Helping Hand: A Continuum Robotic Grasper for Traction in Single-Channel Endoscopic Submucosal Dissection

Abstract

Endoscopic submucosal dissection (ESD) is a minimally invasive approach for removing large polyps from the colon and offers many benefits compared to other surgical approaches. However, ESD is technically very challenging to perform, leading to limited adoption by physicians. We propose the Helping Hand, a two-segment continuum robotic grasper that, together with an electrosurgical knife, enables tissue traction and cutting within a single 3.7 mm working channel. We achieve this functionality by creating a slot at the base of the Helping Hand that lets the electrosurgical knife extend out, enabling countertraction and cutting through one instrument lumen. We evaluate the Helping Hand's workspace and its ability to provide traction forces from different starting configurations. Finally, we demonstrate the full workflow by isolating and removing tissue in an ex vivo chicken thigh, illustrating the potential for the Helping Hand to help improve ESD procedures.

Introduction

Approximately 3.9% of people in the United States will develop colorectal cancer during their lifetime [1]. For patients with precancerous colorectal lesions or selected superficial cancers, endoscopic resection offers a means of removing the lesion while preserving the bowel [2]. Endoscopic mucosal resection (EMR) can capture and remove small lesions with a snare, but large lesions often must be removed in several pieces, which makes it difficult to confirm complete excision and increases the risk of recurrence [2]. In contrast, in endoscopic submucosal dissection (ESD), the physician uses an electrosurgical knife to incise the mucosa around the lesion and dissect the submucosal tissue beneath it, permitting en bloc removal that preserves the resection margins for pathological assessment [2]. ESD thus allows large lesions to be removed without colectomy, which for nonmalignant polyps carries a 14% rate of major adverse events and 0.7% 30-day mortality [3, 4].

Despite the many benefits of ESD, the procedure is technically very challenging to perform. In particular, maintaining a clear view of the tissue immediately ahead of the knife is challenging as the lesion separates from the bowel wall and the partially freed mucosal flap often obscures the plane between

the lesion and the underlying muscle. Applying traction to the flap exposes this plane and places the submucosal fibers under tension, facilitating dissection [5, 6]. The direction of traction is also critical, where a flap that initially exposes the submucosa may become less effective as the cut advances [6]. The procedural challenge is therefore to maintain traction in a direction that keeps the dissection plane both visible to the physician and accessible to the knife.

Several devices have been proposed to maintain traction on the mucosal flap during ESD. One approach has been to mount the grasper externally, leaving the working channel available for the knife. The Olympus FlexiLifter, for example, advances and retracts forceps from a fixed position on a distal cap. This approach results in a dependence between the angle at which the forceps approach the flap and the orientation of the endoscope [7]. Lou et al. have developed an externally mounted soft pop-up robot that articulates a grasper at the tip of the endoscope [8]. However, even when folded, the robot measures approximately 33×32×633 \times32 \times6 mm, adding bulk at the tip that may make it harder to navigate narrow or tortuous anatomy. Other robotic end caps mount to the endoscope tip and articulate instruments passed through the working channel, adding distal dexterity without a second channel [9].

A second approach has been to use separate working channels for traction and cutting. The Fujifilm TRACMOTION uses an articulating, rotatable grasper to adjust the direction of traction while an electrosurgical knife passes through the second working channel [10]. Connor et al. passed two concentric push-pull steerable sheaths through the 3.7 mm and 2.8 mm channels of a standard dual-channel colonoscope, one steering forceps for countertraction and the other an electrosurgical tool, and showed that users with no prior ESD

This article has supplementary downloadable material (Supplementary Video S1) available, provided by the authors. experience could complete ex vivo resections [11]. Gao et al. developed a bimanual telerobotic system that passes a 3.5 mm manipulator for tissue traction and a 2.5 mm manipulator for dissection through the two working channels of a commercial endoscope [12]. However, the externally mounted approaches add bulk at the tip of the endoscope, and the multichannel approaches require a dual-channel endoscope, whose acquisition cost limits its availability in many endoscopy units [13]. These constraints motivate a device that provides articulated traction and knife access through a single working channel.

The contributions of this work are as follows: (1) We present the Helping Hand, a two-degree-of-freedom continuum robotic grasper that houses all the tools needed for endoscopic submucosal dissection (ESD) within a single 3.7 mm colonoscope working channel (see Figure 1). A slot at its base, rather than the usual tip exit, lets the electrosurgical knife extend through the grasper, enabling countertraction and cutting through one instrument lumen without any externally mounted tools or a second working channel. (2) We evaluate the Helping Hand’s workspace and ability to provide retraction forces. We also demonstrate the full workflow by isolating and removing tissue in an ex vivo chicken thigh benchtop model. To the best of our knowledge, this is the first demonstration of robotic ESD with countertraction and cutting performed through a single working channel.

The Helping Hand, a two-degree-of-freedom continuum robotic grasper with an electrosurgical knife

Figure 1. The Helping Hand is a two-degree-of-freedom continuum robotic grasper. It consists of two bending segments, along with a slot at the base that enables a custom electrosurgical knife to be exposed for cutting as the grasper at the tip of the bending segment. The grasper is simultaneously used to provide tissue traction (See Supplementary Video S1).

System Design and Manufacturing

The architecture of the robot, as shown in Figure 2, consists of two main subsystems. The instrument subsystem consists of the Helping Hand—a two-segment, notched continuum robot with an integrated grasper—and an electrosurgical knife that passes through the Helping Hand and exits through its knife slot. The actuation subsystem drives the Helping Hand and the insertion of the knife through four independent linear motion units.

Figure 2. Overview of the Helping Hand and actuation architecture. (a) The Instrument subsystem consists of the Helping Hand and the electrosurgical knife. (b) The actuation subsystem comprises four lead-screw-driven linear motion units for proximal and distal tendon actuation, grasper operation, and knife translation. Load cells at the tendon attachments measure tension during articulation of the Helping Hand. (c) The electrosurgical generator subsystem connects to the knife and return pad through the ESG controller, with activation controlled by a pedal switch.

Instrument Subsystem

Helping Hand

The Helping Hand is a two-segment, notched continuum robot with a slot in the proximal segment

for exposing the knife (see Figure 3). It is fabricated from a hollow Nitinol tube [14], [15], and each segment is independently actuated via a tendon. The proximal segment consists of the knife exit notch p1p_1 and the retraction section p2p_2, with the knife slot on the opposite side of the tube. The single notch p1p_1 sits proximal to the knife slot, so once p1p_1 bends, inserting the knife advances it out through the slot. The four notches in p2p_2 lift the grasped tissue to expose the submucosal plane. The notches in the distal segment dd are oriented 180° from those in the proximal segment, so actuating both segments produces an S bend that directs the grasper toward the polyp. Combined with

Exploded view of the Helping Hand, including detailed views of the medial and proximal routers, and the grasper collet

Figure 3. Exploded view of the Helping Hand, including detailed views of the medial and proximal routers, and the grasper collet.

SymbolDef.Proximal (p1p_1, p2p_2)Distal (dd)
ggNotch depth2.8 mm3.2 mm
hhNotch height1.0 mm1.0 mm
ccUncut height1.0 mm1.0 mm
nnNumber of notches1, 48
rir_iInner radius of tube1.65 mm1.65 mm
ror_oOuter radius of tube1.75 mm1.75 mm

Table 1. Helping Hand parameters.

The uncut lengths at the base and tip are ℓbase=ℓtip=5\ell_{\mathrm{base}}=\ell_{\mathrm{tip}}=5 mm. The gaps between notched sections are ℓp=7\ell_p=7 mm and ℓpd=10\ell_{pd}=10 mm. insertion, this allows the tip pose to be controlled within the bending plane [16]. Table I summarizes the tube dimensions and the geometry of all notched sections.

Building on [14], we developed a cost-effective and repeatable two-operation CNC milling process to fabricate the Helping Hand using a reusable aluminum fixture (Figure 4). A groove supported the thin-walled Nitinol tube along its length, and we used Loctite Super Glue to hold it in place. Relief slots aligned with the notch pattern allowed the end mill to cut through the tube wall without contacting the fixture. In Operation 1, we machined the proximal notches and a semicircular registration feature at the end of the tube. We then rotated the tube 180° and seated the registration feature against a dowel pin to align the distal notch pattern for Operation 2. Using 1 mm miniature square end mills (Harvey Tool, 834139-C3 and 836939-C3), we machined the proximal and distal notch patterns, the knife slot, and an access slot used during fabrication. We ran the end mills at 9,702 rpm with a feed rate of 49.3 mm/min, a chip load of 0.00127 mm/tooth, and a maximum stepdown of 0.5 mm. Finally, we released the tube in acetone and deburred the machined edges.

CNC milling setup for the Helping Hand notched tube

Figure 4. CNC milling setup for the Helping Hand notched tube. (a) The tube is secured in the custom fixture beneath a 1 mm end mill. (b) The CAM setup shows the programmed toolpaths in orange for Operation 1 (left) and Operation 2 (right), between which the tube is rotated 180°. Two operations on one reusable fixture produce both notch patterns with consistent alignment.

To ensure smooth tendon actuation and reliable operation of the grasper and knife, we designed a set of routing disks, SLA 3D printed in Grey V5 resin (Formlabs Inc.) (Figure 3). The proximal router establishes the initial spacing between the Nitinol tendons (0.1524 mm diameter), grasper, and knife. The medial router, located in the uncut length ℓpd\ell_{pd} between the proximal and distal segments, serves three functions. It guides the grasper shaft, anchors the proximal tendon, and redirects the distal tendon. Redirecting the distal tendon keeps it clear of the knife path and away from the inner wall of the proximal bending segment, limiting coupling between distal tendon actuation to proximal bending. The grasper (Vetovation, 1 mm diameter biopsy forceps) passes through the routers and is secured to the distal end of the tube by a custom collet (Tough 1000 resin, Formlabs Inc.) and nut. We bonded the collet to the tube and anchored both tendons with UV curing adhesive (AA 3979, Loctite).

Electrosurgical knife

  1. Electrosurgical knife: The prototype knife visible in Figure 1 fits through the 2.0 mm wide slit at the base of the Helping Hand. It consists of a 1.0 mm diameter stainless steel rod insulated with a silicone conformal coating (1Dfaul), which is applied evenly using a rotary fixture. A proximal brass shoulder is used to connect the knife to the electrosurgical generator (Force 2, Covidien Valleylab) and mounts it to its dedicated linear motion unit. A patient return electrode (PolyHesive, Covidien Valleylab) completes the monopolar circuit through the tissue.

Actuation Subsystem

The actuation subsystem builds on the modular system by Xu et al. [17]. The system consolidates all four linear motion units along a common pair of linear rails (Figure 2). Each linear motion unit is driven by a lead screw and independently controls one degree of freedom of the robot: proximal tendon, distal tendon, knife translation, and grasper activation. The linear motion units that actuate the tendons are fitted with inline load cells between the tendon attachment and the carriage frame, enabling real-time tension measurement during articulation of the Helping Hand. We extended the open source actuation platform of [17] with mechanical components for inline tension sensing and a micro-ROS [18] interface that connects an ESP32 microcontroller to ROS 2, and we provide both additions as open source (GitHub link to be provided upon publication). The most distal linear motion unit controls knife advancement, and the most proximal unit actuates the grasper. We fixed the grasper trigger to the frame of the actuation unit and attached the handle to the linear motion unit, so that carriage translation opens and closes the jaws. The tendons and instruments converge at the proximal to the base of the continuum segments

Modeling

This section presents the forward kinematics of the Helping Hand, which map the proximal and distal tendon displacements to the pose of the tip frame {t}\{t\} relative to the base frame {s}\{s\} (Figure 5). Following [14], we assumed that each notch bends with constant curvature in a steering plane and that the uncut portions of the tube remain rigid. We then combined the individual notch transformations serially across multiple sections, similar to [16].

Side view of the Helping Hand notched tube

Figure 5. Side view of the Helping Hand notched tube, with the base frame {s}\{s\} at the proximal end and the tip frame {t}\{t\} at the distal end. One tendon actuates the proximal segment, which consists of the knife exit notch p1p_1 and retraction section p2p_2, and the second tendon actuates the distal section dd. The knife slot sits between p1p_1 and p2p_2, so two tendons let a single tube expose the knife, aim the grasper, and retract tissue.

Notch Kinematics

The Helping Hand consists of three notched sections (Figure 5). The proximal segment comprises a single knife exit notch, p1p_1, and a section of four notches p2p_2, which all share the same notch geometry and steering tendon. The distal continuum segment, dd, has eight notches actuated by an independent tendon. Throughout this section, the subscript γ∈{p1,p2,d}\gamma\in\{p_1,p_2,d\} denotes the section, and nγn_\gamma is its number of notches (np1=1n_{p_1}=1, np2=4n_{p_2}=4, and nd=8n_d=8). We assume that each tendon displacement is distributed equally among the notches it actuates. For the proximal and distal tendon displacements qpq_p and qdq_d, the displacement per notch is

Δlp1=Δlp2=qpnp1+np2andΔld=qdnd.(1)\Delta l_{p_1}=\Delta l_{p_2}=\frac{q_p}{n_{p_1}+n_{p_2}} \quad\text{and} \quad\Delta l_d=\frac{q_d}{n_d}. \tag*{(1)}

For each section, the distance yˉγ\bar{y}_\gamma from the tube center to the neutral bending plane was calculated from the notch depth and tube radii in Table I using the cross-sectional centroid formulation from [14]. The curvature κγ\kappa_\gamma and arc length sγs_\gamma of a single notch are

κγ≈Δlγhγ(ri+yˉγ)−yˉγΔlγandsγ=hγ1+yˉγκγ(2)\kappa_\gamma\approx\frac{\Delta l_\gamma}{h_\gamma(r_i+\bar{y}_\gamma)-\bar{y}_\gamma\Delta l_\gamma} \quad\text{and} \quad s_\gamma=\frac{h_\gamma}{1+\bar{y}_\gamma\kappa_\gamma} \tag*{(2)}

where hγh_\gamma is the notch height and rir_i is the inner radius of the tube [14]. With bending angle θγ=κγsγ\theta_\gamma=\kappa_\gamma s_\gamma, the transformation across notch jj of section γ\gamma is

Tjcj+1=[10000cos⁡θγ−sin⁡θγ(cos⁡θγ−1)/κγ0sin⁡θγcos⁡θγsin⁡θγ/κγ0001](3)T_{jc}^{j+1}= \begin{bmatrix} 1 & 0 & 0 & 0\\ 0 & \cos\theta_\gamma& -\sin\theta_\gamma& (\cos\theta_\gamma-1)/\kappa_\gamma\\ 0 & \sin\theta_\gamma& \cos\theta_\gamma& \sin\theta_\gamma/\kappa_\gamma\\ 0 & 0 & 0 & 1 \end{bmatrix} \tag*{(3)}

which reduces to a straight translation of hγh_\gamma along the local zz-axis as κγ\kappa_\gamma approaches zero.

Helping Hand Forward Kinematics

Let Tz,ℓT_{z,\ell} denote a translation by ℓ\ell along the local zz-axis and Rz,ϕR_{z,\phi} a rotation by ϕ\phi about that axis. The transformation from the base of the first notch to the end of the last notch in section γ\gamma is

Tγ=(∏j=0nγ−1Tjcj+1Tz,cγ)Tz,−cγ(4)T_\gamma=\left(\prod_{j=0}^{n_\gamma-1}T_{jc}^{j+1}T_{z,c_\gamma}\right)T_{z,-c_\gamma} \tag*{(4)}

where cγc_\gamma is the uncut height between notches. The final translation Tz,−cγT_{z,-c_\gamma} ensures there are nγ−1n_\gamma-1 uncut heights in each section [14].

The base frame {s}\{s\} is located at the proximal end of the Helping Hand, with its zz-axis aligned with the tube and directed toward the tip (Figure 5). The tip frame {t}\{t\} is located at the distal end of the uncut tip and is aligned with section dd. Sections p1p_1 and p2p_2 share the same notch orientation, while the distal notches are oriented 180∘180^\circ from the proximal notches so that section dd bends in the opposite direction. The transformation from the base frame {s}\{s\} to the tip frame {t}\{t\} is therefore

Tst=Tz,ℓbaseTp1Tz,ℓpTp2Tz,ℓpdRz,180∘TdTz,ℓtip(5)T_{st}=T_{z,\ell_{\mathrm{base}}}T_{p_1}T_{z,\ell_p}T_{p_2}T_{z,\ell_{pd}}R_{z,180^\circ}T_dT_{z,\ell_{\mathrm{tip}}} \tag*{(5)}

where ℓbase\ell_{\mathrm{base}} and ℓtip\ell_{\mathrm{tip}} are the uncut lengths at the base and tip, ℓp\ell_p is the uncut length between sections p1p_1 and p2p_2 that contains the knife exit slot, and ℓpd\ell_{pd} is the uncut length between the proximal and distal segments that contains the medial router.

Experiments and Results

We conducted three experiments to evaluate our system. The first two experiments evaluated the Helping Hand alone, and the third experiment evaluated the Helping Hand together with the electrosurgical knife. First, we conducted a workspace characterization (Section IV-A) to track the Helping Hand tip during actuation of the proximal segment and distal segment and compared the results with the kinematic model. Second, we conducted silicone phantom traction trials (Section IV-B) to measure the retraction force applied at several end effector poses. Third, we performed an en bloc resection (Section IV-C), using the Helping Hand and knife together to demonstrate the complete traction and cutting workflow in ex vivo chicken tissue. All experiments used the actuation platform described in Section II-B, mounted on a lead-screw linear stage that provided the insertion axis. Before each trial, both steering tendons were slackened, all sensors were zeroed, and all actuation units were returned to their home positions.

Workspace Characterization

To validate the kinematic model (Section III) and characterize the reachable workspace, we tracked the tip of the Helping Hand with a field generator and an electromagnetic (EM) sensor (Northern Digital Inc.) mounted to the distal end. While actuating the distal tendon, qdq_d, the proximal tendon, qpq_p, was held in slack. The respective segment was actuated in 0.20.2 mm increments (Figure 6(a)) until the notches closed fully and contacted each other. This process was repeated for the proximal tendon, qpq_p (Figure 6(b)). Following [19, 20], we quantified the error with the normalized root mean square error (NRMSE)

Measured Helping Hand tip positions and model predictions during distal-only and proximal-only actuation

Figure 6. Measured Helping Hand tip positions (colored) and model predictions (gray) during (a) distal-only actuation and (b) proximal-only actuation, with the other tendon slack. Each tendon was displaced in 0.2 mm increments. The model captures the shape of each trajectory but overestimates tip travel, consistent with tendon elongation and friction.

eNRMSE=1L1N∑i=1N∥pi−p^i∥2(6)e_{\mathrm{NRMSE}}=\frac{1}{L}\sqrt{\frac{1}{N}\sum_{i=1}^{N}\lVert\mathbf{p}_i-\hat{\mathbf{p}}_i\rVert^2} \tag*{(6)}

where pi\mathbf{p}_i and p^i\hat{\mathbf{p}}_i are the measured and predicted tip positions at step ii, NN is the number of steps, and LL is the length of the actuated segment. The NRMSE was 33.9%33.9\% for the proximal segment and 9.10%9.10\% for the distal segment.

As in prior notched tube designs [14, 15, 16, 19], two modeling assumptions likely accounted for much of this error. First, the model assumes that the full actuator displacement reaches the notched tube. In practice, the thin Nitinol tendons elongate under load, and the actuation system has some compliance, so the tube undergoes less displacement than the actuators command. Second, the model neglects friction. Because friction between the tendon and the notched tube reduces tension along the tube length, the notches bend unevenly rather than with piecewise constant curvature as the model assumes.

Silicone Phantom Traction

During ESD, the position of the polyp relative to the endoscope changes continuously, so the available traction may depend on the Helping Hand configuration at the time of grasping. To investigate this relationship, we mounted a silicone phantom of an 18 mm polyp on a six axis force torque sensor (Mini40, ATI Industrial Automation) (Figure 7). We recorded the tendon tensions and sensor forces and synchronized them through ROS. From the sensor, we report two quantities. The resultant force Fres=Fx2+Fy2+Fz2F_{\mathrm{res}} = \sqrt{F_x^2 + F_y^2 + F_z^2} is the total magnitude of the force that the grasper applied to the phantom. The normal force FzF_z is the component along the sensor zz-axis, perpendicular to the phantom base, and represents the portion of the force that lifted the polyp away from the underlying surface.

Traction test setup

Figure 7. Traction test setup. A silicone phantom of an 18 mm sessile serrated polyp sits on a six axis force and torque sensor below the Helping Hand. We varied the phantom height to change the Helping Hand configuration at grasping. This setup measured the retraction force on the polyp across different Helping Hand configurations.

Figure 8 shows a representative trial. We first articulated the Helping Hand toward the polyp and closed the grasper on a marked point on the phantom. We then retracted the proximal tendon until its measured tension reached 9 N. As the proximal tension increased, the force on the phantom rose. Finally, we returned both tendons to their initial displacements and opened the grasper, and the force on the phantom returned to zero. To determine single force value for each trial, we identified the samples in which the proximal tendon tension was within

Representative phantom traction trial

Figure 8. Representative phantom traction trial. The plots show the resultant and normal force on the phantom and the proximal and distal tendon tensions over time, with dashed lines marking articulation, grasping, traction, and release. The force on the phantom rose as we loaded the proximal tendon and fell when we released it.

Height (mm)FresF_{\mathrm{res}} (N)FzF_z (N)
00.331 ± 0.0150.279 ± 0.029
−40.345 ± 0.0370.299 ± 0.067
−80.316 ± 0.0270.275 ± 0.030
−120.325 ± 0.0160.243 ± 0.037
All0.329 ± 0.0240.274 ± 0.043

Table 2. Force on the phantom at 9 N proximal tendon tension. Values are mean ± SD of three trials per height. Each value is averaged over proximal tension of 9 ± 0.25 N.

9 ± 0.25 N and averaged FresF_{\mathrm{res}} and FzF_z over those samples. This window captured the steady traction phase at the target tension.

We repeated this procedure with the phantom at four

heights, spaced 4 mm apart. The highest position (0 mm) placed the polyp just below the Helping Hand at rest, and the lowest position (−12-12 mm) placed it at the edge of the reachable workspace. At every height, we grasped the same marked point by translating the Helping Hand along the insertion axis. We performed three trials at each height.

Table II summarizes the results. Across all heights, the mean resultant force was 0.33±0.020.33 \pm0.02 N and the mean normal force was 0.27±0.040.27 \pm0.04 N. The Helping Hand applied traction at every tested height, including a substantial component perpendicular to the phantom base, which showed that it could lift tissue throughout its workspace. In this experiment, we controlled only the peak proximal tendon tension in open loop. In future work, a more detailed force model could allow the Helping Hand to actively regulate tendon tension and maintain the desired traction while the physician cuts. To evaluate traction during actual cutting, we next performed an en bloc resection in ex vivo tissue.

En Bloc Tissue Resection

Given that en bloc removal is the main clinical advantage of ESD over other endoscopic colorectal resection techniques, the goal of this test was to evaluate whether the Helping Hand and knife could together complete the full grasp, retract, and cut workflow. To replicate single channel deployment, we fixed the Helping Hand to the tip of a mock endoscope, which constrained its motion to the translation it would undergo while sweeping across the dissection plane. We used chicken thigh tissue as the colon surrogate because its layered connective tissue approximates the thickness and compliance of the colonic wall. We placed the tissue on the return electrode with its top surface level with the base of the Helping Hand and connected the knife to an electrosurgical generator (ForceTriad, Valleylab).

The supplementary video and Figure 9 illustrate the workflow. We routed the Helping Hand to the target, exposed the knife through the knife slot, and closed the grasper on the tissue to apply traction. We then made an initial incision with the knife, lifted the cut flap with the Helping Hand, and advanced the cut while manually actuating the mock endoscope until we had fully removed the target region and retrieved the specimen. The full procedure took 15 min. These results showed that the Helping Hand and knife, deployed together through a single 3.7 mm working channel, provided both traction and cutting for en bloc resection without external tooling.

En bloc resection workflow in ex vivo chicken thigh tissue

Figure 9. En bloc resection in ex vivo chicken thigh tissue with the Helping Hand and electrosurgical knife deployed from the tip of a mock endoscope. The panels show (a) insertion and alignment, (b) repositioning and knife exposure, (c) grasper positioning, (d) traction and initial incision, (e) lifting of the cut flap, (f) extraction of the grasped specimen, and (g) close up views. The Helping Hand and knife together completed grasping, traction, cutting, and retrieval through one working channel (See Supplementary Video S1).

Conclusion

In this paper, we presented the Helping Hand, a continuum robotic grasper for ESD that, together with an electrosurgical knife, enables tissue traction and cutting through the working channel of a standard single-channel colonoscope. We achieved this by creating a two-segment, notched continuum robot with a slot built into the base to expose the knife. We characterized the workspace of the Helping Hand and its ability to provide traction forces, and we demonstrated successful en bloc resection in ex vivo tissue. Overall, these results establish the feasibility of a continuum robotic grasper for ESD that could operate through a single working channel.

There are several directions to be explored in future work. First, the kinematic models could be extended to account for tendon elongation, actuation compliance, and friction, or replaced with a mechanics-based model that relates tendon tension to notch bending. Second, closed-loop control, using tension feedback from the existing load cells or visual feedback from the colonoscope camera, could also be incorporated. Finally, the Helping Hand could enable automation of portions of the ESD workflow, such as adjusting countertraction in response to forces applied by the knife and the tissue moving. Overall, by eliminating the need for a second working channel, an auxiliary endoscope, or intra-procedural tool exchanges, the Helping Hand could enable physicians to perform ESD using the standard colonoscopes already present in most endoscopy suites, lowering the procedural barrier that currently limits ESD adoption outside specialized centers.

References

References

  1. [1]National Cancer Institute, “SEER cancer stat facts: Colorectal cancer,” https://seer.cancer.gov/statfacts/html/colorect.html, accessed: Nov. 10, 2026.
  2. [2]M. Ferlitsch, C. Hassan, R. Bisschops et al., “Colorectal polypectomy and endoscopic mucosal resection: European Society of Gastrointestinal Endoscopy (ESGE) guideline – update 2024,” Endoscopy, 2024.DOI
  3. [3]A. F. Feery, N. J. Shaheen, K. S. Cools, T. H. Baron, M. Koruda, J. A. Galanko, and I. S. Grimm, “Morbidity and mortality after surgery for nonmalignant colorectal polyps,” Gastrointestinal Endoscopy, vol. 87, no. 1, pp. 243–250.e2, 2018.
  4. [4]F. Nakamura, Y. Saito, T. Sakamoto, Y. Otake, T. Nakajima, S. Yamamoto, Y. Murakami, H. Ishikawa, and T. Matsuda, “Potential perioperative advantage of colorectal endoscopic submucosal dissection versus laparoscopy-assisted colectomy,” Surgical Endoscopy, vol. 29, no. 3, pp. 596–606, 2015.DOI
  5. [5]M. Nagata, “Advances in traction methods for endoscopic submucosal dissection: What is the best traction method and traction direction?” World Journal of Gastroenterology, vol. 28, no. 1, pp. 1–22, 2022.DOI
  6. [6]M. Nagata, “Device-assisted traction methods in colorectal endoscopic submucosal dissection and options for difficult cases,” World Journal of Gastrointestinal Endoscopy, vol. 15, no. 4, pp. 265–272, 2023.DOI
  7. [7]Olympus, “Single use endoscopic lifting device LA-400 FlexLifter,” https://www.olympusprofed.com/gi/esd/36728/, OLYMPUS CONTINUUM. Accessed: 2026-10-29.
  8. [8]H. Lou, Y. Shu, J. Yang, Z. Zhou, M. Runciman, and G. Mylonas, “Soft pop-up serial robot with cable-hydraulic hybrid actuation for advanced endoscopy,” in 2026 IEEE 9th International Conference on Soft Robotics (RoboSoft). IEEE, 2026, pp. 1246–1252.DOI
  9. [9]J. Gafford, T. Ranzani, S. Russo, H. Aihara, C. Thompson, R. Wood, and C. Walsh, “Snap-on robotic wrist module for enhanced dexterity in endoscopic surgery,” in 2016 IEEE International Conference on Robotics and Automation (ICRA). IEEE, 2016, pp. 4398–4405.DOI
  10. [10]FUJIFILM Healthcare Americas Corporation, “Third space endoscopy,” https://healthcaresolutions-us.fujifilm.com/products/endoscopy/interventional-gastroenterology/third-space-endoscopy/, tRACMOTION product specifications. Accessed: 2026-10-29.
  11. [11]P. Connor, J. Gafford, P. Anderson, S. Webster, S. D. Herrell, III, K. L. Obstein, N. Kumar, C. Rucker, and R. J. Webster, III, “Steerable sheaths that turn endoscopes into robotic platforms for colon interventions,” Journal of Medical Devices, vol. 20, no. 5, p. 051004, 2026.DOI
  12. [12]H. Gao, X. Yang, T. Zhang, X. Xiao, C. Li, M. Q.-H. Meng, X. Zuo, Y. Li, and H. Ren, “Transendoscopic telerobotic system: Heterogeneous flexible manipulators for bimanual endoscopic submucosal dissection,” IEEE Robotics & Automation Magazine, 2025.DOI
  13. [13]R. F. Knoop, E. Wedi, G. Petzoldt, S. C. B. Bremer, A. Amanzada, V. Ellneried, A. Neesse, and S. Kunsch, “Endoscopic submucosal dissection with an additional working channel (ESD+): a novel technique to improve procedure time and safety of ESD,” Surgical Endoscopy, vol. 35, no. 7, pp. 3506–3512, 2021.DOI
  14. [14]P. A. York, P. J. Swaney, H. B. Gilbert, and R. J. Webster, 3rd, “A wrist for needle-sized surgical robots,” IEEE Int. Conf. Robot. Autom., vol. 2015, pp. 1776–1781, May 2015.pubmed.ncbi.nlm.nih.gov/26405562
  15. [15]K. Chitalia, S. Jeong, K. Y. Yamamoto, J. J. Chern, and J. P. Desai, “Modeling and control of a 2-dof meso-scale continuum robotic tool for pediatric neurosurgery,” IEEE Transactions on Robotics, vol. 37, no. 2, pp. 520–531, 2020.DOI
  16. [16]A. Mangan, S. Kim, N. Jones, M. G. Brandel, J. J. Heit, A. Norbash, J. T. Hwang, E. Hawkes, and T. K. Morimoto, “Serially-connected soft continuum robots for endovascular emergencies,” IEEE Transactions on Medical Robotics and Bionics, vol. 7, no. 3, pp. 1029–1040, 2025.DOI
  17. [17]Y. Xu, A. Mangan, M. Potdar, A. Lu, and T. K. Morimoto, “Design and characterization of a unified continuum robot actuation platform,” arXiv preprint arXiv:2610.19798, 2026.arxiv.org/abs/2610.19798
  18. [18]K. Belsare, A. C. Rodriguez, P. G. Sánchez, J. Hierro, T. Kolcon, R. Lange, I. Lütkebohle, A. Malki, J. M. Sosa, F. Melendez et al., “Micro-ros,” in Robot Operating System (ROS) The Complete Reference (Volume 7). Springer, 2023, pp. 3–55.DOI
  19. [19]A. Mangan, K. Long, K. M. B. Lee, M. Potdar, N. Atanasov, and T. K. Morimoto, “Contact-aware planning and control of continuum robots in highly constrained environments,” arXiv preprint arXiv:2604.15638, 2026.arxiv.org/abs/2604.15638
  20. [20]K. Oliver-Butler, J. A. Childs, A. Daniel, and D. C. Rucker, “Concentric push–pull robots: Planar modeling and design,” IEEE Transactions on Robotics, vol. 38, no. 2, pp. 1186–1200, 2021.

Paper details

vlnc:W_fzb4pw42
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 12

downloads · 1

A view counts once per browser per day, a download each time the PDF is requested.

  • Computer Science

Not recorded for this paper

View statement
Looking for related papers…
Contents