Charge calibration of MALTA2, a radiation hard depleted monolithic active pixel sensor

Abstract

MALTA2 is a depleted monolithic active pixel sensor (DMAPS) designed for tracking at high rates and typically low detection threshold of ∼150 e−\sim150\,\mathrm{e^-}. A precise knowledge of the threshold is crucial to understanding the charge collection in the pixel and specifying the environment for sensor application. A simple procedure is developed to calibrate the threshold to unit electrons making use of a dedicated charge injection circuit and an Fe-55 source with dominant charge deposition of 1600 e−1600\, \mathrm{e^-}. The injection voltage is determined which corresponds to the injection under Fe-55 exposure and is the basis for charge calibration. The charge injection circuit incorporates a capacitance with design value of Cinj=\mathrm{C_{inj}}= 230 aF. Experimentally, the average capacitance value for non-irradiated samples is found to be Cinj,exp=\mathrm{C_{inj,exp}}= 257 aF. The 12 % divergence motivates the need for the presented precise calibration procedure, which is proposed to be performed for each MALTA2 sensor.

Introduction

MALTA2 is the second prototype of the MALTA family of depleted monolithic active pixel sensors (MAPS) designed in Tower 180 nm CMOS imaging sensor technology [1, 2, 3]. The MALTA2 pixel with a pitch of 36.4 µm consists of either high resistivity epitaxial or Czochralski silicon. The front-end in every pixel, as shown in Figure 1, is optimized for threshold settings down to ∼150 e−\sim150\ \mathrm{e}^{-} for detection efficiencies above 99% as demonstrated for charged hadron beams [4]. The threshold is set globally by the different transistors, M0 to M9, by DAC settings corresponding to the gate of each transistor. Every pixel is also equipped with a dedicated charge injection circuit (Figure 2). It makes use of two voltage settings VHIGH and VLOW and a VPULSE signal. VHIGH and VLOW are subtracted by switching transistor M0 off and M1 on at the rising edge of VPULSE. The resulting voltage difference is capacitively coupled to the input node of the front-end through a metal-to-metal connection that has a capacitance extracted from simulation of Cinj=230 aFC_{\mathrm{inj}} = 230\ \mathrm{aF} [5]. The injected charge is therefore:

Image not extracted

Figure 1. MALTA2 front-end schematic including amplification, shaping and digitization of the analog signal per pixel [3]. IBIAS is the main biasing current and accounts for the majority of the power consumption. The current ITHR defines the speed of the feedback loop and is designed to linearly effect the threshold of the discriminator.

Image not extracted

Figure 2. MALTA2 charge injection circuit [5].

Qinj=Cinj ΔV=Cinj(VHIGH−VLOW).(1)Q_{\mathrm{inj}} = C_{\mathrm{inj}} \ \Delta V = C_{\mathrm{inj}}(V_{\mathrm{HIGH}} - V_{\mathrm{LOW}}). \tag*{(1)}

This document describes the procedure to measure the capacitance of the charge injection circuit of MALTA2 assuming a linear behaviour of the injected charge with respect to the difference of the VHIGH and VLOW voltages. ΔVFe55\Delta V_{\mathrm{Fe55}} is the voltage difference that injects the same signal as an Fe-55 source. It is assumed that the charge deposited by the 5.9 keV Kα-line of an Fe-55 source is 1600 e−\mathrm{e}^{-}. Radioactive sources with well defined X-ray lines are commonly selected for silicon sensors as calibration reference because of its point-like charge deposition. For thin sensors such as MALTA2, Fe-55 is selected due to its relatively low peak energy compared to other gamma sources and X-ray fluorescence lines [6, 7]. An alternative setup relying on Compton scattering requires multiple detectors as well as angle variation and is not suitable for frequent calibration of numerous samples [8]. In the following, the injected signals are quantified by the digital amplitude of the signal obtained through dedicated threshold scans, and ΔVFe55\Delta V_{\mathrm{Fe55}} will be measured by dedicated source scans.

Methods

Voltage measurement

The voltages VHIGH and VLOW are controlled by a DAC with a 7-bit range. The voltage produced by the DAC as a function of the value of VHIGH as measured with a Keithley 2400 is shown in Figure 3. A linear regime is observed from DAC value 0 to 90 with a gradient of 13.5 mV/DAC and an offset of 0.45 V. This value will be used to calculate the expected voltage of VHIGH. The behaviour of the VLOW DAC is identical. For DAC values larger than 90 the voltage saturates due to a buffer stage which adds a voltage shift of 0.4 V and restricts the linear voltage generation.

Image not extracted

Figure 3. VHIGHV_{\mathrm{HIGH}} voltage characteristic curve generated at the respective DAC value. The linear responds with a gradient of 13.5 mV/DAC holds up to a DAC value of 90.

Digital amplitude of injected charge

A threshold scan is a variation of the ITHR current DAC that is proportional to the threshold set in the discriminator in the pixel front-end. It is proportional to the speed at which the signal returns to the baseline. From the binary hit data, a signal can be quantified in a threshold scan through the digital amplitude, that is the threshold at which the number of hits are reduced to 50%. Figure 4a parameterises the digital amplitude of charge injected into a single pixel through an s-curve

Image not extracted

Figure 4. (a) Threshold scan of a single pixel under charge injection. (b) Histogram of digital amplitudes for all pixels.

s(x;C,a,b)=C2[1−erf⁡(x−a2b)](2)s(x; C, a, b) = \frac{C}{2}\left[1-\operatorname{erf}\left(\frac{x-a}{\sqrt{2}b}\right)\right] \tag*{(2)}

with the error-function definition

erf⁡(z)=2π∫0ze−t2 dt.(3)\operatorname{erf}(z)=\frac{2}{\sqrt{\pi}}\int_{0}^{z}e^{-t^{2}}\,\mathrm{d}t. \tag*{(3)}

Its differentiation

ddxs(x;C,a,b)=−CN(a,b2)(4)\frac{\mathrm{d}}{\mathrm{d}x}s(x;C,a,b)=-C\mathcal{N}(a,b^{2}) \tag*{(4)}

results in a Gaussian distribution with prefactor CC, mean aa and standard deviation bb. The digital amplitude is quantified through the position parameter aa of the error-function. The histogram of the digital amplitudes for all pixels in Figure 4b is described by a Gaussian distribution with mean amplitude μ=40.0\mu=40.0 and standard deviation σ=5.3\sigma=5.3 that quantifies pixel-to-pixel variations.

According to equation (1), the injected charge ideally depends on ΔV=VHIGH−VLOW\Delta V=V_{\mathrm{HIGH}}-V_{\mathrm{LOW}} but not on the individual setting of the two DAC values. Consequently, a variation of VHIGHV_{\mathrm{HIGH}} and VLOWV_{\mathrm{LOW}} should not affect the digital amplitude when keeping ΔV\Delta V constant. Figure 5 shows the mean digital amplitude as a function of VHIGHV_{\mathrm{HIGH}} and VLOWV_{\mathrm{LOW}} while keeping ΔV\Delta V constant at 10, 15, 20 and 25. A plateau forms for all ΔV\Delta V which defines the range in which all injections yield the same amplitude. At this plateau the amplitude does not depend on the specific value of VHIGHV_{\mathrm{HIGH}} or VLOWV_{\mathrm{LOW}}. The plateau is restricted towards large DAC values by VHIGH≤90V_{\mathrm{HIGH}}\leq90 and towards small DAC values by VLOW≥20V_{\mathrm{LOW}}\geq20. Both restrictions are indicated by dashed lines. At low voltages the PMOS transistors that do switching between VHIGHV_{\mathrm{HIGH}} and VLOWV_{\mathrm{LOW}} saturate. Hence, a reliable injection is obtained when keeping the DAC values within the range [20,90][20,90]. From this follows a maximum injection at ΔV=70\Delta V=70.

Image not extracted

Figure 5. (a) Injected digital amplitude versus VLOWV_{\mathrm{LOW}}. (b) Injected digital amplitude versus VHIGHV_{\mathrm{HIGH}}.

Digital amplitude of Fe-55 source

A gamma source of Fe-55 is used commonly as a calibration source for thin silicon sensors because the charge deposition through photon absorption is around 1600 e−^{-} which corresponds to the most probable charge deposition of a minimum-ionizing particle in 27 μ\mum of silicon [9]. The source emits photons at the Kα_{\alpha}-lines of 5.9 keV. The MALTA2 front-end is designed for thresholds down to ∼150\sim150 e−^{-} to ensure high detection efficiency of minimum-ionizing particles [3]. As a result, the usual threshold range is far below the charge deposition of 1600 e−^{-}. However, a low bias current setting in the front-end allows to suppress the gain and configure the threshold to values >2000>2000 e−^{-}. In such a low-gain setting, a threshold scan is sensitive to the amplitude deposited by soft X-rays. A threshold scan of all pixels of a MALTA2 sensor under exposure of an Fe-55 source is shown in Figure 6. An error-function fit describes the decrease in hits towards larger thresholds and yields a digital amplitude of aFe55=62a_{\mathrm{Fe55}}=62. The width parameter bFe55=11b_{\mathrm{Fe55}}=11 incorporates pixel-to-pixel variations.

Image not extracted

Figure 6. Threshold scan under exposure of an Fe-55 source. Data points (black dots) are the sum of hits from a MALTA2 sensor. The error-function fit (red line) is fitted to the high threshold regime (solid line) to parameterise the absorption of Kα_{\alpha} photons of 5.9 keV. The function is continued outside the fit regime (dashed line). The parameter aFe55a_{\mathrm{Fe55}} marks the position of the falling edge and bFe55b_{\mathrm{Fe55}} the width describing pixel-to-pixel variations. The deviation of fit function and data at low thresholds is due to charge sharing where an absorbed photon near a pixel corner causes multiple hits.

Fig. 4: The digital amplitude from charge injection with ΔV=VHIGH−VLOW=50\Delta V = V_{\mathrm{HIGH}} - V_{\mathrm{LOW}} = 50 DAC is obtained in (a) for a single pixel through the position parameter aa of an error-function fit with width bb. The factor CC scales with the number of injected pulses. As the threshold DAC current ITHRI_{\mathrm{THR}} is raised above the injected signal the number of detected hits decreases. The histogram in (b) of digital amplitudes for all 224×512224 \times512 pixels is described by a Gaussian fit to the core of the distribution. The stated values of the mean μ\mu and the standard deviation σ\sigma are those calculated from the data or obtained through the fit.

Fig. 5: Injected digital amplitude versus VLOWV_{\mathrm{LOW}} and VHIGHV_{\mathrm{HIGH}}. The digital amplitude remains constant in a stable DAC range of [20,90][20,90]. The dashed lines mark the lower boundary for VLOWV_{\mathrm{LOW}} and the upper boundary for VHIGHV_{\mathrm{HIGH}}. Outside that range, saturation effects reduce the injected charge which in turn causes a reduction in digital amplitude.

Calibration Results

Charge calibration through Fe-55 source

Charge is injected for different values of ΔV\Delta V and the mean digital amplitude is reconstructed according to the example for ΔV=50\Delta V = 50 in Figure 4. The amplitude is found to be proportional to ΔV\Delta V as shown in Figure 7. The independently measured digital amplitude aFe55a_{\mathrm{Fe55}} under the exposure of the Fe-55 source is used as a charge calibration point. From the linear fit the corresponding ΔVFe55\Delta V_{\mathrm{Fe55}} value is calculated. The charge injected through the circuit at ΔVFe55\Delta V_{\mathrm{Fe55}} yields the same amplitude as aFe55a_{\mathrm{Fe55}}.

Image not extracted

Figure 7. Average digital amplitude versus injection DAC setting. Data points represent the mean from the Gaussian fit to all pixel amplitudes and the error bars represent the standard deviation. The digital amplitude aFe55a_{\mathrm{Fe55}} from the Fe-55 detection is added as a calibration point to interpolate from the linear fit the corresponding charge injection ΔVFe55\Delta V_{\mathrm{Fe55}} that yields the same digital amplitude.

Further, the injection capacitance is calculated as

Cinj=1600 e−×1.602×10−19 C/e−ΔVFe55×13.5 mV/DAC=19000 aFΔVFe55/DAC(5)C_{\mathrm{inj}} = \frac{1600\,e^- \times1.602 \times10^{-19}\,\mathrm{C}/e^-}{\Delta V_{\mathrm{Fe55}} \times13.5\,\mathrm{mV}/\mathrm{DAC}} = \frac{19000\,\mathrm{aF}}{\Delta V_{\mathrm{Fe55}}/\mathrm{DAC}} \tag*{(5)}

by assuming a voltage input per unit ΔV\Delta V in the injection circuit of 13.5 mV/DAC. Table 1 lists the calibration results of ΔVFe55\Delta V_{\mathrm{Fe55}} and CinjC_{\mathrm{inj}} for all tested samples. Considering only the six non irradiated samples an average injection

sampledopingNIEL [1015 neqcm−2][10^{15}\ \mathrm{n}_{\mathrm{eq}}\mathrm{cm}^{-2}]ΔVFe55\Delta V_{\mathrm{Fe55}} [DAC]CinjC_{\mathrm{inj}} [aF]
W5R21high076±776 \pm7250±23250 \pm23
W8R24low087±987 \pm9218±23218 \pm23
W11R0high063±763 \pm7301±33301 \pm33
W14R11high085±885 \pm8223±21223 \pm21
W18R17very high075±675 \pm6253±20253 \pm20
W18R19very high064±564 \pm5297±23297 \pm23
W12R7high166±866 \pm8288±35288 \pm35
W18R1very high165±865 \pm8292±36292 \pm36
W18R4very high258±858 \pm8327±45327 \pm45
W18R9very high347±647 \pm6404±52404 \pm52
W18R21very high353±653 \pm6358±41358 \pm41
W18R12very high546±1046 \pm10413±90413 \pm90
W18R14very high541±841 \pm8463±90463 \pm90

Table 1. MALTA2 Calibration Results

capacitance of

Cinj,exp=257 aF(6)C_{\mathrm{inj,exp}} = 257\,\mathrm{aF} \tag*{(6)}

is found to be 12% larger than the design value of Cinj=230 aFC_{\mathrm{inj}} = 230\,\mathrm{aF}. For the Fe-55 equivalent injection DAC the average value is determined to be

ΔVFe55,exp=75±10 DAC.(7)\Delta V_{\mathrm{Fe55,exp}} = 75 \pm10\,\mathrm{DAC}. \tag*{(7)}

The stated uncertainty is the standard deviation of the sample and estimates the fluctuation among different sensors originating from the fabrication process.

Calibrated threshold scans

Based on the result of ΔVFe55\Delta V_{\mathrm{Fe55}} any charge injected into a MALTA2 sensor can be calibrated to unit electrons according to

Qinj=ΔV1600 e−ΔVFe55.(8)Q_{\mathrm{inj}} = \Delta V \frac{1600\,e^-}{\Delta V_{\mathrm{Fe55}}}. \tag*{(8)}

For a normal and low gain front-end setting the detection threshold is measured as an average over all pixels and is shown in Figure 8. The low gain is the result of reducing the main front-end bias current IBIAS by a factor of 14 compared to the normal setting and leads to larger threshold values. The thresholds in unit electrons (left y-axes) are obtained through equation (8) based on the measurement of ΔV\Delta V (right y-axes). Error bars quantify the error on the mean threshold and are of the same order than the marker size. The threshold dependence on ITHR can be parameterised by a linear or quadratic function. These threshold values define the detection threshold at testbeam studies and beam telescope applications [4, 10]. The threshold resolution quantifying the standard deviation from pixel to pixel variations is around 10%.

Image not extracted

Figure 8. (a) Thresholds for front-end at normal gain. (b) Thresholds for front-end at low gain.

Irradiation study

A selection of samples from the same wafer of 100 μ\mum thick Czochralski silicon with very high doping of the n- layer has been neutron irradiated at different fluences of non-ionizing energy loss (NIEL) at the Triga reactor in the Institute Jožef Stefan (IJS), Slovenia. The calibration results are compared in Figure 9 and show a decrease in the Fe-55 equivalent injection voltage ΔVFe55\Delta V_{\mathrm{Fe55}} with fluence. In irradiated samples, a lower injection voltage is needed to inject 1600 e−. This can be interpreted as an increase in the apparent injection capacitance. Two sensors are tested for each of the fluences at 1, 3 and 5×1015 neqcm−25 \times10^{15}\ \mathrm{n}_{\mathrm{eq}}\mathrm{cm}^{-2} and show compatibility within the order of the stated uncertainties.

Image not extracted

Figure 9. (a) ΔVFe55\Delta V_{\mathrm{Fe55}} versus NIEL. (b) Apparent injection capacitance versus NIEL.

The study shows that charge calibration has to account for the irradiation fluence for precise threshold determination.

Summary

The charge injection circuit of MALTA2 sensors has been calibrated inside a reliable DAC range of [20,90] for the parameters VHIGHV_{\mathrm{HIGH}} and VLOWV_{\mathrm{LOW}}. A digital amplitude is reconstructed through threshold scans from binary hit data as the position parameter of an error-function fit. The amplitude is proportional to the voltage input of the charge injection circuit ΔV=VHIGH−VLOW\Delta V = V_{\mathrm{HIGH}} - V_{\mathrm{LOW}} and is calibrated through an Fe-55 source. Based on this, the charge injected into a MALTA2 sensor can be calculated:

Qinj=ΔV1600 e−ΔVFe55(9)Q_{\mathrm{inj}} = \Delta V \frac{1600\ e^-}{\Delta V_{\mathrm{Fe55}}} \tag*{(9)}

It is proposed that for precise calibration ΔVFe55\Delta V_{\mathrm{Fe55}} is determined for each sample. For the six tested non irradiated sensors the mean and standard deviation of ΔVFe55,exp=75±10\Delta V_{\mathrm{Fe55,exp}} = 75 \pm10 is obtained. The mean value is used for calibration of non irradiated samples that are not calibrated individually. Neutron irradiated samples show a smaller value of VFe55V_{\mathrm{Fe55}} due to an increase in the effective injection capacitance. Once the charge calibration parameter ΔVFe55\Delta V_{\mathrm{Fe55}} is obtained, the threshold setting of a MALTA2 sensor is quantifiable at

uncertainties. The study shows that charge calibration has to account for the irradiation fluence for precise threshold determination. any front-end setting through charge injection. The described procedure is feasible for any binary sensor that incorporates an injection circuit.

Fig. 8: Calibrated threshold as a function of the ITHR DAC. For each ITHR value, the threshold is measured per pixel in units of injected charge ΔV\Delta V (right y-axes). The calibration input of ΔVFe55\Delta V_{\mathrm{Fe55}} makes a linear calibration to unit electrons according to equation (8) possible (left y-axes). (a) shows a normal gain setting at IBIAS=43 with thresholds down to 200 e−^{-} and a parabolic fit. (b) shows a low gain front-end setting at IBIAS=3 with thresholds above 750 e−^{-} and a linear fit.

Figure 9: Dependence of ΔVFe55\Delta V_{\mathrm{Fe55}} (a) and the corresponding injection capacitance (b) on the irradiation fluence as non-ionizing energy loss (NIEL). All samples originate from the same wafer of 100 μ\mum thick Czochralski silicon with very high doping of n- layer. ΔVFe55\Delta V_{\mathrm{Fe55}} is the voltage DAC value that injects a charge of 1600 e−^{-} corresponding to the main Fe-55 deposition. ΔVFe55\Delta V_{\mathrm{Fe55}} decreases with irradiation. Consequently, the apparent injection capacitance increases with irradiation because a lower voltage is needed to inject the same reference charge.

Acknowledgements

Acknowledgements. This project has received funding from the European Union’s Horizon 2020 Research and Innovation programme under Grant Agreement number 101004761 (AIDAinnova), and number 654168 (IJS, Ljubljana, Slovenia). Furthermore it has been supported by the Marie Sklodowska-Curie Innovative Training Network of the European Commission Horizon 2020 Programme under contract number 675587 (STREAM). This publication was funded by Deutsche Forschungsgemeinschaft (DFG, German Research Foundation) - 491245950.

Calibrated samples

Figure A1-Figure A5 show the calibration results for all 13 calibrated sensors of which six are not irradiated. They correspond to the Fe-55 threshold scan and injection calibration as described for the example sensor in Figure 6, Figure 7. All calibration results are summarized in Table 1.

Image not extracted

Figure A1.

Image not extracted

(a) W18R14 (5E15)

Image not extracted

(b) W18R14 (5E15)

Figure A5. Fig. A5

Image not extracted

(a) Injected digital amplitude [DAC] versus ΔV\Delta V [DAC]. aFe55=59±8a_{\mathrm{Fe55}}=59\pm8, ΔVFe55=85±8\Delta V_{\mathrm{Fe55}}=85\pm8. MALTA2 CZ NGAP. Injection data. y=mx+y0y=mx+y_0, m=1.0±0.2m=1.0\pm0.2, y0=−28±8y_0=-28\pm8.

Image not extracted

(b) Hits versus ITHRI_{\mathrm{THR}} [DAC]. MALTA2 CZ NGAP. Fe-55 data. f(x;C,a,b)f(x;C,a,b), C=3206C=3206, aFe55=59a_{\mathrm{Fe55}}=59 DAC, bFe55=8b_{\mathrm{Fe55}}=8 DAC.

Image not extracted

(c) Injected digital amplitude [DAC] versus ΔV\Delta V [DAC]. aFe55=43±6a_{\mathrm{Fe55}}=43\pm6, ΔVFe55=75±6\Delta V_{\mathrm{Fe55}}=75\pm6. MALTA2 CZ EDPW. Injection data. y=mx+y0y=mx+y_0, m=1.1±0.2m=1.1\pm0.2, y0=−39±11y_0=-39\pm11.

Image not extracted

(d) Hits versus ITHRI_{\mathrm{THR}} [DAC]. MALTA2 CZ EDPW. Fe-55 data. f(x;C,a,b)f(x;C,a,b), C=4731C=4731, aFe55=43a_{\mathrm{Fe55}}=43 DAC, bFe55=6b_{\mathrm{Fe55}}=6 DAC.

Image not extracted

(e) Injected digital amplitude [DAC] versus ΔV\Delta V [DAC]. aFe55=38±5a_{\mathrm{Fe55}}=38\pm5, ΔVFe55=64±5\Delta V_{\mathrm{Fe55}}=64\pm5. MALTA2 CZ EDPW. Injection data. y=mx+y0y=mx+y_0, m=1.1±0.1m=1.1\pm0.1, y0=−33±5y_0=-33\pm5.

Image not extracted

(f) Hits versus ITHRI_{\mathrm{THR}} [DAC]. MALTA2 CZ EDPW. Fe-55 data. f(x;C,a,b)f(x;C,a,b), C=8996C=8996, aFe55=38a_{\mathrm{Fe55}}=38 DAC, bFe55=5b_{\mathrm{Fe55}}=5 DAC.

Figure A2. Fig. A2

Image not extracted

Figure A3. Fig. A3

Image not extracted

Figure A4. (a) W18R9 (3E15); (b) W18R9 (3E15); (c) W18R21 (3E15); (d) W18R21 (3E15); (e) W18R12 (5E15); (f) W18R12 (5E15).

References

  1. [1]Pernegger, H., et al.: First tests of a novel radiation hard CMOS sensor process for Depleted Monolithic Active Pixel Sensors. JINST 12(06), 06008 (2017) https://doi.org/10.1088/1748-0221/12/06/P06008
  2. [2]Berdalovic, I., et al.: MALTA: a CMOS pixel sensor with asynchronous readout for the ATLAS High-Luminosity upgrade. In: Proc. IEEE Nucl. Sci. Symp. Med. Conf., p. 8824349 (2018). https://doi.org/10.1109/NSSMIC.2018.8824349
  3. [3]Piro, F., et al.: A 1 μW radiation-hard front-end in a 0.18 μm CMOS process for the MALTA2 monolithic sensor. IEEE Trans. Nucl. Sci. 69(6) (2022) https://doi.org/10.1109/TNS.2022.3170729
  4. [4]van Rijnbach, M., et al.: Radiation hardness of MALTA2 monolithic CMOS imaging sensors on Czochralski substrates. Eur. Phys. J. C 84(3), 251 (2024) https://doi.org/10.1140/epjc/s10052-024-12601-3 arXiv:2308.13231
  5. [5]Berdalovic, I.: Design of radiation-hard CMOS sensors for particle detection applications. PhD thesis, Zagreb U. (2019). Presented 29 Oct 2019. http://cds.cern.ch/record/2702884
  6. [6]Pohl, D.-L., et al.: Obtaining spectroscopic information with the ATLAS FE-I4 pixel readout chip. NIMA 788, 49–53 (2015) https://doi.org/10.1016/j.nima.2015.03.067
  7. [7]Fasselt, L., et al.: Energy calibration through X-ray absorption of the DECAL sensor, a monolithic active pixel sensor prototype for digital electromagnetic calorimetry and tracking. Frontiers in Physics 11 (2023) https://doi.org/10.3389/fphy.2023.1231336
  8. [8]McCormack, P., et al.: New Method for Silicon Sensor Charge Calibration Using Compton Scattering (2020). https://arxiv.org/abs/2008.11860
  9. [9]Meroli, S., et al.: Energy loss measurement for charged particles in very thin silicon layers. JINST 6(06), 06013 (2011) https://doi.org/10.1088/1748-0221/6/06/P06013
  10. [10]Gustavino, G., et al.: Development of the radiation-hard MALTA CMOS sensor for tracking applications. In: Proc. Vertex 2023, vol. 448, p. 048 (2024). https://doi.org/10.22323/1.448.0048

Paper details

vlnc:W_t8vc95xn
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 0

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…
Contents