Digital confocal imaging in holographic optical coherence tomography at high space-bandwidth product
Abstract
Full-field Fourier-domain optical coherence tomography (FF-FD-OCT) enables rapid volumetric imaging but lacks the confocal gating inherent to point-scanning OCT. By acquiring volumes under multiple oblique illuminations, a digital confocal volume can be synthesized in post-processing, a technique originally developed within the framework of reflection matrix imaging (RMI). Existing implementations, however, are often time-domain, off-axis, or use a Michelson interferometer, which reduces sensitivity, speed and the usable space-bandwidth product (SBP). Here, we present an oblique-illumination FF-FD-OCT system in a Mach-Zehnder configuration. This setup enables optimal use of SBP, providing up to a fourfold increase in effective acquisition throughput. In addition, we pose the scattering and aberrations as an overdetermined system, allowing a closed-form solution of an alternating least squares problem. The resulting iterative optimization is equivalent to the state-of-the-art closed-loop accumulation of single scattering (CLASS) algorithm. We demonstrate the setup and reconstruction method in lens tissue, a PSF phantom, a USAF target, and a model eye. The resulting digital confocal volumes provide locally optimized focusing and detection quality, and show improved resolution and signal-to-noise ratio (SNR) compared to conventional FF-FD-OCT. The technique combines the advantages of FF-FD-OCT and confocal OCT while mitigating the limitations of both approaches.
Introduction
Optical coherence tomography (OCT) is a non-invasive, label-free imaging technique used to create volumetric cross-sections of semi-transparent samples with micrometer resolution. Initially conceived in a confocal time-domain (TD) configuration, it used point-scanning and low-coherence interferometry to enable imaging over depth [1]. Acquisition speed and SNR saw major improvements with the development of Fourier-domain (FD) OCT, which acquired scattering information from all depths simultaneously [2, 3, 4]. A further speed increase was enabled by full-field (FF) OCT, which replaced point-scanning with a global illumination of the area of interest (at the cost of confocality), captured in parallel using a camera [5, 6]. When combined with FD-OCT, the absence of scanning and advancements in camera technology allowed for unprecedented acquisition speeds, making it particularly suitable for certain in vivo applications [7].
Despite these advances in technology, aberrations still pose a challenge in OCT imaging. Aberrations may be induced by both the optical setup and the object (especially deep inside the medium), causing the in- and outgoing waves to be deflected from their ideal single-scattered path [8, 9]. The multiply-scattered light thus prevents optimal delivery and collection of light inside the object. Consequently, image quality suffers due to a disturbed point-spread function (PSF) and lower signal-to-noise ratio (SNR). Confocal systems are more resistant to PSF disturbances than full-field systems, since the confocal gate allows partial rejection of out-of-focus and multiply-scattered light [8, 10]. However, the confocal gate becomes less effective in the face of strong scattering due to an inability to maintain ideal overlapping illumination and detection pinholes.
To optimize light delivery and detection quality, the field of adaptive optics (AO) has been developed. Here, wavefront shaping is physically implemented using spatial light modulators (SLM) or deformable mirrors in the illumination and/or detection arms [9]. It aims to reshape the wavefront to cancel present aberrations. However, this requires the presence of a known guide star (a term originating from astronomy) whose intensity can be optimized, which is often not the case, especially in in vivo applications [11, 12]. In addition, hardware-based AO generally requires elaborate setups, closed-loop control, and is typically limited to a single isoplanatic patch [13].
Computational adaptive optics (CAO) tries to overcome these limitations by correcting for wavefront errors in post-processing. Importantly, this requires the acquired data to be laterally phase-stable. Full-field systems uniquely satisfy this requirement since all lateral points are acquired simultaneously, enabling corrections at every depth and over multiple isoplanatic patches [7, 14]. Though CAO is possible in confocal OCT systems [15, 16, 17], they are inherently sensitive to motion-induced phase errors and may require sophisticated motion-correcting schemes. In its basic implementation, CAO is performed by using a single image and can provide (in vivo) aberration-free imaging by optimizing a guide star [18, 19], image metrics [7, 17, 20] or by using split aperture methods [21, 22, 23].
The most capable form of CAO has been demonstrated using the framework of reflection (and analogous transmission) matrix imaging (RMI), which has been developed over the last decade following work in ultrasound [24, 25, 26, 27]. It stores a set of (complex) wide-field images acquired under different illuminations inside a reflection matrix, which links the inputs and outputs of the imaging system. Essentially, it attempts to acquire the full response from the sample for any input state. Common illumination schemes include point-illumination, (pseudo-)random patterns or, as in this work, oblique illumination (i.e., angled plane waves). Various algorithms can then be used to solve and correct for the scattering behavior, achieving an ideal undisturbed PSF in post-processing and diffraction-limited reconstruction quality.
Most work in RMI has been performed in off-axis holographic setups using low-coherence interferometry, constituting a type of time-domain OCT or microscopy (OCM) [28, 29, 30]. More recent studies have demonstrated RMI in on-axis swept-source FF-FD-OCT, achieving much higher acquisition speeds than time-domain systems [31, 32]. Like much preceding work in time-domain, their Michelson interferometer synchronously scans the reference pupil with the illumination pupil. Consequently, the reference wave is tilted on the camera, together with the illumination incident on the object. In time-domain systems, this improves overlap between the object and reference coherence gate, maintaining interference contrast [33]. In swept-source FD-OCT, the longer instantaneous temporal coherence length makes the synchronous scanning not strictly necessary and is thus only the result of the choice of setup.
However, a synchronous scan of the reference pupil comes at the cost of space-bandwidth product (SBP). This dimensionless quantity is a measure of the information contained in a measurement, and is the product of the acquired FOV (space) and spatial frequencies (bandwidth) of an object [34, 35]. The tilted reference wave induces a phase gradient on the camera, which increases the detected signal’s frequencies. This increase provides no extra information about the object, but does raise the required bandwidth for Nyquist-sampling of the signal. Consequently, not all available bandwidth is used to detect the object’s spatial frequencies, resulting in a suboptimal SBP. If bandwidth were used more efficiently, more object information could be retrieved using the same number of camera pixels.
In this paper, we present an oblique illumination FF-FD-OCT system using a fixed on-axis reference wave in a Mach-Zehnder interferometer. This allows the sampling bandwidth to equal the bandwidth of the object signal. By making full use of the available bandwidth, a
Methods
Setup and data acquisition
The oblique illumination-capable FF-FD-OCT is configured as a Mach-Zehnder interferometer and is shown in Figure 1. Light from the tunable light source (SLDSources.com Broadsweeper BS-840-2-HP) is split by a fiber coupler into object and reference arm. In the object arm, the light is collimated and illuminates a dual-axis galvanometric scanning mirror (ScannerMAX Saturn 9B) that is imaged onto the object using a system. This allows changing the illumination angle on the sample without changing the location of the illuminated area. The backscattered light is imaged via a beam splitter onto the camera. The reference arm contains a separate collimator and the light is recombined via the beam splitter and directed onto the camera. The tunable laser source sweeps from 880 nm to 805 nm. While the laser sweeps, a high-speed camera (Phantom TMX 6410) records the interference between the reference and scattered light from the object. Per sweep and scanner position, 500 interferograms are recorded using pixels of the sensor at 65 000 fps, at an effective A-line rate of 73 MHz or 18.3 gigavoxels per second. In total 128 volumes per sample are acquired, each with a different scanner position, using a Fermat spiral pattern [36]. In addition, 128 volumes were acquired for each object without oblique scanning to compare results to conventional FF-FD-OCT. All data were acquired at an NA of 0.13. The setup is synchronized using an Arduino Giga R1 WiFi, delivering frame-triggers to the camera, triggers to the laser to initiate sweeps, and digital position commands to the scanner between volumes.
Figure 1. Schematic of the oblique illumination FF-FD-OCT system. Galvo mirrors are imaged onto the sample, allowing the angle of illumination on the object to be changed. The independent reference collimator provides a fixed reference wave on the camera.
Detected signal
In this paper, we use the following conventions: for the position-space vector , for its Fourier conjugate and for the unit vector corresponding to . Furthermore, and denote the corresponding projections onto the and plane, respectively. We furthermore use as the Fourier transform of a function .
OCT reconstruction
Oblique illumination induces a phase gradient at every depth in the object, where denotes the th illumination vector at illumination angle and wavenumber . The field detected at the camera plane coming from the object is given by
Here, is the spectrum of the source, the amplitude of the object wave field, the three-dimensional scattering potential of the object, and denotes the two-dimensional convolution with respect to . We further define , , and the pupil functions are given by
with describing the phase error of the input and output aberration, and describing the numerical aperture (NA) of the input and output. While and should in principle be identical, misalignment and motion-induced global phase changes can show up as input aberrations, making a separation crucial. The reference wave is delivered by a fixed reference collimator with amplitude , at a total distance corresponding to the zero-delay line:
In FD-OCT, interferograms are recorded as a function of the wavenumber . This gives the expression
The first two terms are the autocorrelation and DC terms. The last two terms are the signal and the conjugate signal. Concentrating on the signal term and assuming, without loss of generality that , this becomes
with .
Taking the lateral Fourier transform of the signal term and applying the convolution theorem as well as the Fourier shift theorem reveals that the object’s lateral -space shifts inside the detection pupil as the illumination is rotated:
With describing a constant band-pass and being shifted by the illumination , the setup acquires spatial frequencies that are inaccessible with conventional full-field setups [37]. OCT depth-reconstruction is performed by Fourier transforming along the -axis. Assuming the NA is sufficiently small such that and , we obtain
where is the axial PSF. Eq. (9) gives a sheared version of the object’s scattering potential, i.e., different lateral locations from the same layer of the object will be shifted axially by . Physically, this arises due to lateral differences in optical path length (OPL) between the object and reference wave at the camera plane, induced by the oblique illumination.
To compensate for this effect, we applied the conjugate of the original phase gradient to the axial spectrum Eq. (6) before the Fourier transform along the -axis:
This step also compensates the lateral shift in the object’s -space (Eq. (7)), aligning it between measurements.
By summing over the shear-corrected fields in post-processing, we create what we will refer to as a digital confocal volume,
where we used the approximation that would become accurate if densely scanned the pupil. The summation over plane wave illuminations creates a virtual illumination spot at every point of the resulting volume. At the same time, the camera pixels approximate point-detection. These two combine to create a confocal filter in post-processing. The resulting PSF is now the multiplication of the (virtual) mirrored input and output PSF, improving image quality accordingly. In practice, the limited number of illuminations prevents focusing to a single spot. More importantly however, aberrations need to be addressed to obtain optimal overlapping virtual illumination and detection pinholes and achieve maximal image quality.
Aberration correction
A single acquired field obtained for a single incident wave vector contains a mix of both object information and aberrations . This makes the problem of separating the two from a single field fundamentally underdetermined and ill-posed. In our full data, each illumination provides partially overlapping object information while aberrations remain constant, effectively overdetermining the system. In principle, this allows us to separate and solve for object and aberrations objectively, requiring no assumptions or priors. In the following, we omit the depth-index for readability. We can formulate the solution as a least squares problem in -space:
Here, denotes the lateral Fourier transform of the measured fields (from hereon referred to as apertures) after shear correction, functioning as the observations for the model. These apertures are described in terms of a constant amplitude , and the unknown scattering potential , and pupils and to be determined via optimization. The model assumes single scattering.
An efficient way to determine the unknown functions is to use alternating least squares: setting the gradients with respect to , , to 0 while keeping the others fixed and solving for the respective parameters one after the other. Since this analytically determines the extremum in a subspace, it will iterate towards the optimum. This leads to the following update rules, where is the current iteration:
In practice, it was found that modifying (16) such that the output pupil correction was not applied led to better convergence for scattering samples. This is likely due to the large number of degrees of freedom in strongly affecting the small number of degrees of freedom in , which may be problematic if initial estimates are noisy. The following form was therefore used in practice for scattering samples:
In addition, we omit the normalization in the denominators. This leads to a weighting of based on the number of apertures covering each pixel. This makes the subsequent determination of the pupil functions less sensitive to low-amplitude phasors. In addition, it leads to an apodization which suppresses higher frequencies and approximates the transfer function of a confocal system. For the pupil functions, we ultimately correct the observed apertures using the phase functions , whose phase is unaffected by normalization of the pupil functions.
Figure 2 shows a schematic representation of how the estimates , and are determined during the first iteration when no initial aberration correction is applied. We start at (i) with the shear-corrected fields. Per illustration, one on-axis () and two oblique images are shown. Their corresponding apertures are aligned in zero-padded -space. The best estimate of the scattering potential (Eq. (15) for ) is created in (ii) by summing over all the illuminations. Notably, is the lateral Fourier transform of the digital confocal image in Eq. (13), showing a larger NA than the individual apertures. The corresponding image appears less aberrated than any individual image due to the multiplication of the output and input PSF suppressing sidelobes. In (iii) the underlying unknown functions are shown which make up our observed fields and which we try to estimate. The underlying effective pupil function of is the sum of the shifted pupil functions: . This summation averages out fluctuations in the aberrated phase of the individual apertures, equivalently explaining why appears less aberrated.

Figure 2. Schematic showing the principle behind estimating object and aberrations during the first iteration. Shear-corrected apertures (i) are summed to create the estimate (ii). Its effective pupil shows less aberrated phase than the individual pupils (iii). Subsequent phase-comparison between the apertures and shows the difference in aberrations (iv). Summing over estimates (v), while summing over the input-corrected illuminations estimates (vi).
To estimate the aberrations, we first compare the phases of the apertures to in (iv). Since the underlying object’s -space is identical (as seen in (iii)), we observe only the phase difference between the aberrations of the apertures and the effective pupil of . Critically, since the aberrations in the effective pupil appear roughly constant in phase, the differences approximate the output aberration, up to global phase offsets determined by the input aberrations. In (v), we can therefore estimate by summing over phasors (Eq. (16)). We then estimate in (vi) by summing over the illuminations instead, using to remove the global phase offsets (Eq. (17)). Note that for this summation, the phase differences must be centered.
By correcting the apertures using these estimates during the next iteration, the effective aberration of is further reduced, improving the estimates of in turn. The process is repeated until convergence is reached. The optimization can be performed for smaller patches in case of anisoplanatic scattering. The phase of can only be determined up to a piston, tip and tilt [28, 38], since any gradient in can be compensated by a gradient in , resulting in the same loss value. To prevent stitching artifacts between patches, these gradients are removed by centering the center-of-mass of 4.
-registration
The values of need to be known accurately to allow for optimal convergence and image formation. We approach this as a registration problem, treating the apertures at a common depth as laterally shifted images. This avoids relying on calibration measurements, which may not be possible in in vivo imaging. First, an initial shear compensation is applied using theoretical values derived from system specifications and scanner lookup table. Volumes are then axially registered to correct path-length variations induced by dual-axis galvo scanning, which lacks a single pivot point. Apertures are subsequently extracted from a common layer showing strong signal, typically near the surface. Relative shifts for all apertures are then determined using phase correlation on the amplitudes only. The shift values are optimized using a weighted least squares fit, similar to GraphSLAM [39]. Weights are given by the phase correlation peak value, or set to zero if deviations exceed 15 pixels of the relative theoretical shifts.
Comparison to synchronously scanned reference pupil
Previous work in oblique illumination using on-axis FF-FD-OCT placed a mirror in the reference arm, creating a Michelson interferometer. This causes the reference pupil to be scanned synchronously with the object pupil [31], such that the reference wave induces a phase gradient on the camera (Figure 3b).
Figure 3. Schematic comparing a system with a fixed reference arm (ours, a) to a synchronously scanned reference pupil ([31], b). In the latter, the phase gradient from the reference on the camera shifts the spatial frequencies after passing through the detection pupil. This increases the required sampling bandwidth, requiring the detection NA to be lowered to prevent aliasing. DC-component of scattered field shown in blue dotted line.
Space-bandwidth product
This Michelson configuration leads to a much lower SBP compared to a fixed reference. Figure 3 compares the setup and sampled frequencies of a setup using a fixed reference and a synchronously scanned reference. The sampling bandwidth is defined by , with the pixel size of the camera and the magnification from object to camera. The detection pupil transmits frequencies up to . A fixed on-axis reference does not modulate the detected frequencies, requiring for sampling at Nyquist (Figure 3a). This allows optimal use of the available bandwidth.
In contrast, the added phase gradient in a synchronously scanned pupil adds a carrier frequency to the frequencies passing through the pupil (Figure 3b). This increases the required sampling bandwidth to , with a full scan of the pupil requiring in both axes. This requirement can be fulfilled by either decreasing or FOV by a factor of two in each direction. In either case, a synchronously scanned reference pupil reduces SBP by a factor of four compared to our setup, requiring four times more pixels in each camera image compared to our methods. In practice, this will reduce imaging speed by a factor of four assuming a constant pixel digitization rate.
Reconstruction
When scanning the reference pupil, the phase gradient on the camera is -dependent, given by:
At the same time the object arm is the same as in Eq. (1). The object and reference now maintain a common zero-delay line, such that the OCT reconstruction does not appear sheared when transforming over the -axis.
Here, the integral in square brackets is the object term, which only interacts with the reference field after having passed through the detection pupil. The final reconstruction in Eq. (21) is identical to our shear-corrected reconstruction in Eq. (11). Our shear correction is thus the computational counterpart to physically scanning the reference pupil. However, whereas the synchronous scanning requires a larger sampling bandwidth compared to the detection pupil to prevent aliasing, we can zero-pad our -spaces in post-processing to allow the apertures to shift without aliasing. We thus arrive at the same aligned apertures, however, containing up to four times the FOV or resolution.
Relation to CLASS
The aberration correction presented here is equivalent to CLASS [28]. CLASS aims to maximize the coherent sum of the aberration-corrected fields, showing that it is only maximal when all aberrations have been fully corrected [28]. This method was introduced to improve the previously-developed collective accumulation of single-scattered waves (CASS) image, which summed the fields without performing any corrections [40]. Our initial unoptimized estimate (Eq. (15) for , which has no corrections applied) thus equals the CASS image. We then correct the observed fields such that we minimize the difference with respect to the common object . This (partially) phase-aligns the corrected observed fields, since if they were not phase-aligned after correction, they would not correspond to the same object. However, the sum of coherent corrected fields is only maximal if they are phase-aligned. Therefore, both methods ultimately arrive at the same optimization procedure, with the CLASS-image being the improved estimate (Eq. (15)) [29].
Results
We first demonstrate the least squares optimization in a lens tissue (MC-5, Thorlabs). Figure 4a shows the unoptimized digital confocal image . The optimization typically converged after a couple of iterations (at which point aberrations were unchanging), but was run for 100 iterations globally and 20 locally. After iterating, the optimized digital confocal (ODC) enface shows improved image quality and higher SNR (Figure 4b, note the change in dynamic range shown in the image). The corresponding determined aberrations are shown in Figure 4c. Although shows only a slight defocus, appears speckle-like. By optimizing over smaller isoplanatic patches (198 × 198 m, 49 m overlap) we obtain a locally optimized digital confocal (LODC) enface, showing consistent image quality over the full FOV (Figure 4e). The corresponding spatially varying output aberrations are displayed in Figure 4f. Compared to a conventional FF-FD-OCT, consisting of a phase-aligned, coherent average of on-axis volumes having undergone the same corrections (Figure 4d), the (L)ODC enface shows strong improvement in both resolution and SNR.

Figure 4. Lens tissue. a,b. Digital confocal image before and after optimization. c. Corresponding estimated input and output pupil functions used for b. d. Coherent average of locally corrected on-axis volumes e. Locally optimized digital confocal image. f. Local estimated input and output pupil functions.
Next, we evaluated performance over depth by imaging a PSF phantom (OCTPSF01, BioPixS).
Input and output aberrations were first estimated at the surface, and the corresponding corrections were subsequently applied to every depth of the volume. A sliced 3D view of the resulting ODC volume is shown in Figure 5a. Though strong signal is now obtained from the surface, point scatterers blur with increasing depth, as seen in the unoptimized enfaces in Figure 5c and f. The volumes were then corrected sequentially from the surface downwards, using the pupil functions estimated at depth as initial estimates for depth . The recovered input and output aberrations at in-medium (refractive index ) depths of 0.3 mm and 0.6 mm are shown in Figure 5d and Figure 5g, respectively, showing a depth-dependent defocus. The off-center defocus in the input aberration indicates a misalignment between the illumination and detection pupil. After correction, strong signal is recovered throughout the ODC volume (Figure 5b), with scatterers now resolvable at all depths (Figure 5e and Figure 5h).

Figure 5. PSF phantom. a. 3D-sliced digital confocal volume, optimized using only the top layer. b. 3D-sliced digital confocal volume optimized at every depth. c,f Unoptimized enfaces at depths of 0.3 mm and 0.6 mm of volume shown in a.. e,h ODC enfaces at depths of 0.3 mm and 0.6 mm of volume shown in b.. d, g. Estimated input and output pupil functions at depths 0.3 mm and 0.6 mm, respectively.
We investigated potential translation to retina by imaging a model eye (Modell-Augen Manufaktur). For this acquisition, the Fourier transform of the detected object field was set to coincide with the model eye’s pupil since most aberrations are expected to occur there. This was accomplished by moving the lens in the reference arm, inducing a quadratic phase on the image plane, and propagating the Fourier transform in turn. Any residual misalignment between the two planes is corrected computationally [41]. We determined the computational propagation distance by running the algorithm for several values and subsequently selecting the value that maximizes the total energy of the ODC image .
Figure 6a compares the theoretical values of based on the scan protocol to the values obtained using our -registration. Figure 6b and Figure 6c show the output pupil functions estimated using the most reflective layer shown (marked in Figure 6j), using the theoretical values and registered values of , respectively. The resolution in the estimated aberration is improved when using the registered values. The unoptimized enface is shown in Figure 6d. The ODC enface optimized using the registered (Figure 6f) shows improved SNR and reveals the model eye’s texture compared to using the theoretical values (Figure 6e). The estimated input and output pupil functions were applied to each layer and the volumes subsequently refocused over depth. To this end, we manually determined the quadratic phase required to correct defocus at a distant layer and interpolated this correction to all other layers.

Figure 6. Model eye. a. Comparison of theoretical and registered values. b,c. Output pupil functions estimated using the theoretical and registered values, respectively, using layer shown in d. e,f. Corresponding optimized digital confocal (ODC) enfaces using theoretical and registered values. g. Estimated local output pupils for ODC enface shown in h. i. Correction shown in g. applied to conventional FF-FD-OCT enface. j,k. B-scan of locally optimized digital confocal (LODC) volume and same corrections applied to conventional FF-FD-OCT volume.
Next, we iterated through all layers and performed local corrections over smaller isoplanatic patches (99 µm × 99 µm, 25 µm overlap). Figure 6g shows one such map of local output aberrations varying over the FOV. Figure 6h shows the corresponding LODC enface. The local optimization allows us to more clearly resolve scatterers (white inset), but pulls up noise in empty regions (red inset, displayed with increased dynamic range). Figure 6i shows the same corrections applied to the conventional FF-FD-OCT data. The LODC image has better SNR and reveals scatterers that are otherwise not visible. Similar improvements can be seen when comparing an LODC B-scan (Figure 6j) to a conventional FF-FD-OCT B-scan having undergone the same corrections (Figure 6k).
Lastly, we demonstrated improvement in lateral resolution in a USAF target. Figure 7a and Figure 7b show a zoomed-in view of the conventional FF-FD-OCT enface and the ODC enface, respectively. Here, the normalization of Eq. (15) was not applied so that higher frequencies were not suppressed, allowing the resolution improvement to be assessed. Figure 7c and Figure 7d show the mean horizontal line intensity profiles of the boxes marked in Figure 7a and Figure 7b, corresponding to Elements 2, 3 and 4 for Group 7 and Group 8, respectively. The theoretical resolution of the system is given by µm, corresponding to Group 8 Element 3 (1.55 µm), which is indeed the smallest element that can still be resolved in the ODC enface. This is the expected doubling of the lateral resolution compared to using conventional FF-FD-OCT, where the smallest resolvable element is Group 7 Element 3 (3.10 µm).
Figure 7. USAF target showing doubling of the lateral resolution. a, b. Zoomed-in view of USAF target for conventional FF-FD-OCT and optimized digital confocal (ODC) enface, respectively. c, d. Average horizontal intensity in boxes shown in a. and b., respectively.
Processing was performed on a computer with 128GB of RAM and an NVIDIA RTX 4090 (24GB VRAM). Initial reconstruction, -registration, shear correction and alignment of all volumes are performed in under 3 minutes. Subsequent local optimization can be performed efficiently by processing many isoplanatic patches in parallel on GPU. Optimization of the full digital confocal volume for the model eye (90 layers, 2304 isoplanatic patches of 32 × 32 pixels per layer, 20 iterations) took under 19 minutes.
Discussion
We demonstrated the effectiveness of the algorithm in various settings. The optimized digital confocal (ODC) images show substantial improvements over the conventional FF-FD-OCT images. For ODC image quality, correcting input aberrations proved as important as correcting output aberrations. In our setup, motion gives a random phase to each volume, which prevents coherent summing of the single scattering component in the resulting reconstructed images and is in principle indistinguishable from input aberrations. Considering the weak output aberrations
in the lens tissue, the improvement from Figure 4a to b is almost entirely attributable to correction of input aberrations, yielding a 15 dB SNR gain. After the global correction of aberrations, volumes were mutually phase-stable. This is seen in the PSF phantom, where, after correcting the top layer, we find a clear depth-dependent defocus. This also indicates that the optimization finds physically consistent aberrations.
The ODC images showed a marked SNR improvement compared to the on-axis coherent averages (compare Figure 4d and Figure 4e, Figure 6h and Figure 6i, Figure 6j and Figure 6k). This is unexpected since a phase-matched coherent average should provide the same SNR gain. We attribute the improvement to better rejection of autocorrelation and multiple-scattering noise. In on-axis coherent averaging, this noise adds constructively across phase-stable volumes and therefore scales similarly to the single-scattered signal. This leads to suboptimal improvement in SNR, which is theoretically only limited by Poisson-distributed shot noise. Under oblique illumination, however, the phase of multiply scattered and autocorrelation noise varies with the illumination, leading to partial cancellation during coherent summation. Consequently, oblique illumination can potentially suppress non-shot-noise contributions more effectively than conventional FF-FD-OCT. However, this advantage is lost when single-scattered signal is absent or very weak. This is observed in the model eye (Figure 6h, red inset), where local optimization increases the noise floor in the empty regions. This is caused by the algorithm attempting to maximize the coherent sum, without distinguishing signal from noise. This might be mitigated by introducing sparsity-promoting regularization, or by detecting speckle-like output aberrations and replacing them with estimates from neighboring patches.
Small features were much better visible in the ODC images. In the lens tissue, fibers were more clearly separated (Figure 4d and Figure 4e), likely due to improved lateral resolution. In the model eye, many additional scatterers became visible (Figure 6h and Figure 6i, white insets). Because these scatterers are comparable in size to those already visible in the on-axis image, it is unclear to what extent their improved visibility arises from enhanced resolution or increased SNR.
The least squares optimization is susceptible to convergence to local minima, maxima or saddle points, or a local gauge freedom in the reached point. In the PSF phantom, even after having corrected the full volume using the surface aberrations, aberrations at maximal depth could not be recovered without an initial estimate, instead converging to a speckle-like solution. The algorithm is thus not guaranteed to find the desired minimum when signal is weak or if the current solution is far away from this minimum. By working sequentially downwards using as the initial estimate for , we remain close to the minimal loss value. This allowed convergence even at maximum depth.
Non-isoplanatic aberrations introduce additional minima when estimating the global pupil functions. This was observed in the model eye, where global estimation of the pupil functions was only successful when performed using the reflecting layer (marked in Figure 6j), which had most energy concentrated in a smaller number of isoplanatic patches.
Obtaining optimal image quality over the full FOV required a subsequent local optimization. The resolution of the aberration is determined by the size of the isoplanatic patch. Patch sizes must thus be large enough to capture the local PSF distortion, but small enough to allow a smooth transition from one patch to the next to minimize stitching artifacts. An initial global correction and volumetric refocusing are therefore advantageous, as they reduce PSF sidelobes and bring all patches closer to their local minimal loss values. However, small patches remain sensitive to noise and strong local reflections. The latter can be seen in Figure 4f, where a highly reflective fiber at the center produces local aberration estimates that unexpectedly deviate from neighboring patches.
Accurate registration was essential for obtaining optimal image quality in the model eye. Notably, even small deviations from the theoretical values degraded the estimated output aberration. This will be even more important in in vivo retinal imaging. Incorrect values mix phase and amplitude from different parts of the object’s -space when constructing , preventing optimal coherent summation of the apertures. Because the object’s phase is effectively random, even small errors strongly disturb its phase, particularly when considering larger patches due to the higher resolution of each pixel in -space. When subsequently estimating the pupil functions, the object’s phase does not cancel when calculating the terms (Eqs. (16) and (17)), which introduces phase noise.
Creating the oblique-illumination-capable FF-FD-OCT system required few hardware modifications, needing only a galvo and a synchronization device. However, some features of conventional FF-FD-OCT are lost. First, this setup is incompatible with multimode illumination due to its Mach-Zehnder configuration using a fiber coupler. In addition, the scanning changes the phase front of the illumination between volumes beyond the expected phase gradient since each collimated beam passes through different regions of the lenses. Furthermore, strong in vivo sample motion will uniquely disturb the phases contained in each aperture, which will need to be addressed to make aberration correction feasible. Finally, the high volumetric rate that allows FF-FD-OCT to capture dynamic changes, such as optoretinography [42] or dynamic OCT [43], is lost to a certain extent.
Conclusion
FF-FD-OCT offers a substantial speed advantage over point-scanning OCT but lacks a confocal gate. Previous work has demonstrated how oblique illumination may be used in full-field systems to synthesize a confocal volume in post-processing. We demonstrated such an obliqueillumination FF-FD-OCT system in a Mach-Zehnder configuration. Compared with previous work using a Michelson interferometer, our setup samples the object’s spatial frequencies more efficiently, achieving an approximate fourfold improvement in SBP. Combined with a high-speed camera, this enabled volumetric acquisition at 18.3 gigavoxels per second. An illumination-dependent shear introduced by the separation of the object and reference arms was corrected computationally. In addition, an accurate registration of the illumination wave vectors was required in post-processing for optimal image formation.
We addressed aberrations by posing the scattering behavior as an overdetermined system, which could be leveraged to solve jointly for object, input aberrations and output aberrations, without requiring priors or assumptions. To this end, we formulated a closed-form solution of an alternating least squares problem, resulting in an iterative optimization algorithm. Notably, this algorithm is equivalent to the state-of-the-art CLASS algorithm, although it was derived without reference to the reflection matrix framework.
Using this algorithm, we reconstructed digital confocal volumes with globally, depth-dependent, and locally optimized focusing quality in various objects. The resulting images showed substantial improvements in resolution, contrast, and SNR. The reconstructed volumes also showed markedly higher SNR compared to identically corrected conventional FF-FD-OCT data due to better rejection of multiple-scattering and autocorrelation noise. However, the optimization was susceptible to local minima in regions with weak signal or strong anisoplanatism. Expansion of the scattering model or adding regularization may thus be required to improve robustness. In the future, our highly efficient acquisition of oblique illumination datasets at large FOV and high resolution may open up opportunities for applications in human in vivo imaging.
Funding. Dutch Research Council (NWO, OCENW.M.22.087 and OCENW.M.23.316). European Research Council (ERC, FunMicroRetina, grant agreement No. 101170453).
Acknowledgment. We would like to thank ASML for lending us the high-speed camera, and Johannes F. de Boer for useful discussions.
Disclosures. The authors declare no conflicts of interest.
References
- [1]D. Huang, E. A. Swanson, C. P. Lin, et al., “Optical Coherence Tomography,” Science 254, 1178–1181 (1991).
- [2]J. F. d. Boer, B. Cense, B. H. Park, et al., “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett. 28, 2067–2069 (2003).
- [3]R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, “Performance of fourier domain vs. time domain optical coherence tomography,” Opt. Express 11, 889–894 (2003).pubmed.ncbi.nlm.nih.gov/19461802
- [4]S. H. Yun, G. J. Tearney, J. F. d. Boer, et al., “High-speed optical frequency-domain imaging,” Opt. Express 11, 2953–2963 (2003).pubmed.ncbi.nlm.nih.gov/19471415
- [5]E. Beaurepaire, A. C. Boccara, M. Lebec, et al., “Full-field optical coherence microscopy,” Opt. Lett. 23, 244–246 (1998).
- [6]B. Povazay, A. Unterhuber, B. Hermann, et al., “Full-field time-encoded frequency-domain optical coherence tomography,” Opt. Express 14, 7661–7669 (2006).
- [7]D. Hillmann, H. Spahr, C. Hain, et al., “Aberration-free volumetric high-speed imaging of in vivo retina,” Sci. Reports 6, 35209 (2016).arxiv.org/abs/1605.03747
- [8]V. Ntziachristos, “Going deeper than microscopy: the optical imaging frontier in biology,” Nat. Methods 7, 603–614 (2010).
- [9]J. Bertolotti and O. Katz, “Imaging in complex media,” Nat. Phys. 18, 1008–1017 (2022).DOI
- [10]A. Badon, A. C. Boccara, G. Leresoy, et al., “Multiple scattering limit in optical microscopy,” Opt. Express 25, 28914–28934 (2017).
- [11]M. J. Booth, “Adaptive optical microscopy: the ongoing quest for a perfect image,” Light. Sci. & Appl. 3, e165–e165 (2014).DOI
- [12]N. Ji, “Adaptive optical fluorescence microscopy,” Nat. Methods 14, 374–380 (2017).DOI
- [13]J.-H. Park, Z. Yu, K. Lee, et al., “Perspective: Wavefront shaping techniques for controlling multiple light scattering in biological tissues: Toward in vivo applications,” APL Photonics 3, 100901 (2018).DOI
- [14]A. Kumar, W. Drexler, and R. A. Leitgeb, “Numerical focusing methods for full field OCT: a comparison based on a common signal model,” Opt. Express 22, 16061–16078 (2014).DOI
- [15]T. S. Ralston, D. L. Marks, P. S. Carney, and S. A. Boppart, “Interferometric synthetic aperture microscopy,” Nat. physics 3, 129–134 (2007).
- [16]N. D. Shemonski, F. A. South, Y.-Z. Liu, et al., “Computational high-resolution optical imaging of the living human retina,” Nat. Photonics 9, 440–443 (2015).pubmed.ncbi.nlm.nih.gov/26877761
- [17]S. G. Adie, B. W. Graf, A. Ahmad, et al., “Computational adaptive optics for broadband optical interferometric tomography of biological tissue,” Proc. National Acad. Sci. 109, 7175–7180 (2012).DOI
- [18]S. G. Adie, N. D. Shemonski, B. W. Graf, et al., “Guide-star-based computational adaptive optics for broadband interferometric tomography,” Appl. Phys. Lett. 101, 221117 (2012).pubmed.ncbi.nlm.nih.gov/23284179
- [19]C. Liu, X. Yu, and M. K. Kim, “Phase aberration correction by correlation in digital holographic adaptive optics,” Appl. Opt. 52, 2940–2949 (2013).DOI
- [20]T. S. Thurman and J. R. Fienup, “Phase-error correction in digital holography,” JOSA A 25, 983–994 (2008).pubmed.ncbi.nlm.nih.gov/18382499
- [21]L. Ginner, A. Kumar, D. Fechtig, et al., “Noniterative digital aberration correction for cellular resolution retinal optical coherence tomography in vivo,” Optica 4, 924–931 (2017).DOI
- [22]A. Kumar, W. Drexler, and R. A. Leitgeb, “Subaperture correlation based digital adaptive optics for full field optical coherence tomography,” Opt. Express 21, 10850–10866 (2013).DOI
- [23]D. Hillmann, C. Pfäffle, H. Spahr, et al., “Computational adaptive optics for optical coherence tomography using multiple randomized subaperture correlations,” Opt. Lett. 44, 3905–3908 (2019).DOI
- [24]A. Aubry and A. Derode, “Detection and imaging in a random medium: A matrix method to overcome multiple scattering and aberration,” J. Appl. Phys. 106, 044903 (2009).DOI
- [25]W. Lambert, “Reflection Matrix Approach for Quantitative Imaging of Scattering Media,” Phys. Rev. X 10 (2020).DOI
- [26]S. M. Popoff, G. Lerosey, M. Fink, et al., “Controlling light through optical disordered media: transmission matrix approach,” New J. Phys. 13, 123021 (2011).
- [27]A. Badon, D. Li, G. Lerosey, et al., “Smart optical coherence tomography for ultra-deep imaging through highly scattering media,” Sci. Adv. 2, e1600370 (2016).
- [28]S. Kang, P. Kang, S. Jeong, et al., “High-resolution adaptive optical imaging within thick scattering media using closed-loop accumulation of single scattering,” Nat. Commun. 8, 2157 (2017).DOI
- [29]S. Yoon, H. Lee, J. H. Hong, et al., “Laser scanning reflection-matrix microscopy for aberration-free imaging through intact mouse skull,” Nat. Commun. 11, 5721 (2020).DOI
- [30]U. Najar, V. Barolle, P. Balondrade, et al., “Harnessing forward multiple scattering for optical imaging deep inside an opaque medium,” Nat. Commun. 15, 7349 (2024).
- [31]P. Balondrade, V. Barolle, N. Guigui, et al., “Multi-spectral reflection matrix for ultrafast 3D label-free microscopy,” Nat. Photonics 18, 1097–1104 (2024).DOI
- [32]E. Giraudat, V. Barolle, F. Bureau, et al., “Label-free subcellular 3D imaging of oocytes and embryos via reflection matrix microscopy,” (2026). ArXiv:2410.11126 [physics.optics].arxiv.org/abs/2410.11126
- [33]M. Kim, Y. Jo, J. H. Hong, et al., “Label-free neuroimaging in vivo using synchronous angular scanning microscopy with single-scattering accumulation algorithm,” Nat. Commun. 10, 3152 (2019).DOI
- [34]A. W. Lohmann, R. G. Dorsch, D. Mendlovic, et al., “Space–bandwidth product of optical signals and systems,” JOSA A 13, 470–473 (1996).DOI
- [35]J. Park, D. J. Brady, G. Zheng, et al., “Review of bio-optical imaging systems with a high space-bandwidth product,” Adv. Photonics 3, 044001 (2021).DOI
- [36]X. Huang, H. Yan, R. Harder, et al., “Optimization of overlap uniformness for ptychography,” Opt. Express 22, 12634–12644 (2014).DOI
- [37]R. Heintzmann and T. Huser, “Super-Resolution Structured Illumination Microscopy,” Chem. Rev. 117, 13890–13908 (2017).
- [38]W. Lambert, L. A. Cobus, J. Robin, et al., “Ultrasound Matrix Imaging—Part II: The Distortion Matrix for Aberration Correction Over Multiple Isoplanatic Patches,” IEEE Trans. on Med. Imaging 41, 3921–3938 (2022).
- [39]S. Thrun and M. Montemerlo, “The Graph SLAM Algorithm with Applications to Large-Scale Mapping of Urban Structures,” The Int. J. Robotics Res. 25, 403–429 (2006).DOI
- [40]S. Kang, S. Jeong, W. Choi, et al., “Imaging deep within a scattering medium using collective accumulation of single-scattered waves,” Nat. Photonics 9, 253–258 (2015).DOI
- [41]K. Kwon, J. H. Hong, S. Kang, et al., “Computational conjugate adaptive optics microscopy for longitudinal through-skull imaging of cortical myelin,” Nat. Commun. 14, 105 (2023).
- [42]D. Hillmann, H. Spahr, C. Pfäffle, et al., “In vivo optical imaging of physiological responses to photostimulation in human photoreceptors,” Proc. National Acad. Sci. 113, 13138–13143 (2016).DOI
- [43]C. Apelian, F. Harms, O. Thouvenin, and A. C. Boccara, “Dynamic full field optical coherence tomography: subcellular metabolic contrast revealed in tissues by interferometric signals temporal analysis,” Biomed. Opt. Express 7, 1511–1524 (2016).