betaFeedback

WorksPhysics and Astronomy

Imaging reconstruction method on X-ray data of CMOS polarimeter combined with coded aperture

Abstract

X-ray polarization is a powerful tool for unveiling the anisotropic characteristics of high-energy celestial objects. We present a novel imaging reconstruction method designed for hard X-ray polarimeters employing a Si CMOS sensor and coded apertures, which function as a photoelectron tracker and imaging optics, respectively. Faced with challenges posed by substantial artifacts and background noise in the coded aperture imaging associated with the conventional balanced correlation method, we adopt the Expectation-Maximization (EM) algorithm as the foundation of our imaging reconstruction method. The newly developed imaging reconstruction method is validated with imaging polarimetry and a series of X-ray beam experiments. The method demonstrates the capability to accurately reproduce an extended source comprising multiple segments with distinct polarization degrees. Comparative analysis exhibits a significant enhancement in imaging reconstruction accuracy compared to the balanced correlation method, with the background noise levels reduced to 17%. The outcomes of this study enhance the feasibility of Cube-Sat imaging polarimetry missions in the hard X-ray band, as the combination of Si CMOS sensors and coded apertures is a promising approach for realizing it.

Introduction

X-ray polarization is a distinctive method for capturing unprecedented images of high-energy celestial objects. It carries essential information about anisotropic physical phenomena, such as the magnetic field structure responsible for synchrotron radiation, the emission geometry of Compton scattering, and the properties of the strong gravitational field around compact objects. Recent advancements in X-ray polarization studies have been marked by the launch of the Imaging X-ray Polarimeter Explorer (IXPE; [1, 2]) in 2021. This mission, which covers the soft X-ray band of 2–8 keV, has substantially broadened our capabilities in this field (e.g., Crab Nebula observation: [3]). While IXPE focuses on the soft X-ray band through photoabsorption, the Soft Gamma-ray Detector (SGD) onboard Hitomi [4] and the Polarized Gamma-ray Observer+ (PoGO+;[5]) utilize Compton scattering to detect X-ray polarization in the hard X-ray band. Notably, both missions successfully detected polarization from the Crab Nebula in the higher end of the hard X-ray band: 60–160 keV for SGD [6] and 20–160 keV for PoGO+ [7]. However, a significant observational gap exists in the lower end of the hard X-ray band due to the absence of established observational techniques, specifically in the energy range of 10–30 keV. This energy range is particularly important because non-thermal components, carrying anisotropic information, become dominant over unpolarized thermal emission above ∼10\sim10 keV. In addition, the abundance of photon counts in this band, compared to higher energy bands, provides a rich dataset for detailed analysis.

To address this observational gap, a 6U CubeSat mission named the Coded Imaging Polarimeter of High Energy Radiation (cipher; [8]) is under development. It aims to conduct imaging polarimetry in the 10–30 keV band and obtain the polarization map for bright extended sources such as the Crab Nebula. This imaging polarimeter is characterized by a micro-pixel (∼μm\sim\mu\mathrm{m} pixel pitch) Si CMOS sensor and coded apertures. The micro-pixel CMOS sensor serves as a sophisticated photoelectron tracker, enabling the determination of the polarization angle of incident photons. Given the spatial constraints of the CubeSat mission, coded apertures provide the solution for imaging capability, eliminating the need for X-ray mirrors. The CMOS sensor has demonstrated polarization detection capabilities in the 6–30 keV range [9, 10], meeting the fundamental requirements of cipher. The imaging reconstruction of the coded apertures is also established [11], employing a conventional reconstruction technique called the “balanced correlation method” [12]. However, this method introduces significant artifacts and noise levels, which hinder effective imaging polarimetry.

Recognizing the limitation of the balanced correlation method, this paper presents a novel imaging reconstruction method for polarimetry utilizing the Expectation-Maximization (EM) algorithm [13]. It is a statistical approach to efficiently derive maximum likelihood estimates, and has been applied to imaging reconstruction techniques extensively (e.g., [14, 15, 16, 17, 18]). This method holds the potential to achieve imaging polarimetry with enhanced precision, with reduced artifacts and noise levels. The remainder of this paper is organized as follows. We first provide the formulation of the EM algorithm that is applied to imaging polarimetry, and describe the experimental setup to examine the new method in Section 2. The data processing and results of imaging reconstruction are presented in § 3. We then discuss the precision of our new imaging reconstruction method by comparing it with the conventional balanced correlation method in § 4. Conclusions are presented in § 5.

Imaging reconstruction method

Application of EM algorithm to imaging polarimetry

We developed a novel reconstruction method for the coded aperture imaging based on the EM algorithm [13]. This statistical approach involves estimating the maximum likelihood from incomplete experimental data through iterative cycles of Expectation (E-) and Maximization (M-) steps, continuously updating the estimation. The ll-th E-step and M-step in the imaging reconstruction analysis are defined as follows:

(E – step)D~v(l)=∑up(v∣u)S~u(l),(1)\text{(E -- step)}\qquad\widetilde{D}_{v}^{(l)} = \sum_{u} p(v \mid u)\widetilde{S}_{u}^{(l)}, \tag*{(1)}
(M – step)S~u(l+1)=1∑vDv∑vDvp(v∣u)S~u(l)D~v(l),(2)\text{(M -- step)}\qquad\widetilde{S}_{u}^{(l+1)} = \frac{1}{\sum_{v}D_{v}}\sum_{v}D_{v}\frac{p(v \mid u)\widetilde{S}_{u}^{(l)}}{\widetilde{D}_{v}^{(l)}}, \tag*{(2)}

where ll increments by 1 at each iteration [14]. In these equations, S~u(l)\widetilde{S}_{u}^{(l)}, D~v(l)\widetilde{D}_{v}^{(l)}, and DvD_{v} represent the estimated incident source distribution, the estimated event distribution on the detector, and the actual event distribution derived from experimental data, respectively. The term p(v∣u)p(v \mid u) denotes the posterior probability that a photon from the uu-th sky region is detected in the vv-th detector region.

The method is applied to the imaging polarimeter with a CMOS sensor and coded apertures. The sky region is defined as a two-dimensional angular distribution, u=(θx,θy)\boldsymbol{u} = (\theta_x, \theta_y), with respect to the optical axis, while the detector region is defined as a two-dimensional pixel distribution, v=(dx,dy)\boldsymbol{v} = (d_x, d_y). The polarization angle and polarization degree need to be assigned to each sky region. Since the CMOS sensor is sensitive only to the horizontal and vertical directions, we focused on two Stokes parameters, II and QQ, ignoring the effect of UU. These Stokes parameters are linked to the experimental data through the branching ratio in double-pixel events, which is, the ratio between horizontally-elongated (H-type) and vertically-elongated (V-type) events [8]. Equations (1) and (2) can be expressed as

D~v,t(l)=∑up(v∣u)S~u,t(l),(3)\widetilde{D}_{v,t}^{(l)} = \sum_{u}p(v \mid u)\widetilde{S}_{u,t}^{(l)}, \tag*{(3)}
S~u,t(l+1)=1∑vDv,t∑vDv,tp(v∣u)S~u,t(l)D~v,t(l),(4)\widetilde{S}_{u,t}^{(l+1)} = \frac{1}{\sum_{v}D_{v,t}}\sum_{v}D_{v,t}\frac{p(v \mid u)\widetilde{S}_{u,t}^{(l)}}{\widetilde{D}_{v,t}^{(l)}}, \tag*{(4)}

where tt is a two-valued variable representing t=H-typet = \mathrm{H\text{-}type} or t=V-typet = \mathrm{V\text{-}type}. The Stokes parameters II and QQ can be calculated using

\begin{pmatrix} I_{u}^{(l)} \\ Q_{u}^{(l)} \end{pmatrix} = \begin{pmatrix} 1 & 1 \\ 1/m & -1/m \end{pmatrix} \begin{pmatrix} S_{u,\mathrm{H\text{-}type}}^{(l)} \\ S_{u,\mathrm{V\text{-}type}}^{(l)} \end{pmatrix}, \tag*{(5)}\end{pmatrix}

where mm denotes the modulation factor of the CMOS sensor. The posterior probability p(v∣u)p(v \mid u) is calculated from the coded aperture pattern:

p(v∣u)∝{1(u to v pass through aperture)τ(u to v pass through mask)0(otherwise),(6)p(v \mid u) \propto \begin{cases} 1 & (u\ \text{to}\ v\ \text{pass through aperture}) \\ \tau& (u\ \text{to}\ v\ \text{pass through mask}) \\ 0 & (\text{otherwise}), \end{cases} \tag*{(6)}

where τ\tau is the transmittance of the coded aperture.

Beam experiment

X-ray beam experiments were conducted at BL20B2 in SPring-8, a synchrotron radiation facility in Hyogo, Japan, to evaluate the imaging-polarimetry capability of the system. The beamline emits polarized monochromatic X-rays with a polarization degree of nearly 100% (e.g., [9]). The left panel of Figure 1 shows the experimental setup at the beamline. We utilized the GMAX0505RF sensor, manufactured by GPixel Inc., as the polarization detector. This Si CMOS sensor, with a pixel pitch of 2.5 μm2.5\ \mu\mathrm{m}, was originally designed for visible light and near-infrared imaging but also has sensitivity for the X-ray polarization [10]. The sensor was positioned at the farthest part from the beam source. Upstream of it, a 25 cm25\ \mathrm{cm} collimator was mounted to collimate incident X-rays. On the upstream side of the collimator, a 0.1 mm0.1\ \mathrm{mm} thick board made of SUS304 was attached. This board features numerous holes, functioning as eight independent coded apertures with distinct random patterns. The collimator ensures that the shadows of these patterns do not overlap at the detector plane. Each coded aperture comprises of a 64×6464 \times64 pattern with a pitch size of 35 μm35\ \mu\mathrm{m}, covering 896×896896 \times896 pixels on the detector plane (Figure 1 middle). The system has a field of view and angular resolution of 0.5∘×0.5∘0.5^\circ\times0.5^\circ and 29′′29^{\prime\prime}, respectively. The orientation of the collimator (θx\theta_x and θy\theta_y) is adjustable using a rotation stage, facilitating the control of the incident X-ray direction. Moreover, the entire system can rotate around the beam axis, allowing for the adjustment of the polarization angle detected by the CMOS sensor.

Figure 1. (left) Imaging polarimetry setup at Spring-8 beam line. (middle) Picture of coded apertures. (right) Sky region subject to the imaging scan experiment.

A comprehensive sky scan was conducted by manipulating the rotation stage, and an image for an extended source was created by merging the acquired data. The data points for the imaging scan experiments are illustrated in the right panel of Figure 1, covering a 420′′×420′′420^{\prime\prime} \times420^{\prime\prime} plane with a pitch size of 15′′15^{\prime\prime}. A total of 29×29=84129 \times29 = 841 scan points were utilized. Table 1 provides the details on the datasets employed in this paper. To verify the hard X-ray imaging polarimetry, we exposed the system to 16.0 keV16.0\ \mathrm{keV} polarized X-rays with polarization angles of both 0∘0^\circ and 90∘90^\circ. This beam energy was selected by balancing between the polarization detection capability and the quantum efficiency of the sensor. Lower-energy photons would result in low sensitivity for polarization detection because they are more likely to be recorded as single-pixel events. Conversely, higher-energy photons lead to poor statistical quantity due to their smaller cross section with the Si CMOS sensor. The incident beam was attenuated by a 120 μ\mum Cu filter to avoid pile-up. Each dataset has 10 frames per scan point, resulting in a total of 8410 frames. The initial four datasets in Table 1 were merged to generate the “0° polarization data”, while the latter four were merged to generate the “90° polarization data”.

Start time (JST)Beam energy (keV)Polarization angle (degree)Frame exposure (ms)Number of frames
2021-11-03 18:5016.006008410
2021-11-03 20:4016.006008410
2021-11-03 22:4216.006008410
2021-11-04 00:2216.006008410
2021-11-04 03:4216.0906008410
2021-11-04 05:2316.0906008410
2021-11-04 07:0016.0906008410
2021-11-04 08:5416.0906008410

Table 1. List of datasets.

Results

Data processing

We performed standard data processing on the acquired data using ComptonSoft [19, 20, 21], which included pedestal subtraction, bad pixel exclusion, event extraction, and event classification. Subsequent to the data processing, double-pixel events exhibiting either horizontal (H-type) or vertical (V-type) elongation were specifically selected. The polarization angle and polarization degree can be determined by measuring the ratio between the two types of events. The numbers of H-type/V-type events were 9485530/8008517 for 0° polarization and 7622987/9741000 for 90° polarization. The modulation factor was calculated from these values to be m=0.1033±0.0001m = 0.1033 \pm0.0001 for 16.0 keV polarized photons (for more details of the modulation factor, see [10]).

In addition to the 0° and 90° polarization data, a new dataset named “imaging polarimetry test data” with various polarization degrees depending on incident directions was generated. This dataset can easily be generated by appropriately blending the 0° and 90° polarization data; for instance, an unpolarized source can be simulated by mixing equal proportions of the 0° and 90° polarization data. To evaluate the imaging polarimetry capability, we divided the entire scan region into four sub-regions and assigned distinct polarization degrees (Q/IQ/I) to them. Figure 2 illustrates the definition of the segmentation. We assigned Q/I=0.4Q/I = 0.4, −1.0-1.0 (90° polarization), 1.01.0 (0° polarization), and 0.00.0 (unpolarized) to Regions 1, 2, 3, and 4, respectively, where the intensity II was spatially uniform. The first value (Q/I=0.4Q/I = 0.4) was assigned to examine a “realistic” polarization degree for a celestial object, while the latter three values (Q/I=−1.0,1.0,0.0Q/I = -1.0, 1.0, 0.0) were included to examine extreme cases.

Figure 2. Definition of the imaging polarimetry test data (see § 3.1).

Imaging polarimetry

The EM algorithm was applied to the reconstruction of the polarization map from the encoded images on the detector plane. We set S~u,t(0)\tilde{S}^{(0)}_{u,t} as a spatially uniform distribution, and iterated through the expectation and maximization steps using equations (3) and (4). The reconstruction was simultaneously applied to all eight coded apertures, with the transmittance of the apertures set to τ=0.02\tau= 0.02 (corresponding to 0.1 mm SUS304 for 16 keV photons). The encoded image on the detector plane was binned in every 4×44 \times4 pixels due to the limited memory resources, but this did not affect the results as the angular resolution is primarily dominated by the aperture pitch. The reconstructed images were generated with an image pixel size of 20′′×20′′20^{\prime\prime} \times20^{\prime\prime}, which is slightly smaller than the angular resolution of the system. The EM steps were iterated until l=1500l = 1500, ensuring a sufficient convergence. The reconstructed images were obtained after the vignetting correction was applied.

As a simple case, we first conducted the imaging reconstruction on the 0∘0^{\circ} polarization data, where II and QQ distributions should ideally be identical. The upper panels of Figure 3 display the reconstructed images of II, QQ, and Q/IQ/I. The uniformly distributed images of II and QQ were successfully reconstructed, consistent with the actual scan region of 420′′×420′′420^{\prime\prime} \times420^{\prime\prime}. The polarization degrees also yield a convincing value of Q/I=1Q/I = 1 throughout the entire image. It is important to note that the Q/IQ/I image indicates horizontal polarization of incident photons even beyond the scan region. This artifact arises because a portion of the detected photons are inaccurately projected to the outer region. Consequently, careful examination of both II and Q/IQ/I images is necessary, and polarization degrees should not be relied on in regions where the photon intensity II is faint. We also tried various patterns of uniformly distributed Q/IQ/I values (0.5, 0, -0.5, -1.0) and confirmed that the reconstructed images successfully reproduced the uniformly distributed polarization degrees.

Reconstructed images of intensity, Stokes Q, and polarization degree

Figure 3. The upper panels show the reconstructed images from 0∘0^\circ polarization data, while the lower panels show those from imaging polarimetry test data (see text). Reconstructed images include I(θx,θy)I(\theta_x,\theta_y) (left), Q(θx,θy)Q(\theta_x,\theta_y) (middle), and polarization degree Q/IQ/I (right).

Subsequently, we applied the imaging reconstruction to the imaging polarimetry test data, which is described in § 3.1 and defined in Figure 2. The lower panels of Figure 3 display the reconstructed images. A uniform distribution of II extending over 420′′×420′′420^{\prime\prime} \times420^{\prime\prime} similar to the 0∘0^{\circ} polarization data was observed. But a totally different image of QQ, clearly divided into four segments, was observed. Each of the four regions corresponds to the definition in Figure 2. The reconstructed image of Q/IQ/I successfully reflects the distinct polarization degrees of the four regions, displaying ∼0.4\sim0.4, ∼−1.0\sim-1.0, ∼1.0\sim1.0, and ∼0.0\sim0.0 for Regions 1, 2, 3, and 4, respectively.

In Figure 4, a detailed evaluation of the imaging reconstruction results is presented. The left panel shows the distribution of I/Imax⁡I/I_{\max} for the four regions, where Imax⁡I_{\max} corresponds to the maximum value of the entire image. The four regions display comparable distributions, indicating that the spatially uniform distribution of II is successfully reproduced. The right panel exhibits the distribution of the polarization degree Q/IQ/I for the four regions. The four regions display peaks at the predetermined polarization degrees and are significantly distinguishable from each other. Still, non-negligible uncertainties of Δ(Q/I)∼0.2\Delta(Q/I) \sim0.2 remain due to systematic uncertainties stemming from potentially non-uniform modulation factors across the detector plane and biased projections caused by randomized patterns of the coded apertures. While a poorer statistical quantity would result in worse reproducibility, this imaging analysis benefits from an average of ∼40000\sim40000 photon counts contributing each imaging pixel.

Figure 4. Evaluation of imaging reconstruction on the imaging polarimetry test data. The left panel shows the histogram of source intensity plotted for the Regions 1–4. The right panel displays the histogram of polarization degree plotted for Regions 1–4.

Discussion

In Section 3, our formulation of the imaging reconstruction method using the EM algorithm was examined, and the polarization map was successfully reconstructed from the imaging polarimetry test data. In this section, the accuracy of the imaging reconstruction was evaluated by comparing our method with the conventional balanced correlation method [12]. The balanced correlation method is a prominent decoding method for coded apertures, especially suitable for Uniformly Redundant Array (URA; [22, 23]), but it also works for random pattern arrays used in our system (for details of the balanced correlation method in our system, see Section 2 of [11]).

The left panel of Figure 5 shows the reconstructed II image of the 0∘0^\circ polarization data using the balanced correlation method. Despite the incident flux being spatially uniform along the scan region, it exhibits a significant imbalance with unpredictably intense signals in the upper left region. The non-source region also displays large fluctuations with a noisy background in the upper left region. These artifacts and noisy background signals arise from the spatially non-uniform distribution of the apertures, which could back-project more signals to the upper left region of the sky. In the right panel of Figure 5, the intensity distributions of the source and background regions are presented, comparing the results between the EM algorithm and the balanced correlation method. The EM algorithm exhibits sharp distributions in both source and background regions, with the two distributions clearly separated. On the other hand, the balanced correlation method has flat distributions for the source and background regions, and these two distributions are largely overlapped. This difference indicates fewer artifacts and lower noise levels in the EM algorithm compared to the balanced correlation method.

Figure 5. (left) Reconstructed II image of 0∘0^\circ polarization data using the balanced correlation method. (right) Intensity distributions of source (src) and background (bkg) regions compared between the EM algorithm (labeled as EM) and the balanced correlation method (labeled as BC).

The noise levels can be evaluated by

Lnoise=σbkgμsrc,(7)L_{\mathrm{noise}} = \frac{\sigma_{\mathrm{bkg}}}{\mu_{\mathrm{src}}}, \tag*{(7)}

where σbkg\sigma_{\mathrm{bkg}} denotes the standard deviation in the background region, and μsrc\mu_{\mathrm{src}} is the averaged intensity in the source region. Here, the source region represents a collection of the image pixels within the central 420′′×420′′420^{\prime\prime} \times420^{\prime\prime} area while the background region denotes a collection of the image pixels outside this area. We obtain Lnoise=0.13L_{\mathrm{noise}} = 0.13 for the EM algorithm and Lnoise=0.75L_{\mathrm{noise}} = 0.75 for the balanced correlation method, which means that our new imaging reconstruction method reduces the noise level to 17% of the conventional method. The significant artifacts and noise levels in the balanced correlation method had prevented us from creating any efficient polarization map, but the improved method using the EM algorithm has successfully realized it with sufficiently reduced artifacts and noise levels, as depicted in Figure 3 and Figure 4.

Several aspects still require refinement for the application of this system and analysis method to imaging polarimetry. Firstly, while we assumed a uniform modulation factor within the CMOS sensor, this assumption needs careful examination. Deviations of ∼0.01\sim0.01 from the adopted original value (m=0.1033m = 0.1033) could occur when focusing on specific limited areas of the sensor. A better understanding of the dependence of the modulation factor on the sensor pixels would mitigate systematic uncertainties in imaging polarimetry. Secondly, the biased random patterns of the coded apertures also require scrutiny. As observed in the left panel of Figure 5, generated by the balanced correlation method, these coded apertures tend to project more photons to the upper left region of the reconstructed image, potentially impacting the accuracy of the imaging reconstruction in the case of the EM algorithm. It is still underway to develop more sophisticated coded aperture patterns. Furthermore, the performance of imaging polarimetry is influenced by the shape of the polarization distribution. While in this study, the boundaries between different polarization profiles are located around the center of the field of view, we also confirmed that the reconstruction accuracy diminishes when these boundaries are located farther from the center. This is due to reduced sensitivity of the coded apertures in these regions. These considerations are beyond the scope of this paper and will be addressed in future discussions.

Conclusions

We developed a new imaging reconstruction method for hard X-ray imaging polarimetry employing a combination of a CMOS sensor and coded apertures. Motivated by the significant artifacts and background noise levels associated with the conventional balanced correlation method, we introduced the EM algorithm as the foundation of our new imaging reconstruction method. The effectiveness of the newly developed method was confirmed through X-ray beam experiments at Spring-8, where the imaging polarimeter was exposed to the X-ray beam and captured an extended source by a comprehensive scan of the sky. The method exhibited remarkable capabilities in accurately reproducing an extended source comprising multiple segments characterized by distinct polarization degrees. Specifically, it successfully reproduced four regions with Q/I=−1.0Q/I = -1.0, 0.0, 0.4, 1.0. Our developed imaging reconstruction method achieved a significant reduction in artifacts and noise levels compared to the balanced correlation method. The background noise level experienced a significant reduction to 17%. The outcomes of this study demonstrate sufficient feasibility for the hard X-ray imaging polarimetry utilizing the combination of CMOS sensors and coded apertures. This approach represents one of the most promising ways to achieve CubeSat missions on hard X-ray polarimetry.

Acknowledgment

We thank the anonymous referees, who improved this work with their valuable comments. We appreciate the helpful technical support by M.Hoshino and K.Uesugi at SPring-8. This research is supported by the Japan Society for the Promotion of Science (JSPS) KAKENHI Grant No. 20J20050, 23KJ2214 (TT), 18H05861, 19H01906, 19H05185, 22H00128, 22K18277 (HO), 20J00119 (AT), and 23KJ0780 (TI). The synchrotron radiation experiments were performed at BL20B2 in SPring-8 with the approval of the Japan Synchrotron Radiation Research Institute (JASRI) (Proposal No. 2019B1369, 2020A1343, 2021B1542, and 2022B1477). This research is also supported by Society for Promotion of Space Science (TT).

References

  1. [1]M. C. Weisskopf, et al., The Imaging X-ray Polarimetry Explorer (IXPE), in: J.-W. A. den Herder, T. Takahashi, M. Bautz (Eds.), Space Telescopes and Instrumentation 2016: Ultraviolet to Gamma Ray, Vol. 9905 of Society of Photo-Optical Instrumentation Engineers (SPIE) Conference Series, 2016, p. 990517. doi:10.1117/12.2235240.DOI
  2. [2]M. C. Weisskopf, et al., The Imaging X-Ray Polarimetry Explorer (IXPE): Pre-Launch, Journal of Astronomical Telescopes, Instruments, and Systems 8 (2) (2022) 026002. arXiv:2112.01269, doi:10.1117/1.JATIS.8.2.026002.
  3. [3]N. Bucciantini, et al., Simultaneous space and phase resolved X-ray polarimetry of the Crab pulsar and nebula, Nature Astronomy 7 (2023) 602–610. arXiv:2207.05573, doi:10.1038/s41550–023–01936–8.arxiv.org/abs/2207.05573
  4. [4]H. Tajima, et al., Design and performance of Soft Gamma-ray Detector onboard the Hitomi (ASTRO-H) satellite, Journal of Astronomical Telescopes, Instruments, and Systems 4 (2018) 021411. doi:10.1117/1.JATIS.4.2.021411.
  5. [5]M. Chauvin, et al., Calibration and performance studies of the balloon-borne hard X-ray polarimeter PoGO+, Nuclear Instruments and Methods in Physics Research A 859 (2017) 125–133. arXiv:1703.07627, doi:10.1016/j.nima.2017.03.027.DOI
  6. [6]Hitomi Collaboration, et al., Detection of polarized gamma-ray emission from the Crab nebula with the Hitomi Soft Gamma-ray Detector, PASJ70 (6) (2018) 113. arXiv:1810.00704, doi:10.1093/pasj/psy118.
  7. [7]M. Chauvin, et al., Shedding new light on the Crab with polarized X-rays, Scientific Reports 7 (2017) 7816. arXiv:1706.09203, doi:10.1038/s41598–017–07390–7.arxiv.org/abs/1706.09203
  8. [8]H. Odaka, et al., Concept of a CubeSat-based hard x-ray imaging polarimeter: cipher, in: Society of Photo-Optical Instrumentation Engineers (SPIE) Conference Series, Vol. 11444 of Society of Photo-Optical Instrumentation Engineers (SPIE) Conference Series, 2020, p. 114445V. doi:10.1117/12.2560615.DOI
  9. [9]K. Asakura, et al., X-ray imaging polarimetry with a 2.5-μm pixel CMOS sensor for visible light at room temperature, Journal of Astronomical Telescopes, Instruments, and Systems 5 (2019) 035002. arXiv:1906.00012, doi:10.1117/1.JATIS.5.3.035002.DOI
  10. [10]T. Iwata, K. Hagino, H. Odaka, T. Tamba, M. Ichihashi, T. Kato, K. Ishiwata, H. Kuramoto, H. Matsuhashi, S. Arai, T. Minami, S. Takashima, A. Bamba, Development of the X-ray polarimeter using CMOS imager: polarization sensitivity of a 1.5 μm pixel CMOS sensor, arXiv e-prints (2024) arXiv:2405.18907arXiv:2405.18907.DOI
  11. [11]T. Kasuga, et al., Artifact-less coded aperture imaging in the x-ray band with multiple different random patterns, Journal of Astronomical Telescopes, Instruments, and Systems 6 (2020) 035002. arXiv:2007.15278, doi:10.1117/1.JATIS.6.3.035002.DOI
  12. [12]E. E. Fenimore, T. M. Cannon, Coded aperture imaging with uniformly redundant arrays, Applied Optics17 (1978) 337–347. doi:10.1364/AO.17.000337.DOI
  13. [13]A. Dempster, N. Laird, D. Rubin, Maximum Likelihood from Incomplete Data Via the EM Algorithm, Journal of the Royal Statistical Society: Series B 39 (1977) 1. doi:10.1111/j.2517-6161.1977.tb01600.x.
  14. [14]S. Ikeda, H. Odaka, M. Uemura, T. Takahashi, S. Watanabe, S. Takeda, Bin mode estimation methods for Compton camera imaging, Nuclear Instruments and Methods in Physics Research A 760 (2014) 46–56. arXiv:1312.4291, doi:10.1016/j.nima.2014.05.081.
  15. [15]A. J. Reader, et al., One-pass list-mode EM algorithm for high-resolution 3-D PET image reconstruction into large arrays, IEEE Transactions on Nuclear Science 49 (3) (2002) 693–699. doi:10.1109/TNS.2002.1039550.
  16. [16]J. A. Fessler, A. O. Hero, Penalized maximum-likelihood image reconstruction using space-alternating generalized EM algorithms, IEEE Transactions on Image Processing 4 (10) (1995) 1417–1429. doi:10.1109/83.465106.
  17. [17]E. F. Maher, N. M. Laird, EM algorithm reconstruction of particle size distributions from diffusion battery data, Journal of Aerosol Science 16 6 (1985) 557–570. doi:10.1016/0021-8502(85)90007–2.
  18. [18]M. Zanetti, F. Bovolo, L. Bruzzone, Rayleigh–Rice Mixture Parameter Estimation via EM Algorithm for Change Detection in Multispectral Images, IEEE Transactions on Image Processing 24 (12) (2015) 5004–5016. doi:10.1109/TIP.2015.2474710.
  19. [19]H. Odaka, et al., Development of an integrated response generator for Si/CdTe semiconductor Compton cameras, Nuclear Instruments and Methods in Physics Research A 624 (2) (2010) 303–309. doi:10.1016/j.nima.2009.11.052.
  20. [20]H. Suzuki, et al., Development of the detector simulation framework for the Wideband Hybrid X-ray Imager onboard FORCE, Nuclear Instruments and Methods in Physics Research A 979 (2020) 164433. arXiv:2007.07919, doi:10.1016/j.nima.2020.164433.
  21. [21]T. Tamba, et al., Simulation-based spectral analysis of X-ray CCD data affected by photon pile-up, PASJ74 (2) (2022) 364–383. arXiv:2112.14176, doi:10.1093/pasj/psab131.
  22. [22]E. E. Fenimore, T. M. Cannon, Uniformly redundant arrays: digital reconstruction methods, Applied Optics20 (10) (1981) 1858–1865. doi:10.1364/AO.20.001858.DOI
  23. [23]S. R. Gottesman, E. E. Fenimore, New family of binary arrays for coded aperture imaging, Applied Optics28 (20) (1989) 4344–4352. doi:10.1364/AO.28.004344.

Paper details

vlnc:W_r9pt2hwh
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 0

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…