Three-Dimensional Ultrasound Localization Microscopy Using Hadamard-Encoded Multipulse and 4D Denoising on a 256-Channel System with a 1024-element Matrix Array: A Feasibility Study in Human Kidney Transplants

Abstract

Three-dimensional ultrasound localization microscopy (3D ULM) enables volumetric visualization of microvasculature beyond the acoustic diffraction limit, but its clinical translation remains limited by long acquisition times, low microbubble (MB) signal-to-noise ratio (SNR), and insufficient MB contrast, particularly on ultrasound systems with a limited number of acquisition channels. In this study, we developed an in-human 3D ULM framework for a 256-channel research ultrasound system equipped with a 1024-element matrix array. The framework combines diverging-wave Hadamard-encoded multipulse (HEM) transmission to enhance MB SNR with denoising strategy combining MB enhancement and four-dimensional non-local means filtering to improve MB contrast. The framework was implemented on a research ultrasound platform and evaluated in phantom experiments and four in vivo human kidney transplants. In phantom experiments, HEM improved the MB contrast ratio around 2 dB, while the proposed denoising strategy enhanced in vivo MB contrast by up to around 9 dB. 3D ULM reconstruction achieved a spatial resolution of approximately 84 um, as estimated by Fourier shell correlation. The number of reconstructed microvascular voxels increased exponentially with acquisition duration, reaching approximately 74% of the estimated saturated vessel density using 3000 volumes (~12 s). The subject with the highest histopathological injury exhibited quantitatively reduced microvascular vessel density, flow velocities, and mean perfusion index. These results demonstrate that the proposed framework enables high-quality volumetric super-resolution imaging and quantitative characterization of renal microvasculature using a 256-channel ultrasound system, providing a practical step toward the clinical translation of in vivo 3D ULM.

Paper details

vlnc:W_wkb3mrwj
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
arXiv nonexclusive license
Usage

views · 10

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…