Abstract

A Compton camera measures two events in a detector from individual gamma-rays to reconstruct the radiation source. Conventional scintillator-based Compton cameras consist of two separate scintillator volumes or a single segmented scintillator, resulting in drawbacks regarding detector bulkiness and efficiency. Their scintillator readout techniques struggle to resolve two nearby events in one monolithic scintillator. Single-photon avalanche diode (SPAD) cameras have recently been proposed for measuring scintillation events in various configurations. SPAD array sensors enable high-resolution photon-counting acquisition at high rates that is easily scalable to millions of pixels. Prior work has explored different optics and measurement arrangements as well as demonstrated measurement of single interaction locations. In this work, we extend these capabilities to demonstrate a practical radiation imaging device in the form of a Compton camera consisting of a SPAD camera and one compact, monolithic scintillator. Our Compton camera prototype successfully reconstructs a gamma-ray source over multiple locations that span a 4ππ steradian field of view. Our approach is able to resolve two simultaneous events in one monolithic scintillator and enables Compton camera devices with compact, monolithic scintillator volumes for enhanced portability.

Introduction

The Compton camera is an electronic collimation method for reconstructing high-energy gamma-ray sources [1, 2, 3, 4]. A Compton camera measures the 3D position and energy deposition of two events, namely Compton scattering and photoelectric absorption, from individual gamma-rays in a detector. The radiation source is then reconstructed based on the kinematics of Compton scattering. Compton cameras are applicable in nuclear security and nonproliferation [5, 6, 7, 8, 9], medical imaging [10], and astronomy [11, 12, 13].

Current scintillator-based Compton cameras predominantly use silicon photomultiplier (SiPM) arrays or similar sensors coupled to the scintillator’s surface to measure double-events [1, 14]. A Compton camera must be able to spatially resolve and measure two simultaneous events. Accordingly, Compton camera designs have consisted of a single-volume scintillator segmented with waveguides [15, 16, 17, 18, 19] or two separate scintillator volumes [8, 20, 21, 22, 23, 24]. However, inserting waveguides in a scintillator increases cost and decreases event detection efficiency. Increasing distance between two separate scintillator volumes increases the overall space occupied by the detector and reduces sensitivity to wide-angle source locations and scatterings. A single-volume, monolithic scintillator design could address both aforementioned designs’ drawbacks, but current scintillation readout techniques for Compton cameras would struggle to resolve two nearby events in a monolithic scintillator due to overlapping light on the sensor. High-resolution focal plane sensors combined with appropriate optics have been shown to effectively localize single point-like [25, 26, 27] and track-like [28, 29, 30] interactions within a monolithic scintillator, but not double-events.

We propose a camera-based scintillator readout technique to construct a single-volume, monolithic scintillator Compton camera. Specifically, we adopt a single-photon avalanche diode (SPAD) camera. Prior simulations that extrapolated high photon counts have shown that double-events in one monolithic scintillator could be resolved with a SPAD camera, but experimental demonstrations were infeasible due to low light collection [25]. Now, we adopt a kaleidoscopic scintillator to obtain sufficient light collection and event spatial resolution to experimentally demonstrate a single-volume, monolithic scintillator Compton camera [26]. Furthermore, the SPAD camera’s ability to resolve distinct nearby events allows for the miniaturization of scintillator volumes in Compton cameras. Minimizing scintillator volume is important for reducing detector size and weight to enhance portability and integration into field-deployable devices.

The SPAD camera’s sensor, the SPAD array, is fabricated with a highly scalable and integrated CMOS architecture that enables a high number of readout channels and high resolution for detecting single-photon arrivals that alternative sensors do not achieve. Recently, there has been growing interest in adopting SPAD cameras [25, 26, 29, 30] and SPAD arrays [31, 32] for measuring scintillation events, driven largely by major advances in SPAD technology [33]. For example, SPAD arrays have achieved megapixel resolution, photon timetagging resolution of 100 ps, frame rates close to 100 kHz, and overall photon detection efficiencies approaching 70% [34, 35]. Previous works that adopt SPAD cameras have investigated individual components of radiation imaging, such as the camera lens or the localization of individual particle interactions, but demonstrations of actual radiation imaging capabilities are still lacking [25, 26, 27, 29, 30].

We experimentally demonstrate our Compton camera prototype by reconstructing a gamma-ray source over multiple locations that span a 4π4\pi steradian FOV. We test our method with a compact scintillator volume, accepting reduced gamma-ray flux and lower event rates as a tradeoff. Future optimization of the proposed design may enable a Compton camera with source reconstruction performance that is competitive with existing devices and provide benefits in terms of form factor and portability. Our approach for resolving double-events in a monolithic scintillator is broadly applicable to any modality. The design is illustrated in Figure 1.

Compton camera design and backprojection process

Figure 1. Compton camera design and backprojection process. A double-event in a kaleidoscopic scintillator from a gamma-ray is imaged with a SPAD camera and measured to backproject a cone. The source’s angular position is reconstructed by repeating this process over many gamma-rays.

Our contributions are as follows:

  • A Compton camera that measures events with a camera and consists of one compact, monolithic scintillator.

  • An algorithm for measuring two events’ locations and relative energy depositions in one image of a kaleidoscopic scintillator.

Event Image Model

We adopt a square pyramid geometry for the kaleidoscopic scintillator. The world coordinate system origin is located at the scintillator’s apex. The zz-axis is directed toward the camera, and the xx- and yy-axes are perpendicular to the sensor’s edges. The camera’s optical axis is along the zz-axis. The source reconstruction coordinate system is defined by azimuth (Ψ\Psi) and elevation (Φ\Phi) with respect to the origin at the scintillator apex. The (Ψ,Φ)=(0∘,0∘)(\Psi, \Phi) = (0^\circ, 0^\circ) direction is along the −z-z-axis. Azimuth is measured about the yy-axis in the xzxz-plane, increasing counterclockwise when viewed from the +y+y direction. Elevation is measured relative to the xzxz-plane, with positive elevation in the +y+y direction. The source reconstruction map spans the full 4π4\pi steradian FOV. The coordinate systems are illustrated in Figure 2.

Coordinate systems

Figure 2. Coordinate systems. The world coordinate system is denoted by the solid black arrows. The source is reconstructed over azimuth (Ψ\Psi) and elevation (Φ\Phi) denoted by the red arrows.

An event and its mirror reflections are approximated as point sources of light whose locations are constrained according to the scintillator’s geometry. We model the camera with a thin lens and Gaussian point spread function. Therefore, an event or mirror reflection’s (x,y,z)(x,y,z) location corresponds to a photon cluster in the image modeled as photon (x,y)(x,y) coordinates distributed over an isotropic 2D Gaussian with a centroid μ\boldsymbol{\mu} and standard deviation σ\sigma. The correspondence between an event or mirror reflection’s (x,y,z)(x,y,z) location and μ\boldsymbol{\mu} and σ\sigma is derived from depth from defocus and perspective projection and depends on the camera parameters. The equations defining this correspondence are provided in previous work [26]. The image of an event and its mirror reflections is modeled as a Gaussian mixture model (GMM), where all Gaussian components are parameterized by the event’s (x,y,z)(x,y,z) location, as derived in previous work [26]. We assume each specular surface of the kaleidoscopic scintillator generates up to one mirror reflection for a total of up to four mirror reflections for one event. Mirror reflections may be missing from the image due to finite mirrors, depending on their locations.

Event Measurement Algorithm

Our algorithm measures events in post-processing in two stages. First, the algorithm detects whether an image contains two events. Second, the algorithm measures the two events’ 3D locations and energy depositions. The double-event localization algorithm maximizes the GMM’s likelihood using the EM algorithm. The algorithm’s double-event measurements are then used in the Compton camera backprojection method to reconstruct the gamma-ray source.

Double-Event Detection

We use template matching to detect the presence of two events in an image. One template consists of kernels that correspond to one event and its mirror reflections in an image. We define kernel kk, corresponding to event or mirror reflection kk, over a square region centered at μk\boldsymbol{\mu}_k with side length 4σk4\sigma_k. Within this region, pixel intensities follow an isotropic 2D Gaussian distribution centered at μk\boldsymbol{\mu}_k with standard deviation σk\sigma_k. A template’s pixels outside each kernel are zeroed. μk\boldsymbol{\mu}_k and σk\sigma_k are computed based on the event location, scintillator geometry, and camera parameters. σk\sigma_k is assigned a minimum value of 10 pixels to prevent singularities.

We generate a set of templates for event locations over a 3D grid that spans the scintillator volume within the camera’s FOV. The grid’s xx-yy coordinates span the camera’s lateral FOV. The grid’s zz-coordinates range from the scintillator’s apex to the highest zz-coordinate where four mirror reflections are observed in the camera’s FOV. This grid encompasses the measurable 3D region of the scintillator.

To detect two events in an image, we compute the Frobenius inner product between the image and each template and retain the template, T1T_1, that resulted in the maximum value as the first event’s template. Then, we remove all photons in the image that are within the kernels of T1T_1. We repeat this computation on the image with removed photons to obtain the second event’s template, T2T_2. We detect two events when, for both T1T_1 and T2T_2, the original image contains at least 3 kernels with at least 5 photons each, including a kernel corresponding to the event’s photon cluster. This provides initial estimates for event locations and classifies which mirror reflections are present in the image. A mirror reflection is deemed to be present if its corresponding kernel contains at least 5 photons.

Double-Event Measurement

To estimate the locations of two events in one image with NN photons and KK clusters, we optimize the following objective function:

argmax⁡x1,y1,z1,x2,y2,z2,π Q(1)\underset{x_1,y_1,z_1,x_2,y_2,z_2,\boldsymbol{\pi}}{\operatorname{argmax}}\ Q \tag*{(1)}

where QQ is the expected log-likelihood of the complete-data likelihood function.

Q=∑i=1N∑k=1Krik[log⁡(πk)+log⁡(wi)−log⁡(2πσk2)−wi2σk2∥ti−μk∥22](2)Q = \sum_{i=1}^{N}\sum_{k=1}^{K} r_{ik}\left[\log(\pi_k)+\log(w_i)-\log\left(2\pi\sigma_k^2\right)-\frac{w_i}{2\sigma_k^2}\left\lVert\boldsymbol{t}_i-\boldsymbol{\mu}_k\right\rVert_2^2\right] \tag*{(2)}
rik=πkN(ti;μk,σk2)∑k′=1Kπk′N(ti;μk′,σk′2)(3)r_{ik}=\frac{\pi_k\mathcal{N}(\boldsymbol{t}_i;\boldsymbol{\mu}_k,\sigma_k^2)}{\sum_{k'=1}^{K}\pi_{k'}\mathcal{N}(\boldsymbol{t}_i;\boldsymbol{\mu}_{k'},\sigma_{k'}^2)} \tag*{(3)}

(x1,y1,z1)(x_1,y_1,z_1) and (x2,y2,z2)(x_2,y_2,z_2) are the two events’ locations, ti\boldsymbol{t}_i is the xyxy-coordinate for photon ii, π\boldsymbol{\pi} is a vector of mixing weights, wiw_i is the weight applied to photon ii, μk\boldsymbol{\mu}_k and σk\sigma_k are the parameters of Gaussian component kk, and rikr_{ik} is the posterior probability that photon ii belongs to cluster kk. νi=∑j∈Siqexp⁡(−∥ti−tj∥22ν)\nu_i=\sum_{j\in S_i^q}\exp\left(-\frac{\lVert\boldsymbol{t}_i-\boldsymbol{t}_j\rVert_2^2}{\nu}\right), where SiqS_i^q is the set of qq nearest neighbors of photon ii, and ν\nu is a positive scalar. These terms are derived in previous work [26].

The initialized event locations and which KK clusters are present are obtained from the event detection stage. Event location estimates are fine-tuned by optimizing (1) with the EM algorithm.

We assume knowledge of the incident gamma-ray’s energy and proportionally assign energy depositions to the two events based on their assigned photon counts. We assume all the gamma-ray’s energy is deposited in the two events. 5R1C1\frac{5}{R_1}C_1 and 5R2C2\frac{5}{R_2}C_2 counts are assigned to the two events, accounting for missing mirror reflections.

Compton Camera Backprojection

The Compton camera backprojection is based on the Compton scattering kinematics of an incident gamma-ray in the following process. A gamma-ray with energy E0E_0 originating from a source located at p0\boldsymbol{p}_0 propagates into the scintillator and undergoes a Compton scattering event at p1\boldsymbol{p}_1. During this event, the gamma-ray loses some energy, and its trajectory is deflected by an angle θ\theta according to cos⁡θ=1−mc2(1/E1−1/E0)\cos\theta=1-mc^2(1/E_1-1/E_0). E1E_1 is the gamma-ray’s energy after scattering, and mc2mc^2 is the rest mass energy of an electron. The gamma-ray continues propagating in the scintillator and undergoes a photoelectric absorption event at p2\mathbf{p}_2, during which it loses all of its energy and stops in its path.

The backprojection for one gamma-ray consists of a cone with vertex at p1\mathbf{p}_1, axis oriented along the vector from p2\mathbf{p}_2 to p1\mathbf{p}_1, and half-angle θ\theta. The cone is constructed using measurements of the event pair’s locations and energy depositions. In simple backprojection, the source is reconstructed over many gamma-ray backprojections by accumulating the intersection of the cones’ edges with the reconstruction surface. Cones are constructed using our algorithm’s measurements of double-event locations and energy depositions. The edges are assigned a width that adds blur to the reconstruction. Source reconstruction can be deblurred and improved in quality using list-mode maximum likelihood expectation maximization (LM-MLEM) [36, 37]. The LM-MLEM reconstruction is defined as λjn+1=λjnsj∑itij∑ktikλkn\lambda_j^{n+1} = \frac{\lambda_j^n}{s_j} \sum_i \frac{t_{ij}}{\sum_k t_{ik}\lambda_k^n} where λjn\lambda_j^n is the intensity of reconstruction pixel jj at the nnth iteration, sjs_j is the probability that a gamma-ray originating from reconstruction pixel jj is detected anywhere, and tijt_{ij} is the probability that a gamma-ray originating from reconstruction pixel jj results in the scatter-absorption pair measurement ii. The Compton scattering and backprojection processes are shown in Figure 1.

Hardware Prototype

We adopt the same kaleidoscopic scintillator and SPAD camera as in previous work [26]. The scintillator is a GAGG(Ce) crystal (Epic Crystal Co.) in the shape of a square pyramid with 5.77 mm height and a 20 mm base length. All surfaces of the mirror except for the base surface are coated with enhanced specular reflector. The SPAD camera consists of the SPAD512 array (Pi Imaging) with 512 ×\times 512 pixels and 16.4 μ\mum pixel pitch and a 50 mm focal length lens (Nikkor) with f/1.2 aperture. The lens and scintillator are separated by about 30 mm of air. The SPAD camera is focused on the scintillator’s apex, and its FOV spans 5.5 ×\times 5.5 mm. The SPAD camera is oriented such that its optical axis is centered on the scintillator’s apex and its sensor’s edges are parallel with the scintillator’s respective base surface edges. After calibrating the camera parameters, the range in the zz dimension over which events can be measured is 3.5 mm. This results in a measurable scintillator volume that is a square pyramid with 120 degrees opening angle contained within a 5.5 ×\times 5.5 \times 3.5 mm3^3 rectangular prism for a total measurable volume of about 74 mm3^3. The setup is contained within a light-tight box. The top 5% of pixels with the highest dark count rates are zeroed. The experimental setup is shown in Figure 3.

Experimental setup

Figure 3. Experimental setup.

Data is collected by passively capturing 1-bit images with a 1.5 μ\mus global shutter exposure time. This exposure time is selected such that it is long enough to capture scintillation light emitted over scintillator’s 150 ns decay constant and short enough to reduce dark count accumulation. At this setting, the camera reads out images at about 19 kHz. In post-processing, our algorithm analyzes all captured images and detects which images contain double-events for performing the Compton backprojection.

Hardware Experiments

Algorithm Parameters

The calibrated camera parameter values used in this work are S1=45S_1 = 45 mm, S2=64S_2 = 64 mm, A=41.7A = 41.7 mm, and a=0.25a = 0.25, where S1S_1 is the distance between the lens plane and sensor, S2S_2 is the distance between the lens plane and focal plane, AA is the lens diameter, and aa is a scaling factor relating the circle of confusion to σk\sigma_k. These parameters relate an event’s (x,y,z)(x,y,z) location to its corresponding photon cluster’s μ\mu and σ\sigma, as defined in prior work [26].

We generate templates, as described in § 1.2 A, over a grid that consists of 10 points equispaced in each dimension over x∈[−2,2]x \in[-2,2] mm, y∈[−2,2]y \in[-2,2] mm, and z∈[0.1,3]z \in[0.1,3] mm. A template is discarded if any of its kernels overlap or if its event’s location is outside the scintillator.

In one M-step of the EM algorithm, we update (x1,y1,z1)(x_1,y_1,z_1) and (x2,y2,z2)(x_2,y_2,z_2) by running gradient ascent for 1,000 steps with a step size of 1e-7. All σk\sigma_k’s are clipped to a minimum value of 10 pixels. The gradient of σk\sigma_k is zeroed if σk\sigma_k is clipped. We assign weight wiw_i to each photon using q=10q = 10 nearest neighbors and ν=10\nu= 10 pixels. The EM algorithm is terminated once the distance in estimated event locations between consecutive steps is less than 0.01 mm for both events.

Due to unknown ordering in scattering versus absorption events, we include both possible orderings in the list of scattering-absorption pair measurements. We do not include an ordering that results in non-physical energy depositions according to Compton scattering kinematics. The LM-MLEM tijt_{ij} intensity is Gaussian centered on the backprojection’s cone-sphere intersection pixels with standard deviation of 15 degrees. The source is reconstructed over a sphere with a 20 mm radius and an average pixel separation of 0.9 degrees (Healpix nside=64). λj\lambda_j is initialized to all 1’s. sjs_j is set to the uniform distribution. LM-MLEM is run for 10 iterations.

Source Reconstruction

We evaluate our Compton camera prototype over a set of gamma-ray source locations using a 9 μ\muCi Cs-137 source (662 KeV; Spectrum Techniques). The source’s radioactive material is approximately 3.1 mm in diameter. We assign an uncertainty to the source’s true location of ±3\pm3 mm in each dimension. The ambient temperature during capturing is 24 °C, and we obtain a median of 4 dark counts per image. We capture a varying number of images over each source location. We discard images with less than 60 counts.

The tested gamma-ray source locations and number of double-events detected are as follows:

  • (a) (0,0,−20)(0, 0, -20) mm, (0∘,0∘)(0^\circ, 0^\circ), 90 double-events

  • (b) (−20,0,0)(-20, 0, 0) mm, (90∘,0∘)(90^\circ, 0^\circ), 379 double-events

  • (c) (−20,0,−20)(-20, 0, -20) mm, (45∘,0∘)(45^\circ, 0^\circ), 217 double-events

  • (d) (−10,0,−20)(-10, 0, -20) mm, (26.6∘,0∘)(26.6^\circ, 0^\circ), 438 double-events

  • (e) (0,20,−8)(0, 20, -8) mm, (0∘,68∘)(0^\circ, 68^\circ), 250 double-events

  • (f) (0,20,15)(0, 20, 15) mm, (180∘,53.1∘)(180^\circ, 53.1^\circ), 66 double-events

  • (g) (0,10,−20)(0, 10, -20) mm, (0∘,26.5∘)(0^\circ, 26.5^\circ), 140 double-events

  • (h) (20,0,0)(20, 0, 0) mm, (270∘,0∘)(270^\circ, 0^\circ), 278 double-events

  • (i) (10,0,−20)(10, 0, -20) mm, (333.4∘,0∘)(333.4^\circ, 0^\circ), 546 double-events

  • (j) (0,−20,0)(0, -20, 0) mm, (0∘,−90∘)(0^\circ, -90^\circ), 326 double-events

The reconstructions after applying LM-MLEM are shown in Figure 4. Examples of experimental images containing two events and the algorithm’s estimated event locations are shown in Figure 5. We observe a maximum of around 150 counts in an image.

Source reconstruction map for placement (a)

(a)

Source reconstruction map for placement (b)

(b)

Source reconstruction map for placement (c)

(c)

Source reconstruction map for placement (d)

(d)

Source reconstruction map for placement (e)

(e)

Source reconstruction map for placement (f)

(f)

Source reconstruction map for placement (g)

(g)

Source reconstruction map for placement (h)

(h)

Source reconstruction map for placement (i)

(i)

Source reconstruction map for placement (j)

(j)

Figure 4. Source reconstruction results. (a-j) The reconstruction map for each gamma-ray placement. The red outline denotes the angle subtended by the gamma-ray source diameter and placement uncertainty.

Experimental images of double-events

Figure 5. Experimental images of double-events. The algorithm’s estimated Gaussian components are illustrated as dashed circles centered on each Gaussian component’s mean. The inner and outer circles are one and two standard deviations in radius, respectively. The circles’ colors denote which event a Gaussian component belongs to. Pixels with a photon are enlarged for visualization purposes. The number of counts in the image and the algorithm’s estimated event locations are shown in each image.

Discussion

The reconstructions approximately match the source locations, as shown in Figure 4, and the GMM is correctly fit to the double-events in selected images shown in Figure 5. This demonstrates that our algorithm is able to measure two events in one image of a compact, monolithic scintillator, and that our prototype can successfully reconstruct sources over a 4π4\pi steradian FOV.

The reconstruction results are noteworthy in regard to the scintillator’s small form factor. Our prototype’s measurable scintillator volume of 74 mm3\mathrm{mm}^3 is much smaller than that of current Compton camera prototypes, which is typically on the order of 10×10×1010 \times10 \times10 mm3\mathrm{mm}^3 or higher [8, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24]. The small volume constrains scatter-absorption events to short separation distances, causing event localization errors to have a greater impact on backprojection and making accurate source reconstruction more challenging. In addition, the small volume causes most gamma-rays to pass through the scintillator without interacting, reducing event rates. Despite these challenges, the prototype estimates event locations with sufficiently high spatial resolution to reconstruct the source.

Source reconstruction results would likely improve if more events were captured. Double-event counts are limited by long acquisition times due to the small scintillator volume, the low-activity source required by safety protocols, and reducing individual image exposure. Low source activity also forces the source to be placed close to the scintillator, making the source less point-like and increasing the uncertainty of its ground truth angular position. Over all source locations, the algorithm detects an average of 173 double-events per 1 billion captured images, or 1 double-event per 7 minutes of exposure.

There are several aspects of the prototype that future work may improve. Photon counts per image are relatively low, even with adopting a kaleidoscopic scintillator, leading to relatively low energy resolution. Future work could optimize the scintillator geometry by employing a smaller opening angle to image higher-order reflections and further increase light collection. The scintillator volume could be increased to capture more events or decreased for improved portability. Modifying the camera optics would likely accompany such design choices.

The choice of SPAD camera could potentially be improved, specifically regarding photon detection efficiency and dark count rate. We opted to reduce image exposure time to about 3% of the image readout period to reduce dark count rates, resulting in longer acquisition times. The effect of dark count rates on the final measurement is complex and depends on the choice of algorithm and optical configuration among other things. Newer SPAD chips with lower dark count rates are becoming available, and dark count rate could be reduced by better active or passive cooling. In general, a larger SPAD array area would reduce the demands on the optics and make it easier to achieve high energy resolution and position accuracy over larger volumes.

While this work develops a Compton camera, using a SPAD camera to image double-events in a monolithic scintillator is broadly applicable to other radiation imaging modalities, including neutron scatter cameras [38, 39, 40, 41] and positron emission tomography (PET). In PET, many

monolithic scintillator designs assume one event location, whereas most annihilation photons interact two or more times [42]. Our camera-based approach could spatially resolve and locate multiple events from one annihilation photon in a monolithic scintillator and potentially improve PET image reconstruction.

Data Availability

The data that support the findings of this article and the code used to analyze the data may be obtained from the authors upon request.

Acknowledgments

This material is based upon work supported by the Department of Energy / National Nuclear Security Administration under Award Number(s) DE-NA0004196.

References

References

  1. [1]R. K. Parajuli, M. Sakai, R. Parajuli, and M. Tashiro, Development and applications of compton camera—a review, Sensors 22, 7374 (2022).DOI
  2. [2]S. M. Kim and J. S. Lee, A comprehensive review on compton camera image reconstruction: From principles to ai innovations, Biomedical Engineering Letters 14, 1175 (2024).DOI
  3. [3]D. Turecek, J. Jakubek, E. Trojanova, and L. Sefc, Compton camera based on timepix3 technology, Journal of Instrumentation 13 (11), C11022.
  4. [4]L. Mihailescu, K. Vetter, M. Burks, E. Hull, and W. Craig, Speir: a ge compton camera, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 570, 89 (2007).
  5. [5]K. Vetter, R. Barnowski, A. Haefner, T. H. Joshi, R. Pavlovsky, and B. J. Quiter, Gamma-ray imaging for nuclear security and safety: Towards 3-d gamma-ray vision, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 878, 159 (2018).DOI
  6. [6]K. Ziock, Principles and applications of gamma-ray imaging for arms control, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 878, 191 (2018).DOI
  7. [7]H. Al Hamrashdi, S. D. Monk, and D. Cheneler, Passive gamma-ray and neutron imaging systems for national security and nuclear non-proliferation in controlled and uncontrolled detection areas: Review of past and current status, Sensors 19, 2638 (2019).
  8. [8]J. Kataoka, A. Kishimoto, T. Nishiyama, T. Fujita, K. Takeuchi, T. Kato, T. Nakamori, S. Ohsuka, S. Nakamura, M. Hirayanagi, et al., Handy compton camera using 3d position-sensitive scintillators coupled with large-area monolithic mppc arrays, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 732, 403 (2013).DOI
  9. [9]S. Mochizuki, J. Kataoka, L. Tagawa, Y. Iwamoto, H. Okochi, N. Katsumi, S. Kinno, M. Arimoto, T. Maruhashi, K. Fujieda, et al., First demonstration of aerial gamma-ray imaging using drone for prompt radiation survey in fukushima, Journal of Instrumentation 12 (11), P11014.
  10. [10]H. Tashima and T. Yamaya, Compton imaging for medical applications, Radiological Physics and Technology 15, 187 (2022).DOI
  11. [11]V. Schönfelder, A. Hirner, and K. Schneider, A telescope for soft gamma ray astronomy, Nuclear Instruments and Methods 107, 385 (1973).
  12. [12]V. Schonfelder, R. Diehl, G. Lichti, H. Steinle, B. Swanenburg, A. Deerenberg, H. Aarts, J. Lockwood, W. Webber, J. Macri, et al., The imaging compton telescope comptel on the gamma ray observatory, IEEE Transactions on Nuclear Science 31, 766 (1984).DOI
  13. [13]P. Bloser, J. Legere, C. Bancroft, J. Ryan, and M. McConnell, Balloon light test of a compton telescope based on scintillators with silicon photomultiplier readouts, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 812, 92 (2016).DOI
  14. [14]G. Llosá, Sipm-based compton cameras, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 926, 148 (2019).DOI
  15. [15]T. Takahashi, J. Kawarabayashi, H. Tomita, T. Iguchi, and E. Takada, Development of omnidirectional gamma-imager with stacked scintillators, in 2013 3rd International Conference on Advancements in Nuclear Instrumentation, Measurement Methods and their Applications (ANIMMA) (IEEE, 2013) pp. 1–4.DOI
  16. [16]E. L. Swanberg, N. J. Cherepy, B. M. Wihl, P. R. Beck, Z. M. Seeley, S. L. Hunter, S. E. Fisher, S. A. Payne, and J. Kindem, Transparent ceramic garnet gamma-ray spectrometer with directionality, IEEE Transactions on Nuclear Science 65, 2303 (2018).DOI
  17. [17]H. Lee, T. Lee, and W. Lee, Development of a position-sensitive 4π compton camera based on a single segmented scintillator, IEEE Transactions on Nuclear Science 67, 2511 (2020).DOI
  18. [18]H. Lee, J. Park, and W. Lee, Development of modified scintillator-based single-crystal position-sensitive 4π compton camera for a portable radiation imaging device, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 1043, 167485 (2022).DOI
  19. [19]S. Choi, H. Lee, and W. Lee, Detection efficiency and imaging performance of a 25.4× 25.4× 24 mm3 compton camera based on a single voxelized scintillator, Radiation Physics and Chemistry 233, 112754 (2025).DOI
  20. [20]W. Lee and T. Lee, A compact compton camera using scintillators for the investigation of nuclear materials, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 624, 118 (2010).
  21. [21]A. Kishimoto, J. Kataoka, A. Koide, K. Sueoka, Y. Iwamoto, T. Taya, and S. Ohsuka, Development of a compact scintillator-based high-resolution compton camera for molecular imaging, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 845, 656 (2017).DOI
  22. [22]Y.-S. Kim, J. H. Kim, J. Lee, and C. H. Kim, Large-area compton camera for high-speed and 3-d imaging, IEEE Transactions on Nuclear Science 65, 2817 (2018).DOI
  23. [23]H. Hosokoshi, J. Kataoka, S. Mochizuki, M. Yoneyama, S. Ito, H. Kiji, F. Nishi, S. Miyamoto, and T. Shima, Development and performance verification of a 3-d position- sensitive compton camera for imaging mev gamma rays, Scientific reports 9, 18551 (2019).
  24. [24]K. Ogane, M. Unemachi, K. Shimazoe, M. Takahashi, H. Takahashi, Y. Seto, and T. Momose, Simultaneous measurements of single gamma ray of 131I and annihilation radiation of 18f with compton pet hybrid camera, Applied Radiation and Isotopes 176, 109864 (2021).
  25. [25]A. Bocchieri, E. Charbon, and A. Velten, Scintillation event imaging with a single photon avalanche diode camera, Communications Engineering 3, 135 (2024).DOI
  26. [26]A. Bocchieri, J. Mamish, D. Appleyard, and A. Velten, Kaleidoscopic scintillation event imaging, in Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition (CVPR) (2026) pp. 17949–17958.arxiv.org/abs/2512.03216
  27. [27]X. Dai, C.-J. Ho, K. Tandi, C. Lee, A. Bocchieri, D. Parra, F. Peterson, T. Sultan, F. Sutanto, A. Velten, et al., Pleonoptic imaging of particle interactions in scintillation detectors, arXiv preprint arXiv:2607.01123 (2026).arxiv.org/abs/2607.01123
  28. [28]M. Filipenko, T. Ishkakhov, P. Hufschmidt, G. Anton, M. Campbell, T. Gleixner, G. Leuchs, T. Tick, J. Vall-lerga, M. Wagenpfeil, et al., Three-dimensional photograph of electron tracks through a plastic scintillator, The European Physical Journal C 74, 3131 (2014).DOI
  29. [29]T. Dieminger, S. Alonso-Monsalve, C. Alt, C. Bruschini, N. Bührer, E. Charbon, K. Kaneyasu, T. Weber, M. Franks, and D. Sgalaberna, An ultrafast pleonoptic camera system for high-resolution 3d particle tracking in unsegmented scintillators, Nature Communications (2026).
  30. [30]T. Dieminger, M. Barbieri, V. Berardi, S. Alonso-Monsalve, K. Kaneyasu, P. Mos, T. Weber, J. Wüthrich, C. Bruschini, E. Charbon, et al., Event-by-event track imaging of charged particles with an ultrafast pleonoptic camera, arXiv preprint arXiv:2609.35509 (2026).arxiv.org/abs/2609.35509
  31. [31]M. Franks, T. Dieminger, K. Kaneyasu, D. Sgalaberna, C. Bruschini, E. Charbon, U. Kose, B. Li, P. Mos, M. Wayne, et al., Demonstration of particle tracking with scintillating fibres read out by a spad array sensor and application as a neutrino active target, The European Physical Journal C 84, 202 (2024).
  32. [32]N. Bührer, S. Alonso-Monsalve, M. Franks, T. Dieminger, and D. Sgalaberna, AI-based particle track identification in scintillating fibres read out with imaging sensors, Journal of Instrumentation 20 (05), P05041.
  33. [33]C. Bruschini, H. Homulle, I. M. Antolovic, S. Burri, and E. Charbon, Single-photon avalanche diode imagers in biophotonics: review and outlook, Light: Science & Applications 8, 87 (2019).DOI
  34. [34]K. Morimoto, J. Iwata, M. Shinohara, H. Sekine, A. Abdelghafar, H. Tsuchiya, Y. Kuroda, K. Tojima, W. Endo, Y. Maehashi, et al., 3.2 megapixel 3d-stacked charge focusing spad for low-light imaging and depth sensing, in 2021 IEEE International Electron Devices Meeting (IEDM) (IEEE, 2021) pp. 20–2.DOI
  35. [35]A. C. Ulku, C. Bruschini, I. M. Antolovic, Y. Kuo, R. Ankri, S. Weiss, X. Michalet, and E. Charbon, A 512×512 spad image sensor with integrated gating for wide-field flim, IEEE Journal of Selected Topics in Quantum Electronics 25, 1 (2018).
  36. [36]L. Parra and H. H. Barrett, List-mode likelihood: Em algorithm and image quality estimation demonstrated on 2-d pet, IEEE transactions on medical imaging 17, 228 (1998).DOI
  37. [37]S. J. Wilderman, N. H. Clinthorne, J. A. Fessler, and W. L. Rogers, List-mode maximum likelihood reconstruction of compton scatter camera images in nuclear medicine, in 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No. 98CH36255), Vol. 3 (IEEE, 1998) pp. 1716–1720.
  38. [38]N. Mascarenhas, J. Brennan, K. Krenz, P. Marleau, and S. Mrowka, Results with the neutron scatter camera, IEEE Transactions on Nuclear Science 56, 1269 (2009).
  39. [39]J. Braverman, J. Brennan, E. Brubaker, B. Cabrera-Palmer, S. Czyz, P. M. Mattingly, A. Nowack, J. Steele, M. Sweany, et al., Single-volume neutron scatter camera for high-efficiency neutron imaging and spectroscopy, arXiv preprint arXiv:1802.05261 (2018).arxiv.org/abs/1802.05261
  40. [40]K. Weinfurther, J. Mattingly, E. Brubaker, and J. Steele, Model-based design evaluation of a compact, high-efficiency neutron scatter camera, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 883, 115 (2018).DOI
  41. [41]J. J. Manfredi, E. Adamek, J. A. Brown, E. Brubaker, B. Cabrera-Palmer, J. Cates, R. Dorrill, A. Druetzler, J. Elam, P. L. Feng, et al., The single-volume scatter camera, in Hard X-Ray, Gamma-Ray, and Neutron Detector Physics XXII, Vol. 11494 (SPIE, 2020) pp. 121–131.DOI
  42. [42]A. Gonzalez-Montoro, A. J. Gonzalez, S. Pourashraf, R. S. Miyaoka, P. Bruyndonckx, G. Chinn, L. A. Pierce, and C. S. Levin, Evolution of pet detectors and event positioning algorithms using monolithic scintillation crystals, IEEE Transactions on Radiation and Plasma Medical Sciences 5, 282 (2021).

Paper details

vlnc:W_echvtq4m
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 0

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…
Contents