Feasibility of Multifrequency Neural Network-Based Wave Inversion in Ultrasound Time-Harmonic Elastography

Abstract

Ultrasound time-harmonic elastography (THE) is a non-invasive imaging technique that measures soft-tissue stiffness. Relying on the same principle of time-harmonic shear wave excitation as established magnetic resonance elastography (MRE), THE offers full field-of-view stiffness mapping at short acquisition times and sub-second temporal resolution across organs. However, reconstruction of shear wave speed maps from acquired tissue wavefields remains challenging due to speckle noise and sparse spatial coverage of the lateral component. Deep learning methods have shown promising results for image reconstruction in transient-based ultrasound elastography modalities but remain unexplored in THE. In MRE, MF-ElastoNet, a 2D multifrequency neural network-based wave reconstruction method trained on synthetic physics-based wave data, has shown high generalizability across dynamic ranges and resolutions. We investigated the potential of a multifrequency neural network-based wave inversion for THE, by performing a first feasibility analysis. To this end, we adapted the processing pipeline of MF-ElastoNet from MRE to THE. We tested our approach on THE acquired phantom data from the algorithms for mapping elastic modulus Challenge, as well as on in vivo liver THE in 4 healthy volunteers and 14 patients with focal liver lesions, including benign focal nodular hyperplasia, and malignant hepatocellular carcinoma and adenocarcinoma. MF-ElastoNet was successfully adapted to THE with in vivo liver testing. Compared with k-MDEV, the state-of-the-art wave inversion method for THE, Bland-Altman analysis showed a mean bias of -0.02 m/s (95% limits of agreement: -0.36 to 0.33 m/s).

Introduction

Ultrasound elastography is a non-invasive imaging technique that measures soft tissue stiffness [16]. Already standard of care in the staging of liver fibrosis [21], ultrasound elastography is widely evaluated and validated in tumor characterization [18], including focal liver lesions [19] and thyroid nodules [18], as well as in chronic diseases, including chronic kidney disease [18] and heart failure [14].

Different types of ultrasound elastography techniques include strain imaging and shear wave imaging based on acoustic radiation force impulses (ARFI) or on continuous vibrations as in time-harmonic elastography (THE). Relatively novel THE is of particular interest as it allows a wide field-of-view (FOV), high penetration depth (up to 15 cm), millisecond temporal resolution and inherent comparability to MRE. It captures external vibrations at multiple frequencies through conventional B-Mode ultrasound at framerates of 80 to 100 Hz [10]. From the displacement wavefields obtained, a wave inversion reconstruction is performed, obtaining a shear wave speed (SWS) mapping over the full FOV as a proxy for tissue stiffness. Noise and unknown boundary conditions render this reconstruction ill-posed and highly challenging. Furthermore, state-of-the-art reconstruction methods, including the wavenumber-based multifrequency elastovisco inversion (k-MDEV) [21] must be tuned to the acquisition settings used for different ultrasound transducers and in different organs.

Deep learning methods in ultrasound elastography have shown promising results in solving the inverse problem in strain imaging [22, 5, 4, 8, 23] and in ARFI-based elastography [9, 24]. Furthermore, in MRE, where external time-harmonic excitation is widely used, several promising approaches for deep learning-based wave inversion have been developed [15, 17, 11, 3, 1, 2]. However, to the best of our knowledge, there have as of yet been no neural network-based investigations for wave inversion in ultrasound THE.

In this work, we performed the first feasibility analysis of a multifrequency neural network-based wave inversion method for ultrasound THE. To this end, we adapted MF-ElastoNet [1, 2], an uncertainty-aware, multifrequency neural network-based method developed for MRE, to ultrasound THE. We tested the model on phantom datasets from the algorithms for mapping elastic modulus (aMEM) Challenge [6], analyzing the ability of the model to resolve homogeneous regions of interest (ROIs) as well as inclusions of different stiffnesses. We additionally tested our approach on in vivo liver THE in both healthy volunteers and patients with focal liver lesions, including benign (focal nodular hyperplasia) and malignant (hepatocellular carcinoma and adenocarcinoma) cases. In these datasets, we compared our method with k-MDEV [21], the state-of-the-art wave inversion method for THE.

Methods

Data processing

MF-ElastoNet is a 2D2\mathrm{D} multifrequency neural network-based wave inversion approach originally developed for MRE [1, 2]. It maps a multifrequency displacement wavefield to the associated multifrequency SWS and a measure of its uncertainty in a combined processing pipeline. Following an initial preprocessing of the wavefields, the neural network performs the reconstruction in a patch-based approach. Sets of multicomponent and multifrequency 4×44 \times4 image patches are sampled from the wavefield and mapped to multifrequency shear wavelength values in pixels. These are then multiplied by the associated resolution and frequencies to obtain the SWS. The model was trained using a flexible framework of between 1 and 10 input frequencies [1, 2]. As a first feasibility analysis, we used MF-ElastoNet as a pre-trained model, modifying only preprocessing steps to reflect the nature of THE data.

The wavefields were obtained from the ultrasound radiofrequency data using Kasai’s displacement estimation algorithm, followed by a temporal Fourier transform to isolate the induced vibration frequency components [21, 14], and scan conversion to resample to 0.5 mm isotropic Cartesian voxel size [13]. A modified version of the wavefield preprocessing used by MF-ElastoNet for MRE [1] was then performed. Gaussian smoothing with a standard deviation of 2 mm was applied. The smoothing strength was lowered in comparison to the original preprocessing [3] due to the higher resolution of THE. The Laplacian unwrapping and Fourier transform with bin selection, which were part of the original preprocessing for MRE [1], were not included as they are not necessary for THE.

MF-ElastoNet was trained on 4×44 \times4 image patches for wavelengths between 2.5 and 120 pixels [1, 2]. However, at the frequencies and resolution at which THE was performed, the wavelengths present within a 4×44 \times4 patch would be expected to be up to 3 times longer than the training range. As a first feasibility study, we chose not to modify the original model, but to extract larger patches of 12×1212 \times12 and to downsample these to 4×44 \times4. This downsampling mechanism allowed us to obtain data within the original training range, without losing on the resolution of the final image, which would be the case when downsampling the full image.

Finally, MF-ElastoNet provided a spatially resolved measure of the uncertainty of its predictions, allowing the assessment of their trustworthiness. We reported the uncertainty through the coefficient of variation (CV), ensuring its independence with respect to the predicted SWS.

Figure 1 presents the wave inversion processing pipeline. The adapted processing pipeline of MF-ElastoNet for THE has been made publicly available on the open-access elastography platform https://bioiqic-apps.charite.de/ [12].

Wave inversion processing pipeline

Figure 1. Wave inversion processing pipeline: A) Multifrequency complex-valued wavefields were smoothed. 12×1212 \times12 patches were then extracted from the images and downsampled to 4×44 \times4. B) Multifrequency 4×44 \times4 complex-valued patches were processed by MF-ElastoNet, obtaining multifrequency SWS (shear wave speed) and quantifying the associated prediction uncertainty via the CV (coefficient of variation). C) Processed SWS values and associated uncertainty were mapped to frequency-resolved SWS and CV maps. These were averaged to obtain the final SWS and CV maps.

Testing datasets

aMEM Challenge dataset The aMEM Challenge dataset consists of measurements of three CIRS 049 phantoms. Phantom I contains a homogeneous

region, phantom II a 16.68 mm diameter spherical stiff inclusion at a 15 mm depth, and phantom III a 20 mm diameter spherical soft inclusion at a 35 mm depth. These data were acquired using a C5-2v convex array transducer (Verasonics, Inc., Kirkland, WA) and a Verasonics 64 system (Verasonics, Inc., Kirkland, WA) with radiofrequency sampling at 14.28 MHz and at harmonic multifrequency vibration of 100, 120, 140, 160, 180, and 200 Hz. These data are fully described in [6] and are publicly available on the open-access elastography platform https://bioqic-apps.charite.de/ [12].

As there was no ground truth available for these phantoms, we used SWS image quality metrics as proposed in the challenge. In the homogeneous phantom I, spatial variation of SWS was assessed through the standard deviation of the SWS. Phantoms II and III were assessed based on the contrast-to-noise ratio (CNR), edge sharpness (through the edge width), and edge accuracy. The evaluation script, including the regions of interest (ROIs) and the metrics implementation provided by the challenge, was used for this evaluation [6]. A comparison with k-MDEV as submitted to the challenge was performed.

In vivo liver THE We evaluated our approach on in vivo liver THE in 4 healthy volunteers (4 males; age: 33.7±12.033.7 \pm12.0 years) [13] and 14 patients with focal liver lesions, including benign focal nodular hyperplasia (FNH; 4 females; age: 50.5±15.750.5 \pm15.7 years) and malignant hepatocellular carcinoma (HCC; 1 female, 4 males; age: 73.1±7.773.1 \pm7.7 years) and adenocarcinoma (1 female, 4 males; age: 57.7±20.057.7 \pm20.0 years) [19]. In both studies, liver THE was performed using the commercially available THE system (GAMPTmbH, Merseburg, Germany) equipped with a 3.3 MHz convex array transducer. The radiofrequency data were acquired at a framerate of 80 Hz with a penetration depth of 13 cm over 1 s using continuous harmonic vibrations at six superimposed frequencies (27, 33, 39, 44, 50, and 56 Hz). Controlled aliasing was performed to recover the 44, 50, and 56 Hz frequencies due to the Nyquist limit of 40 Hz imposed by the framerate as in [21, 20]. ROIs were manually segmented by a trained radiologist from the B-mode images. A comparison with the SWS maps obtained by k-MDEV [19] was performed, including a Bland-Altman analysis.

Results

aMEM Challenge dataset

The wavefield, SWS, and uncertainty maps for phantoms I, II, and III are displayed in Figure 2, and we report the challenge metrics in Table 1. For MF-ElastoNet, phantom I appeared homogeneous within the identified ROI, except in the far field, where artifacts due to low signal-to-noise ratio (SNR) induced some inhomogeneities. This was well reflected in the uncertainty map, as the CV also increased in those areas. The map produced by k-MDEV had a smoother appearance than that produced by MF-ElastoNet, consistent with its lower standard deviation. For MF-ElastoNet the circular inclusion was clearly identifiable in phantoms II and III. The soft inclusion of phantom III showed better contrast and edge accuracy than the stiff inclusion of phantom II, whereas phantom II had better edge sharpness than phantom III. Across both phantoms, the greater detail captured in the SWS maps produced by MF-ElastoNet relative to k-MDEV was reflected in the higher edge sharpness and edge accuracy in phantom II as well as in the higher CNR and edge accuracy in phantom III.

Wavefields, SWS, and CV maps predicted by MF-ElastoNet for phantoms I, II, and III

Figure 2. Wavefields, SWS (shear wave speed), and CV (coefficient of variation) maps predicted by MF-ElastoNet for phantoms I, II, and III as well as their respective ROIs (regions of interest) as a white dashed line.

PhantomMetrick-MDEVMF-ElastoNet
Phantom ISWS std. (m/s)0.0830.178
Phantom IICNR2.871.57
Phantom IIEdge width (mm)8.163.66
Phantom IIEdge accuracy (mm)-2.371.25
Phantom IIICNR2.092.69
Phantom IIIEdge width (mm)2.936.55
Phantom IIIEdge accuracy (mm)1.690.64

Table 1. aMEM Challenge metrics for k-MDEV and MF-ElastoNet: For phantom I, the spatial variation of the SWS (shear wave speed) is reported through the standard deviation (std.). For phantoms II and III, the CNR (contrast-to-noise ratio), edge sharpness through the edge width, and the edge accuracy are reported.

In vivo liver THE

The B-mode and wavefield images, predicted SWS maps by k-MDEV and MF-ElastoNet, as well as the predicted CV maps by MF-ElastoNet are displayed in Figure 3. The SWS maps produced by MF-ElastoNet appeared more detailed, while those produced by k-MDEV appeared smoother. k-MDEV and MF-ElastoNet showed excellent agreement in SWS values both in liver and tumor ROIs, as shown in Figure 4. Bland-Altman analysis yielded a mean bias of -0.02 m/s (95% limits of agreement: −0.36 to 0.33 m/s).

B-mode, wavefield, SWS, MF-ElastoNet, and uncertainty maps for two healthy volunteers and three patients

Figure 3. B-mode, wavefield, SWS (shear wave speed) maps produced by k-MDEV and MF-ElastoNet, and CV (coefficient of variation) maps predicted by MF-ElastoNet are reported for two healthy volunteers, and three patients, including benign (23-year-old female with focal nodular hyperplasia) and malignant (81-year-old female with hepatocellular carcinoma and 49-year-old male with metastatic colorectal adenocarcinoma) cases. Liver parenchyma (white) and tumor (magenta) are additionally shown overlaid on all images.

Bland-Altman analysis of SWS values

Figure 4. Bland-Altman analysis of SWS (shear wave speed) values: MF-ElastoNet vs k-MDEV. Color denotes the subject groups (green: healthy volunteer, blue: focal nodular hyperplasia (FNH), purple: hepatocellular carcinoma (HCC), brown: adenocarcinoma), and marker shape tissue type (circle: liver parenchyma; cross: tumor).

Discussion and Conclusion

In this work, we performed the first feasibility analysis of a multifrequency neural network-based wave inversion for ultrasound THE. We showed that MF-ElastoNet, a method that was first developed for MRE can be applied to ultrasound THE with minimal changes. We showed in the aMEM Challenge phantoms that MF-ElastoNet was able to resolve homogeneous areas as well as both soft and stiff inclusions. In in vivo liver THE, healthy volunteers and patients with focal liver lesions showed excellent agreement in SWS values with k-MDEV used as our baseline, and even showed improved detail resolution, particularly relevant for tumor characterization [25].

These results were all the more promising considering that the model was applied in a pre-trained manner, with no fine-tuning and only adapting the preprocessing to ultrasound acquisitions. Downsampling was a necessary preprocessing step in this application of the pre-trained model. This step resulted in a loss of information, and adapting the model architecture and training data to THE should therefore be performed to fully leverage the detail resolution in this modality.

Other elastography modalities, such as optical multifrequency time-harmonic elastography [7], show similar challenges in that their resolution is even higher than in THE. Testing and adapting MF-ElastoNet to these challenging datasets

could open the way for a unified cross-modality framework for multifrequency neural network-based wave inversion in elastography.

Acknowledgments. Financial support of the German Research Foundation (DFG, CRC1340 Matrix in Vision, GRK2260 BIOQIC, FOR5628) and the Medical Scientist Pilot Program of the Charité are gratefully acknowledged.

Disclosure of Interests. The authors have no competing interests to declare that are relevant to the content of this article.

References

  1. [1]Bustin, H., Meyer, T., Jordan, J., Walczak, L., Tzschätsch, H., Sack, I., Hennemuth, A.: Multifrequency Neural Network-Based Wave Inversion in MR Elastography, pp. 418–427. Springer Nature Switzerland (Sep 2025). https://doi.org/10.1007/978-3-032-04947-6_40
  2. [2]Bustin, H., Meyer, T., Jordan, J., Walczak, L., Tzschätsch, H., Sack, I., Hennemuth, A.: Multifrequency neural network-based wave inversion in mr elastography with uncertainty quantification. Medical Image Analysis 115, 104317 (Jan 2027). https://doi.org/10.1016/j.media.2026.104317
  3. [3]Bustin, H., Meyer, T., Reiter, R., Jordan, J., Walczak, L., Tzschätsch, H., Sack, I., Hennemuth, A.: Elastonet: Neural network-based multicomponent mr elastography wave inversion with uncertainty quantification. Medical Image Analysis 105, 103642 (Oct 2025). https://doi.org/10.1016/j.media.2025.103642
  4. [4]Delaunay, R., Hu, Y., Vercauteren, T.: An Unsupervised Approach to Ultrasound Elastography with End-to-end Strain Regularisation, pp. 573–582. Springer International Publishing (2020). https://doi.org/10.1007/978-3-030-59716-0_55
  5. [5]Gao, Z., Wu, S., Liu, Z., Luo, J., Zhang, H., Gong, M., Li, S.: Learning the implicit strain reconstruction in ultrasound elastography using privileged information. Medical Image Analysis 58, 101534 (Dec 2019). https://doi.org/10.1016/j.media.2019.101534
  6. [6]Harris, E., Civale, J., Tzschaetzsch, H., Bamber, J., Meyer, T., Palmeri, M., Urban, M.: Algorithms for mapping elastic modulus (ammem) challenge pp. 1–3 (Sep 2024). https://doi.org/10.1109/uffc-js60046.2024.10793812
  7. [7]Jordan, J., Jaitner, N., Meyer, T., Bramè, L., Ghrayeb, M., Köppke, J., Böhm, O., Chandia, S.K., Zaburdaev, V., Chai, L., Tzschätzsch, H., Mura, J., Braun, J., Hagemann, A.I., Sack, I.: Rapid Stiffness Mapping in Soft Biologic Tissues With Micrometer Resolution Using Optical Multifrequency Time-Harmonic Elastography. Advanced Science 12(8) (Dec 2024). https://doi.org/10.1002/advs.202410473
  8. [8]K. Z. Tehrani, A., Mirzaei, M., Rivaz, H.: Semi-supervised Training of Optical Flow Convolutional Neural Networks in Ultrasound Elastography, pp. 504–513. Springer International Publishing (2020). https://doi.org/10.1007/978-3-030-59716-0_48
  9. [9]Kafaei Zad Tehrani, A., Sunethra Dayavansha, E.G., Gu, Y., Candel, I., Wang, M., Tadross, R., Xiao, Y., Rivaz, H., Thomenius, K., Samir, A.: Unsupervised Physics-Inspired Shear Wave Speed Estimation in Ultrasound Elastography, pp. 3–13. Springer Nature Switzerland (Oct 2024). https://doi.org/10.1007/978-3-031-73647-6_1
  10. [10]Klemmer Chandia, S., Schattenfroh, J., Brinker, S.T., Tzschätzsch, H., Sack, I., Meyer, T.: Multimodal assessment of brain stiffness variation in healthy subjects using magnetic resonance elastography and ultrasound time-harmonic elastography. Scientific Reports 14(1) (Nov 2024). https://doi.org/10.1038/s41598-024-79991-y
  11. [11]Ma, S., He, Z., Wang, R., Zhang, A., Sun, Q., Liu, J., Yan, F., Sacks, M.S., Feng, X.Q., Yang, G.Z., Feng, Y.: Measurement of biomechanical properties of transversely isotropic biological tissue using traveling wave expansion. Medical Image Analysis 101, 103457 (Apr 2025). https://doi.org/10.1016/j.media.2025.103457
  12. [12]Meyer, T., Marticorena Garcia, S., Tzschätzsch, H., Herthum, H., Shahryari, M., Stencel, L., Braun, J., Kalra, P., Kolipaka, A., Sack, I.: Comparison of inversion methods in MR elastography: An open-access pipeline for processing multifrequency shear-wave data and demonstration in a phantom, human kidneys, and brain. Magnetic Resonance in Medicine. 88, 1840–1850 (2022). https://doi.org/10.1002/mrm.29320
  13. [13]Meyer, T., Tzschätzsch, H., Wellge, B., Sack, I., Kröncke, T., Martl, A.: Valsalva maneuver decreases liver and spleen stiffness measured by time-harmonic ultrasound elastography. Frontiers in Bioengineering and Biotechnology 10 (May 2022). https://doi.org/10.3389/fbioe.2022.886363
  14. [14]Meyer, T., Wellge, B., Barzen, G., Klemmer Chandia, S., Knebel, F., Hahn, K., Elgeti, T., Fischer, T., Braun, J., Tzschätzsch, H., Sack, I.: Cardiac elastography with external vibration for quantification of diastolic myocardial stiffness. Journal of the American Society of Echocardiography 38(5), 431–442 (May 2025). https://doi.org/10.1016/j.echo.2024.11.009
  15. [15]Murphy, M.C., Manduca, A., Trzasko, J.D., Glaser, K.J., Huston, J., Ehman, R.L.: Artificial neural networks for stiffness estimation in magnetic resonance elastography. Magnetic Resonance in Medicine. 80, 351–360 (2017). https://doi.org/10.1002/mrm.27019
  16. [16]Sack, I., Schaeffter, T.: Quantification of Biophysical Parameters in Medical Imaging. Springer International Publishing (2024). https://doi.org/10.1007/978-3-031-61846-8
  17. [17]Scott, J.M., Arani, A., Manduca, A., McGee, K.P., Trzasko, J.D., Huston, J., Ehman, R.L., Murphy, M.C.: Artificial neural networks for magnetic resonance elastography stiffness estimation in inhomogeneous materials. Medical Image Analysis. 63, 101710 (2020). https://doi.org/10.1016/j.media.2020.101710
  18. [18]Sigrist, R.M., Liau, J., Kaffas, A.E., Chammas, M.C., Willmann, J.K.: Ultrasound elastography: Review of techniques and clinical applications. Therapeutics 7(5), 1303–1329 (2017). https://doi.org/10.7150/thno.18650
  19. [19]Spiesecke, P., Wolff, M., Fischer, T., Sack, I., Meyer, T.: Ultrasound time-harmonic elastography: Habitat viscosity and tumor stiffness heterogeneity for differentiation of benign and malignant liver lesions (Mar 2026). https://doi.org/10.64898/2026.03.12.26348218
  20. [20]Tzschätzsch, H., Kreft, B., Schrank, F., Bergs, J., Braun, J., Sack, I.: In vivo time-harmonic ultrasound elastography of the human brain detects acute cerebral stiffness changes induced by intracranial pressure variations. Scientific Reports 8(1) (Dec 2018). https://doi.org/10.1038/s41598-018-36191-9
  21. [21]Tzschätzsch, H., Nguyen Trong, M., Scheuermann, T., Ipek-Ugay, S., Fischer, T., Schultz, M., Braun, J., Sack, I.: Two-dimensional time-harmonic elastography of the human liver and spleen. Ultrasound in Medicine & Biology 42(11), 2562–2571 (Nov 2016). https://doi.org/10.1016/j.ultrasmedbio.2016.07.004
  22. [22]Wu, S., Gao, Z., Liu, Z., Luo, J., Zhang, H., Li, S.: Direct Reconstruction of Ultrasound Elastography Using an End-to-End Deep Neural Network, pp. 374–382. Springer International Publishing (2018). https://doi.org/10.1007/978-3-030-00928-1_43
  23. [23]Yin, Z., Li, G.Y., Zhang, Z., Zheng, Y., Cao, Y.: Swenet: A physics-informed deep neural network (pinn) for shear wave elastography. IEEE Transactions on Medical Imaging 43(4), 1434–1448 (Apr 2024). https://doi.org/10.1109/tmi.2023.3338178
  24. [24]Zhang, Z., Ma, S., Yin, Z., Qiu, J., Hu, Z., Li, G.Y., Feng, X.Q., Cao, Y.: Unveiling hidden features of acoustic radiation forces in soft tissues via physics-informed neural network-based full shear wave inversion. Journal of the Mechanics and Physics of Solids 205, 106326 (Dec 2025). https://doi.org/10.1016/j.jmps.2025.106326
  25. [25]Zhao, W., Skrip, L., Tzschätzsch, H., Meyer, T., Aghamiry, H.S., Arnold, A., Böhne, L.Ä., Jordan, J., Boehm, O., Böhm, A.K., Goerner, S., Schattenfroh, J., Wu, Y., de Moraes, P.A.D., Pratschke, J., Braun, J., Sauer, I.M., Guo, J., Sack, I., Hillebrandt, K.H.: Compact tabletop magnetic resonance elastography for mapping soft tissue viscoelasticity. Advanced Science (May 2026). https://doi.org/10.1002/advs.75728

Paper details

vlnc:W_myn9jq7a
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 1

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Engineering

Not recorded for this paper

View statement
Looking for related papers…
Contents