Reconstructing 3D+t Cardiovascular Flow From Sparse Planar Velocity Fields With Neural Fields

Abstract

Reconstructing time-resolved three-dimensional (3D+t) cardiovascular flow from sparse planar measurements is challenging because each plane observes only the two in-plane velocity components, on a small part of the volume. We propose a plane-aware neural-field framework that learns a continuous 3D velocity field from space and time coordinates through a per-plane projection, and then investigate the potential benefits of physics constraints. We evaluate the framework on computational aortic and aneurysm datasets, and on planes resampled from a real cardiac 4D Flow MRI scan. A plain multi-layer perceptron (MLP) was more accurate than classical radial basis function (RBF) interpolation on both computational datasets. For the aorta, a physics-constrained SIREN (PC-SIREN) reduced the rRMSE further to 0.37560.3756, compared with 0.42730.4273 for the plain MLP and 0.72560.7256 for RBF, and it was the most accurate model in every anatomical segment of the MRI data. For the aneurysms the plain MLP remained the most accurate model, with an rRMSE of 0.070.07 to 0.160.16 against 0.550.55 to 0.710.71 for PC-SIREN. Further experiments investigated scanning configuration, number of scans, and spatial resolution. These findings show that continuous neural fields can reconstruct volumetric cardiovascular flow from sparse planar measurements, and that physics constraints have to be matched to the flow.

Introduction

Cardiovascular disease remains the leading cause of death globally, highlighting the need for accurate and accessible medical imaging tools [5]. Although magnetic resonance imaging (MRI) and computed tomography (CT) are commonly used for diagnosis due to their 3D imaging capabilities, echocardiography is a faster and less invasive alternative, while also being accessible to patients with medical implants. Significant advances in ultrasound hardware and computational methods have enabled capable 3D and time-resolved imaging [10, 16]. Despite these advances, 3D ultrasound remains susceptible to artifacts such as stitching, blooming, and signal dropout, particularly when volumetric information is reconstructed from sparse observations [27]. The larger size of matrix-array transducers can also increase susceptibility to acoustic shadowing, while the greater number of elements introduces a more pronounced trade-off between spatial and temporal resolution [19].

As an alternative to relying solely on raw anatomical images, blood flow dynamics provide a sensitive indicator of cardiovascular abnormalities, motivating diagnostic approaches based on haemodynamic measurements for conditions such as aortic coarctation and heart failure [29, 38]. Blood velocity is clinically assessed using Doppler ultrasound, although this only measures a single velocity component along the ultrasound beam. Many techniques have been developed to recover 2D velocity fields, including vector flow mapping (VFM), blood speckle tracking (BST), and echocardiographic particle image velocimetry (EchoPIV) [2, 11, 22]. Current 3D ultrasound flow imaging, such as 3D colour Doppler, is typically limited to measuring a single component of the velocity field, while approaches such as 3D EchoPIV aim to recover three-dimensional flow information but remain constrained by the trade-off between spatial and temporal resolution [4, 33]. Consequently, 2D techniques can provide more accurate and temporally resolved flow measurements, but these planar observations do not fully capture the inherently three-dimensional dynamics of cardiovascular flow.

These limitations raise the question of whether the accuracy of planar velocimetry can be combined with the volumetric representation of 3D flow by reconstructing the unobserved velocity field from sparse planar measurements. We propose a neural field framework that represents cardiovascular flow as a continuous function of space and time, enabling reconstruction of transient 3D velocity fields from a limited number of 2D observations. In this work, the planar velocity fields are sampled from computational simulations and from a 4D Flow MRI scan, standing in for the output of 2D ultrasound velocimetry, so that each reconstruction can be evaluated against a known 3D reference. The resulting representation provides continuous access to the reconstructed flow field throughout the volume and over time, while providing a potential alternative to dense volumetric flow acquisition that is used in 3D ultrasound. We systematically evaluate the influence of measurement density and scanning configuration on reconstruction accuracy and computational cost, and further investigate robustness to increasingly complex flow geometries. Our framework is benchmarked against a classical interpolation baseline. Furthermore, the continuous representation also enables the incorporation of physics-based constraints.

In summary, our contributions are

  • Physics-free 3D+t reconstruction from planes: We propose a plane-aware neural-field framework that reconstructs time-resolved 3D cardiovascular velocity fields from two-component measurements on freely oriented planes. Unlike prior learning-based reconstructions from planes, it requires no governing equations, boundary conditions, or pressure data, and the network is interchangeable.

  • Acquisition design: We compare four handheld-probe scanning patterns and scan counts from 4 to 100, and identify two-axis rotation with 20 scans as an accurate and practical protocol.

  • Validation across flows and on real data: Across aortic coarctation, intracranial aneurysms, and real whole-heart 4D Flow MRI, the framework reduces the rRMSE of RBF interpolation by 13 to 89%, and we show where physics constraints add further accuracy.

Related Work

Implicit Neural Representations Neural fields encode a continuous signal as a network mapping spatial coordinates, and optionally time, to a quantity of interest such as colour, density or velocity [12, 23, 37]. NeRF [21] learns a continuous scene representation from posed 2D images by optimising a photometric loss through differentiable volume rendering, and has since been adapted to cross-sectional imaging [28]. In medical imaging, MedNeRF [6] learns continuous CT representations from sparse projections with an adversarial objective. For ultrasound, Ultra-NeRF [36] models planar measurements through tissue-dependent attenuation, reflectance, and scattering, and UIRe-NeRF [14] additionally incorporates reflected signals. Li et al. [20] reconstruct scoliotic spines from 3D ultrasound, separating soft tissue and bone better than voxel-based nearest-neighbour methods. Dynamic extensions condition the field on time: D-NeRF [25] factorises a scene into a static canonical field and a learned time-dependent deformation, and EndoNeRF [34] applied this to binocular endoscopic video to remove instrument occlusions. Dynamic scene reconstruction from 2D ultrasound with neural fields remains largely unexplored.

3D Flow Reconstruction Neural fields have recently been used for 3D+t cardiovascular flow. Garzia et al. [12] model wall motion to constrain 4D Flow MRI reconstruction, and Wong et al. [35] reconstruct 3D velocity fields from 3D colour Doppler. Both, however, assume volumetric input and do not address sparse 2D measurements. For the latter, function-based interpolation such as radial basis functions (RBFs) has proven effective for reconstructing volumes from 2D slices [16]. Poelma et al. [24] recover out-of-plane velocity from 2D acquisitions by combining EchoPIV in-plane fields with correlation peak strength, but rely on regularly spaced scans of a curved tube and a parabolic profile fit. RBFs can accurately approximate more complex profiles, motivating our investigation of whether continuous neural fields offer improved accuracy and flexibility for reconstructing transient 3D flow from sparse planar measurements.

Flow Reconstruction From Planes Physics-informed networks have been used to infer 3D flow from planar velocity data. Cai et al. [3] reconstruct a cylinder wake from a few axis-aligned planes, and Epshtein et al. [8] use a projection loss for tilted planes in steady, low-Reynolds-number cerebrospinal fluid flow. Both depend on the Navier–Stokes residuals, and the latter also on wall boundary data, to recover the unobserved component. Both are evaluated on simulated data, and neither handles time-resolved flow on freely oriented planes. Triplane vector flow mapping [32] measures one velocity component per plane and is specific to the left ventricle, while implicit representations for 4D Flow MRI [9, 26] require volumetric input. Our framework removes these restrictions: it reconstructs time-resolved flow from freely oriented two-component planes, requires neither governing equations nor boundary data, and is validated on three anatomies including real whole-heart measurements.

Methodology

Figure 1 shows a schematic of the proposed framework.

Our proposed framework for reconstructing 3D velocity fields from 2D velocity fields

Figure 1. Our proposed framework takes 2D velocity fields, converts local coordinates into global 3D coordinates using the transducer position xiUS\mathbf{x}_i^{\mathrm{US}} and orientation ni\mathbf{n}_i, and a MLP predicts the 3D velocity vector. The loss function optimises the reconstruction of the 2D projection in accordance with physical constraints.

Problem Formulation and Preprocessing

Given a set of scans D={Ii∈RT×H×W}i=1N\mathcal{D} = \{I_i \in\mathbb{R}^{T \times H \times W}\}_{i=1}^{N}, velocimetry is performed to produce a 2D velocity field Ui∈RT×H×W×2\mathbf{U}_i \in\mathbb{R}^{T \times H \times W \times2} for each scan. Additionally, the position of the transducer xiUS\mathbf{x}_i^{\mathrm{US}} and the normal to the imaging plane ni\mathbf{n}_i are recorded. The Cartesian grid produced by velocimetry exists in a local 2D image coordinate system, with coordinates xi2D\mathbf{x}_i^{2\mathrm{D}} for each scan IiI_i. To transform the points to a global coordinate system, a plane basis (e1i,e2i)(\mathbf{e}_1^i,\mathbf{e}_2^i) was formed from ni\mathbf{n}_i and the transformation was defined as

xi3D=xiUS+(e1ie2i)T⋅xi2D,(1)\mathbf{x}_i^{3\mathrm{D}} = \mathbf{x}_i^{\mathrm{US}} + \begin{pmatrix} \mathbf{e}_1^i \\ \mathbf{e}_2^i \end{pmatrix}^{\mathsf{T}} \cdot\mathbf{x}_i^{2\mathrm{D}}, \tag*{(1)}

where xi3D\mathbf{x}_i^{3\mathrm{D}} is the point in the global 3D coordinate system. Points were masked based on manual segmentation of boundary walls, with points lying in the region of interest denoted by x~i\widetilde{\mathbf{x}}_i.

To render 2D views from the 3D flow field u:R3×R→R3\mathbf{u} : \mathbb{R}^{3} \times\mathbb{R} \to\mathbb{R}^{3}, the projection operator was defined as

Pj(v)=(e1je2j)⋅v,(2)\mathcal{P}_j(\mathbf{v}) = \begin{pmatrix} \mathbf{e}_1^j \\ \mathbf{e}_2^j \end{pmatrix} \cdot\mathbf{v}, \tag*{(2)}

where (e1j,e2j)(\mathbf{e}_1^j,\mathbf{e}_2^j) are a plane basis from a known normal nj\mathbf{n}_j, and v∈R3\mathbf{v} \in\mathbb{R}^{3} is a sample from u\mathbf{u}. The inverse problem consists of recovering u\mathbf{u} from a collection of sparse 2D observations acquired at different probe poses.

The models were evaluated with the relative root mean squared error (rRMSE) defined as

rRMSE=∑i=1N∥vi−vi∗∥2∑i=1N∥vi∗∥2,(3)\mathrm{rRMSE} = \sqrt{\frac{\sum_{i=1}^{N} \lVert\mathbf{v}_i - \mathbf{v}_i^* \rVert^2}{\sum_{i=1}^{N} \lVert\mathbf{v}_i^* \rVert^2}}, \tag*{(3)}

where vi,vi∗∈R3\mathbf{v}_i, \mathbf{v}_i^* \in\mathbb{R}^{3} are the predicted and ground truth velocity vectors, respectively. The sums run over all evaluation points and time frames before the ratio is taken, so regions of low velocity are not weighted more heavily than regions of high velocity.

Neural Field Model

We use a SiLU activated neural network parameterised by θ\theta, fθ:R4→R3f_\theta: \mathbb{R}^{4} \to\mathbb{R}^{3}, to represent the 3D+tt velocity field u\mathbf{u} with the transformed 3D coordinates, xi3D\mathbf{x}_i^{3\mathrm{D}}, and time as inputs. The plain MLP uses SiLU activations, and the physics-constrained SIREN (PC-SIREN) uses sine activations [30]. The two-branch architecture used in D-NeRF [25] was not an appropriate choice for our problem as it operates under the assumption that the transient velocity field can be explained by deforming a canonical field, which is not generally a physically natural assumption for fluid flow. The deformation network may also be susceptible to warping the field in ways that fit observations but produce physically implausible dynamics. Therefore, most suitable was a network expressing the velocity field as a single function of space and time, in conjunction with the physics-based loss function comprising of three parts:

Data Reconstruction Loss The reconstruction loss was defined as

Ldata=1N∑k=1N∥Pi(k)[fθ(x~k,tk)]−uk2D∥22,(4)\mathcal{L}_{\mathrm{data}} = \frac{1}{N}\sum_{k=1}^{N}\left\lVert\mathcal{P}_{i(k)}\left[f_\theta(\widetilde{\mathbf{x}}_k,t_k)\right] - \mathbf{u}_k^{2\mathrm{D}}\right\rVert_2^2, \tag*{(4)}

where uk2D∈R2\mathbf{u}_k^{2\mathrm{D}} \in\mathbb{R}^{2} is the in-plane velocity measured at sample kk, (x~k,tk)(\widetilde{\mathbf{x}}_k,t_k) is its position and time, i(k)i(k) is the scan it belongs to, and ∥⋅∥2\lVert\cdot\rVert_2 is the L2L_2-norm.

Smoothness Loss To mitigate the effect of noise in the 2D measurements, a first-order smoothness loss, denoted LTV\mathcal{L}_{\mathrm{TV}}, was defined as

LTV=1M∑i=1M∥fθ(x~i3D,t)−fθ(x~i3D+δx,t)∥22,(5)\mathcal{L}_{\mathrm{TV}} = \frac{1}{M}\sum_{i=1}^{M}\left\lVert f_\theta(\widetilde{\mathbf{x}}_i^{3\mathrm{D}},t) - f_\theta(\widetilde{\mathbf{x}}_i^{3\mathrm{D}}+\delta\mathbf{x},t)\right\rVert_2^2, \tag*{(5)}

where MM is the number of samples and δx\delta\mathbf{x} is a small perturbation in each spatial component.

Physics Loss To ensure the reconstructed velocity field conserved mass and momentum, the divergence loss was defined as

Ldiv=1M∑i=1M∣∇⋅fθ(x~i3D,t)∣,(6)\mathcal{L}_{\mathrm{div}} = \frac{1}{M}\sum_{i=1}^{M}\left\lvert\nabla\cdot f_\theta(\widetilde{\mathbf{x}}_i^{3\mathrm{D}},t)\right\rvert, \tag*{(6)}

and the velocity-vorticity formulation of the Navier-Stokes loss [17] was defined as

LNS=1M∑i=1M∣∂ω∂t+∇×(ω×fθ(x~i3D,t))+1Re∇×∇×ω∣,(7)\mathcal{L}_{\mathrm{NS}} = \frac{1}{M}\sum_{i=1}^{M}\left\lvert\frac{\partial\boldsymbol{\omega}}{\partial t} + \nabla\times\left(\boldsymbol{\omega} \times f_\theta(\widetilde{\mathbf{x}}_i^{3\mathrm{D}},t)\right) + \frac{1}{\mathrm{Re}}\nabla\times\nabla\times\boldsymbol{\omega} \right\rvert, \tag*{(7)}

where ω=∇×fθ\boldsymbol{\omega} = \nabla\times f_\theta is the vorticity.

The total loss function was

L=λdataLdata+λTVLTV+λdivLdiv+λNSLNS,(8)\mathcal{L} = \lambda_{\mathrm{data}}\mathcal{L}_{\mathrm{data}} + \lambda_{\mathrm{TV}}\mathcal{L}_{\mathrm{TV}} + \lambda_{\mathrm{div}}\mathcal{L}_{\mathrm{div}} + \lambda_{\mathrm{NS}}\mathcal{L}_{\mathrm{NS}}, \tag*{(8)}

where λi∈R\lambda_i \in\mathbb{R} are loss weights. The plain MLP uses Ldata\mathcal{L}_{\mathrm{data}} only, so it needs no governing equations or boundary conditions.

Projection-Aware RBF Interpolation Baseline

Given a set of observed positions and their associated velocity measurements, RBF methods approximate the underlying field using distance-dependent basis functions at the observations [16]. In the standard formulation, a scalar field f(x)f(\mathbf{x}) is represented as

f(x)=∑i=1Nwiϕ(∣x−xi∣),(9)f(\mathbf{x}) = \sum_{i=1}^{N} w_i \phi(\lvert\mathbf{x} - \mathbf{x}_i \rvert), \tag*{(9)}

where xi\mathbf{x}_i denotes the ii-th measurement location, ϕ\phi is a radial basis function, and wiw_i are coefficients determined from the observed data, potentially with a smoothing term. For velocity reconstruction, however, measurements acquired on differently oriented planes provide components in different local coordinate systems. Consequently, these components can not be directly interpolated as independent global velocity components. Instead, at a query location x\mathbf{x}, the 3D velocity estimate v^(x,t)\hat{\mathbf{v}}(\mathbf{x},t) is obtained by combining neighbouring observations according to their spatial proximity while accounting for each measurement plane’s orientation. Let Pi∈R2×3\mathbf{P}_i \in\mathbb{R}^{2 \times3} denote the projection matrix whose rows are the two in-plane unit basis vectors, and let ui∈R2\mathbf{u}_i \in\mathbb{R}^{2} denote the measured in-plane velocity. The local estimate can then be formulated as a regularised weighted least-squares problem

v^(x,t)=arg⁡min⁡v∈R3 ∑i∈N(x,t)wi(x,t)∥Piv−ui∥22+λ∥v∥22,(10)\hat{\mathbf{v}}(\mathbf{x},t) = \underset{\mathbf{v}\in\mathbb{R}^{3}}{\arg\min}\ \sum_{i\in\mathcal{N}(\mathbf{x},t)} w_i(\mathbf{x},t)\lVert\mathbf{P}_i\mathbf{v} - \mathbf{u}_i\rVert_2^2 + \lambda\lVert\mathbf{v}\rVert_2^2, \tag*{(10)}

where N(x,t)\mathcal{N}(\mathbf{x},t) denotes the selected neighbouring measurements, wiw_i are distance-dependent weights, and λ\lambda controls regularisation. This formulation accounts for the orientations of the measurement planes when estimating the global velocity vector. The resulting estimate is evaluated at query locations throughout the reconstruction domain to obtain a continuous approximation of the flow field.

Experimental Setup

Datasets

Coarctation of Aorta (CoA) The AortaPINNs dataset [18] contained computational fluid dynamics (CFD) simulations of an aorta geometry at different levels of coarctation, quantified by the relative diameter of the narrowed region. In this work, simulations with 30%, 40%, 60%, 80%, and 100% of the full diameter were used to capture a broad variation in haemodynamics.

Intracranial Aneurysms (IA) The AnXplore dataset [13] contained fluid-structure interaction (FSI) simulations of various aneurysm geometries obtained from real patient cases with the vessel geometry kept constant to ensure that variation in dynamics were caused by aneurysm sac geometry alone. In this work, volume of the aneurysm sac was varied as it has been shown to be a strong indicator of risk of rupture and significantly alters blood flow [7].

Full Heart MRI Arshad et al. [1] performed a thoracic 4D Flow MRI scan of a patient at rest and manual segmentation of the main vessels and chambers was performed with 3D Slicer (v5.12.4). Velocity vectors were masked according to the result of the segmentation.

Implementation

Each instance of the model was trained using two NVIDIA RTX A5000 GPUs with 24GB memory and a hold-out scan was removed from the training set for evaluation of novel-view synthesis. For the CoA and IA datasets, the sliced 2D flow fields were 200 ×\times 200 pixels, but the slices of the MRI dataset were 200 ×\times 400 pixels. Each instance of the model was evaluated on 30000 points in the CoA dataset, 10000 points in the IA dataset, and 38985 points in the MRI dataset. The values in Table 1, Table 2 and Table 3 are means over three training runs with different random seeds. Implementation details, including the training settings, are given in the supplementary material, and we defer further results to the supplementary material.

Model3D rRMSEHold-out rRMSE
RBF0.72560.7696
Plain MLP0.42730.7623
PC-SIREN0.37560.4950

Table 1. rRMSE of predicted 3D velocity field and hold-out scan for aorta geometry with 60% radius.

SampleRBFPlain MLPPC-SIREN
#110.55650.11830.5662
#1170.58670.16350.5543
#1610.57240.12220.6640
#1830.61960.07060.6171
#1950.57190.07730.7141

Table 2. rRMSE of 3D reconstruction of aneurysm flow.

Model50×5050 \times50100×100100 \times100200×200200 \times200
Plain MLP0.52310.43870.4273
PC-SIREN0.38980.36590.3756

Table 3. Effect of scan resolution on rRMSE for aorta with 60% radius.

Scanning Configurations

Figure 2 shows the different ways that scans could be acquired by either translating the transducer along an axis or by rotating it about various axes. For each scanning mode, training data was generated with every combination of axes for a certain geometry (CoA with 60% coarctation) and a plain multi-layer perceptron (MLP) was trained for each scanning mode and axis. The best configuration was used for the remaining experiments.

Visualisation of the different scanning modes

Figure 2. Visualisation of the different scanning modes that were tested.

Ablations

Experiments initially used a plain MLP, trained using only Ldata\mathcal{L}_{\mathrm{data}}, without coordinate encoding or physics-based regularisation. First, the number of scans was fixed at 20 to compare different scanning configurations and identify the most effective acquisition strategy. The selected configuration was then fixed while the number of scans was varied between 4 and 100. The final scan count was chosen as a compromise between reconstruction accuracy and training time. The same procedure was repeated for the classical RBF baseline on the same scans, independently selecting its optimal scanning configuration and number of scans to ensure a fair comparison.

With the scan configuration and number of scans fixed, different coordinate representations were investigated to assess the contribution of encoding to the representation of complex flow fields. These included using only scaled spatial and temporal coordinates without additional encoding, Fourier feature encoding [31] (applied to spatial and temporal coordinates separately and jointly), and a SIREN [30] network. Finally, the loss weights in Eq. (8) were optimised. A sequential approach was adopted in which individual loss weights were varied while the remaining weights were held fixed. The reconstruction weight λdata\lambda_{\mathrm{data}} was maintained at 1 throughout. First, λNS\lambda_{\mathrm{NS}} was varied with λdiv=λTV=0\lambda_{\mathrm{div}} = \lambda_{\mathrm{TV}} = 0. Next, the selected λNS\lambda_{\mathrm{NS}} was fixed while λdiv\lambda_{\mathrm{div}} was varied. Finally, the selected λNS\lambda_{\mathrm{NS}} and λdiv\lambda_{\mathrm{div}} were fixed while λTV\lambda_{\mathrm{TV}} was varied. The order was chosen based on the complexity of the individual loss function, and the procedure was used to determine the final loss weighting for the proposed model. All of the selection described above was carried out on the 60% coarctation case, and the selected configuration was applied unchanged to all other cases.

Results and Discussion

Acquisition Strategy and Model Selection

The two-axis rotational scanning configuration achieved the best 3D reconstruction accuracy for both the MLP-based and RBF models. In particular, sampling across two rotational axes provided more diverse measurement orientations, enabling better reconstruction of out-of-plane velocity components compared with the layered scanning configuration. Reconstruction accuracy improved rapidly as the number of scans increased, with performance largely converging at approximately 50 scans. However, the improvement between 20 and 50 scans was marginal, despite the latter requiring over twice the training time. Therefore, 20 scans were selected as a compromise between reconstruction accuracy and computational cost.

For the aorta dataset, SIREN encoding achieved substantially better reconstruction accuracy than the Fourier feature encoding strategies and a modest improvement over the plain MLP. Consequently, SIREN was selected for the proposed model. The final loss weights were determined using the sequential optimisation procedure described above, yielding λdata=1\lambda_{\mathrm{data}} = 1, λTV=10−3\lambda_{\mathrm{TV}} = 10^{-3}, λdiv=10−2\lambda_{\mathrm{div}} = 10^{-2}, and λNS=10−3\lambda_{\mathrm{NS}} = 10^{-3}. This physics-constrained SIREN model is denoted as PC-SIREN.

Reconstruction Performance

Table 1 shows that the PC-SIREN model achieves lowest rRMSE for both the full 3D volume and the held-out measurements when reconstructing the aorta with 60% coarctation severity, outperforming both the plain MLP and the RBF baseline. This quantitative improvement is also reflected in the reconstructions shown in Figure 3 across multiple timesteps. While the plain MLP captures the overall flow structure, its reconstruction exhibits slightly greater spatial variation, whereas PC-SIREN produces a smoother and more coherent velocity field while retaining prominent flow features, including the vortex. In contrast, the RBF baseline provides a less continuous representation of the underlying flow.

3D reconstructions of the aorta geometry for different coarctation levels and vortex flow

Figure 3. 3D Reconstruction of the aorta geometry with 60% radius. Though similar, PC-SIREN captures the transition between the vortex and downstream flow smoother than the plain MLP.

Figure 4 shows that the reconstruction quality improves as the level of coarctation approaches a healthier regime (100% is a healthy aorta), suggesting that smoother flows with fewer disturbances are less challenging for the reconstruction framework. Importantly, the PC-SIREN model consistently achieves a lower rRMSE than the plain MLP at all coarctation levels.

rRMSE of Plain MLP and PC-SIREN across aorta radius values

Figure 4. PC-SIREN produced more accurate reconstructions than the plain MLP at all coarctation levels. Both models saw improved reconstruction quality as the geometry approached a healthier regime.

For the aneurysms, the plain MLP is the most accurate model on all five samples, with an rRMSE between 0.07 and 0.16, a reduction of 72 to 89% relative to the RBF baseline (Table 2). The framework therefore reconstructs these flows accurately from the projection loss alone, with no governing equations and no tuning beyond the aorta. PC-SIREN, whose configuration was selected on the 60% coarctation case and applied unchanged, does not improve on the plain MLP here. Furthermore, the aneurysm dataset exhibits relatively organised and less spatially varying flow patterns, which may be adequately represented by the lower-frequency functions learned by a plain MLP. In comparison, the higher-frequency representational capac-

ity of SIREN may be unnecessary for these smoother flow fields and could make optimisation more challenging. Because the network is interchangeable, the representation can be chosen per flow type without changing the framework.

Sensitivity of Resolution and Noise

The contrasting results across the aorta and aneurysm datasets suggest that model performance depends on the underlying flow characteristics. Across all tested resolutions, PC-SIREN outperforms the plain MLP for the aorta, whereas the opposite is observed for the aneurysm (Table 3, Figure 5). Increasing resolution improves aneurysm reconstruction, although gains plateau at 100×100100 \times100px. For the aorta, PC-SIREN is robust to coarse sampling: at 50×5050 \times50px, with sixteen times fewer samples per scan, its rRMSE of 0.39 is still lower than that of the plain MLP (0.43) and RBF (0.73) at 200×200200 \times200px. The contrast between the two datasets may reflect differences in flow complexity: the more spatially varying aortic flow may benefit from SIREN’s higher-frequency representational capacity and physics-based regularisation, while the smoother aneurysm flow may be better suited to a plain MLP. These findings highlight the importance of considering flow characteristics when selecting both network architecture and measurement resolution.

Effect of scan resolution on rRMSE for the aneurysm data

Figure 5. Effect of scan resolution on rRMSE for the aneurysm data. The plain MLP is accurate at all resolutions. Table 3 gives the corresponding results for the aorta.

The effect of multiplicative Rayleigh noise was evaluated at three scales (1.0, 5.0, and 10.0) for both datasets (Table 4). Improvement is the reduction in rRMSE relative to the noisy input. At scales 5.0 and 10.0 the noise is several times larger than the signal, so these rows are a stress test. RBF shows negative improvement at all noise levels. For the aneurysm, the plain MLP improves on the noisy input at low noise but deteriorates at higher levels, while PC-SIREN remains below the noisy input but becomes more competitive with the MLP as noise increases. In contrast, PC-SIREN outperforms both alternatives for the aorta, except at the highest level of noise where RBF achieves

ScaleInput rRMSERBFMLPPC-SIREN
1.00.4735−37.119648.4460−22.1539
5.04.1830−26.0619−21.7422−2.8498
10.08.9440−25.9106−24.8533−4.8789
1.00.4195−91.5792−59.59485.0322
5.03.7051−25.1175−35.6834−22.6175
10.07.9241−24.4633−31.4638−25.9701

Table 4. Effect of Rayleigh noise on rRMSE for aneurysm #11 (top) and aorta with 60% radius (bottom).

marginally better performance. Nevertheless, PC-SIREN consistently outperforms the plain MLP for the aorta across all tested noise levels. This supports the earlier findings that physics-based regularisation benefits the more complex aortic flow and may help mitigate measurement noise. Its limited benefit for aneurysm reconstruction, however, is also reflected in its response to noise. Overall, the effectiveness of PC-SIREN in handling noisy measurements appears dependent on the underlying flow characteristics and noise level.

Generalisation to Real Cardiac Data

For the real cardiac MRI dataset, measurement planes were positioned using subcostal and suprasternal [15] transducer locations (below and above the sternum, respectively), as these provided convenient access for establishing two-axis rotational scanning with approximately perpendicular rotation axes. The in-plane sampling resolution was also changed from square to rectangular to accommodate the greater depth of the cardiac region of interest. Finally, measurements were masked using the overall region of interest rather than individual anatomical segments. This ensured that the reconstructed domain remained a single connected region, avoiding the introduction of multiple disconnected subdomains that could otherwise be treated as separate closed systems by the reconstruction framework.

Table 5 shows that the PC-SIREN model achieves lower rRMSE than both the RBF baseline and plain MLP across the full heart and all evaluated anatomical segments, indicating improved reconstruction accuracy. The accuracy is also reflected in the qualitative comparisons in Figure 6, where PC-SIREN closely reproduces the ground-truth velocity fields across the full heart, aorta, and left ventricle. However, the higher-frequency representational capacity of SIREN may also allow it to fit local variations arising from measurement noise in the original 4D Flow MRI data. In contrast, the plain MLP’s bias towards lower-frequency representations can suppress some of these noisy variations, producing smoother reconstructions, albeit with slightly reduced agreement with the measured ground truth.

Qualitative comparison of RBF baseline and PC-SIREN model with ground truth 4D Flow MRI measurements of the full heart, aorta, and left ventricle

Figure 6. Qualitative comparison of RBF baseline and PC-SIREN model with ground truth 4D Flow MRI measurements of the full heart, aorta, and left ventricle.

SegmentRBFPlain MLPPC-SIREN
Full Heart0.76870.71840.6101
Aorta0.78270.72590.6303
Left Atrium0.70810.72480.6195
Left Ventricle0.86060.77000.6340
Pulmonary Artery0.66990.61850.5234
Right Atrium0.71170.67860.5920
Right Ventricle0.81720.71410.5729

Table 5. Quantitative comparison of models (rRMSE) on the real cardiac 4D Flow MRI dataset.

As the reference 4D Flow MRI data contain measurement noise, part of the remaining error reflects noise in the reference rather than in the reconstruction. From scans acquired through only two acoustic windows, the framework reduces the rRMSE of RBF interpolation in every anatomical segment, by 13 to 30%. Overall, these results demonstrate that the proposed framework can reconstruct cardiovascular velocity fields to a level of agreement comparable to 4D Flow MRI measurements.

Conclusion

We presented a plane-aware neural-field framework that reconstructs time-resolved 3D cardiovascular flow from sparse, freely oriented planar velocity measurements, without requiring governing equations, boundary conditions, or pressure data. It reduced the rRMSE of RBF interpolation by 48% for aortic coarctation, 72 to 89% for intracranial aneurysms, and 13 to 30% across the whole heart in real 4D Flow MRI. We investigated different acquisition strategies, sampling densities, and network architectures. A plain MLP was sufficient for the aneurysms, while a physics-constrained SIREN added accuracy at each coarctation level and in every cardiac segment, and remained accurate with sixteen times fewer samples per scan. Under increasing noise, PC-SIREN generally became more competitive, although RBF remained marginally better at the highest noise level for the aorta. Future work will address probe pose estimation, ultrasound-derived velocities, and a larger cohort.

LLM Usage Statement

LLM usage (Claude) was limited to editing and refining author-produced content. We can confirm that neither the full manuscript, or the major parts of it, were given to an LLM for reviewer-style feedback. This conception of this work was entirely author-produced.

References

  1. [1]Syed M. Arshad, Lee C. Potter, Chong Chen, Yingmin Liu, Preethi Chandrasekaran, Christopher Crabtree, Matthew S. Tong, Orlando P. Simonetti, Yuchi Han, and Rizwan Ahmad. Motion-robust free-running volumetric cardiovascular MRI. Magnetic Resonance in Medicine, 92(3):1248–1262, 2024. 4DOI
  2. [2]Martina Avesani, Bastien Degrelle, Giovanni Di Salvo, Jean-Benoit Thambo, and Xavier Iriart. Vector flow mapping: A review from theory to practice. Echocardiography, 38(8): 1405–1413, 2021. 1DOI
  3. [3]Shengze Cai, Zhiping Mao, Zhicheng Wang, Minglang Yin, and George Em Karniadakis. Physics-informed neural networks (PINNs) for fluid mechanics: A review. Acta Mechanica Sinica, 37:1727–1738, 2021. 2arxiv.org/abs/2105.09506
  4. [4]R. Chaoui, J. Hoffmann, and K. S. Heling. Three-dimensional (3D) and 4D color Doppler fetal echocardiography using spatio-temporal image correlation (STIC). Ultrasound in Obstetrics & Gynecology, 23(6):535–545, 2004. 1pubmed.ncbi.nlm.nih.gov/15170792
  5. [5]Bryan Chong, Jayanth Jayabaskaran, Silingga Metta Jauhari, Siew Pang Chan, Rachel Goh, Martin Tze Wah Kueh, Henry Li, Yip Han Chin, Gwyneth Kong, Vickram Vijay Anand, Jiong-Wei Wang, Mark Muthiah, Vardhmaan Jain, Anurag Mehta, Shir Lynn Lim, Roger Foo, Gemma A. Figtree, Stephen J. Nicholls, Mamas A. Mamas, James L. Januzzi, Nicholas W. S. Chew, A. Mark Richards, and Mark Y. Chan. Global burden of cardiovascular diseases: Projections from 2025 to 2050. European Journal of Preventive Cardiology, 32(11):1001–1015, 2025. 1DOI
  6. [6]Abril Corona-Figueroa, Jonathan Frawley, Sam Bond Taylor, Sarath Bethapudi, Hubert P. H. Shum, and Chris G. Willcocks. MedNeRF: Medical Neural Radiance Fields for Reconstructing 3D-aware CT-Projections from a Single X-ray. In 2022 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC), pages 3843–3848, 2022. 2
  7. [7]Sujan Dhar, Markus Tremmel, J Mocсо, Minsuok Kim, Junichi Yamamoto, Adnan H. Siddiqui, L. Nelson Hopkins, and Hui Meng. Morphology Parameters for Intracranial Aneurysm Rupture Risk Assessment. Neurosurgery, 63(2):185–197, 2008. 4DOI
  8. [8]Mark Epshtein, Tirosh Mekler, Mohammed Salman Shazeeв, Clifford Lindsay, Matthew J. Gounis, Netanel Korin, and Vania Anagnostakou. High-resolution 3D flow reconstruction of cerebrospinal fluid microcirculation using physics-informed neural network: Conceptualization and application to a large animal model. Computer Methods and Programs in Biomedicine, 283:109405, 2026. 2
  9. [9]Mojtaba F. Fathi, Isaac Perez-Raya, Ahmadreza Baghaie, Philipp Berg, Gabor Janiga, Amirhossein Arzani, and Roshan M. D’Souza. Super-resolution and denoising of 4D-Flow MRI using physics-informed deep neural nets. Computer Methods and Programs in Biomedicine, 197:105729, 2020. 2DOI
  10. [10]Aaron Fenster and Donal B. Downey. Three-Dimensional Ultrasound Imaging. Annual Review of Biomedical Engineering, 2:457–475, 2000. 1
  11. [11]F Forsberg, DA Merton, JB Liu, L Needleman, and BB Goldberg. Clinical applications of ultrasound contrast agents. Ultrasonics, 36(1-5):695–701, 1998. 1pubmed.ncbi.nlm.nih.gov/9651599
  12. [12]Simone Garzia, Patryk Rygiel, Sven Dummer, Filippo Cademartiri, Simona Celi, and Jelmer M. Wolterink. Neural Fields for Continuous Periodic Motion Estimation in 4D Cardiovascular Imaging. In Statistical Atlases and Computational Models of the Heart. Workshop, CMRxRecon and MBAS Challenge Papers., pages 378–389, Cham, 2025. Springer Nature Switzerland. 2arxiv.org/abs/2407.20728
  13. [13]Aurèle Goetz, Pablo Jeken-Rico, Ugo Pelissier, Yves Chau, Jacques Sédat, and Elie Hachem. AnXplore: A comprehensive fluid-structure interaction study of 101 intracranial aneurysms. Frontiers in Bioengineering and Biotechnology, 12, 2024. 4DOI
  14. [14]Ziwen Guo, Zi Fang, and Zhuang Fu. UIRe-NeRF: 3D Ultrasound Imaging through Neural Rendering with Ultrasound Reflection Direction Parameterization, 2024. 2arxiv.org/abs/2408.00860
  15. [15]W L Henry, A DeMaria, R Gramiak, D L King, J A Kisslo, R L Popp, D J Sahm, N B Schiller, A Tajik, L E Teichholz, and A E Weyman. Report of the American Society of Echocardiography Committee on Nomenclature and Standards in Two-dimensional Echocardiography. Circulation, 62(2):212–217, 1980. 7pubmed.ncbi.nlm.nih.gov/7397962
  16. [16]Qinghua Huang and Zhaozheng Zeng. A Review on Real-Time 3D Ultrasound Imaging Technology. BioMed Research International, 2017(1):6027029, 2017. 1, 2, 4DOI
  17. [17]Xiaowei Jin, Shengze Cai, Hui Li, and George Em Karniadakis. NSFnets (Navier-Stokes flow nets): Physics-informed neural networks for the incompressible Navier-Stokes equations. Journal of Computational Physics, 426:109951, 2021. 3DOI
  18. [18]Owais Khan. AortaPINNsData, 2025. 4
  19. [19]Roberto M. Lang, Luigi P. Badano, Wendy Tsang, David H. Adams, Eustachio Agricola, Thomas Buck, Francesco F. Faletra, Andreas Franke, Judy Hung, Leopoldo Pérez de Isla, Otto Kamp, Jaroslaw D. Kasprzak, Patrizio Lancellotti, Thomas H. Marwick, Marti L. McCulloch, Mark J. Monaghan, Petros Nihoyannopoulos, Natesa G. Pandian, Patricia A. Pellikka, Mauro Pepi, David A. Roberson, Stanton K. Shennan, Girish S. Shirali, Lissa Sugeng, Folkert J. Ten Cate, Mani A. Vannan, Jose Luis Zamorano, and William A. Zoghbi. EAE/ASE Recommendations for Image Acquisition and Display Using Three-Dimensional Echocardiography. Journal of the American Society of Echocardiography, 25(1):3–46, 2012. 1
  20. [20]Honggen Li, Hongbo Chen, Wenke Jing, Yuwei Li, and Rui Zheng. 3D Ultrasound Spine Imaging with Application of Neural Radiance Field Method. In 2021 IEEE International Ultrasonics Symposium (IUS), pages 1–4, 2021. 2
  21. [21]Ben Mildenhall, Pratul P. Srinivasan, Matthew Tancik, Jonathan T. Barron, Ravi Ramamoorthi, and Ren Ng. NeRF: Representing scenes as neural radiance fields for view synthesis. Communications of the ACM, 65(1):99–106, 2021. 2arxiv.org/abs/2003.08934
  22. [22]Siri A. Nyrnes, Solveig Fadnes, Morten Smedsrud Wigen, Luc Mertens, and Lasse Lovstakken. Blood Speckle-Tracking Based on High-Frame Rate Ultrasound Imaging in Pediatric Cardiology. Journal of the American Society of Echocardiography, 33(4):493–503.e5, 2020. 1DOI
  23. [23]Viraj Patel, Lisa Kreusser, and Katharine Fraser. Dynamic Reconstruction of Ultrasound-Derived Flow Fields With Physics-Informed Neural Fields, 2025. 2arxiv.org/abs/2511.01804
  24. [24]C. Poelma, J. M. Mari, N. Foin, M.-X. Tang, R. Krams, C. G. Caro, P. D. Weinberg, and J. Westerweel. 3D Flow reconstruction using ultrasound PIV. Experiments in Fluids, 50 (4):777–785, 2011. 2
  25. [25]Albert Pumarola, Enric Corona, Gerard Pons-Moll, and Francesc Moreno-Noguer. D-NeRF: Neural Radiance Fields for Dynamic Scenes, 2020. 2, 3arxiv.org/abs/2011.13961
  26. [26]Simone Saitta, Marcello Carioni, Subhadip Mukherjee, Carola-Bibiane Schönlieb, and Alberto Redaelli. Implicit neural representations for unsupervised super-resolution and denoising of 4D flow MRI. Computer Methods and Programs in Biomedicine, 246:108057, 2024. 2DOI
  27. [27]Muhamed Saric, Anita Sadeghpour, Leila Alizadeh, Azin Alizadeh, Philippe B. Bertrand, Kristen Billick, Bindu Chebrolu, Francesco Faletra, Anita Kelsey, Roberto Lang, Robert A. Levine, Leili Pourafkari, and Steve Walling. Recommendations for the Identification and Mitigation of Cardiac Ultrasound Artifacts: A Guideline from the American Society of Echocardiography. Journal of the American Society of Echocardiography, 39(5):435–475, 2026. 1DOI
  28. [28]Haim Sawdayee, Amir Vaxman, and Amit H. Bermano. OReX: Object Reconstruction From Planar Cross-Sections Using Neural Fields. In Proceedings of the IEEE/CVF Conference on Computer Vision and Pattern Recognition, pages 20854–20862, 2023. 2arxiv.org/abs/2211.12886
  29. [29]F. Secchi, A. Iozzelli, G. D. E. Papini, A. Aliprandi, G. Di Leo, and F. Sardanelli. MR imaging of aortic coarctation. La radiologia medica, 114(4):524–537, 2009. 1DOI
  30. [30]Vincent Sitzmann, Julien Martel, Alexander Bergman, David Lindell, and Gordon Wetzstein. Implicit Neural Representations with Periodic Activation Functions. In Advances in Neural Information Processing Systems, pages 7462–7473. Curran Associates, Inc., 2020. 3, 5arxiv.org/abs/2006.09661
  31. [31]Matthew Tancik, Pratul P. Srinivasan, Ben Mildenhall, Sara Fridovich-Keil, Nithin Raghavan, Utkarsh Singhal, Ravi Ramamoorthi, Jonathan T. Barron, and Ren Ng. Fourier Features Let Networks Learn High Frequency Functions in Low Dimensional Domains, 2020. 5arxiv.org/abs/2006.10739
  32. [32]Florian Vixège, Alain Berod, Pierre-Yves Courand, Simon Mendez, Franck Nicoud, Philippe Blanc-Benon, Didier Vray, and Damien Garcia. Full-volume three-component intraventricular vector flow mapping by triplane color Doppler. Physics in Medicine & Biology, 67(9):095004, 2022. 2DOI
  33. [33]Jason Voorneveld, Hicham Saaid, Christiaan Schinkel, Nikola Radeljic, Boris Lippe, Frank J. H. Gijsen, Antonius F. W. van der Steen, Nico de Jong, Tom Claessens, Hendrik J. Vos, Sasa Kenjeres, and Johan G. Bosch. 4-D Echo-Particle Image Velocimetry in a Left Ventricular Phantom. Ultrasound in Medicine & Biology, 46(3):805–817, 2020. 1DOI
  34. [34]Yuehao Wang, Yonghao Long, Siu Hin Fan, and Qi Dou. Neural Rendering for Stereo 3D Reconstruction of Deformable Tissues in Robotic Surgery. In Medical Image Computing and Computer Assisted Intervention – MICCAI 2022, pages 431–441, Cham, 2022. Springer Nature Switzerland. 2arxiv.org/abs/2206.15255
  35. [35]Hong Shen Wong, Wei Xuan Chan, Wenbin Mao, and Choon Hwai Yap. 3D velocity and pressure field reconstruction in the cardiac left ventricle via physics informed neural network from echocardiography guided by 3D color Doppler. Computer Methods and Programs in Biomedicine, 263:108671, 2025. 2DOI
  36. [36]Magdalena Wysocki, Mohammad Farid Azampour, Christine Eilers, Benjamin Busam, Mehrdad Salehi, and Nassir Navab. Ultra-NeRF: Neural Radiance Fields for Ultrasound Imaging. In Medical Imaging with Deep Learning, pages 382–401. PMLR, 2024. 2arxiv.org/abs/2301.10520
  37. [37]Yiheng Xie, Towaki Takikawa, Shunsuke Saito, Or Litany, Shiqin Yan, Numair Khan, Federico Tombari, James Tompkin, Vincent Sitzmann, and Srinath Sridhar. Neural fields in visual computing and beyond. In Computer Graphics Forum, pages 641–676. Wiley Online Library, 2022. 2arxiv.org/abs/2111.11426
  38. [38]Yue Zhong, Wei Zhu, Chun-Mei Li, and Li Rao. Assessment of cardiac dysfunction by dissipative energy loss derived from vector flow mapping. Journal of Cardiology, 67 (1):122, 2016. 1DOI

Paper details

vlnc:W_hy6ztqpn
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 7

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Computer Science

Claude

View statement
Looking for related papers…
Contents