Exploring the proximity effect of thin tungsten-aluminum films for phonon-mediated kinetic inductance detectors

Abstract

Lowering the energy threshold of phonon-mediated kinetic inductance detectors is a key step toward next-generation cryogenic calorimeter arrays with large target mass and high sensitivity. In this work, we demonstrate the feasibility of integrating thin tungsten films, so far used for transition-edge sensors, into the fabrication of Kinetic Inductance Detectors (KIDs). Here we present the results from the characterization of a single KID prototype composed of a tungsten-aluminum bilayer of respectively 15 nm and 35 nm thicknesses. The detector was successfully calibrated, and the superconducting film characteristics were measured by operating the device at different temperatures. We show that, by exploiting the proximity effect of superconductors, the device combines the relatively high critical temperature of aluminum with the large kinetic inductance contribution of tungsten. The main feature is a high kinetic inductance, which accounts for 36% of the total inductance, combined with a superconducting critical temperature of around 630 mK.

Introduction

In the past decade, phonon-mediated superconducting cryogenic detectors have been the state-of-the-art technology to detect small energy depositions produced by rare events[1, 2, 3, 4, 5, 6]. These devices are usually composed of a superconductive sensor coupled to a crystal absorber. The phonon mediation allows sensing of a large absorber mass, typically ranging from a fraction to tens of grams, with a small superconductive metal film. Moreover, the low energy cost to produce the signal carriers (phonons and quasiparticles) allows these detectors to achieve thresholds as low as 1 eV[7].

Currently, Transition-Edge Sensors (TESs) based on thin tungsten films[1, 3, 5, 6, 7] are among of the best performing devices in terms of energy threshold and resolution. On the other hand, TESs suffer greatly from mass scaling issues, since their multiplexing is complicated, and currently no phonon-mediated experiment has demonstrated such a feature. For this reason, the use of phonon-mediated KIDs, like the one developed by the Calder, BULLKID and KIPM projects[2, 8, 9], has been investigated and new detector concepts are being proposed. While phonon-mediated KIDs have proven to be multiplexable[10], further developments are needed to improve their energy threshold to be on par with their TES counterpart.

Different strategies can be devised to improve the response of KIDs. One is to explore different detector geometries like the one proposed in Refs.[11, 12]. Another way is to employ different superconductors, as demonstrated in Refs.[13, 14, 15]. Here, we investigate this second strategy by combining the high responsivity of thin tungsten films, typically used for TESs, with the aluminum films typical of KIDs. We realized a W-Al KID phonon-mediated detector, which, to the authors’ knowledge, represents the first device of its kind.

Signal generation in a Kinetic Inductance Detector

A KID consists of a superconducting LC resonator[16], where the inductance LL is the sum of two terms: the magnetic inductance LgL_g and the kinetic inductance LkL_k. An energy deposition EE in the crystal produces athermal phonons that are absorbed by the KID and can break Cooper-pairs, releasing quasiparticles. This process transiently reduces the Cooper-pair density, increasing the kinetic inductance LkL_k and reducing the resonant frequency frf_r[16, 17, 18, 19].

Usually, the transient change of the resonance of a KID is detected by monitoring the phase and magnitude modulation of the complex single pole transmission function S21(y)S_{21}(y) of the resonator[20].

S21(y)=1−QQc11+2jy(1)S_{21}(y)=1-\frac{Q}{Q_c}\frac{1}{1+2jy} \tag*{(1)}

where Q=(Qc−1+Qi−1)−1Q=(Q_c^{-1}+Q_i^{-1})^{-1}, QcQ_c is the coupling quality factor between the KID and the radio-frequency signal, and QiQ_i is the internal quality factor[16], which quantifies the internal losses of the sensor. The parameter y=Qfr−freadoutfry=Q\frac{f_r-f_{\mathrm{readout}}}{f_r} describes the frequency difference between the readout tone and the resonant frequencies in units of the resonance width fr/Qf_r/Q. Usually, the phase readout φ\varphi provides a better signal-to-noise ratio (SNR)[16] with respect to the magnitude one and therefore, it is adopted in this work.

The signal phase responsivity of a KID is [16, 17, 20, 21, 22]:

dφdE=ηVKIDαS2(fr,T)QN0Δ0211+4y2≈ηVKIDαS2(fr,T)QN0Δ02(2)\frac{d\varphi}{dE} = \frac{\eta}{V_{\mathrm{KID}}}\frac{\alpha S_2(f_r,T)Q}{N_0\Delta_0^2}\frac{1}{1+4y^2} \approx\frac{\eta}{V_{\mathrm{KID}}}\frac{\alpha S_2(f_r,T)Q}{N_0\Delta_0^2} \tag*{(2)}

where η\eta is the energy to pair-breaking conversion efficiency, VKIDV_{\mathrm{KID}} is the active inductor volume,

α=LkLk+Lg(3)\alpha= \frac{L_k}{L_k+L_g} \tag*{(3)}

is the kinetic inductance fraction, and S2(fr,T)S_2(f_r,T) is a dimensionless factor of order one from Mattis-Bardeen theory. The approximation in Eq. (2) was done under the reasonable approximation that the readout tone and the resonant frequency of the device are very close and no correction needs to be applied. Finally, N0N_0 is the single-spin volume density of electron states at the Fermi energy and Δ0≃3.52kBTc\Delta_0 \simeq\frac{3.5}{2} k_\mathrm{B}T_c is the superconducting energy gap, with kBk_\mathrm{B} the Boltzmann constant and TcT_c the superconducting critical temperature.

To understand how the superconducting film properties affect the responsivity of a KID, it is worth noting that the kinetic inductance LkL_k of a superconducting film roughly scales as [23]:

Lk∝Lk□∝R□Tc(4)L_k \propto L_k^{\square} \propto\frac{R^{\square}}{T_c} \tag*{(4)}

where R□R^{\square} and Lk□L_k^{\square} are, respectively, the sheet resistance in the non-superconducting state and the sheet kinetic inductance. In Eq. (2), α\alpha and Δ0\Delta_0 are the parameters most strongly influenced by the choice of superconductor. Consequently, the responsivity of a KID scales approximately as:

dφdE∝R□Tc3(5)\frac{d\varphi}{dE} \propto\frac{R^{\square}}{T_c^3} \tag*{(5)}

From the above equation it is clear that choosing a low TcT_c superconductor, possibly with a high R□R^{\square} could greatly enhance the responsivity of a KID. In this context, thin α\alpha-tungsten films (Tc∼15T_c \sim15 mK [1, 24]) emerge as a compelling candidate to boost KID responsivity when compared to the standard aluminum film with Tc∼1.2T_c \sim1.2 K, considering also that RW□/RAl□∼4R^{\square}_{\mathrm{W}}/R^{\square}_{\mathrm{Al}} \sim4 (see [25, 26]). On the other hand, to achieve optimal SNR conditions a KID must be operated at around Tc/6T_c/6 [27], since the minimum temperature reachable with a dry dilution refrigerator is around 10 mK, a Tc≥60T_c \ge60 mK is required. For this reason in this study we produced a tungsten-aluminum bilayer, which depending on the relative film thicknesses would allow to tune the superconductive critical temperature from 15 mK to 1.2 K by exploiting the proximity effect of superconductors [28].

Detector Description The design of the KID used in this study, shown in Figure 1, has an inductive meander (L) area of 4.8 mm2^2 and is inductively coupled with a coplanar waveguide feedline. Both these features make it similar to the Calder17 aluminum device discussed in Ref. 8, whose main parameters are shown in Tab. I, which therefore will be used as comparison. The Calder17 resonator featured high QiQ_i and QcQ_c was designed in order to have high Q. On the other hand, the QcQ_c of the W-Al KID was intentionally lowered to face a possible low QiQ_i due to the new superconductors involved; the reason being that, as visible from Eq. (1), it is advisable to have Q∼QcQ \sim Q_c.

Design and photograph of the KID sensor

Figure 1. Top: Design of the sensor used in this work: in blue the feedline and guard ring are highlighted, while the KID is displayed in red. The letter C and L respectively highlight the capacitor and the inductor of the design. The blue and red areas indicate a different thickness of the metal films as presented in Table III. Bottom: Picture of the detector with its dimensions. In the picture it is visible how the different elements have different film thicknesses.

This resonator was simulated using SONNET [29], which allows to predict the resonant frequency and the coupling quality factor QcQ_c of the resonator, provided the geometry of the KID and the estimation of Lk□L_k^{\square}. Since no reliable expectation for the kinetic inductance was available at the design stage, the first simulation was performed with this value set to zero. This approach provides an upper limit on the expected resonance frequency and accurately describes the coupling quality factor QcQ_c, while QiQ_i is set to infinity. The corresponding results are reported in the first row of Tab. II.

The KID was produced at the microfabrication facilities of the Max-Planck-Institut für Physik where TESs for the CRESST and NUCLEUS experiments [1, 5] are also produced. While the Calder17 aluminum device was produced using an electron-gun evaporator, the W-Al KID presented here was produced by sputter deposition. Tungsten and aluminum were deposited in the same vacuum chamber, without breaking the vacuum between the two depositions. The full KID pattern was then defined by positive lithography and etching. The feedline and ground plane (blue elements in Figure 1, top) were subsequently thickened with a second aluminum deposition structured through a lift-off process. This last step, while not required for operating a KID, was done to reduce phonon absorption by the inhert structures through the phonon recycling phenomenon³⁰. The final result of the production cycle yielded the resonator in Figure 1 (bottom). Using a profilometer, it was also possible to measure the thickness of the deposited metal films which are reported in Tab. III. The thicknesses of the two metal layers were chosen to have Tc∼0.5T_c\sim0.5 K and the same active volume as Calder17.

Calder17W-Al KID
Active Area [mm2^2]4.04.8
VKIDV_{\mathrm{KID}} [10−6 mm310^{-6}\,\mathrm{mm}^3]240240
VMetalV_{\mathrm{Metal}} [10−6 mm310^{-6}\,\mathrm{mm}^3]5801290
Frequency [MHz]∼2600\sim2600715
QQ [k]1478
QcQ_c [k]15611
QiQ_i [k]>2⋅103>2\cdot10^331
α\alpha0.02540.3568
dφdE\frac{d\varphi}{dE} [mrad/keV]5.85.9
Noise RMS [mrad]0.611.02

Table 1. Main characteristics of the Calder17⁸ aluminum device. Anticipating the results of this work, the table contains also all the relevant parameters for the W-Al KID for an easier comparison.

LkL_k [pH]Frequency [MHz]QcQ_c [k]
00891.8±6.8syst891.8\pm6.8^{\mathrm{syst}}7.4±1.0syst7.4\pm1.0^{\mathrm{syst}}
4.33±0.15syst4.33\pm0.15^{\mathrm{syst}}71511.2±1.9syst11.2\pm1.9^{\mathrm{syst}}

Table 2. SONNET simulation parameters and results. The LkL_k parameter is set as input to the simulation, while the resonant frequency and the quality factor are taken from analyzing the simulation results. The estimated statistical errors are negligible, while the systematic errors are evaluated by varying the sensor size of ±1%\pm1\%, which corresponds to the fabrication precision. The case with Lk=0L_k=0 pH is used to estimate the resonator parameters in the absence of kinetic inductance, while the simulation with Lk=4.33L_k=4.33 pH reproduces the measured resonant frequency, as described later in this work. For this reason, in the second simulation the errors are associated to the value of LkL_k and not to the resonant frequency.

Experimental Setup The device was operated in a dilution 3^3He/4^4He cryostat, at a base temperature T0≃30T_0\simeq30 mK. During the operation two magnetic shields were placed around the detector. The first is a case made

ComponentW ThicknessAl Thickness
KID (red)15 nm35 nm
Feedline (blue)15 nm150 nm

Table 3. Metal thicknesses measured using a profilometer after detector fabrication. The colors next to each component refer to the colors in the top panel of Figure 1.

of Cryophy® thermalized at the same temperature as the detector, while the second, made of MuMETAL®, was placed outside the cryostat. This is the usual setup employed when characterizing the Calder and BULLKID detector arrays²,⁸,¹⁰,¹³, which is able to attenuate the vertical component of earth’s magnetic field of around a factor 100. This is required since the presence of a magnetic field, especially directed orthogonally to the superconducting film, lowers the internal quality factor of the resonators.

The detector was mounted into a copper holder and was mechanically fixed in place by Teflon supports. The device was electrically coupled to the readout coaxial line, running through the cryostat from the outside down to the device, by aluminum wire bonding to the two 50 Ω\Omega launchers. The readout line was equipped with 46 dB input attenuation, deployed at various temperature stages, to reduce the effective noise temperature of the system. The output of the line was amplified through an HEMT low-noise amplifier thermalized at the 4 K stage. The radio-frequency signal used to excite the KID was generated and read back at room temperature by a commercial Ettus X300 board³¹ with custom firmware and software³².

To calibrate the detectors, an optical fiber was placed facing the center of the crystal substrate on the side opposite to the lithography. The fiber was pulsed with 400 nm photon bursts generated at room temperature using the LEDs of the LANTERN calibration system³³,³⁴.

Temperature Scan and Film Properties At base temperature, corresponding to around 30 mK, the resonators’ parameters are:

fr30 mK=(715.185±0.010) MHzf_r^{30\,\mathrm{mK}}=(715.185\pm0.010)\ \mathrm{MHz}
Q=8000Q=8000
Qc=10700Q_c=10700
Qi=31700Q_i=31700

where the reported error on fr30 mKf_r^{30\,\mathrm{mK}} is the frequency spacing used for the transmission scan. By comparing the resonant frequency fr30 mKf_r^{30\,\mathrm{mK}} with the simulation performed without kinetic inductance (see first row of Tab. II) it is possible to estimate the kinetic inductance fraction α\alpha of Eq. (6):

α=1−(fr30 mKfrLk=0)2=0.3569±0.0096syst(6)\alpha=1-\left(\frac{f_r^{30\,\mathrm{mK}}}{f_{r}^{L_k=0}}\right)^2=0.3569\pm0.0096^{\mathrm{syst}} \tag*{(6)}

where frLk=0=(891.8±6.8syst)f_r^{L_k=0}=(891.8\pm6.8^{\mathrm{syst}}) MHz is the resonant frequency simulated assuming Lk=0L_k = 0. This value is more than an order of magnitude higher than the one reported for Calder17, amounting to α=0.0254±0.0028\alpha= 0.0254 \pm0.0028, and such an increase is due to the presence of tungsten. By introducing a non-zero value of the kinetic inductance in the simulation to match the measured resonance frequency, one can estimate the sheet kinetic inductance to be Lk=(4.33±0.15syst) pH/□L_k = (4.33 \pm0.15^{\mathrm{syst}})\,\mathrm{pH}/\square (see simulation results in Tab. II), which, in line with the previous result, is around an order of magnitude greater than the typical values for an aluminum film which is approximately 0.5 pH/□0.5\,\mathrm{pH}/\square^{35}.

A second characterization consisted in performing a transmission measurement of the resonator operated at different temperatures. This temperature scan was performed using a heater placed at the coldest cryostat stage, and the detector was operated with a −81-81 dBm readout tone, a value sufficiently low to reduce self-heating effects. The measured quality factors and resonant frequencies from the scan can be used to fit for the Cooper-pair binding energy Δ0\Delta_0 (see Eq. (2)), using the BCS formalism^{22}. In Figure 2, the best-fit results are shown along with the measured data; in the fitting procedure, the value α\alpha was fixed to the previous estimation. Moreover, since 2Δ0≃3.5kBTc2\Delta_0 \simeq3.5k_{\mathrm{B}}T_{\mathrm{c}}, the fitted value shows a critical transition temperature of Tc≃630T_{\mathrm{c}} \simeq630 mK, which, as expected, lies well below the 1.2 K critical temperature of aluminum. It is worth noting that the BCS theory might be limited in the case of inhomogeneous and multilayer superconductors. A more detailed treatment is beyond the scope of this work and can be found in Ref.36.

Figure 2. Shift of the resonant frequency with temperature. The data points are collected by measuring the resonance at each temperature stage. The plot shows the trend of fractional shift of the resonant frequency with respect to the one taken at the lowest temperature (30 mK). The points are fitted using the BCS prediction; the best-fit line and results are shown in the plot. The errors in the plot are given by the frequency resolution used during the measurement of the resonance, and a 3 mK uncertainty is used for the temperature scale.

Calibration

Before performing the calibration of the detector, the signal-to-noise conditions were optimized

both in terms of temperature and operating power. The optimization was done by measuring, for each working condition, the SNR of the reconstructed pulses in the KID following fixed energy depositions of photons from the calibration system. The highest SNR was found to be at a temperature of 110 mK with an operating power of −66-66 dBm. An example pulse acquired in the best SNR condition is shown in Figure 3 along with its time constants and energy.

Figure 3. Pulse acquired during the detector calibration. The pulse reconstructed energy and its time constants are reported as well. The rise time of the pulse is a factor 2 higher than the resonator ring time Q/πfr≈3 μsQ/\pi f_r \approx3\,\mu\mathrm{s}, which can be due to the phonon lifetime in the substrate. The fast decay time instead points to a fast quasiparticle recombination time possibly linked to the low internal QiQ_i.

The detector calibration was performed following the procedure described in Refs.33,34, which consists of studying the broadening of the amplitude distributions produced by LED-induced pulses. A matched filtering algorithm was used to achieve the optimal SNR conditions, as described in Refs.37,38. The results of this calibration are:

dφdE=(5.90±0.21) mradkeVσ0=(172.2±6.4) eV\frac{d\varphi}{dE} = (5.90 \pm0.21)\,\frac{\mathrm{mrad}}{\mathrm{keV}} \qquad\sigma_0 = (172.2 \pm6.4)\,\mathrm{eV}

where dφdE\frac{d\varphi}{dE} is the responsivity of the device and σ0\sigma_0 is the calibrated noise level after the application of the matched filter. The responsivity ratio between this detector and Calder17 is:

\frac{\left.\frac{d\varphi}{dE}\right|_{\mathrm{W-Al}}}{\left.\frac{d\varphi}{dE}\right|_{\mathrm{Calder17}}_{\mathrm{Measured}}} = \frac{5.9}{5.8} \sim1 \tag*{(7)}

The expected responsivity ratio can be obtained by substituting the parameters of Tab. I in Eq. (2) for both the W-Al KID and Calder17:

dφdE∣W−AldφdE∣Calder17Expected=(ηαS2(fr,T)QN0Δ02)W−Al(ηαS2(fr,T)QN0Δ02)Calder17∼1.6(8)\frac{\left.\frac{d\varphi}{dE}\right|_{\mathrm{W-Al}}}{\left.\frac{d\varphi}{dE}\right|_{\mathrm{Calder17}}}_{\mathrm{Expected}} = \frac{\left(\frac{\eta\alpha S_2(f_r,T)Q}{N_0\Delta_0^2}\right)_{\mathrm{W-Al}}}{\left(\frac{\eta\alpha S_2(f_r,T)Q}{N_0\Delta_0^2}\right)_{\mathrm{Calder17}}} \sim1.6 \tag*{(8)}

where N0Calder17=1.72⋅1010 eV−1 μm−3N_0^{\mathrm{Calder17}} = 1.72 \cdot10^{10}\,\mathrm{eV}^{-1}\,\mu\mathrm{m}^{-3} and N0W−Al=1.46⋅1010 eV−1 μm−3N_0^{\mathrm{W-Al}} = 1.46 \cdot10^{10}\,\mathrm{eV}^{-1}\,\mu\mathrm{m}^{-3} were calculated by following Ref. 28,39, while S2(fr,T)Calder17=3.02S_2(f_r,T)^{\mathrm{Calder17}} = 3.02 and S2(fr,T)W−Al=3.20S_2(f_r,T)^{\mathrm{W-Al}} = 3.20 and ηW−Al/ηCalder17=0.45\eta_{\mathrm{W-Al}}/\eta_{\mathrm{Calder17}} = 0.45. The contribution of ηW−Al/ηCalder17\eta_{\mathrm{W-Al}}/\eta_{\mathrm{Calder17}} to Eq. (8) is the one affected by the greatest uncertainty since the phonon transmission properties between silicon and tungsten are not known; assuming them similar to the Si-Al interface of Calder17, the ratio is then dominated by the phonon absorption competition between the active and the passive volumes of the sensors[40], meaning that:

ηW−AlηCalder17∼(VKIDVMetal)W−Al(VKIDVMetal)Calder17=0.45\frac{\eta_{\mathrm{W-Al}}}{\eta_{\mathrm{Calder17}}} \sim\frac{\left(\frac{V_{\mathrm{KID}}}{V_{\mathrm{Metal}}}\right)_{\mathrm{W-Al}}}{\left(\frac{V_{\mathrm{KID}}}{V_{\mathrm{Metal}}}\right)_{\mathrm{Calder17}}} = 0.45

where VMetalV_{\mathrm{Metal}} is the sum of the active volume VKIDV_{\mathrm{KID}} and the remaining metalized volume of the passive structures of the sensor.

The discrepancy between the measured value of 1 and the expected value of 1.6 of the responsivity ratio then points to a lower phonon transmission between Si and W with respect to the Si-Al case. It is possible to roughly estimate the phonon transmission in the two cases by using the acoustic wave propagation formalism described in Ref. 41, which gives a 1.9 times smaller transmission for Si-W than for the case of Si-Al. This shifts the expectation of the responsivity ratio of Eq. (8) to 0.8, in line with the measured value.

Conclusions and Outlook. This work presents the first realization of a kinetic inductance detector integrating tungsten thin-film technology typical of transition-edge sensors. The sensor, based on a sputtered aluminum-tungsten bilayer, shows a superconductive band gap energy of about 95 μ\mueV, corresponding to a superconducting critical temperature of 630 mK. The device was successfully fabricated and operated in a dry dilution cryostat, achieving a high kinetic inductance fraction of 36%, about a factor of 10 above typical aluminum values. These results highlight the strong potential of this approach for highly responsive kinetic inductance detectors. The measured responsivity of the W-Al KID is 5.9 mrad keV−1\mathrm{mrad}\,\mathrm{keV}^{-1} and is currently limited by the low resonator quality factor. While a low coupling quality factor was chosen by design, the internal quality factor remains below that of aluminum KIDs such as Calder17, possibly due to the sputtering procedure of the superconducting films. As visible from Eq. (8), reaching Calder17-like quality factors is expected to increase the responsivity up to a factor of 20, mostly due to the α\alpha and Δ0\Delta_0 contributions. Several clear paths can further improve this proof of concept device. The first is to raise the resonator quality factor through fabrication optimization. In addition,

tuning the aluminum and tungsten thicknesses can better optimize the critical temperature. The phonon collection efficiency is still to be measured, and improving the interface metal between tungsten and the crystal substrate, possibly turning to a Al-W-Al trilayer, should enhance phonon transmission. Finally, more advanced KID geometries and dedicated phonon-collecting structures[11] can further improve responsivity and noise performance. Taken together, these results provide strong motivation to pursue follow-up studies and fully exploit the potential of this technology.

Acknowledgments

This work was supported by the INFN, Sapienza University of Rome, the Max-Planck-Institut für Physik, and co-funded by the Deutsche Forschungsgemeinschaft (DFG, German Research Foundation) through the Sonderforschungsbereich (Collaborative Research Center) SFB1258 and by the European Union (ERC, DANAE, Grant No. 101087663). Views and opinions expressed are however those of the authors only and do not necessarily reflect those of the European Union or the European Research Council. Neither the European Union nor the granting authority can be held responsible for them. We thank A. L. De Santis for the useful discussions on the phonon transmission properties.

References

  1. [1]G. Angloher et al., “Doublets detectors to investigate the cresst low energy background: results from above-ground prototypes,” The European Physical Journal C 84, 1001 (2024).
  2. [2]D. Delicato et al., “Low-energy spectrum of the bulkid detector array operated on surface,” Eur. Phys. J. C 84, 353 (2024).DOI
  3. [3]R. Anthony-Petersen et al., “Low energy backgrounds and excess noise in a two-channel low-threshold calorimeter,” Applied Physics Letters 126, 102601 (2025).
  4. [4]M. Albakry et al. (SuperCDMS Collaboration), “Search for low-mass electron-recoil dark matter using a single-charge sensitive supercdms-hvev detector,” Phys. Rev. D 113, 032001 (2026).DOI
  5. [5]H. Abele et al. (NUCLEUS Collaboration), “Commissioning of the nucleus experiment at the technical university of munich,” Phys. Rev. D 112, 072013 (2025).DOI
  6. [6]G. Angloher et al. (COSINUS Collaboration), “Deep-underground dark matter search with a cosinus detector prototype,” Phys. Rev. D 110, 043010 (2024).DOI
  7. [7]R. Anthony-Petersen, C. L. Chang, Y.-Y. Chang, et al., “Low energy backgrounds and excess noise in a two-channel low-threshold calorimeter,” Applied Physics Letters 126, 102601 (2025).
  8. [8]L. Cardani et al., “High sensitivity phonon-mediated kinetic inductance detector with combined amplitude and phase read-out,” Applied Physics Letters 110, 033504 (2017).DOI
  9. [9]D. J. Temples, O. Wen, K. Ramanathan, et al., “Performance of a phonon-mediated kinetic inductance detector at the nexus cryogenic facility,” Phys. Rev. Appl. 22, 044045 (2024).DOI
  10. [10]D. Delicato et al., “Background of the bulkid detector array operated with moderate shield on surface,” (2026), arXiv:2605.02468.arxiv.org/abs/2605.02468
  11. [11]L. Pesce et al., “Preliminary tests of a phonon-mediated kinetic inductance detector with phonon-funneling volume,” Nuovo Cimento della Società Italiana di Fisica C 49 (2026), 10.1393/ncc/i2026-26021-8.DOI
  12. [12]L. Pesce et al., “Enhanced athermal phonon responsivity in a kinetic inductance detector with integrated phonon collectors,” Applied Physics Letters 128, 163504 (2026).
  13. [13]L. Cardani et al., “Final results of calder: kinetic inductance light detectors to search for rare events,” The European Physical Journal C 81, 636 (2021).DOI
  14. [14]J. Gao, M. R. Vissers, M. O. Sandberg, et al., “A titanium-nitride near-infrared kinetic inductance photon-counting detector and its anomalous electrodynamics,” Applied Physics Letters 101, 142602 (2012).DOI
  15. [15]X. Li, A. Suzuki, and M. Garcia-Sicreveres, “Low tc hafnium kinetic inductance device with high internal quality factor,” (2025), arXiv:2502.19818 [hep-ex].arxiv.org/abs/2502.19818
  16. [16]J. Zmuidzinas, “Superconducting microresonators: Physics and applications,” Annual Review of Condensed Matter Physics 3, 169–214 (2012).DOI
  17. [17]B. Mazin, Microwave Kinetic Inductance Detectors, Phd thesis, California Institute of Technology (2005).
  18. [18]M. Khalil et al., “An analysis method for asymmetric resonator transmission applied to superconducting devices,” Journal of Applied Physics 111, 054510 (2012).DOI
  19. [19]L. J. Swenson et al., “Operation of a titanium nitride superconducting microresonator detector in the nonlinear regime,” Journal of Applied Physics 113, 104501 (2013).DOI
  20. [20]D. Delicato et al., “Germanium target sensed by phonon-mediated kinetic inductance detectors,” Applied Physics Letters 126, 153502 (2025).DOI
  21. [21]D. Moore et al., “Position and energy-resolved particle detection using phonon-mediated microwave kinetic inductance detectors,” Applied Physics Letters 100, 232601 (2012).DOI
  22. [22]J. Gao et al., “Equivalence of the effects on the complex conductivity of superconductor due to temperature change and external pair breaking,” Journal of Low Temperature Physics 151, 557–563 (2008).DOI
  23. [23]A. J. Annunziata et al., “Tunable superconducting nanoinductors,” Nanotechnology 21, 445202 (2010).DOI
  24. [24]G. Anglhofer, S. Banik, A. Bento, et al., “Optimization of test design for the cresset experiment,” IEEE Transactions on Applied Superconductivity 36, 1–7 (2026).
  25. [25]A. Kaidatzis, V. Psycharis, K. Mergia, and D. Niarchos, “Annealing effects on the structural and electrical properties of sputtered tungsten thin films,” Thin Solid Films 619, 61–67 (2016).DOI
  26. [26]M.-H. Lee and R. Collier, “The sheet resistance of thin metallic films and stripes at both dc and 130 ghz,” Microelectronic Engineering 73–74, 916–919 (2004), micro and Nano Engineering 2003.DOI
  27. [27]E. Sattistelli, F. Bellini, C. Bucci, et al., “Calder: neutrinoless double-beta decay identification in teo2 bolometers with kinetic inductance detectors,” The European Physical Journal C 75, 353 (2015).
  28. [28]A. Catalano, A. Goupy, J. le Sueur, H. et al., “Bi-layer kinetic inductance detectors for space observations between 80–120 ghz,” A&A 580, A15 (2015).DOI
  29. [29]Software, “Sonnet software website,” https://www.sonnetsoftware.com/, accessed: 2025-07-29.
  30. [40]M. Martinez, L. Cardani, N. Casali, et al., “Measurements and simulations of athermal phonon transmission from silicon absorbers to aluminum sensors,” Phys. Rev. Appl. 11, 064025 (2019).DOI

Paper details

vlnc:W_wjh49jff
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 10

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…
Contents