Integrated optomechanical ultrasonic sensors with nano-Pascal-level sensitivity

Abstract

Ultrasonic sensors are widely used for object detection and localization in underwater and biological settings. The operational range and spatial resolution are inherently limited by sensor sensitivity, in which conventional piezoelectric transducers have been overwhelmed by advanced photonic sensors. Here, we demonstrate an optomechanical ultrasonic sensor integrated into a photonic platform, which comprises a suspended SiO2 membrane embedded with a high-Q Si3N4 microring resonator. By exploiting simultaneous optical and mechanical resonances, the sensor achieves a record low noise-equivalent pressure (NEP) of 218 nPa/Hz^1/2 at 289 kHz in air and 9.6 nPa/Hz^1/2 at 52 kHz in water. We demonstrate its versatility through photoacoustic gas spectroscopy in air and underwater ultrasound imaging, achieving a minimum detectable C2H2 concentration of 2.9 ppm (integration time 1 s) and an imaging resolution of 1.89 mm, respectively. Our work represents a significant advancement in compact CMOS-compatible ultrasound sensing, unlocking new possibilities in biomedical imaging, environmental monitoring, industrial testing, and underwater communications.

Ultrasound sensing has become a cornerstone of modern diagnostic and monitoring technologies, with crucial applications spanning medical imaging[1, 2], non-destructive testing[3, 4], underwater acoustics[5], and industrial process monitoring[6, 7]. Conventional piezoelectric transducers[8, 9, 10], though widely used, face critical challenges in miniaturization, integration, and sensitivity, limiting their applications in scenarios that require compact designs or high resolutions[11, 12]. These constraints are particularly evident in miniaturized systems such as portable medical devices[13] and intravascular ultrasound probes[14, 15], where sensor arrays are often restricted to sub-centimeter scale to meet spatial and operational requirements. Recent advances in micromachined sensors, such as piezoelectric micromachined ultrasound transducers (PMUTs)[16] and capacitive micromachined ultrasound transducers (CMUTs)[17, 18, 19], have improved array density and electronic integration. However, these devices remain susceptible to electromagnetic interference and exhibit inadequate sensitivities required for next-generation applications. For instance, their inability to operate close to high-voltage transmission lines hinders the monitoring of corona discharge noise[20]. Furthermore, their insufficient sensitivity poses significant challenges in applications like cranial ultrasound imaging[21, 22].

Optical microcavity-based ultrasonic sensors have emerged as a transformative alternative[10, 23, 24], offering unprecedented sensitivity, immunity to electromagnetic interference, broad bandwidth, and chip-scale integration capability. Fiber-based microcavity sensors, including Fabry-Pérot cavities[25, 26, 27], Bragg gratings[15, 28, 29, 30], and microsphere cavities[31, 32, 33], have shown particular promise in endoscopy applications but face scalability challenges due to limited arraying capabilities and reliance on bulky optical components[34]. The growing demand for miniaturized, high-performance sensing systems has spurred significant research toward chip-scale sensing platforms[35], where CMOS-compatible microring resonators show significant promise. Various microring resonator platforms, fabricated from polymer[36, 37, 38], silicon[39, 40], and chalcogenide[41], have been developed for ultrasound sensing. Despite their broad detection bandwidth, these integrated sensors achieved noise equivalent pressures (NEPs) only in the mPa/Hz\sqrt{\mathrm{Hz}} range, primarily due to their modest optical quality (QQ) factors and restricted mechanical displacements caused by the thick substrates.

Optomechanical sensors offer a powerful approach for ultrasensitive ultrasound detection by harnessing the strong interaction between light and mechanical vibrations in high-QQ microresonators. Suspended microdisk resonators have achieved unprecedented low NEPs at the μ\muPa/Hz\sqrt{\mathrm{Hz}} level[42, 43, 44]. However, the reliance on fiber-taper coupling presents significant challenges in practical applications, including alignment complexity, reduced robustnessness, and difficulty working in aqueous environments. Here, we develop a novel integrated, mass-produced, ultrasensitive optomechanical ultrasonic sensor with nano-Pascal-level sensitivity and further demonstrate the sensor’s versatility through photoacoustic gas spectroscopy and underwater ultrasound imaging. Beyond prototype demonstrations, this fully integrated, CMOS-compatible, highly sensitive ultrasonic sensor enables scalable production and paves the way for future multi-scene sensing applications, advancing medical diagnostics and industrial precision measurements.

Results

Principle

As depicted in Figure 1a, the proposed integrated ultrasonic sensor comprises a suspended SiO2\mathrm{SiO_2} membrane and a high-QQ Si3N4\mathrm{Si_3N_4} microring resonator. The SiO2\mathrm{SiO_2} membrane is clamped at its periphery to the silicon substrate and suspended at its center to maximize mechanical displacement under ultrasound excitation. The Si3N4\mathrm{Si_3N_4} microring is embedded within the SiO2\mathrm{SiO_2} membrane, enabling optomechanical readout by transducing the mechanical displacement of the membrane into optical resonance shifts of the ring resonator. To maximize the readout signal, we systematically optimize the geometries of the sensor, yielding final radii of 450 μm450\ \mu\mathrm{m} for the membrane and 235 μm235\ \mu\mathrm{m} for the microring resonator (see Methods and Extended Data Figure 1). As shown in the right inset of Figure 1a, by locking the frequency of an input continuous wave

Schematic and working principle of the integrated ultrahigh-sensitivity optomechanical ultrasonic sensor

Figure 1. Schematic and working principle of the integrated ultrahigh-sensitivity optomechanical ultrasonic sensor. a Schematic of the ultrasonic sensor, where the intensity of the continuous wave (CW) input laser is modulated by the ultrasound. Left inset: Schematic of the mechanical resonance-enhanced displacement response. Right inset: Optical readout principle of the sensor, illustrating the optical resonance-enhanced sensitivity. b Schematic comparison of integrated microring resonator-based ultrasonic sensors with three configurations: (i) non-suspended structure, (ii) suspended structure (off mechanical resonance), and (iii) suspended structure (on mechanical resonance), illustrating the enhanced response of the suspended structure with mechanical resonance.

Device design

Extended Data Figure 1. Device design. a Schematic of the sensor with key geometric parameters. b Microring radius variation δR\delta R for different microring-to-membrane radius ratios Rring/RmemR_{\mathrm{ring}}/R_{\mathrm{mem}}. Inset: Cross-sectional profile of the membrane’s first-order flapping mode ν00\nu_{00}. c,d Simulated intrinsic QQ factor of the Si3N4\mathrm{Si_3N_4} microring resonator in water, as functions of (c) the microring width and (d) SiO2\mathrm{SiO_2} silica cladding thickness. e Measured transmission spectrum of an optical resonance of the Si3N4\mathrm{Si_3N_4} microring resonator (blue dot), along with its Lorentzian-fitting (red curve), indicating an intrinsic optical QQ factor of 1.35×1061.35 \times10^6 in air.

(CW) laser on the blue-detuned side of the optical resonance, the ultrasound-induced displacement is converted into the laser intensity modulation. The higher the optical QQ factor, the steeper the slope of the transmission spectrum, and the greater the readout sensitivity.

Mechanical resonance further amplifies the displacement of the SiO2\mathrm{SiO_2} membrane, significantly enhancing the sensitivity to acoustic signals. The bottom left inset of Figure 1a compares the displacement response of a mechanical oscillator at different frequencies, illustrating the amplification of the mechanical resonance. Figure 1b presents a schematic comparison of integrated ultrasonic sensors with three configurations: (i) conventional non-suspended structure, suspended structure operating at (ii) off-mechanical and (iii) on-mechanical resonance frequencies, highlighting the mechanical resonance-enhanced sensor response.

Sensor design and fabrication

We fabricate the devices at the wafer scale, with the process flow outlined in Figure 2a. The fabrication begins with electron beam lithography (EBL) and reactive ion etching (RIE) to create the Si3N4\mathrm{Si_3N_4} microring resonator, with both processes optimized to achieve near-vertical sidewalls (Extended Data Figure 2). A SiO2\mathrm{SiO_2} cladding layer is deposited on the Si3N4\mathrm{Si_3N_4} layer to preserve resonator optical performance, with Figure 2b showing a photograph of a full 4-inch wafer after the deposition. We then flip the wafer upside down and perform photolithography and deep reactive ion etching (DRIE) processes to selectively etch the silicon underneath Si3_3N4_4 microring resonators. Figure 2c presents a photograph of a 3 cm×\times3 cm chip after releasing the suspended SiO2_2 membranes. For full integration and enhanced robustness, two mode conversion fibers are used to couple light into the microring resonator and are co-packaged with the device (Figure 2d). This results in a compact and portable ultrasonic sensor with reliable operation in diverse environments.

Sensor fabrication and characterization

Figure 2. a Fabrication process of the sensor. b Photograph of the processed 4-inch wafer. c Photograph of a 3 cm×\times3 cm chip after Si deep etching. d photograph of the packaged sensor. e Optical microscopy image of the sensor, showing a SiO2_2 membrane with an embedded Si3_3N4_4 microring resonator. Inset: Optical mode profile ∣E∣|\mathbf{E}|, with E\mathbf{E} denoting the electric field. f Normalized transmission spectrum of the microring resonator measured in water.

Sensor characterization

Extended Data Figure 2. Sensor characterization. a Scanning electron micrograph (SEM) of the Si3N4\mathrm{Si_3N_4} microring resonator before SiO2\mathrm{SiO_2} cladding layer deposition. b SEM of the etched Si3N4\mathrm{Si_3N_4} microring and its coupling section with the adjacent bus waveguide. c Cross-sectional view SEM of Si3N4\mathrm{Si_3N_4} waveguide structure atop the SiO2\mathrm{SiO_2} undercladding layer. d SEM of Si3N4\mathrm{Si_3N_4} waveguide core fully encapsulated by the SiO2\mathrm{SiO_2} cladding layer. The false-colored region highlights the Si3N4\mathrm{Si_3N_4} waveguide.

Device characterization

Figure 2e shows an optical microscopy image of the sensor, with the optical field distribution shown in the inset. The microring resonator exhibits an optical QQ factor of 1.33×1061.33 \times10^{6} in water (Figure 2f), which is similar to that of 1.35×1061.35 \times10^{6} in air (Extended Data Figure 1), confirming that the packaging robustly preserves the microring optical performance in various operational environments.

The experimental set-up for sensitivity characterization is depicted in Extended Data Figure 3. Specifically, a 1550 nm laser is coupled into the microring resonator and detuned to the optimal slope of the optical resonance. A pre-calibrated ultrasound transducer, driven by a vector network analyzer (VNA) and positioned above the

Experimental setup for ultrasound sensitivity measurements

Extended Data Figure 3. Experimental setup for ultrasound sensitivity measurements. UT: ultrasound transducer; PD: photodetector; OSC: oscilloscope; ESA: electronic spectrum analyzer; VNA: vector network analyzer.

sensor, serves as the ultrasound source. The ultrasound-induced optical resonance shift modulates the transmitted light intensity and is detected by a photodetector. The transmission spectrum of the optical mode, noise power spectral density (PSD), and the ultrasound response of the sensor are measured by an oscilloscope, electronic spectrum analyzer (ESA) and VNA, respectively.

Noise PSD across the first-order flapping mode ν(0,0)\nu_{(0,0)} is measured in both air and water, with the measured results and Lorentzian fittings shown in grey dots and black curves in Figure 3a and Figure 3b, respectively. Notably, ν(0,0)\nu_{(0,0)} mode (with its mechanical profile shown in the insets of Figure 3a and Figure 3b) exhibits the greatest spatial overlap with incident ultrasound from above the sensor and therefore is expected to have better sensitivity. Near the mechanical resonance frequency, thermal noise—arising from both surrounding molecule collisions and intrinsic damping of the mechanical resonator—dominates as Lorentzian peaks (orange dashed curves). Resonance frequencies of the ν(0,0)\nu_{(0,0)} mode in air and water are 289 kHz and 52 kHz with linewidths of 10 kHz and 2 kHz, respectively. The resonance frequency shifts downward in water compared to air due to the increased effective mass and viscosity[45]. Away from the mechanical resonance frequency, thermorefractive noise (TRN) dominates (green dashed curves), originating from temperature fluctuations induced refractive index variation via the thermo-optic effect[46]. As investigated in previous works, achieving thermal noise-limited performance is essential for enhanced sensitivity[42] (Supplementary Information IA).

Characterization of the ultrasonic sensor

Figure 3. Characterization of the ultrasonic sensor. a,b Noise power spectral densities (PSDs) of the ν(0,0)\nu_{(0,0)} mode in air (a) and water (b), with insets showing the ν(0,0)\nu_{(0,0)} mode profile of the SiO2_2 membrane. FWHM: full width at half maximum. c,d Ultrasound response of the sensor at different frequencies in air (c) and water (d). Inset in (c): PSD of the sensor in air when the applied ultrasound frequency is on resonance (288 kHz) and off resonance (268 kHz) with the ν(0,0)\nu_{(0,0)} mechanical mode, with the grey curve representing the noise floor. e,f Noise-equivalent pressure (NEP) spectral densities around the ν(0,0)\nu_{(0,0)} mode in air (e) and water (f). The achieved minimum NEPs are 218 nPa/Hz\sqrt{\mathrm{Hz}} at 289 kHz in air and 9.6 nPa/Hz\sqrt{\mathrm{Hz}} at 52 kHz in water, respectively. Insets: NEP spectral densities between 20 kHz-1 MHz in air (e) and water (f).

The sensor’s single-frequency linear dynamic range (LDR) at mechanical resonance is approximately 52 dB in both air and water (Supplementary Information ID), and all subsequent ultrasound pressures are maintained within this linear range. The ultrasound response of the sensor is measured in air and water by sweeping the applied frequency to the transducer using the VNA, with the results shown in red and blue dots in Figure 3c and Figure 3d, respectively. The observed spectral peak matches the mechanical mode identified in Figure 3a and Figure 3b, confirming that the mechanical mode amplifies the membrane’s displacement response. As illustrated in the inset of Figure 3c, the on-resonance case achieves a signal-noise-ratio (SNR) of 42 dB at 30 Hz resolution bandwidth (RBW), a 7 dB improvement over the off-resonance case (35 dB). Deviations in the off-resonance frequency regions from the fitting curves (red and blue curves in Figure 3c and Figure 3d, respectively) using mechanical susceptibility functions result from interference between responses of two mechanical modes (Supplementary Information section IB). Additionally, the observed response undulations are caused by sound waves reflections between the ultrasound transducer and the sensor chip (Supplementary Information section IB).

The NEPs of the sensor at different frequencies are derived from the noise spectra, single-frequency response, and response spectra, as shown in Figure 3e (air) and 3f (water). The NEP spectra exhibit minima at mechanical resonance frequencies, consistent with theoretical predictions (Supplementary Information sections IA and IC). Remarkably, our sensor achieves optimal NEPs of 218 nPa/Hz\sqrt{\mathrm{Hz}} at 289 kHz in air and 9.6 nPa/Hz\sqrt{\mathrm{Hz}} at 52 kHz in water, representing a record in microcavity-based ultrasonic sensors. The superior sensitivity in water compared to that in air likely arises from improved acoustic impedance match at the water-sensor interface. We also measured the NEPs in a larger frequency range from 20 kHz to 1 MHz, displayed in the insets of Figure 3e and Figure 3f, where the shaded areas correspond to the ν(0,0)\nu_{(0,0)} mode, and the pentagrams mark the best sensitivities.

Device reproducibility is confirmed across nine sensors from the same batch, showing 1% variation in mechanical resonance frequencies and 38% deviations in sensitivities (Supplementary Information IIA). Furthermore, the ν(0,0)\nu_{(0,0)} mode demonstrates excellent directional uniformity, with a 3 dB angular bandwidth exceeding 100° (Supplementary Information IIB). Crucially, suspended-membrane sensors outperform non-suspended (without DRIE process) devices by two orders of magnitude in NEP (Extended Data Figure 4), highlighting the critical role of membrane suspension.

Performance comparison of non-suspended and suspended sensor structures in water

Extended Data Figure 4. Performance comparison of non-suspended and suspended sensor structures in water. a Schematic illustrations of the non-suspended (left) and suspended (right) sensor configurations. b-d Comparative results of noise power spectra densities (PSDs) (b), frequency-dependent ultrasound responses (c), and noise-equivalent pressures (NEPs) (d) of the sensors with both non-suspended (black curves) and suspended (blue curves) structures.

Photoacoustic gas spectroscopy

Highly sensitive air-coupled ultrasonic sensors are indispensable for gas detection in applications ranging from breath analysis and environmental monitoring to hazardous gas detection[47]. Among spectroscopic techniques, photoacoustic spectroscopy (PAS) has emerged as a powerful approach[48, 49], leveraging the photoacoustic effect where gas molecules convert intensity-modulated light into acoustic waves through periodic thermal expansion. PAS offers unique advantages, including background-free detection, high sensitivity, and broad spectral coverage, making it particularly suitable for trace gas analysis.

We demonstrate the application of our highly sensitive integrated ultrasonic sensor in PAS of C2H2\mathrm{C_2H_2} gas, with the experimental setup shown in Figure 4a (details in Extended Data Fig. 5 and Methods). Specifically, a pump laser with its intensity modulated by an electro-optic modulator (EOM), is injected into the gas cell containing a mixture of 1% C2H2\mathrm{C_2H_2} and 99% N2\mathrm{N_2}, while the integrated ultrasonic sensor within the gas cell detects the generated acoustic waves. PA signals are measured at two modulation frequencies: 288 kHz (on mechanical resonance) and 268 kHz (off resonance), at different pump powers. As the PA signal scales linearly with pump power[50], the noise-equivalent concentration (NEC) is inversely proportional to the pump power (Figure 4b). At 350 mW pump power and 1 s integration time, NECs of 2.9 ppm (on resonance) and 13.1 ppm (off resonance) are achieved, demonstrating significant sensitivity enhancement through mechanical resonance. The inset of Figure 4b quantifies this improvement, showing SNRs of 61 dB at 288 kHz (on resonance) versus 48 dB at 268 kHz (off resonance). The background-free nature of PAS is confirmed by comparing PA signals from C2H2/N2\mathrm{C_2H_2/N_2} mixture and pure N2\mathrm{N_2} (Supplementary Information IIIA), with PA responses exclusively emerging at C2H2\mathrm{C_2H_2} absorption wavelengths. The PA spectrum of our sensor is obtained by scanning the pump laser wavelength with a step of 0.01 nm, as shown in the blue curve of Figure 4c, which matches well with the high-resolution transmission molecular absorption database (HITRAN) simulations (red curve). The relative residuals between the two are below 5% (black curve), validating the system’s spectroscopic accuracy.

Demonstrations of photoacoustic gas spectroscopy and underwater ultrasound imaging

Figure 4. Demonstrations of photoacoustic gas spectroscopy and underwater ultrasound imaging. a Experimental setup for photoacoustic detection of gaseous acetylene (C2H2\mathrm{C_2H_2}). EOM: electro-optic modulator, PD: photodetector. b Noise-equivalent concentration (NEC) as a function of pump power. Inset: Photoacoustic signal for 1% C2H2\mathrm{C_2H_2} mixed with 99% N2\mathrm{N_2} at 350 mW pump power. c Measured (blue) and HITRAN-simulated (red) absorption spectra of C2H2\mathrm{C_2H_2} from 1530 nm to 1533 nm. Bottom panel: Residual between experimental and simulated spectra. d Schematic of the underwater ultrasound transmission imaging system. Inset: Cross-sectional view of the sample structure. e (i,ii) Ultrasound images at 0.3 mPa pressure with an ultrasound frequency of (i) 517 kHz (on mechanical resonance) and (ii) 550 kHz (off mechanical resonance). (iii) Ultrasound image from a commercial hydrophone, with an ultrasound frequency of 517 kHz and pressure of 0.7 Pa.

Underwater ultrasound imaging

Ultrasound imaging technology is a powerful non-destructive technique widely used in medical diagnostics, industrial inspection, and marine acoustics. We demonstrate these capabilities through underwater ultrasound imaging using our highly sensitive sensors. As depicted in Figure 4d, the experimental setup includes a water tank containing a 500 kHz focused ultrasound transducer, our sensor, and an acrylic sample featuring an ”F” shaped groove. When immersed in water, the groove traps air, creating a localized acoustic impedance mismatch. The sample is raster-scanned using a two-dimensional horizontal translational stage.

When the ultrasound beam generated by the ultrasound transducer is focused on the ”F” region, the air-filled groove reflects the ultrasound waves, resulting in significant attenuation of the transmitted ultrasound signal. This generates a strong contrast between the groove and the surrounding homogeneous regions, enabling high-resolution imaging. For optimal performance, we operate the sensor at the ν(3,0)\nu_{(3,0)} mechanical mode of 517 kHz (mode profile shown in the inset of Extended Data Fig. 5b), achieving higher spatial resolution than the ν(1,0)\nu_{(1,0)} due to its shorter wavelength.

The sample is scanned with a 1 mm step size, and the sensor’s response is recorded at each position. Figure 4e(Figure 4i) and Figure 4e(ii) represent the imaging results obtained with a driving ultrasound pressure of 0.3 mPa and driving frequencies of 517 kHz (on-resonance) and 550 kHz (off-resonance), respectively, revealing significantly improved contrast and clarity under resonant conditions, a direct consequence of mechanical resonance-enhanced sensitivity. Further quantitative analysis (Supplementary Information IIIB) confirms a spatial resolution of 1.89 mm, demonstrating the potential of our sensor for high-resolution imaging and long-range target detection in underwater applications. Remarkably, our sensor outperforms a commercial hydrophone even at ultrasound pressures three orders of magnitude lower (0.3 mPa versus 0.7 Pa), as shown in Figure 4e(iii).

Conclusion and discussion

Figure 5 compares the NEP versus sensing area AA for various ultrasonic sensors. Our sensor achieves the highest sensitivity (lowest NEP) while maintaining a compact footprint. As the NEP of optomechanical sensors scales approximately as 1/A1/\sqrt{A}, we evaluate the figure of merit NEP×A\mathrm{NEP}\times\sqrt{A} across different technologies. Remarkably, our sensor achieves NEP×A\mathrm{NEP}\times\sqrt{A} values at the 10−810^{-8} Pa mm/Hz1/2\mathrm{Hz}^{1/2} level, surpassing conventional electric and optical sensors by several orders of magnitude, highlighting its compelling advantages in both miniaturization and sensitivity. A comprehensive performance comparison with existing ultrasonic sensors is provided in Supplementary Information IV.

Performance benchmarking of various ultrasonic sensors

Figure 5. Performance benchmarking of various ultrasonic sensors. Noise-equivalent pressures (NEP) as a function of the sensing area AA. See Supplementary Information IV for data references.

Several key directions will drive further development of integrated ultrasound sensing technology. First, the sensitivity of the sensor can be further enhanced by geometric optimization. The working bandwidth can be significantly expanded by suppressing photorefractive noise to enable thermal-noise-limited sensitivity across wider frequency ranges and utilizing multiple mechanical modes for broader spectral coverage. Additionally, advances in photonic integrated circuits[51, 52, 53] enables co-integration of complementary on-chip components, including lasers[54, 55], spectrometers[56], and photodetectors[57], facilitating fully integrated sensor systems. Beyond device optimization, we envision practical applications of these highly sensitive sensors in multimodal sensing and wearable technologies, transitioning these sensitive platforms from laboratory prototypes to field-deployable solutions for mobile sensing[58]. Our work establishes a foundation for next-generation compact, high-performance sensing systems with transformative potential across diverse domains, from biomedical diagnostics and industrial monitoring to real-time environmental surveillance.

Methods

Device design

We systematically optimize the device geometry to enhance ultrasound sensitivity (Extended Data Figure 1). The key design parameter is the radius ratio Rring/RmemR_{\mathrm{ring}}/R_{\mathrm{mem}}, where RringR_{\mathrm{ring}} and RmemR_{\mathrm{mem}} denote the radii of the microring and membrane, respectively. Our optimization focuses on the first-order flapping mode ν(0,0)\nu_{(0,0)}, as it has the largest spatial overlap with the incident ultrasound from above the sensor. The mechanical mode profile of ν(0,0)\nu_{(0,0)} is derived by modeling the SiO2_2 membrane as a thin plate[59], yielding the normalized displacement:

w0,0=J0(λ00RringRmem)−J0(λ00)I0(λ00)I0(λ00RringRmem),(1)w_{0,0}=J_0\left(\lambda_{00}\frac{R_{\mathrm{ring}}}{R_{\mathrm{mem}}}\right)-\frac{J_0(\lambda_{00})}{I_0(\lambda_{00})}I_0\left(\lambda_{00}\frac{R_{\mathrm{ring}}}{R_{\mathrm{mem}}}\right), \tag*{(1)}

where J0J_0 and I0I_0 represent the 0-th Bessel function and the modified Bessel function of the first kind, respectively. λ00\lambda_{00} is a dimensionless parameter which is related to the mode shape, determined from Ref.[60]. The normalized radial change ΔR\Delta R is calculated as[61]:

ΔR=−Hpos∂w0,0∂Rring,(2)\Delta R=-H_{\mathrm{pos}}\frac{\partial w_{0,0}}{\partial R_{\mathrm{ring}}}, \tag*{(2)}

where HposH_{\mathrm{pos}} is the vertical position of the waveguide. Extended Data Figure 1 shows ΔR\Delta R versus Rring/RmemR_{\mathrm{ring}}/R_{\mathrm{mem}}, revealing a maximum at Rring/Rmem=0.52R_{\mathrm{ring}}/R_{\mathrm{mem}}=0.52. Considering an EBL write field of 500 µm, we implement Rring=450R_{\mathrm{ring}}=450 µm and Rmem=235R_{\mathrm{mem}}=235 µm.

We systematically investigate the intrinsic QQ factor of a Si3_3N4_4 microring resonator with a radius of 235 µm and a thickness of 265 nm, as a function of microring width and SiO2_2 cladding thickness (Extended Data Figure 1,Extended Data Figure 1). The intrinsic QQ factor exhibits strong geometric dependence, showing a sharp decline for microring width below 6 µm (Extended Data Figure 1) while remaining high for cladding thickness above 2 µm (Extended Data Figure 1). While thicker cladding enhances optical QQ factor, it comprises mechanical compliance and thus decreases sensitivity. We therefore optimize the design with 6 µm microring width and 2 µm cladding thickness, achieving both intrinsic QQ factors (1.35×106^6 in air, Figure 1e) and environmental stability without sacrificing sensitivity.

Device fabrication

The ultrasonic sensors are fabricated on a 4-inch wafer comprising a 265 nm-thick Si3_3N4_4 layer on a 4 µm-thick buried-oxide layer above a 500 µm-thick silicon substrate. The fabrication begins with patterning Si3_3N4_4 microring resonators and bus waveguides using electron beam lithography (EBL) with AR-P 6200 resist, followed by reactive ion etching (RIE) employing CHF3_3/O2_2 plasma. A protective 2-µm-thick SiO2_2 cladding layer is subsequently deposited via inductively coupled plasma chemical vapor deposition (ICP-CVD) to shield the microring resonator from environmental contamination. We then flipped the chip upside down and patterned the SiO2_2 membrane by double-sided aligned photolithography To ensure precise concentric alignment between the Si3_3N4_4 microring and SiO2_2 membrane structures, we perform double-sided aligned photolithography after flipping the wafer upside down. The SiO2_2 membrane release is achieved through a Bosch process utilizing alternating C4_4F8_8 passivation and SF6_6 etching cycles to remove the underlying silicon substrate. Finally, the completed devices are diced and packaged with mode conversion fibers to create compact, portable sensor units.

Ultrasound sensitivity characterization

A 1550 nm laser (TOPTICA CTL1500) is coupled into the microring resonator, with its frequency detuned relative to the optical mode to maintain 25% transmission power for optimal sensitivity[62]. The transmitted optical signal is detected using a photodetector (KEYANG KY-PF-RM-10M-1-FC) and recorded via an oscilloscope. Ultrasound waves are generated by a pre-calibrated ultrasound transducer[43], positioned above the sensor. Ultrasound pressure calibration at the sensor location is performed using: (i) a needle hydrophone (Onda HNR-1000) for aqueous measurements, and (ii) a scanning laser vibrometer (Sunnyinnovation optical Intelligence LV-SC400) for measurements in air. For single-frequency response characterization, the ultrasound transducer is driven by a sinusoidal signal generated by an arbitrary function generator, with the sensor response recorded using an electrical spectrum analyzer (ROHDE&SCHWARZ FPS4). Broadband frequency response of the sensor is measured via a vector network analyzer (ROHDE&SCHWARZ ZNL3), which sweeps ultrasound frequencies while recording the corresponding sensor response.

Experimental details for photoacoustic gas spectroscopy

Photoacoustic gas spectroscopy measurements are performed at room temperature and atmospheric pressure using a sealed gas cell with fiber-optic feedthroughs for optical access. The cell features a gas inlet port connected to acetylene (C2_2H2_2) and nitrogen (N2_2) cylinders for precise gas mixing, and an outlet port connected with a built-in barometer for pressure monitoring. Before measurements, the gas cell is purged with 1% C2_2H2_2 mixed with 99% N2_2, by opening the C2_2H2_2 supply valve (valve 1) and vent valve (valve 2) to ensure complete air displacement and concentration accuracy. During measurements, both valves are closed to maintain a static gas environment, with identical procedures followed for N2_2 measurements. A CW pump laser (TOPTICA CTL1550) is used to excite the PA signals of the gas, with its intensity modulated by an electro-optic modulator (Exail MX-LN-10). The modulation frequency is controlled by a function generator, while the modulation depth is adjusted via a DC voltage. Before entering the gas cell, the pump power is amplified by an erbium-doped fiber amplifier (Amonics AEDFA-37-R-FA) and converted to a free-space beam with a 1-2 mm diameter spot using a fiber collimator. The integrated ultrasonic sensor, positioned below the optical path, detects the resulting PA signal of the gas.

Data availability

The data that support the plot within this paper and other findings of this study are available upon publication.

Code availability

The codes that support the findings of this study are available upon publication.

Acknowledgments

The authors thank Professors Yun-Feng Xiao, Zhixiong Gong, and Xue-Feng Jiang for useful discussions, and Prof. Tuo Liu for calibrating the ultrasound pressure in air. This work was supported by the Synergetic Extreme Condition User Facility (SECUF, https://cstr.cn/31123.02.SECUF).

Funding: Innovation Program for Quantum Science and Technology (2023ZD0301100); National Natural Science Foundation of China (NSFC) (62222515, 12174438, 11934019, 12274438); The National Key Research and Development Program of China (2021YFA1400700, 2021YFB3501400); Basic frontier science research program of Chinese Academy of Sciences (ZDSL-Y- JSC003); CAS Project for Young Scientists in Basic Research (YSBR-100).

Author contributions

Experiments were conceived and designed by X.C., H.Y., M.W., B-B.L., and Q-F.Y. Measurements and data analysis were performed by X.C., H.Y., and M.W., with assistance from Z-G.H. and Z-L. W. Devices were designed by X.C., H.Y., and M.W. Devices were fabricated by X.C., H.Y., and M.W., with assistance from Y.W. and C.L. The project was supervised by B-B.L. and Q-F.Y. All authors participated in preparing the manuscript.

Competing interests

The authors declare no competing interests.

Additional information

Supplementary information is available for this paper.

Extended data is available for this paper.

Correspondence and requests for materials should be addressed to B-B.L.

Experimental setup for the photoacoustic gas spectroscopy measurement

Figure Extended Data Fig. 5. Experimental setup for the photoacoustic gas spectroscopy measurement. PC: polarization controller; AFG: arbitrary function generator; EOM: electro-optical modulator; EDFA: erbium-doped fiber amplifier; PD: photodetector; ESA: electronic spectrum analyzer.

References

  1. [1]Hu, H. et al. A wearable cardiac ultrasound imager. Nature 613, 667–675 (2023).DOI
  2. [2]Wang, C. et al. Bioadhesive ultrasound for long-term continuous imaging of diverse organs. Science 377, 517–523 (2022).DOI
  3. [3]Schumacher, T. et al. Confirmation of the scanpyramids north face corridor in the great pyramid of giza using multi-modal image fusion from three non-destructive testing techniques. Sci. Rep. 15, 9275 (2025).DOI
  4. [4]Gervillié-Mouravieff, C., Bao, W., Steingart, D. A. & Meng, Y. S. Non-destructive characterization techniques for battery performance and life-cycle assessment. Nat. Rev. Electr. Eng. 1, 547–558 (2024).
  5. [5]Gao, Y. et al. Hydrogen microphones for stealthy underwater listening. Nat. Commun. 7, 12316 (2016).DOI
  6. [6]Zhao, J. et al. Self-focusing high-frequency ultrasonic transducers for non-destructive testing applications. Sci. Rep. 15, 8845 (2025).DOI
  7. [7]Wasylowski, D. et al. Operando visualisation of lithium plating by ultrasound imaging of battery cells. Nat. Commun. 15, 10237 (2024).DOI
  8. [8]Schwarz, M. et al. Optoacoustic dermoscopy of the human skin: Tuning excitation energy for optimal detection bandwidth with fast and deep imaging in vivo. IEEE Trans. Med. Imaging 36, 1287–1296 (2017).DOI
  9. [9]Cho, S. et al. An ultrasensitive and broadband transparent ultrasound transducer for ultrasound and photoacoustic imaging in-vivo. Nat. Commun. 15, 1444 (2024).DOI
  10. [10]Wissmeyer, G., Pleitez, M. A., Rosenthal, A. & Ntzi-achristos, V. Looking at sound: optoacoustics with all-optical ultrasound detection. Light Sci. App. 7, 53 (2018).DOI
  11. [11]Qiu, Z., Lu, Y. & Qiu, Z. Review of ultrasonic ranging methods and their current challenges. Micromachines 13 (2022).DOI
  12. [12]Winkler, A. M., Maslov, K. I. & Wang, L. V. Noise-equivalent sensitivity of photoacoustics. J. Biomed. Opt. 18, 097003 (2013).DOI
  13. [13]Jiang, X. et al. Monolithic ultrasound fingerprint sensor. Microsyst. Nanoeng. 3, 17059 (2017).DOI
  14. [14]Wang, J., Zheng, Z., Chan, J. & Yeow, J. T. W. Capacitive micromachined ultrasound transducers for intravascular ultrasound imaging. Microsyst. Nanoeng. 6, 73 (2020).DOI
  15. [15]Wang, L. et al. Ultrawide-bandwidth high-resolution all-optical intravascular ultrasound using miniaturized photoacoustic transducer. Sci. Adv. 9, eadg8600 (2023).DOI
  16. [16]Wang, H. et al. MEMS ultrasound transducers for endoscopic photoacoustic imaging applications. Micromachines 11 (2020).DOI
  17. [17]Ilkhechi, A. K., Ceroici, C., Li, Z. & Zemp, R. Transparent capacitive micromachined ultrasonic transducer (CMUT) arrays for real-time photoacoustic applications. Opt. Express 28, 13750–13760 (2020).DOI
  18. [18]Dew, E. B., Kashani Ilkhechi, A., Maadi, M., Haven, N. J. M. & Zemp, R. J. Outperforming piezoelectric ultrasonics with high-reliability single-membrane CMUT array elements. Microsyst. Nanoeng. 8, 59 (2022).DOI
  19. [19]Chen, K., Lee, H.-S., Chandrakasan, A. P. & Sodini, C. G. Ultrasonic imaging transceiver design for CMUT: A three-level 30-Vpp pulse-shaping pulser with improved efficiency and a noise-optimized receiver. IEEE J. Solid-State Circuits 48, 2734–2745 (2013).DOI
  20. [20]Fischer, B. Optical microphone hears ultrasound. Nat. Photon. 10, 356–358 (2016).DOI
  21. [21]Ren, J., Li, J., Chen, S., Liu, Y. & Ta, D. Unveiling the potential of ultrasound in brain imaging: Innovations, challenges, and prospects. Ultrasonics 145, 107465 (2025).DOI
  22. [22]Chen, T. et al. Enhanced ultrasound transmission through skull using flexible matching layer with gradual acoustic impedance. ACS Appl. Mater. Interfaces 15, 55510–55517 (2023).DOI
  23. [23]Cao, X., Yang, H., Wu, Z.-L. & Li, B.-B. Ultrasound sensing with optical microcavities. Light Sci. App. 13, 159 (2024).DOI
  24. [24]Bao, K. et al. Photoacoustic imaging sensors based on integrated photonics: Challenges and trends. Las. Photon. Rev. n/a, 2400414.DOI
  25. [25]Xu, F. et al. Fiber-optic acoustic pressure sensor based on large-area nanolayer silver diaphragm. Opt. Lett. 39, 2838–2840 (2014).DOI
  26. [26]Guggenheim, J. A. et al. Ultrasensitive plano-concave optical microresonators for ultrasound sensing. Nat. Pho- ton 11, 714–719 (2017).
  27. [27]Ma, J. et al. Flexible microbubble-based Fabry-Pérot cavity for sensitive ultrasound detection and wide-view photoacoustic imaging. Photonics Res. 8, 1558–1565 (2020).DOI
  28. [28]Shnaiderman, R. et al. Fiber interferometer for hybrid optical and optoacoustic intravital microscopy. Optica 4, 1180–1187 (2017).DOI
  29. [29]La, T. A., Ülgen, O., Shnaiderman, R. & Ntziarchristos, V. Bragg grating etalon-based optical fiber for ultrasound and optoacoustic detection. Nat. Commun. 15, 7521 (2024).DOI
  30. [30]Shnaiderman, R. et al. A submicrometre silicon-on-insulator resonator for ultrasound detection. Nature 585, 372–378 (2020).DOI
  31. [31]Yang, J. et al. Multiphysical sensing of light, sound and microwave in a microcavity brillouin laser. Nanophotonics 9, 2915–2925 (2020).DOI
  32. [32]Sun, J., Meng, J.-W., Tang, S.-J. & Li, C. An encapsulated optical microsphere sensor for ultrasound detection and photoacoustic imaging. Science China Physics, Mechanics & Astronomy 65, 224211 (2021).DOI
  33. [33]Meng, J.-W. et al. Dissipative acousto-optic interactions in optical microcavities. Phys. Rev. Lett. 129, 073901 (2022).DOI
  34. [34]Ansari, R., Zhang, E. Z., Desjardins, A. E. & Beard, P. C. All-optical forward-viewing photoacoustic probe for high-resolution 3d endoscopy. Light Sci. App. 7, 75 (2018).DOI
  35. [35]Ma, X. et al. Integrated microcavity electric field sensors using pound-drever-hall detection. Nat. Commun. 15, 1386 (2024).DOI
  36. [36]Zhang, C., Ling, T., Chen, S.-L. & Guo, L. J. Ultrabroad bandwidth and highly sensitive optical ultrasonic detector for photoacoustic imaging. ACS Photon. 1, 1093–1098 (2014).DOI
  37. [37]Dong, B. et al. Isometric multimodal photoacoustic microscopy based on optically transparent micro-ring ultrasonic detection. Optica 2, 169–176 (2015).
  38. [38]Li, H. et al. Disposable ultrasound-sensing chronic cranial window by soft nanoimprinting lithography. Nat. Commun. 10, 4277 (2019).DOI
  39. [39]Leinders, S. M. et al. A sensitive optical micro-machined ultrasound sensor (omus) based on a silicon photonic ring resonator on an acoustical membrane. Sci. Rep. 5, 14328 (2015).DOI
  40. [40]Westerveld, W. J. et al. Sensitive, small, broadband and scalable optomechanical ultrasound sensor in silicon photonics. Nat. Photon. 15, 341–345 (2021).DOI
  41. [41]Pan, J. et al. Parallel interrogation of the chalcogenide-based micro-ring sensor array for photoacoustic tomography. Nat. Commun. 14, 3250 (2023).DOI
  42. [42]Basiri-Esfahani, S., Armin, A., Forster, S. & Bowen, W. P. Precision ultrasound sensing on a chip. Nat. Commun. 10, 132 (2019).
  43. [43]Yang, H. et al. High-sensitivity air-coupled megahertz-frequency ultrasound detection using on-chip microcavities. Phys. Rev. Appl. 18, 034035 (2022).DOI
  44. [44]Yang, H. et al. Micropascal-sensitivity ultrasound sensors based on optical microcavities. Photonics Res. 11, 1139–1147 (2023).arxiv.org/abs/2211.07965
  45. [45]Vančura, C., Dufour, I., Heinrich, S. M., Josse, F. & Hierle​mann, A. Analysis of resonating microcantilevers operating in a viscous liquid environment. Sens. Actuators A Phys. 141, 43–51 (2008).DOI
  46. [46]Huang, G. et al. Thermorefractive noise in silicon-nitride microresonators. Phys. Rev. A 99, 061801 (2019).DOI
  47. [47]Barik, P. & Pradhan, M. Selectivity in trace gas sensing: recent developments, challenges, and future perspectives. Analyst 147, 1024–1054 (2022).DOI
  48. [48]Xiong, S. et al. Photoacoustic spectroscopy gas detection technology research progress. Appl. Spectrosc. 78, 139–158 (2024). PMID: 37936290.DOI
  49. [49]Palzer, S. Photoacoustic-based gas sensing: A review. Sensors 20 (2020).DOI
  50. [50]Wang, F. et al. Techniques to enhance the photoacoustic signal for trace gas sensing: A review. Sens. Actuators A Phys. 345, 113807 (2022).DOI
  51. [51]Shu, H. et al. Microcomb-driven silicon photonic systems. Nature 605, 457–463 (2022).DOI
  52. [52]Tran, M. A. et al. Extending the spectrum of fully integrated photonics to submicrometre wavelengths. Nature 610, 54–60 (2022).DOI
  53. [53]Hu, Y. et al. Integrated electro-optics on thin-film lithium niobate. Nat. Rev. Phys. 7, 237–254 (2025).arxiv.org/abs/2404.06398
  54. [54]Xiang, C. et al. High-performance lasers for fully integrated silicon nitride photonics. Nat. Commun. 12, 6650 (2021).DOI
  55. [55]Snigirev, V. et al. Ultrafast tunable lasers using lithium niobate integrated photonics. Nature 615, 411–417 (2023).
  56. [56]Finco, G. et al. Monolithic thin-film lithium niobate broadband spectrometer with one nanometre resolution. Nat. Commun. 15, 2330 (2024).DOI
  57. [57]Peng, Y. et al. An 8 × 160 Gb s-1 all-silicon avalanche photodiode chip. Nat. Photon. 18, 928–934 (2024).DOI
  58. [58]Bylinkin, A. et al. On-chip phonon-enhanced IR near-field detection of molecular vibrations. Nat. Commun. 15, 8907 (2024).DOI
  59. [59]Tufoi, M., Gillich, G.-R., Praisach, Z.-I., Ntakpe, J. L. & Hatiegan, C. An analysis of the dynamic behavior of circular plates from a damage detection perspective. Romanian Journal of Acoustics and Vibration 11, 41–46 (2014).
  60. [60]Leissa, A. W. Vibration of plates, vol. 160 (Scientific and Technical Information Division, National Aeronautics and Space Administration, 1969).
  61. [61]Zunic, M. et al. Design of a micro-opto-mechanical ultrasound sensor for photoacoustic imaging. In 2020 21st International Conference on Thermal, Mechanical and Multi-Physics Simulation and Experiments in Microelectronics and Microsystems (EuroSimE), 1–8 (2020).DOI
  62. [62]Hu, Z.-G. et al. Picotesla-sensitivity microcavity optomechanical magnetometry. Light Sci. App. 13, 279 (2024).DOI

Paper details

vlnc:W_xxeh493b
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 0

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…
Contents