Finesse evolution in a bichromatic cavity under blue and red resonant illumination in vacuum
Abstract
The long-term performance of high-finesse optical cavities depends on the stability of their mirror coatings under vacuum and under continuous exposure to high circulating power. In cavities supporting multiple wavelengths, a central question is whether degradation at one wavelength compromises performance at another. Understanding this interplay is essential for those cavity quantum electrodynamics experiments that need diverse optical fields for cavity stabilization, trapping, and light–matter coupling. Here, we investigate wavelength-dependent degradation and recovery of a bichromatic cavity with high-reflectivity coating at 455 nm and 671 nm in vacuum. Cavity ring-down measurements reveal a decrease in the 455 nm finesse from 7.710^3 to 1.910^3 after 11 hours of continuous exposure to high circulating power. Simultaneous measurements show no measurable degradation at 671 nm. Continuous operation at 671 nm, instead, produces no measurable degradation at either wavelength under the tested conditions. Comparison of oxygen depletion and Avrami-type loss models indicates kinetics compatible with nucleation and growth, without uniquely identifying the microscopic mechanism. We demonstrate partial recovery of the finesse with illumination at 455 nm in air. Our results indicate a wavelength dependence of laser-induced optical damage in bichromatic cavities and provide an estimate of the blue light intensity that can be safely used in such cavities.
Introduction
High-finesse optical cavities placed under vacuum are critical tools in atomic molecular and optical (AMO) physics. Thanks to their high finesse and narrow cavity linewidth, they can provide the strong light-matter coupling required to interface the narrow optical transitions of single trapped ions, neutral atoms, or solid-state emitters with photons [1]. As such, they play a key role in fundamental physics and metrology, not only as a precise frequency reference [2] but also as tool for metrological enhancements via squeezing in optical atomic clocks [3, 4] or as a mean to induce long interactions among particles for quantum simulations [5, 6]. In quantum technology applications, instead, such systems serve as key building blocks for efficient entangled single-photon sources [7, 8] and nodes in optical quantum networks [9, 10, 11]. In these systems, bichromatic and multi-wavelength optical cavities - cavities that have a high reflectivity for more than one frequency range - are becoming more and more common with the increasing complexity of AMO experiments and quantum technologies. They are used for various purposes, from providing narrow ULE frequency references of several laser sources [12], to decoupling the stabilization of the cavity from the coupling of atoms/ions to the relevant cavity field [13, 14].
In all these applications, placing high-reflectivity cavity mirrors inside high-vacuum environments is necessary to mitigate spurious thermal, acoustic, and pressure fluctuations which degrade the cavity finesse and the cavity stability and, in turn, the precision and accuracy of the cavity frequency reference. Despite this clear advantage of vacuum environment, the long-term performance of high-finesse optical cavities under vacuum is frequently degraded. In fact, thin-film dielectric coatings — typically consisting of alternating layers of silica () and tantalum pentoxide () — have been observed to undergo significant degradation over time when placed under vacuum [15, 16, 17] and also when illuminated with laser light [18, 19, 20, 21]. This is especially common for cavities with high-reflectivity in the ultraviolet (UV) and near UV (NUV) where degradation is usually more severe. Two main causes of degradation are reported in literature. The first concerns vacuum-induced oxygen depletion in the surface layer, identified as a common mechanism driving optical loss in both the ultraviolet and visible/near infrared (NIR) regimes [16, 17, 22, 23]. This effect is purely vacuum related and can be affected by temperature [16, 22]. In all the works the degradation is reversible with oxygen exposure and can be strongly reduced with the choice of the topmost coating layer [24]. The second cause is hydrocarbon contaminants that photochemically react under intense illumination and alter the micro-chemical environment of the coating [24, 25, 26, 27]. This relates to the combination of vacuum environment with a high intracavity intensity which circulates in the resonator for an extended period of time. The degradation caused by hydrocarbons and its recovery with oxygen has been observed [24, 26, 27]. Nevertheless, just a few studies report on the finesse degradation after cavity illumination for extended periods of time [27], mostly because the degradation is extremely fast in the deep UV regime [26, 28], and this makes it hard to compare the degradation with the hydrocarbon nucleation model. Some studies have already shown the possibility of preventing degradation with clean room facilities [29], with micro-chemical environment control of the topmost coating layer [26] and or with initial UV conditioning in an oxygen environment [24].
Although cavity degradation has been documented across several individual spectral regions, systematically testing multi-wavelength cavities, and especially those designed to maintain simultaneously high finesse for a near ultraviolet radiation or blue alongside a secondary visible or near-infrared band, remains an unaddressed challenge. Consequently, it remains unclear whether vacuum and/or light-induced damage localized on the high-reflectivity coating layer translates into optical losses at the secondary wavelength for which the dielectric stack has also high reflectivity and, viceversa, if high power at the VIS/NIR affects the finesse in the shorter wavelength range. Resolving this issue is increasingly vital for novel cutting-edge quantum systems.
Here, we investigate the degradation of a bichromatic high finesse Fabry-Pérot cavity placed under vacuum that has a high reflectivity at 455 nm and 671 nm. The cavity has been developed to address Ba ions and Li atoms optical transitions [30, 31]. We observe the long-term effect of a high intracavity circulating power either at 455 nm or 671 nm on the finesse measured at both 455 nm and 671 nm via cavity ring-down. In this way, we test whether light-induced changes in optical losses affect both high reflectivity bands and find what mechanism governs the degradation of such optical cavity.
Methods
Finesse detection and experimental setup
When the reflectivity of the cavity mirrors is sufficiently high, the cavity round-trip time is much smaller than the decay time of the cavity , with being the cavity length and the speed of light. In this scenario, cavity ring-down measurements provide a direct measure of the total optical losses of a cavity [32, 33]. We monitor the decay of the light stored in the cavity by using a photodiode in transmission after the input light is suddenly turned off. The resulting exponential decay of the cavity transmission signal , as detected by the photodiode, is described by
where is the offset of the transmission signal on the photodiode, is the initial transmission value, is the time the decay begins. For two identical high reflectivity mirrors, the cavity lifetime is linked to the reflectivity of the cavity mirrors according to
The lifetime obtained from a fit to Eq. (1) therefore allows one to extract the finesse of the cavity through
In our experiment we use a linear high finesse optical cavity with a high reflectivity coating at 455 nm and 671 nm. It is composed of two superpolished curved mirrors with radii of curvature 50 mm with a custom high reflectivity coating (FiveNine Optics) which is composed of layers of SiO and TaO, last one being TaO. The nominal losses are 180 ppm for 455 nm and 671 nm wavelengths. The cavity has a nominal mirror separation of mm and it is operated in vacuum. For the cavity used in this work, the expected cavity lifetime is on the order of a few microseconds. The cavity is mounted inside a UHV chamber, a turbo pump is used to lower the chamber pressure initially down to about mbar, which then slowly decreases down to mbar over the course of the days. The pressure is monitered by using a gauge.
The cavity finesse evolution is tested with two lasers, a home-made external cavity diode laser operating at 455 nm, and a Toptica external cavity diode laser (DL pro) at 671 nm. The optical setup used for coupling and locking to the optical cavity is illustrated in Figure 1. The 455 nm and 671 nm beams are combined on a short-pass dichroic mirror (DM) and co-propagate in the optical cavity. Both wavelengths setups include an acousto-optic modulator (AOM), and an electro-optic modulator (EOM) for fast switching and error-signal generation. Mode-matching lenses and incoupling lenses are also shown in figure and allowed us to couple about 35% of the 455 nm and 671 nm light into the fundamental transverse mode () before starting the degradation measurements. The mode was identified from the observed transmission via a Thorlabs CMOS camera (CS165MU1/M) and this resonance was then used for locking. Cavity-reflected light is routed to the photodetector PD2. The 455 nm or the 671 nm photodetection signal is then mixed with the RF reference driving the corresponding EOM in order to generate the error signal needed for locking the laser onto the cavity. The error signal is then fed to a Moku PID controller to lock the lasers to the selected cavity resonance. At the cavity output, instead, a flip-mirror assembly (FM) allows us to detect the light either on a transmission photodetector (PD1) for recording the cavity ring-down, or on the camera for monitoring the cavity mode. Thorlabs PDA10A2 photodetectors were used for optical detection.

Figure 1. Experimental setup: The 455 nm light is combined with the 671 nm on a dichroic mirror (DM). The combined beams are mode-matched to the Fabry–Pérot cavity using lenses , and and the dielectric mirrors (M1 and M2). Photodiodes PD1 and PD2 detect the transmitted and reflected cavity signals, respectively, along with a Thorlabs camera CS165MU1/M. A flip mirror (FM) directs the transmitted beam to either the camera or PD1. Acousto-optic modulators (AOM) are used for switching off and on the light coupled to the cavity in order to perform the ring-down measurements. Electro-optic modulators (EOMs) are used to generate the error signals.
In order to test the cavity degradation dynamics, we keep the lasers, either at 455 nm or 671 nm, locked to the cavity for an extended period of time so that the cavity is subjected to high circulating power. Each ring-down measurement is recorded after the laser has been locked to the cavity for a variable amount time. This is done by monitoring the cavity transmission on PD1 while scanning the laser frequency across the resonance peak. At resonance, the oscilloscope sends a trigger signal to the AOM driver that turns off the laser light for 500 s. The turn off time is 12.8 0.2 ns. This procedure is used for recording both the 455 nm and 671 nm cavity decay. Regarding the data analysis, an exponential decay function of the form of Eq. (1) is fitted to the ring-down data. The cavity lifetime is extracted from the fit, and the finesse is quantified via Eq. (3). The results are reported in detail in the section 3.
Physical Models of Finesse Degradation
To characterize the temporal evolution of the cavity finesse, we compare two models for the increase in optical loss: an exponential saturating model motivated by oxygen depletion [17, 22] and an Avrami model [34] motivated by nucleation-and-growth kinetics. These models provide two alternative descriptions of the observed dynamics, without assuming that fitting either functional form uniquely identifies the underlying microscopic degradation mechanism. We write the total round-trip power loss as
where is the total loss at the beginning of the measurement, including mirror transmission, absorption, and scattering, and is the time-dependent change in round-trip loss. Both quantities are dimensionless power fractions. In particular, . Using the high-finesse relation [35], and defining , we obtain the following evolution of the finesse normalized to its initial value
Gangloff et al. [22] modeled vacuum-induced degradation of tantalum-oxide-terminated mirrors through the formation of oxygen vacancies that increase optical absorption. At fixed temperature, their model assumes that the vacancy concentration approaches a limiting value via first-order kinetics, with the resulting optical loss directly proportional to this concentration. By expressing this model in terms of round-trip loss, one gets [22]
where the increment of losses is the additional round-trip loss in the limit of long exposure to vacuum as predicted by this model. The parameter is the characteristic depletion timescale. In their work, Gangloff et al. describe their model in terms of excess loss per mirror. For two mirrors undergoing the same loss increase, the corresponding increment of round-trip loss is twice the single-mirror increment, assuming that the mirror transmission and losses remain constant. By defining the dimensionless relative loss amplitude as
we obtain the following time evolution of the finesse due to oxygen vacancies
The parameters and describe the magnitude and timescale of the loss increase, respectively. The predicted steady-state finesse is .
As an alternative, we can describe the finesse degradation using an Avrami-type model, motivated by nucleation-and-growth kinetics [34]:
Here, is the characteristic timescale, and is the Avrami exponent, which controls the temporal shape of the loss increase. The dimensionless amplitude is the ratio of the additional round-trip losses in the long-time limit to the initial total losses. The equation follows from Eq. (5) by assuming that the additional loss term evolves according to the Avrami equation. In our data analysis, we fix the Avrami exponent at , a value consistent with continuous nucleation and two-dimensional, interface-controlled growth of defects on a mirror surface [34]. This is because, in this model, new defect regions appear at a constant rate with their dimensions increasing linearly with time. The area of an individual region, therefore, grows quadratically over time. To this contribution one has to add the continuous formation of new defect regions, giving an Avrami exponent of three. This provides a possible physical picture for the formation and lateral growth of loss-inducing regions on a mirror surface. In our experiments, we use both the oxygen depletion (Eq. (8)) and the Avrami/Nucleation (Eq. (9)) models to fit the data of the finesse degradation.
Results
We performed a series of experiments to investigate how prolonged resonant illumination affects the cavity finesse at 455 nm and 671 nm. These measurements address both the wavelength dependence of degradation in vacuum and the recovery of the finesse under illumination in air. We considered three experimental conditions:
Resonant illumination at 455 nm in vacuum, to characterize degradation at 455 nm and its possible impact on the finesse at 671 nm.
Resonant illumination at 671 nm in vacuum, to assess the stability of the finesse at both wavelengths.
Resonant illumination at 455 nm in air, to investigate the recovery of the blue finesse while monitoring any changes in the finesse at 671 nm.
Finesse degradation at 455 nm
First, we tested the cavity finesse degradation at 455 nm by locking this laser to the optical cavity for an extended period of time. The input power is set to mW, whereas the output power measured in transmission is on average μW. Initial finesse is . We observed a 75% decrease in finesse, which degraded down to a final value of after 660 mins (11 hours) of illumination, as shown in Figure 2. The finesse data is normalized with respect to the maximum finesse value, , that in this case is . To analyze the finesse evolution we applied two degradation models, a nucleation/Avrami model, and a oxygen depletion model, both introduced in the previous section. As Figure 2 shows, the nucleation/Avrami model reproduces the overall degradation trend, whereas the oxygen-depletion model fails to capture the observed decay. Moreover, during the time where the laser is not locked to the cavity (no circulating power), we observe no decrease in the finesse, see inset in figure. Our findings are suggestive of hydrocarbon contamination as cause for cavity degradation.

Figure 2. Cavity finesse degradation at 455 nm. Evolution of the cavity finesse while locked at 455 nm under vacuum, with an output power of μW. The finesse is referenced to the initial measured value . Each data point corresponds to an average of 5 ring-down measurements. The error bars indicate one standard deviation error of the mean of the respective averaged measurements. Each ring-down measurement at 455 nm was fitted from the turn-off time following Eq. (1). The progressive decrease in finesse is fitted to the nucleation/Avrami model in Eq. (9) (purple curve) and to the oxygen-depletion model in Eq. (8) (orange curve) are also plotted The inset shows the same data as a function of elapsed time, where the time intervals in which the laser was locked to the cavity (green) or not locked to the cavity (gray) are highlighted. The cavity remained in vacuum at all times.
Effect of the finesse degradation at 455 nm on the finesse at 671 nm
After partial recovery of the cavity finesse with 455 nm light illumination in air, see § 3.4, we run a second experiment in order to test the effect of the degradation caused by 455 nm illumination on the finesse measured at 671 nm. We lock again the 455 nm laser on the cavity and observe the evolution of the finesse at both wavelengths. The average input power is mW and mW, with corresponding average output power at 455 nm of W. The initial 455 nm finesse value is , and this value is used to normalize the 455 nm finesse data. The evolution of the fractional change in finesse at both wavelengths is shown in Figure 3. As in the previous run, we observe a degradation of the finesse at 455 nm to the value over 558 mins (9.3 hours) of locking time. The finesse at 671 nm, however, does not show degradation as the finesse data fluctuate around a mean value of , used to normalize the data. The data of 455 nm degradation are fitted with the nucleation and oxygen depletion models. In this set of data both models qualitatively agree with the trend, nucleation model still being the one with lowest variance and therefore best fitting. After data acquisition, we found that ring-down curves were convolved with the instrumentation time-response due to an internal low-pass filter in the oscilloscope. To account for this convolution, we fitted the ring-down data over two normalized amplitude windows centered around 50% of the decay: from 0.66 to 0.34 and from 0.815 to 0.185. The finesse reported in Figure 3 is the mean of the two estimates, and its uncertainty is the mean of their uncertainties.

Figure 3. Finesse evolution at 455 nm and 671 nm while locking at 455 nm under vacuum. 455 nm output power is . Finesse values at 455 nm (blue data) were normalized with respect to the initial finesse , while finesse values at 671 nm (red data) were normalized with respect to average value . The first 3 data points represent the average of 5 measurements, while the last three correspond to an average of 10 measurements. The error bars indicate one standard deviation. While the finesse at 455 nm progressively decreases, the finesse at 671 nm remains unchanged throughout the measurement. The finesse depletion at 455 nm is more consistent with the nucleation model (purple curve) than with the oxygen-depletion model (orange curve). The inset shows the same data as a function of elapsed time, where the time intervals in which the laser was locked to the cavity (green) or not locked to the cavity (gray) are highlighted. The cavity remained in vacuum at all times.
Effect of continuous illumination at 671 nm on the finesse at 455 nm
After observing the degradation with high 455 nm circulating intensity, we tested the cavity degradation effect when locking the 671 nm laser to the cavity. We monitor the evolution of the finesse at both wavelengths. This measurement was done with two different conditions. In the first one, we set the input powers at mW and mW. The average power out of the cavity measured at 671 nm is W. In this case the experiment was done without previously recovering the finesse at 455 nm. During this measurement, the 671 nm laser was locked to the cavity for a total time of 680 mins (11.3 hours). The results are shown in Figure 4. Despite the high circulating power at 671 nm, the data show no finesse degradation, since the finesse values at both wavelength fluctuate around a certain mean value, measured to be and .

Figure 4. Cavity finesse stability under continuous illumination at 671 nm. Evolution of the cavity finesse at 455 nm (blue points) and 671 nm (red squares) when the cavity is locked at 671 nm, with an output power of from the cavity when locked. Finesse values at both wavelengths are normalized with respect to the mean finesse values and , respectively. The blue and red lines indicate the weighted average finesse at 455 nm and 671 nm respectively. No measurable degradation is observed at either wavelength despite the high 671 nm circulating power. Each data point represents the average of 10 measurements and the error bars indicate one standard deviation of the averaged ring-down measurements.
On a second set of measurements, we repeated the same experiment after a partial recovery of the cavity finesse at 455 nm. In this case, the input powers of both beams were on average mW and mW. The average output power at 671 nm was measured to be W. We observed once more that the finesse at both wavelengths fluctuates around an average value, found to be and , even when the cavity finesse has been recovered. This time the laser was locked to the cavity for a total time of 812 mins (13.5 hours). These two measurements show that no coating degradation is caused by 671 nm light running into the cavity and, moreover, they confirm once more that there is no evidence of a degradation purely caused by having the cavity under vacuum.
Finesse recovery
Following the work done by Ref. [22], the cavity finesse was partially recovered by locking the 455 nm laser to the cavity in air. During the procedure we monitored both the 455 nm finesse for checking the recovery and that at 671 nm with ring down measurement in order to monitor any potential effect of the recovery on the 671 nm finesse. The average input power of both lasers used during the oxygen recovery is mW and mW. The average output power at 455 nm was measured to be W. The results of the recovery are shown in Figure 5. The initial finesse value at 455 nm is . After the recovery which lasts a total illumination time of 1223 mins (20.4 hours), it reaches a steady state final value of . The finesse at 671 nm, instead, fluctuates around an average value of during the whole recovery. In Figure 5 the finesse values at 455 nm are normalized with respect to the maximum finesse value measured during recovery, . The finesse values at 671 nm, instead, are normalized with respect to the average value obtained from the whole set of data. The data of the 455 nm recovery are found to fit a model of the form

Figure 5. Cavity finesse recovery in air with 455 nm laser locked to the cavity. Blue points correspond to the cavity finesse at 455 nm, while red squares represent the finesse at 671 nm, both measured with cavity ring-down. Each data point represents the average of 10 measurements and the error bars indicate one standard deviation. We observe a recovery of the cavity finesse at 455 nm when the laser is locked to the cavity in air, with an average output power of . Finesse values at 455 nm are normalized with the maximum measured finesse, in this case . Instead, 671 nm finesse values are normalized with respect to the mean finesse (red line). The finesse at 671 nm remains unchanged throughout the measurement. The recovery (green line) are fitted with the phenomenological saturation growth given in Eq. (10).
where is the asymptotic finesse associated to the recovery process, is the initial finesse and is the characteristic recovery time. The recovery is described phenomenologically by a cubicexponent saturation function. An exponent of three is compatible with continuous nucleation and two-dimensional growth under classical Avrami assumptions, however, neither these spatial kinetics nor a linear relationship between recovered fraction and finesse is independently established in this work.
Conclusions
In this work, we investigated the degradation and recovery of a bichromatic Fabry-Perot cavity resonant at both 455 nm and 671 nm. Under vacuum, with the cavity locked at 455 nm and operating at moderate input optical power, the first degradation run showed a decrease in finesse of about 75%, from to , after approximately 11 hours of accumulated illumination. Comparison of oxygen-depletion and Avrami-type models showed that the observed degradation dynamics were better described by the Avrami model, consistent with nucleation-and-growth kinetics. Moreover, no measurable finesse degradation was observed during intervals without circulating optical power, suggesting that the vacuum environment alone is insufficient to account for the observed degradation. The final finesse in steady state suggest that that the cavity can be operated safely with an intensity of on the cavity mirrors.
Despite the substantial degradation observed at 455 nm, the finesse at 671 nm remained unchanged within experimental uncertainty. This suggests that the processes responsible for the increased optical losses at 455 nm have little or no measurable effect on the cavity finesse at 671 nm. Furthermore, when high 671 nm power was circulating in the cavity under vacuum, no measurable degradation was observed at either wavelength. This finding confirmed once more that the degradation is caused by blue radiation circulating in the cavity under vacuum, suggestive of degradation due to hydrocarbon contamination. Finally, a partial recovery of the finesse at 455 nm was observed with the cavity locked at 455 nm in air.
These results demonstrate that high circulating powers in the red (671 nm) can be safely used for applications such as optical trapping and cavity locking, without measurable degradation of the cavity finesse under the conditions investigated. In contrast, blue light (455 nm) applications should aim to limit the circulating photon number, for example by using short, low-power pulses or restricting blue light illumination to applications requiring enhanced coupling to the vacuum electromagnetic field. Minimizing potential sources of contamination inside the vacuum chamber [29] or performing annealing of the cavity mirrors [16] may also help mitigate long-term degradation under continuous blue laser light illumination. These findings are very relevant for the design of experiments with optical cavities that need to be used both for trapping and for cavity coupling, or for CQED with atomic mixtures.
Acknowledgments
We thank M. Keller for helpful discussions. This work has been supported by the project 23FUN03 HIOC, which has received funding from the European Partnership on Metrology, cofinanced from the European Union’s Horizon Europe Research and Innovation Programme and by the Participating States. This work was supported by the Cascading Grant Spoke 3 under the PNRR MUR project PE0000023-NQSTI, funded by the European Union – Next Generation EU. The raw data used in this manuscript are publicly available in Ref. [36].
References
- [1]H. J. Kimble, “Strong interactions of single atoms and photons in cavity qed,” Phys. Scripta 1998, 127 (1998).DOI
- [2]T. Kessler, C. Hagemann, C. Grebing, et al., “A sub-40-mhz-linewidth laser based on a silicon single-crystal optical cavity,” Nat. Photonics 6, 687 – 692 (2011).
- [3]E. Pedrozo-Peñafiel, S. Colombo, C. Shu, et al., “Entanglement on an optical atomic-clock transition,” Nature 588, 414–418 (2020).
- [4]J. M. Robinson, M. Miklos, Y. M. Tso, et al., “Direct comparison of two spin-squeezed optical clock ensembles at the 10−17 level,” Nat. Phys. 20, 208–214 (2024).
- [5]R. Landig, L. Hruby, N. Dogra, et al., “Quantum phases from competing short- and long-range interactions in an optical lattice,” Nature 532, 476–479 (2016).
- [6]N. Sauerwein, F. Orsi, P. Uhrich, et al., “Engineering random spin models with atoms in a high-finesse cavity,” Nat. Phys. 19, 1128–1134 (2023).DOI
- [7]M. Hijlkema, B. Weber, H. P. Specht, et al., “A single-photon server with just one atom,” Nat. Phys. 3, 253–255 (2007).
- [8]A. Stute, B. Casabone, P. Schindler, et al., “Tunable ion-photon entanglement in an optical cavity,” Nature 485, 482–485 (2012).DOI
- [9]S. Ritter, C. Nölleke, C. Hahn, et al., “An elementary quantum network of single atoms in optical cavities,” Nature 484, 195–200 (2012).
- [10]V. Krutyanskyi, M. Galli, V. Krcmarsky, et al., “Entanglement of trapped-ion qubits separated by 230 meters,” Phys. Rev. Lett. 130, 050803 (2023).
- [11]M. Büki, P. Malik, F. Fertig, et al., “Metropolitan entanglement distribution between an atom and a near-visible photon,” Phys. Rev. Lett. 137, 090803 (2026).DOI
- [12]M. Barbiero, D. Calonico, F. Levi, and M. G. Tarallo, “Optically loaded strontium lattice clock with a single multi-wavelength reference cavity,” IEEE Trans. on Instrum. Meas. 71, 1–9 (2022).DOI
- [13]J. Ye, D. W. Vernooy, and H. J. Kimble, “Trapping of single atoms in cavity qed,” Phys. Rev. Lett. 83, 4987–4990 (1999).DOI
- [14]J. Lee, G. Vrijsen, I. Teper, et al., “Many-atom–cavity qed system with homogeneous atom&#x–cavity coupling,” Opt. Lett. 39, 4005–4008 (2014).
- [15]H. Demiryont, J. R. Sites, and K. Geib, “Effects of oxygen content on the optical properties of tantalum oxide films deposited by ion-beam sputtering,” Appl. Opt. 24, 490–495 (1985).pubmed.ncbi.nlm.nih.gov/18216975
- [16]B. Brandstätter, A. McClung, K. Schüppert, et al., “Integrated fiber-mirror ion trap for strong ion-cavity coupling,” Rev. Sci. Instruments 84, 123104 (2013).DOI
- [17]A. Rudelis, B. Hu, J. Sinclair, et al., “Degradation of ta2o5 / sio2 dielectric cavity mirrors in ultra-high vacuum,” Opt. Express 31, 39670–39680 (2023).DOI
- [18]P. Wagner, ““laser-induced contamination on high-reflective optics,” ms thesis, university of applied science darmstadt, 2014,”.
- [19]R. R. Kunz, V. Liberman, and D. K. Downs, “Experimentation and modeling of organic photcontamination on lithographic optics,” J. Vac. Sci. Technol. B: Microelectron. Nanometer Struct. Process. Meas. Phenom. 18, 1306–1313 (2000).
- [20]N. Koster, B. Mertens, R. Jansen, et al., “Molecular contamination mitigation in euvl by environmental control,” Microelectron. Eng. 61-62, 65–76 (2002). Micro- and Nano-Engineering 2001.DOI
- [21]A. Pereira, E. Quesnel, and M. Reymermier, “Dynamic measurements of ultraviolet-enhanced silica contamination by photoluminescence-based diagnostic,” J. Appl. Phys. 105, 013109 (2009).DOI
- [22]D. Gangloff, M. Shi, T. Wu, et al., “Preventing and reversing vacuum-induced optical losses in high-finesse tantalum (v) oxide mirror coatings,” Opt. Express 23, 18014–18028 (2015).DOI
- [23]T. Ballance, H. Meyer, P. Kobel, et al., “Cavity-induced backaction in purcell-enhanced photon emission of a single ion in an ultraviolet fiber cavity,” Phys. Rev. A 95, 033812 (2017).DOI
- [24]Z. Burkley, L. de Sousa Borges, B. Ohayon, et al., “Stable high power deep-uv enhancement cavity in ultra-high vacuum with fluoride coatings,” Opt. Express 29, 27450–27459 (2021).DOI
- [25]J. Hollenshead and L. Klebanoff, “Modeling radiation-induced carbon contamination of extreme ultraviolet optics,” J. Vac. Sci. Technol. B: Microelectron. Nanometer Struct. Process. Meas. Phen. 24, 64–82 (2006).DOI
- [26]C. Brand, C. Knobloch, K. Simonović, and M. Arndt, “Multi-watt cavity for 266 nm light in vacuum,” Phys. Scripta 98, 085521 (2023).DOI
- [27]D. Das, E. Kassa, and H. Takahashi, “Stability of a high-finesse optical cavity at 493nm in vacuum for cavity qed with barium ions,” Appl. Opt. 65, 5082–5090 (2026).
- [28]M. Keller, “Private communications.”
- [29]J. Schmitz, H. M. Meyer, and M. Köhl, “Ultraviolet fabry-perot cavity with stable finesse under ultrahigh vacuum conditions,” Rev. Sci. Instruments 90, 063102 (2019).DOI
- [30]E. Perego, L. Duca, and C. Sias, “Electro-optical ion trap for experiments with atom-ion quantum hybrid systems,” Appl. Sci. 10, 2222 (2020).arxiv.org/abs/2001.11968
- [31]F. Berto, Ph.D. thesis: A new setup and methods for the fast production of ultracold atoms for atom-ion experiments (2022).
- [32]D. Romanini, A. A. Kachanov, N. Sadeghi, and F. Stoeckel, “CW cavity ring down spectroscopy,” Chem. Phys. Lett. 264, 316–322 (1997).DOI
- [33]G. Rempe, R. J. Thompson, H. J. Kimble, and R. Lazeari, “Measurement of ultralow losses in an optical interferometer,” Opt. Lett. 17, 363–365 (1992).
- [34]K. Shirzad and C. Viney, “A critical review on applications of the Avrami equation beyond materials science,” J. R. Soc. Interface 20, 20230242 (2023).DOI
- [35]A. Yariv, Quantum Electronics (John Wiley & Sons, New York, 1989), 3rd ed.
- [36]https://doi.org/10.5281/zenodo.23798276.