WorksPhysics and Astronomy
Absolute Calibration of a Broadband Transmission Grating Spectrometer from 5 to 400 nm with Uncertainty Estimates
Abstract
We present the absolute calibration of a transmission grating spectrometer (TGS) with uncertainty estimates in the 5–400 nm range using synchrotron radiation at the BEAR beamline at Elettra light source in Trieste, Italy. Firstly, the spectral responsivity of a Si photodiode in the spectrometer was absolutely calibrated with respect to the absolutely-calibrated beamline photodiode. Subsequently, this photodiode was used to calibrate the transmittance of Zr and Al thin-film filters, along with bulk MgF and UV-grade fused silica windows at their operating positions in the spectrometer to minimize mounting-induced uncertainties. Next, the quantum efficiency of the spectrometer's CCD camera was calibrated over the 5–400 nm range. Lastly, three gratings with line densities of 10000, 3000, and 1000 lines/mm were calibrated for spectroscopy in the ranges extending from soft X-ray to extreme-, vacuum- and deep-ultraviolet (DUV) wavelengths, respectively. The combination of various filters and gratings enables absolutely calibrated spectroscopy with high versatility across a broad wavelength range from soft X-rays to DUV.