WorksEngineering
Bayesian Joint Reconstruction and Uncertainty Quantification for Limited-Data Multi-Contrast X-Ray CT
Abstract
Multi-contrast X-ray computed tomography (CT) combines attenuation contrast with phase and dark-field contrast, which are sensitive to refraction and ultra-small-angle scattering, respectively. Common implementations record several intensity measurements per projection angle by phase or mask stepping, which lengthens acquisition and, at a fixed exposure per measurement, raises the dose. Limited-data acquisition via sparse angular sampling or reduced stepping lowers these costs but makes the data less informative, calling for prior information and uncertainty quantification. We study joint Bayesian reconstruction of attenuation, phase, and dark-field images from separately retrieved sinograms. A joint total variation term in the prior couples edge locations across the images but not their values. Posterior sampling with the No-U-Turn Sampler complements maximum a posteriori estimates with pixel-wise uncertainty. We compare this joint prior with channel-wise total variation alone on synthetic and experimental Talbot-Lau interferometry data under full, sparse-angle, sparse-step, and combined acquisition (up to 24 times fewer exposures). Both maximum a posteriori reconstructions suppressed undersampling artifacts, with synthetic relative errors 1.6-11 times lower than those of filtered backprojection. The joint prior gave modest synthetic accuracy gains, a cleaner experimental dark-field background, 23-30 % lower mean dark-field posterior standard deviation, and 4-38% higher median effective sample size in experimental limited-data runs. In the synthetic study, posterior standard deviations matched or exceeded the errors in flat regions but underestimated them at boundaries, especially in phase. Experimental phase chains mixed slowly under angular undersampling. The formulation also applies to edge illumination with adapted forward and noise models.
Introduction
Conventional X-ray computed tomography (CT) provides limited contrast between tissues and materials with similar attenuation properties [1, 2, 3]. Phase-contrast and dark-field imaging provide complementary sensitivity to refraction and ultra-small-angle scattering, respectively [4, 5]. We refer to the combined use of attenuation, phase, and dark-field information for tomographic reconstruction as multi-contrast X-ray CT.
Talbot–Lau interferometry (TLI) and edge illumination (EI) provide these contrasts using conventional X-ray sources [6, 7]. Common implementations acquire multiple intensity measurements at each projection angle through phase stepping in TLI [4, 5] or mask stepping in EI [7]. These measurements sample an intensity curve at each detector pixel. Relative to a reference scan without the object, attenuation lowers the mean of this curve, and refraction shifts it laterally, whereas ultra-small-angle scattering smears it out, which lowers its visibility (amplitude relative to the mean) in TLI and broadens it in EI [5, 8]; the three signals can therefore be separated. The measured intensities can be reconstructed directly in the intensity domain [8, 9, 10, 11], or by a two-step pipeline: signal retrieval first extracts attenuation, differential phase, and dark-field sinograms, which are then reconstructed by filtered backprojection (FBP) [5, 12, 13] or by iterative regularized methods [14].
The additional measurements increase acquisition time compared with conventional attenuation-based CT [11, 15] and, at a fixed exposure per measurement, the dose; gratings or masks behind the sample further raise the dose needed for a given image quality, because they absorb photons that have already passed through it [8, 13]. Sparse angular sampling [14, 16] and reduced stepping [9, 10] can lower the acquisition time and the dose per scan; we refer to these strategies collectively as limited-data multi-contrast CT. The two strategies degrade the data in different ways. Sparse angular sampling leaves the tomographic problem underdetermined, so that FBP produces streak artifacts [17], whereas reduced stepping at a fixed exposure per step records fewer photons and therefore increases the noise in the retrieved sinograms [18, 19]. Regularized iterative reconstruction suppresses streaks and noise by introducing prior information [17]. The prior, however, also shapes features that the data constrain only weakly, and a regularized image alone does not reveal which of its features are supported by the measurements. Limited-data multi-contrast CT therefore raises two questions: what prior information should be imposed on the contrast images, and how the uncertainty that remains after regularization can be quantified.
For the first question, shared structure is a natural source of prior information: the contrast images describe different physical properties of the same object but may share boundaries despite differences in their amplitudes. Several joint reconstruction methods estimate all contrast channels directly from the measured intensities [8, 9, 10, 11, 20]. Because each intensity depends on every channel, these methods couple the channels through the measurement model rather than through a prior; regularization, when used, is often applied to each channel separately [9, 11, 20, 21]. In EI simulations, fixing one channel to a noisy or blurred estimate had little influence on the reconstruction of the other [21]; likewise, in grating-based reconstructions, each channel’s penalty mainly affected its own image [9]. These observations indicate robustness in the studied settings, but they do not show that shared structure is used to support a weaker channel.
Other joint reconstruction methods couple the channels explicitly. One penalizes differences between the phase image and a tissue-dependent mapping of the attenuation image [22], and another imposes such a material relationship exactly to reconstruct a single fused image [23]. These methods couple voxel values through a prescribed material model. Structural regularizers instead couple the locations or orientations of edges and leave the image values free. They have been used in joint positron emission tomography/magnetic resonance imaging [24] and multi-energy CT [25, 26], and for attenuation and phase in grating-based radiography [27]. For multi-energy CT with fewer, non-overlapping projections per energy, a comparison of several joint regularizers, among them joint total variation (JTV), found that joint regularization outperformed channel-by-channel reconstruction [26]. Phase-guided structural coupling has recently been used for attenuation reconstruction in near-field holotomography [28]. These studies motivate priors that share spatial information across attenuation, phase, and dark-field images while allowing their values to differ.
Bayesian reconstruction addresses the second question: the posterior distribution of the unknown images, which combines the prior with the likelihood of the data, characterizes the uncertainty that remains under the assumed models and complements point estimates with pixel-wise uncertainty maps [29]. Previous CT studies have quantified uncertainty in reconstructed images [30, 31, 32], inclusion boundaries [33], and suspected emboli [34], including under sparse-angle and limited-angle acquisition. Posterior uncertainty quantification (UQ) for jointly reconstructed attenuation, phase, and dark-field images remains comparatively little explored.
In this work, we study a joint Bayesian reconstruction for limited-data multi-contrast CT from separately retrieved sinograms. Because each retrieved sinogram depends on a single contrast image, the channels interact only through the prior when the channel noise is modeled as independent. The prior combines channel-wise total variation (TV) with a smoothed JTV term that couples edge locations across the three contrast images but not their values. Maximum a posteriori (MAP) estimates are complemented by posterior sampling with the No-U-Turn Sampler (NUTS) [35], which provides pixel-wise uncertainty. Using synthetic and experimental TLI data, we compare channel-wise and joint priors under four acquisition conditions (full data, sparse angles, sparse steps, and combined undersampling) and assess reconstruction accuracy, posterior variability (pixel-wise standard deviation), its calibration against actual error in the synthetic study, and sampling efficiency. These ingredients are established; our contribution is their combination for multicontrast TLI tomography and an assessment of what the joint prior and the posterior summaries deliver, and where they fall short.
Methods
We formulate multi-contrast CT reconstruction from retrieved sinograms as a Bayesian inverse problem, combining channel-wise forward and noise models with prior information to estimate the attenuation, phase, and dark-field images and quantify their uncertainty.
Forward Problem
The forward problem is to predict the noise-free attenuation, differential phase, and dark-field sinograms from given cross-sectional parameter images and a known acquisition geometry. Let denote the discretized attenuation coefficient, refractive-index decrement, and dark-field scattering coefficient on an image grid with pixels. After signal retrieval, contain the logarithmic attenuation, differential phase-contrast (DPC), and logarithmic dark-field sinograms. The discrete line-integral projector is , where for projection angles and detector positions. It approximates each line integral by quadrature along the ray, evaluating the image at the quadrature points by nearest-neighbor or bilinear interpolation. The forward maps for , , and are
For DPC, is the detector-coordinate derivative in the interferometer’s sensitivity direction, applied separately at each angle [4, 36]. The geometry-dependent constant that converts the refraction angle into the measured phase shift is absorbed into , which is therefore reconstructed in scaled units; the sign convention depends on the setup. The dark-field model assumes that the visibility, like the intensity, decays exponentially along each ray [5, 37].
Let stack , and let stack . The observation model is
where stacks , , and in that order, and includes measurement and signal-retrieval errors.
Bayesian Inverse Problem
The inverse problem is to estimate the cross-sectional parameters from the noisy multi-contrast observations in (4). In the Bayesian formulation, the parameters and noise are modeled as random vectors and , giving the random data through
The observed data are a realization of .
Let denote the prior density of the parameters and the noise density. Assuming is independent of , the additive model (5) gives the likelihood
The Bayesian solution is the posterior distribution of the parameters conditioned on the observed data. By Bayes’ formula, its density is [29]
where is the normalizing constant, also called the evidence. Below, we specify the likelihood and prior and describe the resulting posterior.
Likelihood
We model the blocks of for , , and as mutually independent Gaussian random vectors, , , where is the identity matrix and is the corresponding channel variance, treated as known (Section III). The likelihood is
with denoting the Euclidean norm.
Prior
For , we consider two TV-based edge-preserving priors [17], [38]. TV priors are not discretization-invariant [39], so we use them at a fixed image resolution. For , represented on an grid with , the smoothed isotropic TV functional is
Here contains the forward differences of at pixel in the two grid directions, and the sum runs over the pixels at which both differences are defined. The smoothing parameter makes TV differentiable for optimization and sampling.
Because the channels differ in physical units and magnitude, the priors act on the scaled images , , and , with fixed channel scales , given in Section III; for brevity, the scales are omitted below. The channel-wise baseline prior is
where the channel weights set the strength of the TV term in each channel. The TV penalties are invariant to constant shifts and therefore do not define a proper density on their own [40]; the amplitude penalty with weight bounds the density by a Gaussian, so the prior is proper on .
Structural coupling is introduced through
which groups the channel gradients at each pixel. Because the Euclidean norm of the stacked gradients never exceeds the sum of the individual gradient magnitudes, edges that coincide across channels cost less than the same edges placed at different pixels. therefore favors co-located edges without tying the size or sign of the jumps across channels. The joint prior is
where is the coupling weight; recovers .
The channel weights , the coupling weight , the amplitude weight , and the smoothing parameter are referred to collectively as the prior parameters.
Posterior
For , let denote the posterior obtained by setting in (7). Up to an additive constant independent of , the negative log-posterior is
where is the negative log-prior. The likelihood (8) and factorize over channels, so only the JTV term couples the channels. Since the forward maps are linear and the smoothed TV and JTV terms are convex, the quadratic amplitude penalty makes strongly convex, so both posteriors are log-concave with a unique mode.
Inference
MAP Estimate
The MAP estimate minimizes the negative log-posterior:
This is a variational regularization problem with regularization functional [29], so the MAP estimate is also a regularized reconstruction and the prior weights act as regularization parameters.
Sampling and UQ
Posterior samples are generated using the No-U-Turn Sampler (NUTS) [35], a gradient-based Hamiltonian Monte Carlo method previously applied to single-channel CT [31]. The samples , , are the post-warmup draws of several independent NUTS chains. We estimate the posterior mean and pixel-wise standard deviation by
where powers are evaluated component-wise. The map quantifies posterior variability conditional on the assumed forward model, noise model, and prior. We assess convergence with the Gelman–Rubin statistic and sampling efficiency with the effective sample size (ESS) [41], both computed per pixel. Optimization and sampling settings for the synthetic and experimental datasets are specified in Section III.
Numerical Experiments and Results
We compare FBP with the channel-wise TV (10) and JTV (12) priors using synthetic data with known ground truth and experimental TLI data. Each dataset is tested with full data, sparse angles, sparse steps, and combined undersampling. We assess MAP reconstructions and posterior UQ against the ground truth for synthetic data; experimental evaluation uses data consistency and, as surrogate references, full-data reconstructions.
All reconstructions and posterior sampling ran on an NVIDIA RTX 6000 Ada Generation GPU (48 GB) under Ubuntu 22.04.5 LTS, using Python 3.14.2, NumPy 2.4.2, scikit-image 0.26.0 [42], PyTorch 2.10.0 [43] with CUDA 13.0, ASTRA Toolbox 2.4.1 [44], and Pyro 1.9.1 [45]. The code, including the experimental data used here, is available at https://github.com/hassanyazdanian/Bayes_mc_CT.
Synthetic Phantom
Phantom and Data Simulation
The synthetic phantom is a piecewise-constant disc in normalized units (Figure 1). Inclusions A and B share boundaries across channels, with contrast changes relative to the disc of and in , respectively. They test coupling of edge locations across different contrast amplitudes. Cross-talk is tested by a foam-like inclusion F, weak in and ( and ) but dominant in (), and an attenuation-matched inclusion P visible only in . A ladder of five decreasing-size discs, a low-contrast ellipse L ( in and , absent in ), and a one-pixel-wide arc-shaped crack, strong only in , provide small or low-contrast features for the UQ study.

Figure 1. Synthetic multi-contrast phantom: ground-truth attenuation , phase , and dark-field images. Inclusions A and B share their boundaries across the channels, F is -dominant, and P differs from the disc only in ; the low-contrast ellipse L, the ladder of decreasing-size discs, and the thin -dominant crack serve the UQ study.
Phase-stepping intensities are simulated with the Talbot–Lau model at visibility and processed by the harmonic analysis used for the experimental data (see Sections III-B2 and III-B4). To avoid an inverse crime, the phantom is rendered on a grid and block-averaged to for simulation with a bilinear-interpolation projector. Ground truth is block-averaged to the reconstruction grid, where a nearest-neighbor projector is used. All three images are scaled by one common factor so that , below the phase-wrapping limit of . Full data use 180 views over and 10 phase steps; the sparse settings use 30 views and 4 steps (the experimental study uses 5 steps, since only divisors of its 10 recorded steps give evenly spaced subsets).
Independent zero-mean Gaussian noise is added to the retrieved attenuation, DPC, and dark-field sinograms at 1%, 4.7%, and 4.8% of their noiseless root-mean-square (RMS) values. The DPC and dark-field levels are the attenuation level times the phase-stepping noise factors and [18, 19, 46], which are derived for absolute noise levels and applied here to relative levels because the phantom is in normalized units. For steps, the levels are scaled by . On full data, the RMS difference between the noiseless simulated sinograms and the reconstruction model applied to the ground truth is approximately 1.2, 5.1, and 0.4 noise standard deviations in attenuation, DPC, and dark-field, respectively; differentiation amplifies discretization differences in DPC.
Reconstruction and Evaluation Settings
The synthetic fan-beam geometry has source-to-object and source-to-detector distances of 0.20 and 0.40 m, with 256 detector elements spanning 120 mm. Reconstruction uses a grid over a 60 mm field of view and quadrature points per ray. DPC follows (2) with second-order central differences, and the likelihood uses the standard deviations of the added noise. The priors act on the unscaled images (), whose channels have comparable magnitudes. Bayesian inference is restricted to a circular support of radius 0.98 in normalized image coordinates, in which the reconstruction grid spans . Outside this support, the images are fixed at zero. Fan-beam FBP uses a Ram–Lak filter on the full square grid, with integrated phase obtained by Fourier integration of the noisy DPC.
MAP estimates use L-BFGS with a strong-Wolfe line search [47], at most 1000 iterations, and zero initialization for both priors. The same prior parameters are used in all acquisition conditions. A joint sweep of and the channel weights on full data gave similar errors for (within 3.4% of each channel’s best value). We use to avoid the dark-field background texture at and the stiffer posterior, which is harder to sample, at smaller values. Minimizing the relative error (RE) averaged over all four conditions gives , within 4% of each per-condition optimum. Tuning on the full data alone would over-regularize the undersampled cases: the full-data optimum more than doubles the combined-undersampling phase error. The amplitude weight is the largest tested value that changes RE by less than 0.3% in every channel and condition. The coupling weight minimizes the average change in summed per-channel RE relative to TV, reducing that sum in every condition. Tuning against ground truth on the evaluated noise realization gives best-case performance for each prior.
Posterior samples are drawn with NUTS in normalized coordinates, each channel divided by the 99th percentile of its MAP magnitude. Two chains of 1000 samples are drawn after 500 warmup iterations, during which the step size (initial value ) and a diagonal mass matrix are adapted, with a maximum tree depth of six. Each chain starts from the MAP estimate of the corresponding prior with an independent perturbation of standard deviation . The sampling dimension is 147 192, i.e. 49 064 active pixels per channel.
Reconstructions are scored per channel against the ground truth by the RE, computed on the full image, and by the structural similarity (SSIM) index [48] and the peak signal-to-noise ratio (PSNR). For SSIM and PSNR, the true and reconstructed images are mapped by the same affine transformation, taking the range of the true image to , and the reconstruction is clipped to that interval; both indices use the scikit-image [42] implementation at its default settings with unit data range.
MAP Reconstructions
Figure 2 and tbl:I compare the reconstructions. Under combined undersampling, the RE of FBP reaches 0.425 in and 1.56 in ; the phase channel remains more robust after Fourier integration (0.170). The Bayesian RE is 0.066–0.071 in and and 0.111–0.146 in , 1.6–11 times lower than FBP. For FBP and for the Bayesian dark-field channel, reducing the number of views degrades the reconstruction more than reducing the phase steps, whereas the Bayesian and errors vary by less than 5% across conditions.

Figure 2. Synthetic-data reconstructions of the attenuation (), phase (), and dark-field () channels under different acquisition conditions. (a) Full data: 180 projection angles and 10 phase steps; (b) sparse angles: 30 projection angles and 10 phase steps; (c) sparse steps: 180 projection angles and 4 phase steps; and (d) combined undersampling: 30 projection angles and 4 phase steps. For each acquisition condition, the rows show FBP, channel-wise TV, and JTV reconstructions, respectively. Within each panel, the three reconstructions of a given contrast are displayed using the same intensity scale, indicated by the corresponding colorbar.
JTV lowers RE in 11 of the 12 channel–condition pairs; the exception is under combined undersampling (+0.4%). Gains are modest, reaching 3.1% in full-data , with both priors tuned near their individual optima. Dark-field SSIM increases by 0.008–0.013 in every condition, consistent with improved low-contrast boundaries. JTV also does not transfer channel-specific features: the magnitude of the contrast error of P in and and of F in and , measured against the surrounding disc, stays within 0.01 percentage points of TV or decreases, by up to 0.25 points.
MAP estimation took 3–8 s per reconstruction on the GPU, against 2–5 ms for FBP.
| Channel | Method | (a) Full data RE ↓ | (a) Full data SSIM ↑ | (a) Full data PSNR ↑ (dB) | (b) Sparse angles RE ↓ | (b) Sparse angles SSIM ↑ | (b) Sparse angles PSNR ↑ (dB) | (c) Sparse steps RE ↓ | (c) Sparse steps SSIM ↑ | (c) Sparse steps PSNR ↑ (dB) | (d) Combined RE ↓ | (d) Combined SSIM ↑ | (d) Combined PSNR ↑ (dB) |
| FBP | 0.1472 | 0.265 | 23.03 | 0.3066 | 0.114 | 17.41 | 0.1903 | 0.174 | 20.96 | 0.4253 | 0.079 | 14.71 | |
| TV | 0.0673 | 0.963 | 29.52 | 0.0677 | 0.948 | 29.46 | 0.0678 | 0.958 | 29.45 | 0.0705 | 0.923 | 29.11 | |
| JTV | 0.0669 | 0.964 | 29.57 | 0.0674 | 0.948 | 29.51 | 0.0674 | 0.959 | 29.50 | 0.0700 | 0.921 | 29.17 | |
| FBP | 0.1097 | 0.740 | 24.32 | 0.1687 | 0.432 | 21.61 | 0.1100 | 0.728 | 24.30 | 0.1695 | 0.419 | 21.57 | |
| TV | 0.0676 | 0.937 | 28.59 | 0.0656 | 0.957 | 28.80 | 0.0666 | 0.954 | 28.68 | 0.0672 | 0.953 | 28.58 | |
| JTV | 0.0674 | 0.940 | 28.60 | 0.0655 | 0.958 | 28.81 | 0.0665 | 0.955 | 28.70 | 0.0675 | 0.953 | 28.54 | |
| FBP | 0.4110 | 0.156 | 23.16 | 1.0385 | 0.051 | 16.15 | 0.6325 | 0.092 | 19.82 | 1.5636 | 0.030 | 13.02 | |
| TV | 0.1147 | 0.949 | 33.34 | 0.1284 | 0.933 | 32.32 | 0.1177 | 0.950 | 33.09 | 0.1456 | 0.904 | 31.23 | |
| JTV | 0.1111 | 0.961 | 33.59 | 0.1282 | 0.941 | 32.33 | 0.1176 | 0.958 | 33.09 | 0.1429 | 0.917 | 31.39 |
Table 1. Synthetic-data reconstruction accuracy against the ground truth. The better of TV and JTV is shown in bold where the displayed values differ. TV and JTV use at most 1000 L-BFGS iterations.
| Scenario | Prior | flat | edge | flat | edge | flat | edge |
| Full data | TV | 0.45 | 2.92 | 1.01 | 10.33 | 0.47 | 1.30 |
| JTV | 0.44 | 3.21 | 1.02 | 10.54 | 0.45 | 1.51 | |
| Sparse angles | TV | 0.51 | 2.50 | 0.79 | 5.74 | 0.37 | 1.27 |
| JTV | 0.56 | 2.82 | 0.82 | 5.90 | 0.42 | 1.63 | |
| Sparse steps | TV | 0.43 | 2.58 | 0.79 | 8.76 | 0.44 | 1.21 |
| JTV | 0.43 | 2.89 | 0.79 | 8.92 | 0.42 | 1.48 | |
| Combined | TV | 0.62 | 2.77 | 0.92 | 5.34 | 0.38 | 1.35 |
| JTV | 0.74 | 3.10 | 0.95 | 5.48 | 0.46 | 1.76 |
Table 2. Calibration of the synthetic posteriors: RMS error of the posterior mean divided by RMS posterior standard deviation, for pixels on true edges and in flat regions. A calibrated posterior gives one; larger values indicate overconfidence.
Posterior Sampling and UQ
Diagnostics over 2000 randomly selected coordinates gave a maximum of 1.02–1.04 for 180 views and 1.10–1.17 for 30 views, with median . Median ESS values were 480–745 and 143–189, respectively, with minima of 142–311 and 5–13. Thus, some sparse-angle coordinates remained slowly mixing. Two-chain runs took 19 min at 30 views and 94 min at 180 views (0.39 and 1.89 s per iteration).
Figure 3 shows posterior means , standard deviations , and absolute standardized errors , computed pixel-wise, for each channel. From full data to combined undersampling, the mean posterior standard deviation of the TV posterior increases by 30 %, 57 %, and 36 % in , , and , with uncertainty concentrated near boundaries. Relative to TV, JTV reduces mean standard deviation by 12–14 % in , about 2 % in , and 23–27 % in . For full-data , JTV lowers the posterior-mean RE from 0.160 to 0.138, compared with 0.115 to 0.111 for the MAP estimate. The posterior mean remains less accurate than MAP in and , while for they differ by at most 1.1 %.

Figure 3. Synthetic-data posterior UQ for the four acquisition scenarios: (a) full data, (b) sparse angles, (c) sparse steps, and (d) combined undersampling. For each scenario, TV and JTV are compared in terms of the posterior mean, the pixel-wise posterior standard deviation , and the absolute standardized error , i.e. the posterior-mean error in units of (Section III-A4), for the attenuation , phase , and dark-field channels. Mean and standard deviation use a common scale per channel across all scenarios and priors; uses a logarithmic scale centered on the calibrated value , with blue indicating a posterior wider than its error and red indicating overconfidence, saturating at 0.25 and 4. The colorbars at the foot of the figure apply to all panels.
Table II reports the RMS error of the posterior mean divided by the RMS posterior standard deviation, on edge pixels and on the remaining flat pixels. In flat regions, the ratio is 0.37–0.74 for and , with 90–99 % of pixels within one standard deviation, and 0.79–1.02 for , with 73–83 %. At edges, the ratios are 2.5–3.2 for , 5.3–10.5 for , and 1.2–1.8 for ; two-standard-deviation coverage falls to 35–81 % in and 49–65 % in . Against Gaussian reference coverages of 68 % and 95 %, these results indicate conservative uncertainty for and in flat regions, roughly matched uncertainty for there, and underestimated errors at boundaries, also visible in the standardized error maps.
For , the edge-error ratio decreases from 10.3 to 5.3 as data are removed, while the edge error stays near 0.026: posterior variability grows while systematic error persists. Reconstructing with the projector used for data simulation (Section III-A1) reduces the edge error to 0.016, about 40 % lower. This points to model discrepancy and grid-representation errors that the assumed-exact forward model does not include in the posterior. These are per-pixel marginal diagnostics and do not assess joint posterior coverage.
Real Data: Talbot–Lau Interferometry
TLI Setup and Phantom
Measurements were acquired at the University of Helsinki with the TLI setup of [13] (Figure 4) and are openly available [49]. The gratings G0, G1, and G2, have vertical line structures with periods (14, 3.5, 2.0) m, heights (50, 36, 22) m, and duty cycles (0.3, 0.5, 0.5). The G0 absorption grating provides spatial coherence at the tube window. G1 is 140 cm downstream and produces a phase shift at 28 keV. The G2 absorption grating, 20 cm beyond G1 at the fifth fractional

Figure 4. Schematic of the Talbot–Lau interferometry setup and phantom used in the experiments. The vertical cavities (A, B, and C) are filled with sugar–water solutions of 10, 20 and 30 m-%. The water container surrounding the sample is omitted for clarity.
Talbot distance, analyzes the interference pattern. The gratings were fabricated at the Paul Scherrer Institut on silicon using photolithography, wet etching, and gold electroplating for G0 and G2 [50].
The source (XRD Glass Tube, Malvern Panalytical) has a stationary tungsten anode, a 0.4 × 8 mm focal spot viewed at a take-off angle, giving an effective focal spot of 0.4 mm vertical by 0.8 mm horizontal, and a 0.3 mm beryllium window.
Object and reference scans (without the object) used 10 phase steps over one G2 period at each of the 361 projection angles over a full rotation, with 18 s exposure per step at 35 kV and 40 mA, i.e. about 18 h of exposure per scan and about 40 h in total, including sample repositioning.
The phantom (Figure 4), also used in [13], is a 15 mm-diameter PMMA rod with three 4 mm-diameter vertical cavities filled approximately with 10, 20, and 30 m-% sugar–water solutions. At 5 cm before G1, it was submerged in a rectangular water container to reduce phase wrapping and beam-hardening artifacts. With the container in the beam, the mean reference visibility was about 15%.
Data Retrieval and Preprocessing
Slice reconstruction uses 843 pixels of the central detector row. At each of the 361 projection angles, the stepping intensity is modeled as [5]
where are the step positions and , , and are mean intensity, visibility, and modulation phase. Single-harmonic Fourier analysis estimates these parameters from the steps for object and reference scans, giving
with wrapped to . and are line integrals, while is the derivative along the detector of the line integral of , matching the forward model of Section II-A.
Discarding one outermost detector bin leaves 842 bins. For each projection and channel, the median of the 50 sample-free bins at either edge is subtracted to remove the detector baseline; the sign of is set to the forward-model convention. FBP obtains integrated phase from by Fourier integration, using the same background bins to fix the integration constant. Bayesian reconstruction uses directly.
Intensity noise is estimated from reference scan stepping-curve content outside the zeroth (DC) and first harmonics. Monte Carlo propagation through the complete retrieval pipeline uses 50 noisy realizations of the object and reference scans; empirical standard deviations give , , and , including nonlinear noise amplification at finite visibility. Noise propagation is repeated for the five-step retrieval, giving increases of 1.413, 1.416, and 1.416 in , , and , respectively, approximately . Each acquisition condition uses the noise levels estimated for its step count, held fixed during reconstruction.
Full data use 361 angles and 10 phase steps. Sparse-angle data use 30 evenly spaced projections shared across channels; because both the and views are included, they cover 29 distinct angles. Sparse-step data use 5 evenly spaced steps spanning one grating period; combined undersampling applies both reductions. The three subsets contain 12, 2, and 24 times fewer object exposures than the full scan (3610 exposures, 18.1 h at 18 s each); at fixed tube settings, exposure time and dose fall by the same factors.
Reconstruction and Evaluation Settings
The fan-beam model uses source-to-object and source-to-detector distances of 1.35 and 1.63 m, 842 detector bins of 74.8 µm pitch, and a calibrated detector-center offset of four pixels. Reconstruction uses a grid over 52.2 mm (102 µm pixels), 842 quadrature points per ray, and a nearest-neighbor projector; FBP uses a Ram–Lak filter. Bayesian inference is restricted to a circular support of normalized radius 0.5 (13.1 mm), outside which the images are fixed at zero; the support contains the rod and a margin of the surrounding water, with 51 288 active pixels per channel. The likelihood uses and the estimated noise levels.
The priors use channel scales equal to half the range between the 5th and 95th percentiles of each channel’s full-data FBP image, so the channel and coupling weights and the smoothing parameter are not directly comparable with the synthetic values. Prior parameters are selected as described in Section III-B5. MAP estimates for both priors use zero-initialized L-BFGS with a strong-Wolfe line search and at most 150 iterations. Posterior sampling uses the NUTS settings of the synthetic study (Section III-A2), two chains of 1000 samples, but with 200 warmup iterations, and is performed on the scaled images , , and . The sampling dimension is 153 864.
Without ground truth, we assess noise-normalized sinogram residuals and visual agreement with the full-data reconstructions, and compare posterior means and standard deviations between priors.
Forward-Model Adequacy
We assess the channel models (1)–(3) and calibrated geometry by a round trip through the measured intensity, retrieval, and reconstruction pipeline.
Full-data FBP of the retrieved attenuation, integrated-phase, and dark-field sinograms gives . Inverting (18) generates simulated object intensities using the measured reference Fourier coefficients and :
evaluated entry-wise. Measured reference coefficients retain the instrument response, including source spectrum, grating imperfections, and beam-profile variations, so the test isolates the object-dependent model. The simulated intensities pass through the same retrieval and FBP pipeline. We compare measured and simulated intensities, sinograms (Figure 5), and FBP images using the relative discrepancy, normalized by the measured quantity.

Figure 5. Round-trip comparison of the full-data experimental sinograms. Rows: attenuation , differential phase , and dark-field ; columns: retrieved from the measured intensities, retrieved from the simulated intensities, and simulated minus measured. Within each row the measured and simulated panels share a color scale; the difference panel has its own, centered on zero.
The intensity discrepancy is 0.022. Independent noise of equal size in the measured and simulated sinograms would give a nominal noise-only discrepancy of , where is , , or (Section III-B2) relative to the RMS of the measured sinogram. For , discrepancies against are approximately 10%, 15%, and 11% above this reference. Residual structure is concentrated near strong gradients, consistent with differences in projector and detector-derivative discretization.
The phase-image discrepancy is 0.25, with SSIM 0.68 and an approximately uniform offset of 15% inside the rod. Attenuation and dark-field FBP comparisons are noise-dominated, with discrepancies of 1.31 and 1.40, close to . The round trip checks consistency among measured data, retrieval, geometry, and channel models; it does not establish the FBP images as ground truth.
Prior Parameter Selection
Without ground truth, channel weights are selected by held-out prediction: every fifth angle is withheld (73 of 361), reconstructions use the remaining 288, and candidates are scored by the held-out error, the relative error of the predicted held-out sinograms. This angular-stride split follows [51]. The score does not involve the noise levels, but the reconstructions, and hence their ranking, depend on them through the likelihood; they are held fixed throughout the sweep.
We select and on sparse-angle data using SSIM against full-data FBP as an imperfect reference. Only the phase channel is sensitive to , with SSIM peaking at . We use , which adds curvature in weakly constrained directions; relative to , it moves only the phase MAP, 4% further from FBP in relative with unchanged SSIM.
A logarithmic grid of 29 channel weights from to gives and . Phase held-out error is minimal at and within 0.42% of that over ; we take the lower end, . This reduces the local curvature scale of the smoothed TV term from 593 to 333) at negligible cost in held-out error; in otherwise identical sparse-angle runs of the channel-wise posterior, it raises the median per-pixel phase ESS from 38.9 to 99.9. This choice also reflects the synthetic study’s sensitivity to excessive phase regularization with undersampled data (Section III-A2).
Held-out error varies by less than 0.1% over . We therefore compare sparse-angle JTV reconstructions with the full-data channel-wise TV reconstruction: mean SSIM is nearly constant for and decreases for larger values, so we take . This full-data reference is a surrogate, and each parameter group is selected with the others fixed, which need not be jointly optimal.
MAP Reconstruction Results
Figure 6 shows severe sparse-angle FBP streaks, especially in phase, and noise-dominated dark-field images. TV and JTV suppress these artifacts and resolve all three cavities in and . Their attenuation and phase images are similar; JTV’s clearest effect is in the dark-field background outside the rod, where its standard deviation is 21–39% lower than with TV. The dark-field rod boundary is also more continuous. MAP takes about 7 s for 30 views and 40–50 s for 361 views; FBP takes 3–11 ms.

Figure 6. Real-data reconstructions of attenuation (), phase (), and dark-field (): (a) full data, 361 angles and 10 phase steps; (b) sparse angles, 30 angles and 10 steps; (c) sparse steps, 361 angles and 5 steps; (d) combined undersampling, 30 angles and 5 steps. Rows show FBP, TV, and JTV. Within each panel, TV and JTV share a scale per channel; FBP uses its own. Scales differ between panels. All panels are cropped to the bounding box of the reconstruction support, 26.1 mm across: TV and JTV are identically zero outside it, whereas FBP, which uses no support constraint, extends beyond the crop.
| Channel | Method | Full data | Sparse angles | Sparse steps | Combined |
| FBP | 1.553 | 4.064 | 1.529 | 4.020 | |
| TV | 1.248 | 1.242 | 1.225 | 1.243 | |
| JTV | 1.249 | 1.248 | 1.226 | 1.252 | |
| FBP | 1.471 | 4.040 | 1.428 | 3.477 | |
| TV | 1.209 | 1.227 | 1.217 | 1.294 | |
| JTV | 1.211 | 1.241 | 1.220 | 1.311 | |
| FBP | 1.559 | 3.987 | 1.538 | 3.977 | |
| TV | 1.224 | 1.251 | 1.223 | 1.266 | |
| JTV | 1.235 | 1.268 | 1.232 | 1.273 |
Table 3. Per-channel sinogram residual: the RMS misfit between predicted and measured sinograms in units of the channel’s noise standard deviation, estimated for each step count. A value of 1 means a residual at that noise level.
Sinogram residuals, the RMS misfit of each channel in units of its noise standard deviation (Table III), decrease from 1.43–4.06 for FBP to 1.21–1.31 for TV and JTV. Because each step count uses its own noise estimate, the residuals are comparable across conditions. TV gives smaller residuals throughout, with JTV values up to 1.4% higher. These residuals measure data consistency, not reconstruction accuracy; values above unity indicate that the residual exceeds the estimated noise scale, potentially reflecting prior-induced bias, model discrepancy, or limitations of the noise model.
Posterior Sampling and UQ
Using the settings in Section III-B3 and the prior parameters in Section III-B5, the channel-wise and joint posteriors are sampled under all four conditions with the step-specific noise estimates. Adapted step sizes range from to , with mean acceptance probabilities of 0.82–0.89. A chain takes approximately 9.5 h for 361 views and 50 min for 30 views.
Diagnostics cover every sampled pixel. Full-data median ESS values are high (Table IV), with median in each channel. With sparse steps, median ESS falls by 19–41 % across channels and priors; maximum remains below 1.07. The larger noise variance weakens the likelihood relative to the fixed prior and changes posterior geometry, but these diagnostics alone do not establish the cause of slower mixing.
| Scenario | Prior | |||
| Full data | TV | 801.5 | 819.5 | 1043.6 |
| JTV | 867.6 | 889.7 | 1137.6 | |
| Sparse angles | TV | 186.9 | 99.9 | 209.5 |
| JTV | 224.9 | 114.2 | 289.0 | |
| Sparse steps | TV | 547.2 | 486.3 | 778.6 |
| JTV | 657.6 | 580.0 | 927.0 | |
| Combined | TV | 157.9 | 56.3 | 215.7 |
| JTV | 174.6 | 58.7 | 256.3 |
Table IV. Median per-pixel ESS over the reconstruction support for the real-data posteriors, using two chains of 1000 retained draws.
Angular undersampling particularly affects phase: TV median ESS falls to 99.9 for sparse angles and 56.3 for combined undersampling. For the channel-wise phase posterior, median reaches 1.007–1.011 and the 99th percentile of reaches 1.12–1.18, with spatially coherent elevations and minimum ESS values of 2.7 and 3.0, respectively. The 99th percentile of remains below 1.06 in and . Although the medians and the and diagnostics are favorable, these tail values show unresolved slow mixing in parts of the phase posterior. The detector derivative in (2) suppresses low detector frequencies, so the data constrain slowly varying phase components only weakly; this plausibly contributes to the difficult phase conditioning, although the diagnostics do not identify the slowly mixing components directly.
JTV raises median ESS in every limited-data channel–condition pair by 4–38%. Under sparse angles, the phase 99th percentile of decreases from 1.12 to 1.08, its maximum from 1.33 to 1.17, and minimum ESS increases from 2.7 to 5.1. Under combined undersampling, the phase 99th percentile of is essentially unchanged at 1.18, while median ESS rises only from 56.3 to 58.7, without improvement in its minimum. These gains are conditional on the different target posteriors and leave slow phase mixing unresolved.
Figure 7 shows posterior means and pixel-wise standard deviations. Posterior-mean residuals range from 1.12 to 1.24, below the MAP residuals of Table III in every case. Averaged over the reconstruction support, JTV lowers the standard deviation by 7–8% in , 1–3% in , and 23–30% in . For these acquisitions, reducing angular sampling has a larger effect on posterior spread than halving the phase steps. Relative to the full-data channel-wise posterior, halving the steps raises mean posterior standard deviation by 6% in , 20% in , and 12% in . Reducing 361 views to 30 raises it by 18%, 76%, and 38%, respectively, and combined undersampling by 21%, 89%, and 42%. The joint posterior follows the same pattern, with smaller dark-field increases of 26% for sparse angles and 30% for combined undersampling, each relative to its own full-data result. Phase is the most sensitive channel to both reductions, consistent with its derivative measurement. The posterior standard deviations are conditional on the assumed forward model, noise model, and prior, and carry additional Monte Carlo error where mixing is slow; the narrower JTV posteriors alone do not establish better calibration.

Figure 7. Real-data posterior means and pixel-wise standard deviations for TV and JTV under (a) full data, (b) sparse angles, (c) sparse steps, and (d) combined undersampling. Each quantity shares a scale per channel across all scenarios and priors; the bottom colorbars apply to every panel. All panels are cropped to the bounding box of the reconstruction support, 26.1 mm across, outside which the posterior is identically zero.
Discussion
The joint formulation combines MAP estimation with posterior UQ for limited-data multi-contrast CT. Relative to channel-wise TV, JTV improves synthetic MAP accuracy modestly but consistently: RE decreases in 11 of 12 channel–condition pairs, and the summed per-channel RE by 0.8% on average. Its clearest benefit is in the experimental dark-field channel, where it reduces the background standard deviation of the MAP image by 21–39% and the mean posterior standard deviation by 23–30%, while MAP data-fit residuals increase slightly in all channels. These reductions in image noise and posterior spread do not by themselves imply improved reconstruction accuracy or calibrated uncertainty.
The synthetic results expose this distinction. In flat regions, RMS error is 0.37–0.74 times RMS posterior standard deviation in and , and 0.79–1.02 times in , whereas at boundaries the ratio reaches 2.5–3.2 in , 1.2–1.8 in , and 5.3–10.5 in (Table II). These spatial summaries from one noise realization are not repeated-sampling coverage estimates. Reconstructing phase with the simulation projector reduces edge error by about 40%, supporting a contribution from model discrepancy, while residual error can also reflect grid representation and prior-induced bias. The phase ratio decreases as data are removed because posterior spread increases while edge error persists. Thus, the standard deviation maps describe posterior variability under the assumed model and prior but can substantially understate boundary error.
The posterior mean was less accurate than the MAP estimate for and in every synthetic condition, most strongly for dark-field (RE 0.160 versus 0.115 with TV on full data). The weights were tuned for MAP accuracy, so this is an empirical comparison for the evaluated phantom, not a general ordering of Bayesian estimators. Parameter selection also matters: weights tuned to full synthetic data can over-regularize undersampled phase, whereas tuning across acquisition conditions keeps RE within 4% of each per-condition optimum. Ground-truth tuning on the evaluated noise realization gives best-case synthetic performance; the experimental selection combines held-out prediction with surrogate full-data references and is sequential. Hierarchical estimation of the weights and noise levels [30], or empirical Bayesian estimation of the weights [52], could replace this selection, at additional sampling cost.
Operating after signal retrieval makes the framework transferable beyond TLI, for example to EI, when separately retrieved attenuation, differential phase, and dark-field sinograms are available and the forward operators and noise model are adapted. This modularity has a statistical cost: independent additive Gaussian noise in the retrieved channels approximates the nonlinear propagation of intensity noise and neglects the dependence of the noise level on transmission and visibility [18] and possible correlations; where strong scattering saturates the dark-field signal, retrieval itself becomes unreliable [53]. The experimental residuals exceed the estimated noise scale, and the round-trip discrepancies are approximately 10–15 % above the nominal independent-noise reference (Section III-B4). These checks indicate remaining disagreement, without uniquely identifying its
source. Intensity-domain inference could propagate uncertainty through retrieval within a joint model, but would require a more demanding nonlinear likelihood and would still need to address forward-model error.
At fixed tube settings, the sparse-angle, sparse-step, and combined experimental subsets correspond to 12-, 2-, and 24-fold reductions in exposure time and dose (Section III-B2), while the largest increase in channel-wise mean posterior standard deviation is 76, 20, and 89%, respectively, each in phase. These retrospective estimates of potential savings apply to the water-immersed phantom, without dosimetry or validation at equal image quality. The water container increases exposure requirements [13]; performance on heterogeneous specimens scanned in air remains untested, particularly when phase wrapping violates the assumed observation model.
Sampling cost and convergence limit quantitative UQ. A 361-view experimental chain takes about 9.5 h for 153 864 unknowns. JTV raises median ESS by 4–38% in the limited-data runs and lowers the sparse-angle phase 99th percentile of from 1.12 to 1.08. Slow phase mixing persists, however, and under combined undersampling the phase 99th percentile of remains approximately 1.18 for both priors. The smoothed TV term has its largest local curvature in flat regions, where it scales as the channel weight divided by ; reducing from 593 to 333 raised the median per-pixel ESS of the sparse-angle channel-wise phase posterior by a factor of 2.6, while that of and changed by at most 9%. Within each geometry, the four experimental posteriors took the same time per iteration to within 0.3%, consistent with trajectories reaching the tree-depth limit in every run. Five-step retrieval approximately doubles the noise variance, halving the likelihood’s weight relative to the fixed prior, while median ESS falls by 19–41% relative to full data across channels and priors. These observations suggest sensitivity to the balance between likelihood and prior, although posterior geometry and sampler adaptation also affect mixing. Further validation should quantify the Monte Carlo error of the reported uncertainty summaries with longer or additional chains and assess their stability across noise realizations.
Extension to three dimensions will require more scalable inference. Candidate approaches include whitened latent Gaussian fields, such as Whittle–Matérn models represented by truncated Karhunen–Loève expansions, combined with nonlinear maps that represent sharp interfaces [54]; channel coupling could then act on the latent fields or on the reconstructed images. Intensity-domain noise models, repeated-sampling coverage studies, and further biomedical and industrial datasets are complementary priorities.
Conclusion
We studied joint Bayesian reconstruction of attenuation, phase, and dark-field images from retrieved sinograms under limited-data acquisition. A JTV prior provides structural coupling within MAP estimation and posterior sampling. Across synthetic and experimental TLI studies, MAP reconstruction with either prior suppresses undersampling artifacts relative to FBP, while JTV gives modest synthetic accuracy gains and a cleaner experimental dark-field background than channel-wise TV.
For the experimental data, JTV lowers mean dark-field posterior standard deviation by 23–30% and raises median per-pixel ESS by 4–38% under limited data. These gains describe the posterior under the assumed model and sampling efficiency, not calibration: synthetic posterior spread matched or exceeded the error in flat regions but understated it at boundaries, and slow phase mixing persists under angular undersampling. Future work will address uncertainty calibration, intensity-domain likelihoods, and scalable inference for three-dimensional multi-contrast CT.
Acknowledgment
The authors used ChatGPT (OpenAI) and Claude (Anthropic) for language editing and polishing, and Claude Code for assistance with software development and repository organization. The authors reviewed and edited the assisted material and take full responsibility for the final manuscript, code, and results.
References
- [1]M. Bech, T. H. Jensen, R. Feidenhans’l, O. Bunk, C. David, and F. Pfeiffer, “Soft-tissue phase-contrast tomography with an X-ray tube source,” Physics in Medicine & Biology, vol. 54, no. 9, pp. 2747–2753, 2009.DOI
- [2]S. T. Taba, T. E. Gureyev, M. Alakhras, S. Lewis, D. Lockie, and P. C. Brennan, “X-ray phase-contrast technology in breast imaging: principles, options, and clinical application,” American Journal of Roentgenology, vol. 211, no. 1, pp. 133–145, 2018.DOI
- [3]J. Glinz, B. Plank, J. Gutekunst, M. Scheerer, S. Zabler, J. Kastner, and S. Senck, “A comparison of X-ray attenuation, differential phase, and dark-field contrast imaging for the detection of porosity in carbon fiber reinforced cyanate ester,” NDT & E International, vol. 147, p. 103194, 2024.DOI
- [4]F. Pfeiffer, T. Weitkamp, O. Bunk, and C. David, “Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources,” Nature Physics, vol. 2, no. 4, pp. 258–261, 2006.DOI
- [5]M. Bech, O. Bunk, T. Donath, R. Feidenhans’l, C. David, and F. Pfeiffer, “Quantitative X-ray dark-field computed tomography,” Physics in Medicine & Biology, vol. 55, no. 18, pp. 5529–5539, 2010.DOI
- [6]F. Pfeiffer, J. Herzen, M. Willner, M. Chabior, S. Auweter, M. Reiser, and F. Bamberg, “Grating-based X-ray phase contrast for biomedical imaging applications,” Zeitschrift für Medizinische Physik, vol. 23, no. 3, pp. 176–185, 2013.DOI
- [7]A. Olivo, “Edge-illumination X-ray phase-contrast imaging,” Journal of Physics: Condensed Matter, vol. 33, no. 36, p. 363002, 2021.DOI
- [8]N. Six, J. Renders, J. De Beenhouwer, and J. Sijbers, “Joint multi-contrast CT for edge illumination X-ray phase contrast imaging using split Barzilai-Borwein steps,” Optics Express, vol. 32, no. 2, pp. 1135–1150, 2024.DOI
- [9]B. Brendel, M. von Teuffenbach, P. B. Noël, F. Pfeiffer, and T. Koehler, “Penalized maximum likelihood reconstruction for X-ray differential phase-contrast tomography,” Medical Physics, vol. 43, no. 1, pp. 188–194, 2016.DOI
- [10]A. Ritter, G. Anton, and T. Weber, “Simultaneous maximum-likelihood reconstruction of absorption coefficient, refractive index and dark-field scattering coefficient in X-ray Talbot-Lau tomography,” PLOS ONE, vol. 11, no. 10, p. e0163016, 2016.DOI
- [11]J. Xu, Z. Wang, S. van Gogh, M. Rawlik, S. Spindler, and M. Stampanoni, “Intensity-based iterative reconstruction for helical grating interferometry breast CT with static grating configuration,” Optics Express, vol. 30, no. 8, pp. 13847–13863, 2022.DOI
- [12]F. Pfeiffer, C. Kottler, O. Bunk, and C. David, “Hard X-ray phase tomography with low-brilliance sources,” Physical Review Letters, vol. 98, no. 10, p. 108105, 2007.pubmed.ncbi.nlm.nih.gov/17358572
- [13]H. Määttänen, H. Suhonen, T. Siiskonen, C. David, and S. Huotari, “Optimization of contrast and dose in X-ray phase-contrast tomography with a Talbot-Lau interferometer,” Biomedical Physics & Engineering Express, vol. 10, no. 4, p. 045045, 2024.
- [14]Q. Xu, E. Y. Sidky, X. Pan, M. Stampanoni, P. Modregger, and M. A. Anastasio, “Investigation of discrete imaging models and iterative image reconstruction in differential X-ray phase-contrast tomography,” Optics Express, vol. 20, no. 10, pp. 10724–10749, 2012.DOI
- [15]M. Viemetz, N. Guschin, C. Schmid, J. Haeusele, M. von Teuffenbach, P. Meyer, F. Bergner, T. Lasser, R. Proksa, T. Koehler, and F. Pfeiffer, “Dark-field computed tomography reaches the human scale,” Proceedings of the National Academy of Sciences, vol. 119, no. 8, p. e2118799119, 2022.
- [16]J. Bian, J. H. Siewerdsen, X. Han, E. Y. Sidky, J. L. Prince, C. A. Pelizzari, and X. Pan, “Evaluation of sparse-view reconstruction from flat-panel-detector cone-beam CT,” Physics in Medicine & Biology, vol. 55, no. 22, pp. 6575–6599, 2010.
- [17]P. C. Hansen, J. Jørgensen, and W. R. Lionheart, Computed Tomography: Algorithms, Insight, and Just Enough Theory. SIAM, 2021.
- [18]V. Revol, C. Kottler, R. Kaufmann, U. Straumann, and C. Urban, “Noise analysis of grating-based X-ray differential phase contrast imaging,” Review of Scientific Instruments, vol. 81, no. 7, p. 073709, 2010.DOI
- [19]T. Weber, P. Bartl, F. Bayer, J. Durst, W. Haas, T. Michel, A. Ritter, and G. Anton, “Noise in X-ray grating-based phase-contrast imaging,” Medical Physics, vol. 38, no. 7, pp. 4133–4140, 2011.pubmed.ncbi.nlm.nih.gov/21859014
- [20]Y. Chen, H. Guan, C. K. Hagen, A. Olivo, and M. A. Anastasio, “Single-shot edge illumination X-ray phase-contrast tomography enabled by joint image reconstruction,” Optics Letters, vol. 42, no. 3, pp. 619–622, 2017.DOI
- [21]Y. Chen and M. A. Anastasio, “Properties of a joint reconstruction method for edge-illumination X-ray phase-contrast tomography,” Sensing and Imaging, vol. 19, no. 1, p. 7, 2018.DOI
- [22]N. Pato Montemayor, S. van Gogh, M. Rawlik, Z. Varga, and M. Stampanoni, “Intensity-based iterative reconstruction with cross-channel regularization for grating interferometry breast CT,” Optics Continuum, vol. 3, no. 6, pp. 995–1011, 2024.
- [23]S. van Gogh, M. Rawlik, A. Pereira, S. Spindler, S. Mukherjee, M.-C. Zdora, M. Stauber, R. Alaifari, Z. Varga, and M. Stampanoni, “Towards clinical-dose grating interferometry breast CT with fused intensity-based iterative reconstruction,” Optics Express, vol. 31, no. 5, pp. 9052–9071, 2023.DOI
- [24]M. J. Ehrhardt, K. Thielemans, L. Pizarro, D. Atkinson, S. Ourselin, B. F. Hutton, and S. R. Arridge, “Joint reconstruction of PET-MRI by exploiting structural similarity,” Inverse Problems, vol. 31, no. 1, p. 015001, 2015.DOI
- [25]D. S. Rigie and P. J. La Riviere, “Joint reconstruction of multi-channel, spectral CT data via constrained total nuclear variation minimization,” Physics in Medicine & Biology, vol. 60, no. 5, pp. 1741–1762, 2015.
- [26]J. Toivanen, A. Meaney, S. Siltanen, and V. Kolehmainen, “Joint reconstruction in low-dose multi-energy CT,” Inverse Problems and Imaging, vol. 14, no. 4, pp. 607–629, 2020.arxiv.org/abs/1904.05671
- [27]M. Nilchian, E. Bostan, Z. Wang, M. R. Nilchiyan, M. Stampanoni, and M. Unser, “Joint absorption and phase retrieval in grating-based X-ray radiography,” Opt. Express, vol. 24, pp. 7253–7265, Apr 2016.DOI
- [28]J. Liu, J. Hagemann, and M. Burger, “Joint near-field holotomography reconstruction with a phase-guided Bregman TV regularization.” arXiv:2608.17621, 2026.arxiv.org/abs/2608.17621
- [29]J. P. Kaipio and E. Somersalo, Statistical and Computational Inverse Problems. Springer, 2005.DOI
- [30]J. M. Bardsley, “MCMC-based image reconstruction with uncertainty quantification,” SIAM Journal on Scientific Computing, vol. 34, no. 3, pp. A1316–A1332, 2012.DOI
- [31]J. Suuronen, M. Emzir, S. Lasanen, S. Särkkä, and L. Roininen, “Enhancing industrial X-ray tomography by data-centric statistical methods,” Data-Centric Engineering, vol. 1, p. e10, 2020.arxiv.org/abs/2003.03814
- [32]A. Horst, B. Maboudi Afkham, Y. Dong, and J. Lemvig, “Uncertainty quantification for linear inverse problems with Besov prior: A randomize-then-optimize method,” Statistics and Computing, vol. 35, no. 4, p. 101, 2025.DOI
- [33]B. M. Afkham, Y. Dong, and P. C. Hansen, “Uncertainty quantification of inclusion boundaries in the context of X-ray tomography,” SIAM/SASA Journal on Uncertainty Quantification, vol. 11, no. 1, pp. 31–61, 2023.arxiv.org/abs/2107.06607
- [34]A. M. Ramboujn, H. Komber, J. Rossdale, J. Suntharalingam, J. C. Rodrigues, M. J. Ehrhardt, and A. Repetti, “Uncertainty quantification in computed tomography pulmonary angiography,” PNAS Nexus, vol. 3, no. 1, p. pgad404, 2024.
- [35]M. D. Hoffman and A. Gelman, “The No-U-Turn sampler: Adaptively setting path lengths in Hamiltonian Monte Carlo,” Journal of Machine Learning Research, vol. 15, no. 47, pp. 1593–1623, 2014.DOI
- [36]M. von Teuffenbach, T. Koehler, A. Fehringer, M. Viemetz, B. Brendel, J. Herzen, R. Proksa, E. J. Rummeny, F. Pfeiffer, and P. B. Noël, “Grating-based phase-contrast and dark-field computed tomography: a single-shot method,” Scientific Reports, vol. 7, no. 1, p. 7476, 2017.
- [37]Z. Wang, K. Kang, Z. Huang, and Z. Chen, “Quantitative grating-based X-ray dark-field computed tomography,” Applied Physics Letters, vol. 95, no. 9, p. 094105, 2009.DOI
- [38]V. Kolehmainen, E. Somersalo, P. Vauhkonen, M. Vauhkonen, and J. Kaipio, “A Bayesian approach and total variation priors in 3D electrical impedance tomography,” in Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Vol. 20 Biomedical Engineering Towards the Year 2000 and Beyond (Cat. No.98CH36286), vol. 2, pp. 1028–1031, IEEE, 1998.
- [39]M. Lassas and S. Siltanen, “Can one use total variation prior for edge-preserving Bayesian inversion?,” Inverse Problems, vol. 20, no. 5, pp. 1537–1563, 2004.DOI
- [40]D. Lin, A. Garpebring, and T. Löfstedt, “A proper structured prior for Bayesian T1 mapping,” in International Workshop on Uncertainty for Safe Utilization of Machine Learning in Medical Imaging, pp. 224–233, Springer, 2025.DOI
- [41]A. B. Owen, Monte Carlo Theory, Methods and Examples. https://artowen.su.domains/mc/, 2013. Chapter 11.
- [42]S. Van der Walt, J. L. Schönberger, J. Nunez-Iglesias, F. Boulogne, J. D. Warner, N. Yager, E. Gouillart, and T. Yu, “Scikit-image: image processing in python,” PeerJ, vol. 2, p. e453, 2014.
- [43]A. Paszke, S. Gross, F. Massa, A. Lerer, J. Bradbury, G. Chanan, T. Killeen, Z. Lin, N. Gimelsh ein, L. Antiga, A. Desmaison, A. Köp f, E. Yang, Z. DeVito, M. Raison, A. Tejani, S. Chilamkurthy, B. Steiner, L. Fang, J. Bai, and S. Chintala, “PyTorch: An imperative style, high-performance deep learning library,” in Advances in Neural Information Processing Systems, vol. 32, pp. 8024–8035, 2019.arxiv.org/abs/1912.01703
- [44]W. Van Aarle, W. J. Palenstijn, J. Cant, E. Janssens, F. Bleichrodt, A. Dabravolski, J. De Beenhouwer, K. Joost Batenburg, and J. Sijbers, “Fast and flexible X-ray tomography using the ASTRA toolbox,” Optics express, vol. 24, no. 22, pp. 25129–25147, 2016.DOI
- [45]E. Bingham, J. P. Chen, M. Jankowiak, F. Obermeyer, N. Pradhan, T. Karaletsos, R. Singh, P. Szerlip, P. Horsfall, and N. D. Goodman, “Pyro: Deep universal probabilistic programming,” Journal of machine learning research, vol. 20, no. 28, pp. 1–6, 2019.arxiv.org/abs/1810.09538
- [46]M. Chabior, T. Donath, C. David, M. Schuster, C. Schroer, and F. Pfeiffer, “Signal-to-noise ratio in X ray dark-field imaging using a grating interferometer,” Journal of Applied Physics, vol. 110, no. 5, p. 053105, 2011.DOI
- [47]J. Nocedal and S. J. Wright, Numerical Optimization. Springer Series in Operations Research and Financial Engineering, New York, NY: Springer, 2nd ed., 2006.
- [48]Z. Wang, A. C. Bovik, H. R. Sheikh, and E. P. Simoncelli, “Image quality assessment: from error visibility to structural similarity,” IEEE Transactions on Image Processing, vol. 13, no. 4, pp. 600–612, 2004.pubmed.ncbi.nlm.nih.gov/15376593
- [49]H. Mäkinen and H. Yazdanian, “Data for paper: Bayesian Joint Reconstruction and Uncertainty Quantification for Limited-Data Multi-Contrast X-ray CT.” University of Helsinki, 2026. [Online]. Available: https://doi.org/10.23729/fd-769dae34-045-386f-a856-92e1adc60f18.
- [50]C. David, J. Bruder, T. Rohbeck, C. Grünzweig, C. Kottler, A. Diaz, O. Bunk, and F. Pfeiffer, “Fabrication of diffraction gratings for hard X-ray phase contrast imaging,” Microelectronic Engineering, vol. 84, no. 5–8, pp. 1172–1177, 2007.DOI
- [51]A. A. Hendriksen, D. M. Pelt, and K. J. Batenburg, “Noise2Inverse: Self-supervised deep convolutional denoising for tomography,” IEEE Transactions on Computational Imaging, vol. 6, pp. 1320–1335, 2020.DOI
- [52]A. F. Vidal, V. De Bortoli, M. Pereyra, and A. Durrmus, “Maximum likelihood estimation of regularization parameters in high-dimensional inverse problems: An empirical Bayesian approach. Part I: Methodology and experiments,” SIAM Journal on Imaging Sciences, vol. 13, no. 4, pp. 1945–1989, 2020.DOI
- [53]H. Mäkinen, H. Suhonen, and S. Huotari, “Adaptive filtering method for reducing signal saturation induced artifacts in X-ray dark-field tomography,” Scientific Reports, vol. 16, p. 26283, 2026.DOI
- [54]H. Yazdanian, G. Hillers, Y. Lu, and B. M. Afkham, “A Bayesian framework with geology-informed structural priors for tomographic reconstruction and uncertainty quantification,” Geophysical Journal International, vol. 247, no. 2, p. ggag374, 2026.DOI