abs/2304.06955
  4. [4]Mahmoud Assran, Quentin Duval, Ishan Misra, Piotr Bojanowski, Pascal Vincent, Michael Rabbat, Yann LeCun, and Nicolas Ballas. Self-supervised learning from images with a joint-embedding predictive architecture. In CVPR, pages 15619–15629, 2023. 2arxiv.org/abs/2301.08243
  5. [5]Federico Bolelli, Luca Lumetti, Shankeeth Vinayahalingam, Mattia Di Bartolomeo, Arrigo Pellacani, Kevin Marchesini, Niels van Nistelrooij, Pieter van Lierop, Tong Xi, Yusheng Liu, Rui Xin, Tao Yang, Lisheng Wang, Haoshen Wang, Chenfan Xu, Zhiming Cui, Marek Wodzinski, Henning Müller, Yannick Kirchhoff, Maximilian R. Kokuss, Klaus Maier-Hein, Jaehwan Han, Wan Kim, Hong-Gi Ahn, Tomasz Szczepański, Michal K. Grzeszczyk, Przemyslaw Korzeniowski, Vicent Caselles-Ballester, Xavier Paolo Burgos-Artizzu, Ferran Prados Carrasco, Stefaan Berge, Bram van Ginneken, Alexandre Anesi, and Costantino Grana. Segmenting the inferior alveolar canal in CBCT volumes: The ToothFairy challenge. IEEE Transactions on Medical Imaging, 44(4):1890–1906, 2025. 5,DOI
  6. [6]Yuanhao Cai, Yixun Liang, Jiahao Wang, Angtian Wang, Yulun Zhang, Xiaokang Yang, Zongwei Zhou, and Alan Yuille. Radiative Gaussian splatting for efficient X-ray novel view synthesis. In ECCV, pages 283–299, 2024. 2, 6, 7, 11
  7. [7]Yuanhao Cai, Jiahao Wang, Alan Yuille, Zongwei Zhou, and Angtian Wang. Structure-aware sparse-view X-ray 3D reconstruction. In CVPR, pages 11174–11183, 2024. 2, 6, 7, 11
  8. [8]Marco Cipriano, Stefano Allegretti, Federico Bolelli, Mattia Di Bartolomeo, Federico Pollastri, Arrigo Pellacani, Paolo Minafra, Alexandre Anesi, and Costantino Grana. Deep segmentation of the mandibular canal: A new 3D annotated dataset of CBCT volumes. IEEE Access, 10:11500–11510, 2022. 5, 11DOI
  9. [9]Rolf Clackdoyle and Frédéric Noo. Quantification of tomographic incompleteness in cone-beam reconstruction. IEEE Transactions on Radiation and Plasma Medical Sciences, 4(1):63–80, 2020. 2DOI
  10. [10]L. A. Feldkamp, L. C. Davis, and J. W. Kress. Practical cone-beam algorithm. Journal of the Optical Society of America A, 1(6):612–619, 1984. 6, 7, 11, 21DOI
  11. [11]Tilmann Gneiting and Adrian E. Raftery. Strictly proper scoring rules, prediction, and estimation. Journal of the American Statistical Association, 102(477):359–378, 2007. 2, 4, 25
  12. [12]Kaiming He, Xinlei Chen, Saining Xie, Yanghao Li, Piotr Dollár, and Ross Girshick. Masked autoencoders are scalable vision learners. In CVPR, pages 15979–15988, 2022. 2, 4, 23arxiv.org/abs/2111.06377
  13. [13]Kyong Hwan Jin, Michael T. McCann, Emmanuel Froustey, and Michael Unser. Deep convolutional neural network for inverse problems in imaging. IEEE Transactions on Image Processing, 26(9):4509–4522, 2017. 1, 2DOI
  14. [14]Avinash C. Kak and Malcolm Slaney. Principles of Computerized Tomographic Imaging. SIAM, 2001. 1, 2
  15. [15]Wei-An Lin, Haofu Liao, Cheng Peng, Xiaohang Sun, Jingdan Zhang, Jiebo Luo, Rama Chellappa, and Shaohua Kevin Zhou. DuDoNet: Dual domain network for CT metal artifact reduction. In CVPR, pages 10504–10513, 2019. 2arxiv.org/abs/1907.00273
  16. [16]Frank Natterer. The Mathematics of Computerized Tomography. SIAM, 2001. 1, 2
  17. [17]Arnaud Arindra Adiyoso Setio, Alberto Traverso, Thomas de Bel, Moira S. N. Berens, Cas van den Bogaard, Piergiorgio Cerello, Hao Chen, Qi Dou, Maria Evelina Fantacci, Bram Geurts, Robbert van der Gugten, Pheng Ann Heng, Bart Jansen, Michael M. J. de Kaste, Valentin Kotov, Jack Yu-Hung Lin, Jeroen T. M. C. Manders, Alexander Sóñora-Mengana, Juan Carlos García-Naranjo, Evgenia Papavasileiou, Mathias Prokop, Marco Saletta, Cornelia M. Schaefer-Prokop, Ernst T. Scholten, Luuk Scholten, Miranda M. Snoeren, Ernesto Lopez Torres, Jef Vandemeulebroucke, Nicole Walasek, Guido C. A. Zuidhof, Bram van Ginneken, and Colin Jacobs. Validation, comparison, and combination of algorithms for automatic detection of pulmonary nodules in computed tomography images: The LUNA16 challenge. Medical Image Analysis, 42:1–13, 2017. 5, 11
  18. [18]Emil Y. Sidky and Xiaochuan Pan. Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimization. Physics in Medicine and Biology, 53(17):4777–4807, 2008. 11, 21DOI
  19. [19]Ce Wang, Kun Shang, Haimiao Zhang, Shang Zhao, Dong Liang, and S. Kevin Zhou. Active CT reconstruction with a learned sampling policy. In ACM Multimedia, pages 7226–7235, 2023. 2arxiv.org/abs/2211.01670
  20. [20]Tianyuan Wang, Felix Lucka, and Tristan van Leeuwen. Sequential experimental design for X-ray CT using deep reinforcement learning. IEEE Transactions on Computational Imaging, 10:953–968, 2024. 2DOI
  21. [21]Tianyuan Wang, Felix Lucka, Daniël M. Pelt, K. Joost Batenburg, and Tristan van Leeuwen. Dynamic angle selection in X-ray CT: A reinforcement learning approach to optimal stopping. Applied Mathematics for Modern Challenges, 5:36–63, 2025. 2arxiv.org/abs/2503.12688
  22. [22]Robert L. Winkler. A decision-theoretic approach to interval estimation. Journal of the American Statistical Association, 67(337):187–191, 1972. 2, 4, 25DOI
  23. [23]Yulun Wu, Ruyi Zha, Wei Cao, Yingying Li, Yuanhao Cai, and Yaoyao Liu. Active view selection with perturbed Gaussian ensemble for tomographic reconstruction. In ECCV, pages 405–424, 2026. 2arxiv.org/abs/2603.06852
  24. [24]Zefan Yang, Ge Wang, James Hendler, Mannudeep K. Kalra, and Pingkun Yan. X-WIN: Building chest radiograph world model via predictive sensing. In Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition, pages 6920–6930, 2026. 1, 2arxiv.org/abs/2511.14918
  25. [25]Ruyi Zha, Yanhao Zhang, and Hongdong Li. NAF: Neural attenuation fields for sparse-view CBCT reconstruction. In MICCAI, pages 442–452, 2022. 2, 6, 7, 11DOI
  26. [26]Ruyi Zha, Tao Jun Lin, Yuanhao Cai, Jiwen Cao, Yanhao Zhang, and Hongdong Li. R²-Gaussian: Rectifying radiative Gaussian splatting for tomographic reconstruction. In NeurIPS, pages 44907–44934, 2024. 2, 6, 7, 11arxiv.org/abs/2405.20693
  27. [27]Chulin Zhao, Yiran Xu, and Shu Liu. When variance is not an error map: Calibrated uncertainty for radiative Gaussian splatting in sparse-view CT. arXiv preprint arXiv:2607.13682, 2026. 2arxiv.org/abs/2607.13682

Paper details

vlnc:W_ea7zzmfg
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 13

downloads · 1

A view counts once per browser per day, a download each time the PDF is requested.

  • Computer Science

ChatGPT

View statement
Looking for related papers…
Contents