CMOS based high-resolution dynamic X-ray imaging with inorganic perovskite

Abstract

High-resolution dynamic X-ray detector is crucial for time-resolved digital radiography (DR) imaging and fast 3D medical computed tomography (CT) imaging. Recently, perovskites have become promising alternatives to conventional semi-conductor materials, e.g., Si, a-Se and CdTe, for direct X-ray detection. However, the feasibility of their combination with high-speed pixelated complementary metal-oxide-semiconductor (CMOS) arrays remains unknown. This work originally reports an innovative direct-conversion X-ray detector fabricated with 300 micrometer thick inorganic perovskite film printed on a tailored CMOS array. In-house measurements demonstrate that the CsPbBr3 film has excellent optoelectric properties of an electron mobility-lifetime product of 3.40x10−5^{-5} cm2^2 V−1^{-1}, and the X-ray detector exhibits high sensitivity of 9341uC Gyair−1_{\rm air}^{-1} cm−2^{-2}, and low detection limit of 588 nGyair−1_{\rm air}^{-1}. This CMOS X-ray imaging detector achieves a high spatial resolution up to 5.5 lp/mm (close to the resolution limit of 6.0 lp/mm), and >300 frame per second (fps) readout speed. DR image of a resolution pattern phantom and a anesthesia mice, CT images of a biological specimen are acquired for the first time.

Since its discovery in 1895, X-ray imaging has shown superb performance in unveiling the anatomical structures of the human body. By far, X-ray imaging has been widely used in many medical applications, such as the diagnosis and treatment of cardiovascular and cancer diseases[1, 2]. In order to generate sufficient X-ray imaging information for clinicians to make precise diagnoses, usually, X-ray detectors with unique features such as high spatial resolution, fast imaging speed are highly demanded. In addition, to minimize its cancerous damage to human body, its low dose X-ray imaging performance is also an important factor to be considered to use as low as reasonably achievable (ALARA) radiation dose[3].

Over the past two decades, dedicated academic and clinical investigations have been performed to develop the most appropriate X-ray detectors for ultra-high quality medical imaging. Results demonstrate the direct-conversion type X-ray detectors made of semi-conductor materials show better spatial and temporal resolution at low radiation dose X-ray imaging than the indirect-conversion ones made of scintillator materials[4]. However, the current semi-conductor materials used in commercial direct-conversion type X-ray detectors are less satisfactory for generic X-ray imaging purposes. For example, the amorphous selenium (a-Se)[5, 6] only works at low energy X-ray photons (<40<40 keV) due to its low stopping power (Z=34Z=34), which are merely restricted for breast imaging. Moreover, the cadmium zinc telluride (CdZnTe)[7] or cadmium telluride (CdTe) crystals have shown its advancements in fabricating high-energy (>140>140 keV) X-ray detectors with energy-resolving capability. However, the CdZnTe/CdTe are are difficult to grow into large dimensions, and thus are hard to be used for large area X-ray imaging. The high cost also prohibits their wide spreads in medical X-ray imaging applications. In addition, the requirement of flip chip assembly using Indium solder bumps to interconnect the CdZnTe/CdTe crystal slab to the back-end ASIC readout circuit complicates the detector fabrication as well.

As an emerging candidate, metal halide perovskites[8, 9, 10, 11], e.g., single crystal, polycrystalline wafer and solvent-processed thick film, show excellent potentials in realizing high-sensitive (≥50000 μC Gyair−1cm−2\ge50000\ \mu\mathrm{C}\ \mathrm{Gy}_{\mathrm{air}}^{-1}\mathrm{cm}^{-2}) direct X-ray detection, such as high X-ray absorption, high charge carrier mobility (μ\mu) and long carrier lifetime (τ\tau). By integrating with pixelated sensor arrays, prototypes of perovskite-based X-ray imaging detectors have been successfully fabricated. In 2017, Park et al. first reported the promise of high-resolution X-ray imaging with blade-coating CH3NH3PbI3\mathrm{CH_3NH_3PbI_3} films on a large-scale thin-film transistor (TFT) backplane[12]. In 2021, Sarah et al. reported a two-step procedure to manufacture X-ray detector with microcrystalline CH3NH3PbI3\mathrm{CH_3NH_3PbI_3} on TFT backplane[13]. Recently, Tang et al. fabricated a flat-panel X-ray imager by combination of soft-pressing CH3NH3PbI3\mathrm{CH_3NH_3PbI_3} films and TFT backplane[14].

Technically, the direct-conversion perovskite X-ray detectors having the most ideal imaging performance should be fabricated upon the Complementary Metal-Oxide-Semiconductor (CMOS) sensors[15], which have become the majority of consumer and prosumer cameras these days due to their much more distinct advantages[16, 17, 18] in achieving ultra-high image resolution and data readout speed[19]. For example, the pixel size of CMOS sensor can be easily made smaller than 5 micrometers, whereas, the pixel size of TFT array is usually larger than 70 micrometers. Besides, the CMOS sensor also has stronger noise immunity and lower static power utilization. As a consequence, CMOS sensor arrays should be selected to completely fulfill the advancements of perovskites in achieving high-end direct-conversion X-ray imaging with unprecedented spatial and temporal resolution. Moreover, special electrical circuits can be designed and integrated into the CMOS pixel unit to suppress the leakage dark current of the perovskites[20], which is still a remaining issue in developing X-ray imaging detector array[21]. Unfortunately, by far no such studies have been demonstrated on a CMOS sensor based perovskite X-ray detector.

In this work, we report an innovative direct-conversion X-ray detector fabricated with 300 micrometer thick inorganic CsPbBr3\mathrm{CsPbBr_3} perovskite film printed on a tailored CMOS sensor array[22, 23]. The main structure of the X-ray imager was illustrated in Figure 1a. As shown, certain electric field is applied to propel the X-ray photon stimulated electrons within the perovskite to drift towards the signal collection electrode, namely, the CMOS pixel. The perovskite film covered CMOS device with total number of 72×7272 \times72 pixels is depicted in Figure 1b. The perovskite film was printed directly on the CMOS chip via the silk-screen priting process. The cross-section SEM image shows that such in-situ grown CsPbBr3\mathrm{CsPbBr}_{3} film has good affinity with the CMOS array, in which multi-layered parasitic capacitance with multifunctional complex circuits of amplifier and leakage current compensation are integrated to enhance high SNR signal detections. The detector could endure more than four months without significantly degrading its X-ray imaging performance, indicating a excellent working stability of CsPbBr3\mathrm{CsPbBr}_{3} film under X-ray exposure and high electric field conditions. This CMOS based direct-conversion X-ray detector can resolve objects with a very fine structure of 5.5 lp/mm, see Figure 1c. It was obtained at a bias voltage of 15V, corresponding to an electric field of 0.05 V/μm0.05\ \mathrm{V}/\mu\mathrm{m} inside the perovskite film. This experiment was validated by a bar pattern plate having 33 groups of different Pb lines with varied widths and spacing distances. Since the limit spatial resolution of this CMOS array is found to be 6.0 lp/mm (corresponds to the 83.2 μm\mu\mathrm{m} pitch dimension), therefore, this X-ray detector shows an unprecedented high-resolution imaging performance. For this special CMOS array, the gate voltage of the feedback transistor, denoted as VV, controls the decay time of the Charge Sensitive Amplifier (CSA) for every single pixel (see Methods for details). To inhibit the pixel-by-pixel varied leakage current, especially to avoid the signal saturations, the gate voltage was adjusted according to the amplitude of the leakage current without turning on the X-ray beam. The 3D distribution map of the pixel response is illustrated in Figure 1d. In general, the pixel response decreases as the used gate voltage increases. To suppress the internal dark current of the perovskite film, certain gate voltages are calibrated and adopted before X-ray exposures. For larger dark current signals, higher gate voltages are needed to maintain similar pixel responses.

CMOS based perovskite X-ray detector

Figure 1. CMOS based perovskite X-ray detector. a, diagram of the X-ray imaging panel based on CMOS array. Electron-hole pairs are stimulated by the X-ray photons within the perovskite film. The electrons drift swiftly towards the CMOS pixel under the internal electric field. b, photograph of the CMOS based perovskite X-ray detector sealed inside a costumized Teflon holder. Desiccant beads are added to keep the detector away from moist air. c, the obtained X-ray image of a resolution pattern phantom. The bars with 5.5 lp/mm spatial resolution can be clearly distinguished. d, 3D response map of a single CMOS pixel versus the added gate voltage and the input dark current. The pixel response decreases as the gate voltage increases. For larger dark current signals, higher gate voltages are needed.

The inorganic CsPbBr3\mathrm{CsPbBr}_{3} perovskite thick film was fabricated through silk-screen priting process, see Figure 2a. First, a viscous CsPbBr3\mathrm{CsPbBr}_{3} precursor paste was prepared by mixing CsBr\mathrm{CsBr} and PbBr2\mathrm{PbBr}_{2} with equal molar mass in DMF/DMSO primary polar solvent through planetary ball milling. Second, the silk-screen prited CsPbBr3\mathrm{CsPbBr}_{3} precursor thick film was heated at 100 °C for 10 min to obtain an intermediate CsPbBr3\mathrm{CsPbBr}_{3} thick film, which still contains trace amount of residual DMSO according to the Fourier-transform infrared (FTIR) spectra. Afterwards, the intermediate CsPbBr3\mathrm{CsPbBr}_{3} film was hot-pressed at 150 °C and 0.5 MPa through a self-desioned equipment. The hot-pressed CsPbBr3\mathrm{CsPbBr}_{3} thich film owns flat surface and compact internal structure in contrast to conventional CsPbBr3\mathrm{CsPbBr}_{3} film with rough surface and pinholes in Figure 2b,Figure 2c. In general, the hot-presseded CsPbBr3\mathrm{CsPbBr}_{3} thick film with compact and uniform surface has better capability to inhibit leak current, ionic migration, photo-induced carrier trapping, and recombination than the porous film that is produced without hot-pressing. Moreover, the heat-treated CsPbBr3\mathrm{CsPbBr}_{3} thick film shows significantly increased XRD intensity at the main (001), (011) and (002) peaks, suggesting better crystallinity in line with the SEM images in Figure 2d. Be aware that the thickness of CsPbBr3\mathrm{CsPbBr}_{3} film can be easily adjusted from tens to hundreds micrometers if repeating the silk-screen priting procedures. The CsPbBr3\mathrm{CsPbBr}_{3} film was X-rayed themselves, and their X-ray absorption can be adjusted by controlling the CsPbBr3\mathrm{CsPbBr}_{3} film thickness. For X-ray imaging, large-area fabrication of perovskite film is important, we therefore prepared a large area CsPbBr3\mathrm{CsPbBr}_{3} film of 10×1010 \times10 cm by silk-screen printing on a ITO/glass substrate, the film exhibits compact and uniform morphology. In addition, the obtained CsPbBr3\mathrm{CsPbBr}_{3} film has excellent environmental stability with stable optical property and crystal structure, which is benefitial for enhancing the working stability of the X-ray detector.

Figure 2. Inorganic CsPbBr3 perovskite thick film. (a) Schematic illustration of the silk-screen priting CsPbBr3 film. (b) top-view and (c) cross-section SEM images of CsPbBr3 film. (d) XRD

The steady-state photoluminescence (PL) and time-resolved photoluminescence (TRPL) behaviors of the CsPbBr3\mathrm{CsPbBr}_{3} before and after the heating procedures were characterized with the purpose to elucidate their optical property (Figure 1e and Extended Data Fig. S). The treated CsPbBr3\mathrm{CsPbBr}_{3} film shows significantly higher PL intensity with PL peak red-shifted from 515 nm to 518 nm, indicating the reduced defects density and enhanced crystallinity. The measured TRPL values were fitted with exponential decay functions to measure the quenching (including charge transfer) and defects related fast decay process, i.e., short lifetime (τ1\tau_{1}), the recombination related slow decay process, i.e., long lifetime (τ2\tau_{2}), and their occupation fractions A1A_{1} and A2A_{2}. The CsPbBr3_{3} before treating has a short lifetime τ1\tau_{1} = 2.29 ns and a long lifetime τ2\tau_{2} = 11.74 ns, the CsPbBr3_{3} after treating has a short lifetime τ1\tau_{1} = 5.70 ns and a long lifetime τ2\tau_{2} = 33.07 ns, and the τave\tau_{\mathrm{ave}} improved significantly from 6.80 to 25.92 ns, indicating reduced defect density and enhanced carrier transportation capability after treatment. The μτ\mu\tau product of the treated CsPbBr3_{3} thick film is obtained by fitting the following Hecht equation from the measured voltage-dependent photocurrents:

I=I0μτVL2(1−exp⁡(−L2μτV)),I=I_{0}\frac{\mu\tau V}{L^{2}}\left(1-\exp\left(-\frac{L^{2}}{\mu\tau V}\right)\right),

where I0I_{0} is the saturated current, LL is the MAPbI3_{3} wafer thickness, and VV is the bias voltage. The μτ\mu\tau product of our CsPbBr3_{3} film is determined to be 3.40×10−5 cm2 V−13.40 \times10^{-5}\ \mathrm{cm}^{2}\ \mathrm{V}^{-1}, which is significantly higher than the amorphous selenium and thus can benefit the X-ray detection.

spectra of CsPbBr3 film. (e) TRPL spectra of CsPbBr3 film. (f) photoconductivity of the CsPbBr3 film.

The experimental current density-voltage (J-V) curves of the CsPbBr3 thick film are shown in Figure 3a. From the top to the bottom, the device contains glass-ITO/SnO2/CsPbBr3/ Au. The dark current density is as low as 1.41×10−9 A mm−21.41 \times10^{-9}\,\mathrm{A}\,\mathrm{mm}^{-2} and the corresponding light current is up to 2.10×10−7 A mm−22.10 \times10^{-7}\,\mathrm{A}\,\mathrm{mm}^{-2} at the electric field of 80 V/mm, resulting a switch ratio of 150. Time-resolved current densities of the CsPbBr3 detector at various X-ray dose rates varied from 160 μGyair s−1\mu\mathrm{Gy}_{\mathrm{air}}\,\mathrm{s}^{-1} to 1620 μGyair s−1\mu\mathrm{Gy}_{\mathrm{air}}\,\mathrm{s}^{-1} are measured, see Figure 3b. As seen, the detector has a linear response to the X-ray dose rates, and owns smaller dark current densities and larger values of X-ray on-off ratios. Different current densities under varied dose rates are shown in Figure 3c. It is estimated that the CsPbBr3 detector has a sensitivity of 1547, 4334, 7266 and 9341 μCGyair−1 cm−2\mu\mathrm{CGy}_{\mathrm{air}}^{-1}\,\mathrm{cm}^{-2} at electric fields of 20, 40, 60 and 80 V/mm, respectively.

Figure 3. Perovskite X-ray detector properties. (a) J-V characteristic and (b) X-ray response profiles of the CsPbBr3 detector. (c) X-ray current densities as a function of dose rate at different applied electric fields. (d) Dose rate dependent SNR of the CsPbBr3 detector.

The signal-to-noise ratios (SNR), defined as SNR=Isignal/Inoise=(Iphoto−Idark)/Inoise\mathrm{SNR}=I_{\mathrm{signal}}/I_{\mathrm{noise}}=(I_{\mathrm{photo}}-I_{\mathrm{dark}})/I_{\mathrm{noise}}, were calculated, see Figure 4d. Herein, IsignalI_{\mathrm{signal}} denotes the signal, InoiseI_{\mathrm{noise}} denotes the noise, IphotoI_{\mathrm{photo}} denotes the average current under X-ray irradiation, and IdarkI_{\mathrm{dark}} denotes the average dark current calculated from several parallel experiments at each bias. If assuming the minimum detective SNR=3, the lower limit-of-detection (LoD) of our CsPbBr3 detector is determined to be 588 nGyair s−1\mathrm{nGy}_{\mathrm{air}}\,\mathrm{s}^{-1}.

Figure 4. Perovskite X-ray imaging results. a, the DR imaging result of a mice. Image in the first column is obtained from a commercial indirect-conversion CMOS X-ray detector (74.8 μm pixel size). Images in the rest three columns are obtained from the new direct-conversion perovskite CMOS X-ray detector (83.2 μm pixel size) at three dose levels. The sacle bar denotes 5.0 mm. b, the CT imaging results of a chicken drumettes specimen. The upper image shows one single CT image slice, and the lower image shows the 3D volume rendering effect. The sacle bar denotes 10.0 mm. c, the line profile comparison results of the highlighted region on the mice. As seen, the developed new CMOS detector has better image contrast than the commercial COMS detector (solid blue line). d, MTF and NPS plots of the new direct-conversion perovskite CMOS detector. The former detector was measured at 50 kVp with 1.0 mm aluminum filtration, and the

The 2D digital radiography (DR) and the 3D computed tomography (CT) imaging results of biological specimen are depicted in Figure 4, correspondingly. First of all, a three-month old mice after anesthesia was scanned at 50 kVp tube voltage under three different dose levels using this perovskite CMOS detector. To compare, the same mice was also imaged with a commercial scintillator (CsI:TI) based indirect-conversion CMOS flat panel detector (Model: Dexela 2329 NDT, PerkinElmer, USA) under the same imaging condition. Overall, the DR images obtained from this novel direction-converson perovskite CMOS detector shows superior performance than the conventional indirect-conversion CMOS detector. For example, the ribs and forehand bones can be clearly recognized, see Figure 4a. Moreover, images obtained from the new detector also look much more crisp. This demonstrates that the perovskite CMOS detector has higher X-ray conversion efficiency, and thus is able to significantly reduce the imaging radiation dose level. Quantitatively, at least 70% radiation dose can be saved without dramatically degrading the image quality. Results show that a low detection limit of this CMOS detector in generating X-ray images is less than 50 nGyair_{\mathrm{air}} (measured with 50 kV X-ray spectra). In Figure 4c, line profiles of the ribs are compared. As seen, the developed new CMOS detector has better image contrast than the commercial CMOS detector. To demonstrate its dynamic imaging capability, 3D CT imaging were performed on a chicken drumstick specimen with this perovskite CMOS detector, see results in Figure 4b. In total, 180 projections were acquired in a sequence by rotating the sample continuously with two degree angular interval (see Methods for details). After that, CT images were reconstructed via the standard Feldkamp-Davis-Kress (FDK) algorithm with the Ramp filter[24, 25]. As seen, the bony structures and the tissue edges can all be clearly recovered. To quantify the spatial resolution performance, the modulation transfer function (MTF) was measured with a 1.0 mm thick tungsten plate[26], one of whose edges was manually tilted by 3.0 degrees relative to the vertical direction. The obtained MTF and normalized noise power spectra (NPS) plots are illustrated in Figure 4d. In addition, the NPS curves were also characterized[27] from the acquired 200 frames of object-free images (air scans), which have subtracted the dark background and been gain corrected via the routine signal processing procedure. Suprisingly, the NPS of this perovskite CMOS detector is almost flatten, indicating a negligible signal correlations (cross-talking) between neighbouring pixels. In other words, the movements of the stimulated electrons inside the material from X-ray photons are nicely confined by the electric field within a single pixel.

latter detector was measured with RQA3 beam quality (50 kVp with 10.0 mm aluminum filtration).

In summary, we report an innovative CMOS based direct-conversion high-resolution dynamic X-ray imaging detector fabricated from inorganic perovskite. The 300 micrometer thick CsPbBr3\mathrm{CsPbBr}_{3} film was directly printed on the CMOS array via the silk-screen priting process. Experiments demonstrate that the CsPbBr3\mathrm{CsPbBr}_{3} film has excellent optoelectric properties of a μτ\mu\tau product of 3.40×10−5 cm2 V−13.40 \times10^{-5}\ \mathrm{cm}^{2}\ \mathrm{V}^{-1}, high sensitivity of 9341 μ\muC Gyair−1_{\mathrm{air}}^{-1} cm−2^{-2}, and low detection limit of 588 nGyair_{\mathrm{air}} s−1^{-1}. Both resolution pattern and W-plate imaging results show that this CMOS detector is able to achieve high spatial resolution up to 5.5 lp/mm, which is very close to its resolution limit of 6.0 lp/mm. In addition, signals collected from this CMOS detector are less correlated, and thus leading to an almost flatten NPS curve. Benefited by the >300>300 fps fast signal readout speed, samples with size larger than 60 mm×\times60 mm can be quickly scanned. Both DR and CT images are acquired for different biological specimen to demonstrate its high performance 2D and 3D bio-imaging capability.

In conculusion, perovskites provide a completely new blueprint in developing the next generation X-ray imaging detector to reform the conventional bio-medical X-ray imaging applications. As the detector spatial resolution, readout speed and dose conversion efficiency become significantly increased, which have already been demonstrated in this study, the medical X-ray imaging would become more gentle and safe. In the near future, bio-medical X-ray imaging applications such as the small animal imaging, dental imaging and breast imaging would be greatly benefitted if assembling multiple CMOS detector components into a linear detector array or fabricating a large area flat panel from stitched CMOS arrays.

Supplementary material

PbBr2\mathrm{PbBr_2} (lead bromide, 99%, Aladdin), CsBr\mathrm{CsBr} (cesium bromide, 99.5%, Aladdin). Organic solvents including dimethyl formamide (DMSO, AR, 99%) and dimethyl sulfoxide (DMSO, AR, 99%) were purchased from Sigma-Aldrich. The CsBr\mathrm{CsBr} (6.4 g) and PbBr2\mathrm{PbBr_2} (11.0 g) with equal molar mass were mixed with 6 ml DMF/DMSO (1:1) solvent in agate jar, and then the mixture were grinded in ball-milling for three hours, obtaining a viscous CsPbBr3\mathrm{CsPbBr_3} precursor paste for silk-screen printing. The susbtrates (glass/ITO or CMOS) were cleaned with detergent, ultrasonicated in acetone and isopropyl alcohol, and subsequently dried overnight in an oven at 100 °C. The SnO2\mathrm{SnO_2} solvent was spin-coated on substrate (4000 rmp, 40 s), then the substrate was annealing at 140 °C for 1 hour, forming a electron transport layer. Then, the viscous CsPbBr3\mathrm{CsPbBr_3} precursor paste was screen-printed on the substrate, obtaining an CsPbBr3\mathrm{CsPbBr_3} precursor film, and its thickness can be adjusted by controlling the printing times. The CsPbBr3\mathrm{CsPbBr_3} precursor film was pre-heated at 100 °C for 10 min, resulting an intermediate CsPbBr3\mathrm{CsPbBr_3} thick film, which still contains trace amount of residual DMSO. Afterwards, the intermediate CsPbBr3\mathrm{CsPbBr_3} film was hot-pressed at 150 °C and 0.5 MPa for 10 min through a self-desioned equipment, the hot-pressed CsPbBr3\mathrm{CsPbBr_3} thich film was further heated at at 150 °C for 30 min, achieving a complete CsPbBr3\mathrm{CsPbBr_3} thich film. Finally, an Au anode was deposited by thermal evaporation in a vacuum of about 5×10−75 \times10^{-7} Torr.

The used CMOS chip consists of a 72×7272 \times72 active pixel array, a scan module, a switch array and a buffer. The analog signal of each pixel is transmitted off chip pixel by pixel through the switch array and the buffer. The switch array is controlled by the scan module. The charge collection electrode, denoted as TT, is an exposed top-most metal. Another top-most metal, denoted as GG, covered by an insulation material surrounds around the Topmetal to form a focused electric field hence improving the charge collection efficiency. The detailed circuit of each pixel is composed of a charge collection electrode, a Charge Sensitive Amplifier (CSA) and a two-staged cascaded source follower. The charge collected by the charge collection electrode is fed into the CSA and then converted into a voltage signal. The gate voltage labeled FB_VREF of the feedback transistor can be adjustable to adjust the decay time of the CSA. Hence the leakage current caused by the perovskite thick film can be partially absorbed by increasing the FB_VREF, but the signal is also decreased.

The rear signal readout circuit is composed of a CMOS chip binding board, a mixed signal converter board and a Field-Programmable Gate Array (FPGA) based control board. The binding board carries a bare die of the CMOS chip. Filtering capacitors are placed on the binding board to suppress noise. The binding board is connected with the mixed signal converter board through a pin header. The single-end analog signal output from the CMOS chip is fed into the mixed signal converter board and then converted to a differential signal to match the input of the analog-to-digital converter (ADS5282, TI, USA). The ADS5282 features a digital resolution of 12 bit, a sampling rate of 65 MSps and an interface of the serialized Low Voltage Differential Signaling (LVDS) outputs. A 16-bit digital-to-analog converter (DAC8568, TI, USA) on the mixed signal converter board is designed to configure the bias voltages of the CMOS chip. the mixed signal converter board is connected with the FPGA based control board through FPGA Mezzanine Card (FMC) connector. The Xilinx kintex-7 FPGA is the core logic control in a scalable and minimum hardware including an ethernet transceiver module, two Double-Data-Rate Three Synchronous Dynamic Random Access Memory (DDR3 SDRAM) module and a flash configuration module on the FPGA based control board. The FPGA based control board receives the serialized data output from the ADS5282 and transmits the data to PC through the ethernet.

The in-house built X-ray CT imaging benchtop has a medical-grade X-ray tube (G-242, Varex, UT, USA). The maximum tube voltage is 125 kVp, and the tube current is fixed at 12.5 mA during the continuous exposures. A beam collimator is installed to regulate the beam shape. The fixed beam filtration is 1.5 mm aluminum. The distance from the X-ray focal spot to the rotation center is 444.0~mm, and is 476.0~mm to the photosensitive layer of the detector. The horizontal linear stage (Model: LS12-X200, Hanjiang, China) has a travel distance of 200.0 mm, and the vertical linear stage (Model: STS06-X20, Hanjiang, China) has a travel distance of 40.0 mm. For CT imaging, the object is scanned on a rotation stage(Model: URS100BCC, Newport, USA) by 360 degree with 2.0 degree interval. The MTF and NPS experiments are measured with RQA3 beam, which is produced from a tube voltage of 50 kVp and filtered by 10.0 mm thick aluminum. Note that the CMOS detector is not placed 1500.0 mm away from the X-ray source as required by the IEC 62220-1 standard. The tungsten (W) plate has a thickness of 1.0 mm and is placed on the surface of the detector. The W plate blocks nearly half of the CMOS detector array. The CMOS detector array is positioned to align with the X-ray source focal spot when performing the NPS measurements. The acquired thousands of X-ray projection images from the CMOS were postprocessed to stitch a full DR projection of the object. In particular, the signals correspond to the same spatial positions are combined together, and the signals belong to the different spatial positions are orderly rearranged.

References

  1. [1]Reck, Martin, and Klaus F. Rabe. "Precision diagnosis and treatment for advanced non–small-cell lung cancer." New England Journal of Medicine 377.9 (2017): 849-861.
  2. [2]SCOT-Heart Investigators. "Coronary CT angiography and 5-year risk of myocardial infarction." New England Journal of Medicine 379.10 (2018): 924-933.DOI
  3. [3]Joseph, et al. "Screening for lung cancer with low-dose CT--translating science into Medicare coverage policy. " New England Journal of Medicine (2015).DOI
  4. [4]Zhao, Wei, Goran Ristic, and J. A. Rowlands. "X-ray imaging performance of structured cesium iodide scintillators." Medical physics 31.9 (2004): 2594-2605.pubmed.ncbi.nlm.nih.gov/15487742
  5. [5]Stone, Mary F., et al. “The x-ray sensitivity of amorphous selenium for mammography.” Medical Physics 29.3 (2002): 319-324.pubmed.ncbi.nlm.nih.gov/11929014
  6. [6]Huang, H. & Abbaszadeh, S. Recent Developments of Amorphous Selenium-Based X-Ray Detectors: A Review. IEEE Sensors Journal 20, 1694-1704, (2020).DOI
  7. [7]Chen, Z., Zhu, Y. & He, Z. Intrinsic photopeak efficiency measurement and simulation for pixelated CdZnTe detector. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 980, 164501, (2020).DOI
  8. [8]Zhou, F. et al. Halide Perovskite, a Potential Scintillator for X-Ray Detection. Small Methods 4, 2000506, (2020).DOI
  9. [9]Wei, H. et al. Sensitive X-ray detectors made of methylammonium lead tribromide perovskite single crystals. Nature Photonics 10, 333-339, (2016).DOI
  10. [10]He, Y., Hadar, I. & Kanatzidis, M. G. Detecting ionizing radiation using halide perovskite semiconductors processed through solution and alternative methods. Nature Photonics 16, 14-26, (2022).DOI
  11. [11]Shrestha, S. et al. High-performance direct conversion X-ray detectors based on sintered hybrid lead triiodide perovskite wafers. Nature Photonics 11, 436-440, (2017).DOI
  12. [12]Kim, Y. C. et al. Printable organometallic perovskite enables large-area, low-dose X-ray imaging. Nature 550, 87-91, (2017).DOI
  13. [13]Deumel, S. et al. High-sensitivity high-resolution X-ray imaging with soft-sintered metal halide perovskites. Nature Electronics 4, 681-688, (2021).DOI
  14. [14]Xia, M. et al. Compact and Large-Area Perovskite Films Achieved via Soft-Pressing and Multi-Functional Polymerizable Binder for Flat-Panel X-Ray Imager. 32, 2110729, (2022).DOI
  15. [15]Laborde, Cecilia, et al. "Real-time imaging of microparticles and living cells with CMOS nanocapacitor arrays." Nature nanotechnology 10.9 (2015): 791-795.DOI
  16. [16]Zentai, George. “Comparison of CMOS and a-Si flat panel imagers for X-ray imaging.” 2011 IEEE International Conference on Imaging Systems and Techniques. IEEE, 2011.DOI
  17. [17]Cao, Qian, et al. "Modeling and evaluation of a high-resolution CMOS detector for cone-beam CT of the extremities." Medical physics 45.1 (2018): 114-130.DOI
  18. [18]Sheth, Niral Milan, et al. "Technical assessment of 2D and 3D imaging performance of an IGZO-based flat-panel X-ray detector." Medical physics 49.5 (2022): 3053-3066.DOI
  19. [19]D. Kang et al., Large area X-ray CMOS digital pixel sensor based on pulse width modulation for high frame rate applications, IEEE Nuclear Science Symposium and Medical Imaging Conference, 2013, pp. 1-4.
  20. [20]M. M. Dehnavi, Y. Audet and E. Khamsehashari, Differential integrator pixel architecture for dark current compensation in CMOS image sensors, 14th IEEE International New Circuits and Systems Conference, 2016, pp. 1-4.DOI
  21. [21]Zhou, Y. et al. Heterojunction structures for reduced noise in large-area and sensitive perovskite x-ray detectors. 7, eabg6716, (2021).
  22. [22]An, Mangmang, et al. A low-noise CMOS pixel direct charge sensor, Topmetal-II. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 810 (2016): 144-150.DOI
  23. [23]C, Gao, et al. Topmetal-II: a direct charge sensor for high energy physics and imaging applications. Journal of Instrumentation, 2016, 11(01):C01053.DOI
  24. [24]Feldkamp, Lee A., Lloyd C. Davis, and James W. Kress. “Practical cone-beam algorithm.” Josa a 1.6 (1984): 612-619.DOI
  25. [25]Ge, Yongshuai, et al. "ADAPTIVE-NET: deep computed tomography reconstruction network with analytical domain transformation knowledge." Quantitative Imaging in Medicine and Surgery 10.2 (2020): 415.DOI
  26. [26]Fujita, Hiroshi, et al. “A simple method for determining the modulation transfer function in digital radiography.” IEEE Transactions on medical imaging 11.1 (1992): 34-39.pubmed.ncbi.nlm.nih.gov/18218354
  27. [27]Tapiovaara, M. J. "SNR and noise measurements for medical imaging. II. Application to fluoroscopic x-ray equipment." Physics in Medicine & Biology 38.12 (1993): 1761.DOI

Paper details

vlnc:W_y4pj6x3v
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 0

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…
Contents