Interaction-Gated Time-Frequency Learning for Parameter-Efficient PPG-Based Heart Rate Estimation
Abstract
Photoplethysmography (PPG) enables low-cost and non-invasive heart rate (HR) monitoring but remains sensitive to waveform distortion, physiological variability, and measurement interference. Temporal waveform characteristics and frequency-domain periodicity provide complementary information for HR estimation; however, effectively exploring and integrating these representations within a compact model remains challenging. This paper proposes TF-IG, a parameter-efficient interaction-gated time-frequency learning framework for PPG-based HR estimation. TF-IG employs a dual-branch architecture in which a temporal branch captures pulse morphology via dilated depthwise-separable convolutions, while a frequency branch extracts cardiac periodicity information from bounded log-spectrogram representations. To combine the two domains, we introduce an interaction-gated fusion module that explicitly models feature agreement and discrepancy, while also learning feature-wise weights to balance temporal and spectral information. A parameter-shared regression head further maps the temporal, spectral, and fused representations into the final HR prediction space. Experiments on the benchmark BIDMC PPG and Respiration Dataset using a subject-exclusive training, validation, and test protocol show that TF-IG achieves a mean absolute error (MAE) of 1.227 BPM, a root mean square error (RMSE) of 2.060 BPM, a Pearson correlation coefficient of 0.9934, and 92.28 percent of estimates within 3 BPM, while remaining parameter-efficient with only 84.6K parameters. Ablation results further demonstrate the complementary contributions of the temporal and frequency branches, interaction-gated fusion, and shared regression head. These results indicate that explicitly modeling cross-domain interactions provides an effective accuracy-complexity trade-off for lightweight PPG-based HR estimation.
Introduction
Continuous heart-rate (HR) monitoring plays an important role in clinical observation, fitness assessment, ambulatory healthcare and long-term physiological monitoring [13, 1]. Although electrocardiography (ECG) remains a standard reference for cardiac monitoring, photoplethysmography (PPG) provides an attractive alternative for continuous and wearable sensing because it measures peripheral blood-volume variations using inexpensive and non-invasive optical hardware [2], [9]. Its compact sensing configuration has consequently enabled widespread integration into wearable devices, pulse oximeters, smartphones, and other resource-constrained monitoring platforms. PPG therefore represents a practical sensing modality for scalable and continuous cardiovascular monitoring [2], [1]. Despite these advantages, reliable HR estimation from PPG remains challenging. The measured waveform is influenced not only by cardiac pulsation but also by physiological and acquisition-related factors such as respiration, body site, tissue characteristics, ambient illumination, applied pressure, and artifacts [4].
A broad range of signal-processing and learning-based approaches has therefore been developed for PPG-based HR estimation. Classical methods typically rely on filtering, peak detection, spectral analysis, or handcrafted signal descriptors [12], [6]. More recently, deep learning has enabled HR-relevant features to be learned directly from PPG observations rather than relying entirely on manually designed processing rules. For example, PPGNet [17] proposed an end-to-end network operating directly on short PPG segments for device-independent HR estimation; Deep PPG [15] leverages convolutional neural networks over spectrogram of PPG and accelerometer signals; Ismail et al. [7] further explored a set of features characterizing the signal, modeling by a convolutional-recurrent framework, etc. These studies demonstrates that temporal and spectral representations of PPG can provide useful information for HR estimation.
The complementarity of temporal and spectral domain follows naturally from the characteristics of PPG. The temporal domain preserves local pulse morphology and short-term waveform dynamics, whereas the frequency domain makes periodic cardiac activity more explicit. Time-frequency representations further characterize how this spectral structure evolves over the observation interval, emphasizing different aspects of the same underlying physiological process [7]. Although prior work has demonstrated the utility of signal, spectral, and time-frequency representations, effectively structuring the interaction between complementary representations remains an important design problem. The availability of multiple representations does not by itself determine how they should be combined [5]. Under signal variability or corruption, temporal and spectral features may provide partly consistent yet partly different evidence about the underlying HR. Therefore, an effective fusion mechanism should explicitly characterize where they agree, where they differ, and how strongly each should jointly contribute to the final representation.
Motivated by these observations, we propose TF-IG, a parameter-efficient Interaction-Gated Time–Frequency learning framework for PPG-based HR estimation. First, TF-IG learns complementary HR representations through two lightweight branches: a temporal branch based on dilated depthwise-separable convolutions and temporal attention pooling, and a frequency branch that processes a bounded log-spectrogram using spectro-temporal attention pooling. Second, these representations are integrated through a structured interaction-gated mechanism. Element-wise multiplication and absolute difference explicitly encode cross-domain co-activation and discrepancy, while a feature-wise gate adaptively balances the temporal and spectral embeddings. The resulting fused representation, together with the individual branch embeddings, is processed by a shared regression head that provides a common mapping to HR, with the three predictions subsequently combined through learned normalized weights. In this way, TF-IG jointly addresses complementary representation learning, explicit cross-domain interaction, and parameter-efficient prediction within a unified lightweight framework. Accordingly, the main contributions of this work are summarized as follows:
Compact dual-domain representation learning. We propose a parameter-efficient dual-branch architecture that jointly captures temporal pulse morphology and spectro-temporal cardiac periodicity using dilated depthwise-separable convolutions and attention-based pooling.
Structured cross-domain fusion and shared prediction. We introduce an interaction-gated fusion mechanism that explicitly models temporal–spectral co-activation and discrepancy through element-wise multiplication and absolute difference, while a feature-wise gate adaptively balances the contribution of the two domains. A parameter-shared regression head is further applied to the temporal, frequency, and fused embeddings, providing a common space for HR prediction mapping while avoiding separate prediction networks.
Empirical validation of the proposed design. On a subject-independent BIDMC split, TF-IG achieves an MAE of 1.227 BPM, an RMSE of 2.060 BPM, a Pearson correlation of 0.9934, and a of 92.28% using only 84.6K parameters. Ablation experiments further demonstrate the complementary contributions of dual-domain representation learning, interaction-gated fusion, and shared regression to the final performance.
Methodology
This section first formulates PPG-based heart-rate estimation and introduces the notation used throughout the paper § 2.1. § 2.2 describes the overview of the proposed architecture. Subsequently, § 2.3 describes the signal preprocessing stage. Individual components within the framework such as time-domain branch, frequency-domain branch, fusion module, and regression head are detailed in § 2.4, § 3, § 3.1, and § 3.2, respectively.
Problem Definition and Notation
Let denote a single-channel PPG segment of duration seconds recorded at an original sampling rate . Each segment is associated with a reference heart rate , expressed in beats per minute (BPM). After preprocessing and resampling, the segment is represented as .
The objective is to learn a regression function , parameterized by , that maps the input PPG segment to its corresponding heart-rate estimate:
where denotes the predicted heart rate. The parameters are optimized by minimizing a regression loss between and the reference value over the training set.
Throughout this work, the notation represents concatenation along the feature dimension, denotes the Hadamard product, and denotes the sigmoid function.
Overall Architecture
As illustrated in Figure 1, TF-IG adopts two complementary processing branches. The time-domain branch operates directly on the PPG waveform and captures local pulse morphology and temporal dependencies. In parallel, the frequency-domain branch transforms the waveform into a bounded log-magnitude spectrogram and models cardiac periodicity over time-frequency transformed representation.
Figure 1. Overview structure of proposed framework
To reduce the computational operations, both streams use depthwise-separable convolutions [3] as their primary building block. Both branches share a structured processing flow comprising a convolutional stem, domain-specific feature refinement, and targeted attention pooling. Their outputs are converted into fixed-dimensional embeddings through domain-specific attention pooling. An interaction-gated fusion module then combines these embeddings by jointly modeling their agreement, discrepancy, and relative contribution. Finally, a shared regression head generates predictions from the time-domain, frequency-domain, and fused representations, which are combined through a learnable ensemble.
Preprocessing
Continuous PPG recordings are first partitioned into fixed-duration windows of seconds using a predefined window stride. Each window is resampled to a target sampling rate Hz and then normalized using z-score. The resulting signal is provided to both the time-domain and frequency-domain branches.
Time-Domain Branch
1D Convolutional Stem and Dilated Residual Blocks: The resampled waveform is first arranged as a single-channel sequence, . A one-dimensional convolutional stem extracts low-level morphological features while preserving the temporal resolution. The stem consists of a standard Conv1D layer () followed by two dilated depthwise-separable Conv1D layers (, with dilations and , respectively). Batch normalization and activations are applied after each layer, yielding .
Furthermore, to expand the temporal receptive field without losing resolution, is subsequently refined through three cascaded residual blocks. The transformation at the -th block is formulated as:
where . The function denotes a residual mapping comprising a dilated depthwise-separable Conv1D with kernel size and dilation rate , followed by a pointwise projection, batch normalization, and dropout. The output feature map is denoted as .
Temporal Attention Pooling (TAP): Temporal attention pooling converts into a fixed-dimensional embedding while allowing the model to emphasize informative portions of the PPG waveform. Attention logits are generated via a Temporal Transfer block including a two-layer convolutional network:
where the first projection reduces the channel dimension from to a hidden dimension , and the second produces one scalar logit per temporal position.
The normalized attention weights are
The time-domain embedding is then computed as
Unlike uniform global pooling, the TAP operation aims to assign greater importance to waveform intervals containing reliable pulse morphology.
Frequency-Domain Branch
Bounded Log-Spectrogram Representation: The frequency branch started by transforming the resampled signal into spectro-representation via the Short-Time Fourier Transform (STFT) using a Hann window. Since the dominant cardiac periodicity is expected to lie within a bounded physiological frequency range, we therefore retain only the frequency bins within the cardiac band of [0.5, 5.0] Hz to suppress irrelevant low-frequency drift and high-frequency components and to reduce the dimensionality of the spectral input.
Logarithmic dynamic-range compression is then applied as:
The resulting bounded log-spectrogram is represented as
where is the number of retained frequency bins and is the number of STFT frames.
2D Convolutional Stem: Equivalent to its 1D counterpart, is processed by a 2D convolutional stem, which share an identical structural topology. It consists of a standard Conv2D () followed by two dilated depthwise-separable Conv2D layers (, dilations ). Operating over the spatial plane, this block extracts multi-scale spectro-temporal features, outputting .
Spectro-Temporal Attention Pooling: As a non-stationary rhythmic signal, heart rate exhibits time-varying spectral energy distributed along instantaneous-frequency trajectories in the time-frequency plane [8]. Driven by these dynamic properties, Spectro-Temporal Attention Pooling module aggregates the joint spectro-temporal feature map into a fixed-length embedding to explicitly capture those joint time-frequency relationships.
Given , a Spectral Transfer block comprising a convolution, batch normalization, ReLU activation, and a second convolution produces a logit map . A spatial softmax is applied jointly over all positions:
The frequency-domain embedding is obtained as
Interaction-Gated Fusion Module
While direct concatenation or linear projection can effectively combine time- and frequency-domain representations, it does not explicitly model the cross-domain interactions between these two domains. Therefore, our aim is to achieve an explicit module that can interactively aggregated features from both time and frequency domain. Specifically, Interaction-Gated Fusion Module decomposes cross-domain fusion into two complementary components: an explicit interaction pathway that models feature relationships and a sample-dependent gate that balances the contribution of the two domains.
Explicit Cross-Domain Interaction To capture joint cross-domain relationships, we construct an extended interaction vector by combining linear terms, element-wise correlation, and absolute disparity:
Here, the multiplicative term of emphasizes the co-occurring activations across both domains, while the point-wise difference measures feature discrepancies. The extended vector is subsequently projected back to dimension via a non-linear projection block:
where , , and . Batch normalization and dropout regularization are applied following the projection.
Feature-Wise Reliability Gating In parallel, a feature-wise gate is leveraged to estimate the relative channel-wise weight of each domain directly from the concatenated representations:
where denotes the sigmoid function, , and . The learned gate modulates the combination between time and frequency-domain embeddings along each feature channel. The gated representation is computed as
Fused Representation Finally, the gated features and the interaction embedding are concatenated and projected to form the unified representation:
where , , and , followed by batch normalization and dropout.
Shared Regression Head with Learnable Ensemble
The domain-specific features and , together with the unified interaction-gated representation , pass through a regression head to produce the HR estimate. Rather than employing a distinct predictor for each representation, we formulate heart rate regression using a single parameter-shared multi-layer perceptron:
where denotes the input feature embedding, and is the predicted scalar heart rate in BPM. The shared mapping comprises a linear projection to a bottleneck dimension , followed by a ReLU activation, dropout, and a final linear layer outputting the estimated heart rate.
The final HR estimate is obtained by a learnable weighted combination of the three predictions:
where the ensemble weights are constrained to sum to 1 via a softmax function applied over learnable logit parameters:
The shared head encourages all representations to lie in a consistent HR-predictive space, while the normalized ensemble learns the overall contribution of each prediction.
Experimental Settings
Dataset and Evaluation Metrics
The experiments were conducted on the BIDMC PPG and Respiration Dataset (BIDMC), which contains physiological waveforms, including PPG, together with corresponding vital-sign numerics derived from recordings in the MIMIC-II database [14]. To prevent subject-level information leakage and evaluate generalization, we adopted a subject-exclusive split with a fixed random seed of 36. Specifically, the 53 subjects in the dataset were partitioned into a training set of 31 subjects, a validation set of 11 subjects for hyperparameter tuning and early stopping, and a held-out test set of 11 subjects for final evaluation.
Model performance was evaluated using six complementary metrics: Mean Absolute Error (MAE), Root Mean Squared Error (RMSE), Mean Squared Error (MSE), Pearson correlation coefficient (), and the coverage percentages and . The first three metrics quantify the magnitude of HR estimation errors, while measures the linear agreement between estimated and reference HR values. In addition, denotes the percentage of predictions whose absolute error is within BPM of the reference value.
Implementation Details
Training Setup and Hyperparameters: The proposed model was implemented using the PyTorch framework and trained on an NVIDIA GeForce RTX 4050 GPU. Continuous PPG signals were windowed into 30-second segments with a 2-second shift between adjacent frames. All models were trained for 150 epochs using the Adam optimizer [10] with a batch size of 128, an initial learning rate of , and a weight decay coefficient of . The hidden channel is set to 64. To facilitate stable convergence, ReduceLROnPlateau scheduler was employed to scale the learning rate by a factor of 0.5 whenever the validation loss stagnated for 8 consecutive epochs. To maintain reproducible evaluation protocol across all experiment runs, the global random seed was fixed to 36. The detail code implementation will be released upon acceptance.
The optimization process was driven by a Mean Squared Error (MSE) loss function defined as:
where and represent the ground-truth and estimated heart rate values (in BPM) for the -th sample, respectively. Best model checkpointing was dictated by the lowest validation Root Mean Squared Error (RMSE) to evaluate performance on the unseen test set.
Results and Discussion
Main Performance Comparison
To evaluate the effectiveness of the proposed TF-IG architecture, we conducted comparative experiments against both a conventional signal-processing baseline, which estimate heart rate from the dominant spectral component of the PPG signal, refered to as Welch PSD [11], and three learning-based PPG baselines: FCN, FCN_Residual, and FCN_DCT [16]. All methods were evaluated using the same preprocessing pipeline and subject-level split of the BIDMC dataset to ensure a fair comparison. Table 1 summarizes the performance of all methods across six evaluation metrics.
| Model | Params | FLOPs/sample | MAE (BPM) | RMSE (BPM) | MSE | Pearson | (%) | (%) |
| TF-IG (Ours) | 84,591 | 13.46 M | 1.2270 | 2.0601 | 4.2440 | 0.9934 | 95.98 | 92.28 |
| Welch PSD [11] | 0 | N/A | 1.0823 | 3.2015 | 10.2494 | 0.9810 | 93.24 | 90.79 |
| FCN_Residual [16] | 53,713 | 99.52 M | 1.5584 | 2.1180 | 4.4860 | 0.9918 | 97.79 | 87.45 |
| FCN [16] | 8,945 | 8.47 M | 2.9695 | 3.7681 | 14.1986 | 0.9824 | 85.72 | 61.10 |
| FCN_DCT [16] | 3,266 | 0.573 M | 5.8389 | 8.0721 | 65.1582 | 0.9122 | 57.68 | 36.85 |
Table 1. Performance and complexity comparison on the BIDMC test split under identical experimental protocols.
As shown in Table 1, TF-IG achieved strong performance across nearly all metrics, with an MAE of 1.2270,BPM, an RMSE of 2.0601,BPM, and a Pearson correlation coefficient of . Compared with the Welch PSD baseline, TF-IG produced a slightly higher MAE but substantially lower RMSE and MSE, indicating fewer large estimation errors and a more concentrated overall error distribution. This is further reflected in the strict accuracy metric, where TF-IG achieved , compared with 90.79% for Welch PSD. Moreover, TF-IG substantially outperformed all learning-based baselines in terms of MAE, RMSE, correlation, and . In particular, its exceeded the strongest learning-based baseline, FCN_Residual by 4.83 percentage points.
Model Parameter Analysis
As shown in Table 1, TF-IG contains only 84,591 trainable parameters. Although this is higher than the parameter counts of FCN-DCT, FCN, and FCN-Residual due to its dual-domain encoders, attention modules, and interaction-gated fusion, the complete model remains below 0.1 million parameters. Together with its superior estimation performance, this compact model size demonstrates a favorable trade-off between predictive accuracy and computational complexity.
Ablation Study
To evaluate the contribution of each individual component in the proposed TF-IG architecture, we conduct an ablation study across systematically isolated variants. The quantitative results are summarized in Table 2.
| Time Branch | Freq Branch | Int-Gated Fusion | Shared Head | MAE ↓ | RMSE ↓ | MSE ↓ | Pearson ↑ | (%) ↑ | (%) ↑ |
| ✓ | × | × | ✓ | 2.1677 | 3.3984 | 11.5492 | 0.9862 | 78.52 | 87.77 |
| × | ✓ | × | ✓ | 2.347 | 4.5638 | 20.8283 | 0.9713 | 83.31 | 89.02 |
| ✓ | ✓ | × | ✓ | 1.5878 | 2.464 | 6.0713 | 0.9925 | 87.17 | 92.24 |
| ✓ | ✓ | ✓ | × | 1.502 | 2.4475 | 5.9904 | 0.9926 | 84.75 | 93.85 |
| ✓ | ✓ | ✓ | ✓ | 1.227 | 2.0601 | 4.244 | 0.9934 | 92.28 | 95.98 |
Table 2. Ablation study evaluating the contribution of individual architectural components.
The single-branch variants demonstrate that both temporal and spectral representations can independently support HR estimation, although neither is sufficient to achieve the performance of the complete model. The time-only variant obtains lower overall errors than the frequency-only variant, with an MAE of 2.1677 BPM and an RMSE of 3.3984 BPM, compared with 2.3470 BPM and 4.5638 BPM, respectively. In contrast, the frequency-only variant achieves higher and values, suggesting that the two branches capture different but complementary characteristics of the PPG signal.
Combining the time and frequency branches without the interaction-gated fusion module substantially improves performance over both single-branch configurations, reducing the MAE to 1.5878 BPM and increasing the Pearson correlation to 0.9925. This result confirms that jointly exploiting waveform morphology and spectral periodicity provides a more informative representation than relying on either domain alone. Incorporating the proposed fusion module further reduces the MAE and RMSE to 1.2270 BPM and 2.0601 BPM, respectively, while increasing from 87.17% to 92.28%. These improvements indicate that explicitly modeling cross-domain interactions and adaptively balancing the two representations significantly improve model performance.
The shared regression head also contributes to the final performance. Replacing it with three distinct prediction heads increases the MAE from 1.2270 to 1.5020 BPM and decreases from 92.28% to 84.75%. Overall, the full TF-IG configuration achieves the best result across all reported metrics, validating the complementary contributions of the dual-branch design, interaction-gated fusion, and parameter-shared regression head.
Figure 2. t-SNE visualization of the representations learned by TF-IG on the BIDMC test split: (a) temporal features , (b) frequency features , and (c) fused features . Colors indicate reference-HR ranges.
Window Size Sensitivity Analysis
To investigate the effect of the observation duration on HR estimation, we evaluated TF-IG using raw PPG windows of 16 and 30 seconds. Here, the window size refers to the duration of each raw signal segment before resampling; after segmentation, both configurations were processed using the same resampling pipeline and experimental protocol described above. The results are reported in Table 3.
| Raw Window | MAE (BPM) ↓ | RMSE (BPM) ↓ | MSE ↓ | Pearson ↑ | (%) ↑ | (%) ↑ |
| 16 seconds | 1.9331 | 3.2854 | 10.7938 | 0.9856 | 82.56 | 89.58 |
| 30 seconds | 1.2270 | 2.0601 | 4.2440 | 0.9934 | 92.28 | 95.98 |
Table 3. Performance of TF-IG with different raw PPG window durations before resampling.
Increasing the raw window duration from 16 to 30 seconds consistently improves performance across all evaluation metrics. In particular, the MAE decreases from 1.9331 to 1.2270 BPM, while the RMSE decreases from 3.2854 to 2.0601 BPM. The longer window also increases and by 9.72 and 6.40 percentage points, respectively, and raises the Pearson correlation from 0.9856 to 0.9934.
These results suggest that the 30-second raw window provides the model with a broader temporal context and a larger number of cardiac cycles, supporting more stable extraction of waveform morphology and spectral periodicity after resampling. Consequently, the 30-second configuration is adopted for the main experiments. Nevertheless, the 16-second setting remains a potential alternative when a shorter observation interval is preferred, albeit with reduced estimation accuracy.
Feature Space Representation Analysis
To investigate how the two modality-specific branches and the proposed fusion module organize heart-rate information, we extracted the temporal feature , frequency feature , and fused feature for all windows in the held-out BIDMC test set. Each representation is a -dimensional vector obtained before the regression head. For qualitative visualization, the features were standardized independently and projected to two dimensions using t-SNE. Points were colored according to five reference-HR ranges: , 60–80, 80–100, 100–120, and BPM with their corresponding number of windows belong to that range.
The temporal representation is distributed across several separated groups, with samples from the same HR range appearing in different regions. This suggests that the temporal branch can capture not only HR-related dynamics but also variations in waveform morphology, signal quality, and subject-specific characteristics. In contrast, the frequency representation exhibits more compact and localized organization according to HR, consistent with its direct encoding of cardiac periodicity. However, overlap between neighboring HR ranges remains. It may be due to their continuous boundaries and the effects of spectral noise, harmonics, and physiological variability between subjects. The fused representation integrates temporal morphology and spectral periodicity, reducing ambiguities present in either individual branch and producing a more balanced organization of HR-related information. Overall, the visualization supports the complementary roles of the temporal and frequency branches and the effectiveness of their fusion.
Conclusion
This work introduced TF-IG, a parameter-efficient framework for PPG-based heart-rate estimation that jointly models temporal morphology and spectral periodicity. Its interaction-gated fusion captures cross-domain agreement and discrepancy while adaptively balancing their contributions, and a shared regression head provides a compact common mapping to the final HR prediction. Evaluation on the subject-exclusive BIDMC split demonstrates that TF-IG consistently outperforms the evaluated baselines across most of error-based, correlation, and strict accuracy metrics, while maintaining a compact parameter footprint. Ablation studies further validate the complementary roles of dual-domain representation learning, interaction-gated fusion, and shared regression in improving estimation performance. Overall, these results demonstrate that explicitly modeling temporal–spectral interactions provides an effective accuracy–complexity trade-off for efficient PPG-based HR estimation.
References
- [1]Almarshad, M.A., Islam, M.S., Al-Ahmadi, S., BaHammam, A.S.: Diagnostic features and potential applications of ppg signal in healthcare: a systematic review. In: Healthcare. vol. 10, p. 547. MDPI (2022)DOI
- [2]Castaneda, D., Esparza, A., Ghamari, M., Soltanpur, C., Nazeran, H.: A review on wearable photoplethysmography sensors and their potential future applications in health care. International journal of biosensors & bioelectronics 4(4), 195 (2018)DOI
- [3]Chollet, F.: Xception: Deep learning with depthwise separable convolutions. In: 2017 IEEE conference on computer vision and pattern recognition (CVPR). pp. 1800–1807. IEEE (2017)arxiv.org/abs/1610.02357
- [4]Fine, J., Branan, K.L., Rodriguez, A.J., Boonya-anant, T., Ajmal, Ramella-Roman, J.C., McShane, M.J., Coté, G.L.: Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring. Biosensors 11(4) (2021). https://doi.org/10.3390/bios11040126, https://www.mdpi.com/2079-6374/11/4/126
- [5]Gao, J., Li, P., Chen, Z., Zhang, J.: A survey on deep learning for multimodal data fusion. Neural computation 32(5), 829–864 (2020)DOI
- [6]Ismail, S., Akram, U., Siddiqi, I.: Heart rate tracking in photoplethysmography signals affected by motion artifacts: A review. EURASIP Journal on Advances in Signal Processing 2021(1), 5 (2021)DOI
- [7]Ismail, S., Siddiqi, I., Akram, U.: Heart rate estimation in ppg signals using convolutional-recurrent regressor. Computers in Biology and Medicine 145, 105470 (2022)DOI
- [8]Khan, N.A., Jönsson, P., Sandsten, M.: Performance comparison of time-frequency distributions for estimation of instantaneous frequency of heart rate variability signals. Applied Sciences 7(3) (2017). https://doi.org/10.3390/app7030221, https://www.mdpi.com/2076-3417/7/3/221
- [9]Kim, K.B., Baek, H.J.: Photoplethysmography in wearable devices: a comprehensive review of technological advances, current challenges, and future directions. Electronics 12(13), 2923 (2023)DOI
- [10]Kingma, D.P., Ba, J.: Adam: A method for stochastic optimization. arXiv preprint arXiv:1412.6980 (2014)arxiv.org/abs/1412.6980
- [11]Lee, H., Ko, H., Chung, H., Nam, Y., Hong, S., Lee, J.: Real-time realizable mobile imaging photoplethysmography. Scientific reports 12(1), 7141 (2022)DOI
- [12]Pankaj, Kumar, A., Komaragiri, R., Kumar, M.: A review on computation methods used in photoplethysmography signal analysis for heart rate estimation. Archives of Computational Methods in Engineering 29(2), 921–940 (2022)
- [13]Park, J., Seok, H.S., Kim, S.S., Shin, H.: Photoplethysmogram analysis and applications: an integrative review. Frontiers in physiology 12, 808451 (2022)DOI
- [14]Pimentel, M.A.F., Johnson, A.E.W., Charlton, P.H., Birrenkott, D., Watkinson, P.J., Tarassenko, L., Clifton, D.A.: Toward a robust estimation of respiratory rate from pulse oximeters. IEEE Transactions on Biomedical Engineering 64(8), 1914–1923 (2017). https://doi.org/10.1109/TBME.2016.2613124
- [15]Reiss, A., Indlekofer, I., Schmidt, P., Van Laerhoven, K.: Deep ppg: Large-scale heart rate estimation with convolutional neural networks. Sensors 19(14), 3079 (2019)DOI
- [16]Samavati, T., Farvardin, M., Ghaffari, A.: Parameter-efficient deep learning models for vital sign estimation from ppg. Healthcare Technology Letters 13(1), e70083 (2026). https://doi.org/10.1049/htl2.70083, https://ietresearch.onlinelibrary.wiley.com/doi/abs/10.1049/htl2.70083
- [17]Shyam, A., Ravichandran, V., Preejith, S., Joseph, J., SivaprakSam, M.: Ppgnet: Deep network for device independent heart rate estimation from photoplethysmogram. In: 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). pp. 1899–1902. IEEE (2019)arxiv.org/abs/1903.08912