betaFeedback

WorksPhysics and Astronomy

A method to measure the quenching factor for recoil energy of oxygen in bismuth germanium oxide scintillators

Abstract

Bismuth germanium oxide (Bi4Ge3O12\rm Bi_{4} Ge_{3} O_{12}, BGO) scintillation crystals are widely used as detectors in the fields of particle physics and astrophysics due to their high density, and thus higher efficiency for gamma-ray detection. Owing to their good chemical stability, they can be used in any environment. For rare-event searches, such as dark matter and coherent elastic neutrino-nucleus scattering, BGO crystals are essential to comprehend the response of nuclear recoil. In this study, we have analyzed the events of neutron elastic scattering with oxygen in BGO crystals. Then, we have measured the quenching factor for oxygen recoil energy in the BGO crystal as a function of recoil energy by using a monoenergetic neutron source.

Introduction

Bismuth Germanium Oxide (Bi4Ge3O12\mathrm{Bi}_{4}\mathrm{Ge}_{3}\mathrm{O}_{12}, BGO) crystal is a scintillation material that is widely used as a detector in particle physics and astrophysics. Some of the advantages are rich light yield and large atomic number to enhance detection efficiency of high energy photons. It is also suitable for calorimetry [1, 2]. Recently, rare-event searches, such as dark matter [3, 4] and coherent elastic neutrino-nucleus scattering [5] have been conducted using BGO crystal. In the events of nuclear recoil with lower energy, it should be considered with light emission quench of the recoil energy.

The yield of scintillation light for deposited energy of recoil nuclei in the inorganic crystal is generally smaller than that for deposited energy of recoil electrons. A ratio of the visible energy EvisE_{\mathrm{vis}} to the energy of recoil nuclei ERE_{\mathrm{R}} is called quenching factor (QFQF). It is defined as:

QF=Evis/ER,(1)QF = E_{\mathrm{vis}}/E_{\mathrm{R}}, \tag*{(1)}

where QFQF depends on the material and energy of recoil nuclei. In the scintillating bolometric techniques at 20 mK, the QFQF was determined to be 12–15% for several dozen keV in BGO crystals, which is the ratio of the relation between light yield and heat signals for gamma-ray and nuclear recoil events [4, 6, 7]. In this study, we have reported a method to measure the QFQF for oxygen recoil in BGO crystal using monoenergetic neutron irradiation that is capable of expanding the recoil energy up to MeV scale of recoil energy.

The rest of the article is organized as follows. In § 2, we have briefly described the experimental setup and detector calibration. In § 3, we have explained the details of simulation and compared the results with that of the measured spectrum. Analysis of the proposed method and summary of results for QFQF are shown in § 4 and 5, respectively. Finally, we have presented the conclusion.

Experiment

Setup

To measure the QFQF of BGO, we have performed an experiment with accelerator-based neutron facility at the National Institute of Advanced Industrial Science and Technology (AIST), Japan [8]. Figure 1 shows the graphical and schematic view of our experimental setup. The monoenergetic neutrons are produced by deuterium interacting with a titanium-tritium target (for 14.8 MeV) and titanium-deuterium target (for 3 MeV) through the process of T(d,nd,n)4He^{4}\mathrm{He} and d(d,n)3Hed(d,n)^{3}\mathrm{He}, respectively. The neutron fluence is a few hundred cm−2\mathrm{cm}^{-2}. To monitor the 14.8 MeV neutron fluence, the associated alpha particles produced by T(d,nd,n)4He^{4}\mathrm{He} toward 90.7° and 131.9° are collimated with tantalum and detected by silicon surface barrier detectors (SSBD). To monitor the 3.0 MeV neutron fluence, a proton recoil is detected by a 0.5 mm thick polyethylene disk mounted SSBD. A BGO detector is placed at the center of the deuterium beamline at a distance of 1 m from the Ti-T (Ti-D) target and liquid scintillation (LS) detector is placed at a distance of 1 m from the BGO detector by changing the scattering angle θ\theta defined in Figure 1. The BGO detector is set to measure the quenching factor for nuclei recoiled by neutrons and LS detector is set to detect the neutrons scattered from the BGO detector. The energy of recoil nucleus ERE_{\mathrm{R}} is correlated with the neutron scattering angles by a function given by:

Graphical and schematic views of the setup for neutron irradiation

Figure 1. Graphical view (top) and schematic view (bottom) of setup for neutron irradiation at AIST, Japan. The neutron scattering angle θ\theta is varied up to 90°.

ER=En{1−(mncos⁡θ+mN2−mn2sin⁡2θmn+mN)2}(2)E_{\mathrm{R}} = E_{\mathrm{n}} \left\{ 1 - \left( \frac{m_{\mathrm{n}}\cos\theta+ \sqrt{m_{\mathrm{N}}^{2} - m_{\mathrm{n}}^{2}\sin^{2}\theta}}{m_{\mathrm{n}} + m_{\mathrm{N}}} \right)^{2} \right\} \tag*{(2)}

where EnE_{\mathrm{n}} is the energy of the incident neutron, and mnm_{\mathrm{n}} and mNm_{\mathrm{N}} are neutron and nuclei mass, respectively. We have measured the QFQF for various energies of recoil nuclei. The energies are selected by changing the position of the LS detector at different angles with respect to the BGO detector.

Detector configuration

We have used a BGO scintillation detector composed of a BGO crystal (density of 7.13 g/cm3\mathrm{g/cm}^{3}) with depth and radius of 2 inches each. It is connected to phototubes (R6231-100, Hamamatsu) on both sides and photo-shielded by aluminized Mylar and black sheet. Typically, a light yield of 1,000 photons/MeV is achieved for the energy deposition of charged particles.

To identify the neutron scattering events, we have used a liquid scintillation detector. It consists of a vessel with depth and radius of 3 inches each filled with a liquid scintillator (BC501A, Bicron) and phototube (9822B, ET Enterprises) connected to the window of the vessel.

Front-end electronics and data acquisition

The front-end electronics is shown in Figure 2. The waveforms of analog signals are recorded by the flash ADC when both BGO and LS detectors detect signals. The coincidence signals from both phototubes on the BGO detector are required to suppress dark noise and lower the threshold to the level below 1 p.e. for each phototube. Therefore, these thresholds are set to a voltage of 0.5 p.e. equivalent. The width of coincidence signal is set to 200 ns. A trigger signal is generated by the coincidence signals from the BGO and LS detectors. The trigger is fed to the flash ADC (DRS4, PSI) and the waveform of analog signals from these detectors are stored with a precision of 14 bits ADC for a particular pulse height and 0.7 GSPS1 sampling rate for a full range of 1463 ns.

Schematic diagram of front-end electronics and data acquisition

Figure 2. Schematic diagram of front-end electronics and data acquisition.

Calibration and performance check

Typical waveforms of the signals are displayed in Figure 3. The waveform of PMT 1 and 2 of the BGO detector are shown by shifting +400 mV+400\ \mathrm{mV} and +200 mV+200\ \mathrm{mV} in this figure, respectively. The expected energy of the recoil nuclei signals is as low as a few dozen to several hundred keV. With such a low energy, the waveform looks like a jagged line made of several single photoelectron signals following the time distribution of an exponential function with a decay constant of 300 ns. The energy deposit in the BGO scintillator is evaluated with the cumulative charge of two PMTs that is an integration of the waveforms from 143 to 1463 ns. The waveform of the LS detector is not shifted in Figure 3. As most of the signal is distributed between 0 and 571 ns and the peak width is narrow (∼30\sim30 ns), the charge is calculated with an integration window of 571 ns. However, the waveform overlapping with the other neutron signals was not observed in this experiment. The arrival time difference dtdt is calculated as dt=tLS−tBGOdt = t_{LS} - t_{BGO} where tLSt_{LS} is the rise time of the waveform from the LS detector and tBGOt_{BGO} is an average of the rise times for two phototubes of the BGO detector. The rise time is defined as a time when an absolute voltage is over one-third of the pulse height from the baseline.

Typical waveforms of the phototubes from the BGO and liquid scintillation detectors

Figure 3. Typical waveforms of the phototubes from the BGO and liquid scintillation detectors.

Energy calibration for the BGO detector was performed using various gamma-ray sources for the order of ten to hundreds keV, including 57Co{}^{57}\mathrm{Co} (47 and 122 keV), 133Ba{}^{133}\mathrm{Ba} (31, 81, and 356 keV), 137Cs{}^{137}\mathrm{Cs} (32 and 662 keV), and 241Am{}^{241}\mathrm{Am} (16.1 and 59.5 keV). We have determined the calibration coefficient as (1.788±0.001)×102 keVee/nC2(1.788 \pm0.001) \times10^{2}\ \mathrm{keV}_{\mathrm{ee}}/\mathrm{nC}^{2}2 by a linear fit to the relation between the sum of the integrated charges from two phototubes on both ends of the BGO crystal and gamma-ray energy as shown in Figure 4 (a). The energy resolution for deposited energy EvisE_{\mathrm{vis}} is evaluated as shown in Figure 4 (b), here, it is defined as the width of one sigma of the Gaussian fit on each photo-electron peak. The resolution curve is consistent with an inversely proportional function Evis\sqrt{E_{\mathrm{vis}}}. These indicate the energy threshold of ∼7 keVee\sim7\ \mathrm{keV}_{\mathrm{ee}} and lower limit of the energy peak at 16 keVee16\ \mathrm{keV}_{\mathrm{ee}} with a resolution of 38%.

Relation between the integrated charge and energy for BGO detector, and evaluated energy resolution

Figure 4. Relation between the integrated charge of the signal waveform and energy for BGO detector (a). The evaluated energy resolution as a function of visible energy (b). Open circles are data and red lines are a fit function. Temperature has been stable at 18 °C during the measurement.

The timing resolution of our detector is evaluated with 22Na{}^{22}\mathrm{Na} source by applying its positron annihilation gamma rays. The result is shown in Figure 5. It was observed that the timing resolution worsens at lower keVee\mathrm{keV}_{\mathrm{ee}} as expected, and the resolution is estimated by the exponential fit in the energy region less than hundred keVee\mathrm{keV}_{\mathrm{ee}}.

Timing resolution as a function of visible energy

Figure 5. Timing resolution as a function of visible energy. Black open circles are the data. Red line is a fit result by a function composed of exponential and constant terms.

Energy and intensity of irradiating neutrons

Energy of the irradiating neutrons are monitored by the 3He{}^{3}\mathrm{He} detector in a polyethylene Bonner sphere placed at the same position as the BGO detector at a distance of 1 m away from the beam target. Figure 6 shows the monitored energy spectra of neutrons generated from T(d,n)4He\mathrm{T}(d,n){}^{4}\mathrm{He} and d(d,n)3Hed(d,n){}^{3}\mathrm{He} reactions. The measured spectrum has been adopted in the simulation as the initial neutron energy distributions.

Neutron energy spectra for $d(d,n){}^{3}\mathrm{He}$ and $T(d,n){}^{4}\mathrm{He}$ reactions

Figure 6. Neutron energy spectra for d(d,n)3Hed(d,n){}^{3}\mathrm{He} (left) and T(d,n)4HeT(d,n){}^{4}\mathrm{He} (right) reactions.

The intensity of irradiating neutrons is relevant to a ratio of the number of signal events to the accidental coincident background. To evaluate the impact, we have measured two different intensities ∼670\sim670 and ∼200 cm−2 s−1\sim200\ \mathrm{cm}^{-2}\ \mathrm{s}^{-1} for the neutron energy of 14.8 MeV. For 3 MeV neutron irradiation, we have considered the data with the neutron intensity of ∼100 cm−2 s−1\sim100\ \mathrm{cm}^{-2}\ \mathrm{s}^{-1}.

Monte Carlo Simulation

In this study, we have performed Monte Carlo simulation using GEANT4 toolkit [9] (Geant4.9.6p04 + G4NDL4.2) and compared the results with that of the experiment. Geometric configuration of the experimental hall was reproduced in the simulation with a beam pipe, detector stage, wall, mesh floor, and ceiling in AIST. The detector, composed of phototubes, BGO crystal, and liquid scintillator, was irradiated by the neutrons. The detector response was simulated based on the measured energy resolution described in § 2.4. The neutron dataset of ENDF-VIII.0 model was adopted for the simulation. Model dependence of the cross sections for elastic and inelastic scattering of oxygen, germanium, and bismuth is negligible while determining QFQF in this study. The neutron energy distribution shown in Figure 6 was used as the input to the simulation.

In this study, the energy of recoil nuclei is determined by the scattering angle of neutrons. Therefore, bias of the scattering angle θ\theta due to the secondary scattering in other materials surrounding the detectors causes a shift of the recoil energy, thus resulting in the variation of the measured QFQF. In addition, width of the initial neutron energy causes variation in the energy of recoil nuclei ERE_{\mathrm{R}} in the detector. These effects were evaluated and accounted for as systematic uncertainty on the dependence of ERE_{\mathrm{R}} in the QFQF measurement. Details of the systematic uncertainty will be explained in § 4.5.

Analysis

This section presents the procedural analysis on various factors, such as systematic uncertainty, QFQF, energy distribution, and difference between neutron and gamma rays. Elastic scattering events have been selected based on the criteria of neutron identification in the LS detector and relative timing cuts for the BGO and LS detectors. To evaluate net signal events, contamination of the background events was evaluated from the off-time coincidence sample and compared with the events in the signal time window after normalization. Then, the QFQF was determined by a ratio of the measured peak energy to the estimation of recoil energy.

Discrimination of neutron and gamma ray

Neutrons and gamma rays are discriminated by the signal waveforms of the LS detector and are known as pulse shape discrimination (PSD). Figure 7 shows the relation between totaltotal and slow/totalslow/total for neutron energies of 3 and 14.8 MeV where totaltotal and slowslow are the integrated charges of the waveform in the nominal time window and shorter time window, respectively. The latter starts 21 ns later. The neutron-like and gamma-like events are distributed in isolation. The boundaries are represented as red lines as shown in Figure 7 to discriminate between the two events.

Relation between total and slow/total for neutron energies of 3.0 and 14.8 MeV

Figure 7. Relation between totaltotal and slow/totalslow/total for neutron energies of 3.0 and 14.8 MeV. Color at the pixel indicates the count number of events. Red lines are the boundaries to discriminate neutron and gamma-ray events.

Neutron time of flight

Figure 8 shows the distributions of the arrival time difference (dtdt) when the LS detector is set at θ=90∘\theta= 90^\circ and the detectors are irradiated to neutrons with En=3E_n = 3 MeV. The peak value is obtained at approximately 0 ns due to gamma rays induced by inelastic scattering of neutrons in the BGO crystal. After the PSD cut, the remaining peak is identified at approximately 40 ns. Since the scattered neutrons have ∼\sim3.0 MeV of kinetic energy and distance between BGO and LS is 1 m, ∼\sim50 ns of additional time than gamma rays is required to reach the LS detector after being scattered in the BGO detector. However, a flat component of dtdt indicates the existence of accidental coincidence of neutrons in the BGO and LS detectors.

Figure 8. Distributions of arrival time difference (dtdt) for En=3E_n = 3 MeV and θ=90∘\theta= 90^\circ. Black open and red filled histograms distributions before and after the PSD cut are applied to select neutron-like events, respectively. On-time window is set from +5 to +45 ns and the off-time window is set from −200 to 0 ns and from +50 to +100 ns.

Energy distribution in BGO

Significant improvement in performance was recorded in the dtdt distributions of the neutron-like events in various configurations. The signal window was set from +5 to +45 ns (on-time) to include the elastic scattering events. The background events due to the accidental coincidence were evaluated with two time-windows set at both sides of on-time window; that is, one from −200 to 0 ns and the other from +50 to +100 ns (off-time). These time windows are set for En=3.0E_n = 3.0 MeV while in the case of En=14.8E_n = 14.8 MeV, the on-time window was set from −40 to +30 ns, and the off-time window was set at both sides as −200 to −40 ns and +30 to +100 ns, respectively. The number of events in the off-time sample was normalized by the ratio of the ranges of on-time to off-time windows for a comparison with the on-time sample. Figure 9 shows the energy distributions from on-time and off-time samples observed in the BGO detector. We found a significantly excess value with a peak at approximately Evis=40E_{\mathrm{vis}} = 40 keVee_{\mathrm{ee}}, which is due to oxygen recoil. For En=3.0E_n = 3.0 MeV and θ=90∘\theta= 90^\circ, the recoil energy of oxygen nuclei was calculated as ER=334±17E_R = 334 \pm17 keV where the uncertainty includes the width of the neutron energy and range of the scattering angle due to the finite size of the BGO crystal and LS container.

Energy distributions from on-time and off-time samples

Figure 9. Energy distributions from on-time (black dots) and off-time (blue shade histogram) sample for En=3.0E_n = 3.0 MeV and θ=90∘\theta= 90^\circ (top panel). Bottom panel is the residual with a Gaussian fit represented by the red dashed line.

Quenching factor determination

QFQF is defined as QF=Evis/ERQF = E_{\mathrm{vis}}/E_R. The visible energy for the recoil oxygen is given as Evis=38.5±1.3E_{\mathrm{vis}} = 38.5 \pm1.3 keVee_{\mathrm{ee}} from the Gaussian fit to the peak. Then, QFQF was evaluated to be (11.5±0.4)×10−2(11.5 \pm0.4) \times10^{-2} from the value of ER=334E_R = 334 keV for a configuration of En=3.0E_n = 3.0 MeV and θ=90∘\theta= 90^\circ.

Further, we have performed another approach to determine the QFQF more precisely. The measured energy spectra in the on-time sample were fitted by a combination of the simulated energy distribution for oxygen recoil and background distribution in the off-time sample with three free parameters: 1) normalization factor for the simulated energy distribution for oxygen recoil; 2) QFQF value to change the energy scale of simulation; and 3) normalization factor for the off-time distribution. The minimum χ2\chi^2 value, χ2/dof=25.4/37\chi^2/dof = 25.4/37, was obtained with QF=(11.49−0.41+0.59)×10−2QF = (11.49^{+0.59}_{-0.41}) \times10^{-2} in the fit for ER=334E_R = 334 keV. The value of minimum χ2\chi^2 is consistent with the best fit value of QFQF.

Figure 10 shows the fit results for five configurations that generate different energies for recoil oxygen: a) En=3.0E_n = 3.0 MeV, θ=58∘\theta= 58^\circ, and ER=163E_R = 163 keV, b) En=3.0E_n = 3.0 MeV, θ=90∘\theta= 90^\circ, and ER=334E_R = 334 keV, c) En=14.8E_n = 14.8 MeV, θ=60∘\theta= 60^\circ, and ER=899E_R = 899 keV, d) En=14.8E_n = 14.8 MeV, θ=75∘\theta= 75^\circ, and ER=1305E_R = 1305 keV, and e) En=14.8E_n = 14.8 MeV, θ=90∘\theta= 90^\circ, and ER=1726E_R = 1726 keV. Quenching factors are determined by analyzing the fit to each distribution. The observed spectra in the on-time sample are in good agreement with the combination of off-time data and simulated spectrum for the recoil oxygen nuclei if the quenching factors are applied.

Fit results for five different configurations

Figure 10. Fit results for five different configurations. Neutron energy EnE_n and angle of the LS detector θ\theta are specified in each plot. Black dots are the on-time data overlaid with the off-time data (blue dashed-line), simulated energy distribution for elastic scattering scaled by the best fit value of QFQF (green dotted line), and combination of off-time data and simulated spectrum (red solid lines). Simulated spectra and off-time data are normalized with the best fit parameters.

Systematic uncertainties

In this study, we have considered systematic uncertainties to determine the QFQF of oxygen recoil energy. The main sources of uncertainties are off-time window selection and energy linearity. For the former source, the variation of QFQF was evaluated with the off-time window expanded by a factor of ∼2\sim2 and was computed as 0.87% in the setup of θ=90∘\theta= 90^\circ in En=3.0E_n = 3.0 MeV. For the latter source, the uncertainty was estimated from the fit error in the calibration (see Figure 4) as 6.8% in the same setup. Uncertainties for the other setups of θ\theta and EnE_n were estimated using this approach. It was observed that the total systematic uncertainties in the QFQF measurement were comparable to the statistical uncertainties for all setups. In addition, the uncertainties in the recoil energy (ERE_R) were also accounted. As the main sources, uncertainties due to the detector volume and initial neutron energy width were estimated with simulation. For the setup of θ=90∘\theta= 90^\circ in En=3.0E_n = 3.0 MeV, total uncertainty on ERE_R was evaluated as ±17\pm17 keV for 334 keV. Systematic uncertainties for all setups are summarized in Table 1.

EnE_n (MeV)θ\theta (deg.)ERE_R (keV)EvisE_{\mathrm{vis}} (keVee_{\mathrm{ee}})QF×102QF \times10^2
3.058±158 \pm1163±11163 \pm1125.2±2.425.2 \pm2.415.12−1.32+1.7415.12^{+1.74}_{-1.32} (stat.) ±\pm 1.87 (syst.)
90±190 \pm1334±17334 \pm1738.6±5.838.6 \pm5.811.49−0.41+0.5911.49^{+0.59}_{-0.41} (stat.) ±\pm 0.79 (syst.)
14.860±160 \pm1899±46899 \pm4676.9±1.176.9 \pm1.18.60−0.16+0.508.60^{+0.50}_{-0.16} (stat.) ±\pm 0.31 (syst.)
75±175 \pm11305±531305 \pm5396.6±2.196.6 \pm2.17.78−0.27+0.247.78^{+0.24}_{-0.27} (stat.) ±\pm 0.49 (syst.)
90±190 \pm11726±581726 \pm58116.9±10.2116.9 \pm10.26.91−0.21+0.226.91^{+0.22}_{-0.21} (stat.) ±\pm 0.24 (syst.)

Table 1. Summary of quenching factor measurements for oxygen recoil in BGO crystals. The scattering angle θ\theta, initial neutron energy EnE_n, and recoil energy ERE_R are listed with the measured values of the visible energy EvisE_{\mathrm{vis}} and quenching factor QFQF for oxygen nuclei.

Results

The measured QFQF values and uncertainties are summarized in Table 1 and also shown in Figure 11 as a function of nuclear recoil energy (ERE_R). The energy dependence of QFQF, especially the increase at lower energy, was clearly observed.

Quenching factor as a function of oxygen recoil energy

Figure 11. Quenching factor as a function of oxygen recoil energy. Black dots are the data with the statistical uncertainties (black bar) and systematic uncertainties (red band).

Conclusion

In this study, we conducted an experiment to measure the QFQF of recoil energy for oxygen nuclei in BGO scintillators by the irradiation of monochromatic neutrons. The QFQF measured at different nuclei recoil energies was in the range of 163–1726 keV and the significant energy dependence was confirmed.

Acknowledgments

This work is supported by the Neutron measurement consortium for Underground Physics and JSPS KAKENHI Grants Grant-in-Aid for Scientific Research (C) No. 20K03998 and Grant-in-Aid for Scientific Research on Innovative Areas 19H05808. We would like to thanks to National Institute of Advanced Industrial Science and Technology (AIST), Japan for providing the well-controlled neutron irradiation.

References

  1. [1]G. Ambrosi, et al., Direct detection of a break in the teraelectronvolt cosmic-ray spectrum of electrons and positrons, Nature 552 (2017) 63–66. arXiv:1711.10981, doi:10.1038/nature24475.DOI
  2. [2]S. Alef, et al., The BGOOD experimental setup at ELSA, Eur. Phys. J. A 56 (4) (2020) 104. arXiv:1910.11939, doi:10.1140/epja/s10050-020-00107-x.DOI
  3. [3]A. Calleja, et al., Recent Performance of Scintillating Bolometers Developed for Dark Matter Searches, J. Low Temp. Phys. 151 (3) (2008) 848–853. doi:10.1007/s10909-008-9752-3.
  4. [4]N. Coron, et al., A BGO scintillating bolometer as dark matter detector prototype, Opt. Mater. 31 (10) (2009) 1393–1397, including Special Issue: Papers from the Fourth Workshop on Cryogenic Scintillation CryoScint08. doi:https://doi.org/10.1016/j.optmat.2008.09.016.
  5. [5]M. Biassoni, C. Martinez, Study of supernova ν-nucleus coherent scattering interactions, Astropart. Phys. 36 (1) (2012) 151–155. doi:https://doi.org/10.1016/j.astropartphys.2012.05.009.
  6. [6]S. Cebrián, et al., The ROSEBUD experiment at Canfranc : 2001 report, Nucl. Phys. B - Proceedings Supplements 110 (2002) 97–99. doi:https://doi.org/10.1016/S0920-5632(02)01463-9.
  7. [7]N. Coron, et al., Scintillating and particle discrimination properties of selected crystals for low-temperature bolometers: from LiF to BGO, Nucl. Instr. Meth. A 520 (1) (2004) 159–162. doi:https://doi.org/10.1016/j.nima.2003.11.282.
  8. [8]H. Harano, et al., Accelerator-based Neutron Fluence Standard of the National Metrology Institute of Japan, AIP Conf. Proc. 1099 (2009) 915. doi:10.1063/1.3120189.DOI
  9. [9]V. N. Ivanchenko, Geant4 toolkit for simulation of HEP experiments, Nucl. Instrum. Meth. A 502 (2003) 666–668. doi:10.1016/S0168-9002(03)00538-2.DOI

Paper details

vlnc:W_zawfck7v
Version
v1 · latest