betaFeedback

WorksPhysics and Astronomy

Phonon-mediated crystal detectors with metallic film coating capable of rejecting αα and ββ events induced by surface radioactivity

Abstract

Phonon-mediated particle detectors based on single crystals and operated at millikelvin temperatures are used in rare-event experiments for neutrino physics and dark-matter searches. In general, these devices are not sensitive to the particle impact point, especially if the detection is mediated by thermal phonons. In this Letter, we demonstrate that excellent discrimination between interior and surface ββ and αα events can be achieved by coating a crystal face with a thin metallic film, either continuous or in the form of a grid. The coating affects the phonon energy down-conversion cascade that follows the particle interaction, leading to a modified signal shape for close-to-film events. An efficient identification of surface events was demonstrated with detectors based on a rectangular 20×20×1020 \times 20 \times 10 mm3^3 Li2_2MoO4_4 crystal coated with a Pd normal-metal film (10~nm thick) and with Al-Pd superconductive bi-layers (100~nm-10~nm thick) on a 20×2020 \times 20 mm2^2 face. Discrimination capabilities were tested with 238^{238}U sources emitting both αα and ββ particles. Surface events are identified for energy depositions down to millimeter-scale depths from the coated surface. With this technology, a substantial improvement of the background figure can be achieved in experiments searching for neutrinoless double-beta decay.

Phonon-mediated particle detectors1 (often defined “bolometers”) have nowadays important applications in neutrino physics,[2, 3] dark-matter searches[4] and rare nuclear decay investigations.[5] They also provide outstanding α\alpha, β\beta, γ\gamma, X-ray and neutron spectroscopy.[1, 4, 5, 6]

Neutrinoless double-beta (0ν2β0\nu2\beta) decay[7] is a hypothetical rare nuclear transition of an even-even nucleus to an isobar with two more protons, with the emission of just two electrons. Its observation would provide a unique insight into neutrino physics.[8] Bolometers based on Li2_2MoO4_4 crystals are promising detectors for a next-generation 0ν2β0\nu2\beta decay experiment.[6, 9, 10] They embed the favorable candidate 100^{100}Mo, maximising the detection efficiency. The 0ν2β0\nu2\beta decay signature is a peak in the sum energy spectrum of the two emitted electrons, expected at 3.034 MeV for 100^{100}Mo. In a bolometer, the energy deposited by a particle in the crystal is converted into phonons, which are then detected by a suitable sensor.

The highest challenge in 0ν2β0\nu2\beta decay search is the control of the radioactive background, due to the long expected lifetime of the process (>1025−1026>10^{25}-10^{26} y).[11, 12, 13] The experiments are located underground under heavy shielding. To reduce the current background level of bolometric experiments, it is mandatory to reject α\alpha or β\beta events — defined “surface α\alpha’s or β\beta’s” in the following for brevity — induced by radioactive impurities located either close to the surface of the crystal itself or to that of the surrounding structure.[14, 15] Surface α\alpha’s can be rejected in scintillating materials — such as Li2_2MoO4_4 — by detecting simultaneously scintillation and phonons for the same event[14, 16, 17] and exploiting the generally lower light yield of α\alpha’s with respect to β\beta’s,[18] but the rejection of surface β\beta’s requires dedicated techniques capable of tagging surface events in bolometers.[19, 20, 21, 22, 23]

In this letter, we report an effective method to identify both surface α\alpha’s and β\beta’s in Li2_2MoO4_4 bolometers. The discrimination is achieved by coating a Li2_2MoO4_4 crystal side with a metallic film acting as a pulse-shape modifier for events that release energy close to the coated face. When an ionizing event occurs in a dielectric crystal kept at mK temperature, the deposited energy is readily converted to athermal phonons with typical energies of the order of tens of meV, to be compared with the few μ\mueV thermal-bath energy. The energy down-conversion of these athermal phonons occurs mainly by anharmonic decay and is progressively slowing down, as the phonon lifetime scales as the fifth power of the energy.[24, 25] If a sensor sensitive mainly to thermal phonons is used (as in this work), the rise time of the signal is in the ∼10\sim10 ms range, which corresponds to the typical thermalization time of the deposited energy. However, thermalization can speed up via a metallic film covering a crystal side. If the particle is absorbed close to the film, a significant fraction of its energy is trapped in the metal in the form of hot electrons, excited by the absorption of the particle-generated athermal phonons. The energy is quickly thermalised in the electron system, so that phonons of much lower energies are re-injected in the crystal from the film. Signals from events occurring close to the film will present therefore a shorter rise time and a modified time evolution. We show here that surface events can be tagged according to this approach.

All the detectors in this work share a common basic structure, which is similar to that used in the 0ν2β0\nu2\beta experiments CUORE,[13] LUMINEU,[9] CUPID-Mo,[10, 26] CUPID-0[27] and in the dark-matter experiment EDELWEISS[28] as far as the phonon readout is concerned. The surface sensitivity was studied above ground with prototypes of reduced size with respect to the final 0ν2β0\nu2\beta bolometers. The energy absorber of the bolometers described here is a single Li2_2MoO4_4 crystal[29] with a size of 20×20×1020 \times20 \times10 mm3^3 and a mass of ∼12\sim12 g. All the tests involve just a single 20×2020 \times20 mm2^2 coated side.

The phonon sensor is a neutron transmutation doped Ge thermistor[30] (NTD) with a size of 3×3×13 \times3 \times1 mm3^3. Its resistivity increases exponentially as the temperature decreases.[31] The NTD is glued on the crystal by means a two-component epoxy. The glue provides a slow transmission interface, making the NTD sensitive mainly to thermal phonons.

We used uranium radioactive sources to test the detector surface sensitivity. They were obtained by drying up a drop of uranium acid solution on a copper foil. These sources provide two main α\alpha lines at ∼4.2\sim4.2 and ∼4.7\sim4.7 MeV from 238U{}^{238}\mathrm{U} and 234U{}^{234}\mathrm{U} respectively, affected by a significant straggling due to partial α\alpha absorption in the source residues and/or in the copper substrate. 238U{}^{238}\mathrm{U} disintegration is followed by two consecutive β\beta emissions, from 234Th{}^{234}\mathrm{Th} (with a half-life of 24.1 d and an end-point of 0.27 MeV) and from 234Pa{}^{234}\mathrm{Pa} (with a half-life of 1.2 min and an end-point of 2.27 MeV). The 238U{}^{238}\mathrm{U} α\alpha rate and the 234mPa{}^{234}\mathrm{m}\mathrm{Pa} β\beta rate are extremely close.

The detector assembly is shown in Figure 1. A first test was conducted with a Li2_2MoO4_4 crystal without coating to establish the bare detector performance. All subsequent tests have adopted the configuration shown in Figure 1, where the metal-coated side, which is optically polished before the film deposition, faces the radioactive source. A bolometric light detector based on a Ge wafer[9, 10] is used to separate α\alpha from β/γ\beta/\gamma/muon events by detecting the scintillation light.[14, 16, 17]

Image not extracted

Figure 1. Scheme (left) and photograph (right) of the detector assembly with a bare Li2_2MoO4_4 crystal. A Ge thermistor and a Si:P heater (used to stabilize the bolometric response) are glued on the upper face of the crystal, which is held by polytetrafluoroethylene (PTFE) elements, not shown in the scheme. A uranium source — visible in transparency in the photograph — is placed below the crystal. A bolometric light detector — removed to take the photograph — faces the upper side of the crystal. The reflecting foil forms an optical cavity that aids light collection.

The detectors were cooled down in a dilution refrigerator located in IJCLab, Orsay, France.[32] All the data discussed here have been collected with the detector thermalized to a copper plate at ∼22\sim22 mK (Figure 1). A current of the order of ∼5\sim5 nA is injected in the NTD, rising the detector temperature to about ∼25\sim25 mK, at which the NTD resistance is about 0.5 MΩ\Omega. The voltage signal amplitude across the NTD is ∼60\sim60 μ\muV/MeV for the bare crystal, corresponding to an NTD temperature change of ∼0.5\sim0.5 mK/MeV. The pulse rise time (from 10% to 90% of maximum amplitude) is typically in the 3–10 ms range and the pulse decay time (from 90% to 30% of maximum amplitude) is tens of ms. The signals are read out by a DC-coupled low-noise voltage-sensitive amplifier.[33] In all the tests, the Li2_2MoO4_4 detector is energy-calibrated using γ\gamma peaks of the environmental background, and the light detector using the cosmic muons crossing the Ge wafer.[34] In the Li2_2MoO4_4 heat channel, we obtained routinely good energy resolution of 5–10 keV FWHM for environmental γ\gamma peaks in the 0.2–1 MeV region.

The first test to attain surface sensitivity was performed with a 10-μ\mum-thick Al-film coating. The details of the achieved results are reported elsewhere.[35, 36] We remind here that an excellent separation of surface α\alpha particles was demonstrated thanks to pulse-shape discrimination.

The best separation between surface α\alpha's and any type of interior events was obtained via a specially developed pulse-shape parameter — extensively used here — that we will designate as m/Smm/S_m.[35] To construct it, the signals are passed through a digital optimal filter,[37] whose transfer function is built using the noise power spectrum and the pulse shape of an interior event. This filter provides the best estimator of the signal amplitude SmS_m (i.e. energy). An individual pulse S(t)S(t) is plotted point by point against an average pulse A(t)A(t) — formed from a large sample of interior events and normalized to 1 — obtaining approximately a straight line. The related slope parameter mm is an estimator of the pulse amplitude as well. The ratio m/Smm/S_m turns out to be very sensitive to the pulse shape. Interior events have m/Sm∼1m/S_m \sim1, as expected. On the contrary, m/Smm/S_m deviates from 1 for surface α\alpha events.

For the Al-coated Li2_2MoO4_4 crystal, the separation between the interior and the surface α\alpha events from a uranium source is better than 10σ10\sigma in terms of m/Smm/S_m distributions.[35] Unfortunately, only a slight hint of separation of the surface β\beta events emitted by the same source was observed,[35, 38] ruling out Al coating as a viable method for a complete surface-event tagging.

Aluminum was chosen as it is superconductive at the bolometer operation temperature, with a critical temperature TC(Al)∼1.2T_\mathrm{C}(\mathrm{Al}) \sim1.2 K.[39] This leads to a negligible contribution to the heat capacity of the full bolometer, as the electron specific heat of superconductors vanishes exponentially with the temperature. We remark that the heat capacity of a bolometer must be as low as possible to achieve high signal amplitudes. In fact, no deterioration of the detector sensitivity was observed with respect to the bare Li2_2MoO4_4 crystal. However, the behaviour of superconductors can spoil surface particle tagging. The prompt absorption of athermal phonons by the film breaks Cooper pairs and forms quasi-particles. Theoretically, quasi-particle lifetime diverges as the temperature of the superconductor decreases,[40, 41] although it is often experimentally found to saturate at low temperatures.[42, 43] In aluminum, at very low temperatures such as ours (T/TC<0.02T/T_\mathrm{C} < 0.02), we expect the quasi-particle lifetime to be as large as several ms,[42, 43, 44, 45] similar to the thermalization time of interior events. This mechanism competes with the faster thermalization that should be provided by the film.

Driven by these considerations, we tested a Li2_2MoO4_4 bolometer with a normal-metal coating. At low temperatures, the electron specific heat of normal metals is proportional to the temperature and tends to dominate over the crystal heat capacity, which scales as T3T^3 according to the Debye law. The thickness of normal-metal films must be substantially smaller than the aluminum ones. We chose palladium as a coating material as it can provide continuous thin films down to 2 nm thickness and no challenging radioactive isotopes are present in its natural composition. A thickness of 10 nm was chosen as a good compromise between heat capacity reduction and phonon absorption probability. The particle-identification results are encouraging, as shown in Figure 2: both surface α\alpha’s and β\beta’s are well separated from the interior events.

Image not extracted

Figure 2. Particle identification obtained by a Li2_2MoO4_4 detector with a 10-nm-thick Pd coating (see inset on the left) exposed to a uranium source. Left: The pulse-shape parameter m/Smm/S_m is plotted as a function of the heat energy (estimated by SmS_m) deposited in the Li2_2MoO4_4 crystal, with a γ\gamma-based calibration. Surface events appear as a population with lower m/Smm/S_m values. They are selected by visual inspection and highlighted in red. The neutron-capture line from the reaction 6Li(n,t)α{}^{6}\mathrm{Li}(n,t)\alpha lays in the interior-event band. The α\alpha particles are mis-calibrated by about 20 % due to both pulse-shape effects and intrinsic different responses for α\alpha’s and β/γ\beta/\gamma’s. Right: The Li2_2MoO4_4 scintillation light yield (LY) is plotted for the same pulses on the same energy scale. The LY is expressed by the ratio of the energy deposited in the light detector by scintillation photons (in keV) to that deposited in the Li2_2MoO4_4 crystal as heat (in MeV), for the same event. The same surface events highlighted in the left panel are shown in red. Surface β\beta’s lay in the high-LY band, while α\alpha’s and the neutron capture events are well separated in the low-LY band.

Unfortunately, the heat capacity of the Pd film[46] competes with that of the Li2_2MoO4_4 crystal[47] affecting seriously the sensitivity of the detector, which was only ∼23\sim23 μ\muV/MeV, about one third of that achieved with the bare crystal. Therefore, this option is not viable for a full coating of the crystal.

To overcome the heat-capacity problem, we developed a detector coated with an Al–Pd bi-layer (100 nm and 10 nm thick respectively, with Al on the top), which is superconducting by proximity effect below TC(AlT_\mathrm{C}(\mathrm{Al}-Pd) = 0.65 K. The superconductive gap induced in Pd by the Al film reduces substantially the Pd specific heat with respect to the normal state. This gap is however low enough to ensure the fast thermalization of the energy deposited by surface events. In fact, the surface-event discrimination capability was fully maintained (see Figure 3, left). The detector sensitivity was measured to be 43 μ\muV/MeV, almost doubled with respect to the pure Pd film.

Image not extracted

Figure 3. Particle identification obtained by a Li2_2MoO4_4 detector with an Al-Pd coating exposed to a uranium source. The α\alpha events are removed by a light-yield cut. Left: in a plot of the pulse-shape parameter m/Smm/S_m versus energy, the surface events (in red) lay below a black curve defining a 3σ3\sigma acceptance region for the interior events (in blue). The analysis is carried out in the [0–3000] keV energy interval, which is divided in several sub-intervals. For each of them, a double Gaussian fit of the m/Smm/S_m distribution is performed to separate the two populations. An example is provided in the inset. The black point is located 3σ3\sigma at the left of the mean of the Gaussian of the interior events. The black curve fits the black points by a power-law function. Right: Energy spectra (with and without source) of the surface events selected according to the procedure illustrated on the left. The live times of the two measurements are normalized. The fit of the source data accounts for the two simulated β\beta contributions of the uranium source and that of the background.

We performed two runs with the bi-layer detector. In the first, a uranium source was present, while the second was a background measurement in the same configuration. The trigger rate was ∼0.2\sim0.2 Hz with the source. The contribution of the source is at the level of ∼0.03\sim0.03 Hz. First, we developed a method to separate the surface β\beta component. The events below the black curve in the left panel of Figure 3 — collected in a source run — are selected as surface events, while those above represent more than 99% of the interior event population. The same analysis was performed for the background run. By means of Geant4-based[48] Monte-Carlo simulations (using the G4RadioactiveDecay and Decay0[49] event generators), we were then able to confirm that the surface β\beta events isolated at low energies come actually from the radioactive source. We built a model to predict the β\beta spectrum shape considering the observed α\alpha straggling (Figure 4) and the β\beta interactions in the detector. We fitted then the experimental β\beta spectrum using the predicted shape and taking the background into account (right panel in Figure 3). The total number of 234mPa{}^{234\mathrm{m}}\mathrm{Pa} decay events returned by the fit — the only free parameter — is 3526(81). To build the source model, we set a uranium-source depth profile capable of reproducing the observed α\alpha spectrum and the related straggling, as shown in Figure 4. From the model and the experimental number of α\alpha counts it was possible to predict independently the expected total number of 234mPa{}^{234\mathrm{m}}\mathrm{Pa} events, which resulted to be 3455(273), in excellent agreement with that deduced from the selection of the source β\beta events. The efficiency in selecting surface β\beta events can be estimated as 102(8)%.

Image not extracted

Figure 4. Energy spectrum collected by a Li2_2MoO4_4 detector with an Al-Pd coating exposed to a uranium source after selection of the α\alpha events by a light-yield cut with ∼100%\sim100\% efficiency. The spectrum is calibrated using the α\alpha-line positions. The measurement is the same that provided the source data shown in Figure 3. The straggling can be reproduced by assuming five source components in copper. Each component is a 6 mm diameter disk with a given thickness. The active nuclei are assumed to be uniformly distributed in each disk. The exact source structure is unknown, but our goal here is to set up a phenomenological model capable of explaining the observed straggling. 238U{}^{238}\mathrm{U} and 234U{}^{234}\mathrm{U} are not in secular equilibrium, as already observed in these types of liquid sources.

The β\beta-particle range in Li2_2MoO4_4 is of the order of 2 mm at 1 MeV and 4 mm at 2 MeV. Therefore, we can separate events that deposit a significant amount of energy up to ∼4\sim4 mm from the film, well beyond its thickness. We performed then a last test by replacing the continuous Al-Pd film with an Al-Pd grid. The width of the grid lines was 70 μ\mum and the spacing between each line was 700 μ\mum (see inset in Figure 5). The purpose of using a grid is manifold: (1) further reduction of the heat capacity of the coating; (2) possibility to extract scintillation light through the coating; (3) availability of geometrical parameters to possibly tune the discrimination depth. The grid was tested with another uranium source, prepared with the same method as the first one, but about twice less intense. The detector with grid coating can separate surface β\beta events (see Figure 5). The β\beta selection efficiency resulted to be 93(10)%, in good agreement with the continuous-film results. In addition, we measured a discrimination power of about 4.5σ4.5\sigma for surface α\alpha events using the m/Smm/S_m parameter. In terms of detector performance, we observed an almost full recovering of the detector sensitivity, that was ∼51 μV/MeV\sim51\ \mu\mathrm{V/MeV} for β/γ\beta/\gamma events. Therefore, the grid method is currently our protocol for surface event discrimination.

Image not extracted

Figure 5. Particle identification obtained by a Li2_2MoO4_4 detector with an Al-Pd grid coating exposed to a uranium source. The event selection is performed as in Figure 3, left. In the top inset, the grid-coated crystal is shown. In the bottom inset, pulses from a surface (red) and an interior (blue) event are shown, corresponding to a deposited energy of about 1 MeV.

In conclusion, we have shown that both α\alpha and β\beta particles absorbed close to a metal-coated surface of a Li2_2MoO4_4 bolometer can be rejected with high efficiency by pulse-shape discrimination. The prospects of this approach for 0νββ0\nu\beta\beta searches are promising. In fact, the current background model of the future 0νββ0\nu\beta\beta experiment CUPID-50[50] predicts a background level of 0.1 counts/(tonne y keV). Next-to-next generation experiments aim to a reduction of an additional factor 10. Since surface β\beta events contribute significantly to the current background level, a necessary condition to achieve the desired reduction is to reject them with an efficiency up to 90%. This is achievable with the technique here described.

This work is supported by the European Commission (Project CROSS, Grant ERC-2016-ADG, ID 742345). The ITEP group was supported by the Russian Scientific Foundation (grant No. 18-12-00003). F.A.D., V.I.T. and M.M.T. were supported in part by the National Research Foundation of Ukraine (grant No. 2020.02/0011). The PhD fellowship of H.K. has been partially funded by the P2IO LabEx (ANR-10-LABX-0038) managed by the Agence Nationale de la Recherche (France). The dilution refrigerator used for the tests and installed at IJCLab (Orsay, France) was donated by the Dipartimento di Scienza e Alta Tecnologia of the Insubria University (Como, Italy).

The data that support the findings of this study are available from the corresponding author upon reasonable request.

References

  1. [1]1C. Enss and D. McCammon, J. Low Temp. Phys. 151, 5–24 (2008).
  2. [2]2A. Giuliani, J. Low Temp. Phys. 167, 991–1003 (2012).
  3. [3]3A. Nucciotti, J. Low Temp. Phys. 176, 848–859 (2014).
  4. [4]4S. Pirro and P. Mauskopf, Annu. Rev. Nucl. Part. Sci. 67, 161–181 (2017).
  5. [5]5P. Bellini, R. Bernabei, F. Danevich, et al., Eur. Phys. J. A 55, 140 (2019).
  6. [6]6T. B. Bekker, N. Coron, F. A. Danevich, et al., Astropart. Phys. 72, 38–45 (2016).
  7. [7]7M. J. Dolinski, A. W. P. Poon, and W. Rodejohann, Annu. Rev. Nucl. Part. Sci. 69, 219–251 (2019).
  8. [8]8J. D. Vergados, H. Ejiri, and F. Šimkovic, Int. J. Mod. Phys. E 25, 1630007 (2016).
  9. [9]9E. Armengaud, C. Augier, A. S. Barabash, et al., Eur. Phys. J. C 77, 785 (2017).
  10. [10]10E. Armengaud, C. Augier, A. S. Barabash, et al., Eur. Phys. J. C 80, 44 (2020).
  11. [11]11A. Gando, Y. Gando, T. Hachiya, et al., Phys. Rev. Lett. 117, 082503 (2016).
  12. [12]12M. Agostini, G. R. Araujo, A. M. Bakalyarov, et al., Phys. Rev. Lett. 125, 252502 (2020).
  13. [13]13D. Q. Adams, C. Alduino, K. Alfonso, et al., Phys. Rev. Lett. 124, 122501 (2020).
  14. [14]14D. R. Artusa, F. T. Avignone III, O. Azzolini, et al., Eur. Phys. J. C 74, 3096 (2014).
  15. [15]15C. Alduino, K. Alfonso, D. R. Artusa, et al., Eur. Phys. J. C 77, 543 (2017).
  16. [16]16S. Pirro, J. W. Beeman, S. Capelli, et al., Phys. At. Nucl. 69, 2109–2116 (2006).
  17. [17]17D. Poda and A. Giuliani, Int. J. Mod. Phys. A 32, 1743012 (2017).
  18. [18]18V. I. Tretyak, Astropart. Phys. 33, 40–53 (2010).
  19. [19]19L. Foggetta, A. Giuliani, C. Nones, et al., Appl. Phys. Lett. 86, 134106 (2005).
  20. [20]20S. Marnieros, L. Berge, A. Broniatowski, et al., J. Low Temp. Phys. 151, 835–840 (2008).
  21. [21]21C. Nones, Nuovo Cimento B 125, 417–437 (2010).
  22. [22]22C. Nones, L. Bergé, L. Dumoulin, et al., J. Low Temp. Phys. 167, 1029–1034 (2012).
  23. [23]23R. Agnese, A. J. Anderson, D. Balakishiyeva, et al., Appl. Phys. Lett. 103, 164105 (2013).
  24. [24]24R. Orbach and L. A. Vredevoe, Physics Physique Fizika 1, 91–94 (1964).
  25. [25]25W. E. Bron, Y. B. Levinson, and J. M. O’Connor, Phys. Rev. Lett. 49, 209–211 (1982).
  26. [26]26E. Armengaud, C. Augier, A. S. Barabash, et al., “A new limit for neutrinoless double-beta decay of ¹⁰⁰Mo from the CUPID-Mo experiment,” (2020), arXiv:2011.13243.DOI
  27. [27]27O. Azzolini, J. W. Beeman, F. Bellini, et al., Phys. Rev. Lett. 123, 032501 (2019).
  28. [28]28E. Armengaud, Q. Arnaud, C. Augier, et al., JINST 12, P08010 (2017).
  29. [29]29V. D. Grigorieva, V. N. Shlegel, T. B. Bekker, et al., J. Mat. Sci. Eng. B 7, 63 (2017).
  30. [30]30E. E. Haller, N. P. Palaio, M. Rodder, et al., “NTD Germanium: A Novel Material for Low Temperature Bolometers,” in Neutron Transmutation Doping of Semiconductor Materials, edited by R. D. Larrabee (Springer US, Boston, MA, 1984) pp. 21–36.DOI
  31. [31]31A. L. Efrós and B. I. Shklovskii, Phys. C 8, L49–L51 (1975).
  32. [32]32M. Mancuso, D. M. Chernyak, F. A. Danevich, et al., J. Low Temp. Phys. 176, 571–577 (2014).
  33. [33]33C. Arnaboldi, C. Bucci, J. W. Campbell, et al., IEEE Trans. Nucl. Sci. 49, 2440–2447 (2002).
  34. [34]34V. Novati, L. Bergé, L. Dumoulin, et al., Nucl. Instr. Meth. Phys. Res. A 940, 320–327 (2019).
  35. [35]35I. C. Bandac, A. S. Barabash, L. Bergé, et al., J. High Energy Phys. 2020, 18 (2020).
  36. [36]H. Khalife, L. Bergé, M. Chapellier, et al., J. Low Temp. Phys. 199, 19–26 (2020).
  37. [37]E. Gatti and P. F. Manfredi, Riv. Nuovo Cimento 9, 1–146 (1986).
  38. [38]H. Khalife, CROSS and CUPID-Mo: future strategies and new results on bolometric search for 0νββ, Ph.D. thesis, Université Paris-Saclay (2021), http://www.theses.fr/s191647 (see Figs. 4.14 and 4.15).DOI
  39. [39]J. F. Cochran and D. E. Mapother, Phys. Rev. 111, 132–142 (1958).
  40. [40]S. B. Kaplan, C. C. Chi, D. N. Langenberg, et al., Phys. Rev. B 14, 4854–4873 (1976).
  41. [41]R. Barends, J. J. A. Baselmans, S. J. C. Yates, et al., Phys. Rev. Lett. 100, 257002 (2008).
  42. [42]P. J. de Visser, D. J. Goldie, P. Diener, et al., Phys. Rev. Lett. 112, 047004 (2014).
  43. [43]A. Eyhrie, P. Day, J. Glenn, et al., J. Low Temp. Phys. 199, 688–695 (2020).
  44. [44]J. Schnagl, G. Anglhover, F. v. Feilitzsch, et al., Nucl. Instr. Meth. Phys. Res. A 444, 245–248 (2000).
  45. [45]J. A. Baselmans and S. J. C. Yates, AIP Conf. Proc. 1185, 160–163 (2009).
  46. [46]U. Mizutani, T. Massalski, and J. Bevk, J. Phys. F: Met. Phys. 6, 1–9 (1976).
  47. [47]A. Musikhin, V. Naumov, M. Bespyatov, et al., J. Alloys Compd. 639, 145–148 (2015).
  48. [48]S. Agostinelli, J. Allison, K. Amako, et al., Nucl. Instr. Meth. Phys. Res. A 506, 250–303 (2003).
  49. [49]A. A. Ponk ratenko, V. I. Tretyak, and Y. Zdesenko, Phys. At. Nucl. 63, 1282–1287 (2000).
  50. [50]The CUPID Group of Interest, “CUPID pre-CDR,” (2019), arXiv:1907.09376.

Paper details

vlnc:W_yqchqd4t
Version
v1 · latest