Physics-Informed Deep Learning for Shear Wave Speed Estimation in MR Elastography
Abstract
Objective: Magnetic Resonance Elastography (MRE) is a non-invasive imaging technique for mapping biomechanical properties of in vivo tissue, including shear wave speed (SWS), but involves intrinsically slow data acquisition and an ill-posed wave inversion. Instead of relying on handcrafted image priors, we propose a data-driven approach jointly combining image reconstruction and MRE inversion for robust SWS estimation from undersampled k-space data. Methods: Our physics-informed reconstruction framework comprises two blocks: a model-based neural network (NN)-regularized reconstruction module and a phase-gradient inversion (k-MDEV) calculating SWS from the reconstructed images. Concatenating both blocks yields an end-to-end trainable method to estimate SWS directly from measured k-space data. We evaluated the method on retrospectively highly undersampled brain MRE data and compared it to a total variation (TV) minimization-based approach. We assessed the impact of end-to-end training (qualitative images and SWS maps as targets) versus pre-training (qualitative images as targets) and applied the method also to in vivo data. Results: Our approach significantly reduces NRMSE by 30% compared to TV. End-to-end training improves SWS estimation over separate image reconstruction and SWS calculation. Conclusion: Accurate SWS quantification is possible at acceleration factors up to 19. Our method significantly outperforms TV, highlighting the need for data-driven regularization in this challenging MR problem. Further, our approach successfully generalizes to in vivo data. Significance: We present the first end-to-end trainable MRE reconstruction method for estimating SWS maps directly from k-space. NN-based reconstruction can enable rapid stiffness mapping for dynamic studies, functional imaging, and real-time clinical feedback.
© 2026 The Authors. This work is licensed under a Creative Commons Attribution 4.0 License. For more information, see https://creativecommons.org/licenses/by/4.0/.
This is the authors’ accepted version of an article published in IEEE Transactions on Biomedical Engineering. Changes were, or will be, made to this version by the publisher prior to final publication.
DOI: 10.1109/TBME.2026.3666306
IEEE Xplore: https://ieeexplore.ieee.org/document/11407998
Please cite this paper as:
S. Martin, J. Schattenfroh, P. Schuenke, F. F. Zimmermann, I. Sack, C. Kolbitsch, and A. Kofler, “Physics-Informed Deep Learning for Shear Wave Speed Estimation in MR Elastography,” in IEEE Transactions on Biomedical Engineering, 2026, doi: 10.1109/TBME.2026.3666306.
Code: https://github.com/Stef-Martin/E2E-MRE
Introduction
Magnetic Resonance Imaging (MRI) is a versatile and powerful imaging technique that allows the acquisition of high-resolution qualitative images of the internal anatomy of the human body. In quantitative MRI (qMRI), various quantitative parameters with specific biophysical correspondences can then be estimated from the image data. MR Elastography (MRE) [1, 2, 3] is a qMRI technique that provides detailed maps of biomechanical properties, enabling a non-invasive assessment of tissue stiffness. Externally induced mechanical waves result in tissue displacements, which are encoded as proportional phase shifts in the MR signal by motion encoding gradients (MEG) that are incorporated into specially designed MRI sequences. These phase shifts are tracked over time, enabling the calculation of shear wave speed (SWS) [4], which generally involves three steps: acquisition of -space data, reconstruction of complex-valued MR images from the data, and estimation of quantitative SWS maps from these images with an inversion method.
The brain is a fascinating target for MRE, particularly for detecting and monitoring neurological disorders that are associated with localized changes in tissue stiffness, such as Alzheimer’s disease [5]. Conventional MRE techniques typically require multiple time-resolved acquisitions, resulting in extended acquisition protocols with prolonged scan times and vibration exposure. These factors can increase patient discomfort, reduce clinical efficiency, and limit temporal resolution, making it difficult to investigate dynamic physiological processes that require fast, continuous measurements. Real-time MRE addresses these limitations by enabling high frame-rate stiffness mapping [6]. This approach improves patient comfort, reduces motion-related artifacts, and enables the monitoring of dynamic changes in brain tissue properties – such as those occurring during the Valsalva maneuver [7] or cardiac-cycle-related brain pulsations, as demonstrated in ultrasound elastography [8]. Recent studies have shown the feasibility and clinical value of rapid MRE in various applications, including subsecond mapping of liver stiffness [9], evaluation of the response in skeletal muscle to loading at frame rates of Hz [6], and functional brain MRE [10].
Real-time stiffness mapping can be achieved by reducing scan time through undersampled data acquisition. In this work, we utilize a motion-robust single-shot spiral readout for MRE data acquisition, enabling the capture of all required data in a single shot per timestep and MEG. However, data undersampling introduces challenges into MRI reconstruction, yielding an ill-posed inverse problem necessitating the use of regularization techniques to stabilize the reconstruction. Importantly, the subsequent generation of MRE maps from wave images is itself inherently ill-posed due to noise and unknown boundary conditions.
Undersampled MRE images have been reconstructed using established non-learning approaches, which are typically applied prior to elastography inversion. These include parallel imaging and compressed sensing with sparsity priors, such as MRE-specific phase or displacement regularization [11], as well as Bayesian models that impose inter–offset constraints [12]. Further, machine learning (ML) has significantly influenced and improved MR image reconstruction [13] in recent years and currently defines the state-of-the-art both in MRI and qMRI reconstruction, see e.g. [14] for an extensive review. Despite these advances, several challenges persist. Frame rates in current rapid MRE techniques are still limited, and developments in neural network (NN)-based reconstruction methods and real-time MRE have largely progressed as separate fields. This work integrates both approaches to enhance the stiffness mapping frame rate while maintaining the accuracy of the reconstructed MR images and SWS maps. So far, ML in MRE has been applied to segmentation tasks [15], phase unwrapping optimization [16], and learning the inversion method for the calculation of stiffness maps [17, 18, 19, 20].
In this work, we employ a NN-based regularization in a physics-informed image reconstruction framework [21], [22] that can be trained end-to-end [23], [24] – from -space data to stiffness map. By treating qualitative image reconstruction and SWS quantification as joint tasks, we enable the NN to learn a task-specific regularization tailored to the quantification of SWS by the inversion method -MDEV [25].
Problem Formulation
The process of data acquisition in MRI is typically modeled as
Here, represents the measured -space data, the MRI operator, the complex-valued MR image that has to be reconstructed, and denotes Gaussian noise, which accounts for measurement errors. In qMRI, the considered forward problem is extended by performing several different measurements of the same underlying quantity, thereby collecting complementary data by, for example, applying different preparation pulses or by varying the MR sequence parameters. Formally, this can be described by a (typically non-linear) model , which models all relevant scanner and MR sequence parameters defined by (see e.g. [26]). It relates the quantities of interest to the prepared image, i.e.
where is the reference parameter associated with the reference image and accounts for non-trivial modeling errors inherently present in . The corresponding -space data are acquired according to (1). In MRE, specifically, the model describes the wave propagation as reflected in tissue displacement. The entire acquisition can thus be expressed as
where and each represents the -th wave field generated by the forward model with set-up parameters defined by (more details on that later in Subsection III-A). Note that the employed forward models and are assumed to be sufficiently accurate relative to the true models, such that the modeling error term is considered negligible compared to the sensor noise and, therefore, omitted from the acquisition model of (3). External mechanical vibrations are applied, generating shear waves propagating through the tissue. The resulting displacement depends on the underlying viscoelastic properties, which are characterized by the complex shear modulus . The motion is captured by MEGs and encoded into the phase of the MR signal, which is read out during MR acquisition. The forward model in (2) maps the viscoelastic parameters to the measured MR phase, reflecting shear wave propagation. This wave is governed by the Navier equation, which describes the propagation of waves in a medium. With several simplifying key assumptions – time-harmonic motion, incompressibility (neglecting longitudinal wave components), and local homogeneity of the viscoelastic parameters – we can simplify the Navier equation to the Helmholtz equation:
where is the shear wave field (obtained from processed MR phase data: applying filters or the curl operator to suppress compressional components), is the angular frequency of the vibration, and is the SWS at voxel . The homogeneity assumption allows spatial derivatives of material parameters to be neglected and enables this formulation. The inverse problem, which is the one of interest in our application, aims to estimate the viscoelastic parameters from the shear wave field } and is commonly approached in three ways [4], [27]. One approach solves a weak formulation via finite elements (FEM); another directly inverts the Helmholtz equation and typically invokes local homogeneity; and a third relies on phase-gradient or wavenumber-based methods. Many FEM and NN approaches do not require the homogeneity assumption and explicitly accommodate heterogeneous media by solving full/weak forms of the governing equations or learning mappings from displacement fields to spatially varying properties. In this work, we use a phase-gradient inversion such as -MDEV [25]. It assumes locally planar waves and homogeneity and estimates the clinically relevant SWS from the displacement field encoded in the MR image phase. The SWS is given as , with the bulk density and the complex shear modulus at voxel . Consequently, for our problem, the vector of the parameters sought in (3) is given by .
Methods
Since problem (3) is typically non-convex due to the non-linear model and large-scale due to the acquisition of complementary measurements that allow for the quantification of , different solution strategies exist to tackle the problem, see [14]. In this work, we consider the widely used two-step approach for solving qMRI problems, i.e., decoupling the image reconstruction from the parameter quantification problem by
Thereby, (5) corresponds to reconstructing an estimate of the image in the ill-posed inverse problem in (1), and (4) addresses the estimation of the relevant quantitative parameters from the reconstructed image data as in (2). Based on the assumption that the only measurement error is sensor noise , the data-fidelity term in (5) is formulated using an -norm, which corresponds to the log-likelihood function under an additive Gaussian noise model. Note that the noise component in (2) is contained in in (4) and is inherently linked to the choice of the regularization method in (5) and its capability to approximate the unknown target image. When solving (5) and (4) in a sequential fashion, once the image data in (5) are fixed after the reconstruction step, residual errors directly impact the quantification process in (4). As such, it is crucial to choose an expressive regularization method that can represent complex priors or constraints on the solution.
Here, we propose to construct an end-to-end trainable NN based on algorithm unrolling [28] that comprises two main blocks. The first block consists of a model-based deep learning approach [22] that constructs a reconstruction network alternating between CNN application and data consistency (DC) layers. The second block corresponds to the -MDEV inversion method [25] for computing SWS maps from the reconstructed image data. Notably, both blocks can be trained end-to-end, i.e. from -space data to SWS maps. As a result, the regularizer in (5) is optimized for the downstream task of SWS quantification – the final quantity of interest in MRE – rather than for reconstructing precise qualitative images, which serve only as an intermediate step in the parameter estimation.
Image Reconstruction
First, we expand upon the forward problem introduced in (1) to precisely formulate our method for solving problem (5). We consider the general setup of multi-coil MRI, where the -space is acquired using multiple receiver coils to be able to accelerate the acquisition by data undersampling while, at the same time, maintaining comparable image quality when reconstructing the images.
In the specific case of MRE, vibration phase offsets are typically used to track wave propagation over time. Additionally, MEG directions and different frequencies can be acquired per slice and per subject, allowing for a more precise SWS computation, as demonstrated in [3]. The image data from these various MEG directions, frequencies, and vibration phase offsets can then be used to compute a single SWS map. For a fixed MEG , frequency and time point , we consider the linear operator , which maps a complex-valued 2D image to its corresponding multi-coil -space data by
where with with denotes the -th coil sensitivity map (CSM), is a 2D Fourier operator that maps a 2D image to its -space coefficients indexed by the set , and represents an identity operator. Note that in this work, corresponds to a non-uniform fast Fourier operator that samples -space data along spiral trajectories. In our approach, to be able to exploit the correlation between the images acquired for different time points , the precise form of the operator in (1) is
with each and given in the form of (6), meaning that we tackle the reconstruction of the data simultaneously for all time points, i.e. with , but separately for the different frequencies and MEGs . Crucially, the sets of indices vary for different acquisition time points , thus yielding complementary information in the -space data, which can be used for regularization. Consequently, the problem in (1) is rather large, involving -space data that are acquired with multiple receiver coils and for different time steps, requiring a computationally efficient reconstruction algorithm.
We approach the reconstruction of the undersampled MRE images in (5) by a model-aware reconstruction network that follows MoDL [22], i.e. given an initial reconstruction , we construct an alternating sequence of iterates
of length , where denotes a learnable network and are learnable scalar parameters. The term enforces DC between estimated image and measured data in raw data space through a weighted -norm involving a positive definite density compensation (DCF) operator that corrects for non-uniform sampling via Voronoi tessellation. The Tikhonov regularization term enforces proximity of the current iterate to an NN-based image prior that is obtained as the output of an expressive CNN-block, reducing noise and artifacts. Since problem (1) is in general overdetermined due to multiple receiver coils, we can initialize the reconstruction network with an approximate pseudo-inverse solution obtained by approximately solving the normal equations, i.e. with . Computing in (8) corresponds to solving a linear system of the form
which we tackle by means of a conjugate gradient (CG) method [29]. We note that the number of iterations that are used to unroll the network defined by (6) and (8) is a hyperparameter that impacts both the reconstruction quality as well as the GPU memory footprint and the training times of the considered method. In [22], it was shown that a larger tends to lower the achievable reconstruction error; however, it also increases training times and hardware requirements. In our case, we found to provide a good balance between reconstruction performance, GPU memory usage, and training time. Further, it is worth noting that setting corresponds to a conceptually different approach, namely the one presented in [30], that – although exhibiting some theoretically favorable properties (convexity of the thereby considered reconstruction problem (5)) – empirically falls short compared to unrolling iterations, especially for more complex problems with high acceleration factor as the one considered in this work.
In our specific problem, we can leverage the temporal correlation in the phase of the complex MR images generated by the mechanical wave traveling through the tissue. Therefore, we use the network presented in [31], which exploits the underlying spatio-temporal correlation in the data by extracting - and -spatio-temporal slices, applying a 2D U-Net [32] to reduce noise and artifacts, and reassembling the slices to the image volume. We refer to this network as XTYT. The framework of this physics-informed reconstruction network – which incorporates the physical forward model and the acquired -space data through DC blocks – is denoted as and is illustrated in Figure 1.

Figure 1. Yellow block: Image reconstruction block : Starting from undersampled spiral -space data, an initial MR image estimate is computed by approximating the pseudo-inverse solution (purple block). Grey block: An image is fed into a CNN, whose output is refined via a DC step according to (8). CNN and DC steps are repeated times. Pre-training minimizes the loss between reconstructed and target images. Orange block: -MDEV: The phase is extracted, unwrapped, and temporally Fourier transformed. The harmonic component of the wave field is filtered and inverted to yield SWS maps, as described in (7). Green block: The entire reconstruction network, comprising the image reconstruction and the -MDEV module, is fine-tuned by minimizing a combined loss that uses reference images and reference SWS maps.
k-MDEV for SWS Estimation
In the following, we briefly describe -MDEV [25], which is a method that estimates SWS maps from the reconstructed image data, recovering all necessary wave information from the wave vector, and extending classical phase-gradient approaches [2] by accounting for wave superposition and using directional filtering [33]. The method transitions from a continuous wave description to a discrete image-domain representation compatible with the nature of MR imaging. For details, see [3], [25].
Phase
Let denote the polar representation of the complex-valued MR image reconstructed as explained in III-A. Mechanical vibrations induce shear waves, and MEGs encode this periodic motion proportionally into the image phase , which approximates the displacement field under the simplifying transverse wave assumptions.
Phase Unwrapping
Oscillations with large amplitudes can cause phase jumps that can hinder the estimation of the SWS. As in the original work (see [3], [25]), we tackle these phase wraps by Laplacian-based unwrapping [34].
Temporal Fast Fourier Transform (FFT)
To isolate the fundamental harmonics of the wave corresponding to the excitation frequency used during MRE acquisition, we compute the FFT of the image phase and consider only the harmonic frequency, which we denote by .
Remark: In general, in a continuous setting, a harmonic shear wave can be modeled as a complex-valued plane wave
Directional Filtering and Wave Vector Estimation
, where the complex wave vector includes both propagation and attenuation, and is the angular frequency of the vibration. Here, we focus on estimating , which determines the wave speed. The real part of the phase of the planar shear wave (as encoded by the MEGs and captured in the MR image phase) is given by . Applying the temporal FFT and taking the phase gradient at the fundamental harmonic provides an estimate of the real part of the wave vector .
4) Directional Filtering and Wave Vector Estimation: For each of the MEG directions and frequencies , the wavefield consists of superimposed waves. To isolate wavefronts propagating in several directions, the wavefield is filtered by , with being the linear operators associated with the directional filters and a 2D FFT. Then a full wave vector can be calculated for all , , and of the discretized filtered wavefields . To account for geometric attenuation, each filtered wave component is normalized, and the magnitude of the real wave vector in each direction is componentwise given by the phase-gradient:
Weighted Inversion
Finally, the SWS is computed from the real part of the estimated wave vectors. For voxel , a weighted averaging is applied to combine multiple MEG directions and frequency components. Each measurement is weighted by [3], which favors higher signal amplitudes. The SWS reads
where each image is given by the CNN-based reconstruction with denoting the -space data for the MEG and frequency . This yields spatially resolved SWS maps, which are used as the primary tissue parameter in this work. Additionally, it should be noted that due to GPU memory limitations in our end-to-end pipeline as described in IV, we retain frequency-resolved SWS maps using only one image per MEG rather than using . The entire -MDEV pipeline (shown in Figure 1) is based on [3], [25] and was implemented in PyTorch to allow for gradient backpropagation during joint image reconstruction and SWS estimation.
Experiments
We evaluated our proposed method on a brain MRE imaging problem. The experimental setup and the dataset are described in the following. The code for the image reconstruction pipeline is available at https://github.com/Stef-Martin/E2E-MRE. All operators were implemented using our in-house developed PyTorch-compatible Python package MRpro [35].
Data
In total, in vivo MRE was performed in 13 healthy volunteers without a history of neurological disease (10M, 3F; mean age SD: 28.6 8.6 years). Data from 12 of 13 subjects were used in the retrospective study, as detailed in Subsection IV-A.1. In addition, another subject was scanned for the prospective study; see Subsection IV-A.2 for acquisition details. The study was approved by the ethics committee of Charité – Universitätsmedizin Berlin in accordance with the Declaration of Helsinki. Every participant gave oral and written informed consent. The data were acquired using a clinical 3T MRI scanner (Magnetom Lumina, Siemens Healthineers, Erlangen, Germany) equipped with a 32-channel excite/receive head coil. Shear waves were induced using a cradle actuator system placed beneath the head and driven by compressed air at consecutive frequencies (20, 25, 30, and 35 Hz), with vibration initiated 2 s before acquisition to ensure a steady state of harmonic motion. Motion was encoded into the phase of the complex MRE signal using trapezoidal, flow-compensated MEGs [36] with an amplitude of 34 mT/m, a slew rate of 125 mT/m/ms, and a duration of 26.48 ms, resulting in motion-encoding efficiencies of 69.48 (20 Hz), 98.61 (25 Hz), 125.10 (30 Hz), and 145.80 rad/mm (35 Hz). The MEGs were consecutively applied along directions [37] (read, phase, and slice) sequentially encoding three-dimensional displacement fields. Data were acquired at frequencies with vibration phase offsets equally spaced over a vibration period.
In the retrospective study, Cartesian spin-echo MR images served as ground-truth for simulating spiral acquisitions according to model (1), with trajectories incrementally rotated by the golden angle (GA, 137.5°). In the prospective study, spirals (also rotated by the GA) were acquired directly. To optimize -space coverage, all spirals acquired for the different , , and were further rotated by the GA, ensuring complementary information collection in both spatial and temporal dimensions.
Retrospectively Undersampled Data
1) Retrospectively Undersampled Data: Brain data from 12 volunteers were acquired with a Cartesian spin-echo MRE sequence [37] with a field of view (FOV) of 202 mm 202 mm 80 mm and a resolution of 1.6 mm 1.6 mm 2 mm. The scan-time per subject was 451 s with an echo time (TE) of 70 ms and a repetition time (TR) of 4700 ms. These Cartesian reference data were acquired with an image size of . Based on this, we retrospectively simulated accelerated MRE data acquisition according to (1) using spiral trajectories with 863 -space points each (identical to those employed for in vivo scans in IV-A.2), yielding undersampling factors of 19, 9.5, and 6.3, respectively. Note that the forward operator in (1) inherently depends on the CSMs, i.e., , which must be estimated before the approach reconstruction. To mimic a realistic situation found in clinical practice, -space data were retrospectively simulated according to (1) with , where were computed from spiral MRE data acquired with 25 spirals and 863 -space points per spiral using the Walsh algorithm. An estimate of these CSMs was obtained from retrospectively simulated Cartesian calibration -space data, which were generated using the same CSMs, but only a small calibration set of -space points, thereby introducing realistic calibration uncertainties in the form of a model-mismatch, i.e. . The operator was used in the image reconstruction, see III-A. The DCF was derived using the Voronoi method. The dataset comprised 5760 samples – 12 subjects 40 slices 4 frequencies 3 MEGs – each associated with 8 distinct vibration phase offsets. The dataset was partitioned into training, validation, and test subsets using 8, 1, and 3 subjects, respectively, which correspond to 3840, 480, and 1440 samples.
Prospective In Vivo Study
To evaluate our pipeline on in vivo data, we designed an MRE sequence using PyPulseq [38] and acquired data from a single volunteer. The acquisition parameters were set to an FOV of 220 mm 220 mm 5 mm, with an in-plane resolution of 1.7 mm, a flip angle of , a TE of 21 ms, and a TR of 56 ms. Data were acquired using 25 spirals, each consisting of 863 -space points, of which only a subset of was used in the model. CSMs were estimated from a reference dataset acquired using 10 spiral interleaves with a total acquisition time of 2.5 s. Prior to reconstruction with any method, the considered -space data were normalized by their standard deviation.
Network Architectures
As the CNN (see (6)) employed in the block, we use the XTYT network [31] – specifically designed to reconstruct spatiotemporal slices. For the ablation study, we compared models with different numbers of unrolled iterations using XTYT blocks of different sizes. Both configurations employed three encoding and decoding stages with two convolutions per stage. For our proposed method with , an XTYT-block with about 2.1 million parameters was used, whereas for the ablation study and , the network used around 8.3 million trainable parameters, allowing for a more meaningful comparison of the approaches, see [39].
Loss Functions and Training
Here, we consider two primary training strategies. Let and -MDEV denote the image reconstructed with the network described in Subsection III-A and the SWS maps estimated by -MDEV described in Subsection III-B, respectively, with and representing their corresponding targets. Both depend on the trainable parameters , which, in turn, are determined by the network architecture as described in IV-B. Importantly, although only the image reconstruction block contains learnable parameters, we have the flexibility to choose the target in our network optimization to be either the complex-valued MR image or the SWS maps (or a combination thereof). By doing so, we can tailor the reconstruction network to the ultimate specific task, i.e. the estimation of the SWS maps. Here, we consider a loss function that combines contributions from qualitative MR images and SWS maps, given by
where and are defined as
and
with the deviation between image estimate and target being the perpendicular loss as given by [40] and with MAE denoting the mean absolute error. In (10), the root mean squared error (RMSE in m/s) quantifies the average magnitude of errors between SWS maps, while the structural similarity index (SSIM, unitless) measures their perceptual similarity – both metrics are on a comparable scale. A weight of has been assigned to the RMSE term in (12). Note that while we use RMSE for training, we additionally report the normalized RMSE (NRMSE) for evaluation; that is, the RMSE normalized by the reference image/SWS map energy, which provides a dimensionless and scale-invariant error measure.
Pre-Training
The most straightforward approach is to train the network directly with fully sampled complex MRI images serving as the target [41]. In this method, -MDEV is treated as an independent module and does not influence the training process, as visualized in Figure 1, yellow box. This corresponds to setting in (8). AdamW [42] was chosen as the optimizer with a learning rate of and a weight decay of . After each epoch, the SWS maps were calculated from the image estimate and compared to the corresponding reference SWS maps using (10). This loss was tracked over the whole pre-training process. The initialization block (see purple block in Figure 1) uses 24 iterations to approximate the Moore-Penrose pseudo-inverse solution. Pre-training required approximately 1.5 hours per epoch, totaling 3 days for 50 epochs of training on a single Nvidia A6000 with GPU requirements of 13.2 GB.
End-to-End Fine-Tuning
Our final proposed end-to-end approach, displayed in Figure 1, green box, integrates the entire -MDEV module into the model as a post-processing step applied to the network output (the estimated MR images) and used in the training process. The SWS loss (10) can then be computed by comparing the resulting SWS maps to the target SWS maps, which were obtained by applying -MDEV to the reference MR images. End-to-end training corresponds to setting in (8). In this subsequent task-specific fine-tuning stage, the pre-trained weights from IV-C.1 are further adapted, shifting the optimization towards the final quantitative objective of accurate SWS estimation, while maintaining clinical interpretability of the qualitative images due to the data-consistent reconstruction framework and the choice of , in our case . Since the network is already pre-trained according to IV-C.1, a very small initial learning rate of was chosen (decreasing linearly to over the 50 epochs) in the AdamW optimizer with a weight decay of . The learning rate of the regularization parameter was chosen to be . We included MR images from multiple MEG directions for SWS calculation, as this yields more reliable estimates. Frequency-resolved maps, however, provide a sufficiently accurate basis for training. Their discrepancy from weighted multi-frequency maps according to (7) is 0.19 in terms of (10). While this difference is not negligible, weighted maps can be reconstructed retrospectively from frequency-resolved ones using (7). Furthermore, since the weighting depends on the magnitude of the SWS while our loss (10) measures RMSE, the improvements achieved on the frequency-resolved maps extend directly to the weighted maps. To minimize the influence of artifacts in the SWS maps outside the tissue, a binary brain mask was applied during loss calculation. The application of this mask is crucial, as it ensures that the network focuses primarily on the brain region, preventing it from merely removing artifacts in the non-brain regions. Fine-tuning required approximately 1.5–2 hours per epoch, corresponding to about 4 days for 50 epochs on a single Nvidia A6000 GPU, with 35.4 GB memory usage. Overall, pre-training and fine-tuning together amounted to roughly one week of computation time. After training, the feed-forward time during inference to compute an SWS map from undersampled -space data with 8 vibration phase offsets is approximately 0.75 s.
Methods of Comparison
For comparison, we included two additional reconstruction methods. First, an iterative SENSE [29] reconstruction was employed as a conventional parallel imaging approach. Second, we implemented a TV-regularized reconstruction promoting temporal gradient sparsity of the images, following [43]. The problem was solved using a variable splitting scheme that alternates between CG and PDHG [44], which was implemented with MRpro [35].
Different Acceleration Factors
To address varying trajectories and the resulting undersampling artifacts, the model was trained using trajectories containing one, two, or three spiral arms with 863 -space points each. These correspond to undersampling factors of approximately 19, 9.5, and 6.3, respectively. The trajectories were randomly selected for retrospective undersampling during training, enabling the network to handle different acceleration factors within a unified framework. The trained model was evaluated on simulated data and applied to real acquisitions to demonstrate its robustness and in vivo applicability across varying sampling conditions. Additionally, its performance was evaluated against our comparison methods as introduced in IV-D.
Results
Pre-Training vs. End-to-End Fine-Tuning
As outlined in Section IV-C, the training procedure comprises two stages: an initial pre-training phase, where the model is trained using qualitative MR images as targets, followed by a fine-tuning phase that leverages the pre-trained weights and uses the SWS maps as targets. Here, we illustrate the impact of each stage by comparing the model performance before and after fine-tuning.
Pre-Training
1) Pre-Training: The left section of Figure 2 (up to the vertical dashed line) shows the training and validation loss curves for both reconstructed images and SWS maps throughout the 50-epoch pre-training phase. Blue lines indicate losses for estimated images relative to reference images, while orange lines represent SWS map losses, which are not used in training but monitored for comparison. Both losses decrease steadily, indicating improved model performance for both tasks, with convergence observed near 50 epochs. The image reconstruction loss drops significantly to 56% of its initial value, whereas the SWS map loss decreases more modestly to 83%. While these losses are not directly comparable due to differing loss functions, the observed discrepancy remains noteworthy.
Figure 2. Training and validation loss of complex MR images and SWS maps over 100 epochs with . Perpendicular loss (9) was used for images, RMSE combined with SSIM for SWS maps (10). The first 50 epochs correspond to pre-training with the qualitative images as reference, followed by 50 epochs of fine-tuning with according to (8).
End-to-End Fine-Tuning
2) End-to-End Fine-Tuning: Building upon the converged pre-trained network, the deviation from the SWS map targets can be further reduced through additional fine-tuning within an end-to-end optimization framework. The right segment of Figure 2, beyond the vertical dashed line, shows the training and validation loss curves over an additional 50 epochs, with optimization focused on minimizing SWS loss. This fine-tuning stage yields a further reduction of approximately 5% in the validation loss of the SWS maps. Concurrently, the image reconstruction loss is monitored, revealing a slight increase as the model prioritizes minimizing the SWS error. Given that SWS maps derived from the combination of multiple MEGs yield more accurate representations compared to those obtained from a single MEG, the network was trained using data from all acquired MEGs. However, to maintain consistency and comparability with the image pre-training phase, the loss computed for each individual MEG direction is presented.
Figure 3 shows the improvements in SWS maps, illustrating SSIM, NRMSE, and PSNR for qualitative images and SWS maps with using pre-trained and fine-tuned models applied to the test set undersampled with 1 spiral arm. A negligible model performance decline is observed exclusively for the qualitative images. In contrast, the metrics calculated for the SWS maps reflect improved model performance following fine-tuning.
Figure 3. SSIM, NRMSE, and PSNR computed on the test set with pre-trained and fine-tuned model (, ) applied. The metrics were calculated on qualitative images and SWS with . Except for the qualitative images, the fine-tuned model outperformed the pre-trained one.
Comparison to Standard Methods
Comparison to Standard Methods
We assess the performance of our proposed end-to-end approach in comparison to conventional reconstruction techniques, including iterative SENSE and our TV-regularized comparison method. Figure 4 presents magnitude and phase images of a representative slice, along with the SWS map derived from three MEGs, illustrating the outcomes of different reconstruction methods. Specifically, we compare a straightforward adjoint reconstruction, the pseudo-inverse, the comparison method, and our proposed method against the reference target images. Our approach achieves superior reconstruction quality, delivering clearer and more accurate results than conventional techniques for both qualitative and quantitative images. Table I summarizes the quantitative results for SSIM, NRMSE, and PSNR computed on the masked SWS maps, showing that our approach substantially outperforms all compared methods. While the TV-based comparison method produces clear improvements in magnitude images as seen in Figure 4, the masked SWS maps exhibit only minor enhancements relative to iterative SENSE.
| Spirals | Method | SSIM | NRMSE | PSNR (dB) |
| 1 | Adjoint | 0.75 ± 0.072 | 0.43 ± 0.054 | 21.10 ± 2.96 |
| 1 | it. SENSE | 0.83 ± 0.051 | 0.23 ± 0.040 | 26.79 ± 2.69 |
| 1 | TV | 0.87 ± 0.049 | 0.20 ± 0.041 | 27.81 ± 2.58 |
| 1 | Proposed | 0.93 ± 0.030 | 0.14 ± 0.033 | 30.92 ± 2.34 |
| 2 | Adjoint | 0.83 ± 0.055 | 0.26 ± 0.030 | 25.37 ± 2.88 |
| 2 | it. SENSE | 0.92 ± 0.035 | 0.16 ± 0.036 | 30.10 ± 2.65 |
| 2 | TV | 0.93 ± 0.035 | 0.15 ± 0.036 | 30.61 ± 2.64 |
| 2 | Proposed | 0.96 ± 0.019 | 0.11 ± 0.031 | 33.03 ± 2.39 |
| 3 | Adjoint | 0.88 ± 0.041 | 0.20 ± 0.027 | 27.81 ± 2.61 |
| 3 | it. SENSE | 0.94 ± 0.027 | 0.14 ± 0.036 | 31.30 ± 2.69 |
| 3 | TV | 0.95 ± 0.026 | 0.13 ± 0.037 | 31.83 ± 2.67 |
| 3 | Proposed | 0.97 ± 0.015 | 0.10 ± 0.031 | 34.02 ± 2.44 |
Table I. Comparison of quantitative metrics (SSIM, NRMSE, PSNR) computed from SWS maps reconstructed using different methods, applied to simulated undersampled data with 1, 2, and 3 spirals. Masked metrics are shown, using the brain mask as explained in IV-C.2. Best values shown in bold.
Figure 4. Magnitude images (left), phase images (middle), and resulting SWS maps (right) from reconstructions using various methods on retrospectively undersampled single-spiral data. From top to bottom: Adjoint, iterative SENSE [29], the TV minimization-based comparison method, and the proposed method. Fully sampled data (bottom row) serves as reference.
Different Undersampling Factors
Table I presents SWS map evaluation results for using SSIM, NRMSE, and PSNR metrics on the test set. As expected, performance improves across all metrics with an increasing number of spirals. A substantial gain is observed when moving from one to two spirals. These trends are consistent with the results of the reconstructed qualitative images.
Ablation Study
To evaluate the impact of key design choices of our model, we conducted an ablation study examining two critical parameters: the number of unrolled iterations and the loss weighting coefficient , which balances the combined loss (8). Our analysis reveals that increasing to improves SWS estimation quality by 18% when keeping constant. Regarding the loss weighting, we observe a trade-off ends_mid=1 between SWS accuracy and image quality: setting close to zero yields improved SWS metrics but results in substantial degradation of qualitative image reconstruction (12% when moving from to ), whereas higher values preserve image fidelity at the expense of SWS accuracy. Figure 5 and Table II summarize the described results.
Figure 5. Comparison between different numbers of unrolled iterations ( vs ) as well as different hyperparameters for the loss function (10) used for fine-tuning the entire end-to-end trainable network. Employing a larger yields a lower validation error. Additionally, reducing the effect of catastrophic forgetting with respect to the task of image reconstruction is possible by setting . The proposed parameter combination and yields a good tradeoff between image- and SWS-accuracy.
In Vivo Data
Figure 6 displays the SWS maps obtained by applying the proposed method to the undersampled -space data of the in vivo data for one to three spirals, compared to iterative SENSE, the TV-regularized comparison method, and the reference SWS maps obtained using 25 spirals. The SWS maps displayed were computed with all acquired MEGs, frequencies, and time steps. The resulting SWS maps are sufficiently accurate even with a single spiral; however, they lack finer detail. For two and three spirals, they closely match the reference.
Figure 6. SWS maps resulting from the application of different methods to in vivo data acquired with one, two or three spirals compared to the target SWS maps of an acquisition with 25 spirals.
Discussion
In this work, we presented a novel end-to-end approach for computing SWS maps directly from highly undersampled -space data. In contrast to ultrasound elasticity imaging, where physics-informed neural networks estimate stiffness directly from measured wavefields [20] – which are subject to depth limitations and acoustic artifacts – our MRE framework integrates physics-informed learning into the MR image
| SSIM | NRMSE | PSNR (dB) | |||
| Image | 1 | 0.00 | 0.90 0.041 | 0.16 0.031 | 31.45 2.64 |
| Image | 1 | 0.10 | 0.91 0.042 | 0.14 0.033 | 31.69 2.74 |
| Image | 1 | 0.15 | 0.91 0.041 | 0.14 0.032 | 31.75 2.75 |
| Image | 1 | 0.50 | 0.91 0.041 | 0.14 0.033 | 31.90 2.75 |
| Image | 6 | 0.00 | 0.95 0.030 | 0.10 0.029 | 34.97 3.07 |
| Image | 6 | 0.10 | 0.95 0.029 | 0.09 0.028 | 35.34 3.09 |
| SWS | 1 | 0.00 | 0.90 0.039 | 0.17 0.032 | 29.47 2.38 |
| SWS | 1 | 0.10 | 0.90 0.040 | 0.17 0.032 | 29.35 2.40 |
| SWS | 1 | 0.15 | 0.89 0.040 | 0.17 0.032 | 29.31 2.41 |
| SWS | 1 | 0.50 | 0.89 0.041 | 0.18 0.032 | 29.21 2.42 |
| SWS | 6 | 0.00 | 0.93 0.030 | 0.14 0.033 | 30.92 2.32 |
| SWS | 6 | 0.10 | 0.93 0.030 | 0.14 0.033 | 30.92 2.34 |
Table 2. Reconstruction metrics for different and configurations applied to the 19-fold undersampled test set. SWS maps are calculated for all MEGs and frequencies combined.
reconstruction stage. Here, the SWS maps – derived from an established MRE inversion – guide the reconstruction toward high displacement fidelity under standardized slice-selective encoding and low SNR conditions. This approach naturally extends to 3D acquisitions beyond the 2D implementation presented in this work. This is, to the best of our knowledge, the first end-to-end trainable physics-informed reconstruction network in MRE that jointly performs image reconstruction and SWS estimation directly from -space. We discuss key aspects of this framework in the following sections.
Reconstruction Method Design
Our proposed method integrates -MDEV [25] as a differentiable layer within a model-based deep learning approach for MR image reconstruction, following MoDL [22]. We employ an XTYT architecture [31] to exploit temporal correlations in the phase (compared to magnitude in cardiac cine MRI) across vibration phase offsets, with loss functions [40] designed to preserve both magnitude and phase fidelity critical for MRE. The sampling patterns vary temporally, enabling effective regularization by leveraging redundancies across both spatial and temporal dimensions. In doing so, the reconstruction problem remains relatively small, consisting of 2D MR reconstruction problems that are solved simultaneously. This setup allows for the construction of unrolled methods such as the one considered here, where we used unrolling steps as defined by (7) and (8). Exploiting redundancies along different dimensions (e.g., the -dimension across slices) could be an interesting alternative direction. Given that typically ( vs. in our case), constructing unrolled networks with a larger number of outer iterations might be computationally prohibitive. Setting and opting for a much larger network , as proposed in [30] and [23], and also studied here in the ablations study in V-D, provides computational benefits and certain theoretical advantages (namely, convexity of the implicitly defined reconstruction problem). However, as demonstrated in our ablation study (see Figure 5), we find that choosing yields substantially improved reconstruction performance.
Regarding the choice of the loss function used for end-to-end fine-tuning, setting leads to a very mild form of catastrophic forgetting in the network, as illustrated in Figure 5. In this case, the qualitative image produced by the model deteriorates noticeably – compared to the pre-training results – to enhance the quality of the estimated SWS maps. Although the data-consistent architecture of our model mostly mitigates this decrease in reconstruction quality, we observe that represents an optimal trade-off: the image quality remains largely preserved, while the model continues to prioritize accurate SWS reconstruction without noticeable compromise (see also Table II, with and , SSIM for SWS maps decreases while image quality is mostly the same compared to and ).
Training Strategies
For training, we explored two strategies visualized in Figure 1: using qualitative images alone versus using both SWS maps and qualitative images as targets (see loss function in (10)). The latter enables the reconstruction network to learn representations that facilitate SWS estimation rather than solely optimizing image quality. Our results indicate that, consistent with other studies, e.g. for brain T1-mapping in [24], the achievable performance of reconstruction networks for qMRI problems can be improved if they are additionally trained for the subsequent downstream task – here, SWS quantification – rather than separating image reconstruction problem and SWS estimation, as seen in Figure 3. The level of improvement is expected to vary between different qMRI problems and, although visible here, was less prominent than in [24]. We suspect that this relates to the nature and level of ill-posedness of the considered problem. Future research could explore which problems benefit from this approach and to what extent.
Reconstruction Performance
Compared to conventional image reconstruction techniques such as iterative SENSE [29] and TV minimization [45], our approach consistently produces more accurate qualitative images and SWS maps, as illustrated in Figure 4. Remarkably, all deep learning-based reconstruction approaches tested outperformed classical methods by a large margin, independent of network architecture and size or training strategy, highlighting the effectiveness of incorporating a CNN block within a regularized image reconstruction framework. Furthermore, conventional reconstruction techniques focus solely on producing qualitative images and do not consider the calculation of SWS maps. In contrast, our proposed end-to-end model integrates -MDEV as a fixed, differentiable module, enabling direct estimation of SWS maps within the reconstruction framework. To assess the impact of this integration, we compared our pre-training approach – formulated as a regularized image reconstruction model – against the full end-to-end model. Our results demonstrate that the end-to-end approach outperforms the standalone reconstruction model, reducing SWS map deviation by 5%. These findings suggest that jointly optimizing image reconstruction and SWS estimation within an end-to-end framework better preserves the quantitative information in the phase, which is essential for accurate SWS mapping. This indicates that by incorporating regularization that jointly considers both tasks, the model achieves more reliable and consistent results compared to methods that treat these tasks separately. To achieve real-time MRE down to single-shot spiral acquisition, we trained our model on undersampled data with 1-3 spiral arms, demonstrating its feasibility with up to 19-fold acceleration in data acquisition. This significant decrease in acquisition time minimizes the effects of motion and enables high frame-rate stiffness mapping, making it possible to track dynamic physiological processes. Furthermore, this approach facilitates free-breathing acquisitions, which are particularly beneficial for applications such as abdominal MRE or pediatric scans.
In Vivo & Different Number of Spiral Arms
By varying the data-acquisition protocol using different acceleration factors, we demonstrated that, as anticipated, increasing the number of spirals resulted in progressively improved performance, as shown in Table I. Testing on in vivo data further validated the robustness of our model, despite inevitable variations in spiral trajectories and CSMs that lead to different undersampling artifacts than those in the simulated training data. Despite these differences, the proposed model demonstrated strong generalizability and maintained high performance (Figure 6), confirming its feasibility for potential clinical and real-world applications.
Limitations
While our proposed method demonstrates promising results, several limitations must be considered. First, the training process for our deep learning model requires significant computational resources and time, which can limit its practical application, particularly in resource-constrained environments such as clinical settings. Additionally, like all supervised training-based methods, the approach heavily depends on the availability of high-quality training data, making it difficult to apply in new contexts where such data are scarce or difficult to acquire. As a result, transferring the method to other body parts may present a significant challenge. However, as we have shown, it is possible to train on simulated data and successfully apply the model to newly acquired real data, potentially alleviating some of these limitations. An additional possible solution to this problem could be given by self-supervised approaches that do not require fully sampled reference data, as demonstrated in [46, 47]. Another limitation of this study is the potential presence of measurement errors already present in the reference phase images acquired with 25 spirals, which could likely originate from inaccuracies in the acquisition process, given that spiral trajectories are particularly sensitive to gradient imperfections. These errors may affect the reference SWS maps seen in Figure 6. Addressing these errors, for example, through gradient impulse response function (GIRF) correction, is beyond the scope of this work but will be pursued in future studies. A further limitation is that the evaluation was conducted exclusively on healthy volunteer data. Nevertheless, owing to the data-consistent design of the reconstruction framework and the use of an established stiffness estimation method, we consider it unlikely that the approach would fail in patient cohorts with abnormal anatomy or pathological stiffness. A systematic clinical validation in a larger patient population will, however, be an essential next step to confirm its broader applicability.
Future Work
Future work could explore extending our approach to 3D, which would be computationally feasible with and particularly relevant for addressing the more complex wave fields observed in brain elastography. Further, an exciting direction would be integrating learnable components within the -MDEV framework – for instance, replacing fixed filters with a trainable counterpart or incorporating a data-driven strategy for phase-unwrapping. Additionally, regularization during fine-tuning could be improved – for example, by adding a term penalizing the difference between the weights in pre-trained and fine-tuned models. Alternatively, a LoRA-like approach [48] could be employed, where a low-rank trainable matrix is added parallel to the frozen pre-trained model instead of fine-tuning all parameters, see [49] for an application to MRI. Moreover, our model utilizes the temporal dimension as the primary source of redundant information for regularization. Expanding on this, future work could investigate leveraging correlations within the MEG and frequency dimensions to refine the regularization framework. Further, while in this work we followed an algorithm unrolling strategy, based on the observation that training times amounted to several days, one could opt for other approaches such as Plug and Play [50] or Regularization by Denoising [51], which employ off-the-shelf Gaussian denoisers based on deep NNs, see e.g. [52, 53]. Lastly, while this work focused on rapid brain MRE, future research could apply similar techniques in the abdomen, e.g., the liver, where managing respiratory motion artifacts is critical for the detection and staging of liver fibrosis [54].
Conclusion
We introduced a novel method for estimating SWS maps from highly undersampled -space data by integrating the -MDEV inversion method as a trainable module within a CNN-regularized image reconstruction framework. This data-driven regularization approach demonstrated superior performance in both qualitative and quantitative image reconstruction, outperforming conventional methods such as TV regularization, as shown on a retrospectively undersampled brain dataset. Crucially, embedding -MDEV into the model enables end-to-end optimization of network parameters for the ultimate clinical goal: accurate estimation of tissue mechanical properties via SWS mapping, rather than simply reconstructing intermediate image representations. Although trained exclusively on retrospectively undersampled data, our model successfully generalized to in vivo applications, underscoring its potential for clinical deployment. In particular, it enabled a 19-fold acceleration in data acquisition, facilitating real-time MRE and broader clinical adoption. Future work will explore self-supervised learning strategies to alleviate the dependence on large, labeled training datasets and enhance scalability.
References
- [1]R. Muthupillai et al. Magnetic resonance elastography by direct visualization of propagating acoustic strain waves. Science, 269(5228):1854–1857, 1995.pubmed.ncbi.nlm.nih.gov/7569924
- [2]A. Manduca et al. Magnetic resonance elastography: Non-invasive mapping of tissue elasticity. Medical Image Analysis, 5(4):237–254, 2001.pubmed.ncbi.nlm.nih.gov/11731304
- [3]S. Hirsch et al. Magnetic resonance elastography: Physical background and medical applications. John Wiley & Sons, 2017.
- [4]I. Sack. Magnetic resonance elastography from fundamental soft-tissue mechanics to diagnostic imaging. Nature Reviews Physics, 5(1):25–42, 2023. 10.1038/s42254-022-00543-2.DOI
- [5]M. C. Murphy et al. Regional brain stiffness changes across the Alzheimer’s disease spectrum. NeuroImage: Clinical, 10:283–290, 2016. 10.1016/j.nicl.2015.12.007.DOI
- [6]F. Schrank et al. Real-time MR elastography for viscoelasticity quantification in skeletal muscle during dynamic exercises. Magnetic Resonance in Medicine, 84(1):103–114, 2020. 10.1002/mrm.28095.DOI
- [7]H. Herthum et al. Real-time multifrequency MR elastography of the human brain reveals rapid changes in viscoelasticity in response to the Valsalva maneuver. Frontiers in Bioengineering and Biotechnology, 9:666456, 2021. 10.3389/fbioe.2021.666456.DOI
- [8]T. Meyer et al. Stiffness pulsation of the human brain detected by non-invasive time-harmonic elastography. Frontiers in Bioengineering and Biotechnology, 11:1140734, 2023. 10.3389/fbioe.2023.1140734.DOI
- [9]M. Anders et al. Rapid MR elastography of the liver for subsecond stiffness sampling. Magnetic Resonance in Medicine, 91(1):312–324, 2024. 10.1002/mrm.29859.DOI
- [10]S. Patz et al. Imaging localized neuronal activity at fast time scales through biomechanics. Science Advances, 5(4):eaav3816, 2019. 10.1126/sciadv.aav3816.DOI
- [11]S. Mohammed et al. Phase-regularized and displacement-regularized compressed sensing for fast magnetic resonance elastography. NMR in Biomedicine, 36(7):e4899, July 2023. 10.1002/nbm.4899.DOI
- [12]C. Ebersole et al. A Bayesian method for accelerated magnetic resonance elastography of the liver. Magnetic Resonance in Medicine, 80(3):1178–1188, September 2018. 10.1002/mrm.27083.DOI
- [13]G. Wang et al. Image reconstruction is a new frontier of machine learning. IEEE Transactions on Medical Imaging, 37(6):1289–1296, 2018. 10.1109/TMI.2018.2833635.DOI
- [14]C. Kofler et al. Machine Learning for Quantitative MR Image Reconstruction, pp. 171–213. Springer International Publishing, Cham, 2024.arxiv.org/abs/2402.19396
- [15]N. Aldoj et al. Fully automated quantification of in vivo viscoelasticity of prostate zones using magnetic resonance elastography with Dense U-net segmentation. Scientific Reports, 12(1):2001, 2022. 10.1038/s41598-022-05878-5.DOI
- [16]S. Ma et al. MR Elastography with Optimization-Based Phase Unwrapping and Traveling Wave Expansion-Based Neural Network (TWENN). IEEE Transactions on Medical Imaging, 42(9):2631–2642, 2023. 10.1109/TMI.2023.3261346.DOI
- [17]H. Bustin et al. Elastonet: Neural network-based multicomponent MR elastography wave inversion with uncertainty quantification. Medical Image Analysis, pp. 103642, 2025. 10.1016/j.media.2025.103642.DOI
- [18]J. M. Scott et al. Artificial neural networks for magnetic resonance elastography stiffness estimation in inhomogeneous materials. Medical Image Analysis, 63:101710, 2020. 10.1016/j.media.2020.101710.DOI
- [19]M. C. Murphy et al. Artificial neural networks for stiffness estimation in magnetic resonance elastography. Magnetic Resonance in Medicine, 80(1):351–360, 2018. 10.1002/mrm.27019.DOI
- [20]Z. Yin et al. SWENet: a physics-informed deep neural network (PINN) for shear wave elastography. IEEE Transactions on Medical Imaging, 43(4):1434–1448, 2023. 10.1109/TMI.2023.3338178.DOI
- [21]J. Schluemper et al. A deep cascade of convolutional neural networks for MR image reconstruction. In Information Processing in Medical Imaging: 25th International Conference, IPMI 2017, Boone, NC, USA, June 25-30, 2017, Proceedings 25, pp. 647–658. Springer, 2017. 10.1007/978-3-319-59050-9_51.
- [22]H. Aggarwal et al. MoDL: Model-based deep learning architecture for inverse problems. IEEE Transactions on Medical Imaging, 38(2):394–405, 2018. 10.1109/TMI.2018.2865356.DOI
- [23]A. Kofler et al. An end-to-end-trainable iterative network architecture for accelerated radial multi-coil 2D cine MR image reconstruction. Medical Physics, 48(5):2412–2425, 2021. 10.1002/mp.14809.DOI
- [24]F. F. Zimmermann et al. PINQI: An End-to-End Physics-Informed approach to Learned Quantitative MRI Reconstruction. IEEE Transactions on Computational Imaging, 2024. 10.1109/TCI.2024.3388869.DOI
- [25]H. Tzschätzsch et al. Tomoelastography by multifrequency wave number recovery from time-harmonic propagating shear waves. Medical Image Analysis, 30:1–10, 2016. 10.1016/j.media.2016.01.001.
- [26]J. J. Tamir et al. Computational MRI with physics-based constraints: Application to multicontrast and quantitative imaging. IEEE Signal Processing Magazine, 37(1):94–104, 2020. 10.1109/MSP.2019.2940062.
- [27]M. McGarry et al. Comparison of iterative and direct inversion MR elastography algorithms. In Mechanics of Biological Systems and Materials, Volume 2: Proceedings of the 2011 Annual Conference on Experimental and Applied Mechanics, pp. 49–56. Springer, 2011.DOI
- [28]V. Monga et al. Algorithm Unrolling: Interpretable, Efficient Deep Learning for Signal and Image Processing. IEEE Signal Processing Magazine, 38(2):18–44, 2021. 10.1109/MSP.2020.3016905.arxiv.org/abs/1912.10557
- [29]K. P. Prasanna et al. Advances in sensitivity encoding with arbitrary k-space trajectories. Magnetic Resonance in Medicine, 46(4):638–651, 2001. 10.1002/mrm.1241.
- [30]A. Kofler et al. Neural networks-based regularization for large-scale medical image reconstruction. Physics in Medicine & Biology, 65(13):135003, 2020. 10.1088/1361-6560/ab990e.DOI
- [31]A. Kofler et al. Spatio-Temporal Deep Learning-Based Undersampling Artefact Reduction for 2D Radial Cine MRI With Limited Training Data. IEEE Transactions on Medical Imaging, 39(3):703–717, 2019. 10.1109/TMI.2019.2930318.DOI
- [32]O. Ronneberger et al. U-Net: Convolutional networks for biomedical image segmentation. In Medical image computing and computer-assisted intervention–MICCAI 2015: 18th international conference, Munich, Germany, October 5-9, 2015, proceedings, part III 18, pp. 234–241. Springer, 2015.arxiv.org/abs/1505.04597
- [33]A. Manduca et al. Spatio-temporal directional filtering for improved inversion of MR elastography images. Medical Image Analysis, 7(4):465–473, 2003. 10.1016/S1361-8415(03)00038-0.
- [34]M. A. Schofield and Y. Zhu. Fast phase unwrapping algorithm for interferometric applications. Optics letters, 28(14):1194–1196, 2003. 10.1364/OL.28.001194.pubmed.ncbi.nlm.nih.gov/12885018
- [35]F. F. Zimmermann et al. MRpro - open PyTorch-based MR reconstruction and processing package. arXiv preprint, 2025. 2507.23129.DOI
- [36]F. Dittmann et al. In Vivo Wideband Multifrequency MR Elastography of the Human Brain and Liver. Magnetic Resonance in Medicine, 76(4):1116–1126, 2016. 10.1002/mrm.26006.DOI
- [37]H. Herthum et al. Cerebral tomoelastography based on multifrequency MR elastography in two and three dimensions. Frontiers in Bioengineering and Biotechnology, 10:1056131, 2022. 10.3389/fbioe.2022.1056131.DOI
- [38]K. S. Ravi et al. PyPulseq: A Python Package for MRI Pulse Sequence Design. Journal of Open Source Software, 4(42):1725, 2019. 10.21105/joss.01725.
- [39]A. Kofler et al. The more the merrier?—on the number of trainable parameters in iterative neural networks for image reconstruction. In Proceedings of the Joint Annual meeting of ISMRM-ESMRMB and SMRT 31st Annual Meeting, London, 2022. 10.58530/2022/0051.DOI
- [40]M. L. Terpstra et al. L-loss: A symmetric loss function for magnetic resonance imaging reconstruction and image registration with deep learning. Medical Image Analysis, 80:102509, 2022. 10.1016/j.media.2022.102509.DOI
- [41]S. Martin et al. MR Elastography Image Reconstruction using Spatio-Temporal Neural Networks-based Regularization. In ISMRM & ISMRT Annual Meeting & Exhibition, Singapore, 2024.DOI
- [42]I. Loshchilov and F. Hutter. Decoupled Weight Decay Regularization. arXiv preprint, 5(5):5, 2019. 1711.05101.
- [43]J. Caballero et al. Dictionary learning and time sparsity for dynamic MR data reconstruction. IEEE Transactions on Medical Imaging, 33(4):979–994, 2014. 10.1109/TMI.2014.2301271.DOI
- [44]A. Chambolle and T. Pock. A first-order primal-dual algorithm for convex problems with applications to imaging. Journal of mathematical imaging and vision, 40:120–145, 2011. 10.1007/s10851-010-0251-1.DOI
- [45]K. T. Block et al. Undersampled radial MRI with multiple coils. Iterative image reconstruction using a total variation constraint. Magnetic Resonance in Medicine, 57(6):1086–1098, 2007. 10.1002/mrm.21236.pubmed.ncbi.nlm.nih.gov/17534903
- [46]B. Yaman et al. Zero-Shot Self-Supervised Learning for MRI Reconstruction. In Proceedings of the International Conference on Learning Representations (ICLR), 2022. 2102.07737.arxiv.org/abs/2102.07737
- [47]S. Martin et al. Single-spirit MRI image reconstruction using zero-shot self-supervised deep learning towards real-time stiffness mapping. In ISMRM & ISMRT Annual Meeting & Exhibition, Honolulu, Hawai’i, 2025.
- [48]E. J. Hu et al. LoRA: Low-Rank Adaptation of Large Language Models. ICLR, 1(2):3, 2022. 2106.09685.arxiv.org/abs/2106.09685
- [49]N. Yismaw et al. Domain Expansion via Network Adaptation for Solving Inverse Problems. IEEE Transactions on Computational Imaging, 2024. 10.1109/TCI.2024.3337710.arxiv.org/abs/2310.06235
- [50]S. V. Venkatakrishnan et al. Plug-and-play priors for model based reconstruction. In 2013 IEEE Global Conference on Signal and Information Processing, pp. 945–948. IEEE, 2013. 10.1109/GlobalSIP.2013.6737048.
- [51]Y. Romano et al. The little engine that could: Regularization by denoising (RED). SIAM Journal on Imaging Sciences, 10(4):1804–1844, 2017. 10.1137/16M102884.arxiv.org/abs/1611.02862
- [52]W. Dong et al. Denoising Prior Driven Deep Neural Network for Image Restoration. IEEE Transactions on Pattern Analysis and Machine Intelligence, 41(10):2305–2318, 2018. 10.1109/TPAMI.2018.2873610.DOI
- [53]U. S. Kamilov et al. Plug-and-play methods for integrating physical and learned models in computational imaging: Theory, algorithms, and applications. IEEE Signal Processing Magazine, 40(1):85–97, 2023. 10.1109/MSP.2022.2319955.DOI
- [54]P. Kennedy et al. Quantitative Elastography Methods in Liver Disease: Current Evidence and Future Directions. Radiology, 286(3):738–763, 2018. 10.1148/radiol.2018170601.DOI