How many events does entanglement-based positronium imaging need? Information limits of Compton polarimetry with CZT detectors

Abstract

The degree of polarisation correlation of the two annihilation photons, measured by Compton polarimetry, has been proposed as a PET biomarker of tissue microstructure, but the statistics it requires have not been quantified. We parameterise the measurement by the linear-polarisation correlation coefficient κκ (κ=1κ=1 maximally entangled, 1/21/2 separable), derive the classical Fisher information of double-Compton events and compare it with the quantum Fisher information, relate κκ to the annihilation channels of positronium in tissue, and simulate pixelated CdZnTe (CZT) rings. The model is checked against three published measurements and an independent GATE/Geant4 simulation. A double-Compton event carries at most 0.12-0.14 units of Fisher information about κκ and 0.015 on average, 0.1-0.6% of the quantum limit. Event-wise maximum likelihood reaches the Cramer-Rao bound, whereas the conventional ratio R needs 1.2-2.3 times more events. The observed coefficient is κobs=1−α(1−κpick)κ_{obs}=1-α(1-κ_{pick}), with a pick-off share αα of 0.13-0.29 in tissue; it is sensitive to lipid fraction but not to oxygen, and carries no tissue information if pick-off photons are maximally entangled. Published porous-polymer data imply κpickκ_{pick} = 0.34-0.58. In a 15-mm CZT module, ordering ambiguity, position errors and electron transport roughly halve the information per event, and sequence reconstruction of multi-site events recovers this loss. With 18^{18}F and an eight-module ring, κκ can be measured to ±0.01\pm0.01 in a 10-mm tumour, but separating tissues by κκ alone requires 6-74 times more decays than are available.

Introduction

Positronium imaging extends positron emission tomography (PET) from where a tracer accumulates to the environment in which positrons annihilate (Bass et al 2023, Moskal et al 2024, 2021). Three observables are under study: the mean ortho-positronium (o-Ps) lifetime, the ratio of three- to two-photon annihilations (Fujimoto et al 2026) and the polarisation correlation of the two annihilation photons (Moskal 2026, Moskal et al 2025, Watts et al 2021). The last, measured by the azimuthal correlation of two Compton scatterings [25] (Snyder et al 1948), has been interpreted as a degree of quantum entanglement. In a porous polymer the measured correlation was smaller than expected for the maximally entangled state but larger than for separable photons, which was explained by assuming that photons from o-Ps pick-off annihilation are not entangled (Moskal et al 2025). Because the pick-off fraction depends on free-volume size and composition, the correlation has been proposed as a biomarker of tissue type, and the first in vivo images of this quantity in a human subject have been reported [23]; a roadmap towards quantum-entanglement PET has been outlined [24].

Three questions remain open. The first is what the measurement determines. A Compton polarimeter is insensitive to circular polarisation, and separable states can reproduce the double-Compton cross section of the entangled state (Clarke et al 2026, Hiesmayr and Moskal 2019). Single-scattering observables built on the Pryce–Ward cross section therefore admit local hidden-variable descriptions and cannot certify entanglement on their own (Clarke et al 2026); lower bounds on Bell-state fidelities that can be extracted from coarse-grained coincidences were derived only recently (Sutter and Hiesmayr 2026), and it has been argued that the full double-Compton distribution of the maximally entangled state is reproduced only by that state (Joos and Kling 2026). The second concerns what happens to the correlation when a photon scatters before detection: experiments disagree (Bordes et al 2024, Ivashkin et al 2023, Tkachev et al 2025, Watts et al 2021), and theoretical treatments of the evolution of the two-photon state are still being developed (AskariPour Ravari and Riazi 2026, Caradonna and Shimazoe 2025, Caradonna et al 2024). The third concerns the pick-off channel: the interpretation of the J-PET result assumes that pick-off photons are separable [23] (Moskal et al 2025), but this has not been derived from the quantum electrodynamics of annihilation with bound molecular electrons and remains to be verified.

Underlying all three is a more practical question that has received little attention: how many events are needed. In X-ray astronomy the statistics of single-photon polarimetry are well developed, through the minimum detectable polarisation and per-event Stokes-parameter estimators (Kislat et al 2015, Weisskopf et al 2010). For two-photon correlations the corresponding analysis has not been done. The most recent proposal for a Bell test with annihilation photons gives an order-of-magnitude rate estimate and defers the required number of events to a dedicated feasibility study (Clarke et al 2026). Counts required to discriminate between two prescribed modulation levels, with finite acceptance, resolution and uniform background, have been derived within an idealised framework for coincidence-conditioned single-arm modulation (Caradonna et al 2026). An estimation-theoretic treatment of the correlation coefficient itself, referred to the quantum limit and to realistic detectors, has not been given. For imaging, the question is sharper still, because tissue contrasts in the correlation coefficient are expected to be of order 10−210^{-2}.

In a related study that is under review (Zhou et al 2026), we combined positronium lifetime, the three-to-two-photon ratio and the polarisation correlation in a source model and found that lifetime plus correlation is the combination that separates microstructure from oxygenation. That work assumed a constant analysing power and a usable fraction of double-Compton events. Here we derive these quantities from first principles and detector simulation and ask: (i) how much information a double-Compton event carries about the correlation coefficient, compared with the quantum limit; (ii) which estimator attains it; (iii) how the observed coefficient depends on annihilation channels, microstructure and oxygen, and on the unresolved pick-off state; (iv) how many usable events a pixelated CZT system delivers per decay; (v) whether in-body scattering biases the measurement; and (vi) whether tissue-level contrasts are reachable in a small-animal setting with 18F^{18}\mathrm{F}. We test the model against published measurements before using it for these predictions.

Methods

Measurement model

We consider two back-to-back 511-keV photons, each undergoing a Compton scattering through polar angle θi\theta_i and azimuth φi\varphi_i, with Δφ\Delta\varphi the angle between the two scattering planes. A single Compton scattering is described by a positive-operator-valued measure parameterised by the analysing power (Clarke et al 2026)

β(θ)=sin⁡2θε+ε−1−sin⁡2θ,ε=E′E=11+(E/mec2)(1−cos⁡θ)(1)\beta(\theta)=\frac{\sin^2\theta}{\varepsilon+\varepsilon^{-1}-\sin^2\theta}, \qquad\varepsilon=\frac{E'}{E}=\frac{1}{1+(E/m_ec^2)(1-\cos\theta)} \tag*{(1)}

which reaches its maximum at θopt≈82∘\theta_{\mathrm{opt}}\approx82^\circ for E=mec2E=m_ec^2. Let F(θ)=ε2(ε+ε−1−sin⁡2θ)F(\theta)=\varepsilon^2(\varepsilon+\varepsilon^{-1}-\sin^2\theta) be the unpolarised Klein–Nishina cross section (constants omitted). For a source without net polarisation, invariant under rotation about the line of flight and under parity, the joint distribution is

p(θ1,θ2,Δφ)∝F(θ1)F(θ2)[1−κβ(θ1)β(θ2)cos⁡2Δφ],(2)p(\theta_1,\theta_2,\Delta\varphi)\propto F(\theta_1)F(\theta_2)[1-\kappa\beta(\theta_1)\beta(\theta_2)\cos2\Delta\varphi], \tag*{(2)}

which defines the linear-polarisation correlation coefficient κ\kappa. For the maximally entangled annihilation state κ=1\kappa=1; for a separable mixture of orthogonally polarised photons with uniformly distributed orientation κ=1/2\kappa=1/2; for uncorrelated photons κ=0\kappa=0. The widely used ratio R=p(Δφ=90∘)/p(Δφ=0∘)=(1+a)/(1−a)R = p(\Delta\varphi= 90^\circ)/p(\Delta\varphi= 0^\circ) = (1+a)/(1-a) with a=κβ1β2a = \kappa\beta_1\beta_2 follows directly. A uniform background fraction bb in Δφ\Delta\varphi multiplies β1β2\beta_1\beta_2 by 1−b1-b. Because the marginal distribution of (θ1,θ2)(\theta_1,\theta_2) in (2) does not depend on κ\kappa, all information on κ\kappa resides in Δφ\Delta\varphi.

Equation (2) is the general form under the stated symmetries: the measurement determines one number, the common linear-polarisation correlation. Appendix § 12 shows that, for the Bell-diagonal states compatible with these symmetries, κ\kappa constrains the concurrence CC only to the interval [max⁡(0,2κ−1),κ][\max(0,2\kappa-1),\kappa], because the circular correlation is not measured. Throughout this paper we therefore treat κ\kappa as the measured quantity and refer to entanglement only where this interval is invoked.

Fisher information and quantum limit

For fixed θ1,θ2\theta_1,\theta_2, the conditional density of Δφ\Delta\varphi is (1−acos⁡2Δφ)/2π(1-a\cos2\Delta\varphi)/2\pi with a=κβ1β2a=\kappa\beta_1\beta_2. Its Fisher information about κ\kappa is (Appendix § 13)

Iκ(θ1,θ2)=(β1β2)21a2(11−a2−1)→a→0(β1β2)22.(3)I_\kappa(\theta_1,\theta_2)=(\beta_1\beta_2)^2\frac{1}{a^2}\left(\frac{1}{\sqrt{1-a^2}}-1\right)\xrightarrow{a\to0}\frac{(\beta_1\beta_2)^2}{2}. \tag*{(3)}

The information per double-Compton event is the average of (3) over the scattering-angle distribution F(θ1)sin⁡θ1F(θ2)sin⁡θ2F(\theta_1)\sin\theta_1 F(\theta_2)\sin\theta_2, optionally restricted to an angular window. With NN independent events, the Cramér–Rao bound is σ(κ)≥(NIˉκ)−1/2\sigma(\kappa)\ge(N\bar{I}_\kappa)^{-1/2}.

To assess how much of the available information Compton polarimetry extracts, we computed the quantum Fisher information (QFI) for two one-parameter families of Bell-diagonal states. In family A, the maximally entangled state is mixed with the separable state described above, so that κ∈[1/2,1]\kappa\in[1/2,1]; in family B, it is mixed with white noise, so that κ∈[0,1]\kappa\in[0,1]. In both families the eigenbasis does not depend on the parameter, and the QFI equals the classical Fisher information of the Bell-basis weights:

FQ(A)=1κ(1−κ),FQ(B)=94(1+3κ)+34(1−κ).(4)F_Q^{(A)}=\frac{1}{\kappa(1-\kappa)},\qquad F_Q^{(B)}=\frac{9}{4(1+3\kappa)}+\frac{3}{4(1-\kappa)}. \tag*{(4)}

As an intermediate reference we also evaluated (3) with β1=β2=1\beta_1=\beta_2=1, which corresponds to an ideal linear-polarisation measurement with a continuous azimuth.

Estimators and toy Monte Carlo

We compared three estimators: event-wise maximum likelihood (ML) based on (2); a moment estimator weighted by the analysing power, κ^=−2∑jβ1jβ2jcos⁡2Δφj/∑j(β1jβ2j)2\hat{\kappa}=-2\sum_j\beta_{1j}\beta_{2j}\cos2\Delta\varphi_j/\sum_j(\beta_{1j}\beta_{2j})^2, which is the two-photon analogue of the per-event Stokes estimators used in X-ray polarimetry ((Kislat et al 2015)); and the ratio RR formed from counts within ±22.5∘\pm22.5^\circ of Δφ=0∘\Delta\varphi=0^\circ and 90∘90^\circ and converted to κ\kappa with the mean β1β2\beta_1\beta_2 of the accepted events. Events were drawn from (2) for κ=0.5\kappa=0.5, $0.75 and 0.95, with 2×1042 \times10^{4} events per experiment and 300–400 repetitions, using scattering-angle windows of 60–110° and 20–160°.

Annihilation channels and tissue model

Photons from direct annihilation, para-positronium (p-Ps) self-annihilation and O2_2-induced conversion of o-Ps to p-Ps were taken to be maximally entangled (κ=1\kappa= 1). Pick-off photons were assigned an unknown coefficient κpick\kappa_{\mathrm{pick}}. The observed coefficient of the two-photon sample is then

κobs=1−α(1−κpick),(5)\kappa_{\mathrm{obs}} = 1 - \alpha\left(1 - \kappa_{\mathrm{pick}}\right), \tag*{(5)}

where α\alpha is the pick-off share of all two-photon annihilations. This is identical to the entanglement observable of Zhou et al (2026), in which the partial-entanglement parameter η\eta corresponds to κpick=(1+η)/2\kappa_{\mathrm{pick}} = (1 + \eta)/2. We evaluated three hypotheses: H0_0, κpick=1/2\kappa_{\mathrm{pick}} = 1/2 (the working assumption of Moskal et al (2025)); Hη_\eta, κpick=3/4\kappa_{\mathrm{pick}} = 3/4; and H1_1, κpick=1\kappa_{\mathrm{pick}} = 1.

The channel shares were computed with the source model of Zhou et al (2026), an absorbing Markov chain in which o-Ps and p-Ps interconvert through O2_2 until annihilation, with parameters derived from published tissue positron-lifetime data (Avachat et al 2024, Moskal 2026, Moskal et al 2021, 2023, Stepanov et al 2020). Microstructure was varied through a shift of the Tao–Eldrup free-volume radius ([30] (Eldrup et al 1981)) and through the lipid fraction, at pO2_2 = 30 mmHg. To test how the pick-off hypothesis propagates to oxygen estimation, we recomputed the lower bounds (floors) on σ\sigma(pO2_2) when four microstructure nuisance parameters are free within priors set by the spread between measured tissues (Zhou et al (2026)).

CZT detector Monte Carlo

We modelled a commercial pixelated CZT Compton-camera module (DT-PA05): CdZnTe (Cd0.9_{0.9}Zn0.1_{0.1}Te, 5.78 g cm−3^{-3}), 44×44×1544 \times44 \times15 mm3^{3}, 22×2222 \times22 pixels of 2 mm, sub-pixel lateral resolution 0.5 mm and depth resolution 1 mm (both treated as FWHM), and 1% energy resolution at 662 keV. Compton attenuation was computed from the Klein–Nishina cross section with Z/A=0.4173Z/A = 0.4173; photoabsorption was parameterised as τ/ρ=0.0136(511 keV/E)2.8\tau/\rho= 0.0136(511\ \mathrm{keV}/E)^{2.8} cm2^{2} g−1^{-1}, which gives a total μ/ρ=0.0856\mu/\rho= 0.0856 cm2^{2} g−1^{-1} at 511 keV. Coherent scattering, fluorescence escape and charge sharing were neglected. Transport within the crystal was unpolarised; the effect of the polarisation correlation on escape probabilities is of second order.

A photon was counted as a usable polarimetric event if (i) its full energy was deposited, (ii) its first interaction was Compton scattering, (iii) its interactions formed exactly two pixel clusters, the first containing only the first interaction, and (iv) both clusters exceeded an energy threshold EthE_{\mathrm{th}} (10, 15 or 30 keV). The scattering angle was reconstructed from the energy of the first cluster, and the azimuthal resolution was propagated as a modulation factor exp⁡(−2σφ2)\exp(-2\sigma_{\varphi}^{2}) with σφ=2 σ⊥/L⊥\sigma_{\varphi} = \sqrt{2}\,\sigma_{\perp}/L_{\perp}, where L⊥L_{\perp} is the transverse cluster separation. Events with three or more clusters, which would require sequence reconstruction, were counted separately as an upper bound. This idealised model, which assumes correct cluster ordering, was used for the ring geometry and the pre-scattering study.

We simulated a single module, then two rings with a point source at the centre: G8, eight modules in an octagon with crystal front faces at R=10.5R = 10.5 cm, which reflects the 83.5-mm housing; and G16, sixteen crystals of the same size forming a gapless polygon (R=11.06R = 11.06 cm), which represents tighter packing. Photons could cross between modules. The per-decay Fisher information was the probability that both photons gave usable events multiplied by the mean of (3) over these pairs, with b=0.05b = 0.05 and a time-averaged live fraction of 0.596 for double coincidences at 10 MBq, as in Zhou et al (2026). Random coincidences between usable single events in opposite modules were estimated for a 24-ns window at 10 MBq.

Detector realism

A second single-module simulation removed the main idealisations. (i) Bound-electron Doppler broadening was included in the impulse approximation: the scattered energy was shifted by δE′=pz∣q∣/[137(1+E(1−cos⁡θ)/mec2)]\delta E' = p_z |\mathbf{q}|/[137(1 + E(1 - \cos\theta)/m_e c^2)], with ∣q∣|\mathbf{q}| the momentum transfer and the electron momentum projection pzp_z drawn from a Gaussian of standard deviation 1.5 atomic units (1 and 2 a.u. as variants). (ii) Positions were the energy-weighted centroids of the interactions within each 2-mm pixel, smeared by the sub-pixel lateral (0.5 mm FWHM) and depth (1 mm FWHM) resolutions; energies were smeared by the energy resolution and a full-energy window of ±2σ\pm2\sigma was applied. (iii) The order of two clusters was not assumed. Two rules were compared: the larger deposit is the first scattering ((Watts et al 2021)), and, using the incident direction known from the line of response, the order whose geometric scattering angle agrees best with the angle computed from the energies. (iv) Events with three or four clusters were reconstructed by testing all orderings and choosing the one that minimised the summed squared difference between geometric and kinematic scattering cosines along the sequence.

For each accepted event we recorded the true analysing power β(θtrue)\beta(\theta_{\rm true}) of the first scattering, the reconstructed angle θrec\theta_{\rm rec} and the error ε\varepsilon of the reconstructed scattering-plane axis, which includes ordering errors. The modulation retained by reconstructed events is βcos⁡2ε\beta\cos2\varepsilon. Because a real analysis can use only θrec\theta_{\rm rec}, the effective analysing power was calibrated in 18 bins of θrec\theta_{\rm rec}, βeff(θrec)=⟨β(θtrue)cos⁡2ε⟩\beta_{\rm eff}(\theta_{\rm rec}) = \langle\beta(\theta_{\rm true})\cos2\varepsilon\rangle, as would be done with a calibration source; the information per pair was then the average of (3) with βeff\beta_{\rm eff} over random pairs of accepted events (κ=0.926\kappa= 0.926, b=0.05b = 0.05). The ratio of the information per pair of interacting photons to that of the idealised model, taken from the independent simulation of § 2.7, was used to rescale the ring results. One parameter at a time was varied to obtain the sensitivity to the photoabsorption cross section (±10%\pm10\%), pixel size (1 and 3 mm), crystal thickness (10 and 20 mm), threshold (10 and 30 keV), energy resolution (2%) and Doppler width.

The total attenuation coefficient of the model was compared with the tabulated values for Cd0.9Zn0.1Te\mathrm{Cd}_{0.9}\mathrm{Zn}_{0.1}\mathrm{Te} obtained from the elemental tables of Hubbell and Seltzer (2004) by mass-weighted addition.

Independent Monte Carlo

The detector model and the correlation model were checked with GATE 10 (Sarrut et al 2022), which is based on Geant4 (Allison et al 2016), using the Livermore polarised electromagnetic physics, which includes Doppler broadening, fluorescence and electron transport. Two simulations were run. In the first, 511-keV photons from a point 10.5 cm in front of a DT-PA05 module were tracked; the deposits were processed with the same clustering, ordering and calibration as above. In the second, 2×1062 \times10^{6} positrons annihilated at rest in a 1-mm water sphere between two 30×30×1.5 cm330 \times30 \times1.5\ \mathrm{cm}^{3} CZT slabs at ±10.5\pm10.5 cm, with Geant4’s entangled annihilation-photon option enabled or disabled; κ\kappa was fitted by maximum likelihood with (2) to the true first-Compton angles and scattering planes of both photons.

Comparison with published measurements

Three published measurements were reproduced with the model. (i) Watts et al (2021) measured the modulation ratio RR with two CZT polarimeters (1×1 cm21 \times1\ \mathrm{cm}^{2} detectors with sub-millimetre pixels, 4.35 cm from the source), selecting events with two pixels and assigning the first scattering to the larger deposit; they reported R=1.85±0.04R = 1.85 \pm0.04 for scattering angles of 7070–110∘110^{\circ} and 1.95±0.071.95 \pm0.07 for 9393–103∘103^{\circ}. We simulated their geometry with the transport model of § 2.5, reconstructed angles and azimuths from pixel centres with their ordering rule, and computed RR for κ=1\kappa= 1 and 1/21/2. (ii) Ivashkin et al (2023) used scintillator scatterers with 16 azimuthal counters of 22.5∘22.5^{\circ} and scattering angles of 8080–100∘100^{\circ}, reporting R=2.435±0.018R = 2.435 \pm0.018; we folded (2) with their angular acceptance. (iii) Moskal et al (2025) measured RR for photons from the porous polymer XAD-4 at 82∘82^{\circ} and 94∘94^{\circ}, with circular angular regions of 20∘20^{\circ} radius, and quoted ideal and detector-simulated ratios for the maximally entangled and separable cases. We used their simulated ratios as a scale, checked that our model reproduces their separable reference, and inverted their measured ratios to κobs\kappa_{\mathrm{obs}} and, through (5) with their pick-off share, to κpick\kappa_{\mathrm{pick}}.

In-body pre-scattering

A photon that scatters by a small angle in the body may remain inside the photopeak window, and its correlation after scattering is disputed [23] (Ivashkin et al 2023; Tkachev et al 2025; Watts et al 2021). We therefore treated the post-scattering coefficient κs\kappa_{\mathrm{s}} as unknown in [0,κ][0,\kappa], so that the bias is bounded by fsκf_{\mathrm{s}}\kappa, where fsf_{\mathrm{s}} is the fraction of accepted pairs with at least one pre-scattered photon. The mouse was modelled as an infinite water cylinder of radius 1.0 or 1.5 cm with the source on the axis or 0.5–1.0 cm off axis. Only Compton scattering in water was tracked, and photons were accepted if they reached a module of G8. The photopeak window was ±2σ\pm2\sigma of the CZT energy resolution at 511 keV (FWHM 5.8 keV), which admits scattering angles below 8.0∘8.0^\circ; a case without an energy window was simulated for comparison with detectors of low energy resolution. The quantity of interest is the difference in fSκf_S\kappa between source positions, since a depth-dependent bias could mimic a tissue contrast.

Imaging scenarios

We considered 18F{}^{18}\mathrm{F} with 10 MBq injected in a mouse and acquisition until complete decay. Two regions were used: a 10-mm tumour (0.5 mL at 400 kBq mL−1^{-1}, 1.9×1091.9 \times10^{9} decays) and a liver of about 1.5 mL (5.7×1095.7 \times10^{9} decays). The decays required per region were N=1/(Idec δκ2)N = 1/(I_{\mathrm{dec}}\,\delta\kappa^{2}) for a target precision δκ\delta\kappa, and N=50/(IdecΔκ2)N = 50/(I_{\mathrm{dec}}\Delta\kappa^{2}) for a 5σ5\sigma separation of two regions whose coefficients differ by Δκ\Delta\kappa.

Implementation

All calculations used Python with NumPy; the source model of Zhou et al (2026) additionally used SciPy, and the independent simulations used GATE 10.1. Analytical Fisher information was checked against direct numerical integration, and the QFI expressions against numerical differentiation of the Bell weights. Code and numerical results will be released with the preprint.

Results

Benchmarks

The analysing power reached βmax⁡=0.6918\beta_{\max} = 0.6918 at 81.66∘81.66^\circ, in agreement with Clarke et al (2026). With both photons at 82∘82^\circ, (2) gave R=2.835R = 2.835 for κ=1\kappa= 1 and R=1.629R = 1.629 for κ=1/2\kappa= 1/2, reproducing the values usually quoted (2.85 and 1.63) (Ivashkin et al 2023, Moskal et al 2025). Equation (3) agreed with numerical integration to six digits for κ\kappa between 0 and 1, and (4) with numerical differentiation to five digits.

Comparison with published measurements

Table 1 summarises the comparison. For the CZT polarimeter of Watts et al (2021), the model predicted R=1.91R = 1.91 (70–110°) and 1.98 (93–103°) for κ=1\kappa= 1, against measured values of 1.85±0.041.85 \pm0.04 and 1.95±0.071.95 \pm0.07; for κ=1/2\kappa= 1/2 it predicted 1.37 and 1.40. The larger-deposit rule ordered 69% of events correctly, close to the 60% reported, and pixelisation and misordering lowered RR from 2.06 and 2.15. For the scintillator polarimeter of Ivashkin et al (2023), folding with the counter width gave R=2.46R = 2.46, against 2.435±0.0182.435 \pm0.018. Both measurements are thus consistent with κ=1\kappa= 1 for direct annihilation in these sources and with our treatment of angular acceptance.

For the J-PET data ((Moskal et al 2025)), our model reproduced the separable reference from their simulated ratios (κ=0.508\kappa= 0.508 and 0.5120.512 at 82∘82^\circ and 94∘94^\circ). Their measured ratios correspond to κobs=0.796±0.016\kappa_{\mathrm{obs}} = 0.796 \pm0.016 and 0.869±0.0190.869 \pm0.019; with their pick-off share this gives κpick=0.34±0.05\kappa_{\mathrm{pick}} = 0.34 \pm0.05 and 0.58±0.060.58 \pm0.06. Both values exclude κpick=1\kappa_{\mathrm{pick}} = 1, but they differ from each other by about three standard deviations, and the XAD-4 mixture implied by their own simulation at 94∘94^\circ (0.895) differs from that implied by their stated pick-off share (0.845). The inference therefore depends on how the published ratios were corrected and should be taken as indicative.

How much information a double-Compton event carries

Figure 3 shows IκI_\kappa over the (θ1,θ2)(\theta_1,\theta_2) plane. The information peaks where both photons scatter near 82∘82^\circ, at 0.120 for κ=0.5\kappa= 0.5 and 0.139 for κ=1\kappa= 1, so that δκ=0.01\delta\kappa= 0.01 would need 77–8×1048 \times10^{4} such events. The Klein–Nishina distribution, however, places most scatterings at forward angles where β\beta is small. Averaged over all angles, a double-Compton event carries 0.0146–0.0157, and δκ=0.01\delta\kappa= 0.01 requires 6.46.4–6.9×1056.9 \times10^{5} events, about nine times more than the optimal-angle estimate.

Figure 3. Fisher information about κ\kappa per double-Compton event as a function of the two scattering angles, for (a) κ=0.5\kappa=0.5 and (b) κ=0.926\kappa=0.926. White contours: relative density of double-Compton events (25, 50, 75% of maximum). Cross: optimal angle 81.7°.

Figure 4 places these numbers on the scale of the quantum limit. For family A at κ=0.5\kappa= 0.5 the QFI is 4.0; an ideal linear-polarisation measurement extracts 0.62 (15.5%); Compton polarimetry at the optimal angles extracts 0.12 (3.0%); and Klein–Nishina-averaged Compton polarimetry extracts 0.015 (0.37%). For family B the corresponding fractions at κ=0.5\kappa= 0.5 are 25.8%, 5.0% and 0.61%. The first step reflects the blindness to circular polarisation and the restriction to a single azimuthal outcome; the second reflects β<0.7\beta< 0.7; the third reflects the scattering-angle distribution. As κ→1\kappa\to1 the QFI of family A diverges, whereas the Compton information stays finite, so the efficiency falls further (0.14% at κ=0.9\kappa= 0.9).

Figure 4. Information about κ\kappa per photon pair: quantum Fisher information, ideal linear-polarisation measurement, Compton polarimetry at the optimal angle and Compton polarimetry averaged over the Klein–Nishina distribution, for (a) family A and (b) family B.

Angular selection and estimators

Selecting a window of scattering angles raises the information per accepted event but discards events, and the product always decreased (Figure 5(a)). Relative to using all angles, the windows 6060–110∘110^\circ, 7070–95∘95^\circ and 7575–90∘90^\circ required 1.9, 5.4 and 14 times as many double-Compton events. A window of 3030–150∘150^\circ lost only 1%.

Figure 5. (a) Double-Compton events needed for a given precision when a scattering-angle window is imposed, relative to using all angles (κ=0.75\kappa=0.75). (b) Efficiency of three estimators (Cramér–Rao bound divided by Monte Carlo variance) with a 20–160° window and 2×1042 \times10^{4} events.

Event-wise ML and the β\beta-weighted moment estimator were unbiased and reached the Cramér–Rao bound within Monte Carlo error at all three values of κ\kappa (efficiency 0.96–1.03; Figure 5(b)). The ratio RR had an efficiency of 0.43–0.45 with the 2020–160∘160^\circ window, that is it needed 2.2–2.3 times as many events. With the narrower 6060–110∘110^\circ window the penalty was smaller (variance ratio 1.24) because β\beta varies less within the window.

Annihilation channels, microstructure and the pick-off hypothesis

Across the tissue parameter set, the pick-off share of two-photon annihilations α\alpha ranged from 0.13 (porcine liver) to 0.29 (human adipose, (Moskal et al. 2023)), giving κobs\kappa_{\mathrm{obs}} between 0.854 and 0.934 under H0H_0 (Table 2). Under HηH_\eta the coefficients moved halfway towards 1, and under H1H_1 they were all exactly 1.

Tissue (source of lifetime data)α\alphaH0 (κpick=1/2\kappa_{\mathrm{pick}} = 1/2)Hη\eta (κpick=3/4\kappa_{\mathrm{pick}} = 3/4)H1 (κpick=1\kappa_{\mathrm{pick}} = 1)
Human adipose (Moskal et al (2021))0.2170.8910.9461
Human cardiac myxoma (Moskal et al (2021))0.2280.8860.9431
Human adipose (Moskal et al (2023))0.2920.8540.9271
Human cardiac myxoma (Moskal et al (2023))0.1890.9050.9531
Porcine adipose (Avachat et al (2024))0.1870.9070.9531
Porcine liver (Avachat et al (2024))0.1320.9340.9671
Porcine muscle (Avachat et al (2024))0.1480.9260.9631

Table 2. Pick-off share of two-photon annihilations α\alpha and observed correlation coefficient κobs\kappa_{\mathrm{obs}} at pO2=30\mathrm{pO}_2 = 30 mmHg under three hypotheses for the pick-off coefficient.

The coefficient was sensitive to composition but almost insensitive to oxygen (Figure 2). In muscle under H0H_0, raising pO2\mathrm{pO}_2 from 20 to 40 mmHg changed κobs\kappa_{\mathrm{obs}} by 1.0×10−41.0 \times10^{-4}, a free-volume radius shift of ±0.01\pm0.01 nm by 4.0×10−44.0 \times10^{-4}, and a lipid fraction of 0.1 by −3.6×10−3-3.6 \times10^{-3}. Contrasts between tissue pairs were Δκ=0.0193\Delta\kappa= 0.0193 (muscle vs porcine adipose), 0.0080 (muscle vs liver) and 0.0055 (cardiac myxoma vs adipose) under H0H_0, and half of these under HηH_\eta.

Figure 2. Observed correlation coefficient as a function of (a) pO2, (b) free-volume radius shift and (c) lipid fraction, for three pick-off hypotheses. Muscle base tissue in (b) and (c), pO2 = 30 mmHg.

Table 3 shows how the hypothesis propagates to oxygen estimation. In the analysis of Zhou et al. (2026), adding the correlation to lifetime events lowers the floor on σ(pO2)\sigma(\mathrm{pO}_2) from 2214 to 100 mmHg in muscle and from 339 to 14 mmHg in adipose. These floors were identical under H0H_0 and HηH_\eta, because the floor depends on how the observables vary with the parameters and not on the size of that variation; the number of decays needed to approach the floor scales as (1−κpick)−2(1-\kappa_{\mathrm{pick}})^{-2}. Under H1H_1 the correlation coefficient carries no information and the floor reverts to that of lifetime alone.

HypothesisMuscleAdipose
LL+CL+R+CLL+CL+R+C
H022141008.8339142.7
Hη\eta22141008.8339142.7
H1221422141661339339275

Table 3. Floors on σ(pO2)\sigma(\mathrm{pO}_2) (mmHg; 101610^{16} decays, microstructure nuisances with inter-tissue priors, long-axial-field-of-view scenario of Zhou et al (2026)) under the three pick-off hypotheses. L: lifetime events; C: double-Compton events; R: three-photon events.

Usable events in CZT

In a single 15-mm module (Table 4), 47.0% of incident 511-keV photons interacted; of these, 50.4% deposited their full energy and 84.1% interacted first by Compton scattering. Only 17.3% of interacting photons produced a clean two-cluster, full-energy event, and 6.6% did so within the 82±20∘82 \pm20^\circ window commonly used for RR. The main loss is lateral escape: a photon scattered near 82∘82^\circ carries about 250 keV sideways and must be absorbed within the same 44-mm crystal. Thicknesses of 10 and 20 mm gave windowed fractions of 6.1% and 6.8%, and thresholds between 10 and 30 keV changed the fractions by less than 0.002. Clean events had an approximately symmetric distribution of reconstructed angles centred near 90∘90^\circ (Figure 6(a)), unlike the forward-peaked distribution of first scatterings. Within the window, the effective analysing power ⟨β1β2⟩\langle\beta_1\beta_2\rangle was 0.391, and the 2-mm pixels reduced the modulation by less than 1% (median factor 0.994). A further 8.5% of interacting photons produced three or more clusters with full energy.

Table 4: Usable polarimetric events in DT-PA05 CZT (Eth=15E_{\mathrm{th}} = 15 keV). Single-module fractions are relative to photons that interact in the crystal; ring quantities are per decay. Information per decay includes a live fraction of 0.596 and b=0.05b = 0.05.

QuantitySingle moduleG8 ringG16 gapless ring
Interaction probability (photon / both photons)0.4702.56×10−22.56 \times10^{-2}4.64×10−24.64 \times10^{-2}
Clean two-cluster fraction, all angles0.1730.180–
Clean two-cluster fraction, 82±20∘82 \pm20^\circ0.0660.067–
Usable pairs per decay, all angles–8.7×10−48.7 \times10^{-4}1.8×10−31.8 \times10^{-3}
Cross-module share of usable events–1.7%5.1%
Information per usable pair, all angles0.0230.0220.024
Fisher information per decay, all angles (ML)–1.15×10−51.15 \times10^{-5}2.55×10−52.55 \times10^{-5}
Fisher information per decay, 82±20∘82 \pm20^\circ–5.8×10−65.8 \times10^{-6}–
Random coincidences at 10 MBq–0.6%0.5%

Table 4.

Figure 6. (a) Reconstructed scattering angle of clean two-cluster, full-energy events in a DT-PA05 module, with the Klein–Nishina distribution of first scatterings (peak-scaled) and the 82±20∘82 \pm20^\circ window. (b) Fisher information about κ\kappa per decay (live time included) assumed previously and obtained here with the idealised model and with the GATE-corrected response using two clusters only or including sequence reconstruction of events with three or four clusters.

In the eight-module ring, the probability that both photons interacted was 2.56×10−22.56 \times10^{-2} per decay; usable pairs occurred at 8.7×10−48.7 \times10^{-4} per decay over all angles and 1.3×10−41.3 \times10^{-4} within the window. Only 1.7% of usable events involved two modules, because the housings leave wide gaps between crystals. The Fisher information per decay including live time was 1.15×10−51.15 \times10^{-5} with all angles and ML, and 5.8×10−65.8 \times10^{-6} with the window (Figure 6(b)). In the gapless 16-module ring, both photons interacted with probability 4.6×10−24.6 \times10^{-2}, 5.1% of usable events crossed modules, and the information per decay rose to 2.55×10−52.55 \times10^{-5}. Random coincidences contributed 0.5–0.6% of usable pairs at 10 MBq in both geometries.

Zhou et al (2026) assumed a usable single-photon fraction of 0.22 (range 0.15–0.30) within the window. The simulation gives 0.066, which reduces the information per decay by a factor of 11 with windowed analysis and of 5.4 with all angles and ML. Its per-event information (effective analysing power 0.41) was consistent with the Klein–Nishina average within the window (0.415).

The realistic model changed these numbers in two opposite directions (table 5). With positions from sub-pixel centroids and the kinematic ordering rule, 19% of interacting photons gave usable two-cluster events, and the order was correct for 70% of them. Misordering is intrinsic near 90°, where both deposits are close to 256 keV, and it reverses the scattering-plane axis only partly; together with position errors it reduced the mean cos⁡2ε\cos2\varepsilon to 0.84, and 14% of events had axis errors above 20°. After calibration in θrec\theta_{\mathrm{rec}}, the information per usable pair was 0.013, against 0.023 in the idealised model, so that two-cluster events alone gave 0.68 times the idealised information per pair of interacting photons. Doppler broadening of 1–2 a.u. had no measurable effect (relative change below 3%), because the angular resolution was dominated by the lever arm between clusters. Ordering by the larger deposit lowered the information by a further 19%. Including events with three or four clusters raised the usable fraction to 27%; their first pair was identified correctly in 87% of cases, and the information per pair of interacting photons rose to 1.26 times the idealised two-cluster value.

Table 6 lists the sensitivity. Crystal thickness had the largest effect (0.75 and 1.21 for 10 and 20 mm), followed by pixel size (1.12 and 0.87 for 1 and 3 mm) and threshold (1.03 and 0.90 for 10 and 30 keV); a ±10% change in photoabsorption changed the information by −4% and +6%, and 2% energy resolution by +2%. The model’s total attenuation coefficient was 6% below the tabulated value at 511 keV and 5–13% below it between 100 and 600 keV, mainly because coherent scattering was omitted; raising photoabsorption by 25%, more than this deficit, increased the information per pair of interacting photons by 10%, so the estimates are conservative by about this amount.

The independent GATE simulation supported both models. For 2.3×1052.3 \times10^{5} annihilation pairs in which both photons first scattered by Compton scattering, the fitted coefficient was κ=1.039±0.016\kappa= 1.039 \pm0.016 with entanglement enabled and 0.536±0.0170.536 \pm0.017 with it disabled, close to the values 1 and 12\frac{1}{2} of the maximally entangled and the separable orthogonally polarised states; the common excess of about 4% indicates a slightly larger azimuthal modulation in the Livermore polarised Compton model than in the free-electron Klein–Nishina analysing power used here. For 5×1055 \times10^{5} photons incident on a module, GATE gave the same usable fraction as our model (0.266 versus 0.265 with sequence reconstruction; 0.181 versus 0.188 for two clusters), but lower correct-ordering rates (0.65 versus 0.70 for two clusters; 0.74 versus 0.87 for the first pair of multi-cluster events) and a lower information per usable pair (0.0100 versus 0.0126). We attribute the difference to electron ranges, fluorescence and bremsstrahlung, which displace cluster centroids and add spurious clusters. Relative to the idealised model, the information per pair of interacting photons in GATE was 1.03 with sequence reconstruction and 0.58 with two clusters only; we used these more conservative factors for the ring results.

In-body pre-scattering

Without an energy window, 3–6% of accepted pairs contained a photon scattered in the water cylinder (Table 7). The CZT photopeak window reduced this to 0.23–0.40%. The worst-case bias fsκf_s\kappa was then at most 0.0037, and its variation between source positions at most 0.0004, below 10% of the smallest tissue contrast considered. Without an energy window, the worst-case bias reached 0.052 and varied by up to 0.013 between source positions in the 1.5-cm cylinder, which exceeds the muscle–liver contrast of 0.008.

Radius / offset (cm)CZT photopeak window (±2σ\pm2\sigma)No energy window
fsf_sfsκf_s\kappadifferencefsf_sfsκf_s\kappadifference
1.0 / 00.00230.0022–0.0270.025–
1.0 / 0.50.00250.00230.00010.0300.0280.003
1.5 / 00.00360.0034–0.0420.039–
1.5 / 0.50.00400.00370.00040.0490.0460.007
1.5 / 1.00.00380.00350.00020.0560.0520.013

Table 7. In-body pre-scattering in a water cylinder. fsf_s: fraction of accepted pairs with at least one pre-scattered photon. Differences are relative to an on-axis source in the same cylinder.

Feasibility in a small-animal setting

Figure 7 and Table 8 combine these results, using the GATE-corrected detector response with sequence reconstruction. With G8 the Fisher information per decay was 1.18×10−51.18 \times10^{-5}. Then δκ=0.01\delta\kappa= 0.01 required 8.5×1088.5 \times10^{8} decays per region, about half of the decays available in a 10-mm tumour; δκ=0.005\delta\kappa= 0.005 required 3.4×1093.4 \times10^{9}, 1.8 times the available decays. Separating muscle from adipose at 5σ5\sigma under H0H_0 needed 1.1×10101.1 \times10^{10} decays per region (6.0 times the tumour’s decays, 2.0 times the liver’s), muscle from liver 6.6×10106.6 \times10^{10} (35 and 12 times), and myxoma from adipose 1.4×10111.4 \times10^{11} (74 and 25 times). With two-cluster events only, all requirements increased by a factor of 1.8. The gapless ring reduced all requirements by a factor of 2.2; under HηH_\eta they increased by a factor of 4.

TargetGeometryDecays per regionTumour ratioLiver ratio
δκ=0.01\delta\kappa= 0.01G88.5×1088.5 \times10^{8} (1.5×1091.5 \times10^{9})0.440.15
G163.8×1083.8 \times10^{8} (6.8×1086.8 \times10^{8})0.200.07
δκ=0.005\delta\kappa= 0.005G83.4×1093.4 \times10^{9} (6.0×1096.0 \times10^{9})1.80.59
G161.5×1091.5 \times10^{9} (2.7×1092.7 \times10^{9})0.800.27
Muscle vs adipose (Δκ=0.019\Delta\kappa= 0.019)G81.1×10101.1 \times10^{10} (2.0×10102.0 \times10^{10})6.02.0
G165.1×1095.1 \times10^{9} (9.1×1099.1 \times10^{9})2.70.90
Muscle vs liver (Δκ=0.008\Delta\kappa= 0.008)G86.6×10106.6 \times10^{10} (1.2×10111.2 \times10^{11})3512
G163.0×10103.0 \times10^{10} (5.3×10105.3 \times10^{10})165.2
Myxoma vs adipose (Δκ=0.0055\Delta\kappa= 0.0055)G81.4×10111.4 \times10^{11} (2.5×10112.5 \times10^{11})7425
G166.3×10106.3 \times10^{10} (1.1×10111.1 \times10^{11})3311
Muscle vs liver, Hη\eta (Δκ=0.004\Delta\kappa= 0.004)G82.6×10112.6 \times10^{11} (4.7×10114.7 \times10^{11})13946

Table 8. Decays required per region with 18F^{18}\mathrm{F} (GATE-corrected detector response with sequence reconstruction; in parentheses, two-cluster events only), and their ratio to all decays in a 10-mm tumour (1.9×1091.9 \times10^{9}) and a mouse liver (5.7×1095.7 \times10^{9}). Tissue separations are 5σ5\sigma under H0 unless stated.

Standard deviation of kappa versus decays per region

Figure 7. Standard deviation of κ\kappa versus decays per region for the eight-module and gapless 16-module rings (GATE-corrected detector response with sequence reconstruction; dashed, two clusters only). Dotted lines: precision per region required for a 5σ5\sigma separation of tissue pairs (H0). Dashed lines: all decays of 18F{}^{18}\mathrm{F} in a 10-mm tumour and a mouse liver at 400 kBq mL−1^{-1}.

Discussion

What is measured. Double-Compton polarimetry of annihilation photons measures one number, the linear-polarisation correlation κ\kappa. Its relation to entanglement requires assumptions about the unmeasured circular correlation: for Bell-diagonal states, κ>1/2\kappa> 1/2 implies entanglement and bounds the concurrence below by 2κ−12\kappa- 1, but the same κ\kappa is compatible with concurrences up to κ\kappa (Appendix § 12). This is consistent with recent analyses showing that separable states can reproduce the double-Compton cross section and that single-scattering observables admit local hidden-variable models ((Clarke et al 2026, Hiesmayr and Moskal 2019, Sutter and Hiesmayr 2026)). For imaging, the distinction matters less than it might seem: what carries tissue information is the mixture of channels in (5), and κ\kappa is a well-defined, model-independent observable of that mixture. We therefore suggest reporting κ\kappa (or RR together with the analysing power) and reserving the term entanglement degree for statements supported by the bounds above.

Statistical requirements. The information per double-Compton event is small for reasons that cannot be engineered away: Compton scattering does not resolve circular polarisation, the analysing power at 511 keV does not exceed 0.69, and most scatterings are forward. In the small-modulation limit, (3) gives σ(κ)≈2/(N⟨(β1β2)2⟩)\sigma(\kappa) \approx\sqrt{2/(N\langle(\beta_1\beta_2)^2\rangle)}, the two-photon analogue of the per-event Stokes uncertainty 2/(μ2N)\sqrt{2/(\mu^2N)} of X-ray polarimetry (Kislat et al 2015) with μ2\mu^2 replaced by the mean squared product of analysing powers. Two practical consequences follow. First, scattering-angle windows should not be imposed for estimation; every event should enter the likelihood with its own analysing power. Second, the RR ratio, although convenient for display, should not be the estimator: it needed up to 2.3 times as many events as ML in our simulations, because a single mean analysing power is applied to events with widely different β\beta. The ML and β\beta-weighted estimators are simple to implement and reached the Cramér–Rao bound.

Detector design. The usable fraction in CZT was limited not by energy or position resolution, which were more than adequate, but by containment of the laterally scattered photon. Tighter packing helped: a gapless ring more than doubled the information per decay, mostly through geometric coverage. Once ordering and position errors are included, two further properties matter: sequence reconstruction of events with three or more clusters, which nearly doubled the information per pair of interacting photons, and the use of the known incident direction for ordering, which gained 19% over ordering by deposit size. Doppler broadening and energy resolution between 1% and 2% were unimportant. Thicker crystals (20 mm: +21%), finer pixels (1 mm: +12%) and layered geometries in which the scattered photon is absorbed in a second detector, as in Compton cameras, would add further gains. Table 6 can serve as a list of design requirements. Rates of random coincidences were negligible at 10 MBq, and polarisation-correlated Compton events have been measured to have a higher signal-to-random ratio than photoelectric events at clinical activities (Kožuljević et al 2026).

Table 6: Sensitivity of the information per pair of interacting photons to detector parameters (realistic model with sequence reconstruction; reference: 15 mm, 2-mm pixels, 15 keV, 1% at 662 keV, Doppler 1.5 a.u.).

ParameterValueRelative information
Crystal thickness10 / 20 mm0.75 / 1.21
Pixel size1 / 3 mm1.12 / 0.87
Energy threshold10 / 30 keV1.03 / 0.90
Photoabsorption cross section−10% / +10%0.96 / 1.06
Energy resolution at 662 keV2%1.02
Doppler width1 / 2 a.u.1.00 / 0.97
Cluster orderinglarger deposit instead of kinematic0.81
Sequence reconstructiontwo clusters only0.55

Table 6.

Energy resolution and the decoherence question. Whether a photon keeps its polarisation correlation after scattering in the body is disputed. One view holds that scattering below about 40° does not reduce the correlation [23] (Ivashkin et al 2023); another measurement reported a reduction (Watts et al 2021). With the energy resolution of CZT, pre-scattered photons were reduced to below 0.4% of accepted pairs, and the worst-case bias varied by less than 4×10−44 \times10^{-4} across positions, so tissue contrasts can be measured whichever view is correct. Without energy discrimination, the worst case would vary by more than the muscle–liver contrast across a mouse-sized object, and more in humans. Energy resolution therefore makes κ\kappa imaging robust to an open question in the physics of the measurement.

The pick-off question. (5) makes explicit that κ\kappa carries tissue information only through the pick-off channel, and only to the extent that κpick<1\kappa_{\mathrm{pick}} < 1. The interpretation of the JPET result relies on separable pick-off photons (Moskal et al 2025), an assumption that remains to be verified. Our inversion of their data gave κpick\kappa_{\mathrm{pick}} between 0.34 and 0.58, which disfavours maximally entangled pick-off photons; if confirmed, tissue contrasts would be 0.8–1.3 times those computed under H0\mathrm{H}_0. The inference, however, rests on the published correction from measured to ideal ratios, and the two angles disagree. A first-principles calculation of κpick\kappa_{\mathrm{pick}} for annihilation with bound molecular electrons is therefore the key theoretical input for the whole programme. Our results show what is at stake: the floors on oxygen estimation are unchanged for any κpick<1\kappa_{\mathrm{pick}} < 1, but the required statistics scale as (1−κpick)−2(1-\kappa_{\mathrm{pick}})^{-2}, and the biomarker vanishes at κpick=1\kappa_{\mathrm{pick}} = 1. Measurements in materials with known pick-off fractions and different electronic structure, analysed with the estimators recommended here, could discriminate between these hypotheses.

Feasibility. With 18F{}^{18}\mathrm{F} and an eight-module CZT ring, κ\kappa can be measured to about ±0.01\pm0.01 in a 10-mm tumour, which is sufficient to distinguish separable from maximally entangled pick-off photons in phantoms with large pick-off fractions. It is not sufficient for tissue classification by κ\kappa alone, which needs 6–74 times the available decays. Classification in such systems will rely on lifetime and three-photon observables, as in Zhou et al (2026), with κ\kappa contributing as a complementary observable in combined estimation. For that study the present results leave the identifiability floors unchanged but increase the number of decays required in the CZT scenario: by factors of 1.1–1.3 for myxoma versus adipose, where lifetime and three-photon events dominate, by 3–7 for muscle versus liver, and by 5–11 when κ\kappa is used alone.

Limitations. Transport in CZT was unpolarised and neglected coherent scattering, charge sharing and fluorescence escape; Doppler broadening was approximated by a Gaussian momentum profile. The effects of cluster ordering, sequence reconstruction and electron transport were computed for a single module and transferred to the rings as a common factor, and the effective analysing power was calibrated with the true scattering angles, which in practice requires a calibration measurement or a validated simulation. The comparison with published data used reported ratios rather than raw data. The mouse was a uniform water cylinder, and the source was a point at the centre of the ring for the detector simulation. The channel shares rely on ex vivo positron-lifetime data and on the source model of Zhou et al (2026), including its treatment of oxygen-converted p-Ps. The post-scattering coefficient was bounded rather than modelled. Detector parameters not provided by the manufacturer (timing, count-rate behaviour) were taken from Zhou et al (2026) and must be measured.

Conclusion

Compton polarimetry of annihilation photons extracts less than 1% of the quantum Fisher information about their polarisation correlation, and a double-Compton event carries about 0.015 units of information on κ\kappa. Using all scattering angles with event-wise maximum likelihood is optimal; angular windows and the RR ratio waste up to half the data. The model reproduces published modulation ratios of CZT and scintillator polarimeters and agrees with an independent Geant4 simulation. In a pixelated CZT ring, lateral containment limits the usable events, ordering errors, position errors and electron transport halve the information per event, and sequence reconstruction of multi-site events recovers this loss; energy resolution makes the measurement robust to in-body scattering. With 18F{}^{18}\mathrm{F} in a mouse, κ\kappa can be measured to ±0.01\pm0.01 in centimetre-sized regions, but tissue classification by κ\kappa alone is out of reach. The biomarker value of κ\kappa depends entirely on the polarisation state of pick-off photons; published porous-polymer data suggest κpick<1\kappa_{\mathrm{pick}} < 1, and this should be the next target of theory and experiment.

Acknowledgments

This work was supported by the National Natural Science Foundation of China (Nos. 62005150, U22A6008 and 82027804), the Shanxi Province Higher Education “Billion Project” Science and Technology Guidance Project, the Fundamental Research Program of Shanxi Province (No. 202503021211260), the Natural Science Foundation for Young Scientists of Shanxi Province (No. 202303021222344) and the First Hospital of Shanxi Medical University Hospital Fund Project (No. SYYYRC-2023004).

Data availability statement

The code, simulation scripts and numerical results that support the findings of this study are openly available at Zenodo (DOI to be assigned on acceptance; an anonymised copy is provided to the reviewers).

Author contributions

Haitao Zhou: conceptualization, methodology, software, formal analysis, validation, visualization, funding acquisition, writing – original draft, writing – review and editing. Ruonan Wang: data curation (tissue parameter set), funding acquisition, writing – review and editing. Jianyong Hu: methodology (quantum optics and polarimetry), validation, writing – review and editing. Zhifang Wu: resources, writing – review and editing. Sijin Li: conceptualization, supervision, project administration, funding acquisition, writing – review and editing.

Declaration of generative AI use

During the preparation of this work the authors used Claude (Anthropic) to assist in writing and debugging the Python code for the Fisher-information calculations, the toy Monte Carlo and the detector simulations; to help organise and summarise the literature; and to draft and edit the English text. All assumptions, parameter choices, results and conclusions were designed, checked and verified by the authors, and all cited references were read and verified by the authors. The authors take full responsibility for the content of this article.

Ethics statement

This study is entirely theoretical and computational. No human participants, human samples, animals or patient data were involved; tissue parameters were taken from published measurements.

Conflict of interest

The authors declare no competing interests.

Identifiability: from κ\kappa to concurrence

Write the two-photon polarisation state in the Pauli basis as ρ=14(I+∑iriσi⊗I+∑jsjI⊗σj+∑ijTijσi⊗σj)\rho= \frac{1}{4}(I + \sum_i r_i \sigma_i \otimes I + \sum_j s_j I \otimes\sigma_j + \sum_{ij} T_{ij}\sigma_i \otimes\sigma_j), with σ1,2\sigma_{1,2} the linear Stokes components and σ3\sigma_3 the circular one. We assume no single-photon polarisation (r=s=0r=s=0) and a correlation matrix that is diagonal in a frame in which the annihilation state has T=diag⁡(−1,−1,−1)T=\operatorname{diag}(-1,-1,-1), i.e. a Bell-diagonal state. Rotational invariance about the line of flight requires T11=T22≡−κT_{11}=T_{22}\equiv-\kappa, and the double-Compton distribution (2) depends on TT only through κ\kappa; the circular correlation t3=T33t_3=T_{33} is not measured. The Bell weights are p0=(1+2κ−t3)/4p_0=(1+2\kappa-t_3)/4 for the annihilation state, p1,2=(1+t3)/4p_{1,2}=(1+t_3)/4 and p3=(1−2κ−t3)/4p_3=(1-2\kappa-t_3)/4. Positivity requires −1≤t3≤1−2κ-1\le t_3\le1-2\kappa. The concurrence of a Bell-diagonal state is C=max⁡(0,2pmax⁡−1)C=\max(0,2p_{\max}-1), which over the allowed range of t3t_3 gives

max⁡(0,2κ−1)≤C≤κ,κ≤F0=p0≤1+κ2,(6)\max(0,2\kappa-1)\le C\le\kappa,\qquad\kappa\le F_0=p_0\le\frac{1+\kappa}{2}, \tag*{(6)}

where F0F_0 is the fidelity with the annihilation Bell state (Figure 8). Thus κ>1/2\kappa>1/2 certifies entanglement within this family, and κ=1\kappa=1 fixes the maximally entangled state, in agreement with Joos and Kling (2026); for κ≤1/2\kappa\le1/2 a separable state with the same κ\kappa exists. The family A and B states of § 2.2 correspond to t3=0t_3=0 and t3=−κt_3=-\kappa, respectively.

Figure 8. Concurrence compatible with a measured κ\kappa for Bell-diagonal states (Appendix A).

Fisher information

For p(x)=(1−acos⁡x)/2πp(x) = (1-a\cos x)/2\pi on [0,2π)[0,2\pi) with x=2Δϕx = 2\Delta\phi and a=κβ1β2a = \kappa\beta_1\beta_2, ∂κp=−β1β2cos⁡x/2π\partial_\kappa p = -\beta_1\beta_2\cos x/2\pi and

Iκ=(β1β2)212π∫02πcos⁡2x1−acos⁡x dx=(β1β2)21a2(11−a2−1)(7)I_\kappa= (\beta_1\beta_2)^2 \frac{1}{2\pi}\int_{0}^{2\pi}\frac{\cos^2x}{1-a\cos x}\,dx = (\beta_1\beta_2)^2\frac{1}{a^2}\left(\frac{1}{\sqrt{1-a^2}}-1\right) \tag*{(7)}

using 12π∫02π(1−acos⁡x)−1dx=(1−a2)−1/2\frac{1}{2\pi}\int_{0}^{2\pi}(1-a\cos x)^{-1}dx = (1-a^2)^{-1/2}. For the quantum Fisher information, the states of families A and B are diagonal in the Bell basis with parameter-independent eigenvectors, so FQ=∑k(∂κpk)2/pkF_Q = \sum_k(\partial_\kappa p_k)^2/p_k. Family A has weights (κ,(1−κ)/2,(1−κ)/2,0)(\kappa,(1-\kappa)/2,(1-\kappa)/2,0), giving FQ=1/[κ(1−κ)]F_Q = 1/[\kappa(1-\kappa)]; family B has ((1+3κ)/4,(1−κ)/4,(1−κ)/4,(1−κ)/4)((1+3\kappa)/4,(1-\kappa)/4,(1-\kappa)/4,(1-\kappa)/4), giving (4).

Tables

MeasurementSelectionModel, κ=1\kappa= 1Model, κ=1/2\kappa= 1/2Measured
Watts et al (2021) , CZT7070–110∘110^\circ1.911.371.85±0.041.85 \pm0.04
9393–103∘103^\circ1.981.401.95±0.071.95 \pm0.07
Ivashkin et al (2023) , scintillators8080–100∘100^\circ, 22.5∘22.5^\circ counters2.46–2.435±0.0182.435 \pm0.018
Separable reference κ\kappa (expected 0.5)Inferred κobs/κpick\kappa_{\mathrm{obs}} / \kappa_{\mathrm{pick}}
Moskal et al (2025), XAD-482∘82^\circ0.5080.796±0.016/0.34±0.050.796 \pm0.016 / 0.34 \pm0.05
94∘94^\circ0.5120.869±0.019/0.58±0.060.869 \pm0.019 / 0.58 \pm0.06

Table 1. Comparison with published measurements of the modulation ratio RR and inferred coefficients.

Table 5: Single DT-PA05 module: idealised and realistic detector models (Eth=15E_{\mathrm{th}} = 15 keV). Fractions are relative to photons that interact in the crystal. Information per pair of interacting photons is the product of the squared usable fraction and the information per usable pair.

ModelUsable fractionCorrect orderInfo per usable pairRelative infob
Idealised (correct order, two clusters)0.17310.0231
Realistic, no Doppler, kinematic order, two clusters0.1890.700.0130.68
Realistic, Doppler 1.5 a.u., larger-deposit order, two clusters0.1460.660.0180.55
Realistic, Doppler 1.5 a.u., kinematic order, two clusters0.1900.700.0130.69
Realistic, Doppler 1.5 a.u., kinematic order, ≤4\le4 clusters0.2660.70/0.87a0.70 / 0.87^{a}0.0121.26
GATE, two clusters0.1810.650.0120.58
GATE, ≤4\le4 clusters0.2660.65/0.74a0.65 / 0.74^{a}0.0101.03

Table 5. Two-cluster events / first pair of events with three or four clusters. Per pair of interacting photons, relative to the idealised model.

Figures

Figure 1. (a) Annihilation channels and their polarisation-correlation coefficients; the pick-off coefficient is unknown. (b) Double-Compton measurement in an eight-module CZT ring with a source in a water cylinder.

References

  1. [23]Moskal P et al 2026a First-in-human quantum entanglement imaging arXiv:2606.29421arxiv.org/abs/2606.29421
  2. [24]Moskal P et al 2026b Roadmap towards quantum entanglement positron emission tomography (QE-PET) arXiv:2607.12182arxiv.org/abs/2607.12182
  3. [25]Pryce M H L and Ward J C 1947 Angular correlation effects with annihilation radiation Nature 160 435 (https://doi.org/10.1038/160435a0)
  4. [30]Tao S J 1972 Positronium annihilation in molecular substances J. Chem. Phys. 56 5499–5510 (https://doi.org/10.1063/1.1677067)

Paper details

vlnc:W_mbfhxnte
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 0

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Claude (Anthropic)

View statement
Looking for related papers…
Contents