Ultra-wideband optical coherence elastography from acoustic to ultrasonic frequencies

Abstract

Visualizing elastic waves by noninvasive imaging has been useful for analyzing the mechanical properties of materials and tissues. However, the maximum wave frequency of elastography has been limited to ~10 kHz due to the finite sensitivity to small vibration and finite imaging speed. Here, we present an optical coherence elastography technique that extends the frequency range to MHz by noise reduction, anti-aliasing demodulation, and advanced wave analysis. Our system can measure the stiffness of hard (GPa) materials including bones with mm-scale resolution and characterize soft, viscoelastic materials from 100 Hz to 1 MHz. The dispersion of Rayleigh surface waves over the wide frequency range allowed us to profile depth-dependent shear modulus (10 kPa to 100 MPa) in cartilages ex vivo and the human skin in vivo. This technique opened a new window for the characterization of materials in situ with 3-dimensional resolution.

Introduction

Measurement of the mechanical properties of materials is routinely performed in many places in sciences, engineering and industries, as well as hospitals[1, 2, 3, 4]. Widely used tools include strain-stress testing[5] and dynamic mechanical analysis[6] for measuring bulk properties, and atomic force microscopy (AFM)[7, 8] and microrheology[9, 10] for local measurements. In clinical medicine, elastography has been adopted for disease diagnosis[11, 12, 13, 14]. Elastography allows to noninvasively measure the elasticity of tissues in normal and abnormal states using medical imaging modalities, such as ultrasound and magnetic resonance imaging (MRI). In all these tools, samples under test are deformed by some force, their responses are measured, and the mechanical properties are calculated from the data[15, 16]. The timescale or frequency range of the measurement spans from quasi static (as slow as 10−510^{-5} Hz) to acoustic (as fast as 10310^{3} Hz) ranges. Higher speeds up to a few tens of kHz have been used in AFM[17] and in our recent work on dynamic optical coherence elastography (OCE)[18]. In the latter, elastic waves of short impulse or continuous-wave monotones are generated, their propagation within a tissue is visualized, and from the data the wave velocities are determined and related to the elastic moduli of the tissue[15, 19].

While the material analysis thus far has been focused on the quasi-static to acoustic ranges, we hypothesized that a higher frequency range beyond 10 kHz can offer a window of opportunity that was previously underappreciated especially for elastography. First, the higher frequency data can reveal the viscoelastic characteristics of materials in the shorter time scale. Dynamic mechanical analysis (DMA) is widely used to characterize viscoelasticity, but it has a limited frequency range (1 to 100 Hz). The time-temperature superposition technique can mimic the high-frequency responses of simple, homogeneous materials[20] but is not applicable to composite materials and living matters[9]. Ultrasound non-destructive testing (NDT)[21, 22] is an established technique using ultrasound waves, typically at 500 kHz to 20 MHz, to measure the time of flight across a material or structure with well-defined boundaries to determine bulk elastic properties, but this technique is not directly applicable to elastography and especially for complex materials such as tissues. Second, since the spatial resolution of elastography is approximately given by the wavelength of elastic waves[23], the higher frequency can lead to the higher resolution. Third, wave velocity dispersion over an extended frequency range is useful to extract more detailed mechanical information such as the depth-dependent variation of elasticity and internal stress. The depth information is not accessible using laser Doppler vibrometry or AFM-based rheology that only probes the surface[24, 25, 26].

Despite all these anticipated benefits mentioned above, it is technically challenging to extend the upper frequency limit. At the same amplitude of vibration, the energy dissipation or damping of a wave increases with ∼f3\sim f^{3} (ff, frequency) because the wave energy is proportional to f2f^{2} and the viscosity tends to grow with ff. So, with increasing frequency, the wave amplitude must be reduced by ∼f3/2\sim f^{3/2} to avoid excessive sample heating or damage. Detecting the reduced amplitude requires improved sensitivity, for example, by a factor of a thousand when comparing 100 kHz to 1 kHz. To analyze waves with submicron amplitudes, a nanometer-scale sensitivity is required to an instrument. OCE is an emerging elastography technique[11, 27] built on optical coherence tomography (OCT)[28]. OCE has been developed with a variety of different system architectures and applied to various tissue types, such as cornea[18, 29], sclera[30], breast[31], brain[32], and skin[33]. As OCE uses optical interferometry, it offers superior detection sensitivity in the order of nanometers, compared to micrometers for ultrasound and millimeters for MRI. The high sensitivity of OCE make it a good candidate for elastography beyond the acoustic range.

Here, we demonstrate an OCE system capable of visualizing elastic waves from the acoustic to ultrasonic frequencies up to a few MHz. We developed a novel aliasing technique, demodulation algorithm, and jitter-correction method to handle such high frequencies far beyond the typical axial line-scan (A-line) rates of OCT. After verifying the system with homogeneous materials, we apply it to dynamic shear analysis of soft materials, and stiffness mapping of complex tissues in a knee joint ex vivo and human skin in vivo with high, 3-dimensional resolution. Our results demonstrate the benefits and broad applicability of ultra-wideband OCE.

Results

Signal demodulation of high-frequency vibration in swept-source OCE

Figure 1a depicts a basic interferometer used in swept source OCT, where the optical wavenumber is tuned repeatedly with a period of TT: k=k0+k1[t]k = k_{0} + k_{1}[t], where [t]=mod⁡(t,T)−T/2[t] = \operatorname{mod}(t,T) - T/2, k0=2π/λk_{0} = 2\pi/\lambda (λ\lambda, center wavelength), and k1k_{1} is a tuning rate[34]. Imagine a mirror-like sample that vibrates with an amplitude δ\delta, angular frequency ωm(=2πfm)\omega_{m}(=2\pi f_{m}), and reflection coefficient rr at a mean depth of z0z_{0}. As the depth is modulated by δcos⁡(ωmt)\delta\cos(\omega_{m}t), we obtain an interferometric signal I(t)=rcos⁡(2k0z0+2k1z0t+2k0δcos⁡(ωmt)+2k1tδcos⁡(ωmt))I(t) = r\cos(2k_{0}z_{0} + 2k_{1}z_{0}t + 2k_{0}\delta\cos(\omega_{m}t) + 2k_{1}t\delta\cos(\omega_{m}t)), where 2k1z0T2k_{1}z_{0}T and ωmT\omega_{m}T are assumed to be multiples of 2π2\pi for convenience. Figure 1b illustrates I(t)I(t) for δ≪λ\delta\ll\lambda. For the typical case of k1T≪k0k_{1}T \ll k_{0} and introducing ω0=2k1z0\omega_{0} = 2k_{1}z_{0}, we get I(t)≈rcos⁡(2k0z0+ω0t+2k0δcos⁡(ωmt))I(t) \approx r\cos(2k_{0}z_{0} + \omega_{0}t + 2k_{0}\delta\cos(\omega_{m}t)). In the frequency domain, it consists of an amplitude rJ0(k0δ)rJ_{0}(k_{0}\delta) at the carrier frequency ω0\omega_{0} and two first harmonic sidelobes at ω0±ωm\omega_{0} \pm\omega_{m} with an amplitude of rJ1(k0δ)≈rk0δrJ_{1}(k_{0}\delta) \approx rk_{0}\delta. This is depicted in Figure 1c. In OCT, the frequency-domain analysis is performed by discrete Fourier transform F(ω)=FFT{I(t)}F(\omega) = FFT\{I(t)\} for each A-line data. The frequency resolution is about equal to the half of the A-line rate fA=1/Tf_{A} = 1/T. So, the first harmonic sidelobes overlap with the main peak when fm<0.5fAf_{m} < 0.5f_{A}. This condition has always been met in conventional OCE. The carrier and modulated components interfere with each other in the pixel at ω0\omega_{0} with a phase ϕ≈2k0z0+2k0δcos⁡(ωmt)\phi\approx2k_{0}z_{0} + 2k_{0}\delta\cos(\omega_{m}t). From the amplitude of the phase variation, δ\delta is readily measured. This method has been commonly used in OCE to date.

Image not extracted

Figure 1. Principle of high-frequency OCE. a, Schematic of a swept-source interferometric setup. A piezoelectric actuator excites elastic waves with angular frequency ωm\omega_m and amplitude profile δ(x,z)\delta(x,z). b, Example of time-domain detector signals without and with vibration of a mirror. The primary carrier frequency is given by the depth location (zz) of the reflector and the wavelength sweep speed. c, Frequency-domain transform of the trace in b. d, Overlap of the carrier and modulation components for low and high frequencies compared to the A-line rate (fAf_A). e, Graphical relationship between δ(z)\delta(z) and r(z)r(z) in the Fourier-domain at ω0−ωm\omega_0-\omega_m. Depending on whether r(z0−zm)r(z_0-z_m) is greater or smaller than the magnitude of rotating phasor r(z0)k0δ(z0)r(z_0)k_0\delta(z_0), the pixel phase undergoes an oscillatory or diverging pattern, respectively. For demodulation, the rotating phase circles (i) is brought to the origin (ii) by subtracting the offset r(z0−zm)r(z_0-z_m), and δ(z0)\delta(z_0) is determined from the magnitude of the counterclockwise rotating phasor.

We realized that the condition fm<fA/2f_{m} < f_{A}/2 is not necessarily required. When fm>fA/2f_{m} > f_{A}/2, the modulation sidelobes separate from the carrier (Figure 1d). In this case, one cannot directly apply the conventional method. We rewrite I(t)I(t)

I(t)→k0δ≪1Re⁡{rei2k0z0[eiω0t+k0δei(ω0−ωm)t+k0δei(ω0+ωm)t]}.I(t) \xrightarrow{k_{0}\delta\ll1} \operatorname{Re}\{re^{i2k_{0}z_{0}}[e^{i\omega_{0}t} + k_{0}\delta e^{i(\omega_{0}-\omega_{m})t} + k_{0}\delta e^{i(\omega_{0}+\omega_{m})t}]\}.

Now that the modulation component is separated from the carrier, the vibration amplitude k0δk_{0}\delta manifest itself in the amplitude of the Fourier component at ω0±ωm\omega_{0} \pm\omega_{m}. In this case, δ\delta can be measured from the amplitude, not the phase. However, this is only possible for the mirror-like sample when there is no other signal (background) present at ω0±ωm\omega_{0} \pm\omega_{m}. Consider a tissue-like sample that has continuous scattering points along its depth with r(z)r(z). The problem of retrieving δ\delta from the scatterer at z0z_{0} becomes more complicated. Now the Fourier component F(ω0−ωm)F(\omega_{0} - \omega_{m}) contains not only the vibration signal (negative sideband) originated from the scatterer at ω0\omega_{0} but also the time-independent reflection signal (carrier signal) from a different scatterer at z0−zmz_{0} - z_{m}, where zm=ωm/(2k1)z_{m} = \omega_{m}/(2k_{1}). Depending on whether r(z0−zm)>r(z0)kδr(z_{0} - z_{m}) > r(z_{0})k\delta or not, the phasor circle may or may not exclude the origin, and δ\delta is embedded differently in the phase and amplitude of F(ω0−ωm)F(\omega_{0} - \omega_{m}). Furthermore, F(ω0−ωm)F(\omega_{0} - \omega_{m}) also contains a vibration signal (positive sideband) originated from another scatterer located at z0−2zmz_{0} - 2z_{m}. Figure 1e (Figure 1i) shows a phasor diagram illustrating this situation.

Our solution to this problem is as follows. In one embodiment, we may add a large constant bias to F(ω0−ωm)F(\omega_{0} - \omega_{m}) to move the rotating phase circles far away from the origin and then retrieve δ(z0)\delta(z_{0}) from the time-varying amplitude of phase after proper renormalization with the bias and r(z0)=⟨F(ω0)⟩r(z_{0}) = \langle F(\omega_{0}) \rangle, where ⟨ ⟩\langle\ \rangle denotes time average. In another embodiment, the background ⟨F(ω0−ωm)⟩\langle F(\omega_{0} - \omega_{m}) \rangle is subtracted from F(ω0−ωm)F(\omega_{0} - \omega_{m}). This brings the phasor circles to the origin as depicted in Figure 1e (ii). δ(z0)\delta(z_{0}) and δ(z0−2zm)\delta(z_{0} - 2z_{m}) can then be identified from a Fourier transform of F−⟨F⟩F - \langle F \rangle. We used this second algorithm in experiments. Its rigorous mathematical description is provided in Methods. The situation of fm>fA/2f_{m} > f_{A}/2 is equivalent to signal aliasing that occurs when the signal frequency is greater than the sampling rate. In fact, our algorithm works for any fmf_{m} values regardless of whether it is aliased or not. Thus, it is a general algorithm for OCE.

Experimental demonstration of anti-aliasing demodulation

We used a swept-source OCT system previously built using a polygon-scanner wavelength-swept laser providing a center wavelength of 1307 nm, 3-dB bandwidth of 80 nm, axial resolution of 16 μ\mum, A-line rate fAf_{A} of 43.2 kHz, and maximum average optical power of 12 mW on samples (Supplementary Fig. S1). To test our demodulation scheme, we used a piezoelectric transducer (PZT) actuator block as a mirror-like sample and apply a sinusoidal waveform synchronously with a master clock in the OCT system (Supplementary Fig. S2). To generate mirror-like reflection, a flat glass plate was attached to the PZT. The laser power on the sample was attenuated to generate a signal-to-noise (SNR) of 40 dB at the air-glass interface. Figure 2a shows a typical A-line profile in the Fourier domain when fmf_{m} was 679.6 kHz. The signal appeared at the 107{sup}th pixel. The interval between pixels is 21.5 $\mu$m in depth or 52.7 kHz. Two first harmonic sidelobes emerged at about 12 pixels away from the main peak. The side lobes were lower by 20 dB in signal power (10 dB in amplitude) than the main lobe. Since the power ratio is $20\log_{10}(k_{0}\delta)$, we get $\delta = 20.4$ nm. The sideband position increased linearly with the vibration frequency, with the expected slope of 0.019 pixel / kHz ({xref}Fig. 2b`).

Image not extracted

Figure 2. Implementation of aliasing in OCE. a, Signal from the surface of a stationary PZT in the frequency (depth) domain without modulation (top) and with modulation at 679.6 kHz (bottom). b, Measured position of modulation peak at different modulation frequency. c, Several M-scan traces at 679.6 kHz before and after time jitter correction. d, Phase data of the M-scan traces showing noises before and after jitter correction. e, Phase noise performance measured with optical SNR of 40 dB and N=108N = 108. The SNR-limited phase noise corresponds to -94 dB/Hz. The 1/f1/f noise is also seen. Note that this graph is for 40 dB SNR. For lower SNR, the optical phase noise level would increase. For example, for 20 dB optical SNR, the optical noise level is moved up to 0.07 rad, crossing the 1/f1/f noise line at 200 Hz.

Noises and time jitter correction for ultrasonic frequencies

Various noises ultimately limits the system’s sensitivity to vibration and the accuracy of wave velocity measurement. The fundamental limit comes from optical SNR, XX, in A-lines. The length of the modulation-induced phasor is r(z0)k0δ(z0)r(z_{0})k_{0}\delta(z_{0}). The minimum detectable amplitude δmin⁡\delta_{\min} is then given by r(z0)k0δmin⁡(z0)=rmin⁡(z0−zm)r(z_{0})k_{0}\delta_{\min}(z_{0})=r_{\min}(z_{0}-z_{m}), where rmin⁡(z0−zm)=rmin⁡(z0)=r(z0)X(z0)−0.5r_{\min}(z_{0}-z_{m})=r_{\min}(z_{0})=r(z_{0})X(z_{0})^{-0.5} is the minimum detectable reflection coefficient by definition[35]. So, k0δmin⁡=X−0.5k_{0}\delta_{\min}=X^{-0.5}, same for non-aliased and aliased regimes. We note that although δ(z0)\delta(z_{0}) is obtained from sidebands at z0±zmz_{0}\pm z_{m}, its error is governed by the SNR of the main peak at z0z_{0}. In OCE, δ\delta is typically extracted from a set of M-mode data consisting of NN A-lines. Then, the sensitivity is improved by N\sqrt{N}, and k0δmin⁡=X−0.5N−0.5k_{0}\delta_{\min}=X^{-0.5}N^{-0.5}. For the above experiment, where N=108N=10^{8} and X=104X=10^{4}, we find δmin⁡=195\delta_{\min}=195 pm. Another fundamental noise source is 1/f electrical noise. SNR at low frequencies may be limited by the 1/f noise rather than the reflectivity-limited phase noise. Mechanical vibrations of optical components add noises. In our system, the mechanical jitters of beam scanners and rotating polygon filters were suppressed to practically acceptable levels.

Additionally, time jitters in electronics boards and signal generators generate phase noise. This noise increases in proportion to the modulation frequency and can become dominant at the ultrasonic range. In our system, we found that the sweep period TT has a Gaussian jitter with a standard deviation σT\sigma_{T} of 4.9 ns (Supplementary Fig. S3). The time error accumulates through successive sweeps (Figure 2c), resulting in a phase error σL\sigma_{L}. Let τi\tau_{i} denote the period jitter in the ii-th sweep, where ⟨τi⟩=0\langle\tau_{i}\rangle=0 and ⟨τi−⟨τi⟩⟩=σT\langle\tau_{i}-\langle\tau_{i}\rangle\rangle=\sigma_{T}. The timing of the mm-th A-line is given by tm=mT+∑i=1mτit_{m}=mT+\sum_{i=1}^{m}\tau_{i}. The output from the mm-th A-line is Fm=k0δeiωmtmF_{m}=k_{0}\delta e^{i\omega_{m}t_{m}}. δ\delta is determined from the inverse discrete Fourier transform of FmF_{m} at ωm\omega_{m}. The result is k0δ(1+iωm∑m=1N∑i=1mti/N)k_{0}\delta(1+i\omega_{m}\sum_{m=1}^{N}\sum_{i=1}^{m}t_{i}/N). The standard variation of the second term is σL≈ωmσT(N+1)(2N+1)/6N2\sigma_{L}\approx\omega_{m}\sigma_{T}\sqrt{(N+1)(2N+1)/6N^{2}} (Supplementary Note 1). For N≫1N\gg1, we get σL≈N/3 ωmσT\sigma_{L}\approx\sqrt{N}/3\ \omega_{m}\sigma_{T}. The total phase noise Δϕ\Delta\phi is a square sum of the SNR- and jitter-induced noises: Δϕ=1/NX2+0.58N(ωmσT)2\Delta\phi=\sqrt{1/NX^{2}+0.58N(\omega_{m}\sigma_{T})^{2}}. To measure the time jitter in real time and correct for it, we recorded the modulation waveform by using another digitizer channel in the system. The waveform was generated by a trigger synchronized with a particular laser wavelength, so it is supposed to contain the same time jitters as in the sweep period. We obtained the phase of the modulation waveform (applied to the PZT) and subtracted it from the phase of the demodulated signal of each M-scan (Supplementary Note 1). This phase correction was effective (Figure 2d). The total phase noise after the correction coincides with the SNR limited noise (Figure 2e).

Finally, the phase noise causes uncertainty in determining the wave velocity, vv, from the phase variation along the propagation distance. The error Δv\Delta v in velocity is given by Δv/v=Δϕ/2π(λ/L)\Delta v/v=\Delta\phi/2\pi(\lambda/L), where LL is the effective measurement length. In soft materials, elastic waves are attenuated, and LL is typically a couple of wavelengths. Therefore, Δϕ=0.01\Delta\phi=0.01 yields a high accuracy of Δv/v<1%\Delta v/v<1\%.

Ultrasonic OCE of hard materials

To verify our system at ultra-high frequencies, we measured various hard materials. Mechanical waves were excited on the surface by using a custom-made contact probe with the piezoelectric actuator (see Methods). For a 5 mm-thick acrylate plastic block, we obtained a flat dispersion curve over 1 to 2 MHz (Figure 3a). Since the wavelength (< 1.4 mm) is smaller than the thickness, the excited wave is the Rayleigh surface wave with a velocity cR=(0.862+1.14ν)/(1+ν)csc_R = (0.862 + 1.14\nu)/(1 + \nu)c_s, where csc_s is the pure (bulk) shear wave speed and ν\nu denotes the Poisson’s ratio. The shear elastic modulus μ\mu is given by μ=ρcs2\mu= \rho c_s^2, where ρ\rho is the density. From the measured cR=1285±16c_R = 1285 \pm16 m/s and known ρ=1.2\rho= 1.2 g/cm3^3 and ν=0.37\nu= 0.37, we obtained μ=2.26±0.06\mu= 2.26 \pm0.06 GPa. This value at MHz is slightly higher than the literature value of 1.7 GPa measured by quasi-static mechanical tools[36]. The difference is presumably due to the weak viscoelasticity of the material. Next, we measured a 1 mm-thick polystyrene plate. Figure 3b shows a representative kk-domain plot measured at 787.6 kHz. The dispersion relations for the A0 and S0 waves were fitted using the Lamb wave model[37]. The result was μ=1.64±0.05\mu= 1.64 \pm0.05 GPa, in comparison to quasi-static shear modulus of ~ 1.3 GPa reported in literature[38]. We also performed OCE on glass and metal (Supplementary Fig. S4). We obtained μ=24.0±0.61\mu= 24.0 \pm0.61 GPa for borosilicate glass and 45.7 ±1.79\pm1.79 GPa for copper (Fig. 3c). These values were consistent with their quasi-static shear moduli[39, 40], indicating their near-pure elasticity. We also performed OCE on the cortical surface of a bovine tibia bone specimen (ρ=1.6\rho= 1.6 g/cm3^3 and ν=0.3\nu= 0.3) and found μ\mu to vary from 2.36 ±0.04\pm0.04 GPa at 485 kHz to 3.60 ±0.12\pm0.12 GPa at 2 MHz (Fig. 3d). This frequency dependence is in part due to the viscoelasticity of the bone but also contributed by its depth-dependent variation of modulus. Later we will show how the wave velocity dispersion can be used to measure depth-dependent elasticity.

Image not extracted

Figure 3. Ultrasonic OCE of hard materials. a, Measured Rayleigh wave speeds (circles) at ultrasonic frequencies on an acrylate plastic block. Inset, a theoretical Rayleigh surface wave pattern. The measured surface displacement at 1 MHz is displayed. b, Measured asymmetric (A0) and symmetric (S0) Lamb wave modes on a 1 mm thick polystyrene bottom plate (circles) and the fitting result (dashed lines). Wavenumber domain plot displayed at 787.6 kHz resolves the two wave modes. c, Comparison of shear moduli measured by MHz OCE and literature values obtained from static mechanical testing. d, Measured wave speed on the cortical surface of a bovine tibia bone (circles) and the fitting result (dashed line). Vertical error bars represent the standard deviation of three measurements at different locations.

Dynamic shear analysis of uniform soft materials

Next, we measured soft polymer samples using surface waves excitation in a frequency range of 0.1 kHz to 100 kHz. Figure 4 displays the experimental and analysis data from a rubber (Ecoflex 00-50), polydimethylsiloxane (PDMS), and a tough hydrogel[41], respectively. Figures 4 (a, d, g) show measured surface displacement profiles at 40 kHz. The profiles contain not only the Rayleigh waves but also supershear surface waves[42, 43], which is evidenced by the two separated peaks in the wavenumber domain (Supplementary Fig. S5). We isolated the Rayleigh wave and derived the complex wave number kr+ikik_r + ik_i of the bulk shear wave (Supplementary Fig. S6). The measured speed (v=ωm/krv = \omega_m/k_r) and the coefficient of attenuation (kik_i) are shown in Fig. 4 (b, e, h). The complex shear modulus, μ′+iμ′′\mu' + i\mu'', is related to the complex wave numbers via:

μ′=ρωm2(kr2−ki2)(kr2−ki2)2−4kr2ki2(1)\mu' = \rho\omega_m^2 \frac{(k_r^2-k_i^2)}{(k_r^2-k_i^2)^2-4k_r^2k_i^2} \tag*{(1)}
μ′′=2ρωm2krki(kr2−ki2)2−4kr2ki2.(2)\mu'' = 2\rho\omega_m^2 \frac{k_rk_i}{(k_r^2-k_i^2)^2-4k_r^2k_i^2}. \tag*{(2)}

The measured storage and loss shear moduli are plotted in Figure 4 (c, f, i). They show characteristic frequency dependencies due to viscoelasticity. For the rubber, μ′\mu' and μ′′\mu'' increase from 50 to 160 kPa and from 9 to 323 kPa, respectively, over 0.2 kHz to 50 kHz. μ′′\mu^{\prime\prime} grows at a faster rate above 10 kHz and becomes greater than μ′\mu^{\prime} above 40 kHz. The crossover of μ′\mu^{\prime} and μ′′\mu^{\prime\prime} causes strong wave attenuation (ki/kr=0.41k_i/k_r = 0.41). When ki/kr>1k_i/k_r > 1, waves become overdamped, and their velocities cannot be determined. At low frequencies, our data agreed with previous measurements by MRI elastography, which showed μ′\mu^{\prime} and μ′′\mu^{\prime\prime} to increase from 30 to 80 kPa and from 10 to 40 kPa over 0.2 to 7.8 kHz[44]. However, the dramatic increase of μ′′\mu^{\prime\prime} beyond 10 kHz is only revealed by the ultrasonic OCE. For PDMS, our data up to 130 kHz show no crossover between μ′\mu^{\prime} and μ′′\mu^{\prime\prime}. The viscoelastic modulus of PDMS in the high frequency range was previously estimated only by the time-temperature superposition method using DMA data below 100 Hz[45]. Among the three materials, the tough hydrogel sample showed the lowest viscosity. The elasticity of this type of hydrogel has recently been measured using DMA below 100 Hz[41]. The low viscosity is due to the water that lubricates highly-entangled polymer chains and reduces energy dissipation[46]. Our data reveals the unusually low μ′′\mu^{\prime\prime} over the entire acoustic range, suggesting that this material is good at transmitting sounds and even ultrasounds.

Image not extracted

Figure 4. Ultra-wideband shear rheological analysis of uniform soft materials: a, b, c: Rubber, d, e, f: PDMS, g, h, i: Hydrogel. a, Picture of the rubber and its surface wave displacement profile at 40 kHz (Re: real part, Im: imaginary part). b, Rubber’s wave dispersion relation and attenuation. c, Rubber’s storage modulus and loss modulus. d, Picture of the PDMS and its Rayleigh surface wave displacement profile at 40 kHz. e, PDMS’s wave dispersion relation and attenuation. f, PDMS’s storage modulus and loss modulus. g, Picture of the hydrogel and its Rayleigh surface wave displacement profile at 40 kHz. h, Hydrogel’s wave dispersion relation and attenuation. i, Hydrogel’s storage modulus and loss modulus. Data are represented as mean values ±\pm std for measurements at three different locations. Dashed lines in (i) are curve fits using a power law μ∝fα\mu\propto f^{\alpha}.

Depth-resolved stiffness mapping of cartilage

We applied OCE to bovine cartilage in the knee joint ex vivo (Figure 5a). The firm, connective tissue consists of three layers with different mechanical properties: the noncalcified cartilage (NCC), calcified cartilage (CC) and the bone (Figure 5b). In our three-layer model, the thicknesses and shear moduli of NCC, CC and bone layer are denoted by hih_i and μi\mu_i (i=1,2,3i = 1, 2, 3), respectively. The NCC layer is readily distinguishable in the OCT image, from which we determined h1=300h_1 = 300 μ\mum. Unfortunately, the lower part of the CC layer and the underlying bone are not visible due to the limited optical penetration depth. We estimated the thickness of the CC layer to be h2=2.5h_2 = 2.5 mm from cross-sectional cuts of equivalent samples (see Methods) and assumed h3≫h2h_3 \gg h_2. We measured the elastic wave motions over 10 to 500 kHz. Figure 5d shows the dispersion of the Rayleigh waves, which shows several features. Besides a dramatic transition between 20 and 40 kHz, there appears to be a weak transition between 80 and 100 kHz. The Rayleigh surface wave has a 1/e amplitude decay depth equal to, approximately, a half wavelength. Therefore, as the frequency increases, the wave is increasingly confined near the surface, and its speed reflects the average stiffness of the region. The high speed below 20 kHz is due to the hard bone (μ3=3.27\mu_3 = 3.27 GPa, see Figure 2d). The first downward transition from 20 to 40 kHz is related to the bone-CC interface as the Rayleigh wave moves away from the bone. And the downward transition from 80 to 100 kHz is attributed to the CC-NCC interface and indicates that μ1<μ2\mu_1 < \mu_2. We developed a three-layer finite element analysis (FEA) model (see Method) and plotted the best-fit result in Figure 5c, from which we determined μ1=5.6±0.2\mu_1 = 5.6 \pm0.2 MPa and μ2=13.2±0.9\mu_2 = 13.2 \pm0.9 MPa.

Image not extracted

Figure 5. Depth-resolved stiffness mapping of joint tissue ex vivo. a, Picture of a bovine joint sample and the PZT contact probe. b, Cross-sectional OCT image with three layers labeled: noncalcified cartilage (NCC), calcified cartilage (CC), and the bone. The shear moduli and thickness of the NCC, CC and bone layers are denoted by μi\mu_i and hih_i (i=1,2,3i = 1, 2, 3), respectively. c, Measured Rayleigh surface wave speeds (circles) and the best-fit FEA result using the three-layer model (dashed curve). Error bars represent the standard deviation of four measurements from two samples at two locations. d-f, Side by side comparison of the measured cross-sectional displacement image and the FEA simulation at 100 kHz, 140 kHz, and 500 kHz, respectively. Dashed lines indicate the approximate boundaries of the NCC layer.

Our demodulation method allowed us to generate subsurface wave maps (Supplementary Fig. S7). Figure 5d shows cross-sectional displacement images obtained at 100, 140, and 500 kHz, respectively. It is apparent that the wavelength is longer in NCC than CC. We performed FEA using the geometrical and elastic parameters obtained from the curve fitting to generate displacement maps. The simulation reproduces the measured displacement fields quite well.

Depth-resolved stiffness mapping of in vivo human skin

We next examined the stiffness of human fingertip skin in vivo over 100 Hz to 500 kHz. We have previously measured the moduli of the epidermis, dermis (D), and hypodermis (H) in the forearm skin by OCE at acoustic frequencies[33]. With ultrasonic frequencies, we were interested in resolving the stratum corneum (SC)[47] and the viable epidermis (VE) layers within the epidermis (Figure 6a). Figure 6b show cross-sectional displacement images at different frequencies. At 20 kHz, the Rayleigh wave is confined predominantly in the SC and VE layers. Above 300 kHz, the majority of the elastic energy resides in the SC layer.

Image not extracted

Figure 6. Depth-resolved stiffness mapping of human skin in vivo. a, Cross-sectional OCT image of the fingertip skin in the normal condition. The stratum corneum (SC), viable epidermis (VE), dermis (D), and hypodermis (H) layers are labeled. b, Cross-sectional displacement images (only real part is shown) of normal fingertip at different frequencies. The layer interfaces and the half wavelength of the Rayleigh surface wave (λa/2\lambda_a/2) are marked. c, Measured Rayleigh surface wave speed over 2–500 kHz and the theoretical fitting curves for normal and hydration conditions. Error bars represent the standard deviation of three measurements from approximately same tissue regions. d, Calculated depth profiles of shear modulus for the normal and hydrated tissues. e, Cross-sectional OCT image of the same region after immersing in warm water for 20 min

The wave velocity dispersion curve has a strong frequency dependence in the normal skin condition (Figure 6c, red circles). To relate the dispersion to shear modulus, let us consider a depth-profile of elastic shear modulus, μ(z)\mu(z). There are no analytic solutions for wave propagation for arbitrary nonuniform modulus. For small nonuniformity, the wavenumber can be estimated from k2=ρω2∫0∞ψ∗1μ(z)ψ dzk^2 = \rho\omega^2 \int_{0}^{\infty} \psi^* \frac{1}{\mu(z)} \psi\, dz,

where ψ(z)\psi(z) is the wave function in a uniform medium with a space-averaged modulus: ∣ψ(z)∣2≈2kze−2kzz\lvert\psi(z)\rvert^2 \approx2k_z e^{-2k_z z}, where kz≈0.31 kk_z \approx0.31\,k. We may extend the perturbation theory to large variation with a simplified form:

k2≈ρω2Lz∫0Lz1μ(z) dz(3)k^2 \approx\frac{\rho\omega^2}{L_z} \int_{0}^{L_z} \frac{1}{\mu(z)}\,dz \tag*{(3)}

where LzL_z denotes the penetration depth of the Rayleigh wave. We may choose Lz=1/2kz≈0.25 λL_z = 1/2k_z \approx0.25\,\lambda or more generally, Lz=aλ=av/fL_z = a\lambda= av/f, where aa is a fitting constant. By differentiating both sides of Eq. (3) with respect to ff, we obtain

μ(z)≈ρv2(v−fv′)(v+fv′)(4)\mu(z) \approx\rho v^2 \frac{(v-fv')}{(v+fv')} \tag*{(4)}

where z=av/fz = av/f, v=1.048 vRv = 1.048\,v_R is bulk shear velocity obtained from the measured Rayleigh wave velocity vR(f)v_R(f), and v′=dv/dfv' = dv/df. Figure 6d shows μ(z)\mu(z) obtained from the experimental data. We find μ(z)\mu(z) to decay rapidly from 56 MPa at the surface to 100 kPa at the asymptotic depth (using a=0.25a=0.25). This finding is consistent with the fact that water content decreases exponentially from the VE to the surface of SC[48] and that the elastic modulus decreases exponentially with depth in SC.

We measured the same skin immediately after the fingertip was immersed in warm water for 20 min. The thickness of the SC layer increased from 0.31 mm to 0.41 mm due to swelling (Figure 6e). The wave velocity at high frequencies decreased significantly (Figure 6c, yellow dots), indicating a softening of the SC layer. The Rayleigh waves became overdamped at frequencies above 50 kHz due to high viscous damping. Using Eq. (4), we obtained μ=2.6\mu=2.6 MPa at the surface of the hydrated SC.

To verify, we performed FEA simulation by modeling the skin as a four-layer structure. We measured the shear moduli of the dermis (μ2=9\mu_2 = 9–1717 kPa) and hypodermis (μ3=2\mu_3 = 2–33 kPa) layers from wave dispersion below 1 kHz[33]. The shear modulus in the SC layer at the pre-hydrated condition was assumed to have an exponentially decreasing function from μ0=50\mu_0 = 50 MPa at the surface to μ1=2\mu_1 = 2 MPa at the interface between the SC and VE layers (Supplementary Fig. S8). In the hydrated condition, the stiffness of the swollen SC layer was assumed to be uniform and the same as the VE layer (μ0=μ1=2 MPa\mu_0 = \mu_1 = 2\ \mathrm{MPa}). In Figure 6c we plot the dispersion relations obtained by the FEA simulation, which captures the salient features of the dispersion relations.

Discussion

We have demonstrated an OCE system capable of visualizing elastic waves within various materials and tissues over a wide range of shear modulus from 1 kPa to 10 GPa and across an unprecedented frequency range from static to a few MHz. The fundamental reflectivity-limited performance of the system provided sufficient SNR to determine wave velocities with high accuracy. The measured speed dispersion over the wide frequency range was critical to extract depth-dependent stiffness from multi-layered tissues commonly found in various tissues. The contact area of the PZT probe was optimized to be less than half the wavelength to maximize acoustic impedance matching from the actuator motion to the elastic waves. Throughout the experiments, the peak wave amplitude ranged from several hundreds of nm at acoustic frequencies to as small as 10 nm at ultrasonic frequencies.

The demodulation scheme was highly effective and allowed the upper frequency limit to be no longer limited by the A-line rate. This relaxes the system requirement. The demodulation scheme should be applicable to any swept source OCT systems regardless of their A-line rates. It should be noted that the aliasing scheme does not work for spectra-domain OCT because signal modulation during the A-line acquisition period is averaged out due to the integrating nature of CCD spectrometers (equivalent detector bandwidth is only fA/2f_A/2). This is analogous to the phase washout motion artifact in spectral-domain OCT[34]. Swept-source OCT is free from the phase washout as long as fmf_m is lower than the detector bandwidth, which is typically 1000 times fAf_A.

We envision ultra-wideband OCE to be a powerful tool for mechanical characterizations of soft and hard tissues with high spatio-temporal resolution. For example, ultra-wideband OCE may be useful in capturing the full spectrum of the complex rheology of tissues[49, 50, 51] and biomaterials[52, 53]. With the capability of 3D high-resolution stiffness mapping, ultra-wideband OCE has the potential to enable novel biomechanics-based clinical diagnostics[54, 55, 56]. Besides, the ability to detect up to MHz excitation may be useful to study fast neuronal activities through neuro-mechanical coupling and develop novel therapeutics to treat brain diseases[57]. Furthermore, ultra-wideband OCE may be used to extract mechanical stress in thin structures without prior knowledge of material properties[58]. Finally, the emergence of deep learning offers a new approach to apply the information-rich elastography data to multi-parameter inverse problems[59].

Materials and Methods

Swept-source OCT signal. The photodetector current I(t)I(t) of the OCT interferometry from a point scatterer located at z0z_0 can be expressed as

I(t)=rP(t)cos⁡(2k0z0+2k1tz0).(5)I(t)=rP(t)\cos(2k_0z_0+2k_1tz_0). \tag*{(5)}

where rr denotes the reflectance amplitude, P(t)P(t) denotes the optical power profile, k=k0+k1tk=k_0+k_1t denotes the wavenumber of the swept source. The Fourier transformation of the photocurrent acquired during time interval −T/2≤t≤T/2-T/2\leq t\leq T/2 with respect to k^=2k1t\hat{k}=2k_1t, denoted by F(z)F(z), is given by

F(z0;z)=∫−k1Tk1TI(k^)eik^z dk^,(6)F(z_0;z)=\int_{-k_1T}^{k_1T}I(\hat{k})e^{i\hat{k}z}\,d\hat{k}, \tag*{(6)}

where TT is the period of wavelength sweep (A line period). We assume that the optical power profile has a Gaussian shape with a full width at half maximum (FWHM) σT\sigma T, i.e., P(t)=exp⁡[−4ln⁡2t2(σT)2]P(t)=\exp\left[-4\ln2\frac{t^2}{(\sigma T)^2}\right]. Then we get

F(z0;z)=r2∫−k1Tk1Texp⁡[−4ln⁡2k^2(2σk1T)2]{e−i(2k0z0+k^z0)+c.c.}eik^z dk^,(7)F(z_0;z)=\frac{r}{2}\int_{-k_1T}^{k_1T}\exp\left[-4\ln2\frac{\hat{k}^2}{(2\sigma k_1T)^2}\right]\left\{e^{-i(2k_0z_0+\hat{k}z_0)}+\mathrm{c.c.}\right\}e^{i\hat{k}z}\,d\hat{k}, \tag*{(7)}

where ‘c.c.’ denotes the complex conjugate. When σ<1\sigma<1, we can approximate the integral range to infinity and get

F(z0;z)=r2(πln⁡2σk1T){F0(z)+c.c.}.(8)F(z_0;z)=\frac{r}{2}\left(\sqrt{\frac{\pi}{\ln2}}\sigma k_1T\right)\{F_0(z)+\mathrm{c.c.}\}. \tag*{(8)}

where we introduce F0(z)F_0(z) that is defined as

F0(z)=e−4ln⁡2(z−z0)2(σz)2e−i2k0z0.(9)F_0(z)=e^{-4\ln2\frac{(z-z_0)^2}{(\sigma_z)^2}}e^{-i2k_0z_0}. \tag*{(9)}

where σz≡4ln⁡2σk1T\sigma_z\equiv\frac{4\ln2}{\sigma k_1T} corresponds to the FWHM axial resolution.

Without loss of generality, the displacement of the scatter along the laser beam can be expressed as Δz(t)=δg(t)\Delta z(t)=\delta g(t), where δ\delta is a constant and g(t)g(t) is a dimensionless function of time. Then the location of the scatter is z0+Δzz_0+\Delta z. Since the laser is periodically tuned, we introduce t^=t−mT\hat{t}=t-mT for the mm-th A line acquired during mT−T/2≤t≤mT+T/2mT-T/2\leq t\leq mT+T/2, and gm(t^)=g(t^+mT)g_m(\hat{t})=g(\hat{t}+mT). Substitution of z0z_0 and P(t)P(t) in (5) with z0+δgm(t^)z_0+\delta g_m(\hat{t}) and P(t^)P(\hat{t}), we can get

Im(t^)=rP(t^)cos⁡(2k0z0+2k1t^z0+2k0δgm(t^)+2k1t^δgm(t^)).(10)I_m(\hat{t})=rP(\hat{t})\cos(2k_0z_0+2k_1\hat{t}z_0+2k_0\delta g_m(\hat{t})+2k_1\hat{t}\delta g_m(\hat{t})). \tag*{(10)}

We make assumption that the wavelength tuning range is a small fraction of the mean wavelength, k1t^≪k0k_1\hat{t}\ll k_0, to get

Im(t^)≈rP(t^)cos⁡(2k0z0+2k1t^z0+2k0δgm(t^))=rP(t^){e−i(2k0z0+2k1t^z0+2k0δgm(t^))+c.c.}=rP(t^){e−i(2k0z0+2k1t^z0)[1+(−2ik0δ)gm(t^)+12(−2ik0δ)2gm2(t^)+⋯ ]+c.c.}.(11)\begin{aligned} I_m(\hat{t})&\approx rP(\hat{t})\cos(2k_0z_0+2k_1\hat{t}z_0+2k_0\delta g_m(\hat{t}))\\ &=rP(\hat{t})\{e^{-i(2k_0z_0+2k_1\hat{t}z_0+2k_0\delta g_m(\hat{t}))}+\mathrm{c.c.}\} \\ &= rP(\hat{t})\left\{e^{-i(2k_{0}z_{0}+2k_{1}\hat{t}z_{0})}\left[1+(-2ik_{0}\delta)g_{m}(\hat{t})+\frac{1}{2}(-2ik_{0}\delta)^{2}g_{m}^{2}(\hat{t})+\cdots\right]+\mathrm{c.c.}\right\}. \tag*{(11)} \end{aligned}

We define

Gm(1)(z)=∫−∞+∞gm(t^)eik^z dk^,Gm(2)(z)=∫−∞+∞gm2(t^)eik^z dk^.(12)\begin{aligned} G_{m}^{(1)}(z) &= \int_{-\infty}^{+\infty}g_{m}(\hat{t})e^{i\hat{k}z}\,\mathrm{d}\hat{k},\\ G_{m}^{(2)}(z) &= \int_{-\infty}^{+\infty}g_{m}^{2}(\hat{t})e^{i\hat{k}z}\,\mathrm{d}\hat{k}. \tag*{(12)} \end{aligned}

The Fourier transform of the photocurrent I(t^)I(\hat{t}) is

\begin{aligned} F_{m}(z_{0};z) &\approx\frac{r}{2}\left(\sqrt{\frac{\pi}{\ln2}}\sigma k_{1}T\right)\left\{\left[F_{0}(z)+(-2ik_{0}\delta)F_{0}(z)*G_{m}^{(1)}(z)+\frac{1}{2}(-2ik_{0}\delta)^{2}F_{0}(z)*\right.\\ &\qquad\left.G_{m}^{(1)}(z)+\cdots\right]+\mathrm{c.c.}\right\}, \tag*{(13)} \end{aligned}

where ‘*’ denotes the convolution.

For harmonic vibration studied here, Δz=δsin⁡(ωmt+φ)\Delta z=\delta\sin(\omega_{m}t+\varphi), where δ\delta, φ\varphi, and ωm\omega_{m} denote the vibration amplitude, initial phase, and angular frequency, respectively. Then have gm(t^)=sin⁡(ωmt^+mωmT+φ)g_{m}(\hat{t})=\sin(\omega_{m}\hat{t}+m\omega_{m}T+\varphi), and

Gm(1)(z)=12i[ei(mωmT+φ)δ(z+ωm2k1)−e−i(mωmT+φ)δ(z−ωm2k1)],G_{m}^{(1)}(z)=\frac{1}{2i}\left[e^{i(m\omega_{m}T+\varphi)}\delta\left(z+\frac{\omega_{m}}{2k_{1}}\right)-e^{-i(m\omega_{m}T+\varphi)}\delta\left(z-\frac{\omega_{m}}{2k_{1}}\right)\right],
Gm(2)(z)=12[δ(z)−12(e2i(mωmT+φ)δ(z+ωmk1)+e−2i(mωmT+φ)δ(z−ωmk1))].(14)G_{m}^{(2)}(z)=\frac{1}{2}\left[\delta(z)-\frac{1}{2}\left(e^{2i(m\omega_{m}T+\varphi)}\delta\left(z+\frac{\omega_{m}}{k_{1}}\right)+e^{-2i(m\omega_{m}T+\varphi)}\delta\left(z-\frac{\omega_{m}}{k_{1}}\right)\right)\right]. \tag*{(14)}

where δ(z)\delta(z) is the Dirac delta function. Inserting Eq. (13) we get

Fm(z0;z)≈r2(πln⁡2σk1T){[(1−(k0δ)2)F0(z)−(k0δ)(ei(mωmT+φ)F0(z+ωm2k1)−e−i(mωmT+φ)F0(z−ωm2k1))+12(k0δ)2(e2i(mωmT+φ)F0(z+ωmk1)+e−2i(mωmT+φ)F0(z−ωmk1))+⋯ ]+c.c.}.(15)\begin{aligned} F_{m}(z_{0};z) &\approx\frac{r}{2}\left(\sqrt{\frac{\pi}{\ln2}}\sigma k_{1}T\right)\left\{\left[(1-(k_{0}\delta)^{2})F_{0}(z)-(k_{0}\delta)\left(e^{i(m\omega_{m}T+\varphi)}F_{0}\left(z+\frac{\omega_{m}}{2k_{1}}\right)\right.\right.\right.\\ &\qquad\left.\left.\left.-e^{-i(m\omega_{m}T+\varphi)}F_{0}\left(z-\frac{\omega_{m}}{2k_{1}}\right)\right)+\frac{1}{2}(k_{0}\delta)^{2}\left(e^{2i(m\omega_{m}T+\varphi)}F_{0}\left(z+\frac{\omega_{m}}{k_{1}}\right)\right.\right.\right.\\ &\qquad\left.\left.\left.+e^{-2i(m\omega_{m}T+\varphi)}F_{0}\left(z-\frac{\omega_{m}}{k_{1}}\right)\right)+\cdots\right]+\mathrm{c.c.}\right\}. \tag*{(15)} \end{aligned}

Taking the first order approximation, we have

Fm(z0;z)≈r2(πln⁡2σk1T){[F0(z)−(k0δ)(ei(mωmT+φ)F0(z+ωm2k1)−e−i(mωmT+φ)F0(z−ωm2k1))]+c.c.}.(16)\begin{aligned} F_{m}(z_{0};z) &\approx\frac{r}{2}\left(\sqrt{\frac{\pi}{\ln2}}\sigma k_{1}T\right)\left\{\left[F_{0}(z)-(k_{0}\delta)\left(e^{i(m\omega_{m}T+\varphi)}F_{0}\left(z+\frac{\omega_{m}}{2k_{1}}\right)\right.\right.\right.\\ &\qquad\left.\left.\left.-e^{-i(m\omega_{m}T+\varphi)}F_{0}\left(z-\frac{\omega_{m}}{2k_{1}}\right)\right)\right]+\mathrm{c.c.}\right\}. \tag*{(16)} \end{aligned}

Aliasing of the vibration signal

Aliasing of the vibration signal. To derive the vibration signal from the OCT measurement, we consider one of the side lobes in Eq. (16), Fm(z0;z0−ωm2k1)F_m\left(z_0;z_0-\frac{\omega_m}{2k_1}\right), and find

Fm(z0;z0−ωm2k1)∝(k0δ)e−i2k0z0eimωmTeiφ,(17)F_m\left(z_0;z_0-\frac{\omega_m}{2k_1}\right)\propto(k_0\delta)e^{-i2k_0z_0}e^{im\omega_mT}e^{i\varphi}, \tag*{(17)}

which retains the vibration signal. Denote the frequency fm=ωm2πf_m=\frac{\omega_m}{2\pi} and the A line rate fA=1Tf_A=\frac{1}{T}. When fm>0.5fAf_m>0.5f_A, Eq. (17) gives rise to the aliasing effect. Denote fm=fˉm+0.5nfAf_m=\bar{f}_m+0.5nf_A, where nn is an integer to make 0≤fˉm<0.5fA0\leq\bar{f}_m<0.5f_A. Then,

Fm(z0;z0−ωm2k1)∝(k0δ)e−i2k0z0ei2πmfˉmTeiφ,(18)F_m\left(z_0;z_0-\frac{\omega_m}{2k_1}\right)\propto(k_0\delta)e^{-i2k_0z_0}e^{i2\pi m\bar{f}_mT}e^{i\varphi}, \tag*{(18)}

when nn is an even number, and we can get the apparent frequency fˉm\bar{f}_m.

Fm(z0;z0−ωm2k1)∝(k0δ)e−i2k0z0ei2πm(fˉm−0.5fA)Teiφ,(19)F_m\left(z_0;z_0-\frac{\omega_m}{2k_1}\right)\propto(k_0\delta)e^{-i2k_0z_0}e^{i2\pi m(\bar{f}_m-0.5f_A)T}e^{i\varphi}, \tag*{(19)}

when nn is an odd number and we get the apparent frequency (fˉm−0.5fA)(\bar{f}_m-0.5f_A).

Demodulation of vibrations from internal scatters

Given the axial coordinate z0z_0, without loss of generality, the Fourier transform of the photocurrent for the mm-th scan, denoted by Fm(z0)\mathcal{F}_m(z_0), is

Fm(z0)≈Fm(z0;z0)+Fm(z0+zm;z0)+Fm(z0−zm;z0),(20)\mathcal{F}_m(z_0)\approx F_m(z_0;z_0)+F_m(z_0+z_m;z_0)+F_m(z_0-z_m;z_0), \tag*{(20)}

where zm=ωm2k1z_m=\frac{\omega_m}{2k_1}. The latter two, comes from the contributions of the sidelobes of two scatters located at z0+zmz_0+z_m and z0−zmz_0-z_m (denoted by S’ and S’’), respectively. As the sidelobes contain the vibration signals, we expect to extract the vibrations of S’ and S’’ from Fm(z)\mathcal{F}_m(z).

Note that the sum ∑m=1NFm(z0±zm;z0)=0\sum_{m=1}^{N}F_m(z_0\pm z_m;z_0)=0, given that Nfm/fANf_m/f_A is an integer, where NN is the number of A lines. We have 1N∑m=1NFm(z0)=Fm(z0;z0)\frac{1}{N}\sum_{m=1}^{N}\mathcal{F}_m(z_0)=F_m(z_0;z_0), the static OCT signal of the scatter located at z0z_0. Subtracting this term from Fm(z0)\mathcal{F}_m(z_0) we can get,

Fm(z0+zm;z0)+Fm(z0−zm;z0)≈Fm(z0)−1N∑m=1NFm(z0).(21)F_m(z_0+z_m;z_0)+F_m(z_0-z_m;z_0)\approx\mathcal{F}_m(z_0)-\frac{1}{N}\sum_{m=1}^{N}\mathcal{F}_m(z_0). \tag*{(21)}

The left side of Eq. (21) contains the left sidelobe of the scatter S’ and the right sidelobe of scatter S’’. Here we show the extraction of the vibration for S’ as an example. Note that 1N∑m=1NFm(z0+zm)=Fm(z0+zm;z0+zm)\frac{1}{N}\sum_{m=1}^{N}\mathcal{F}_m(z_0+z_m)=F_m(z_0+z_m;z_0+z_m), the static OCT signal of S’.

(k0δz0+zm)ei(mωmT+φ)+Fm(z0−zm;z0)1N∑m=1NFm(z0+zm)≈Fm(z0)−1N∑m=1NFm(z0)1N∑m=1NFm(z0+zm).(22)(k_0\delta_{z_0+z_m})e^{i(m\omega_mT+\varphi)}+\frac{F_m(z_0-z_m;z_0)}{\frac{1}{N}\sum_{m=1}^{N}F_m(z_0+z_m)}\approx\frac{\mathcal{F}_m(z_0)-\frac{1}{N}\sum_{m=1}^{N}\mathcal{F}_m(z_0)}{\frac{1}{N}\sum_{m=1}^{N}F_m(z_0+z_m)}. \tag*{(22)}

where δz0+zm\delta_{z_0+z_m} denotes the vibration amplitude at z0+zmz_0+z_m. The first term of the left side in Eq. (22) gives the vibration of S′S'. To eliminate the second term that results from the contribution of the right sidelobe of S′′S'', we note that the phase change of the right sidelobe has an opposite direction as the left side lobe. The two terms can be separated in the frequency domain. Then the last step to extract the vibration of S′S' is to perform Fourier transform to right side of Eq. (22). An example of the demodulation algorithm is provided in Supplementary Fig. S1.

Experimental setup. Our experimental setup (Supplementary Fig. S2) is based on a home-built, swept source OCT system[60]. The system uses a rotating-polygon-mirror wavelength-swept laser[61] with a central wavelength of 1307 nm and a 3-dB bandwidth of 80 nm at an A-line rate of 43.2 kHz. The axial resolution is 16 μm. The optical beam is scanned by a pair of galvanometer mirror scanners and focused by a wide-aperture scan lens (Thorlabs, LSM54–1310) yielding a long working distance of 64 mm and a transverse resolution of ∼30 μm. The average optical power on the sample is 12 mW. A fiber Bragg grating (FBG) and photodiode (PD) provide a pulse signal synched to each wavelength sweep cycle of the laser output. This FBG optical clock ensures time synchronization among the modulation waveform to PZT, OCT beam position scan, and OCT data acquisition. The detector signal was digitized by a data acquisition board (Signatec, PX14400) at a sampling rate of 108 MHz. A total 2048 data points were acquired during each A-line. An input/output (I/O) board (National Instruments, USB-6353) is used to generate analog waveforms for the galvanometer scanners. The I/O board or a function generator (Tektronix, AFG3021C) is used to generate stimulus waveforms. The waveforms are smoothed with a reconstruction filter (Thorlabs, EF122) before feeding to the PZT. During data acquisition, the stimulus waveforms applied to the PZT are simultaneously recorded by the I/O board, which was later used for time jitter correction. Figure S2 depicts the timing diagram for the data acquisition protocol. The M-B scan protocol comprises the acquisition of MM consecutive A-lines at each transverse location. The OCT beam is then moved to the next position, and the stimulus waveform is repeated. A total of 96 positions are acquired. Typical, we set MM to be in the range of 100 to 250. For M=108M = 108, a single M-scan at each transverse location takes 2.5 ms, and the total measurement time per frequency is 0.24 s.

Image not extracted

Figure S2. Timing diagrams of the OCE system. The FBG optical clock provides the sync signal for the I/O board that controls the stimulus signal, the OCT beam scan, and data acquisition. For stimulus frequencies below 21.6 kHz, the I/O board generates the harmonic stimulus directly. For stimulus frequencies above 21.6 kHz, the I/O board generates a stimulus trigger to externally trigger a function generator to send the harmonic stimulus.

The harmonic waveforms are sent to a wideband PZT (Thorlabs, PA4CEW) via an amplifier (PiezoDrive, PDm200B for frequencies below 200 kHz or E&I 1020L above 200 kHz). For the ultrasonic OCE of hard materials, the PZT was in direct contact with the sample with a contact area of 2 mm×2 mm to generate sufficient push force. For the dynamic shear analysis of soft uniform materials, a custom mechanical actuator comprising of a PZT and a 3D-printed cylindrical tip with 0.6 mm radius was used. The tip was designed to facilitate the theoretical model of near-field surface wave[62]. For the depth-resolved stiffness mapping of the joint and skin, a 3D-printed prism-shape tip with a line contact of 2 mm is used to reduce wave attenuation and suppress supershear surface wave[33].

Displacement field analysis. All algorithms were developed in MATLAB (MathWorks Inc., MATLAB 2019). The raw data collected from the M-B scan was processed using the standard swept-source phase-stabilized algorithm to obtain the complex-valued OCT tomogram[60]. When fm>fA/2f_m > f_A/2, the correct vibrations occurred at two side lobes. We typically selected the left side lobe, and then applied the demodulated algorithm to extract the surface displacement profiles. Next, we performed a 1-dimensional Fourier transform to move the data from time tt domain to frequency ff domain. The frequency domain data was filtered at the driven frequency (when fm<fA/2f_m < f_A/2) or the aliased frequency (when fm>fA/2f_m > f_A/2) to obtain lower noise waveforms. After we obtained the displacement profiles over the xx coordinate, we performed another 1-dimensional Fourier transform to move the data from the spatial xx domain to the wavenumber kk domain. The wavenumber kk of the surface wave was then determined from the plot by selecting the peak corresponding to the Rayleigh surface wave. This filtering in the kk-xx domain is critical to remove other higher-order modes especially at high frequencies. The phase velocity is then given by ν=2π/k\nu= 2\pi/k. To obtain the cross-sectional displacement image, the same method used for obtaining the surface wave displacement was applied throughout the entire pixels along the depth zz. In addition, the motion artefact caused by the surface wave motion were corrected within the sample[63].

Measure complex wave number from surface waves. For soft materials, the near-field displacement is primarily dominated by the Rayleigh and supershear surface waves in high frequency regime. An analytical approximation solution for the surface waves excited by a cylindrical probe can be obtained. The surface displacement at lateral coordinate xx is[62]

u(x)=iπap0K4ρωm2[J1(KRa)KRF′(KR)H0(1)(−KRx)+J1(KSSa)KSSF′(KSS)H0(1)(−KSSx)],(23)u(x) = i\pi\frac{a p_0 K^4}{\rho\omega_m^2} \left[\frac{J_1(K_R a)K_R}{\mathcal{F}'(K_R)}H_0^{(1)}(-K_R x) + \frac{J_1(K_{SS}a)K_{SS}}{\mathcal{F}'(K_{SS})}H_0^{(1)}(-K_{SS}x)\right], \tag*{(23)}

where J1J_1 is the Bessel function of the first kind of order 1, H0(1)H_0^{(1)} is the Hankel function of the first kind, aa is the radius of the cylindrical probe, p0p_0 is the pressure amplitude exerted to the sample by harmonic vibrations of the probe. ωm\omega_m is the vibration frequency. KK, KRK_R and KSSK_{SS} are wavenumbers of the shear, Rayleigh surface, and supershear surface waves, respectively. KRK_R and KSSK_{SS} are roots of the secular equation F(κ)=(2κ2−K2)2−4κ2κ2(κ2−K2)sign⁡{Re⁡(κ2−K2)}=0\mathcal{F}(\kappa) = (2\kappa^2 - K^2)^2 - 4\kappa^2\sqrt{\kappa^2(\kappa^2-K^2)}\operatorname{sign}\{\operatorname{Re}(\kappa^2-K^2)\} = 0. So, we have KR=1.047KK_R = 1.047K and KSS=(0.4696−0.1355i)KK_{SS} = (0.4696 - 0.1355i)K. F′(κ)=dF/dκ\mathcal{F}'(\kappa) = \mathrm{d}\mathcal{F}/\mathrm{d}\kappa. To derive KK from the experiments, we fit the real and imaginary displacements simultaneously with Eq. (23) using the least-squares method. Representative fitting results can be found in Supplementary Figure S7.

Finite element analysis. Finite element analysis (FEA) was performed with Abaqus 6.12 (Dassault Systèmes). We used plane strain models and performed time domain simulations. A localized harmonic surface pressure was applied to excite elastic waves, simulating the PZT actuator in the experiments. Other sides of the model were completely constrained. The geometry and total time of the model was scaled by the wavelength and stimulus frequency fmf_m, respectively. The time increment and total time were 0.1/fm0.1/f_m and 20/fm20/f_m, respectively. The width and height of the model were about 16 folds of the maximum wavelength to avoid reflection from the boundaries. The model was divided into uniform layers, each with different thickness and elastic parameters. To model a layer with a gradient property, we assigned a temperature-dependent elastic modulus to the layer and then prescribed the temperature field with an analytical distribution. The gradient mesh was adopted to reduce the computation costs. At the surface the mesh size was smaller than one tenth of the wavelength and at the bottom the mesh size was about half of the wavelength. The element type used in this study was 8-node biquadratic element (CPE8RH).

The convergence of the model was checked by making sure the results independent of the mesh size.

Preparation of hard materials. The acrylate plastic block (McMaster-Carr) presented in Figure 3a had a thickness of 5 mm and an area of 50×5050 \times50 mm. The polystyrene petri dish (Fisher Scientific) presented in Figure 3b had a bottom thickness of 1 mm and a diameter of 150 mm. The borosilicate glass coverslip (Fisher Scientific) measured in Supplementary Fig. S5a had a thickness of 0.15 mm and an area of 24×5024 \times50 mm. The copper foil (All Foils Inc.) measured in Supplementary Fig. S5b had a thickness of 0.3 mm and an area of 15×4015 \times40 mm. The glass coverslip and the copper foil were placed on a lens holder so that the sample was bounded by the air on two sides.

Preparation of soft materials. The bulk silicone rubber has a diameter and height of 60 mm and 12 mm, respectively. It was prepared from Ecoflex 0050 material (Smooth-On Inc) by mixing the Ecoflex 1A and 1B at 1:1 ratio by weight. The mixture was poured into a mold and cured at room temperature overnight. The material was then post-cured in an oven at 80°C for 2 hours. The bulk PDMS sample has a diameter and height of 120 mm and 50 mm, respectively. It was prepared by using a 2:1 mixing ratio of base elastomer and curing agent (Sylgard 184, Dow Corning) and cured for 45 mins at 85 °C. The hydrogel sheet had a size of 40×4040 \times40 mm and a thickness of 1.3 mm under fully swollen, highly entangled condition[41]. All soft materials were assumed to have a mass density of ρ≃1000\rho\simeq1000 kg/m3^{3}.

Cartilage tissues. Two fresh bovine tibia bones from two juvenile calves were obtained 1 h post-mortem (Research 87 Inc., Boylston, MA). The tissues were wrapped in a wet towel to keep them well-hydrated until use. Three measurements were performed on the cortical surface of the bone. Four measurements were performed on the femoral condyle region covered by articular cartilage. To measure the thickness of calcified cartilage layer, we made a cross-sectional cut after the OCE experiments.

Human subject. We measured the skin of the left index finger on a 31-year-old male subject. The study was conducted at the Massachusetts General Hospital (MGH) following approval from the Institutional Review Board (IRB) of Massachusetts General Hospital and the Mass General Brigham Human Research Office. Written informed consent was obtained from both subjects prior to the measurement and all measurements were performed in accordance with the principles of the Declaration of Helsinki. The measurement site was marked by a surgical grade skin marker. For the hydration test, the whole index finger was immersed in warm water for 20 min, wiped with a clean towel, and then measured immediately.

Acknowledgements

This study was supported by funding from the National Institutes of Health via grants R01-EY027653, R01-EY033356, and R01-HL098028. The authors thank Dr. Guoguo Zhang for providing the hydrogel sample.

Author contributions

S.H.Y. conceived the idea. G.Y.L. and S.H.Y. developed the theory and signal processing. X.F. and G.Y.L. carried out the experiments and analyzed the results. All authors wrote and reviewed the manuscript.

Figures

Supplementary Materials for Ultra-wideband optical coherence elastography from acoustic to ultrasonic frequencies

Xu Feng; Guo-Yang Li; Seok-Hyun Yun

¹ Co-first authors with equal contribution

Image not extracted

Figure S1. Schematic of the optical imaging system and signal generation and acquisition. The output spectrum of the laser is measured by an optical spectral analyzer (AQ6370C, Yokogawa Electric). Abbreviations used: Mach-Zehnder interferometer (MZI), photodiode (PD), balanced photodiode (BPD), data acquisition (DAQ) system, Input/output (I/O) board, fiber Bragg grating (FBG),circulators (C1, C2), galvanometer mirrors (GM), fiber collimator (FC), objective lens (OBJ), reference mirror (RM), dichroic mirror (DM), piezoelectric transducer (PZT).

Gaussian distribution of the time jitter in sweep period $T$

Figure S3. Gaussian distribution of the time jitter in sweep period TT. Data represents 900 M-scans, with each M-scan consisting of 108 A-lines. By evaluating the phase difference between two consecutive A-lines, we obtained the time jitter over 900×107=96,300900 \times107 = 96{,}300 data points. The distribution is fitted with a Gaussian curve (dashed line). The standard deviation of the time jitter is 4.9 ns.

Vibration noise introduced by laser time jitter

According to Eq. (17) in the main text, the noise for the vibration phase primarily depends on the optical phase noise (e−i2k0z0e^{-i2k_{0}z_{0}}) and the laser timing jitter (e−imωmTe^{-im\omega_{m}T}). The latter one comes from the mechanical instability the sweep-source laser, which results in a timing jitter in the FBG clock, TT. For our system, we get the standard deviation of TT, denoted by σT≈4.9 ns\sigma_{T} \approx4.9\ \mathrm{ns}.

Let’s consider a measurement to a vibrating scatter using NN continuous A scan (M scan). The standard deviation of the vibration phase noise, denoted by σφ\sigma_{\varphi}, is

σφ=(σX′)2+(σL)2,(24)\sigma_{\varphi}=\sqrt{(\sigma_{X'})^{2}+(\sigma_{L})^{2}}, \tag*{(24)}

where σL\sigma_{L} denotes the phase noise introduced by laser timing jitter and σX′\sigma_{X'} is the optical phase noise. σX′\sigma_{X'} is determined by the SNR of the sidelobe (denoted by X′X'),

σX′=1N1X′.(25)\sigma_{X'}=\frac{1}{\sqrt{N}}\frac{1}{\sqrt{X'}}. \tag*{(25)}

According to Eq. (16) in the main text,

10log⁡10(X′)=10log⁡10(X)+20log⁡10(k0δ),(26)10\log_{10}(X')=10\log_{10}(X)+20\log_{10}(k_{0}\delta), \tag*{(26)}

where XX is the SNR of the static OCT signal (main lobe).

To derive the expression for σL\sigma_{L}, we denote the time interval of mm-th A scan with (T+tm)(T+t_{m}) (m=1,2,…,Nm=1,2,\ldots,N). Therefore, the mean and standard deviation of tmt_{m} are 0 and σT\sigma_{T}, respectively. Then the time of the mm-th A scan, denoted by TmT_{m}, is Tm=∑i=1m(T+ti)=mT+∑i=1mtiT_{m}=\sum_{i=1}^{m}(T+t_{i})=mT+\sum_{i=1}^{m}t_{i}. The vibration signal, according to Eq. (17) in the main text, is Fm∼eiωmTmF_{m}\sim e^{i\omega_{m}T_{m}}. We perform Fourier transform to this signal to get

F(ω)∼∑m=1NeiωmTme−ωmT=∑m=1Neim(ωm−ω)Teiωm(∑i=1mti).(27)\mathcal{F}(\omega)\sim\sum_{m=1}^{N}e^{i\omega_{m}T_{m}}e^{- \omega mT}=\sum_{m=1}^{N}e^{im(\omega_{m}-\omega)T}e^{i\omega_{m}(\sum_{i=1}^{m}t_{i})}. \tag*{(27)}

At ω=ωm\omega=\omega_{m}, we have

F(ωm)∼∑m=1Neiωm(∑i=1mti)≈∑m=1N[1+iωm(∑i=1mti)].(28)\mathcal{F}(\omega_{m})\sim\sum_{m=1}^{N}e^{i\omega_{m}(\sum_{i=1}^{m}t_{i})}\approx\sum_{m=1}^{N}\left[1+i\omega_{m}\left(\sum_{i=1}^{m}t_{i}\right)\right]. \tag*{(28)}

The phase of F(ωm)\mathcal{F}(\omega_{m}), denoted by arg⁡{F(ωm)}\arg\{\mathcal{F}(\omega_{m})\}, is

arg⁡{F(ωm)}≈ωm∑m=1N(∑i=1mti)N=ωmN[Nt1+(N−1)t2+⋯+tN].(29)\arg\{\mathcal{F}(\omega_{m})\}\approx\frac{\omega_{m}\sum_{m=1}^{N}(\sum_{i=1}^{m}t_{i})}{N}=\frac{\omega_{m}}{N}[Nt_{1}+(N-1)t_{2}+\cdots+t_{N}]. \tag*{(29)}

Then σL\sigma_{L} is the standard deviation of arg⁡{F(ωm)}\arg\{\mathcal{F}(\omega_{m})\},

σL≈ωm(NN)2(σt1)2+(N−1N)2(σt2)2+⋯+(1N)2(σtN)2.(30)\sigma_{L}\approx\omega_{m}\sqrt{\left(\frac{N}{N}\right)^{2}(\sigma_{t_{1}})^{2}+\left(\frac{N-1}{N}\right)^{2}(\sigma_{t_{2}})^{2}+\cdots+\left(\frac{1}{N}\right)^{2}(\sigma_{t_{N}})^{2}}. \tag*{(30)}

where σtm=σT\sigma_{t_{m}}=\sigma_{T} is the standard deviation of tmt_{m} (m=1,2,…,Nm=1,2,\ldots,N). Therefore, we get

σL≈ωmσTNN(N+1)(2N+1)6≈N3(ωmσT).(31)\sigma_{L}\approx\frac{\omega_{m}\sigma_{T}}{N}\sqrt{\frac{N(N+1)(2N+1)}{6}}\approx\sqrt{\frac{N}{3}}(\omega_{m}\sigma_{T}). \tag*{(31)}
Ultrasonic OCE of the glass and metal

Figure S4. Ultrasonic OCE of the glass and metal. a, Dispersion relation of the coverslip glass (dot) and theoretical fitting with a Lamb wave model (line). b, Dispersion relation of the copper sheet (dot) and the theoretical fitting with a Lamb wave model (line).

Wavenumber domain plots for Rubber, PDMS, and Hydrogel

Figure S5. Wavenumber kk domain plots obtained by the Fourier Transform. Comparison is made at 40 kHz. The Rayleigh surface wave mode (R) and the supershear surface wave (SS) mode are labeled on the plots.

Rheological fitting of the soft materials

Figure S6. Rheological fitting of the soft materials. a, Rubber (20 kHz). b, PDMS (40 kHz). c, Hydrogel (40 kHz). A comparison is made between the experimental data and the theoretical model for the surface displacement. Markers, experiment; lines, theory. Inset: the total wave amplitude.

Cartilage OCE cross-sectional displacement images at 40–140 kHz, before and after applying the demodulation algorithm for inner scatterers

Figure S7. Cartilage OCE. Cross-sectional displacement images at 40 – 140 kHz for a, Without applying the demodulation algorithm for inner scatters, and b, After applying the demodulation algorithm for inner scatterers.

Figure S8. FEA model for the skin. The depth dependent profile of shear modulus in the human skin for the normal, pre-hydrated state (margenta) and hydrated state (turquoise).

An example of signal demodulation

Figure S9. An example of signal demodulation. a, The Fourier components at F(ω0−ωm)F(\omega_{0}-\omega_{m}) contain three signals: (1) The time-independent reflection signal (carrier signal) from a scatterer at z0−zmz_{0}-z_{m} (top); (2) The vibration signal (negative sideband) originated from the scatterer at ω0\omega_{0} (middle); and (3) The vibration signal (positive sideband) originated from the scatterer located at z0−2zmz_{0}-2z_{m} (bottom). b, Steps of signal demodulation. (1) Original signal F(ω0−ωm)F(\omega_{0}-\omega_{m}). (2) Subtracting the original signal from the background ⟨F(ω0−ωm)⟩\langle F(\omega_{0}-\omega_{m})\rangle. (3) Normalizing the signal by ⟨F(ω0−ωm)⟩\langle F(\omega_{0}-\omega_{m})\rangle. (4) Fourier transform to identify δ(z0)\delta(z_{0}) and δ(z0−2ωm)\delta(z_{0}-2\omega_{m}).

References

  1. [1]Nia, H.T., Munn, L.L. & Jain, R.K. Physical traits of cancer. Science 370, eaaz0868 (2020).DOI
  2. [2]Guimarães, C.F., Gasperini, L., Marques, A.P. & Reis, R.L. The stiffness of living tissues and its implications for tissue engineering. Nature Reviews Materials 5, 351-370 (2020).DOI
  3. [3]Cross, S.E., Jin, Y.-S., Rao, J. & Gimzewski, J.K. Nanomechanical analysis of cells from cancer patients. in Nano-Enabled Medical Applications 547-566 (Jenny Stanford Publishing, 2020).
  4. [4]Pelleg, J. Mechanical properties of materials, (Springer Science & Business Media, 2012).
  5. [5]Sharpe, W., Yuan, B. & Edwards, R. A new technique for measuring the mechanical properties of thin films. Journal of Microelectromechanical systems 6, 193-199 (1997).DOI
  6. [6]Menard, K.P. & Menard, N.R. Dynamic mechanical analysis, (CRC press, 2020).
  7. [7]Krieg, M., et al. Atomic force microscopy-based mechanobiology. Nature Reviews Physics 1, 41-57 (2019).
  8. [8]Ando, T., et al. A high-speed atomic force microscope for studying biological macromolecules. Proceedings of the National Academy of Sciences 98, 12468-12472 (2001).pubmed.ncbi.nlm.nih.gov/11592975
  9. [9]Rigato, A., Miyagi, A., Scheuring, S. & Rico, F. High-frequency microrheology reveals cytoskeleton dynamics in living cells. Nature physics 13, 771-775 (2017).DOI
  10. [10]Yamada, S., Wirtz, D. & Kuo, S.C. Mechanics of living cells measured by laser tracking microrheology. Biophysical journal 78, 1736-1747 (2000).pubmed.ncbi.nlm.nih.gov/10733956
  11. [11]Kennedy, B.F., Wijesinghe, P. & Sampson, D.D. The emergence of optical elastography in biomedicine. Nat Photonics 11, 215 (2017).DOI
  12. [12]Itoh, A., et al. Breast disease: clinical application of US elastography for diagnosis. Radiology 239, 341-350 (2006).pubmed.ncbi.nlm.nih.gov/16484352
  13. [13]Chang, J.M., et al. Clinical application of shear wave elastography (SWE) in the diagnosis of benign and malignant breast diseases. Breast cancer research and treatment 129, 89-97 (2011).DOI
  14. [14]Murphy, M.C., et al. Measuring the characteristic topography of brain stiffness with magnetic resonance elastography. PloS one 8, e81668 (2013).DOI
  15. [15]Li, G.-Y. & Cao, Y. Mechanics of ultrasound elastography. Proceedings of the Royal Society A: Mathematical, Physical and Engineering Sciences 473, 20160841 (2017).DOI
  16. [16]Sigrist, R.M., Liau, J., El Kaffas, A., Chammas, M.C. & Willmann, J.K. Ultrasound elastography: review of techniques and clinical applications. Theranostics 7, 1303 (2017).DOI
  17. [17]Cartagena-Rivera, A.X., Wang, W.-H., Geahlen, R.L. & Raman, A. Fast, multi-frequency and quantitative nanomechanical mapping of live cells using the atomic force microscope. Scientific reports 5, 1-11 (2015).
  18. [18]Ramier, A., et al. In vivo measurement of shear modulus of the human cornea using optical coherence elastography. Scientific reports 10, 1-10 (2020).DOI
  19. [19]Larin, K.V. & Sampson, D.D. Optical coherence elastography–OCT at work in tissue biomechanics. Biomedical optics express 8, 1172-1202 (2017).DOI
  20. [20]Li, R. Time-temperature superposition method for glass transition temperature of plastic materials. Materials Science and Engineering: A 278, 36-45 (2000).DOI
  21. [21]Su, Z., Ye, L. & Lu, Y. Guided Lamb waves for identification of damage in composite structures: A review. Journal of sound and vibration 295, 753-780 (2006).DOI
  22. [22]Blitz, J. & Simpson, G. Ultrasonic methods of non-destructive testing, (Springer Science & Business Media, 1995).
  23. [23]Kirby, M.A., et al. Spatial resolution in dynamic optical coherence elastography. J Biomed Opt 24, 096006 (2019).DOI
  24. [24]Xu, H. & Luo, Z.-P. Assessment of viscoelasticity of ex vivo bovine cartilage using Rayleigh wave method in the near-source and far-field region. Journal of Biomechanics 116, 110252 (2021).DOI
  25. [25]Whittemore Jr, K.R., Merchant, S.N., Poon, B.B. & Rosowski, J.J. A normative study of tympanic membrane motion in humans using a laser Doppler vibrometer (LDV). Hearing research 187, 85-104 (2004).pubmed.ncbi.nlm.nih.gov/14698090
  26. [26]Moeendarbary, E., et al. The soft mechanical signature of glial scars in the central nervous system. Nature communications 8, 1-11 (2017).DOI
  27. [27]Zvietcovich, F. & Larin, K.V. Wave-based optical coherence elastography: the 10-year perspective. Progress in Biomedical Engineering 4, 012007 (2022).DOI
  28. [28]Schmitt, J.M. OCT elastography: imaging microscopic deformation and strain of tissue. Opt Express 3, 199-211 (1998).pubmed.ncbi.nlm.nih.gov/19384362
  29. [29]Ramier, A., Tavakol, B. & Yun, S.-H. Measuring mechanical wave speed, dispersion, and viscoelastic modulus of the cornea using optical coherence elastography. Opt Express 27, 16635-16649 (2019).DOI
  30. [30]Vinas-Pena, M., Feng, X., Li, G.-Y. & Yun, S.-H. In situ measurement of the stiffness increase in the posterior sclera after UV-riboflavin crosslinking by optical coherence elastography. Biomedical Optics Express 13, 5434-5446 (2022).
  31. [31]Foo, K.Y., et al. Optical palpation for tumor margin assessment in breast-conserving surgery. Biomedical Optics Express 12, 1666-1668 (2021).DOI
  32. [32]Zvietcovich, F., et al. Longitudinal shear waves for elastic characterization of tissues in optical coherence elastography. Biomedical optics express 10, 3699-3718 (2019).
  33. [33]Feng, X., Li, G.-Y., Ramier, A., Eltony, A.M. & Yun, S.-H. In vivo stiffness measurement of epidermis, dermis, and hypodermis using broadband Rayleigh-wave optical coherence elastography. Acta Biomaterialia (2022).arxiv.org/abs/2201.12258
  34. [34]Yun, S., Tearney, G., De Boer, J. & Bouma, B. Motion artifacts in optical coherence tomography with frequency-domain ranging. Opt Express 12, 2977-2998 (2004).pubmed.ncbi.nlm.nih.gov/19483816
  35. [35]Choma, M.A., Ellerbee, A.K., Yang, C., Creazzo, T.L. & Izatt, J.A. Spectral-domain phase microscopy. Optics Letters 30, 1162-1164 (2005).pubmed.ncbi.nlm.nih.gov/15945141
  36. [36]MatWeb, L. Material property data. MatWeb [Online]. Available: http://www.matweb.com (2022).DOI
  37. [37]Achenbach, J. Wave propagation in elastic solids, (Elsevier, 2012).
  38. [38]Miyake, K., Satomi, N. & Sasaki, S. Elastic modulus of polystyrene film from near surface to bulk measured by nanoindentation using atomic force microscopy. Appl Phys Lett 89, 031925 (2006).DOI
  39. [39]Basavalingappa, A., Shen, M.Y. & Lloyd, J.R. Modeling the copper microstructure and elastic anisotropy and studying its impact on reliability in nanoscale interconnects. Mechanics of Advanced Materials and Modern Processes 3, 1-10 (2017).DOI
  40. [40]Bouras, N., Madjoubi, M., Kolli, M., Benterki, S. & Hamidouche, M. Thermal and mechanical characterization of borosilicate glass. Physics Procedia 2, 1135-1140 (2009).DOI
  41. [41]Kim, J., Zhang, G., Shi, M. & Suo, Z. Fracture, fatigue, and friction of polymers in which entanglements greatly outnumber cross-links. Science 374, 212-216 (2021).DOI
  42. [42]Pitre Jr, J.J., et al. Super-shear evanescent waves for non-contact elastography of soft tissues. Appl Phys Lett 115, 083701 (2019).DOI
  43. [43]Li, G.-Y., Feng, X., Ramier, A. & Yun, S.-H. Supershear surface waves reveal prestress and anisotropy of soft materials. Journal of the Mechanics and Physics of Solids, 105085 (2022).DOI
  44. [44]Yasar, T.K., Royston, T.J. & Magin, R.L. Wideband MR elastography for viscoelasticity model identification. Magnetic resonance in medicine 70, 479-489 (2013).DOI
  45. [45]Rubino, E. & Ioppolo, T. Young's modulus and loss tangent measurement of polydimethylsiloxane using an optical lever. Journal of Polymer Science Part B: Polymer Physics 54, 747-751 (2016).DOI
  46. [46]Hassan, S. & Kim, J. Polyacrylamide hydrogels. IV. Near-perfect elasticity and rate-dependent toughness. Journal of the Mechanics and Physics of Solids 158, 104675 (2022).DOI
  47. [47]Park, A. Rheology of stratum corneum-II: A physico-chemical investigation of factors influencing the water content of the corneum. J. Soc. Cosmet. 23, 13-21 (1972).
  48. [48]Darvin, M.E., Schleusener, J., Lademann, J. & Choe, C.-S. Current Views on Noninvasive in vivo Determination of Physiological Parameters of the Stratum Corneum Using Confocal Raman Microspectroscopy. Skin Pharmacology and Physiology 35, 125-136 (2022).DOI
  49. [49]Bilston, L.E. Soft tissue rheology and its implications for elastography: Challenges and opportunities. NMR in Biomedicine 31, e3832 (2018).DOI
  50. [50]Parker, K., Szabo, T. & Holm, S. Towards a consensus on rheological models for elastography in soft tissues. Physics in Medicine & Biology 64, 215012 (2019).DOI
  51. [51]Rossetti, L., et al. The microstructure and micromechanics of the tendon–bone insertion. Nature materials 16, 664-670 (2017).DOI
  52. [52]Zhao, H., et al. Multiscale engineered artificial tooth enamel. Science 375, 551-556 (2022).DOI
  53. [53]Moutos, F.T., Freed, L.E. & Guilak, F. A biomimetic three-dimensional woven composite scaffold for functional tissue engineering of cartilage. Nature materials 6, 162-167 (2007).pubmed.ncbi.nlm.nih.gov/17237789
  54. [54]Kirby, M.A., et al. Optical coherence elastography in ophthalmology. J Biomed Opt 22, 121720 (2017).DOI
  55. [55]Sun, Q., et al. Elastin imaging enables noninvasive staging and treatment monitoring of kidney fibrosis. Science translational medicine 11, eaat4865 (2019).DOI
  56. [56]Fuhs, T., et al. Rigid tumours contain soft cancer cells. Nature Physics, 1-10 (2022).DOI
  57. [57]Patz, S., et al. Imaging localized neuronal activity at fast time scales through biomechanics. Science advances 5, eaav3816 (2019).DOI
  58. [58]Li, G.-Y., Gower, A.L., Destrade, M. & Yun, S.-H. Non-destructive mapping of stress, strain and stiffness of thin elastically deformed materials. Communications Physics 5, 1-7 (2022).arxiv.org/abs/2201.01847
  59. [59]Raissi, M., Perdikaris, P. & Karniadakis, G.E. Physics-informed neural networks: A deep learning framework for solving forward and inverse problems involving nonlinear partial differential equations. Journal of Computational physics 378, 686-707 (2019).DOI
  60. [60]Ramier, A., Cheng, J.T., Ravicz, M.E., Rosowski, J.J. & Yun, S.-H. Mapping the phase and amplitude of ossicular chain motion using sound-synchronous optical coherence vibrography. Biomedical optics express 9, 5489-5502 (2018).DOI
  61. [61]Yun, S.-H., Tearney, G.J., de Boer, J.F., Iftimia, N. & Bouma, B.E. High-speed optical frequency-domain imaging. Opt Express 11, 2953-2963 (2003).pubmed.ncbi.nlm.nih.gov/19471415
  62. [62]Li, G.-Y., Feng, X., Ramier, A. & Yun, S.-H. Supershear surface waves reveal prestress and anisotropy of soft materials. arXiv preprint arXiv:2107.11482 (2021).DOI
  63. [63]Song, S., Huang, Z. & Wang, R.K. Tracking mechanical wave propagation within tissue using phase-sensitive optical coherence tomography: motion artifact and its compensation. Journal of biomedical optics 18, 121505 (2013).DOI

Paper details

vlnc:W_zhap3kua
Version
v1 · latest
Published
Fresh — added to the Hub within the last 30 days
Added
License
CC BY 4.0
Usage

views · 0

downloads · 0

A view counts once per browser per day, a download each time the PDF is requested.

  • Physics and Astronomy

Not recorded for this paper

View statement
Looking for related papers…
Contents