betaFeedback

WorksPhysics and Astronomy

First Limits on Light Dark Matter Interactions in a Low Threshold Two Channel Athermal Phonon Detector from the TESSERACT Collaboration

Abstract

We present results of a search for spin-independent dark matter-nucleon interactions in a 1 cm2^2 by 1 mm thick (0.233 gram) high-resolution silicon athermal phonon detector operated above ground. For interactions in the substrate, this detector achieves a r.m.s. baseline energy resolution of 361.5 ±\pm 0.4 MeV/c2c^2, the best for any athermal phonon detector to date. With an exposure of 0.233g ×\times 12 hours, we place the most stringent constraints on dark matter masses between 44 and 87 MeV/c2c^2, with the lowest unexplored cross section of 4 ×10−32\times 10^{-32} cm2^2 at 87 MeV/c2c^2. We employ a conservative salting technique to reach the lowest dark matter mass ever probed via direct detection experiment. This constraint is enabled by two-channel rejection of low-energy backgrounds that are coupled to individual sensors.

Introduction

Current and previous generation dark matter (DM) direct detection experiments such as LZ [1], XENONnT [2], DarkSide [3], and PandaX [4] have largely focused on the detection of weakly interacting massive particles (WIMPs), placing continually tightening constraints on DM models with masses above 1 GeV/c2c^2. Increasing interest [5, 6, 7] in searching for DM below a GeV/c2c^2 has motivated some collaborations, including CRESST [8, 9], EDELWEISS [10], SuperCDMS [11], DAMIC-M [12] and SENSEI [13], to develop experiments sensitive to sub-GeV DM masses. This mass regime contains models of thermal relic DM (e.g. ELDERs [14] and SIMPs [15]), freeze-in DM (e.g. FIMPs [16]), secluded DM [17, 18], Hidden Valleys [19], asymmetric DM [20], and supersymmetric hidden sectors [21, 22] that can either fully or partially produce the correct DM relic abundance. Directly detecting interactions from such DM candidates is challenging, given the eV scale nuclear recoils that they would create in a detector.

Previous efforts to search for nuclear recoils from sub-GeV DM interactions were impeded by an unknown low energy background or backgrounds broadly termed the “Low Energy Excess” (LEE) [23]. These backgrounds are hypothesized to be associated with material effects in the detector environment, such as stress relaxation in the detector holding structure [24] or aluminum sensor films [25], annihilation of radiation-induced defects in the detector crystal [26], or scintillation of materials around the detector [27].

Recently, a subclass of LEE backgrounds has been observed to strongly couple to sensor films from which the detector is constructed [28, 29, 30], allowing possible discrimination of this background from DM interactions via a coincidence requirement. Multiple athermal phonon readout channels, coupled to the same detector target, would respond simultaneously to athermal phonon bursts from possible DM interactions in the target, while the response to sensor-coupled backgrounds would primarily be confined to a single channel.

In this Letter, we employ this two-channel technique to obtain novel constraints on low-mass DM interactions. This work is part of the TESSERACT (Transition Edge Sensors with Sub-eV Resolution And Cryogenic Targets) initiative, which aims to search for models of sub-GeV DM using a suite of cryogenic detector materials and technologies optimized for low detector thresholds.

Detector

We search for low energy DM interactions in a 1 cm2^2 by 1 mm thick silicon athermal phonon detector, shown in Figure 1, which uses voltage-biased tungsten Transition Edge Sensors (TESs [31], Tc≈48T_{\mathrm{c}} \approx48 mK) coupled to aluminum athermal phonon collection fins in the common Quasiparticle-trap-assisted Electrothermal-feedback Transition edge sensor (“QET”) architecture [32] to sense phonon bursts in the silicon substrate. We wire 50 QETs into two channels of 25 QETs and read out each channel separately with DC SQUID array amplifiers. This read-out scheme was designed to discriminate backgrounds

The TESSERACT 1 cm² detector

Figure 1. The TESSERACT 1 cm2^2 detector used in this analysis [33]. The detector is supported by wirebonds attached at the top center and bottom corners of the detector. A gold wirebond (left side) is used to cool the detector. The two sensor channels (“left” and “right”) can be seen as the parallel lines, each biased and read out separately (see electrical wire bonds to readout PCB).

which couple primarily to QET metal films from possible DM interactions. These backgrounds were expected to deposit more energy in one readout channel than in the other, whereas DM interactions with the substrate would produce roughly equal responses in both channels. Backgrounds associated with stress relaxation in the detector mount were suppressed by suspending the detector by wire bonds as in Ref. [24]. We describe low energy background and noise observations in this detector in Ref. [28].

As we are primarily interested in searching for low energy DM interactions in the detector, we undertook no special precautions to isolate our detector from high-energy backgrounds (e.g. background radioactivity, cosmic rays). We operated our detector in a dilution refrigerator two floors below ground level at the University of California, Berkeley.

To calibrate our detector’s response, we injected short (∼μ\sim\mus) pulses of small numbers of 3.06 eV photons onto the detector, creating athermal pulses of quantized energy [28]. By observing the responses in the two channels, we measure the expected pulse shape for phonon DM-like events, and by combining the responses in the two channels measure a world-leading baseline phonon energy resolution of σP=361.5±0.4\sigma_P = 361.5 \pm0.4 meV (stat.).

Current signals from the two detector channels were continuously digitized at 1.25 MHz. Twenty-four hours of data were collected, interleaved with periods where the detector state was monitored to ensure stable operation. To mitigate bias in the analysis, the dataset was divided into 96 15-minute periods. Odd-numbered periods were unblinded and used to develop the analysis pipeline. Even-numbered periods were kept blinded and only opened after the analysis was finalized. Results from the blinded dataset are presented here.

Offline Trigger and Multi-Channel Optimal Filtering

Offline analysis is performed in two stages: triggering, in which event times are identified, and feature extraction, in which event type and energy are estimated. Both stages use Optimal (Matched) Filtering [34, 35], a class of algorithms designed to identify small signal pulses in a noisy data stream.

The main analysis challenge is that the QET response depends on where the energy deposition occurs:

  • For interactions in the substrate, where we expect >99%>99\% of DM recoils to occur, the energy is first carried by athermal phonons. These are absorbed in the Al fins creating quasi-particles, which eventually trap on the W TES sensor, heating it up. This 2-step process leads to a relatively slow rise time. Simultaneously, the ∼15%\sim15\% per-channel phonon collection efficiency suppresses the pulse height. For this class of phonon mediated events, we expect coincident pulses in the two channels with roughly the same amplitude. These events are “shared” between both channels.

  • In contrast, energy released in the fins of one channel is not coincident with any significant energy in the other channel. These “singles” pulses have much sharper rise times and do not suffer from phonon collection inefficiency. The combination of these two effects leads to much higher sensitivity. Our modeling indicates that the left and right channels are more sensitive by a factor ∼15\sim15 and ∼28\sim28, respectively.

To deal with this challenge,we have extended the classical Optimal Filter method to NN simultaneous readout channels and MM independent amplitudes.

Building an Optimal Filter

An Optimal Filter (OF) requires two elements: noise and signal models. The noise is assumed to be stationary (i.e., statistical properties such as moments are independent of time) and Gaussian (Gaussian distribution of Fourier amplitudes at each frequency and random phases, leading to no correlations between the Fourier components). The noise is therefore uniquely characterized by a covariance matrix between all channels at discrete Fourier frequencies. For the signal, at energies well below the saturation of our sensors, the normalized pulse shape (template) is independent of energy [31, 35, 36].

Given the independence of noise at different frequencies, we can construct a χ2\chi^2 estimator of signal amplitudes and pulse start times. Minimizing χ2\chi^2 is equivalent to maximizing likelihood, so this procedure is optimal for the identification and estimation of energy depositions down to the lowest possible threshold.

In order to simultaneously fit several sensors using a variety of signal shapes, we have designed an “N×MN \times M” Optimal Filter that simultaneously fits NN readout channels with channel-specific pulse shapes scaled by MM independent amplitudes Appendix § 1.2.1. A straightforward example is when N=MN=M; each energy deposition has a unique start time, but its amplitudes in the NN channels vary independently.

Implementing NxM Optimal Filters

For substrate events (“shared” or “phonon-like”), the best signal-to-noise ratio is obtained by imposing equal pulse amplitudes in both channels, with proper energy normalization of the pulse shapes (templates). This requirement is implemented in a 2×12 \times1 OF. Photon calibration events are used to measure normalized pulse shapes, which are cross checked against the shapes of background shared events and detailed sensor modeling [30, 35, 36].

By selecting “singles” events, we also empirically generate templates for aluminum fin events; sensor modeling is used to normalize pulse energy. We use a 2×12 \times1 OF to fit singles backgrounds, setting the pulse shape for the non-excited channel to a constant zero waveform. The singles OF uses the non-excited channel to estimate and optimally subtract correlated noise from the excited channel.

Our data acquisition is continuous (no online trigger). We use a 2×12 \times1 OF with shared pulse templates to identify events in the recorded data. Events are triggered when the OF amplitude exceeds 4σnoise4\sigma_{\mathrm{noise}}, corresponding to 1.45+0.021.45+0.02 eV, where σnoise\sigma_{\mathrm{noise}} is the baseline sensor noise. Continuous periods of time when the OF amplitude exceeds the trigger threshold are assembled into discrete events, using a maximum window of 2 ms. More details about the trigger can be found in Appendix § 1.2.1.

To extract data features, two types of optimal filters are deployed on each triggered event. First, we inspect energy partitioning between the two readout channels by fitting a 2×22 \times2 OF to each triggered event, with shared pulse templates. The best-fit amplitude (i.e., energy) in each channel is allowed to vary independently, while maintaining start time coincidence. As shown in Figure 2, the events divide into three primary populations: shared events consistent with calibrations and DM, on the x≈yx \approx y diagonal; left and right channel singles, on the xx and yy axes. Figure 2 also displays the average shared and singles pulse shapes in its inserts.

Figure 2. (Main) Two dimensional histogram of observed events. Black to orange points on the diagonal show accepted shared/phonon events; blue to purple points show singles events which are inconsistent with a DM response and excluded in the DM search. For energy reconstruction, we use a 2×22 \times2 OF and assume a shared pulse shape. As explained in the text this leads an overestimate of the single events, which cluster along the two coordinate axes. (Top Left Insert) Averaged shared/phonon event response in the left (blue) and right (red) channels for events in orange dashed box in main figure. (Bottom Right Insert) Averaged left singles event response in the left (blue) and right (red) channels for events in purple dashed box in main figure. In both insert figures, traces are offset and low pass filtered at 50 kHz for clarity.

To further discriminate potential DM signals (shared events) from background events (singles), we fit three 2×12 \times1 OFs to each event. We assume shared, left channel singles, and right channel singles event topologies, respectively. In this way, each event is associated with three best-fit χ2\chi^2 values: χshared2\chi^2_{\mathrm{shared}}, χsingleLeft2\chi^2_{\mathrm{singleLeft}}, and χsingleRight2\chi^2_{\mathrm{singleRight}}. To compare these signal template assumptions, we further define

δχSL2=χshared2−χsingleLeft2δχSR2=χshared2−χsingleRight2δχLR2=χsingleLeft2−χsingleRight2.(1)\begin{aligned} \delta\chi^2_{\mathrm{SL}} &= \chi^2_{\mathrm{shared}}-\chi^2_{\mathrm{singleLeft}} \\ \delta\chi^2_{\mathrm{SR}} &= \chi^2_{\mathrm{shared}}-\chi^2_{\mathrm{singleRight}} \\ \delta\chi^2_{\mathrm{LR}} &= \chi^2_{\mathrm{singleLeft}}-\chi^2_{\mathrm{singleRight}}. \tag*{(1)} \end{aligned}

For example, an event with δχSL2>0\delta\chi^2_{\mathrm{SL}}>0 is more consistent with a singles pulse in the left channel than with a shared pulse, and δχLR2<0\delta\chi^2_{\mathrm{LR}}<0 is more consistent with a left singles event than a right singles event.

The histogram in Figure 2 is colored based on the three 2×12 \times1 OFs—the black-orange heat map represents shared-like events (δχSL2<0\delta\chi^2_{\mathrm{SL}}<0 and δχSR2<0\delta\chi^2_{\mathrm{SR}}<0), and the blue-purple heat map represents other events. Figure 3 depicts two example measured events, compared with best-fit shared and singles templates.

Figure 3. A candidate shared event (left panel) and a candidate left singles event (right panel) observed in our detector. We compare the measured response (thin solid line) with three templates: a shared response in both channels (thick solid line), a left singles (dotted), and a right singles (dashed) in the left and right channels, respectively. By using δχ2\delta\chi^2 statistics, we can determine whether the shape and amplitude in both channels are most consistent with a shared, left singles, or right singles event. Data is filtered with a 50 kHz low-pass filter and offset for clarity. The left-panel and right-panel events have reconstructed energies of 4.97 and 3.31 eV, respectively, using a shared template.

Event Selection and Efficiencies

Our trigger finds 434,090 events, each reconstructed with a trace length of 20 ms. To remove events in periods of poor data quality where the detector response will be abnormal (e.g. periods of high electromagnetic interference (EMI) or vibration induced noise, or “pileup” events with multiple triggers in the same OF window), we impose data quality cuts. To reject periods of high EMI or noise, as well as elevated detector temperature following e.g. high-energy cosmic ray interactions, we cut events where the pre-pulse quiescent current (baseline) or difference between pre- and post-pulse currents (slope) fell outside a predetermined range from the unblinded dataset. These cuts remove 4.5% of events. We additionally perform a cut based on the χ2\chi^2 value extracted from the OF for every given pulse (considering only frequencies below 50 kHz and assuming a phonon template), rejecting events which are inconsistent with the expected pulse shape. This 5.6σ5.6\sigma low frequency χ2\chi^2 cut has the effect of removing pileup, triggers at incorrect times, and events with high noise or abnormal pulse shape. Events passing these cuts in both of our detector channels are preliminarily accepted as DM candidate events.

To reject backgrounds which couple primarily to the phonon sensors, we perform a final selection of events with δχSL2<0\delta\chi^2_{\mathrm{SL}}<0 and δχSR2<0\delta\chi^2_{\mathrm{SR}}<0 (i.e. events that are more shared-like than singles-like). Events which pass this additional δχ2\delta\chi^2 cut are accepted into our final DM analysis.

We lastly place an analysis threshold of 1.5 eV, removing any events below this energy. The final event spectrum can be seen in Figure 4.

Figure 4. (Main) Spectrum of observed events passing quality cuts and shared event cuts (black). Dotted lines show the modeled DM interaction spectra at the limit of exclusion for DM with masses 63, 100, and 178 MeV/c2c^2. (Insert) Calibration spectrum, showing peaks from absorbing zero, one, two, etc. photons.

We measure the impact of these cuts on the detector nuclear recoil (NR) efficiency using injected virtual pulses with an ideal shared pulse shape (“salt”) [37]. Salt pulses at a range of NR energies, as low as 0.361 eV, are injected into the continuous data stream pre-trigger. For each injected energy, we measure the binned difference between the salted and unsalted spectrum and normalize by the total number of salts. This net differential signal response estimates inefficiencies due to triggering, analysis cuts, and measures energy smearing. The response to sub-threshold recoil energies observed through noise boosting [37] is also studied provided that the signal response is linear and has small true signal pileup. See Appendix § 1.2.2 for details. This is convolved with the standard spin-independent NR DM spectrum dR/dEdR/dE [38] to build the model of measured spectrum dR/dE′dR/dE'. The dR/dE′dR/dE' spectra for DM masses of 63, 100, and 178 MeV/c2c^2 can be seen overlaid on the measured event spectrum in Figure 4.

Results

The total exposure is 0.233 g × 12 h. The interaction rate between DM and silicon nuclei is modeled assuming scattering from DM with a local density of 0.3 GeV/cm3\mathrm{cm}^3 and a Maxwell–Boltzmann velocity distribution, using the parametrization of Lewin and Smith [38]. This model is convolved with the detector net differential sensitivity estimated via salting to get the net differential response for a given DM model [37].

The limit on the DM-nucleon spin-independent scattering cross-section σSI\sigma_{SI} is calculated with the Yellin optimum interval method [39, 40] at the 90 % confidence level, using reconstructed energy as the only observable. The background is completely un-modeled and is thus considered potential DM signal. As is standard for DM searches at relatively high cross-sections [11, 41], we take into consideration the shielding effect from the overburden. This includes the atmosphere and 3 m concrete floors above the lab, and is calculated using the Verne code base [42, 43]. The resulting limit given this overburden consideration is seen as the blue dashed line in Figure 5.

Figure 5. The 90 % C.L. limits on spin-independent DM below 1 GeV/c2c^2. The blue shaded region shows the exclusion from this work. The blue dashes line represents the exclusion from the Yellin optimum interval test when considering the effect of the overburden. The red dotted line is from the requirement that there be no DM signal pileup. Previously excluded DM phase space from the CRESST [8, 41] and SuperCDMS [11] collaborations are shown in gray. Constraints from cosmology and astrophysics are shown in light gray [48, 49, 50, 51, 52].

Some of this nominally excluded parameter space has a DM interaction rate that would lead to significant signal pileup. In this regime, the dR/dE′dR/dE' does not scale linearly with σSI\sigma_{SI}, rendering any linear interaction model incorrect [37]. To account for this, we further restrict the exclusion region to include only DM masses and cross-sections that would not produce significant pileup. To estimate the rate at which we begin to see pileup effects, we note that 91 μ\mus after a phonon pulse is triggered, the OF will relax to 10% of its peak amplitude. This suggests that a pileup rate of 1/(2 × 91 μ\mus) = 5.5 kHz is a reasonable upper exclusion boundary, below which our linear differential rate modeling is valid. This is seen as the red dotted curve in Figure 5. Future analyses can be designed specifically to search for DM with high pileup rates. These analyses will likely extend the reaches of this detector and others to substantially lower DM masses and higher interaction cross sections [44, 45].

New bounds on cross sections as low as 4 × 10−3210^{-32} cm2^2 below 87 MeV/c2c^2 DM mass are established. We place constraints on DM-nucleon cross sections down to masses of 44 MeV/c2c^2 at 4.67 × 10−3010^{-30} cm2^2, the lowest mass ever probed by a particle-like DM search, as a consequence of our excellent energy resolution. We place our most stringent bounds for DM with a cross-section of 6.56 × 10−3510^{-35} cm2^2 at 500 MeV/c2c^2. The full DM exclusion region from this work can be seen as the blue shaded region in Figure 5.

Conclusion

This letter presents the lowest-mass sensitivity of any dark matter nuclear recoil search, as a result of our unparalleled energy resolution of 361.5±0.4 meV. By treating our unknown background with the optimum interval, we are able to place world-leading constraints on DM between 44 MeV/c2c^2 and 87 MeV/c2c^2. These results show the potential of low-threshold superconducting sensors for exploring new DM parameter space. Through both the direct minimization of low energy excess backgrounds and the discrimination of these backgrounds with novel cryogenic targets like gallium arsenide and superfluid He [46, 47], TESSERACT aims to substantially improve upon this surface DM search, as well as search for other DM interactions.

Acknowledgments

The athermal phonon detector used in this search builds upon the work that has been done by the CDMS and SuperCDMS collaborations [32, 53, 54]. This work was supported in part by DOE Grants DE-SC0022916, DE-SC0019319, DE-SC0025523 and DOE Quantum Information Science Enabled Discovery (QuantISED) for High Energy Physics (KA2401032) and National Science Foundation Grants 2111375 and 1106400. This material is based upon work supported by the Department of Energy National Nuclear Security Administration through the Nuclear Science and Security Consortium under Award Number(s) DE-NA0003180 and/or DE-NA0000979. Work at Lawrence Berkeley National Laboratory was supported by the U.S. DOE, Office of High Energy Physics, under Contract No. DEAC02-05CH11231. Work at Argonne is supported by the U.S. DOE, Office of High Energy Physics, under Contract No. DE-AC02-06CH11357. Work at Texas A&M University was supported by the Mitchell Institute and by the U.S. DOE, Office of High Energy Physics, under Contract Nos. DE-SC0018981 and DE-SC0021051. W.G. and Y.Q. acknowledge the support by the National High Magnetic Field Laboratory at Florida State University, which is supported by the National Science Foundation Cooperative Agreement No. DMR-2128556 and the state of Florida.

Multi-Channel Optimal Filtering

Our offline analysis trigger extends the Optimal (Matched) Filter algorithm [34, 35] to NN simultaneous readout channels and MM potential signal shapes. It is based on χs2(A⃗,ts)\chi_s^2(\vec{A},t_s), defined below, which represents the consistency of an observed trace with the hypothesis ss under consideration.

χs2(A⃗,ts)=∫dfχs2(A⃗,ts,f),\chi_s^2(\vec{A},t_s)=\int df\chi_s^2(\vec{A},t_s,f),

and

χs2(A⃗,ts,f)=∑i,j=1N(vi∗(f)−∑α=1MAαe2πiftssiα∗(f))Cij−1(f)(vj(f)−∑β=1MAβe−2πiftssjβ(f)).(2)\chi_s^2(\vec{A},t_s,f)=\sum_{i,j=1}^{N}\left(v_i^*(f)-\sum_{\alpha=1}^{M}A_\alpha e^{2\pi i f t_s}s_{i\alpha}^*(f)\right)C_{ij}^{-1}(f)\left(v_j(f)-\sum_{\beta=1}^{M}A_\beta e^{-2\pi i f t_s}s_{j\beta}(f)\right). \tag*{(2)}

Here, A⃗\vec{A} is an MM-length real vector of signal template amplitudes; tst_s is the common starting time of pulses, vi(f)v_i(f) is the Fourier transform of the observed trace in channel ii; siα(f)s_{i\alpha}(f) is the Fourier transform of the signal template for channel ii corresponding to amplitude AαA_\alpha; C(f)C(f) is the noise covariance matrix (i.e., the noise cross-spectral density between pairs of channels); Cij−1(f)C_{ij}^{-1}(f) is the (i,j)(i,j) element of C(f)−1C(f)^{-1}; and we integrate over frequencies.

The Optimal Filter estimates of A⃗\vec{A} and tst_s are obtained by minimizing χs2(A⃗,ts)\chi_s^2(\vec{A},t_s). In practice, the time traces are digitized at a sampling frequency fsampf_{\mathrm{samp}} and a discrete Fourier transform is performed over a trace time-length of TT. The integral then becomes a sum over frequencies between −fsamp/2-f_{\mathrm{samp}}/2 and fsamp/2f_{\mathrm{samp}}/2 in steps of 1/T1/T. In this study, fsamp=1.25f_{\mathrm{samp}}=1.25 MHz and T=20T=20 ms.

Construction of an optimal filter trigger. In order to identify potential events, we let a 2×12\times1 OF with shared templates and a common amplitude run freely along the time sequence of the data stream. We then construct Δχs2(t)\Delta\chi_s^2(t) as below, where A⃗^\hat{\vec{A}} is the value of A⃗\vec{A} that minimizes χs2(A⃗,t)\chi_s^2(\vec{A},t):

Δχs2(t)≡χs2(0,t)−χs2(A⃗^,t)(3)\Delta\chi_s^2(t)\equiv\chi_s^2(0,t)-\chi_s^2(\hat{\vec{A}},t) \tag*{(3)}

This represents, for a given trace, the difference between the minimal χ2\chi^2 in Fourier space and the χ2\chi^2 that would be obtained by assuming no signal. Our N×MN\times M Optimal Filter trigger runs on Δχs2(t)\Delta\chi_s^2(t), optimally selecting pulses that cross a certain threshold. This allows us to cleanly vary the signal template without changing the trigger parameters. It can be shown that Δχs2(t)\Delta\chi_s^2(t) is quadratically dependent on the best-fit signal amplitude A⃗^\hat{\vec{A}}. For the simple case of one readout channel and one signal template, this trigger is thus identical to triggering on the Optimal Filter amplitude, as described in [35].

Events are selected when Δχshared2\Delta\chi_{\mathrm{shared}}^2 exceeds 16, or equivalently when the best-fit amplitude A^\hat{A} exceeds 4σbase4\sigma_{\mathrm{base}}. Here, σbase\sigma_{\mathrm{base}} is the baseline resolution of A^\hat{A}, i.e., the magnitude of fluctuations of A^\hat{A} due only to noise.

For a given pulse, Δχshared2\Delta\chi_{\mathrm{shared}}^{2} will exceed the trigger threshold continuously for some amount of time, so we select triggers by assembling above-threshold regions of time within a 2-ms window into discrete events. This window was selected to capture all unsaturated signals while minimizing deadtime. Within each region, the trigger is placed at the time that maximizes Δχshared2\Delta\chi_{\mathrm{shared}}^{2}.

Analysis optimal filters. We then implement OFs on triggered events to extract pulse features. As the template shape varies, the pulse start time is also allowed to vary within the 2 ms time window. For instance, to discriminate potential DM signals (shared events) from background events (singles), we fit three 2×12 \times1 OFs to each event and construct δχ2\delta\chi^{2}, as explained in the main text. (Note that δχ2\delta\chi^{2} is different from Δχshared2\Delta\chi_{\mathrm{shared}}^{2}.) Cuts based on δχ2\delta\chi^{2} are, in essence, maximum likelihood ratio tests on the data.

Noise modeling with the Salting Method

The upward fluctuation of noise allows additional sensitivity to events with sub-threshold true energies that are boosted above the trigger threshold. This idea has been widely adopted in DM searches [8, 11]. The upward fluctuation is usually limited to 3σ3\sigma of the baseline energy resolution to prevent sensitivities to zero energy events. Discussions in [37] pointed out that a more rigorous modeling of the noise smearing the true event energy without the arbitrary 3σ3\sigma cut-off can be achieved by salting the raw data traces with ideal signal pulses.

The core concept is the net differential response, defined as

Δf(E′∣E)≡f(E′∣E)−f(E′∣0)(4)\Delta f(E'|E) \equiv f(E'|E) - f(E'|0) \tag*{(4)}

where EE is the true energy and E′E' is the measured energy. The quantity f(E′∣E)f(E'|E) represents the probability distribution of E′E' given that events with energy EE are present. The −f(E′∣0)-f(E'|0) term accounts for the reduction of noise-only time periods in the measurement as signals are added. In previous works, this term is ignored because the overlap between f(E′∣E)f(E'|E) and f(E′∣0)f(E'|0) above the trigger threshold is negligible. However, it is problematic if signals with E∼0E \sim0 are considered, resulting in underestimation of DM cross sections at low-masses.

We estimate Δf(E′∣E)\Delta f(E'|E) using the salting method. First, NsN_s pulses (salts) with ideal signal pulse shape and energy EsE_s are injected into the continuously recorded raw traces of total exposure time TT randomly in time. The random injection times are separated by the triggered trace length to prevent pileup of two salt pulses. Second, the salted traces are processed by the offline trigger and filtering algorithm as described in Sec. III, then selected under the same criteria as in Sec. IV. The exact same analysis is done for both salted and unsalted datasets to ensure the accurate measurement of trigger and event selection efficiencies. Then, the measured spectrum dRdE′^(E′∣S+s)\widehat{\frac{dR}{dE'}}(E'|S+s) is compared with the one before salting dRdE′^(E′∣S)\widehat{\frac{dR}{dE'}}(E'|S), and normalized by rs=Ns/Tr_s = N_s/T to estimate Δf(E′∣E)\Delta f(E'|E)

Δf^(E′∣S+Es)=dRdE′^(E′∣S+s)−dRdE′^(E′∣S)rs(5)\widehat{\Delta f}(E'|S+E_s) = \frac{\widehat{\frac{dR}{dE'}}(E'|S+s) - \widehat{\frac{dR}{dE'}}(E'|S)}{r_s} \tag*{(5)}

where SS represents the potential DM signal in the unsalted spectrum, and ss represents the salt. Finally, salting is repeated at energies from 0.361 eV to 30 eV. The limit of DM with mass mχm_\chi is calculated with the Yellin optimum interval method [40], with

ΔdRdE′(E′∣S,s(mχ,σ0))≡∫0∞dRdE(Es∣s(mχ,σ0))Δf^(E′∣S+Es) dEs(6)\Delta\frac{dR}{dE'}(E'|S, s(m_\chi,\sigma_0)) \equiv\int_{0}^{\infty} \frac{dR}{dE}(E_s|s(m_\chi,\sigma_0)) \widehat{\Delta f}(E'|S+E_s)\,dE_s \tag*{(6)}

scaling with σ0\sigma_0 but not dRdE′\frac{dR}{dE'}. Here σ0\sigma_0 is the reference DM cross section. See [37] for full discussion.

References

  1. [1]D. S. Akerib et al. (LZ Collaboration), Nucl. Instrum. Meth. A 953, 163047 (2020), arXiv:1910.09124 [physics.ins-det].DOI
  2. [2]E. Aprile et al. (XENON Collaboration), Phys. Rev. Lett. 131, 041003 (2023), arXiv:2303.14729 [hep-ex].DOI
  3. [3]P. Agnes et al. (DarkSide-50), Phys. Rev. D 107, 063001 (2023), arXiv:2207.11966 [hep-ex].DOI
  4. [4]Y. Meng et al. (PandaX-4T Collaboration), Phys. Rev. Lett. 127, 261802 (2021), arXiv:2107.13438 [hep-ex].DOI
  5. [5]Y. Hochberg, E. Kuflik, T. Volansky, and J. G. Wacker, Physical Review Letters 113, 171301 (2014).
  6. [6]Y. Hochberg, E. Kuflik, H. Murayama, T. Volansky, and J. G. Wacker, Physical Review Letters 115, 021301 (2015).
  7. [7]E. Kuflik, M. Perelstein, N. R.-L. Lorier, and Y.-D. Tsai, Journal of High Energy Physics 2017, 78 (2017).
  8. [8]A. H. Abdulhameed et al. (CRESST), Phys. Rev. D 100, 102002 (2019), arXiv:1904.00498 [astro-ph.CO].
  9. [9]G. Angloher et al. (CRESST), Phys. Rev. D 107, 122003 (2023), arXiv:2212.12513 [astro-ph.CO].DOI
  10. [10]E. Armengaud et al. (EDELWEISS), Phys. Rev. D 99, 082003 (2019), arXiv:1901.03588 [astro-ph.GA].DOI
  11. [11]I. Alkhatīb et al. (SuperCDMS), Phys. Rev. Lett. 127, 061801 (2021), arXiv:2007.14289 [hep-ex].DOI
  12. [12]I. Arnquist, N. Avalos, D. Baxter, X. Bertou, N. Castelló-Mor, A. Chavarria, J. Cuevas-Zepeda, J. C. Gutiérrez, J. Duarte-Campderros, A. Dastgheibi-Fard, et al., Physical Review Letters 130, 171003 (2023).
  13. [13]O. Abramoff et al. (SENSEI), Phys. Rev. Lett. 122, 161801 (2019), arXiv:1901.10478 [hep-ex].DOI
  14. [14]E. Kuflik, M. Perelstein, N. R.-L. Lorier, and Y.-D. Tsai, Phys. Rev. Lett. 116, 221302 (2016), arXiv:1512.04545 [hep-ph].DOI
  15. [15]Y. Hochberg, E. Kuflik, T. Volansky, and J. G. Wacker, Phys. Rev. Lett. 113, 171301 (2014).
  16. [16]L. J. Hall, K. Jedamzik, J. March-Russell, and S. M. West, JHEP 03, 080 (2010), arXiv:0911.1120 [hep-ph].DOI
  17. [17]C. Boehm and P. Fayet, Nucl. Phys. B 683, 219 (2004), arXiv:hep-ph/0305261.DOI
  18. [18]M. Pospelov, A. Ritz, and M. Voloshin, Physics Letters B 662, 53 (2008).
  19. [19]M. J. Strassler and K. M. Zurek, Physics Letters B 651, 374 (2007).
  20. [20]D. E. Kaplan, M. A. Luty, and K. M. Zurek, Phys. Rev. D 79, 115016 (2009).
  21. [21]J. L. Feng and J. Kumar, Phys. Rev. Lett. 101, 231301 (2008).
  22. [22]D. Hooper and K. M. Zurek, Physical Review D 77 (2008), 10.1103/physrevd.77.087302.
  23. [23]P. Adair et al., SciPost Physics Proceedings , 001 (2022).
  24. [24]R. Anthony-Petersen et al., Nature Communications 15 (2024), 10.1038/s41467-024-50173-8.
  25. [25]R. K. Romani, Journal of Applied Physics 136 (2024), 10.1063/5.0222654.
  26. [26]K. Nordlund, F. Kong, F. Djurabekova, M. Heikinheimo, K. Tuominen, and N. Mirabolfathi, “Defect recombination origin of low energy excess in semiconductor detectors,” (2024).
  27. [27]P. Du, D. Egana-Ugrinovic, R. Essig, and M. Sholapurkar, Phys. Rev. X 12, 011009 (2022).
  28. [28]R. Anthony-Petersen et al., “Low energy backgrounds and excess noise in a two-channel low-threshold calorimeter,” (2024).
  29. [29]G. Angloher et al., “Doublets detectors to investigate the cresst low energy background: results from above-ground prototypes,” (2024).
  30. [30]R. K. Romani, Y.-Y. Chang, R. Mahapatra, M. Platt, M. Reed, I. Rydstrom, B. Sadoulet, B. Serfass, and M. Pyle, “A transition edge sensor operated in coincidence with a high sensitivity phonon veto for photon coupled rare event searches,” (2024).DOI
  31. [31]K. D. Irwin, Applied Physics Letters 66, 1998–2000 (1995).
  32. [32]K. D. Irwin, S. W. Nam, B. Cabrera, B. Chugg, and B. A. Young, Review of Scientific Instruments 66, 5322 (1995).
  33. [33]Photo courtesy of Marilyn Sargent/Lawrence Berkeley National Laboratory.
  34. [34]S. R. Golwala, Exclusion limits on the WIMP nucleon elastic scattering cross-section from the Cryogenic Dark Matter Search, Ph.D. thesis, UC, Berkeley (2000).
  35. [35]S. L. Watkins, Athermal Phonon Sensors in Searches for Light Dark Matter, Ph.D. thesis, UC, Berkeley (main) (2022), arXiv:2301.08699 [hep-ex].arxiv.org/abs/2301.08699
  36. [36]K. Irwin and G. Hilton, in Cryogenic Particle Detection, Topics in Applied Physics, edited by C. Enss (Springer, Berlin, Heidelberg, 2005) pp. 63–150.
  37. [37]X. Li, M. Pyle, and B. Sadoulet, “Modeling the differential rate for signal interactions in coincidence with noise fluctuations or large rate backgrounds,” (2024), arXiv:2411.10671 [hep-ex].DOI
  38. [38]J. Lewin and P. Smith, Astropart. Phys. 6, 87 (1996).
  39. [39]S. Yellin, Phys. Rev. D 66, 032005 (2002), arXiv:physics/0203002.DOI
  40. [40]S. Yellin, (2008), arXiv:0709.2701 [physics.data-an].arxiv.org/abs/0709.2701
  41. [41]G. Angloher et al. (CRESST), Phys. Rev. D 110, 083038 (2024), arXiv:2405.06527 [astro-ph.CO].arxiv.org/abs/2405.06527
  42. [42]B. J. Kavanagh, Phys. Rev. D 97, 123013 (2018), arXiv:1712.04901 [hep-ph].DOI
  43. [43]B. J. Kavanagh, “bradkav/verne: Version 2.0,” (2024).
  44. [44]A. Das, N. Kurinsky, and R. K. Leane, Phys. Rev. Lett. 132, 121801 (2024), arXiv:2210.09313 [hep-ph].DOI
  45. [45]work in progress, “Constrain dark matter interactions with noise power,”.
  46. [46]S. A. Hertel, A. Biekert, J. Lin, V. Velan, and D. N. McKinsey, Phys. Rev. D 100, 092007 (2019).
  47. [47]R. Anthony-Petersen et al. (SPICE, HeRALD), Phys. Rev. D 110, 072006 (2024), arXiv:2307.11877 [physics.ins-det].
  48. [48]C. C. Cappiello and J. F. Beacom, Phys. Rev. D 100, 103011 (2019), [Erratum: Phys.Rev.D 104, 069901 (2021)], arXiv:1906.11283 [hep-ph].DOI
  49. [49]T. Bringmann and M. Pospelov, Phys. Rev. Lett. 122, 171801 (2019), arXiv:1810.10543 [hep-ph].DOI
  50. [50]E. O. Nadler et al. (DES), Phys. Rev. Lett. 126, 091101 (2021), arXiv:2008.00022 [astro-ph.CO].DOI
  51. [51]K. K. Rogers, C. Dvorkin, and H. V. Peiris, Phys. Rev. Lett. 128, 171301 (2022).
  52. [52]V. Gluscevic and K. K. Boddy, Phys. Rev. Lett. 121, 081301 (2018).
  53. [53]J. Yen et al., Applied Physics Letters 105, 163504 (2014), http://dx.doi.org/10.1063/1.4899130.
  54. [54]M. C. Pyle, Optimizing the design and analysis of cryogenic semiconductor dark matter detectors for maximum sensitivity, Ph.D. thesis, Stanford U. (2012).DOI

Paper details

vlnc:W_tt5qs6cf
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 1

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…